RF Exposure Info._20160808_v1 - PPD-QCNFA364AH_SAR_19.2_SAR Test Data Plots

FCC ID: PPD-QCNFA364AH

RF Exposure Info

Download: PDF
FCCID_3150694

Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                 Date: 2016/7/20

Wi-Fi 2.4GHz Band

Frequency: 2437 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.5C
                                                                                               3
Medium parameters used (interpolated): f = 2437 MHz; σ = 1.882 S/m; εr = 53.627; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(7.32, 7.32, 7.32); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 1.4mm
(Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Main Ant/802.11b/ch6/Area Scan (6x7x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.65 W/kg

Bottom/Main Ant/802.11b/ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm,
dz=5mm
Reference Value = 5.436 V/m; Power Drift = 0.10 dB
Peak SAR (extrapolated) = 2.19 W/kg
SAR(1 g) = 1.01 W/kg; SAR(10 g) = 0.506 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.72 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 2.4GHz Band
Frequency: 2437 MHz; Duty Cycle: 1:1

Bottom/Main Ant/802.11b/ch6/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 2.15 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 2.4GHz Band

Frequency: 2437 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.5C
                                                                                               3
Medium parameters used (interpolated): f = 2437 MHz; σ = 1.882 S/m; εr = 53.627; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(7.32, 7.32, 7.32); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Aux Ant/802.11b/ch6/Area Scan (6x11x1): Measurement grid: dx=12mm, dy=12mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.46 W/kg

Bottom/Aux Ant/802.11b/ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm,
dz=5mm
Reference Value = 7.746 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 1.87 W/kg
SAR(1 g) = 0.918 W/kg; SAR(10 g) = 0.517 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.48 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 2.4GHz Band
Frequency: 2462 MHz; Duty Cycle: 1:1

NB/Bottom/Main Ant/802.11b/ch6/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Maximum value of SAR (measured) = 2.23 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                 Date: 2016/7/20

Wi-Fi 2.4GHz Band

Frequency: 2437 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.5C
                                                                                               3
Medium parameters used (interpolated): f = 2437 MHz; σ = 1.882 S/m; εr = 53.627; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(7.32, 7.32, 7.32); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 1.4mm
(Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

(Spot check)Bottom/Main Ant/802.11b/ch6/Area Scan (6x7x1): Measurement grid: dx=12mm,
dy=12mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.781 W/kg

(Spot check)Bottom/Main Ant/802.11b/ch6/Zoom Scan (7x7x7)/Cube 0: Measurement grid:
dx=5mm, dy=5mm, dz=5mm
Reference Value = 10.19 V/m; Power Drift = 0.06 dB
Peak SAR (extrapolated) = 1.40 W/kg
SAR(1 g) = 0.707 W/kg; SAR(10 g) = 0.460 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.06 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 2.4GHz Band
Frequency: 2437 MHz; Duty Cycle: 1:1

(Spot check)Bottom/Main Ant/802.11b/ch6/Z Scan (1x1x21): Measurement grid: dx=20mm,
dy=20mm, dz=5mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.853 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 2.4GHz Band

Frequency: 2437 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.5C
                                                                                               3
Medium parameters used (interpolated): f = 2437 MHz; σ = 1.882 S/m; εr = 53.627; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(7.32, 7.32, 7.32); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

(Spot check)Bottom/Aux Ant/802.11b/ch1/Area Scan (6x7x1): Measurement grid: dx=12mm,
dy=12mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.964 W/kg

(Spot check)Bottom/Aux Ant/802.11b/ch1/Zoom Scan (7x7x7)/Cube 0: Measurement grid:
dx=5mm, dy=5mm, dz=5mm
Reference Value = 11.40 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 1.12 W/kg
SAR(1 g) = 0.681 W/kg; SAR(10 g) = 0.467 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.912 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 2.4GHz Band
Frequency: 2437 MHz; Duty Cycle: 1:1

(Spot check)Bottom/Aux Ant/802.11b/ch1/Z Scan (1x1x21): Measurement grid: dx=20mm,
dy=20mm, dz=5mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.375 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5270 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                  3
Medium parameters used: f = 5270.5 MHz; σ = 5.191 S/m; εr = 48.611; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(4.27, 4.27, 4.27); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Main Ant/802.11n HT40/ch54/Area Scan (7x8x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 1.17 W/kg

Bottom/Main Ant/802.11n HT40/ch54/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm,
dy=4mm, dz=2mm
Reference Value = 14.14 V/m; Power Drift = -0.06 dB
Peak SAR (extrapolated) = 2.20 W/kg
SAR(1 g) = 0.748 W/kg; SAR(10 g) = 0.606 W/kg
Maximum value of SAR (measured) = 2.20 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5270 MHz; Duty Cycle: 1:1

Bottom/Main Ant/802.11n HT40/ch54/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm,
dz=5mm
Maximum value of SAR (measured) = 2.80 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5270 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.5C
                                                                                  3
Medium parameters used: f = 5270.5 MHz; σ = 5.191 S/m; εr = 48.611; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(4.27, 4.27, 4.27); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Aux Ant/802.11n HT40/ch54/Area Scan (6x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 1.63 W/kg

Bottom/Aux Ant/802.11n HT40/ch54/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm,
dy=4mm, dz=2mm
Reference Value = 8.878 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 2.60 W/kg
SAR(1 g) = 1.03 W/kg; SAR(10 g) = 0.639 W/kg
Maximum value of SAR (measured) = 1.77 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5270 MHz; Duty Cycle: 1:1

Bottom/Aux Ant/802.11n HT40/ch54/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm,
dz=5mm
Maximum value of SAR (measured) = 1.64 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5690 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.0C; Liquid Temperature: 23.5C
                                                                                               3
Medium parameters used (interpolated): f = 5690 MHz; σ = 5.704 S/m; εr = 47.947; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.63, 3.63, 3.63); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Main Ant/802.11ac/ch138/Area Scan (7x8x1): Measurement grid: dx=10mm, dy=10mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 2.43 W/kg

Bottom/Main Ant/802.11ac/ch138/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm,
dy=4mm, dz=2mm
Reference Value = 9.626 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 2.97 W/kg
SAR(1 g) = 1.2 W/kg; SAR(10 g) = 0.797 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 2.10 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5690 MHz; Duty Cycle: 1:1

Bottom/Main Ant/802.11ac/ch138/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.523 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                 Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5690 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                               3
Medium parameters used (interpolated): f = 5690 MHz; σ = 5.704 S/m; εr = 47.947; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.63, 3.63, 3.63); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Aux Ant/802.11ac/ch138/Area Scan (6x7x1): Measurement grid: dx=10mm, dy=10mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.59 W/kg

Bottom/Aux Ant/802.11ac/ch138/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm,
dy=4mm, dz=2mm
Reference Value = 7.581 V/m; Power Drift = -0.07 dB
Peak SAR (extrapolated) = 5.42 W/kg
SAR(1 g) = 1.13 W/kg; SAR(10 g) = 0.714 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 2.34 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5690 MHz; Duty Cycle: 1:1

Bottom/Aux Ant/802.11ac/ch138/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm, dz=5mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.262 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5795 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                  3
Medium parameters used: f = 5795.2 MHz; σ = 5.833 S/m; εr = 47.779; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.63, 3.63, 3.63); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Main Ant/802.11n HT40/ch159/Area Scan (7x8x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 2.47 W/kg

Bottom/Main Ant/802.11n HT40/ch159/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm,
dy=4mm, dz=2mm
Reference Value = 6.050 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 4.77 W/kg
SAR(1 g) = 1.12 W/kg; SAR(10 g) = 0.441 W/kg
Maximum value of SAR (measured) = 2.58 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5795 MHz; Duty Cycle: 1:1

Bottom/Main Ant/802.11n HT40/ch159/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm,
dz=5mm
Maximum value of SAR (measured) = 0.0262 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                 Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5795 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                  3
Medium parameters used: f = 5795.2 MHz; σ = 5.833 S/m; εr = 47.779; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.95, 3.95, 3.95); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Aux Ant/802.11n HT40/ch159/Area Scan (6x7x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) = 2.02 W/kg

Bottom/Aux Ant/802.11n HT40/ch159/Zoom Scan (7x7x12)/Cube 0: Measurement grid: dx=4mm,
dy=4mm, dz=2mm
Reference Value = 6.876 V/m; Power Drift = -0.06 dB
Peak SAR (extrapolated) = 2.86 W/kg
SAR(1 g) = 1.01 W/kg; SAR(10 g) = 0.600 W/kg
Maximum value of SAR (measured) = 1.77 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5795 MHz; Duty Cycle: 1:1

Bottom/Aux Ant/802.11n HT40/ch159/Z Scan (1x1x21): Measurement grid: dx=20mm, dy=20mm,
dz=5mm
Maximum value of SAR (measured) = 1.19 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5795 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                  3
Medium parameters used: f = 5795.2 MHz; σ = 5.833 S/m; εr = 47.779; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.95, 3.95, 3.95); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Main Ant/802.11n HT40/ch159 Ant2/Area Scan (7x8x1): Measurement grid: dx=10mm,
dy=10mm
Maximum value of SAR (measured) = 2.17 W/kg

Bottom/Main Ant/802.11n HT40/ch159 Ant2/Zoom Scan (7x7x12)/Cube 0: Measurement grid:
dx=4mm, dy=4mm, dz=2mm
Reference Value = 7.336 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 4.10 W/kg
SAR(1 g) = 1.16 W/kg; SAR(10 g) = 0.701 W/kg
Maximum value of SAR (measured) = 2.26 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5795 MHz; Duty Cycle: 1:1

Bottom/Main Ant/802.11n HT40/ch159 Ant2/Z Scan (1x1x21): Measurement grid: dx=20mm,
dy=20mm, dz=5mm
Maximum value of SAR (measured) = 1.58 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5755 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                  3
Medium parameters used: f = 5755.6 MHz; σ = 5.778 S/m; εr = 47.816; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.95, 3.95, 3.95); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Main Ant/802.11n HT40/ch151 Ant2/Area Scan (7x8x1): Measurement grid: dx=10mm,
dy=10mm
Maximum value of SAR (measured) = 1.51 W/kg

Bottom/Main Ant/802.11n HT40/ch151 Ant2/Zoom Scan (7x7x12)/Cube 0: Measurement grid:
dx=4mm, dy=4mm, dz=2mm
Reference Value = 13.08 V/m; Power Drift = 0.09 dB
Peak SAR (extrapolated) = 1.60 W/kg
SAR(1 g) = 0.645 W/kg; SAR(10 g) = 0.423 W/kg
Maximum value of SAR (measured) = 1.33 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5755 MHz; Duty Cycle: 1:1

Bottom/Main Ant/802.11n HT40/ch151 Ant2/Z Scan (1x1x21): Measurement grid: dx=20mm,
dy=20mm, dz=5mm
Maximum value of SAR (measured) = 0.476 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                 Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5690 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                                               3
Medium parameters used (interpolated): f = 5690 MHz; σ = 5.704 S/m; εr = 47.947; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.63, 3.63, 3.63); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Aux Ant/802.11ac/ch138 spot check/Area Scan (6x7x1): Measurement grid: dx=10mm,
dy=10mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.55 W/kg

Bottom/Aux Ant/802.11ac/ch138 spot check/Zoom Scan (7x7x12)/Cube 0: Measurement grid:
dx=4mm, dy=4mm, dz=2mm
Reference Value = 7.035 V/m; Power Drift = 0.05 dB
Peak SAR (extrapolated) = 5.08 W/kg
SAR(1 g) = 1.03 W/kg; SAR(10 g) = 0.674 W/kg
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 1.91 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5690 MHz; Duty Cycle: 1:1

Bottom/Aux Ant/802.11ac/ch138 spot check/Z Scan (1x1x21): Measurement grid: dx=20mm,
dy=20mm, dz=5mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.480 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01                  Date: 2016/7/20

Wi-Fi 5GHz Band

Frequency: 5610 MHz; Duty Cycle: 1:1;     Room Ambient Temperature: 24.5C; Liquid Temperature: 24.0C
                                                                               3
Medium parameters used: f = 5610.4 MHz; σ = 5.6 S/m; εr = 48.08; ρ = 1000 kg/m
DASY5 Configuration:
- Area Scan Setting - Find Secondary Maximum Within: 2.0 dB and with a peak SAR value greater than 0.0012W/kg
- Electronics: DAE4 Sn877; Calibrated: 2016/3/21
- Probe: EX3DV4 - SN3665; ConvF(3.63, 3.63, 3.63); Calibrated: 2016/5/26;
- Sensor-Surface: 1.4mm (Mechanical Surface Detection)
- Phantom: ELI 4.0; Type: QDOVA001BA; Serial: 1056

Bottom/Aux Ant/802.11ac/ch122 spot check/Area Scan (6x7x1): Measurement grid: dx=10mm,
dy=10mm
Maximum value of SAR (measured) = 1.54 W/kg

Bottom/Aux Ant/802.11ac/ch122 spot check/Zoom Scan (7x7x12)/Cube 0: Measurement grid:
dx=4mm, dy=4mm, dz=2mm
Reference Value = 5.350 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 2.59 W/kg
SAR(1 g) = 0.914 W/kg; SAR(10 g) = 0.557 W/kg
Maximum value of SAR (measured) = 1.71 W/kg


Test Laboratory: Compliance Certification Service Inc. SAR Lab 01   Date: 2016/7/20

Wi-Fi 5GHz Band
Frequency: 5610 MHz; Duty Cycle: 1:1

Bottom/Aux Ant/802.11ac/ch122 spot check/Z Scan (1x1x21): Measurement grid: dx=20mm,
dy=20mm, dz=5mm
Maximum value of SAR (measured) = 0.324 W/kg



Document Created: 2016-08-01 14:22:30
Document Modified: 2016-08-01 14:22:30