SAR Appendix D

FCC ID: OVFKWC-KX7

RF Exposure Info

Download: PDF
FCCID_566631

                                                                FCC ID: OVFKWC-KX7




                                        Appendix D:

                                Dipole Calibration Parameters




Dipole Calibration Parameters                                           Model: KX7


Calibration Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland




Client




 Object(s)


 Calibration procedure(s)




 Calibration date:


 Condition of the calibrated item



 This calibration statement documents traceability of M&TE used in the calibration procedures and conformity of the procedures with the ISO/EC 17025
  international standard.


  All calibrations have been conducted in the closed laboratory facility: environment temperature 22 +/— 2 degrees Celsius and humidity < 75%.


  Calibration Equipment used (M&TE critical for calibration)


  Model Typs                             D#                        Cal Date (Calibrated by, Certificate No.)           Scheduled Calibration
  Power meter EPM E442                   GB37480704                6—Nov—03 (METAS, No. 252—0254)                      Nov—04
  Power sensor HP 8481A                  US37292783                6—Nov—03 (METAS, No. 252—0254)                      Nov—04
  Power sensor HP 8481A                  MY41092317                18—0ct—02 (Agilent, No. 20021018)                   Oct—04
  RF generator R&S SML—O3                1006988                   27—Mar—2002 (R&S, No. 20—92389)                     In house check: Mar—OS
  Network Analyzer HP 8753E              US37390585                18—Oct—01 (SPEAG, in house check Nov—03)            In house check:; Oct 05



                                         Name                                   Function                                Signature
  Calibrated by:



  Approved by:



                                                                                                                        Date issued: April 21, 2004



  This calibration certificate is issued as an intermediate solution until the accreditation process (based on ISO/EC 17025 International Standard) for
  Calibration Laboratory of Schmid & Partner Engineering AG is completed.




 880—KPO301061—A                                                                                                                                 Page 1 (1)


Schmid & Partner Engingering AG                        s    E      e   a   g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 245 9779
info@speag.com, http://www.speag.com




                                               DASY
                      Dipole Validation Kuit

                           Type: D835V2

                         Serial:                454

                      Manufactured:             January 31, 2002
                            Calibrated:         April 20, 2004


1.      Measurement Conditions


The measurements were performed in the flat section of the SAM twin phantom filled with head
simulating solution of the following electrical parameters at 835 MHz:

        Relative Dielectricity                 42.8         +5%
        Conductivity                           0.94 mho/m   #5%

The DASY4 System with a dosimetric E—field probe ET3DV6 (SN:1507, Conversion factor 6.3 at 835
MHz) was used for the measurements.

The dipole was mounted on the small tripod so that the dipole feedpoint was positioned below the
center marking of the flat phantom section and the dipole was oriented parallel to the body axis (the
long side of the phantom). The standard measuring distance was 15mm from dipole center to the
solution surface. The included distance spacer was used during measurements for accurate distance
positioning.

The coarse grid with a grid spacing of 1 5mm was aligned with the dipole. The 7x7x7 fine cube was
chosen for cube integration.
The dipole input power (forward power) was 250 mW +3 %. The results are normalized to 1 W input
power.




2.      SAR Measurement with DASY4 System

Standard SAR—measurements were performed according to the measurement conditions described in
section 1. The results (see figure supplied) have been normalized to a dipole input power of 1 W
(forward power). The resulting averaged SAR—values measured with the dosimetric probe ET3DV6
SN:1507 and applying the advanced extrapolation are:


        averaged over 1 cm" (1 g) of tissue:         10.2 mW/g + 16.8 % (k=2)‘

        averaged over 10 em‘ (10 g) of tissue:       6.64 mWi/ig + 16.2 % (k=2)‘




‘ validation uncertainty


3.       Dipole Impedance and Return Loss

The impedance was measured at the SMA—connector with a network analyzer and numerically
transformed to the dipole feedpoint. The transformation parameters from the SMA—connector to the
dipole feedpoint are:

      Electrical delay:           1.378 ns    (one direction)
      Transmission factor:        0.988      (voltage transmission, one direction)


The dipole was positioned at the flat phantom sections according to section 1 and the distance holder
was in place during impedance measurements.


         Feedpoint impedance at 835 MHz:             Re{Z} = 50.9 Q

                                                     Im {Z} = —2.2 Q

         Return Loss at 835 MHz                      —32.3 dB




4.       Measurement Conditions

The measurements were performed in the flat section of the SAM twin phantom filled with body
simulating solution of the following electrical parameters at 835 MHz:

         Relative Dielectricity               55.5           + 5%
         Conductivity                         0.99 mho/m     £5%

The DASY4 System with a dosimetric E—field probe ET3DV6 (SN:1507, Conversion factor 6.13 at 835
MHz) was used for the measurements.

The dipole was mounted on the small tripod so that the dipole feedpoint was positioned below the
center marking of the flat phantom section and the dipole was oriented parallel to the body axis (the
long side of the phantom). The standard measuring distance was 15mm from dipole center to the
solution surface. The included distance spacer was used during measurements for accurate distance
positioning.

The coarse grid with a grid spacing of 1 5mm was aligned with the dipole. The 7x7x7 fine cube was
chosen for cube integration.
The dipole input power (forward power) was 250 mW + 3 %. The results are normalized to 1 W input
power.


5,      SAR Measurement with DASY4 System

Standard SAR—measurements were performed according to the measurement conditions described in
section 4. The results (see figure supplied) have been normalized to a dipole input power of 1 W
(forward power). The resulting averaged SAR—values measured with the dosimetric probe ET3DV6
SN:1507 and applying the advanced extrapolation are:


        averaged over 1 cm" (1 g) of tissue:        10.1 mW/g + 16.8 % (k=2)
        averaged over 10 em" (10 g) of tissue:      6.64 mWig + 16.2 % (k=2)



6.      Dipole Impedance and Return Loss

The dipole was positioned at the flat phantom sections according to section 4 and the distance holder
was in place during impedance measurements.


        Feedpoint impedance at 835 MHz:             Re{Z} = 47.2 Q

                                                     Im {Z} = —1.1 Q

        Return Loss at 835 MHz                       —29.6 dB



7.       Handling


Do not apply excessive force to the dipole arms, because they might bend. Bending of the dipole arms
stresses the soldered connections near the feedpoint leading to a damage of the dipole.


8.       Design

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is
directly connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—
signals.


9.       Power Test

After long term use with 1 00W radiated power, only a slight warming of the dipole near the feedpoint
can be measured.




2 validation uncertainty


                                                                   Date/Time: 04/20/04 12:55:03

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN454

Communication System: CW—835; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: HSL 835 MHz;
Medium parameters used: £= 835 MHz; 0 = 0.94 mho/m; &s, = 42.8; p = 1000 kg/m*
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:

   e Probe: ET3DV6 — SN1507; ConvF(6.3, 6.3, 6.3); Calibrated: 1/23/2004
     Sensor—Surface: 4mm (Mechanical Surface Detection)
     Electronics: DAE3 Sn411; Calibrated: 11/6/2003
     Phantom: SAM with CRP — TP1006; Type: SAM 4.0; Serial: TP:1006;
     Measurement SW : DASY4, V4.2 Build 44; Postprocessing SW: SEMCAD, V1.8 Build 112

Pin = 250 mW; d = 15 mm/Area Scan (81x81x1):; Measurement grid: dx=1 5mm, dy=1 5mm
Reference Value = 55.5 V/m; Power Drift = —0.0 dB
Maximum value of SAR (interpolated) = 2.75 mW/g

Pin = 250 mW; d = 15 mm/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 55.5 V/m; Power Drift =—0.0 dB
Maximum value of SAR (measured) = 2.76 mW/g
Peak SAR (extrapolated) = 3.88 W/kg
SAR(1 g) = 2.56 mW/g; SAR(10 g) = 1.66 mW/g




             0 dB =2.76mW/g


                                                          20 fApr 2004    08:28: 07
CH1    S41    1 U FS           1:50.910 6   2.2402 6   427.99 pH         835.9900 002 MHz




Del


Cor




fivg
16




CH2




Cor




             CENTER $35.9090 000 MHz                               SPAN 400.000 900 MHz


                                                                  Date/Time: 04/16/04 13:28:44

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN454

Communication System: CW—835; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Muscle 835 MHz;
Medium parameters used: f= 835 MHz; 0 = 0.99 mho/m; €, = 55; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)

DASY4 Configuration:

     Probe: ET3DV6 — SN1507; ConvF(6.13, 6.13, 6.13); Calibrated: 1/23/2004
     Sensor—Surface: 4mm (Mechanical Surface Detection)
     Electronics: DAE3 Sn411; Calibrated: 11/6/2003
     Phantom: SAM with CRP — TP1006; Type: SAM 4.0; Serial: TP:1006;
     Measurement SW:; DASY4, V4.2 Build 44; Postprocessing SW: SEMCAD, V1.8 Build 112

Pin = 250 mW; d = 15 mm/Area Scan (81x81x1): Measurement grid: dx=15mm, dy=1 5mm
Reference Value = 54.2 V/m; Power Drift = 0.004 dB
Maximum value of SAR (interpolated) = 2.74 mW/g

Pin = 250 mW; d = 15 mm/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 54.2 V/m; Power Drift = 0.004 dB
Maximum value of SAR (measured) = 2.72 mW/g
Peak SAR (extrapolated) = 3.69 W/kg
SAR(1 g) = 2.52 mW/g; SAR(10 g) = 1.66 mW/g




             0 dB =2.72mW/g


                                                                                            4+ &V\\/


                                                       16 fipr 2004      16:28:103
CHI) S11    1 U FS          1: 47.172 &   —14191 4   1798.31 pF          $35.0940 000 MHz




Del

Cor




fAv
16 9




CH2




Cor




           CENTER §35.000 000 MHz                                 SPAN 4090.000 009 MHz



Document Created: 2005-07-13 14:22:48
Document Modified: 2005-07-13 14:22:48

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC