CALIBRATION DATA_Part 2

FCC ID: POD-DMR2

RF Exposure Info

Download: PDF
FCCID_3422048

DIPOLE CALIBRATION DATA


Calibration Laboratory of                                    «7                 Schweizerischer Kalibrierdienst
Schmid & Partner                                                                Service suisse d‘étaionnage
  Engineering AG                                                                Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                    Swiss Calibration Service
                                                               Fahils®

Accredited by the Swiss Accreditation Service (SAS)                          Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
       devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
   e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
   * Measurement Conditions: Further details are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at the frequency indicated.
   * Antenna Parameters with TSL: The source is mounted in a touch configuration below the
      center marking of the flat phantom.
   * Return Loss: This parameter is measured with the source positioned underthe liquid filled
      phantom (as described in the measurement condition clause). The Return Loss ensures low
      reflected power. No uncertainty required.
      SAR measured: SAR measured at the stated antenna input power.
      SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
   «_ SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.

The reported uncertainty of measurementis stated as the standard uncertainty of measurement
multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
probability of approximately 95%.




Certificate No: CLA150—4008_Jan17                              Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not iven on page 1.
     DASY Version                                            DASYS                                V52.8.8
      Extrapolation                                  Advanced Extrapolation

      Phantom                                           EL14 Flat Phantom             Shell thickness: 2 + 0.2 mm
      EUT Positioning                                    Touch Position
      Zoom Scan Resolution                           dx, dy = mm, dz = mm           Graded Ratio = 1.4 (Z direction)

      Frequency                                         150 MHz + 1 MHz


Head TSL parameters
    The followingparameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity
      Nominal Head TSL parameters                                 22.0 °C          52.3               0.76 mho/m
      Measured Head TSL parameters                             (22.0 £0.2) °C   50.1 £ 6 %         0.75 mho/m + 6 %

      Head TSL temperature change during test                     <0.5 °C                                   ———

SAR result with Head TSL

      SAR averaged over 1 cm‘ (1 g) of Head TSL                 Condition

      SAR measured                                           1 W input power                   3.89 W/kg

      SAR for nominal Head TSL parameters                    normalized to 1W       3.90 W/kg + 18.4 % (k=2)


      SAR averaged over 10 cm‘ (10 g) of Head TSL               condition

      SAR measured                                           1 W input power                   2.59 W/kg
      SAR for nominal Head TSL parameters                    normalized to 1W       2.60 Wikg + 18.0 % (k=2)


Body TSL parameters
    Thefollowing parameters and calculations were applied.
                                                               Temperature      Permittivity         Conductivity

      Nominal Body TSL parameters                                 22.0 °C          61.9               0.80 mho/m
      Measured Body TSL parameters                             (22.0 £0.2) °C   61.4 +6 %          0.82 mho/m x 6 %

      Body TSL temperature change during test                     <0.5 °C           ~—                      —


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                 Condition

      SAR measured                                           1 W input power                   4.03 Wig
      SAR for nominal Body TSL parameters                    normalized to 1W       3.95 Wikg 2 18.4 % (k=2)

      SAR averaged over 10 m‘ (10 g) of Body TSL                condition

      SAR measured                                           1 W input power                   2.67 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W       2.62 Wikg 2 18.0 % (k=2)




Certificate No: CLA150—4008_Jan17                     Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                   44.1 Q — 6.0 t

      Return Loss                                               —20.9 dB


Antenna Parameters with Body TSL

      Impedance, transformedto feed point                    47.6 0 — 9.0 jQ

      Return Loss                                              —20.5 dB


Additional EUT Data

      Manufactured by                                           SPEAG

      Manufactured on                                      December 16, 2013




Certificate No: CLA150—4008_Jan17            Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                       Date: 23.01.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: CLA—150; Type: CLA—150; Serial: 4008

Communication System: UID 0 — CW; Frequency: 150 MHz
Medium parameters used: £ = 150 MHz; 0 = 0.75 $/m; &, = 50.1; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    *    Probe: EX3DV4 — SN3877; ConyF(12.04, 12.04, 12.04); Calibrated: 31.12.2016;

    «+   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    *«   Electronics: DAE4 Sn654; Calibrated: 12.08.2016

    «_   Phantom: ELI v4.0; Type: QDOVAOO1BB; Serial: TP:1003

    +    DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


CLA Calibration for HSL—LF Tissue/CLA150, touch configuration, Pin=1 W/Area Scan
(81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 5.38 W/kg

CLA Calibration for HSL—LF Tissue/CLA150, touch configuration, Pin=1 W/Zoom Scan,
dist=1.4mm (8x10x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 84.22 V/m; Power Drift =0.01 dB
Peak SAR (extrapolated) = 7.04 W/kg
SAR(I g) = 3.89 W/kg; SAR(10 g) = 2.59 W/kg
Maximum value of SAR (measured) = 5.42 W/kg




                   —9.00

                   —12.00

                   —15.00
                            0 dB = 5.38 W/kg =7.31 dBW/kg



Certificate No: CLA150—4008_Jan17               Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                               23 Jan 2017   14:38:13
          CHD si      1 U FS         2440610   —6,0098 0   176.55 pF         150.000 008 MHz




          B
          185


          H1d

          CH2




          Ca



          By
          159


          Hid
                    STaRT 100.000 000 Mz                               ToP 200.000 008 Mtiz




Certificate No: CLA150—4008_Jan17                Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                       Date: 23.01.2017

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: CLA—150; Type: CLA—150; Serial: 4008

Communication System: UID 0 — CW; Frequency: 150 MHz
Medium parameters used: f= 150 MHz; 0 = 0.82 S/m; & = 61.4; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASYS (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

    «_   Probe: EX3DV4 — SN3877; ConvF(11.54, 11.54, 11.54); Calibrated: 31.12.2016;

    *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)

    +    Electronics: DAE4 Sn654; Calibrated: 12.08.2016

    «+   Phantom: ELI v4.0; Type: QDOVAOO1BB; Serial: TP:1003

    +    DASY52 52.8.8(1258); SEMCAD X 14.6.10(7372)


CLA Calibration for MSL—LF Tissue/CLA150, touch configuration, Pin=1 W/Area Scan
(81x81x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 5.72 W/kg

CLA Calibration for MSL—LF Tissue/CLA150, touch configuration, Pin=1 W/Zoom Scan,
dist=1.4mm (8x10x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 82.65 V/m; Power Drift =0.01 dB
Peak SAR (extrapolated) =7.67 W/kg
SAR(I g) = 4.03 W/kg; SAR(10 g) = 2.67 W/kg
Maximum value of SAR (measured) = 5.66 W/kg




                    8.78


                    q1.7



                    1464

                                0 dB = 5.72 W/kg = 7.57 dBW/kg




Certificate No: CLA150—4008_Jan17                Page 7 of 8




      SAR Reference Dipole Calibration Report


                                        Ref : ACR.69.2.17.SATU.A




ATTESTATION OF GLOBAL COMPLIANCE CO.
                LTD.
  1&2F, NO.2 BUILDING, HUAFENG NO.1 INDUSTRIAL
    PARK, GUSHU COMMUNITY XIXIANG STREET
      BAOAN DISTRICT, SHENZHEN, P.R. CHINA
      MVG COMOSAR REFERENCE DIPOLE
                      FREQUENCY: 450 MHZ
                      SERIAL NO.: SN 46/11 DIP 0G450—184


                             Calibrated at MVG US
                  2105 Barrett Park Dr. — Kennesaw, GA 30144

                               osHop>

                           4. im

                              "Crlilgs®         colitretioncest szzeso:
                         Calibration Date: 03/10/2017


    Summary:

    This document presents the method and results from an accredited SAR reference dipole calibration
    performed in MVG USA using the COMOSAR test bench. All calibration results are traceable to
    national metrology institutions.
















                    SAR REFERENCE DIPOLE CALIBRATION REPORT                                                      Ref: ACR.69.2.17.SATU.A




                        1900                 39.7                                      20.5

                        1950                 40.5                                      20.9

                       2000                  41.1                                      21.1
                       2100                  43.6                                      21.9
                       2300                  48.7                                      25.3

                       2450                  52.4                                              24
                       2600                  55.3                                      24.6

                       3000                  63.8                                      25.7
                       3500                  67.1                                              25
                       3700                  67.4                                      24.2



                                         | Zmiem ]




   Sika

Bautrns
                                                                                  San (wikgl




                                                                                                       10 12 14 16 18
                 xiCC vie                                                                                      2 (mal




     7.3   BODY LIQUID MEASUREMENT
                    Fre'clt:-‘ezncy        Relative permittivity (€;)                Conductivity {0) S/m

                                          required           measured            required           measured

                        150               61.9 45 %                             0.80 15 %
                        300               58.2 45 %                             0.92 5 %

                        450               56.7 45 %             PASS            0.94 £5 %             PASS
                        750               55.5 45 %                             0.96 45 %
                        835               55.2 45 %                             0.97 25 %

                        900               55.0 45 %                             1.05 £5 %
                        915               55.0 45 %                             1.06 £5 %
                        1450              54.0 #5 %                             1.30 £5 %
                        1610              53.8 45 %                             1.40 £5 %
                        1800              53.3 45 %                             1.52 £5 %

                        1900              53.3 15 %                             1.52 45 %
                       2000               53.3 45 %                             1.52 £5 %
                       2100               53.2 45 %                             1.62 £5 %

                                                         Page: 9/11

           This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
           The information contained herein is to be used onlyfor the purposefor which it is submitted andis not to
           be released in whole orpart without written approval ofMVG.


                           SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR.69.2.17.SATU.A




                                2300            52.9 45 %                             1.81 £5 %

                                2450            52.7 25 %                             1.95 £5 %
                                2600            52.5 45 %                             2.16 £5 %

                                3000            52.0 £5 %                             2.73 45 %
                                3500            51.3 45 %                             3.31 15 %
                                3700            51.0 45 %                             3.55 45 %
                                5200           49.0 £10 %                            5.30 £10 %
                                5300           48.9 £10 %                             5.42 £10 %
                                5400           48.7 £10 %                             5.53 £10 %
                                5500           48.6 £10 %                             5.65 £10 %

                                5600           48.5 £10 %                             5.77 210 %
                                5800           48.2 £10 %                             6.00 £10 %


           7.4   SAR MEASUREMENT RESULT WITH BODY LIQUID
Software                                                               OPENSAR V4
Phantom                                                                SN 20/09 SAM71
Probe                                                                  SN18/11 EPG122
Liquid                                                                 Body Liquid Values: eps‘ : 57.6 sigma : 0.95
Distance between dipole center and liquid                              15.0 mm
Area scan resolution                                                   dx=8mm/dy=8mm
Zoon Scan Resolution                                                   dx=8mm/dy=8mm/dz=Smm
Frequency                                                              450 MHz
Input power                                                            20 dBm
Liquid Temperature                                                     21 °C
Lab Temperature                                                        21 °C
Lab Humidity                                                           45 %




                          Fm&fi““'                    1 g SAR (W/kg/w)                     10 g SAR (W/ke/w)
                                                         measured                               measured

                                450                      4.78 (0.48)                            3.19 (0.32)


                     td mniy                     snn                                               s           a



          h3                                                                                      (.’



                                                                                         §0.4
   Fram     e                                                                            &
        zesn

  e e                                                                                                         10 dz i4 i6 18
                                                                                                                                         f
                                                                                                                                2 26 28 30
                       xpat 6     vie)                                                                             2 (oenl

                                                              Page: 10/11

                 This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                 The information contained herein is to be used onlyfor the purposefor which it is submitted andis not to
                 be released in whole orpart without written approval ofMVG.


 mvG                       SAR REFERENCE DIPOLE CALIBRATION REPORT                                                    Ref: ACR.69.2.17.SATU.A




       8    LIST OF EQUIPMENT



                                                Equipment Summary Sheet

    Equipment                Manufacturer / Identification No                                  Current                 Next Calibration
    Description                     Model                                           ‘| Calibration Date                       Date
   SAM Phantom                        mvo                  SASomsgamy,                  [felesied Nowal
                                                                                      required.
                                                                                                                     WWelicatad
                                                                                                                     required.
                                                                                                                                     No wal
COMOSAR Test Bench|                Version 3                         NA                 elgalsd. Nogal
                                                                                      required.
                                                                                                                      Weldeled
                                                                                                                     required.
                                                                                                                                       hle mel
  Network Analyzer           Rhode8Schwarz                      sN100132                       ozpoi6                        0222019
      Calipers                       Carrera                  CALIPER—01                       01/2017                       01/2020

  Reference Probe                     MVG                 EPG122 SN 18/11                       10/2016                      10/2017

     Multimeter                 Keithley 2000                    1188656                        01/2017                      01/2020

  Signal Generator             Agilent E4438C                 MY49070581                        01/2017                      01/2020
           s                                                                          Characterized prior to |Characterized prior to
      Apilifiet                  Aethersomm                       $N 048              test. No cal required. |test. No cal required.

    Power Meter                  HP E4418A                    US38261498                        01/2017                      01/2020

    Power Sensor                HP ECP—E26A                   US37181460                        01/2017                      01/2020
   D   megh                                                                           Characterized prior to |Characterized prior to
 Directional Coupler           Narda 4216—20                       01386              test. No cal required. |test. No cal required.
  Temperature and
   Humidity Sensor           Control Company                   150798832                        10/2015                      10/2017




                                                               Page: 11/11

                  This document shall not be reproduced, except in full or in part, without the written approval ofMVG.
                  The information contained herein is to be used onlyfor the purposefor which it is submitted andis not to
                  be released in whole orpart without written approval ofMVG.



Document Created: 2017-06-02 14:10:34
Document Modified: 2017-06-02 14:10:34

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC