RF Exposure Infomation

FCC ID: OU9TMB1872BS

RF Exposure Info

Download: PDF
FCCID_4349014

                                         RF exposure Estimation

1.    Introduction

      Product: Blood Pressure Monitor
      Model no.: TMB-1872-B
      FCC ID: OU9TMB1872BS
      The EUT is Blood Pressure Monitor, which contain Bluetooth function inside.

2.    Limit and Guidelines on Exposure to Electromagnetic Fields
      According to§15.247(e)(i) and §1.1307(b)(1), systems operating under the provisions of this
      section shall be operated in a manner that ensure that the public is not exposed to radio
      frequency energy level in excess of the Commission’s guideline.

      According to KDB558074 D01 Mobile Portable RF Exposure v05r02, no SAR required if power
      is lower than the flowing threshold:
      The 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation
      distances ≤ 50 mm are determined by:
       [(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)]
       [√f(GHz)] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity SAR, where
           • f(GHz) is the RF channel transmit frequency in GHz
           • Power and distance are rounded to the nearest mW and mm before calculation25
           • The result is rounded to one decimal place for comparison
           • 3.0 and 7.5 are referred to as the numeric thresholds in the step 2 below

      The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm
      and for transmission frequencies between 100 MHz and 6 GHz.

3.    Calculation method

      [(max. power of` channel, including tune-up tolerance, mW)/(min. test separation distance,
      mm)] ·[√f(GHz)] ≤ 3.0

      Conducted Power + tune up tolerance =4.7dBm= 2.95mW
      Distance = 5 mm
      f = 2.48 GHz

      [2.95/5] * SQRT (2.402) =0.914
      0.914≤ 3.0
      Therefore, excluded from SAR testing.

- TÜV SÜD Shenzhen LTD. -
Reviewed by:                                                               Prepared By:




Laurent Yuan /EMC Project Manager                                 Mark Chen/EMC Project Engineer
Date: 2019-06-18                                                  Date: 2019-06-18

                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch