SAR Test Report Yageo

FCC ID: PD911230BNHU

RF Exposure Info

Download: PDF
FCCID_1529695

                                                                                                                 AT4 wireless, S.A.
                                                                                                              Parque Tecnológico de Andalucía,
                                                                                                                     c/ Severo Ochoa nº 2
                                                                                                             29590 Campanillas/ Málaga/ España
                                                                                                             Tel. 952 61 91 00 - Fax 952 61 91 13
                                                                                                                MÁLAGA, C.I.F. A29 507 456
                                        51/LE342                                                           Registro Mercantil de Málaga,Tomo 1169,
                                                                                                              Libro 82, Folio 133, Hoja MA3729

                                                                          TEST REPORT
                                                                REFERENCE STANDARDS:
                                                            FCC 47CFR Part 2.1093 (10-1-09 Edition)
                                                   FCC OET Bulletin 65, Supplement C (Edition 01-01)
                                                                  IC RSS-102 Issue 4 (2010-03)
NIE........................................................ :     34184RRF.001                                                       Firmado digitalmente por
                                                                                                       Alejandro Llamas Alejandro Llamas Rodríguez
Approved by                                                                                            Rodríguez        Fecha: 2011.08.24 11:40:26
                                                                                                                        +02'00'
(name / position & signature) ............... :                   A. Llamas / RF Lab. Manager
                                                                                                          ..........................................................
Elaboration date .................................... :           2011-08-24
Identification of item tested ................ :                  Intel® Centrino® Wireless-N 1030 inside a host laptop
Trademark ............................................ :          Intel
Model and/or type reference ................. :                   USA: 11230BNHMW / Canada: 11230BNHU
Serial number ........................................ :          MAC: BC7737794FD7
Other identification of the product ........ :                    FCC ID: PD911230BNHU / IC: 1000M-11230BNHU
Features ................................................. :      802.11 b/g/n
Description ............................................ :        Wireles Module: Intel® Centrino® Wireless-N 1030
                                                                  Antenna Type: Yageo 25.90A72.011 / 25.90A71.011
                                                                  Host platform: Lenovo U400
Applicant ............................................. :         Intel Corporation
Address.................................................. :       100 Center Point Circle, Suite 200, Columbia, SC 29210, USA
CIF/NIF/Passport................................... :             ---
Contact person:                                                   Steven C. Hackett (Wireless Regulatory Engineer)
Telephone / Fax ..................................... :           803-216-2344
e-mail: ................................................... :     steven.c.hackett@intel.com
Test samples supplier ......................... :                 Same as applicant
Manufacturer ...................................... :             Same as applicant




         Report Nº(NIE): 34184RRF.001                                          Page 1 of 33                                                 2011-08-24


Test method requested ........................: See Standard
Standard ................................................: 1. FCC 47 CFR Part 2.1093 (10-1-09 Edition). Radiofrequency radiation
                                                              exposure evaluation: portable devices.
                                                           2. FCC OET Bulletin 65, Supplement C (Edition 01-01), “Evaluating
                                                              Compliance with FCC Guidelines for Human Exposure to Radio
                                                              Frequency Electromagnetic Fields”.
                                                           3. IC RSS-102 Issue 4 (2010-03). Radio Frequency (RF) Exposure
                                                              Compliance of Radiocommunication Apparatus (All Frequency
                                                              Bands)
Application Notes ..................................:      1. FCC OET KDB 248227 – SAR Measurements Procedures
                                                              802.11a/b/g Transmitters (May 2007 – Revised).
                                                           2. FCC OET KDB 616217 D03 – SAR Evaluation Considerations for
                                                              Notebook/Netbook and Laptop Computers (November 2009).
                                                           3. FCC OET KDB 450824 – SAR Probe Calibration and System
                                                              Verification Considerations for measurements at 150 MHz – 3 GHz
                                                              (January 2007).
                                                           4. IC RSS-102 Supplementary Procedures (SPR)-001 (2011-01). SAR
                                                              testing requirements with regard to bystanders for laptop type
                                                              computers with antennas built-in on display screen (laptop mode/tablet
                                                              mode).
Test procedure .......................................: PERF019
Non-standardized test method ...............: N/A
Used instrumentation .............................: 1. Dosimetric E-field probe SPEAG ES3DV3
                                                    2. Data acquisition device SPEAG DAE4
                                                    3. Electro-optical converter SPEAG EOC3
                                                    4. 2450 MHz dipole validation kit SPEAG D2450V2
                                                    5. Robot STÄUBLI RX60BL
                                                    6. Robot controller STÄUBLI CM7MB
                                                    7. Oval flat phantom SPEAG ELI 4
                                                    8. SAR measurement software SPEAG DASY52 V52.6.2.424
                                                    9. Measurement server SPEAG DASY5 SE UMS 011 BS
                                                    10. Body Tissue Equivalent Liquids for 2450MHz band
                                                    11. Vector network analyzer Agilent E5071C
                                                    12. Dielectric probe kit Agilent 85070C
                                                    13. Power meter R&S NRVD
                                                    14. Power Sensor R&S NRV-Z51
                                                    15. Power Sensor R&S NRV-Z1
                                                    16. RF Generator Agilent ESG E4438C
                                                    17. Dual directional coupler NARDA FSCM 99899
                                                    18. Power amplifier MITEQ AMF-4D-00400600-50-30P
                                                    19. Laptop positioning extension SPEAG Laptop Holder
Report template No. .............................: FDT08_12
IMPORTANT: No parts of this report may be reproduced or quoted out of context, in any form or by any means, except in full, without the previous
written permission of AT4 wireless.




Report Nº(NIE): 34184RRF.001                                      Page 2 of 33                                                   2011-08-24


                                                                         INDEX

Competences and guarantees ................................................................................................................ 4
General conditions ................................................................................................................................ 4
Uncertainty............................................................................................................................................ 4
Usage of samples................................................................................................................................... 5
Testing period ....................................................................................................................................... 5
Environmental conditions ..................................................................................................................... 5
Summary ............................................................................................................................................... 6
Remarks and comments ........................................................................................................................ 6
Testing verdicts ..................................................................................................................................... 6
APPENDIX A: Test Configuration....................................................................................................... 7
APPENDIX B: Test results................................................................................................................. 15
APPENDIX C: Measurements Reports .............................................................................................. 20
APPENDIX D: Calibration Data ........................................................................................................ 23
APPENDIX E: Photographs ............................................................................................................... 31




Report Nº(NIE): 34184RRF.001                                             Page 3 of 33                                                         2011-08-24


Competences and guarantees
AT4 wireless is a testing laboratory accredited by the National Accreditation Body (ENAC -Entidad Nacional de
Acreditación), to perform the tests indicated in the Certificate No. 51/LE 342.
In order to assure the traceability to other national and international laboratories, AT4 wireless has a calibration
and maintenance programme for its measurement equipment.
AT4 wireless guarantees the reliability of the data presented in this report, which is the result of the
measurements and the tests performed to the item under test on the date and under the conditions stated on the
report and, it is based on the knowledge and technical facilities available at AT4 wireless at the time of
performance of the test.
AT4 wireless is liable to the client for the maintenance of the confidentiality of all information related to the item
under test and the results of the test.
General conditions
       1.   This report is only referred to the item that has undergone the test.
       2.   This report does not constitute or imply on its own an approval of the product by the Certification
            Bodies or competent Authorities.
       3.   This document is only valid if complete; no partial reproduction can be made without previous
            written permission of AT4 wireless.
       4.   This test report cannot be used partially or in full for publicity and/or promotional purposes without
            previous written permission of AT4 wireless and the Accreditation Bodies.
Uncertainty
       Uncertainty (factor k=2) was calculated according to the following documents:


       1.   FCC OET Bulletin 65, Supplement C (Edition 01-01), “Evaluating Compliance with FCC Guidelines for
            Human Exposure to Radio Frequency Electromagnetic Fields”.
       2.   FCC OET KDB 865664 – SAR Measurements Requirements for 3-6 GHz (October 2006).




Report Nº(NIE): 34184RRF.001                           Page 4 of 33                                        2011-08-24


Usage of samples
Samples undergoing test have been selected by: the client.
Sample M/01 is composed of the following elements:
     Control Nº          Description                Model                  Serial Nº          Date of reception
      34184/01          WiFi module        Intel Centrino® Wireless-
                                                 ®
                                                                            MAC:                2011-08-08
                     inside a Laptop PC     N 1030 / Lenovo U400 /       BC7737794FFA
                                                Yageo Antennas

1.      Sample M/01 has undergone the test(s) specified in subclause “Test method requested”.


Testing period
The performed test started on 2011-08-12 and finished on 2011-08-22.
The tests have been performed at AT4 wireless.


Environmental conditions
In the laboratory for measurements, the following limits were not exceeded during the test:


                  Temperature                                Min. = 22.89 ºC
                                                             Max. = 25.41 ºC
                  Relative humidity                          Min. = 47.03 %
                                                             Max. = 61.46 %




Report Nº(NIE): 34184RRF.001                         Page 5 of 33                                      2011-08-24


Summary
Considering the results of the performed test according to FCC 47CFR Part 2.1093, the item under test is IN
COMPLIANCE with the requested specifications specified in the standard.
The maximum 1g volume averaged SAR found during this test has been 0.502 W/kg, for the 2450 MHz band and
802.11b mode.
NOTE: The results presented in this Test Report apply only to the particular item under test established in page 1
of this document, as presented for test on the date(s) shown in section, “USAGE OF SAMPLES, TESTING
PERIOD AND ENVIRONMENTAL CONDITIONS”.
Remarks and comments
1: 802.11n = 20MHz BW / 802.11n* = 40MHz BW.
2: Testing of 802.11g and 802.11n is not required due to the testing reductions mentioned in FCC OET KDB
248227 – SAR Measurements Procedures 802.11a/b/g Transmitters (May 2007 – Revised), paragraph
“Frequency Channel Configurations”.
3: Testing of other channels in each band is optional when the maximum output channel SAR fulfills the testing
reductions mentioned in FCC OET KDB 248227 – SAR Measurements Procedures 802.11a/b/g Transmitters
(May 2007 – Revised), paragraph “Frequency Channel Configurations”.
4: All the test positions of device relative to body were measured placing the device in direct contact with the
phantom surface, so the requirements mentioned at RSS-102 Supplementary Procedures (SPR)-001 - SAR
TESTING REQUIREMENTS WITH REGARD TO BYSTANDERS FOR LAPTOP TYPE COMPUTERS
WITH ANTENNAS BUILT-IN ON DISPLAY SCREEN (LAPTOP MODE/TABLET MODE) are covered.
Antenna to phantom distance during measurements was 7 mm (< 25 mm).


Testing verdicts
Not applicable .................................................... :          NA
Pass..................................................................... :    P
Fail ..................................................................... :   F
Not measured...................................................... :           NM




2450 MHz band
                                     FCC 47CFR Part 2.1093 Paragraph                                   VERDICT
                                                                                                  NA    P   F  NM
(d)(2)      802.11b                                                                                     P
(d)(2)      802.11g                                                                                            NM2
(d)(2)      802.11n1                                                                                           NM2
(d)(2)      802.11n*1                                                                                   P
1 and 2: See Remarks and Comments.




Report Nº(NIE): 34184RRF.001                                                       Page 6 of 33             2011-08-24


               APPENDIX A: Test Configuration




Report Nº(NIE): 34184RRF.001   Page 7 of 33     2011-08-24


                                                                           INDEX

   1.     GENERAL INTRODUCTION .............................................................................................................. 9
   1.1. Application Standard.............................................................................................................................. 9
   1.2. General requirements ............................................................................................................................. 9
   1.3. Measurement system and phantom requirements ................................................................................... 9
   1.4. Measurement Liquids requirements. ...................................................................................................... 9


   2.     MEASUREMENT SYSTEM............................................................................................................... 10
   2.1. Measurement System............................................................................................................................ 10
   2.2. Test Positions of device relative to body 4 ........................................................................................... 11
   2.3. Test to be performed ............................................................................................................................ 11
   2.4. Description of interpolation/extrapolation scheme............................................................................... 11
   2.5. Determination of the largest peak spatial-average SAR ....................................................................... 11
   2.6. System Validation ................................................................................................................................ 12


   3.     UNCERTAINTY ................................................................................................................................. 13


   4.     SAR LIMIT.......................................................................................................................................... 14




Report Nº(NIE): 34184RRF.001                                               Page 8 of 33                                                              2011-08-24


1. GENERAL INTRODUCTION

1.1.     Application Standard
The Federal Communications Commission (FCC) sets the limits for General Population /
Uncontrolled exposure to radio frequency electromagnetic fields for transmitting devices designed
to be used within 20 centimetres of the user body under FCC 47 CFR Part 2.1093 -
“Radiofrequency radiation exposure evaluation: portable devices”, paragraph (d)(2).
Specific requirements and procedure for SAR assessment are describe under FCC OET Bulletin 65,
Supplement C (Edition 01-01), “Evaluating Compliance with FCC Guidelines for Human Exposure
to Radio Frequency Electromagnetic Fields”, and all the FCC OET Knowledge Database documents
referred at the beginning of this document.

1.2.     General requirements
The SAR measurement has been performed continuing the following considerations and
environment conditions:
       - The ambient temperature shall be in the range of 18ºC to 25ºC and the variation shall not
       exceed +/-2ºC during the test.
       - The ambient humidity shall be in the range of and 30% - 70%.
       - The device battery shall be fully charged before each measurement.

1.3.     Measurement system and phantom requirements
The measurement system used for SAR tests fulfils the procedural and technical requirements
described at the reference standards used.
The phantom is a simplified representation of the human anatomy and comprised of material with
electrical properties similar to the corresponding tissues in human body.

1.4.     Measurement Liquids requirements.
The liquids used to simulate the human tissues, must fulfils the requirements of the dielectric
properties required. These target dielectric properties per FCC OET KDB 450824 instructions come
from the dipole and probe calibration data which are included in Appendix B, Section 2, of this
document.
As indicated in FCC OET KDB 450824, it is allowed a 5% variation of the above mentioned level at
the 2450 MHz band.




Report Nº(NIE): 34184RRF.001                      Page 9 of 33                             2011-08-24


2. MEASUREMENT SYSTEM

2.1.   Measurement System


                    Manufacturer                   Device                  Type

                   Schmid & Partner
                                           Dosimetric E-Fiel Probe       ES3DV3
                    Engineering AG
                   Schmid & Partner           Data Acquisition
                                                                          DAE4
                    Engineering AG              Electronics
                   Schmid & Partner            Electro-Optical
                                                                           EOC5
                    Engineering AG               Converter
                   Schmid & Partner           2450 MHz System
                                                                         D2450V2
                    Engineering AG            Validation Dipole
                        Stäubli                     Robot                RX60BL
                        Stäubli                Robot controller          CM7MB
                   Schmid & Partner
                                              Oval flat phantom            ELI 4
                    Engineering AG
                   Schmid & Partner                                      DASY52
                                            Measurement Software
                    Engineering AG                                      V52.6.2.424
                   Schmid & Partner                                   DASY5 SE UMS
                                             Measurement Server
                    Engineering AG                                       011 BS
                                               Vector Network
                        Agilent                                           E5071C
                                                  Analyser
                        Agilent              Dielectric Probe Kit         85070C
                  Rohde & Schwarz               Power Meter               NRVD
                  Rohde & Schwarz               Power Sensor             NRV-Z51
                  Rohde & Schwarz               Power Sensor             NRV-Z1
                        Agilent                 RF Generator           ESG E4438C
                       NARDA               Dual directional coupler    FSCM 99899
                                                                         AMF-4D-
                       MITEQ                   Power amplifier
                                                                      00400600-50-30P
                   Schmid & Partner
                                                Laptop Holder         SM LH1 001 AC
                    Engineering AG

                                  Table 1: Measurement Equipment




Report Nº(NIE): 34184RRF.001                  Page 10 of 33                             2011-08-24


2.2.   Test Positions of device relative to body 4
The laptop device was tested in one position for all tests, with the bottom face placed directly
against the phantom so the position of the laptop would be used (normal use condition). Further
analysis was performed to determine the location which showed the highest SAR.
The antennas on the laptop are located within the edge screen. According to FCC OET Bulletin 65 –
Supplement C, the antennas which would applied in the test are antennas or radiating structures in
direct contact with the user’s body within 20 centimetres of the body of a user under normal
operating conditions.



2.3.   Test to be performed
In all operating modes and bands the measurements have to be performed on the “default test
channels” defined at FCC OET KDB 248227 – SAR Measurements Procedures 802.11a/b/g
Transmitters (May 2007 – Revised), except those channels defined as “required test channels” at the
same document.



2.4.   Description of interpolation/extrapolation scheme
The local SAR inside the Phantom is measured using small dipole sensing elements inside a probe
element. The probe tip must not be in contact with the Phantom’s surface in order to minimise
measurement errors, but the highest local SAR is obtained from measurements at a certain distances
from the shell trough extrapolation. The accurate assessment of the maximum SAR averaged over 1
gr. and 10 gr. requires a very fine resolution in the three dimensional scanned data array. Since the
measurements have to be performed over a limited time, the measured data have to be interpolated
to provide an array of sufficient resolution.
The interpolation of 2D area scan is used after the initial area scan, at a fixed distance from the
Phantom shell wall. The initial scan data is collected with a proper spatial resolution and this
interpolation is used to find the location of the local maximum for positioning the subsequent 3D
scanning to within a 1mm resolution.
For the 3D scan, data is collected on a spatially regular 3D grid having 5mm steps in both lateral
directions, and 5mm in depth direction for the 2450MHz band. After the data collection by the SAR
probe, the data are extrapolated in the depth direction to assign values to points in the 3D array
closer to the shell wall. A notional extrapolation value is also assigned to the first point outside the
shell wall so that subsequent interpolation schemes will de applicable right up to the shell wall
boundary.



2.5.   Determination of the largest peak spatial-average SAR
To determine the maximum value of the peak spatial-average SAR of a EUT, all device positions,
configurations and operational modes should be tested for each frequency band.
According to FCC 47 CFR Part 2.1093, the averaging volume shall be chosen as 1 g of contiguous
tissue. The cubic volumes, over which the SAR measurements are averaged after extrapolation and
interpolation, are chosen in order to include the highest values of local SAR.
The maximum SAR level for the EUT will be the maximum level obtained of the performed
measurements, and indicated in the previous points.



Report Nº(NIE): 34184RRF.001                     Page 11 of 33                                   2011-08-24


2.6.   System Validation
Prior to the SAR measurements, system verification is done daily to verify the system accuracy. As
FCC OET Bulletin 65 – Supplement C, Appendix D “SAR measurement procedures” Paragraph
“System Verification” specifies, a complete SAR evaluation is done using a half-wavelength dipole
as source with the frequency of the mid-band channel of the operating band, or within 100MHz of
this channel.
The measured 1 gr. and 10 gr. SAR should be within 10% of the expected target values specified in
the calibration certificate of the dipole, for the specific tissue and frequency used.




Report Nº(NIE): 34184RRF.001                  Page 12 of 33                                2011-08-24


  3. UNCERTAINTY

  Uncertainty for 300 MHz – 6 GHz

                                                                                            Standard      Standard
                               Uncertainty   Probability                     (ci)   (ci)                                vi
   ERROR SOURCES                                                 Divisor                   uncertainty   uncertainty
                                value (%)    distribution                    1g     10g                                veff
                                                                                            (1g) (%)      (10g) (%)
Measurement Equipment
     Probe Calibration           ±4.480        Normal                1        1      1       ±4.480        ±4.480          ∞
      Axial Isotropy             ±7.558      Rectangular                 3    1      1       ±4.364        ±4.364          ∞
  Hemispherical Isotropy         ±2.000      Rectangular                 3    1      1       ±1.155        ±1.155          ∞
      Boundary effect            ±4.700      Rectangular                 3    1      1       ±2.714        ±2.714          ∞
         Linearity               ±1.000      Rectangular                 3    1      1       ±0.577        ±0.577          ∞
  System detection limits        ±0.300      Rectangular                 3    1      1       ±0.300        ±0.300          ∞
    Readout electronics          ±1.010        Normal                1        1      1       ±0.583        ±0.583          ∞
      Response time              ±2.600      Rectangular                 3    1      1       ±1.501        ±1.501          ∞
      Integration time           ±3.000      Rectangular                 3    1      1       ±1.732        ±1.732          ∞
  RF Ambien conditions           ±3.000      Rectangular                 3    1      1       ±1.732        ±1.732          ∞
     Probe positioner            ±0.800      Rectangular                 3    1      1       ±0.462        ±0.462          ∞
     Probe positioning           ±9.900      Rectangular                 3    1      1       ±5.716        ±5.716          ∞
Maximum SAR evaluation           ±1.000      Rectangular                 3    1      1       ±0.577        ±0.577          ∞
 Test Sample Related
  Device positioning             ±2.900        Normal                1        1      1       ±2.900        ±2.900      145
      Device Holder              ±3.600        Normal                1        1      1       ±3.600        ±3.600          5
       Power Drift               ±5.000      Rectangular                 3    1      1       ±2.887        ±2.887          ∞
   Phantom and Setup
   Phantom uncertainty           ±4.000      Rectangular                 3    1      1       ±2.309        ±2.309          ∞
   Liquid conductivity
                                 ±5.000      Rectangular                 3   0.64   0.43     ±1.848        ±1.241          ∞
  (deviation from target)
   Liquid conductivity
                                 ±3.100        Normal                1       0.64   0.43     ±1.984        ±1.333          ∞
   (measurement error)
    Liquid permittivity
                                 ±5.000      Rectangular                 3   0.64   0.43     ±1.848        ±1.241          ∞
  (deviation from target)
    Liquid permittivity
                                 ±4.410        Normal                1       0.64   0.43     ±2.822        ±1.896          ∞
   (measurement error)
                                                     m
   Combined standard
      uncertainty                            uc =   ∑c      i
                                                             2
                                                                 ⋅ ui2                       ±11.99        ±11.56      330
                                                     1=1
   Expanded uncertainty
                                                ue =2.00 uc                                  ±23.98        ±23.11
(confidence interval of 95%)


                               Table 2: Uncertainty Assessment for 300 MHz - 6 GHz




  Report Nº(NIE): 34184RRF.001                          Page 13 of 33                                         2011-08-24


4. SAR LIMIT
Having a worst case measurement, the SAR limit is valid for general population/uncontrolled
exposure.
The SAR values have to be averaged over a mass of 1g (SAR 1 gr.) with the shape of a cube. This
level couldn’t exceed the values indicated in the application Standard:


                Standard                         SAR                  SAR Limit (W/Kg)

        FCC 47 CFR Part 2.1093
                                               SAR 1 gr.                      1.6
           Paragraph (d)(2)

                                      Table 4: SAR limit




Report Nº(NIE): 34184RRF.001                 Page 14 of 33                              2011-08-24


                      APPENDIX B: Test results




Report Nº(NIE): 34184RRF.001    Page 15 of 33    2011-08-24


                                                                    INDEX

    1.     TEST CONDITIONS...............................................................................................................................17
    1.1.   Temperature (ºC):.....................................................................................................................................17
    1.2.   Test signal, Output Power and Frequencies..............................................................................................17
    1.3.   DUT information ......................................................................................................................................17

    2.     TISSUE PARAMETERS MEASUREMENTS........................................................................................18

    3.     SYSTEM VALIDATION MEASUREMENTS .......................................................................................18
    3.1.   Validation results in 2450 MHz Band for Body TSL...............................................................................18

    4.     CONDUCTED AVERAGE POWER MEASUREMENTS .....................................................................18

    5.     MEASUREMENT RESULTS FOR SAR (SPECIFIC ABSORPTION RATE) ......................................19
    5.1.   Summary maximum results.......................................................................................................................19
    5.2.   Results for 2450 MHz Band – Lapheld ....................................................................................................19




Report Nº(NIE): 34184RRF.001                                          Page 16 of 33                                                             2011-08-24


1. TEST CONDITIONS


1.1.   Temperature (ºC):
Tn = +22.89 to +25.41

The subscript n indicates normal test conditions.



1.2.   Test signal, Output Power and Frequencies
The device was put into operation by using an own control software to program the test mode
required for select the continuous transmission with 100% duty cycle.
In all operating bands the measurements were performed on the “default test channels” defined at
FCC OET KDB 248227 – SAR Measurements Procedures 802.11a/b/g Transmitters (May 2007 –
Revised), except those that fulfil the frequency channel selection criteria mentioned on paragraph
“Frequency Channel Configuration” at the same document.
The output power of the device was set to maximum power level for all tests; a fully charged battery
was used for every test sequence.
The maximum average conducted power of the device was measured with a Power meter R&S
NRVD and a thermocoupled Power sensor NRV-Z51.



1.3.   DUT information
The device under test was the Intel® Centrino® Wireless-N 1030 card located inside a host device
(Lenovo U400) computer which utilises a set of Yageo antennas (25.90A72.011 / 25.90A71.011).
The card was operated utilizing proprietary software (DRTU version 1.3.9-0235) and each channel
was measured using a broadband power meter to determine the maximum average power.


According to host device manufacturer, the source-based time-averaged output power of the
Bluetooth device is far bellow the SAR threshold (60/f(GHz) mW). Therefore, neither SAR testing
nor co-transmission evaluation is required for the Bluetooth transmitter, following the guidelines
stated at FCC OET KDB 447498 – Mobile and Portable Device, RF Exposure Procedures and
Equipment Authorization Policies (November 2009), paragraph 4) a).


Collocation with WWAN transmitter has not been considered following grantee request because this
collocation will be considered in the WWAN transmitter SAR testing.




Report Nº(NIE): 34184RRF.001                    Page 17 of 33                                2011-08-24


  2. TISSUE PARAMETERS MEASUREMENTS
                    Target Body Tissue:              Target Body Tissue:
                  Parameters used in Probe         Parameters used in Dipole         Measured Body Tissue
Frequency               Calibration                      Calibration                                                  Measured
  (MHz)                                                                                                                Date
                                 Conductivity                     Conductivity                      Conductivity
              Permittivity                        Permittivity                     Permittivity
                                    [S/m]                            [S/m]                             [S/m]

  2450           52.7 ± 5%           1.95 ± 5%     53.2 ± 6%       2.00 ± 6%          51.17            1.98          2011-08-22

  Note: The dielectric properties have been measured by the contact probe method at 24° C.




  3. SYSTEM VALIDATION MEASUREMENTS

  3.1.      Validation results in 2450 MHz Band for Body TSL

                               Target SAR               Measured SAR
         SAR                                                                         Drift (%)                Limit (%)
                                 (W/kg)                    (W/kg)
         1 gr.                    53.00                     51.96                      -1.97                       ± 10
         10 gr.                       25.00                    23.88                   -4.46                       ± 10




  4. CONDUCTED AVERAGE POWER MEASUREMENTS
                                                                                                  Conducted
                                                                       Frequency
                        Mode                     Channel                                           Power
                                                                         (MHz)
                                                                                                    (dBm)
                                                    1                    2412                        16.61
                       802.11b                      6                    2437                       16.98
                                                   11                    2462                       16.64
                                                    1                    2412                       16.9
                       802.11g                      6                    2437                       16.67
                                                   11                    2462                       14.18
                                                    1                    2417                       14.15
                               1
                       802.11n                      6                    2437                       16.79
                                                   11                    2457                       13.92
                                                    3                    2422                       11.99
                                 1
                      802.11n*                      6                    2437                       14.90
                                             9                           2452                       12.43
                  1: See Remarks and Comments.




  Report Nº(NIE): 34184RRF.001                              Page 18 of 33                                          2011-08-24


5. MEASUREMENT RESULTS FOR SAR (SPECIFIC ABSORPTION
   RATE)


5.1.     Summary maximum results

       2450 MHz band:

                                                                                             SAR limit
                                                      Frequency        Measured SAR 1g
            Band          Mode         Channel                                                  1g
                                                        (MHz)              (W/Kg)
                                                                                              (W/Kg)
                         802.11b          6              2437               0.502               1.6

          2450 MHz       802.11g          -                -                NM2                 1.6
            band         802.11n   1
                                          -                -                NM   2
                                                                                                1.6
                         802.11n*1        6              2437               0.334               1.6
             1 and 2: See Remarks and Comments.




5.2.     Results for 2450 MHz Band – Lapheld

                                                            SAR averaged
                                         Frequency                             Power         Limit
              Mode         Channel                            over 1g
                                           (MHz)                              Drift (%)       (%)
                                                              (W/Kg)
                               1              2412             NM3                     -      ±5
             802.11b           6              2437              0.502                0.23     ±5
                               11             2462                NM3                  -      ±5
                                                                       2
                               1              2412                NM                   -      ±5
                                                                       2
             802.11g           6              2437                NM                   -      ±5
                               11             2462                NM2                  -      ±5
                                                                       2
                               1              2417                NM                   -      ±5
                     1                                                 2
             802.11n           6              2437                NM                   -      ±5
                                                                       2
                               11             2457                NM                   -      ±5
                                                                       3
                               3              2422                NM                   -      ±5
             802.11n*1         6              2437              0.334                -0.34    ±5
                                                                       3
                                9           2452                  NM                   -      ±5
             1, 2 and 3: See Remarks and Comments.




