SAR Dipole

FCC ID: PPD-AR5B195

RF Exposure Info

Download: PDF
FCCID_1656549

Calibration Laboratory of                                         \\“\\wl’"/,,                                     Schweizerischer Kalibrierdienst
Schmid & Partner                                                 ib\fi&                                             Service suisse d‘étalonnage
   Engineering AG                                                               >                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                              Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multitateral Agreement for the recognition of calibration certificates

Client       CCS—TW (Auden)                                                                     Certificate No: D2450V2—728_Nov11

CALIBRATION CERTIFICATE
 Object                                D2450V2 — SN: 728



 Calibration procedure(s)              QA CAL—05.v8
                                       Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                     November 22, 2011




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the cerfificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration}


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Power meter EPM—442A                   GB37480704                05—Oct—11 (No. 217—01451)                            Oct—12
 Power sensor HP 8481A                  US37202783                05—Oct—11 (No. 217—01451)                            Oct—12
 Reference 20 dB Attenuator             SN: 5086 (20g)            29—Mar—11 (No. 217—01368)                            Apr—12
 Type—N mismatch combination            SN: 5047.2 / 06827        29—Mar—11 (No. 217—01371)                            Apr—12
 Reference Probe ESSDV3                 SN: 3205                  29—Apr—11 (No. ES3—3205_Apr11)                       Apr—12
 DAE4                                   SN: 601                   04—Jul—11 (No. DAE4—601_Jul11)                       Jul—12

 Secondary Standards                    ID #                      Check Date (in house)                                Scheduted Check
 Power sensor HP 8481A                  MY41092317                18—Oct—02 (in house check Oct—11)                    In house check: Oct—13
 RF generator R&S SMT—06                100005                    04—Aug—99 (in house check Oct—11)                    in house check: Oct—13
 Network Analyzer HP 8753E              US37300585 $4206          18—Oct—01 (in house check Oct—11)                    in house check: Oct—12



                                        Name                                   Function                                 Signature
 Calibrated by:                         Jeton Kastrati                         Laboratory Technician                                ( Z



 Approved by:                           KatJa Pokovic                          Technical Manager
                                                                                                                   ets
                                                                                                                       Issued: November 22, 2011
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D2450V2—728_Nov11                                     Page 1 of 8


Calibration Laboratory of                                                        Schweizerischer Kalibrierdienst
Schmid & Partner                                                                 Service suisse d‘étalonnage
  Engineering AG                                                                 Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                     Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                           Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
         Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
         Communications Devices: Measurement Techniques", December 2003
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
       "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiotfrequency
         Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
         Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
         Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    e    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    *    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    *    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
    e    SAR measured: SAR measured at the stated antenna input power.
    e   SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector.
    +    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
         nominal SAR result.



Certificate No: D2450V2—728_Nov11                               Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not iven on page 1.
      DASY Version                                              DASY5                              V52.6.2
      Extrapolation                                   Advanced Extrapolation
      Phantom                                          Modular Flat Phantom

      Distance Dipole Center — TSL                               10 mm                           with Spacer
      Zoom Scan Resolution                               dx, dy, dz =5 mm

      Frequency                                         2450 MHz £ 1 MHz



Head TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Head TSL parameters                                  22.0 °C           39.2              1.80 mho/m
      Measured Head TSL parameters                              (22.0 x 0.2) °C   37.7 x6 %          1.84 mho/m £ 6 %
      Head TSL temperature change during test                      <0.5 °C


SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                   13.7 mW /g
      SAR for nominal Head TSL parameters                     normalized to 1W        53.7 mW /g + 17.0 % (k=2)

      SAR averaged over 10 cm° (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                6.35 mW /g
      SAR for nominal Head TSL parameters                     normalized to 1W        25.1 mW /qg £ 16.5 % (k=2)


Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity        Conductivity
      Nominal Body TSL parameters                                  22.0 °C           52.7              1.95 mho/m
      Measured Body TSL parameters                              (22.0 £ 0.2) °C   50.9 + 6 %        2.01 mho/m + 6 %
      Body TSL temperature change during test                      <0.5 °C


SAR result with Body TSL

      SAR averaged over 1 cm° (1 g) of Body TSL                  Condition

      SAR measured                                        250 mW input power                   13.1 mW /g
      SAR for nominal Body TSL parameters                     normalized to 1W       51.2 mW / g + 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition
      SAR measured                                        250 mW input power                   6.09 mW / g
      SAR for nominal Body TSL parameters                     normalized to 1W       24.0 mW / g # 16.5 % (k=2)



Certificate No: D2450V2—728_Nov11                      Page 3 of 8


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           55.8 Q + 3.3 jQ

      Return Loss                                                                        — 24.0 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           51.1 2 + 4.3 jQ

      Return Loss                                                                       —27.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.172 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                               January 09, 2003




Certificate No: D2450V2—728_Nov11                        Page 4 of 8


DASY5 Validation Report for Head TSL
                                                                                   Date: 22.11.201 1

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 728

Communication System: CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.84 mho/m; & = 37.7; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

        Probe: ES3DV3 — SN3205; ConvF(4.45, 4.45, 4.45); Calibrated: 29.04.201 1

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 04.07.201 1

        Phantom: Flat Phantom 5.0 (front); Type: QDOOOP50OAA; Serial: 1001

        DASY52 52.6.2(482); SEMCAD X 14.4.5(3634)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 101.4 V/m; Power Drift = 0.08 dB
Peak SAR (extrapolated) = 28.548 W/kg
SAR(I g) = 13.7 mW/g; SAR(10 g) = 6.35 mW/g
Maximum value of SAR (measured) = 17.527 mW/g




                            —12.00




                            —18.00




                            <24.00




                            +30.00

                                     0 dB = 17.530mW/g




Certificate No: D2450V2—728_Nov11                  Page 5 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 22.11.2011

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 728

Communication System: CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 2.01 mho/m; &, = 50.9; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C€63.19—2007)

DASY52 Configuration:

    +   Probe: ES3DV3 — SN3205; ConvF(4.26, 4.26, 4.26); Calibrated: 29.04.2011

        Sensor—Surface: 3mm (Mechanical Surface Detection)

        Electronics: DAE4 $n601; Calibrated: 04.07.201 1

        Phantom: Flat Phantom 5.0 (back); Type: QDOOOP5OAA; Serial: 1002

        DASY52 52.6.2(482); SEMCAD X 14.4.5(3634)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 96.441 V/m; Power Drift = —0.008 dB
Peak SAR (extrapolated) = 27.089 W/kg
SAR(I g) = 13.1 mW/g; SAR(10 g) = 6.09 mW/g
Maximum value of SAR (measured) = 17.326 mW/g




                            5.00



                            —10.00




                            +15.00




                            —20.00




                            —25.00

                                     0 dB = 17.330mW/g




Certificate No: D2450V2—728_Nov11                  Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                            22 Nov 2011      ©9:48158
          CHI si1       1 U Fs               2rSid11 0       4.39145 a   280.27 pH         2 450.000 000 MHz




          Del


          Ca                                 j



          fvg
          16


          HId


          CH2                                REF —20 dB.




          Ca




          y
          169      o   cencfocss es



          H1d    n nfi ons        onl   s c2000ons on ho onn ceprrnen 0 sopsoceen     es en neabeccseenefoseso_


                   ~ETART 2 250.000 000MHz




Certificate No: D2450V2—728_Nov11                             Page 8 of 8



Document Created: 2011-12-09 10:23:37
Document Modified: 2011-12-09 10:23:37