HAC RF Data Plots

FCC ID: OVF-K33BIC01

Test Report

Download: PDF
FCCID_1048549

                                                                                                                       FCC ID: OVF-K33BIC01


                                                                                                                                  Date: 12/14/2008


File Name: FCC H-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


File Name: FCC E-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


Communication System: CDMA-1900; Frequency: 1851.25 MHz;Duty Cycle: 1:1
Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1 kg/m3 Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1000 kg/m3
Phantom section: H Device Section Phantom section: E Device Section


DASY4 Configuration:
- Probe: H3DV6 - SN6123Probe: ER3DV6 - SN2341; ConvF(1, 1, 1); Calibrated: 8/18/2008Calibrated: 4/17/2008
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn530; Calibrated: 4/15/2008
- Phantom: HAC Test Arch; Type: SD HAC P01 BA;
- Measurement SW: DASY4, V4.7 Build 71; Postprocessing SW: SEMCAD, V1.8 Build 184




Ch25_Backlight On/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 0.153 A/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 0.155 A/m; Power Drift = -0.139 dB



                                                              Peak H-field in A/m


                                                               Grid 1       Grid 2    Grid 3
                                                               0.082 M4 0.121 M4 0.120 M4
                                                               Grid 4       Grid 5    Grid 6
                                                               0.111 M4 0.153 M4 0.148 M4
                                                               Grid 7       Grid 8    Grid 9
                                                               0.115 M4 0.151 M4 0.146 M4




Ch25_Backlight On/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 78.3 V/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 78.4 V/m; Power Drift = -0.150 dB

                                                                Peak E-field in V/m


                                                                 Grid 1     Grid 2    Grid 3
                                                                 42.6 M4 68.2 M3 67.4 M3
                                                                 Grid 4     Grid 5    Grid 6
                                                                 55.1 M4 78.3 M3 77.3 M3
                                                                 Grid 7     Grid 8    Grid 9
                                                                 59.1 M4 78.0 M3 76.3 M3


                                                                              +




  HAC RF Data Plots                                                       Page 1 of 11                                    Model: K33BIC-01


                                                     FCC ID: OVF-K33BIC01




                    0 dB = 0.153A/m




HAC RF Data Plots                     Page 2 of 11      Model: K33BIC-01


                                                                                                                       FCC ID: OVF-K33BIC01


                                                                                                                                  Date: 12/14/2008


File Name: FCC H-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


File Name: FCC E-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


Communication System: CDMA-1900; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1 kg/m3 Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1000 kg/m3
Phantom section: H Device Section Phantom section: E Device Section


DASY4 Configuration:
- Probe: H3DV6 - SN6123Probe: ER3DV6 - SN2341; ConvF(1, 1, 1); Calibrated: 8/18/2008Calibrated: 4/17/2008
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn530; Calibrated: 4/15/2008
- Phantom: HAC Test Arch; Type: SD HAC P01 BA;
- Measurement SW: DASY4, V4.7 Build 71; Postprocessing SW: SEMCAD, V1.8 Build 184




Ch600_Backlight On/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 0.167 A/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 0.168 A/m; Power Drift = -0.029 dB


                                                              Peak H-field in A/m


                                                               Grid 1       Grid 2    Grid 3
                                                               0.094 M4 0.136 M4 0.136 M4
                                                               Grid 4       Grid 5    Grid 6
                                                               0.125 M4 0.167 M4 0.164 M4
                                                               Grid 7       Grid 8    Grid 9
                                                               0.131 M4 0.165 M4 0.161 M4




Ch600_Backlight On/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 91.4 V/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 90.7 V/m; Power Drift = -0.158 dB

                                                                Peak E-field in V/m


                                                                 Grid 1     Grid 2    Grid 3
                                                                 52.0 M4 79.2 M3 78.6 M3
                                                                 Grid 4     Grid 5    Grid 6
                                                                 67.1 M3 91.4 M3 90.5 M3
                                                                 Grid 7     Grid 8    Grid 9
                                                                 71.4 M3 90.9 M3 89.1 M3




  HAC RF Data Plots                                                       Page 3 of 11                                    Model: K33BIC-01


                                                     FCC ID: OVF-K33BIC01




                    0 dB = 0.167A/m




HAC RF Data Plots                     Page 4 of 11      Model: K33BIC-01


                                                                                                                       FCC ID: OVF-K33BIC01


                                                                                                                                  Date: 12/14/2008


File Name: FCC H-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


File Name: FCC E-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


Communication System: CDMA-1900; Frequency: 1908.75 MHz;Duty Cycle: 1:1
Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1 kg/m3 Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1000 kg/m3
Phantom section: H Device Section Phantom section: E Device Section


DASY4 Configuration:
- Probe: H3DV6 - SN6123Probe: ER3DV6 - SN2341; ConvF(1, 1, 1); Calibrated: 8/18/2008Calibrated: 4/17/2008
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn530; Calibrated: 4/15/2008
- Phantom: HAC Test Arch; Type: SD HAC P01 BA;
- Measurement SW: DASY4, V4.7 Build 71; Postprocessing SW: SEMCAD, V1.8 Build 184




Ch1175_Backlight On/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 0.124 A/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 0.124 A/m; Power Drift = 0.138 dB



                                                              Peak H-field in A/m


                                                               Grid 1       Grid 2    Grid 3
                                                               0.074 M4 0.104 M4 0.104 M4
                                                               Grid 4       Grid 5    Grid 6
                                                               0.092 M4 0.124 M4 0.123 M4
                                                               Grid 7       Grid 8    Grid 9
                                                               0.091 M4 0.121 M4 0.119 M4




Ch1175_Backlight On/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 66.9 V/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 64.6 V/m; Power Drift = 0.012 dB

                                                                Peak E-field in V/m


                                                                 Grid 1     Grid 2    Grid 3
                                                                 40.0 M4 60.7 M4 60.3 M4
                                                                 Grid 4     Grid 5    Grid 6
                                                                 46.0 M4 66.9 M3 66.7 M3
                                                                 Grid 7     Grid 8    Grid 9
                                                                 46.9 M4 64.7 M3 64.0 M3




  HAC RF Data Plots                                                       Page 5 of 11                                    Model: K33BIC-01


                                                     FCC ID: OVF-K33BIC01




                    0 dB = 0.124A/m




HAC RF Data Plots                     Page 6 of 11      Model: K33BIC-01


                                                                                                                       FCC ID: OVF-K33BIC01


                                                                                                                                  Date: 12/14/2008


File Name: FCC H-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


File Name: FCC E-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


Communication System: CDMA-1900; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1 kg/m3 Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1000 kg/m3
Phantom section: H Device Section Phantom section: E Device Section


DASY4 Configuration:
- Probe: H3DV6 - SN6123Probe: ER3DV6 - SN2341; ConvF(1, 1, 1); Calibrated: 8/18/2008Calibrated: 4/17/2008
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn530; Calibrated: 4/15/2008
- Phantom: HAC Test Arch; Type: SD HAC P01 BA;
- Measurement SW: DASY4, V4.7 Build 71; Postprocessing SW: SEMCAD, V1.8 Build 184




Ch600 Backlight Off/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 0.167 A/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 0.171 A/m; Power Drift = -0.057 dB



                                                              Peak H-field in A/m


                                                               Grid 1       Grid 2    Grid 3
                                                               0.095 M4 0.134 M4 0.134 M4
                                                               Grid 4       Grid 5    Grid 6
                                                               0.127 M4 0.167 M4 0.163 M4
                                                               Grid 7       Grid 8    Grid 9
                                                               0.131 M4 0.164 M4 0.160 M4




Ch600 Backlight Off/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 91.6 V/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 87.6 V/m; Power Drift = 0.065 dB




                                                                Peak E-field in V/m


                                                                 Grid 1     Grid 2    Grid 3
                                                                 51.9 M4 79.3 M3 78.7 M3
                                                                 Grid 4     Grid 5    Grid 6
                                                                 66.1 M3 91.6 M3 90.6 M3
                                                                 Grid 7     Grid 8    Grid 9
                                                                 70.2 M3 91.1 M3 89.8 M3




  HAC RF Data Plots                                                       Page 7 of 11                                    Model: K33BIC-01


                                                     FCC ID: OVF-K33BIC01




                    0 dB = 0.167A/m




HAC RF Data Plots                     Page 8 of 11      Model: K33BIC-01


                                                                                                                       FCC ID: OVF-K33BIC01


                                                                                                                                  Date: 12/14/2008


File Name: FCC H-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


File Name: FCC E-FIELD Ceramix #3139, 1900Mhz, Dec14, 08.da4


Communication System: CDMA-1900; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1 kg/m3 Medium parameters used: σ = 0 mho/m, εr = 1; ρ = 1000 kg/m3
Phantom section: H Device Section Phantom section: E Device Section


DASY4 Configuration:
- Probe: H3DV6 - SN6123Probe: ER3DV6 - SN2341; ConvF(1, 1, 1); Calibrated: 8/18/2008Calibrated: 4/17/2008
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn530; Calibrated: 4/15/2008
- Phantom: HAC Test Arch; Type: SD HAC P01 BA;
- Measurement SW: DASY4, V4.7 Build 71; Postprocessing SW: SEMCAD, V1.8 Build 184




Ch600 Backlight Off (360 Degree)/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 0.167 A/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 0.168 A/m; Power Drift = 0.107 dB



                                                              Peak H-field in A/m


                                                               Grid 1       Grid 2    Grid 3
                                                               0.094 M4 0.135 M4 0.135 M4
                                                               Grid 4       Grid 5    Grid 6
                                                               0.127 M4 0.167 M4 0.164 M4
                                                               Grid 7       Grid 8    Grid 9
                                                               0.132 M4 0.165 M4 0.160 M4




(360 Degree) Ch600 Backlight Off/Hearing Aid Compatibility Test (101x101x1): Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field = 88.6 V/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, 353.7 mm
Reference Value = 89.1 V/m; Power Drift = 0.045 dB

                                                                Peak E-field in V/m


                                                                 Grid 1     Grid 2    Grid 3
                                                                 51.1 M4 77.3 M3 76.9 M3
                                                                 Grid 4     Grid 5    Grid 6
                                                                 64.2 M3 88.6 M3 87.8 M3
                                                                 Grid 7     Grid 8    Grid 9
                                                                 68.5 M3 88.0 M3 86.5 M3




  HAC RF Data Plots                                                       Page 9 of 11                                    Model: K33BIC-01


                                                      FCC ID: OVF-K33BIC01




                    0 dB = 0.167A/m




HAC RF Data Plots                     Page 10 of 11       Model: K33BIC-01


                                    FCC ID: OVF-K33BIC01




HAC RF Data Plots   Page 11 of 11       Model: K33BIC-01