Report Nº(NIE): 34184RRF.001                         Page 19 of 33                                    2011-08-24


          APPENDIX C: Measurements Reports




Report Nº(NIE): 34184RRF.001   Page 20 of 33   2011-08-24


2450 MHz Band – 802.11b Channel 6 – Lapheld
DUT: Lenovo U400 + Intel 12230BNHMW + Yageo; Type: Laptop; Serial: 814PJ01007G0001J

Communication System: 802.11; Frequency: 2437 MHz; Duty Cycle: 1:1
Medium parameters used (interpolated): f = 2437 MHz; σ = 1.967 mho/m; εr = 51.209; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY5 Configuration:
- Probe: ES3DV3 - SN3052; ConvF(4.14, 4.14, 4.14); Calibrated: 22/10/2010
- Sensor-Surface: 4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 4mm
(Mechanical Surface Detection)
- Electronics: DAE4 Sn669; Calibrated: 03/02/2011
- Phantom: Flat Phantom ELI4.0; Type: QDOVA001BA; Serial: SN:1060
- Measurement SW: DASY52, Version 52.6 (2); SEMCAD X Version 14.4.4 (2829)

2450 MHz - Lapheld/Chain A, 802.11b, Channel 6/Area Scan (61x351x1): Measurement grid: dx=10mm,
dy=10mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (interpolated) = 0.580 mW/g

2450 MHz - Lapheld/Chain A, 802.11b, Channel 6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 17.488 V/m; Power Drift = 0.02 dB
Peak SAR (extrapolated) = 1.275 W/kg
SAR(1 g) = 0.502 mW/g; SAR(10 g) = 0.210 mW/g
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.586 mW/g




                               0 dB = 0.590mW/g




Report Nº(NIE): 34184RRF.001                      Page 21 of 33                                     2011-08-24


2450 MHz Band – 802.11n* Channel 6 – Lapheld
DUT: Lenovo U400 + Intel 12230BNHMW + Yageo; Type: Laptop; Serial: 814PJ01007G0001J

Communication System: 802.11; Frequency: 2437 MHz; Duty Cycle: 1:1
Medium parameters used (interpolated): f = 2437 MHz; σ = 1.967 mho/m; εr = 51.209; ρ = 1000 kg/m3
Phantom section: Flat Section

DASY5 Configuration:
- Probe: ES3DV3 - SN3052; ConvF(4.14, 4.14, 4.14); Calibrated: 22/10/2010
- Sensor-Surface: 4mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 4mm
(Mechanical Surface Detection)
- Electronics: DAE4 Sn669; Calibrated: 03/02/2011
- Phantom: Flat Phantom ELI4.0; Type: QDOVA001BA; Serial: SN:1060
- Measurement SW: DASY52, Version 52.6 (2); SEMCAD X Version 14.4.4 (2829)

2450 MHz - Lapheld/Chain A, 802.11n*, Channel 6/Area Scan (61x351x1): Measurement grid: dx=10mm,
dy=10mm
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (interpolated) = 0.383 mW/g

2450 MHz - Lapheld/Chain A, 802.11n*, Channel 6/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 14.368 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 0.862 W/kg
SAR(1 g) = 0.334 mW/g; SAR(10 g) = 0.139 mW/g
Info: Interpolated medium parameters used for SAR evaluation.
Maximum value of SAR (measured) = 0.394 mW/g




                               0 dB = 0.390mW/g




Report Nº(NIE): 34184RRF.001                      Page 22 of 33                                     2011-08-24


                 APPENDIX D: Calibration Data




Report Nº(NIE): 34184RRF.001   Page 23 of 33    2011-08-24


                                                                                                                                            areQ@                     WirceicEss




               ga"brgt;‘a; Laboratnry of                                             3‘1‘:\53_;::?5                               Schweizerischer Kalibrlerd lerast
                 chmi              artner                                            ig           C                               Service suisse d‘étalonnage
                  Engineering AG                                                 T                                                Servizio savizzero di taratura
               Zeughausstrasse 43, 8004 Zurich, Switzorland                          *z                                           Swiss Calibration Service


                Accredited by the Swiss Accreditation Service (SAS)                                                Accreditation No.: SCS 108
                The Swiss Accreditation Service is one of the signatories to the EA
                Multilateral Agreement for the recognition of calibration certificates

                Client      ATAwireless                        y                                                   Certificate No: ES3—3052_Oct10

                CALIBRATION CERTIFICATE
                Object                                Essova=snso52

                Calibration procedure{s)             ‘GA CAL—01.v6, GA CAL—23.vand OA Cal—25v2\
                                                     «Calibration procedure for dosimetric E—field probes

                Calibration date                      October 22) 2010



                 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1}
                The measurements and the uncertainties with confidence probabiliy are given on the following pages and are part of the certificate


                 All calibrations have been conducted in the closed laboratory facilty. environment termperature 1(22 + 3)"C and humidity < 70%.


                 Calibration Equipment used (M&TE critical for calibration}


                 Primary Standards—                     ID #                         Cal Date (Certificate No.)                       ~Scheduled Calibration
                 Power mater E44198                    GB41203674                    1—Apr—10 (No, 217—01138)                          Apr—11
                 Power sensor E44124                   MY41455277                    1—Apr—10 (No. 217—011386)                         Apr—t 1
                 Power sensor E4412A                   MY41498087                    1—Apr—10 (No, 217—01136)                          Apr—41
                 Reference 3 d8 Attenuator             SN: 56054 (3¢)                30—Mar— 140 (No. 217—01159)                       Mar—11
                 Reference 20 dB Attenuator            SN: 55086 (206)               30—Mar—10 (No. 217—01161)                         Mar—11
                 Reference 30 dB Attenuator            SN: 56129 (30b1)              30—Mar—10 (No, 217—01 160)                        Mar— 1
               | Reference Probe ES3OVZ                SN; 3013                      30—Dec—09 (No ES3—3013 DecO9)                     Dec—10
                | DaE4                                 SN: 580                       20—Apr—10 (No. DAE4—660_Apr10)                    Apret1

                 Secondary Standards _                  ID# _                        Check Date (in house)                             Schaduled Check
                 RF generator HP 5648C                 US3E42L01700                  4—Aug—29 (in house check Oct—09)                  In house—check: Oct—11
                 Network Analyzer HP 8753E             US37390545                    18—0ct—01 (in house check Oct10)                  In house check: Oct—11           |

                                                       Name                                      Function                              Signature
                Calibrated by                          Jétan Kastrati                            LaboratoryTechnician                                  La/ _

                                                                                                                                       mt es                   5
                Approved by                            Kitha Pokevic                             Technical Maniager
                                                                                                                                       e—t2] 4@
                                                                                                                                       Issued: Ociober 23. 2010
                This calibration certificate shall not ‘he reproduced except in full without written approval of the laboratory


               Certificate No; ES2—3052_Oct10                                             Page 1 of 11




An~ se aos   (o temp t   t n n   e e > > —


                                                                                                                                                   areQ@                     WinciEss




                  ES3DV3 SN:3052                                                                                                                        October 22, 2010


                  DASY/EASY — Parameters of Probe: ES3DV3 SN:3052

                  Basic Calibration Parameters
                                                                                                                Sensor X Sensor Y Sensor Z |Unc (k=2)
                  Norm (uvi(¥/m}*)*                                                                                110          1123          1113       £101%
                  DCP (mv)"                                                                                        87.8         86.1          970



                  Modulation Calibration Parameters
                  UID             Communication System Name                                PAR                       A           B             C            VR           Unc®
                                                                                                                    dB          dBuV                        mM           (k=2)
                  10000           CW                                                       0.00|       x                 1.00        a.00          1.00| 300.0;         <+1.5%
                                                                                                       h                 0.00        0.00          1.00| 300.0
                                                                                                       2                 0.00        0.00          1.00| 300.0




                   The reported uncertainty of measurement is stated as the standard uncertainty of measurement multiplied
                   by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of
                   approximately 95%.


                  * The uncertainties of NonmX, Y Z do not affect the E—field uncertainty inside T SL (see Pages 5 and 5)
                  * Numericai lineanzation perameter, uncertainty sol requined
                   Uncertainty is determined using the maximum deviabon from linear response applying recatengular cislrbuben aind is expressed for the square of the tield value




                        Certificate No: ES3—3057_Oct10                                     Fage 4 of 11




sns pesia s n n   q r ny e   >                                                            — o ——


                                                                                                                               areQ@                         Winrceicss




ES3DV3 SN:3052                                                                                                                         October 22, 2010




DASY/EASY — Parameters of Probe: ES3DV3 SN:3052

Calibration Parameter Determined in Head Tissue Simulating Media


t [MHz]         Validity (MHz]"        Permittivity       Conductivity       ConvF x        _ConvF¥       ConvF Z             Alpha        Depth Unc (k=2) _

750             + 50/ + 100           43.5 + 5%           087 £5%                    §.25         6.25           6.25           0.88         1.02 +£11.0%
835             +50 /+ 100            41.5 + 8%           090 + 5%                   5.97          5.97          5.97           0 66         115. 4 11,0%
900             £50/ +100             A41:5 £5%           0.97 £5%                   5.82          5.42          5.82          055           1.32) $11.0%
1780            + 50/ + 100           40.1 + 5%            1.37 +5%                  5.16          5.16          5.16          0.41          1.534¥41:0%
1900            + 50/ + 100           40.0 + 5%            140 + 5%                  4.95         4.95           4.95          0.36          1.80 +11.0%
2000            + 50/ + 100           40.0 +5%             140 + 5%                  4.85         4.85           485           0.36          1.70 £11.0%
2450           +501 +100              38.2 + 5%            1.80 4 5%                 4.33         4.33           4,33          0.32         221 £11.0%
2600           +650 /+100             39.0+5%              188 +5%                   4.22         4.22           4.23          0.34         :06 £11.0%



\ The vabdity of + 100 MHz only apples tor DASY v4 4 and higher {see Page 2). The uncertainty is the RSS of ths ConyP uncertainty at callbration frequemcy

and the uncertainty for the indicated fecusncy band.




      Certficate No: ES3—30572_Oct10                                  Page 5 of 11


                                                                                                                                              areQ@                         WircicEsSsS




                ES3DV3 SN:3052                                                                                                                       October 22, 2010




                DASY/EASY — Parameters of Probe: ES3DV3 SN:3052

                Calibration Parameter Determined in Body Tissue Simulating Media


                f [MHz]             Validity [MH?.'C         Permittivity   Conductivity    ConvF X       ConvF¥         ConvF2Z             Alpha       Dapth Unc (k=2)

                750                 +50 / + 100              56.7 + 5%      0.94 + 5%              6.02           6.02          6.02          0.61         1.24 £11.0%

                835                 * 50 /# 100              5§5,2 4 5%     0.97 #5%               5.93           5.93          5.93          0.72         119 +110%

                900                 + 50/ +100               55.0 + 5%      1.05 £5%               5.81           5.81          5.81          0.63         1122 +11,0%

                1750                +50/+100                 53.4 +5%       149 + 5%               4.70           4.70          4.70          0.32         219 £11.0%

                1900                +850 / + 100             53.3 4 5%      1.82 £5%               4.51           4.51          4.51          0.34         2.26 £11.0%

                2000                + 50/ + 100              53.3 + 5%      1.52 + 5%              4.54           4.54          4.54          1.33         2.28 £11.0%

                2450                150/+ 100                52.7 + 5%      1.95 £5%               4.14           4.14          4.14          0.72         1.18— + 11.0%
                2600                4 50 /+100               52.5 + 54      216 £+5%               4.01           4.01          4.01          0.99         0.99° +11.0%




                * The valigity of + 100 MHz only apples for DASY v4 4 and higher (see Page 21 The uncerainly is the RSS of the ConwF uncendainty at celibration frequency
                and the uncertainty for the indlicaled frequency band.




                      Certificate No: ES3—3052_Oct10                                  Page § of 11




exeprva e a n   qA vakorg e   sn   n w n   ba se sn en e n                           — o —


                                                                                                                                         argQ@                  WirceicEss




                Calibration Laboratory of                                        fi@z                                       g      Schwelzerischer Kallbrierdienst
                Schmid & Partner                                                M                                          g Service suissed‘étalonnage
                   Engineering AG                                               T o mng                                      Servizio svizzero di tarature
                Zeughausstrasse 43,8004 Zurich, Switzerland                     ons   eX     o¥                            S swiss Calibration Service
                                                                                   "lcfipont®


                The Swiss Accreditation Service is one of the signatories to the EA
                Multitateral Agreement for the recognition of callbration certificates

                ol mm

                 Calibration procedure(s)




                 Caltrationdate:             aaboaaianrencienmerrn
                 Condition ot t caluratcsion t
                 This calibration cortificate documents the traceability to national standards, which realize the physical units of messuremeants (S1).
                 The measurements and the uncertainties with contidence probatility are given on the fallowing pages and are part of the certificate.


                .All calibrations have been conducted in the closed laboratory facility: environment temperature {22 4 3)°C and humidity <70%,

                 Calibration Equipment used (M&TE ertical tor calibration}

                 Primary Standards                       10 #                     Cal Date (Calibrated by, Certificate No.)           Scheduled Calibration
                 Power meter EPM—442A                    GB37480704               08—Oct—08 (No. 217—00898)                           Oct—O8
                 Power sensor HP 8481A                   us3720?783               08—O0ct—08 (No. 217—00898)                          Oct—09
                 Reference20 dB Attenuator               SN: 5086 (209)           31—Mar—09 (No, 217—01025)                           Mar10
                 Type—N mismaitch combination            SN: 5047.2 / 06327      ~31—Mar—09 (No. 217—01029)                           Mar10
                 Reference Probe ES3OV2                  SN: 30256               28—Apr—08 (No ES3—3025_ Apr08)                       Apr—O8
                 Rotorence Probe ES30v2                  SN: 3025                30—Apr—08 (No. ES3—3025_Apr00)                       Apr—10
                 DAE4                                    SN: 601                 O07—Mar—09 (No, DAE4—601_Mart9)                      Mar10

                 Secondary Standards                     ID #                     Check Date (in house)                               Schoduled Check
                 Power sansor HP 8491A                   Mya10g2317               18—0ct—02 (In haouse chack Oct—07)                  in house check: Oct—08
                 RF genarator A&S SMT—06                   100005                 4—Aug—99(in house check Oct—07)                     in house check: Oct—O§
                 Network Analyzer HP 8753E               US37300585 $4206         18—Oct—01 (in house check Oct—08)                   In house check: Oct—o9

                                                        Name                                      Function                            Signature
                 Calibrated by:                          fiko   Mall                              aborak        hC                     52C



                 Approved by:



                                                                                                                                     Issuad‘ June 19, 2008
                 This calibration certificate shall not be reproduced except in full without wntten approval of the laboratory.


                Certificate No: D2450V2—756_JunO§                                     Page 1 of 9




mneprsn n n n   q n rn y n   t   n n n   nee e n > >                              =ocg> —>         —— — >


                                                                                                                  areQ@                WirceicEss




             Measurement Conditions
                DASY system configuration, as tar as not given on page 1
                   DASY Version                                               DASYS                                   V5.0
                   Extrapolation                                      Advanced Extrapotation
                   Phantom                                          Modular Flat Phantom ¥5.0
                   Distance Dipole Center — TSL                                10 mm                               with Spacer
                   Zoom Scan Resolution                                  dx, dy, dz =5 mm
                   Frequency                                            2450 MHz + 1 MHz



             Head TSL parameters
                 The following parameters and calculations were applied.
                                                                              Temperature         Permittivity          Conductivity
                   Nominal Head TSL parameters                                    22.0 °C            39.2                1.80 mhaoim
                   Measured Head TSL parameters                               (22.0 + 0.2) °C     40.4 +6 %           1.78 mhaim * 6 %
                   Head TSL temperature during test                           (22.4 + 0.2) °C         fke



             SAR result with Head TSL

                   SAR averaged over 1 em" (1 g) of Head TSL                    Condition
                   SAR measured                                           250 mW input power                     13.3 mW /g
                   SAR normalized                                           normalized to 1 W                 53.2 mW /g
                   SAR for nominal Head TSL parameters ‘                    narmalized to 1W          53.9 mW /g + 17.0 % (k=2)

                   SAR averaged over 10 em"‘ (10 g) of Head TSL                 condition
                   SAR measured                                           250 mW input power                  6.29 mW /g
                   SAR normalized                                           normalized to 1W                  25.2 mWw /g
                   SAR for nominal Head TSL parameters ‘                    narmalized to 1W          25.3 mW ig > 16.5 % (k=2)




             ‘ Correction to nominal TSL parameters according to d), chapter "SAR Sensitmties"®

             Cenificate No: D2450V2—756_Jund®                          Page 3 of 9




An~ se aos
             At ty t   mt


Report Nº(NIE): 34184RRF.001   Page 30 of 33   2011-08-24