CALIBRATION DATA_Part 1

FCC ID: POD-DMR2

RF Exposure Info

Download: PDF
FCCID_3422047

CALIBRATION DATA
PROBE CALIBRATION DATA


Calibration
      4     Laboratory of                                                               .    Schweizerischer Kalibrierdienst
Schmid & Partner                                                                        .    Service suisse d‘étalonnage
  Engineering AG                                                                        «g   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                 Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                       Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates
Glossary:
TSL                         tissue simulating liquid
NORMx,y,z                   sensitivity in free space
ConvF                       sensitivity in TSL / NORMxy,z
DCP                         diode compression point
CF                          crest factor (1/duty_cycle) of the RF signal
A, B,C, D                   modulation dependent linearization parameters
Polarization                @ rotation around probe axis
Polarization 8              8 rotation around an axis that is in the plane normal to probe axis (at measurement center),
                            ie., 8 = 0 is normal to probe axis
Connector Angle             information used in DASY system to align probe sensor X to the robot coordinate system

Calibration is Performed According to the Following Standards:
    a)   IEEE Std 1528—2013, "IEEE Recommended Practicefor Determiningthe Peak Spatial—Averaged Specific
         Absorption Rate(SAR) in the Human Head from Wireless Communications Devices: Measurement
         Techniques®, June 2013
    b) IEC 62209—1, "Procedure to measurethe Specific Absorption Rate (SAR) for hand—held devices used in close
         proximity to the ear (frequency range of 300 MHz to 3 GHz)", February 2005
    c)   IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
         used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)®, March 2010
    d)   KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Methods Applied and Interpretation of Parameters:
         NORMx.y,z: Assessed for E—field polarization 3 = 0 (f < 900 MHz in TEM—cell; f > 1800 MHz: R22 waveguide).
         NORMsx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E*—field
         uncertainty inside TSL (see below ConvF).
         NORM(Ax,y,z = NORMx.y,z * frequency_response (see Frequency Response Chart). This linearization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
         in thestated uncertainty of ConvF.
         DCPx.y,z: DCP are numerical linarization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
         PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
         characteristics
         Axy.z: Bxy.z: Cay.2: Dxy,2; VRey.z: A, B, C, D are numerical linearization parameters assessed based on
         the data of power sweepfor specific modulation signal. The parameters do not depend on frequency nor
         media. VR is the maximum calibration range expressed in RMS voltage across the diode.
         ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature Transfer
         Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
         measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy closeto the boundary. The sensitivity in TSL corresponds
         to NORMsxy,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY version 4.4 and higher which allows extending the validity from + 50 MHz to + 100
         MHz.
         Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
         exposed by a patch antenna.
         Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.
         Connector Angle: The angle is assessed using the information gained by determining the NORMsx (no
         uncertainty required).



Certificate No: ES3—3337_Sep16                                 Page 2 of 11


ESSDV3 — SN:3337                                                                September 28, 2016




                      Probe ES3DV3

                                          SN:3337

                                  Manufactured:            January 24, 2012
                                  Calibrated:              September 28, 2016



                                 Calibrated for DASY/EASY Systems
                                    (Note: non—compatible with DASY2 system!)




Certificate No: ES3—3337_Sep16                    Page 3 of 11


ES3DV3— SN:3337                                                                                                          September 28, 2016



DASY/EASY — Parameters of Probe: ES3DV3 — SN:3337
Basic Calibration Parameters
                                                Sensor X                         Sensor Y                    Sensor Z              Une (k=2)
 Norm (uV/(Vim)?)"                                 1.06                           0.93                          0.97               £10.1%
 DCP ({mV)"                                       105.1                           106.0                        102.3

Modulation Calibration Parameters
 uiD           Communication System Name                                    A             B          C          D          VR          Unc®
                                                                           dB        dBvuV                     dB          mV          (k=2)
 0             Cw                                                 x        0.0           0.0        10         o00        1919        #3%
                                                                  y        o0            00         10                    1833
                                                                  z        o0            0.0        10                    1807


 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.

* The uncertainties of Norm X.Y Z do not affect the E*—feld uncertainty inside TSL (see Pages 5 and 6).
* Numerical linearization parameter: uncertainty not required
® Uncertainty is determined using the max. deviation from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: ES3—3337_Sep16                                     Page 4 of 11


ESEDV3— SN:3337                                                                                                           September 28, 2016




DASY/EASY — Parameters of Probe: ES3DV3 — SN:3337

Calibration Parameter Determined in Head Tissue Simulating Media

     1 (mHz)         Fa’r‘l:lnmy'        cn’(‘g;:nc;"my      ConvFX         ConvFY        ConvFZ         Alpha®       r’('.f:.t.?)G    (::;)

        150.              52.3                0.76             787            7.67           7.87         0.08          120           £133%
        450               43.5                0.87              7.04          7.04           7.04         0.19          220           £133%
        900               41.5                0.97              6.56          6.56           6.56         0.80          1.14          £120%
        1810              40.0                140               5.52          5.52           5.52         0.80          112           £120%
       2000               40.0                 1.40             5.47          547            5.47         o.71          125           £120%
       2450               30.2                 1.80            4.83           4.83           4.83         0.80          122|          £120%


© Frequency validty above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), elsei is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at callbration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHiz.
" At frequencies below 3 GHz, the validity of issue parameters (and a) can be relaxed to + 10% i iquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (c and a) is restricted to + 5%. The uncertainty is the RSS of
the Convuncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: ES3—3337_Sep16                                     Page 5 of 11


ES3DV3— SN:3337                                                                                                         September 28, 2016




DASY/EASY — Parameters of Probe: ES3DV3 — SN:3337

Calibration Parameter Determined in Body Tissue Simulating Media
                   Relative              Conductivity                                                                Depth °          Un
     £(MHz)®_|_Permittivity®                (Sim)"           ConvFX        ConvFY        ConvFZ         Aipha®        |__(mm)        (k22)

        150.              61.9                0.80             747           747            747          0.06          1.20        £133%
        450               56.7                0.94             7.37          7.37           7.37         0.12          1.60        £133%
        900               55.0                1.05             644           6.44           644          0.43          1.62_|      £120%
       1810               53.3                1.52             5.15          5.15           5.15         0.52          153         £120%
       2000               53.3                1.52             5.13          5.13           5.13         0.75          1.30        £120%
       2450               52.7                1.95             4.63          4.63           4.63         0.74          123         £120%


© Frequency validty above 300 MHz of + 100 MHz only applies for DASY v4.4 and higher (see Page 2), elsei is restricted to * 50 Miz. The
uncertainty is the RSS of the Convuncertainty at callbration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128, 150 and 220 MHz respectively. Above 5 GHz frequency
validity can be extended to + 110 MHz.
* At frequencies below 3 GHz, the validity of issue parameters (c and a) can be relaxed to * 10% if iquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the valiity of tissue parameters (c and a) is restricted to + 5%. The uncertainty is the RSS of
the Conv: uncertainty for indicated target tissue parameters.
© Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
always less than + 1% for frequencies below 3 GHz and below + 2% for frequencies between 3—6 GHiz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: ES3—3337_Sep16                                    Page 6 of 11


ES3DV3— SN:3397                                                                                          September 28, 2016

                                             Frequency Response of E—Field
                                                  (TEM—Cell:ifi110 EXX, Waveguide: R22)
          Frequency response (normalized)




                                            Uncertainty of Frequency Response of E—field: 2 6.3% (k=2)




Certificate No: ES3—3337_Sep16                                     Page 7 of 11


ES3DV3— SN:3337                                                                              September 28, 2016




                                   Receiving Pattern (¢), 9 = 0°

                      £=600 MHz,TEM                                       £=1800 MHz,R22




                      10%2                     SU%Z                  WfiHHI                255@}‘1


                                 Uncertainty of Axial Isotropy Assessment: 2 0.5% (k=2)




Certificate No: ES3—3337_Sep16                        Page 8 of 11


ES3DV3— SN:3337                                                                                                      September 28, 2016




                                                    Dynamic Range f(SARneaq)
                                                              (TEM cell , fayi= 1900 MHz)


                                      10°
                  Input Signal [uV)




                                      102




                                            102        102          101          10       10‘           10‘         10°
                                                                          SAR [mWicm3]

                                                       not compensated
                                                                                                Ce]
                                                                                          compensated




                                             10#        102         101         109      10           102     103
                                                                          SAR {mW/cm3]
                                                              _e]
                                                       not compensated                   compensated

                                                   Uncertainty of Linearity Assessment: £ 0.6% (k=2)




Certificate No: ES3—3337_Sep16                                            Page9 of 11


ES3DV3— SN:3337                                                                                             September 28, 2016

                                    Conversion Factor Assessment
                 f = 900 MHz,WGLS R9 (H_convF)                                    {= 1810 MHz,WGLS R22 (H_convF)

           10f                                                           ${|
           ss!                                                               t
                                                                         s|
           so:                                                               F4

       & as!                                                         3 =|

       &     P                                                       5
       5*4             *                                             FRG

           1s]                                                               L
                                                                         10|
           Bi                               %
           as                                   F                        s



           so>                                        ncaaaad            s         prora                           :
             S   1k x tifim
                         aA CR CE 44                                         $     T3 % sitm a &#                      %
                  wetthe
                    +
                               nealsee                                              wolihe
                                                                                      al
                                                                                               wattes
                                                                                                 a



                                   Deviation from Isotropy in Liquid
                                                    Error (¢, 8), f= 900 MHz
                       Deviation




                             —1.0     —08   —06     .04   —02   00        02       04      0.6   08   1.0
                                   Uncertainty of Spherical Isotropy Assessment: £ 2.6% (k=2)




Certificate No: ES3—3337_Sep16                             Page 10 of 11




DAE CALIBRATION DATA







Document Created: 2017-06-02 14:10:19
Document Modified: 2017-06-02 14:10:19

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC