Sar report I

FCC ID: P4MAGT100D

Test Report

Download: PDF
FCCID_698536

                   SAR EVALUATION REPORT

                                           For

                          AnyData Corporation
                   18902 Bardeen Ave. Irvine, CA 92612-1522, U.S.A



                               FCC ID: P4MAGT100D



   This Report Concerns:                      Equipment Type:
     Original Report                          CDMA AGPS Tracker

    Test Engineer:       Eric Hong/


      Report No.:        R0607263S


     Report Date:        2006-08-07

     Reviewed By:        Lab Manager: Samuil Lisinker
                         Bay Area Compliance Laboratory Corporation
                         1274 Anvilwood Ave.
     Prepared By:        Sunnyvale, CA 94089
                         Tel: (408) 732-9162
                         Fax: (408) 732 9164




Note: This test report is specially limited to the above client company and this particular sample
      only. It may not be duplicated without prior written consent of Bay Area Compliance
      Laboratory Corporation. This report must not be used by the client to claim product
      certification, approval, or endorsement by NVLAP, NIST or any agency of the U.S.
      Government.


AnyData Corporation                                                                                                        FCC ID: P4MAGT100D



                  DECLARATION OF COMPLIANCE SAR EVALUATION
                      Rule Part(s):                              CFR47 §2.1093
                   Test Procedure(s):                  FCC OET Bulletin 65 Supplement C
                  Device Classification:                  Licensed Mobile Transmitter
                      Device Type:                 CDMA Position Reporting Device
                        FCC ID:                                  P4MAGT100D
                      Modulation:                                    CDMA
                 TX Frequency Range:                     824-849 MHz/1850-1910 MHz
              Max. Conducted Power Tested:         26.15dBm (835MHz) / 26.27dBm (1900MHz)
                    Antenna Type(s):                            Integral Antenna
                Body-Worn Accessories:                              Belt Clip
                 Face-Head Accessories:                               None
              SAR Value Measured: 0.821mW/g (835MHz Band) / 1.46mW/g (1900MHz Band)
BACL Corp. declares under its sole responsibility that this wireless portable device has been determined to be in compliance for localized specific absorption
rate (SAR) for uncontrolled exposure and general population exposure limits specified in EN 50361:2001 and has been tested in accordance with the
measurement procedures specified in ANSI IEEE C95.3:2002


All measurements reported herein were performed under my supervision and believed to be accurate to the best of my knowledge. I further attest for the
completeness of these measurements and vouch for the qualifications any and all personnel performing such measurements.

                            The results and statements contained in this report pertain only to the device(s) evaluated.


Eric Hong

/signature/

Bay Area Compliance Laboratory Corp.




Report#R0607263S                                                  Page 2 of 68                                                SAR Evaluation Report


AnyData Corporation                                                                                                                    FCC ID: P4MAGT100D

TABLE OF CONTENTS
REFERENCE, STANDARDS, AND GUILDELINES ...............................................................................................4
  SAR LIMITS .................................................................................................................................................................5
EUT DESCRIPTION ....................................................................................................................................................6
DESCRIPTION OF TEST SYSTEM ..........................................................................................................................7
  MEASUREMENT SYSTEM DIAGRAM .............................................................................................................................9
  SYSTEM COMPONENTS ...............................................................................................................................................11
TESTING EQUIPMENT............................................................................................................................................20
  EQUIPMENTS LIST & CALIBRATION INFO ...................................................................................................................20
SAR MEASUREMENT SYSTEM VERIFICATION ..............................................................................................21
  SYSTEM ACCURACY VERIFICATION ...........................................................................................................................21
EUT TEST STRATEGY AND METHODOLOGY .................................................................................................23
  SAR EVALUATION PROCEDURE .................................................................................................................................23
CONCLUSION ............................................................................................................................................................24
 SAR TEST DATA ........................................................................................................................................................24
APPENDIX A – MEASUREMENT UNCERTAINTY ............................................................................................25

APPENDIX B – PROBE CALIBRATION CERTIFICATES .................................................................................27
APPENDIX D - TEST SYSTEM VERIFICATIONS SCANS.................................................................................52
  LIQUID MEASUREMENT RESULT ................................................................................................................................52
APPENDIX E - EUT SCANS .....................................................................................................................................57
APPENDIX F – CONDUCTED OUTPUT POWER MEASUREMENT ...............................................................65
  PROVISION APPLICABLE.............................................................................................................................................65
  TEST PROCEDURE ......................................................................................................................................................65
  TEST EQUIPMENT .......................................................................................................................................................65
  TEST RESULTS ...........................................................................................................................................................65
APPENDIX G – EUT TEST POSITION PHOTOS .................................................................................................66
  BODY WORM SETUP PHOTO WITH BELT CLIP ...............................................................................................................66
APPENDIX H – EUT & ACCESSORIES PHOTOS ...............................................................................................67
  EUT - TOP VIEW ........................................................................................................................................................67
  EUT - BOTTOM VIEW ................................................................................................................................................67
APPENDIX I - INFORMATIVE REFERENCES....................................................................................................68




Report#R0607263S                                                         Page 3 of 68                                                     SAR Evaluation Report


AnyData Corporation                                                                  FCC ID: P4MAGT100D


REFERENCE, STANDARDS, AND GUILDELINES

FCC:
The Report and Order requires routine SAR evaluation prior to equipment authorization of portable
transmitter devices, including portable telephones. For consumer products, the applicable limit is 1.6
mw/g as recommended by the ANSI/IEEE standard C95.1-1992 [6] for an uncontrolled environment
(Paragraph 65). According to the Supplement C of OET Bulletin 65 “Evaluating Compliance with FCC
Guide-lines for Human Exposure to Radio frequency Electromagnetic Fields", released on Jun 29, 2001
by the FCC, the device should be evaluated at maximum output power (radiated from the antenna) under
“worst-case” conditions for normal or intended use, incorporating normal antenna operating positions,
device peak performance frequencies and positions for maximum RF energy coupling.

This report describes the methodology and results of experiments performed on wireless data terminal.
The objective was to determine if there is RF radiation and if radiation is found, what is the extent of
radiation with respect to safety limits. SAR (Specific Absorption Rate) is the measure of RF exposure
determined by the amount of RF energy absorbed by human body (or its parts) – to determine how the RF
energy couples to the body or head which is a primary health concern for body worn devices. The limit
below which the exposure to RF is considered safe by regulatory bodies in North America is 1.6 mw/g
average over 1 gram of tissue mass.

CE:
The order requires routine SAR evaluation prior to equipment authorization of portable transmitter
devices, including portable telephones. For consumer products, the applicable limit is 2 mw/g as
recommended by the EN50360 for an uncontrolled environment. According to the Standard, the device
should be evaluated at maximum output power (radiated from the antenna) under “worst-case” conditions
for normal or intended use, incorporating normal antenna operating positions, device peak performance
frequencies and positions for maximum RF energy coupling.

This report describes the methodology and results of experiments performed on wireless data terminal.
The objective was to determine if there is RF radiation and if radiation is found, what is the extent of
radiation with respect to safety limits. SAR (Specific Absorption Rate) is the measure of RF exposure
determined by the amount of RF energy absorbed by human body (or its parts) – to determine how the RF
energy couples to the body or head which is a primary health concern for body worn devices. The limit
below which the exposure to RF is considered safe by regulatory bodies in Europe is 2 mw/g average over
10 gram of tissue mass.

The test configurations were laid out on a specially designed test fixture to ensure the reproducibility of
measurements. Each configuration was scanned for SAR. Analysis of each scan was carried out to
characterize the above effects in the device.

There was no SAR of any concern measured on the device for any of the investigated configurations.




Report#R0607263S                              Page 4 of 68                             SAR Evaluation Report


AnyData Corporation                                                              FCC ID: P4MAGT100D

SAR Limits
FCC Limit (1g)

                                                                 SAR (W/kg)
       EXPOSURE LIMITS                   (General Population /                  (Occupational /
                                         Uncontrolled Exposure                Controlled Exposure
                                            Environment)                         Environment)
         Spatial Average
  (averaged over the whole body)                  0.08                                0.4

            Spatial Peak
(averaged over any 1 g of tissue)                 1.60                                8.0

            Spatial Peak
      (hands/wrists/feet/ankles                    4.0                               20.0
        averaged over 10 g)

CE Limit (10g)

                                                                 SAR (W/kg)
       EXPOSURE LIMITS                   (General Population /                  (Occupational /
                                         Uncontrolled Exposure                Controlled Exposure
                                            Environment)                         Environment)
         Spatial Average
  (averaged over the whole body)                  0.08                                0.4

            Spatial Peak
(averaged over any 1 g of tissue)                  2.0                                10

            Spatial Peak
      (hands/wrists/feet/ankles                    4.0                               20.0
        averaged over 10 g)


Population/Uncontrolled Environments are defined as locations where there is the exposure of individual
who have no knowledge or control of their exposure.

Occupational/Controlled Environments are defined as locations where there is exposure that may be
incurred by people who are aware of the potential for exposure (i.e. as a result of employment or
occupation).

Population/uncontrolled environments Spatial Peak limit 1.6w/kg (FCC) & 2.0w/kg (CE) applied to the
EUT.




Report#R0607263S                            Page 5 of 68                           SAR Evaluation Report


AnyData Corporation                                                                     FCC ID: P4MAGT100D


EUT DESCRIPTION
The AnyDATA Corporation.’s product, FCC ID: P4MAGT100D or the “EUT" as referred to in this report
is a dual-band device that operates on both Code Division Multiple Access (CDMA) frequencies: cellular
services at 800 MHz, and Personal Communication Services (PCS) at 1.9 GHz. Also features soft/softer
handoff, hard handoff, and dynamic RF power control technologies to reduce call interruptions.

Approximate measurement is: 81mmL x 47 mmW x 20.5mmH

  * The test data gathered are from typical production sample, serial number: 001, provided by the manufacturer.




Report#R0607263S                                Page 6 of 68                               SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D


DESCRIPTION OF TEST SYSTEM
These measurements were performed with the automated near-field scanning system DASY4 from
Schmid & Partner Engineering AG (SPEAG) which is the fourth generation of the system shown in the
figure hereinafter:




The system is based on a high precision robot (working range greater than 0.9m), which positions the
probes with a positional repeatability of better than ±0.02mm. Special E- and H-field probes have been
developed for measurements close to material discontinuity, the sensors of which are directly loaded with
a Schottky diode and connected via highly resistive lines to the data acquisition unit.
The SAR measurements were conducted with the dosimetric probe ET3DV6 SN: 1604 (manufactured by
SPEAG), designed in the classical triangular configuration and optimized for dosimetric evaluation. The
probe has been calibrated according to the procedure with accuracy of better than ±10%. The spherical
isotropy was evaluated with the procedure and found to be better than ±0.25dB.




Report#R0607263S                            Page 7 of 68                            SAR Evaluation Report


AnyData Corporation                                                                                     FCC ID: P4MAGT100D

The phantom used was the Generic Twin Phantom”. The ear was simulated as a spacer of 4 mm thickness
between the earpiece of the phone and the tissue simulating liquid. The Tissue simulation liquid used for
each test is in according with the FCC OET65 supplement C as listed below.

   Ingredients                                                    Frequency (MHz)
  (% by weight)              450                  835                     915                         1900                 2450
   Tissue Type        Head         Body    Head     Body          Head          Body          Head       Body       Head      Body
      Water           38.56        51.16   41.45        52.4      41.05         56.0           54.9          40.4    62.7         73.2
    Salt (Nacl)       3.95         1.49    1.45         1.4       1.35          0.76           0.18          0.5     0.5          0.04
      Sugar           56.32        46.78   56.0         45.0      56.5          41.76          0.0           58.0    0.0          0.0
       HEC            0.98         0.52     1.0         1.0        1.0          1.21           0.0           1.0     0.0          0.0
   Bactericide        0.19         0.05     0.1         0.1        0.1          0.27           0.0           0.1     0.0          0.0
   Triton x-100        0.0          0.0     0.0         0.0        0.0           0.0           0.0           0.0     36.8         0.0
      DGBE             0.0          0.0     0.0         0.0        0.0           0.0          44.92          0.0     0.0          26.7
Dielectric Constant   43.42        58.0    42.54        56.1      42.0          56.8           39.9          54.0    39.8         52.5
Conductivity (s/m)    0.85         0.83    0.91         0.95       1.0          1.07           1.42          1.45    1.88         1.78


     IEEE SCC-34/SC-2 P1528 Recommended Tissue Dielectric Parameters

    Frequency                              Head                                                         Body
      (MHz)                        εr                    Ơ (S/m)                         εr                         Ơ (S/m)
        150                    52.3                        0.76                         61.9                         0.80
        300                    45.3                        0.87                         58.2                         0.92
        450                    43.5                        0.87                         56.7                         0.94
        835                    41.5                        0.90                         55.2                         0.97
        900                    41.5                        0.97                         55.0                         1.05
        915                    41.5                        0.98                         55.0                         1.06
       1450                    40.5                        1.20                         54.0                         1.30
       1610                    40.3                        1.29                         53.8                         1.40
    1800-2000                  40.0                        1.40                         53.3                         1.52
       2450                    39.2                        1.80                         52.7                         1.95
       3000                    38.5                        2.40                         52.0                         2.73
       5800                    35.3                        5.27                         48.2                         6.00




Report#R0607263S                                   Page 8 of 68                                              SAR Evaluation Report


AnyData Corporation                                                                  FCC ID: P4MAGT100D

Measurement System Diagram




The DASY4 system for performing compliance tests consists of the following items:

• A standard high precision 6-axis robot (St¨aubli RX family) with controller, teach pendant and
  software. An arm extension for accommodating the data acquisition electronics (DAE).

• A dosimetric probe, i.e., an isotropic E-field probe optimized and calibrated for usage in tissue
  simulating liquid. The probe is equipped with an optical surface detector system.

• A data acquisition electronics (DAE) which performs the signal amplification, signal multiplexing,
  AD-conversion, offset measurements, mechanical surface detection, collision detection, etc. The unit is
  battery powered with standard or rechargeable batteries. The signal is optically transmitted to the EOC.

• The Electro-optical converter (EOC) performs the conversion between optical and electrical of the
  signals for the digital communication to the DAE and for the analog signal from the optical surface
  detection. The EOC is connected to the measurement server.

• The function of the measurement server is to perform the time critical tasks such as signal filtering,
  control of the robot operation and fast movement interrupts.

• A probe alignment unit which improves the (absolute) accuracy of the probe positioning.

• A computer operating Windows 2000 or Windows XP.

Report#R0607263S                              Page 9 of 68                             SAR Evaluation Report


AnyData Corporation                                                                   FCC ID: P4MAGT100D

• DASY4 software.

• Remote control with teach pendant and additional circuitry for robot safety such as warning lamps, etc.

• The SAM twin phantom enabling testing left-hand and right-hand usage.

• The device holder for handheld mobile phones.

• Tissue simulating liquid mixed according to the given recipes.

• Validation dipole kits allowing to validate the proper functioning of the system.




Report#R0607263S                             Page 10 of 68                             SAR Evaluation Report


AnyData Corporation                                                                FCC ID: P4MAGT100D

System Components
• DASY4 Measurement Server
• Data Acquisition Electronics
• Probes
• Light Beam Unit
• Medium
• SAM Twin Phantom
• Device Holder for SAM Twin Phantom
• System Validation Kits
• Robot


DASY4 Measurement Server


The DASY4 measurement server is based on a
PC/104 CPU board with a 166MHz low-power
pentium, 32MB chip disk and 64MB RAM. The
necessary circuits for communication with either
the DAE4 (or DAE3) electronic box as well as
the 16-bit AD-converter system for optical detection
and digital I/O interface are contained on the
DASY4 I/O-board, which is directly connected to
the PC/104 bus of the CPU board.

The measurement server performs all real-time data evaluation for field measurements and surface
detection, controls robot movements and handles safety operation. The PC-operating system cannot
interfere with these time critical processes. All connections are supervised by a watchdog, and
disconnection of any of the cables to the measurement server will automatically disarm the robot and
disable all program-controlled robot movements. Furthermore, the measurement server is equipped with
two expansion slots which are reserved for future applications. Please note that the expansion slots do not
have a standardized pinout and therefore only the expansion cards provided by SPEAG can be inserted.
Expansion cards from any other supplier could seriously damage the measurement server.




Report#R0607263S                             Page 11 of 68                           SAR Evaluation Report


AnyData Corporation                                                                  FCC ID: P4MAGT100D

Data Acquisition Electronics

The data acquisition electronics DAE3 consists of
a highly sensitive electrometergrade preamplifier with
auto-zeroing, a channel and gain-switching multiplexer,
a fast 16 bit AD-converter and a command decoder and
control logic unit. Transmission to the measurement
server is accomplished through an optical downlink for
data and status information as well as an optical uplink
for commands and the clock.




Probes
The DASY system can support many different probe types.

Dosimetric Probes: These probes are specially designed and calibrated for use in liquids with high
permittivity. They should not be used in air, since the spherical isotropy in air is poor (±2 dB). The
dosimetric probes have special calibrations in various liquids at different frequencies.

Free Space Probes: These are electric and magnetic field probes specially designed for measurements
in free space. The z-sensor is aligned to the probe axis and the rotation angle of the x-sensor is specified.
This allows the DASY system to automatically align the probe to the measurement grid for field
component measurement. The free space probes are generally not calibrated in liquid. (The H-field probes
can be used in liquids without any change of parameters.)

Temperature Probes: Small and sensitive temperature probes for general use. They use a
completely different parameter set and different evaluation procedures. Temperature rise features
allow direct SAR evaluations with these probes.


ET3DV6 Probe Specification

Construction Symmetrical design with triangular core
Built-in optical fiber for surface detection System
Built-in shielding against static charges
Calibration In air from 10 MHz to 2.5 GHz
In brain and muscle simulating tissue at
Frequencies of 450 MHz, 900 MHz and
1.8 GHz (accuracy ± 8%)
Frequency 10 MHz to > 6 GHz; Linearity: ± 0.2 dB
(30 MHz to 3 GHz)
Directivity ± 0.2 dB in brain tissue (rotation around
probe axis)
± 0.4 dB in brain tissue (rotation normal probe axis)
Dynamic 5 mW/g to > 100 mW/g;
Range Linearity: ± 0.2 dB
Surface ± 0.2 mm repeatability in air and clear liquids                            Photograph of the probe
Detection over diffuse reflecting surfaces.
Dimensions Overall length: 330 mm
Tip length: 16 mm

Report#R0607263S                              Page 12 of 68                            SAR Evaluation Report


AnyData Corporation                                                                FCC ID: P4MAGT100D

Body diameter: 12 mm
Tip diameter: 6.8 mm
Distance from probe tip to dipole centers: 2.7 mm
Application General dosimetric up to 3 GHz
Compliance tests of mobile phones
Fast automatic scanning in arbitrary phantoms

The SAR measurements were conducted with the dosimetric probe
ET3DV6 designed in the classical triangular configuration and
optimized for dosimetric evaluation. The probe is constructed using
the thick film technique; with printed resistive lines on ceramic
substrates. The probe is equipped with an optical multi-fiber line
ending at the front of the probe tip. It is connected to the EOC box on
the robot arm and provides an automatic detection of the phantom
surface. Half of the fibers are connected to a pulsed infrared
transmitter, the other half to a synchronized receiver. As the probe
approaches the surface, the reflection from the surface produces a
coupling from the transmitting to the receiving fibers. This reflection
increases first during the approach, reaches maximum and then
decreases. If the probe is flatly touching the surface, the coupling is
zero. The distance of the coupling maximum to the surface is
independent of the surface reflectivity and largely independent of the
surface to probe angle. The DASY3 software reads the reflection                    Inside view of
during a software approach and looks for the maximum using a 2nd                ET3DV6 E-field Probe
order fitting. The approach is stopped when reaching the maximum.


E-Field Probe Calibration Process

Each probe is calibrated according to a dosimetric assessment procedure described in [6] with accuracy
better than +/- 10%. The spherical isotropy was evaluated with the procedure described in [7] and found
to be better than +/-0.25dB. The sensitivity parameters (NormX, NormY, NormZ), the diode compression
parameter (DCP) and the conversion factor (ConvF) of the probe are tested.

The free space E-field from amplified probe outputs is determined in a test chamber. This is performed in
a TEM cell for frequencies bellow 1 GHz, and in a waveguide above 1 GHz for free space. For the free
space calibration, the probe is placed in the volumetric center of the cavity and at the proper orientation
with the field. The probe is then rotated 360 degrees.

E-field temperature correlation calibration is performed in a flat phantom filled with the appropriate
simulated brain tissue. The measured free space E-field in the medium correlates to temperature rise in
dielectric medium. For temperature correlation calibration a RF transparent thermistor-based temperature
probe is used in conjunction with the E-field probe.

Data Evaluation

The DASY4 postprocessing software (SEMCAD) automatically executes the following procedures to
calculate the field units from the microvolt readings at the probe connector. The parameters used in the
evaluation are stored in the configuration modules of the software:

Probe parameters: - Sensitivity                  Normi, ai0, ai1, ai2
                  - Conversion factor            ConvFi
                  - Diode compression point      dcpi


Report#R0607263S                             Page 13 of 68                           SAR Evaluation Report


AnyData Corporation                                                                 FCC ID: P4MAGT100D

Device parameters: - Frequency                   f
                   - Crest factor                cf

Media parameters: - Conductivity                 σ
                  - Density                      ρ


These parameters must be set correctly in the software. They can be found in the component
documents or they can be imported into the software from the configuration files issued for the
DASY components. In the direct measuring mode of the multimeter option, the parameters of the
actual system setup are used. In the scan visualization and export modes, the parameters stored
in the corresponding document files are used.
The first step of the evaluation is a linearization of the filtered input signal to account for the
compression characteristics of the detector diode. The compensation depends on the input signal,
the diode type and the DC-transmission factor from the diode to the evaluation electronics. If the
exciting field is pulsed, the crest factor of the signal must be known to correctly compensate for
peak power. The formula for each channel can be given as:




With Vi    = compensated signal of channel i (i =x, y, z)
     Ui    = input signal of channel i (i =x, y, z)
     cf    = crest factor of exciting field (DASY parameter)
    dcpi   = diode compression point (DASY parameter)

From the compensated input signals the primary field data for each channel can be evaluated:




With   Vi    = compensated signal of channel i (i =x, y, z)
       Normi = sensor sensitivity of channel i (i =x, y, z)
             µV/ (V/m)2 for E-field probes
       ConF = sensitivity enhancement in solution
       aij   = sensor sensitivity factors for H-field probes
        f    = carrier frequency [GHz]
       Ei    = electric field strenggy of channel i in V/m
       Hi    = diode compression point (DASY parameter)

The RSS value of the field components gives the total field strength (Hermitian magnitude):




The primary field data are used to calculate the derived field units.




Report#R0607263S                              Page 14 of 68                           SAR Evaluation Report


AnyData Corporation                                                                   FCC ID: P4MAGT100D

With    SAR = local specific absorption rate in mW/g
        Etot = total field strength in V/m
        σ    = conductivity in [mho/m] or [Siemens/m]
        ρ    = equivalent tissue density in g/cm3

Note that the density is normally set to 1, to account for actual brain density rather than the density of the
simulation liquid.




Report#R0607263S                               Page 15 of 68                            SAR Evaluation Report


AnyData Corporation                                                                   FCC ID: P4MAGT100D

Light Beam Unit

The light beam switch allows automatic ”tooling” of the probe. During the process, the actual
position of the probe tip with respect to the robot arm is measured, as well as the probe length and
the horizontal probe offset. The software then corrects all movements, so that the robot coordinates
are valid for the probe tip. The repeatability of this process is better than 0.1 mm. If a position
has been taught with an aligned probe, the same position will be reached with another aligned
probe within 0.1 mm, even if the other probe has different dimensions. During probe rotations, the
probe tip will keep its actual position.



Medium

Parameters

The parameters of the tissue simulating liquid strongly influence the SAR in the liquid The parameters
for the different frequencies are defined in the corresponding compliance standards (e.g., EN 50361, IEEE
1528-2003).

Parameter measurements

Several measurement systems are available for measuring the dielectric parameters of liquids:

• The open coax test method (e.g., HP85070 dielectric probe kit) is easy to use, but has only moderate
  accuracy. It is calibrated with open, short, and deionized water and the calibrations a critical process.

• The transmission line method (e.g., model 1500T from DAMASKOS, INC.) measures the transmission
  and reflection in a liquid filled high precision line. It needs standard two port calibration and is probably
  more accurate than the open coax method.

• The reflection line method measures the reflection in a liquid filled shorted precision lined.The method
  is not suitable for these liquids because of its low sensitivity.

• The slotted line method scans the field magnitude and phase along a liquid filled line. The evaluation is
  straight forward and only needs a simple response calibration. The method is very accurate, but can
  only be used in high loss liquids and at frequencies above 100 to 200MHz. Cleaning the line can be
  tedious.




Report#R0607263S                               Page 16 of 68                            SAR Evaluation Report


AnyData Corporation                                                                  FCC ID: P4MAGT100D

SAM Twin Phantom

The SAM twin phantom is a fiberglass shell phantom with
2mm shell thickness (except the ear region where shell
thickness increases to 6mm). It has three measurement
areas:

• Left hand
• Right hand
• Flat phantom

The phantom table comes in two sizes: A 100 x 50 x 85 cm
(L x W x H) table for use with free standing robots (DASY4
 professional system option) or as a second phantom and a
100 x 75 x 85 cm(L x W x H) table with reinforcements for
table mounted robots (DASY4 compact system option) .
The bottom plate contains three pair of bolts for locking the device holder. The device holder positions
are adjusted to the standard measurement positions in the three sections. Only one device holder is
necessary if two phantoms are used (e.g., for different liquids) A white cover is provided to tap the
phantom during o_-periods to prevent water evaporation and changes in the liquid parameters. Free space
scans of devices on the cover are possible. On the phantom top, three reference markers are provided to
identify the phantom position with respect to the robot.

The phantom can be used with the following tissue simulating liquids:

• Water-sugar based liquids can be left permanently in the phantom. Always cover the liquid if the system
  is not used, otherwise the parameters will change due to water evaporation.

• Glycol based liquids should be used with care. As glycol is a softener for most plastics, the liquid should
  be taken out of the phantom and the phantom should be dried when the system is not used (desirable at
  least once a week).

• Do not use other organic solvents without previously testing the phantom resistiveness.




Report#R0607263S                              Page 17 of 68                           SAR Evaluation Report


AnyData Corporation                                                                  FCC ID: P4MAGT100D

Device Holder for SAM Twin Phantom

The SAR in the phantom is approximately inversely proportional to the square of the distance between the
source and the liquid surface. For a source in 5mm distance, a positioning uncertainty of ±0.5mm would
produce a SAR uncertainty of ±20%. An accurate device positioning is therefore crucial for accurate and
repeatable measurements. The positions, in which the devices must be measured, are defined by the
standards.
The DASY device holder is designed to cope with different positions given in the standard. It has two
scales for the device rotation (with respect to the body axis) and the device inclination (with respect to the
line between the ear reference points). The rotation centers for both scales is the ear reference point ERP).
Thus the device needs no repositioning when changing the angles.




The DASY device holder has been made out of low-loss POM material having the following dielectric
parameters: relative permittivity "=3 and loss tangent _=0.02. The amount of dielectric material has been
reduced in the closest vicinity of the device, since measurements have suggested that the influence of the
clamp on the test results could thus be lowered.




Report#R0607263S                              Page 18 of 68                            SAR Evaluation Report


AnyData Corporation                                                                FCC ID: P4MAGT100D

System Validation Kits

Each DASY system is equipped with one or more system validation kits. These units, together with
the predefined measurement procedures within the DASY software, enable the user to conduct the
system performance check and system validation. For that purpose a well defined SAR distribution
in the flat section of the SAM twin phantom is produced.

System validation kit includes a dipole, tripod holder to fix it underneath the flat phantom and a
corresponding distance holder. Dipoles are available for the variety of frequencies between 300MHz
and 6 GHz (dipoles for other frequencies or media and other calibration conditions are available upon
request).

The dipoles are highly symmetric and matched at the center frequency for the specified liquid and
distance to the flat phantom (or flat section of the SAM-twin phantom). The accurate distance between
the liquid surface and the dipole center is achieved with a distance holder that snaps on the dipole.



Robot

The DASY4 system uses the high precision industrial robots RX60L, RX90 and RX90L, as well as
the RX60BL and RX90BL types out of the newer series from St¨aubli SA (France). The RX robot
series offers many features that are important for our application:

• High precision (repeatability 0.02mm)
• High reliability (industrial design)
• Low maintenance costs (virtually maintenance-free due to direct drive gears; no belt drives)
• Jerk-free straight movements (brushless synchron motors; no stepper motors)
• Low ELF interference (the closed metallic construction shields against motor control fields)

For the newly delivered DASY4 systems as well as for the older DASY3 systems delivered since 1999,
the CS7MB robot controller version from St¨aubli is used. Previously delivered systems have either a
CS7 or CS7M controller; the differences to the CS7MB are mainly in the hardware, but some procedures
in the robot software from St¨aubli are also not completely the same. The following descriptions about
robot hard- and software correspond to CS7MB controller with software version 13.1 (edit S5). The
actual commands, procedures and configurations, also including details in hardware, might differ if an
older robot controller is in use. In this case please also refer to the St¨aubli manuals for further
information.




Report#R0607263S                             Page 19 of 68                           SAR Evaluation Report


AnyData Corporation                                                  FCC ID: P4MAGT100D


TESTING EQUIPMENT

Equipments List & Calibration Info

                      Type / Model                       Cal. Date          S/N:
 DASY4 Professional Dosimetric System                      N/A              N/A
 Robot RX60L                                               N/A        CS7MBSP / 467
 Robot Controller                                          N/A        F01/5J72A1/A/01
 Dell Computer Dimension 3000                              N/A              N/A
 SPEAG EDC3                                                N/A              N/A
 SPEAG DAE3                                             2005-10-18           456
 DASY4 Measurement Server                                  N/A              1176
 SPEAG E-Field Probe ET3DV6                             2006-05-02          1604
 Antenna, Dipole, D900V2                                2006-06-16           122
 Antenna, Dipole, D1800-S-1                             2005-03-04        BCL-049
 SPEAG Generic Twin Phantom                                N/A              N/A
 SPEAG Light Alignment Sensor                              N/A               278
 Brain Equivalent Matter (1900MHz)                       Each Use           N/A
 Muscle Equivalent Matter (1900MHz)                      Each Use           N/A
 Brain Equivalent Matter (835MHz)                        Each Use           N/A
 Muscle Equivalent Matter (835MHz)                       Each Use           N/A
 Robot Table                                             Each Use           N/A
 Phone Holder                                            Each Use           N/A
 Phantom Cover                                           Each Use           N/A
 HP Spectrum Analyzer HP8566A                              N/A          2240A01930
 Microwave Amp. 8349A                                      N/A          2644A02662
 Power Meter Agilent E4419B                             2005-08-31      MY4121511
 Power Sensor Agilent E4412A                            2005-09-08      US38488542
 Network Analyzer HP-8752C                              2005-02-22         820079
 Dielectric Probe Kit HP85070A                           Each Use       US99360201
 Signal Generator HP-83650B                             2005-05-10     3614A002716
 Amplifier, ST181-20                                       N/R           E012-0101




Report#R0607263S                        Page 20 of 68                 SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D


SAR MEASUREMENT SYSTEM VERIFICATION
System Accuracy Verification

Prior to the assessment, the system validation kit was used to test whether the system was operating
within its specifications of ±10%. The validation results are tabulated below. And also the corresponding
SAR plot is attached as well in the SAR plots files.

IEEE P1528 recommended reference value for head

 Frequency                                     Local SAR at surface           Local SAR at surface
                 1 g SAR       10 g SAR
   (MHz)                                        (above feed point)        (v=2cm offset from feed point)
     300            3.0            2.0                 4.4                             2.1
     450            4.9            3.3                 7.2                             3.2
     835            9.5            6.2                14.1                             4.9
     900           10.8            6.9                16.4                             5.4
    1450           29.0           16.0                50.2                             6.5
    1800           38.1           19.8                69.5                             6.8
    1900           39.7           20.5                72.1                             6.6
    2000           41.1           21.1                74.6                             6.5
    2450           52.4           24.0                104.2                            7.7
    3000           63.8           25.7                140.2                            9.5


Validation Dipole SAR Reference Test Result for Body (835 MHz)

 Validation
                   SAR @ 1W Input averaged over 1g               SAR @ 1W Input averaged over 10g
Measurement
   Test 1                       7.71                                           4.91
   Test 2                       7.98                                           5.08
   Test 3                        8.0                                           5.09
   Test 4                       7.98                                           5.06
   Test 5                       7.84                                            5.0
   Test 6                       7.61                                           4.85
   Test 7                       7.89                                           5.02
   Test 8                        7.9                                           5.05
   Test 9                       7.92                                           5.07
  Test 10                       7.99                                           5.09
  Average                       7.873                                          5.022




Report#R0607263S                            Page 21 of 68                           SAR Evaluation Report


AnyData Corporation                                                           FCC ID: P4MAGT100D

Validation Dipole SAR Reference Test Result for Body (1900 MHz)

              SAR @ 0.126W       SAR @ 1W          SAR @ 0.126W       SAR @ 1W
 Validation
               Input averaged   Input averaged      Input averaged   Input averaged
Measurement
                  over 1g          over 1g             over 10g         over 10g
   Test 1            3.1            24.61                1.42            11.27
   Test 2            3.1            24.61                1.41            11.20
   Test 3            3.2            25.41                1.43            11.35
   Test 4            3.2            25.41                1.42            11.27
   Test 5            3.1            24.61                1.42            11.27
   Test 6            3.2            25.61                1.41            11.20
   Test 7            3.2            25.61                1.43            11.35
   Test 8            3.1            24.61                1.42            11.27
   Test 9            3.1            24.61                1.42            11.27
  Test 10            3.1            24.61                1.43            11.35
  Average           3.14            24.97               1.421            11.28




Report#R0607263S                           Page 22 of 68                        SAR Evaluation Report


AnyData Corporation                                                               FCC ID: P4MAGT100D


EUT TEST STRATEGY AND METHODOLOGY
SAR Evaluation Procedure
The evaluation was performed with the following procedure:

Step 1: Measurement of the SAR value at a fixed location above the ear point or central position was
used as a reference value for assessing the power drop.

Step 2: The SAR distribution at the exposed side of the head was measured at a distance of 3.9 mm from
the inner surface of the shell. The area covered the entire dimension of the head or EUT and the
horizontal grid spacing was 20 mm x 20 mm. Based on these data, the area of the maximum absorption
was determined by spline interpolation.

Step 3: Around this point, a volume of 32 mm x 32 mm x 34 mm was assessed by measuring 5 x 5 x 7
points. On the basis of this data set, the spatial peak SAR value was evaluated under the following
procedure:

    1. The data at the surface were extrapolated, since the center of the dipoles is 2.7 mm away from the
       tip of the probe and the distance between the surface and the lowest measuring point is 1.2 mm.
       The extrapolation was based on a least square algorithm [11]. A polynomial of the fourth order
       was calculated through the points in z-axes. This polynomial was then used to evaluate the points
       between the surface and the probe tip.

    2. The maximum interpolated value was searched with a straightforward algorithm. Around this
       maximum the SAR values averaged over the spatial volumes (1 g or 10 g) were computed by the
       3D-Spline interpolation algorithm. The 3D-Spline is composed of three onedimensional splines
       with the “Not a knot"-condition (in x, y and z-directions) [11], [12]. The volume was integrated
       with the trapezoidal-algorithm. One thousand points (10 x 10 x 10) were interpolated to calculate
       the average.

    3. All neighboring volumes were evaluated until no neighboring volume with a higher average
       value was found.

Step 4: Re-measurement of the SAR value at the same location as in Step 1. If the value changed by more
than 5%, the evaluation was repeated.




Report#R0607263S                            Page 23 of 68                           SAR Evaluation Report


AnyData Corporation                                                                 FCC ID: P4MAGT100D

CONCLUSION
This page summarizes the results of the performed dosimetric evaluation. The plots with the
corresponding SAR distributions, which reveal information about the location of the maximum SAR with
respect to the device, could be found in Appendix E.



SAR Test Data
     Environmental Conditions

   Ambient Temperature:     25° C
   Relative Humidity:       77%
   ATM Pressure:            1026 mbar

* Testing was performed by Eric Hong on 2006-08-07.




                                                       Output              Measured      Limit
             EUT position           Freq (MHz)                   Phantom                       Plot #
                                                     Power (dBm)            (mw/g)      (mw/g)

      BODY WORN WITH BELT
                                         824.7           26.15    Flat      0.380        1.6       1
             CLIP
      BODY WORN WITH BELT
                                        836.52           26.03    Flat      0.821        1.6       2
             CLIP
      BODY WORN WITH BELT
             CLIP
                                         848.3           25.97    Flat      0.302        1.6       3
      BODY WORN WITH BELT
                                        1851.25          26.24    Flat      0.943        1.6       4
             CLIP
      BODY WORN WITH BELT
                                         1880            26.14    Flat       1.46        1.6       5
             CLIP
      BODY WORN WITH BELT
             CLIP
                                        1908.75          26.27    Flat       1.31        1.6       6




Report#R0607263S                                  Page 24 of 68                       SAR Evaluation Report


AnyData Corporation                                                           FCC ID: P4MAGT100D

APPENDIX A – MEASUREMENT UNCERTAINTY

The uncertainty budget has been determined for the DASY4 measurement system and is given in the
following Table.




Report#R0607263S                          Page 25 of 68                        SAR Evaluation Report


                                DASY4 Uncertainty Budget
                                According to CENELECEN 50361 [2]
                       Uncotamy            Prob:   Div.   (a) (a | Sid: Th:   Sid Ure:   (n)
 Eizor Deseription     value               Dist           is 10« (19)         (10))      tagy
 Measurement Equipment
 Probe Calibration                         x       7 To                       oX         &
  Tidal Totropy                            r       5 oz |or                   s19%       <
 Spherical otropy                          n       3 oz for                       '      <
 Probe Emc                                 n       OO   Cp                               .
 Detection Ennit                           n       OO                                    <
 Domday Eies                               n       OO                                    .
 Teadon: Electromes                        x       1 C                                   ~
 Teponse Tim                               x       1 C                                   &
 Nor                                       x       1 C                                   ~
 Tntcsmation Thm                           x       1 C                                   &
 Mechanical Constramis
 Scamig System         sou%                n       <Ou          c                        s
 Phantom Shell                   a10%      r       §      |1    1                        <
"Probe Postioune                 w2o%      n       o0                                    ~
 Dovice Postome                  sra%      x       1 To                                  Ts
 Physical Parameters
 Liguid Conuetaiy zy             5500      n       C      oz    us                       ~
 Liquid Conduetivity (mons)                n       3      oz    fos                      <
 Tiqud Ponmtimty (reey)                    n       i      (os   fus                      &
 Liquid Pormittivity: (mens.)              r       Vi     oo    |os                      <
 Pover Drm                                 n       OO  C                                 <
 NP Ambiont Conditions                     n       OuCto|5                               .
 Post—Processma
 Exrap, and Intesrmtion    s1o%            n       @OI          t     0s      zns        —
 Combmal SUT. Uncertam                                                Eon     + mal      o
 Expanded Sid, Unceriainty                                            smy     ce


Calibration Laboratory of                                 «C                                    Schsizerischer Kailoerdionat
Schmid & Partner                                          innmat                                Service suiase cétalonnage
  Engineering AG                                                                                Servizio sviezero dl taratrn
ZReughausstrasse 43, 8004 Zorich, Stzariand                                                     Suiss Caltbation Service



Accrestnd by ns Swiss Federl Ofice of Metrolcgy and Accredtaton                 Accredtation No: SCS 108
‘The Swiss Accrediation Service is one of the signatoris tothe EA
Muittateral Agreement for the recogrition of calbration certficates




 ‘is caltraton certfcat documents he racebity o natonal standarts, wtich realze the physica unts of measuremonts (S1)
 ‘he measurements ans he unceniinies win confidence probabit aregien on thefolowing pages and ae parof e ceriicats

 Allcaltrations rave Seen conducted in the clsed laborator facl: envirnment fenperstne (22 3]°C and hy < 70%.
 Caiteaton Eqiment used (METE crticalfor caliraton)

 Prmacy Standarts                  e                      Cal Date (Caltrtes by. Conticae No)        senedul Caltraton          I
 PovermaterEsa1s8                  conzsere               005 metas, No.251—0os57)                   Apear                     —
 Poversersor E44T2A                mveresserr             4008 guetas, No. 25—00557)                 Apear
 Posersersor E4412A                mvetasensr             sapros wetas, no. 2s—00s57)                Apear                      |
 Reterence 3 68 Atercster          sn ssose ae)           11—209.08 metas, ho. 251—00100)            Aoos
 Rateunce 20 8 Atensator           sc ssoss 2oe)          «Apr0s metas, no. 25t—005se)               Apror                      |
 Reterence 20 8 Atarsator          sh: serze oo           11—809.08 metas, ho. 251—00500)            Aupos
 Reference Prote ESsOV2            sn aors                2Jan.96 (GPEAG. No. E832012lnce)           Jnar
 caee                              sicese                 25e008 (SPEAG, No act—ase_Febos)           revar
 Secondary Sancarts                EB                     Check Date n hause)                        sereduied Check
 RF gereraor no seiec              usseausrco             +409 (GPGAG,in hovae check 0%)             inhouse chock NovaT            |
 Network Anaiyzer P 8753E          ussrasoses             18—09t01 (SFEAG,inhouse check Nov08)       Inhouse check: NovO6       |


 Caierai by:




Cericate No: E73—1604_Ma,06                                  Page 1 o9


Calibration Laboratory of                                                         Sctwaizerischer Kaiiordianst
Schmid & Partner                                                                  Service suiase o‘ttalonnage
  Engineering AG                                                                  Serviio avizzero oi tarnturn
Zeughausstrasse 43 8004 Zurich, Switzerind                                        Swiss Galbration Service

Accredtas by ho Sniss Federal Offce of Metology and Accredtatin                Aceredation No:: SCS 108
‘Tne Swiss Accredtation Service is one of th signatories to the EA
Muittateral Agreement fo the recognition of calbraion certicates

Glossary:
TSL                        tissue simulating liquid
NORMxy.z                   sensitivity in free space
ConF                       sensitivity in TSL / NORMxy.z
DCP                        diode compression point
Polarization q             © rotation around probe axis
Polarization 8             3 rotation around an axis that is in the plane normal to probe axis (at
                           measurement center), ie., 8 = 0 is normal to probeaxis

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", December 2003
   b) CENELEC EN 50361, "Basic standard for the measurement of Specific Absorption Rate
      related to human exposure to electromagnetic fields from mobile phones (300 MHz — 3
         GHz), July 2001

Methods Applied and Interpretation of Parameters:
   + NORMxy,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell; £> 1800 MHz:
     R22 waveguide). NORMx.y,z are only intermediate values, i.e., the uncertainties of
         NORMxy,z does not effect the E*—field uncertainty inside TSL (see below ConvF).
    + NORM(Ax.y.z              NORMxy,z * frequency_response (see Frequency Response Chart). This
         linearization is implemented in DASY4 software versionslater than 4.2. The uncertainty of
        the frequency response is included in the stated uncertainty of ConvF.
    +   DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of
        power sweep (no uncertainty required). DCP does not depend on frequency nor media.
    *    ConvF and Boundary Effect Parameters: Assessed in fat phantom using E—field (or
         Temperature Transfer Standard for f < 800 MHz) and inside waveguide using analytical field
         distributions based on power measurements for f > 800 MHz. The same setups are used for
         assessment of the parameters applied for boundary compensation (alpha, depth) of which
         typical uncertainty values are given. These parameters are used in DASY4 software to
         improveprobe accuracy close to the boundary. The sensitivity in TSL corresponds to
         NORMxy,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A
        frequency dependent Conv is used in DASY version 4.4 and higher which allows
         extending the validity from + 50 MHz to + 100 MHz.
    *   Spherical isotropy (3D deviation from isotropy); in a field of low gradients realized using a
        flat phantom exposed by a patch antenna.
    *   Sensor Offset: The sensor offset corresponds to the offset of virtual measurementcenter
        from the probe tip (on probe axis). No tolerance required.


Cenificate No: ET3—1504_May0$                               Page 2 0t 9


   ET3DV6 SN:1604                                                               May 2, 2006




                      Probe ET3DVGC6

                                         SN:1604
                               Manufactured:                   July 30, 2001
                               Last calibrated:                March 18, 2005
                               Recalibrated:                   May 2, 2006

                                   Calibrated for DASY Systems
                                    (Note: non—compatibe with DASY2 system!}




Ceniicate No: ET3—1604_MayO6                  Page 3 of 0


   ET3DV6 SN:1604                                                                               May 2, 2006


   DASY — Parameters of Probe: ET3DV6 SN:1604

   Sensitivity in Free Space"®                                                  Diode Compression®
            NormxX                 187 a10.1%           pVi(Vim)®              DCP X          93 mV
            NormY                  1.80 £10.1%          uVi(Vim®               DCP Y          93 mV
            Normz                  191 s 101%           uVi(Vim)"               por z         93 mV

   Sensitivity in Tissue Simulating Liquid (Conversion Factors)
   Please see Page 8.


   Boundary Effect
   Tst                   900 MHz         Typical SAR gradient: 5 % per mm

            Sensor Center to Phantom Surface Distance                           3.7 mm 47 mm
            SARs. [            Without Correction Algonthm                         79    41
            SAR., []           With Correction Algorthm                            o1    o2

   Tst                  1810 MHz.        Typical SAR gradient: 10 % per mm

            Sensor Center to Phantom Surface Distance                           3.7 mm 4.7 mm
            SAR.. DK]          Without Correction Algorthm                         70    41
            SARs. Dsl          With Correction Algorithm                           0.1   02


   Sensor Offset
            Probe Tip to Sensor Center                                     2.7 mm



    The reported uncertainty of measurement is stated as the standard uncertainty of
    imeasurement multiplied by the coverage factor k=2, which for a normal distribution
    (corresponds to a coverage probability of approximately 95%.

   * The unceraintes of NomXY.Z do notafecthe Eid uncertainy inaide TSL (see Page 0)
   * Nomeral ineareason paramater unonriiny ns eaured




Certfcate No: E3—1604_May6                          Page 4 of 9


   ET3DV6 SN:1604                                                                                              May 2, 2006



                                                Frequency Response of E—Field
                                                    (TEM—Cell:if110 EXX, Waveguide: R22)


                                 1s                                —              ——         —

                                 14                    —.          i

                                 13
             Frequency response (normalized)
                           s




                                ce                —|—
                                o8                                                                       4

                                or                                                  ——            ~——

                                 as                                    —

                                 as ——                      —                    L           1
                                    o             so        1000       1800       2000     2500         3000
                                                                    t un
                                                       —o—rem                 —o—nze


                                               Uncertainty of Frequency Response of E—field: 2 6.3% (k=Z)




Certiicate No: ET9—1604_May06                                      Page 5 or 0


   ET3DV6 SN:1604                                                                       May 2, 2006




                                 Receiving Pattern (¢), 8 = 0°

                £= 600 MHz, TEM if110EXX                          £=1800 MHz, WG R22




                                                              <e—x     <o—y    <#—20    —0—Tat
   Error {d8)




                               Uncertainty of Axial Isotropy Assessment: 2 0.5% (k=2)




Certfcate No: E73— 1604Mayo6                    Page Sot 0


   ET3DV6 SN:1604                                                                              May 2, 2006


                                      Dynamic Range f(SARneaq)
                                            (Waveguide R22, f = 1800 MHz)
           input Signal fav)
             Error {dB]




                               o00t   cor          0              1          10          100

                                                   SAR [riiem]


                                      Uncertainty of Lineaity Assessment: £.0.6% (k=2)




Cerifcate No: ET3—1604_Mayos                         Page 7 of0


ET3DVG SN:1604                                                                                      May 2, 2006



                            Conversion Factor Assessment

            1= 900 MHz, WGLS R9 (head)                               1= 1810 MHiz, WGLS R22 (head)
                            t                                       300
          35

          30                                                        250

      & 25                                                      E 200


                                                                E
                                                                & 150

      &Eis
      &


       8 i0                                                     § o

          os :                                                       so
          ae 1—L_—‘                                                  so




pubte) Validity [MiHz]®         tst.   Permittvity    Conductivity        Alpha Dopth       ConvE Uncertainty
450        +s0/2100             Head   435 25%        0.8725%             ore     206         T14   £133%(c2)
800        £50/2100             Head   41.525%        0.9725%             ose     181         660   £11.0%(e2)
1810       +50/2100             Head   40.05%         14025%              oso     25e         520   £11.0%(c2)
2450       £50/2100             Head   39.225%        18025%              oss     185         460   £11.8%(e2)


4so        +50/2100             Boty   se.7a5%        09425%              o2s     444         742. £13.3% (c2)
800        +50/2100             Bocy   55.015%        1.0525%             o4r     208         627 £11.0% (c2)
1810       50/2100              Body   §3.315%        1.5225%             ose     286         488 211.0% (n2)
2450       £50/2100             Bogy   52.7%5%        1.9525%             oss     185         427 211.8%(k2)




® the vality o2 100 Mit only applles for DASY v6.t and Nigher (see Page 2 The uncertaint is t RSS
w the Cone® uncertainy at caltration regueney and the uneertainyfor the indlcate trequency band.



                                                   Page 8 0(9



Document Created: 2006-08-28 18:03:13
Document Modified: 2006-08-28 18:03:13

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC