Permissive change request letter

FCC ID: PQC-OBRTBV1

Cover Letter(s)

Download: PDF
FCCID_4316141

                                                                                   PHILIPS
Philips Medizin Systeme Boblingen GmbH




                                                                                    February 12, 2019




Federal Communication Commission
Equipment Authorization Devision, Application Processing Branch
7435 Oakland Mills Road
Columbia, MD 21048



Certification and Engineering Bureau
Industry Canada
Spectrum Engineering Branch
3701 Carling Avenue, Building 94
Ottawa, Ontario K2H 882




PCII Request / Re—assessment

CTG Base Station {(Model Id. 866074),
FCC ID: PQC—OBRBSBV1 granted on November 18, 2013 and
IC: 3549C—OBRBSBV1 granted on November 18, 2013

Wireless Transducer (Model Id. 866075, 866076, 866077),
FCC 1ID: PQC—OBRTBV1 granted on July 11, 2013 and
IC: 3549C—OBRTBV1 granted on July 11, 2013




TO WHOM IT MAY CONCERN

Pursuant to CFR § 2.1043 and RSP—100, Philips Medizin Systeme Boblingen GmbH hereby requests
a Permissive Change Class II / Re—assessment.

Reason:


Instead of using a predefined channel, the system can search for a free channel on its own. There
is no need any more for the customer to set up a frequency plan. Furthermore the system detects
if there is a collision in the used channel, caused by another cordless device, and then changes the
channel automatically.

Philips Medizin Systeme Boblingen GmbH
Hewlett—Packard—Str. 2, 71034 Bdblingen, Germany
Telefon     +49 (0) 7031 463—0
Telefax     +49 (0) 7031 463—2202
Geschdftsfuhrer: Dr. Dieter Haase (Sprecher), Dr. Peter Ziese, Matthijs de Groot
Registergericht Stuttgart Reg.—Nr. HRB 245187, Sitz der Gesellschaft: Boblingen
www .philips.com/healthcare


                                                                                    PHILIPS
 Philips Medizin Systeme Boblingen GmbH


 Modifications:

 — Occupied bandwidth for both, base station and transducers: Old=> approx. 120kHz, New=>
 approx. 6OkHz                                                                 ;
 — Timing base station (worst case transmit time during 100ms): Old=> 38.4ms, New=> 9.6ms
 — Timing ‘transducers (worst case transmit time during 100ms): Old=> 43.2ms, New=> 28.2ms

 The following new exhibits will be uploaded:

 —    Test report on conducted measurements
 —    External Photos
 —    Internal Photos




 Sincerely,


                                       Pnilips Modizin Systeme Boklingen m
                                       Howlett—Packard—Str. 2 ,
                                       71034 Boblingen,                   knfas

 Christian Noguera
 Quality & Regulat@ry Affairs

  Philips Medizin Systeme Boblingen GmbH
— Hewlett—Packard—Str. 2            71034 Boblingen, Germany
  Tel.: +49 7031 463—2316            Fax.: +49 7031 463—2442
  E—Mail: christian.noguera@philips.com




 Philips Medizin Systeme Boblingen GmbH
 Hewleft—Packard—Str. 2, 71034 Bdblingen, Germany
 Telefon     +49 (0) 7031 463—0
 Telefox     +49 (0) 7031 463—2202
 Gesch@ftsfUhrer: Or. Dieter Haase (Sprecher), Dr. Peter Ziese, Matthijs de Groot
 Registergericht Stuttgart Reg.—Nr. HRB 245187, Sitz der Gesellschatft: Boblingen
 www.philips.com/healthcare



Document Created: 2019-06-13 07:05:20