RF Test Report of BLE

FCC ID: OU9TMB1872BS

Test Report

Download: PDF
FCCID_4349013

Report Number: 68.930.19.0006.01




                                                               FCC - TEST REPORT



Report Number                                          : 68.930.19.0006.01                              Date of Issue:                       June 27, 2019


Model                                                  : TMB-1872-B

Product Type                                           : Blood Pressure Monitor

Applicant                                              : Guangdong Transtek Medical Electronics Co., Ltd

Address                                                : Zone A, No.105, Dongli Road Torch Development District

                                                       : Zhongshan, Guangdong China 528437

 Factory                                               : Guangdong Transtek Medical Electronics Co., Ltd

Address                                                : Zone B, No.105, Dongli Road Torch Development District

                                                       : Zhongshan, Guangdong China 528437




Test Result                                            : n Positive                   o Negative



Total pages including
Appendices                                             : 34




TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch is a subcontractor to TÜV SÜD Product Service GmbH according to the principles outlined in
ISO 17025.

TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch reports apply only to the specific samples tested under stated test conditions. Construction of
the actual test samples has been documented. It is the manufacturer's responsibility to assure that additional production units of this model are manufactured with
identical electrical and mechanical components. The manufacturer/importer is responsible to the Competent Authorities in Europe for any modifications made to the
production units which result in non-compliance to the relevant regulations. TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch shall have no
liability for any deductions, inferences or generalizations drawn by the client or others from TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
issued reports.

This report is the confidential property of the client. As a mutual protection to our clients, the public and ourselves, extracts from the test report shall not be
reproduced except in full without our written approval.




EMC_SZ_FR_21.00 FCC                                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                               Page 1 of 34
                                          Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                               Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




1     Table of Contents
1     Table of Contents......................................................................................................................................................... 2
2     Details about the Test Laboratory .............................................................................................................................. 3
3     Description of the Equipment Under Test .................................................................................................................. 4
4     Summary of Test Standards ....................................................................................................................................... 5
5     Summary of Test Results ............................................................................................................................................ 6
6     General Remarks ......................................................................................................................................................... 7
7     Test Setups .................................................................................................................................................................. 8
8     Systems test configuration .......................................................................................................................................... 9
9     Technical Requirement.............................................................................................................................................. 10
    9.1      Conducted Emission ......................................................................................................................................... 10
    9.2      Conducted peak output power.......................................................................................................................... 13
    9.3      Power spectral density ...................................................................................................................................... 16
    9.4      6 dB Bandwidth and 99% Occupied Bandwidth .............................................................................................. 19
    9.5      Spurious RF conducted emissions ................................................................................................................... 23
    9.6      Band edge .......................................................................................................................................................... 27
    9.7      Spurious radiated emissions for transmitter .................................................................................................... 29
10        Test Equipment List ............................................................................................................................................... 33
11        System Measurement Uncertainty ....................................................................................................................... 34




EMC_SZ_FR_21.00 FCC                                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                                      Page 2 of 34
                                           Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




2   Details about the Test Laboratory
Details about the Test Laboratory
Test Site 1

Company name:         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
                      Building 12 & 13, Zhiheng Wisdomland Business Park, Nantou Checkpoint
                      Road 2, Nanshan District
                      Shenzhen 518052
                      P.R. China

Telephone:            86 755 8828 6998
Fax:                  86 755 8288 5299

FCC Registration      514049
No.:




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 3 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




3   Description of the Equipment Under Test

Product:                          Blood Pressure Monitor

Model no.:                        TMB-1872-B

FCC ID:                           OU9TMB1872BS

Options and accessories:          USB Cable, Adapter

Rating:                           6VDC, (Supplied by 4×1.5V AAA batteries)
                                  5VDC, 1000mA (Charged by AC/AD adapter)
Adapter                           Model: BLJ06L050100U-U
                                  Input: 100-240V~50/60Hz 0.2A max
                                  Output: 5V/1000mA

RF Transmission                   2402MHz-2480MHz
Frequency:
No. of Operated Channel:          40

Modulation:                       GFSK

Antenna Type:                     Integrated Antenna

Antenna Gain:                     0dBi

Description of the EUT:           The Equipment Under Test (EUT) is Blood Pressure Monitor
                                  operated at 2.4GHz.




EMC_SZ_FR_21.00 FCC                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 4 of 34
                          Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20               Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




4   Summary of Test Standards

                                         Test Standards
FCC Part 15 Subpart C       PART 15 - RADIO FREQUENCY DEVICES
10-1-2017 Edition           Subpart C - Intentional Radiators

All the test methods were according to KDB558074 D01 v05r02 DTS Measurement Guidance
and ANSI C63.10 (2013).




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 5 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




5   Summary of Test Results
                                           Technical Requirements
FCC Part 15 Subpart C
                                                                                                      Test          Test Result
Test Condition                                                                         Pages
                                                                                                      Site        Pass Fail N/A
§15.207               Conducted emission AC power port                                 10           Site 1

§15.247 (b) (1)       Conducted peak output power                                      13           Site 1

§15.247(a)(1)         20dB bandwidth                                                   ---          ---

§15.247(a)(1)         Carrier frequency separation                                     ---          ---

§15.247(a)(1)(iii)    Number of hopping frequencies                                    ---          ---

§15.247(a)(1)(iii)    Dwell Time                                                       ---          ---
                      6dB bandwidth and 99% Occupied
§15.247(a)(2)                                                                          16           Site 1
                      Bandwidth
§15.247(e)            Power spectral density                                           19           Site 1

§15.247(d)            Spurious RF conducted emissions                                  23           Site 1

§15.247(d)            Band edge                                                        27           Site 1
§15.247(d) &
                      Spurious radiated emissions for
§15.209 &                                                                              29           Site 1
                      transmitter
§15.205
§15.203               Antenna requirement                                              See note 2

Note 1: N/A=Not Applicable.
Note 2: The EUT uses a Integrated antenna, which gain is 0dBi. In accordance to §15.203, it is
considered sufficiently to comply with the provisions of this section.




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 6 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




6   General Remarks
Remarks

This submittal(s) (test report) is intended for FCC ID: OU9TMB1872BS complies with Section
15.205,15.207 15.209, 15.247 of the FCC Part 15, Subpart C.

SUMMARY:

All tests according to the regulations cited on page 5 were

n - Performed

o - Not Performed

The Equipment under Test

n - Fulfills the general approval requirements.

o - Does not fulfill the general approval requirements.


Sample Received Date:                    February 21, 2019

Testing Start Date:                      February 21, 2019

Testing End Date:                            April 22, 2019


- TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch -

Reviewed by:                                Prepared by:                                          Tested by:




    _______________________                      _______________________                           _______________________
        Laurent Yuan                                  Mark Chen                                       Carry Cai
     EMC Project Manager                          EMC Project Engineer                            EMC Test Engineer




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 7 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




7     Test Setups

Below 1GHz




Above 1GHz




Conducted RF test setups


    Measuring                 EUT
    Receiver


AC Power Line Conducted Emission test setups




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 8 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




8   Systems test configuration

Auxiliary Equipment Used during Test:

            DESCRIPTION             MANUFACTURER                     MODEL NO.(SHIELD)                    S/N(LENGTH)
               Notebook                     Lenovo                              X220                               ---
                  ---                         ---                                ---



Test software: EMI_Test_Tool_v1.3.exe test tool, which used to control the EUT in continues
transmitting mode.

The system was configured to channel 0, 19, and 39 for the test.




EMC_SZ_FR_21.00 FCC                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 9 of 34
                          Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20               Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




9   Technical Requirement


9.1 Conducted Emission

Test Method
   1. The EUT was placed on a table, which is 0.8m above ground plane
   2. The power line of the EUT is connected to the AC mains through a Artificial Mains
      Network (A.M.N.).
   3. Maximum procedure was performed to ensure EUT compliance
   4. A EMI test receiver is used to test the emissions from both sides of AC line

Limit

                           Frequency                          QP Limit                        AV Limit
                              MHz                              dBμV                            dBμV
                           0.150-0.500                         66-56*                          56-46*
                             0.500-5                             56                              46
                               5-30                              60                              50

*Decreases with the logarithm of the frequency.




EMC_SZ_FR_21.00 FCC                       TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 10 of 34
                            Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                 Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




Product Type                            :    Blood Pressure Monitor
M/N                                     :    TMB-1872-B
Operating Condition                     :    AC Powered(Adapter)+ Measuring+ BT Link
Test Specification                      :    Line
Comment                                 :    5VDC(Supplies by AC/AD Adapter)



                                 80

                                 70

                                 60
                 Level in dB 礦




                                 50

                                 40

                                 30

                                 20

                                 10

                                 0
                                 150k           300 400 500         800 1M             2M      3M 4M 5M 6           8 10M            20M        30M
                                                                                 Frequency in Hz


Critical_Freqs
  Frequency      MaxPeak*                   Average      Limit     Margin      Line     Corr.**
    (MHz)         (dBµV)                    (dBµV)      (dBµV)      (dB)                (dB/m)
      0.170000       39.58                        ---     64.96      25.39      L1        10.2
      0.334000       33.15                        ---     59.35      26.20      L1        10.2
      0.434000       31.29                        ---     57.18      25.89      L1        10.3
      1.274000       21.27                        ---     56.00      34.73      L1        10.3
      2.690000       20.58                        ---     56.00      35.42      L1        10.3
     15.458000       28.96                        ---     60.00      31.04      L1        10.8

Remark
*Level=Reading Level + Correction Factor
**Correction Factor=Cable Loss + LISN Factor
(The Reading Level is recorded by software which is not shown in the sheet)




EMC_SZ_FR_21.00 FCC                                          TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                  Page 11 of 34
                                               Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                    Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01


Product Type                          :    Blood Pressure Monitor
M/N                                   :    TMB-1872-B
Operating Condition                   :    AC Powered(Adapter)+ Measuring+ BT Link
Test Specification                    :    Neutral
Comment                               :    5VDC(Supplies by AC/AD Adapter)



                               80

                               70

                               60
                Level in dB礦




                               50

                               40

                               30

                               20

                               10

                               0
                               150k           300 400 500         800 1M             2M      3M 4M 5M 6           8 10M            20M        30M
                                                                               Frequency in Hz


Critical_Freqs
  Frequency     MaxPeak*                  Average      Limit     Margin      Line     Corr.**
    (MHz)        (dBµV)                   (dBµV)      (dBµV)      (dB)                (dB/m)
     0.162000       39.91                       ---     65.36      25.45      N         10.2
     0.318000       33.70                       ---     59.76      26.06      N         10.3
     0.438000       31.25                       ---     57.10      25.85      N         10.3
     0.794000       25.95                       ---     56.00      30.05      N         10.3
     2.014000       27.02                       ---     56.00      28.98      N         10.3
     3.934000       25.63                       ---     56.00      30.37      N         10.4

Remark
*Level=Reading Level + Correction Factor
**Correction Factor=Cable Loss + LISN Factor
(The Reading Level is recorded by software which is not shown in the sheet)




EMC_SZ_FR_21.00 FCC                                        TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                  Page 12 of 34
                                             Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                  Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01


9.2 Conducted peak output power

Test Method
   1. Use the following spectrum analyzer settings:
       RBW > the 6dB bandwidth of the emission being measured, VBW≥3RBW, Span≥3RBW
       Sweep = auto, Detector function = peak, Trace = max hold.
   2. Add a correction factor to the display.
   3. Allow the trace to stabilize. Use the marker-to-peak function to set the marker to the
       peak of the emission. The indicated level is the peak output power.
Limits
According to §15.247 (b) (1), conducted peak output power limit as below:

                      Frequency Range                        Limit                           Limit
                            MHz                               W                              dBm
                         2400-2483.5                          ≤1                              ≤30

Test result as below table

                                                               Conducted
                                                               Peak Output                 Result
                                Frequency                        Power
                                   MHz                            dBm
                          Low channel 2402MHz                     4.70                      Pass
                         Middle channel 2440MHz                   4.29                      Pass
                          High channel 2480MHz                    3.79                      Pass




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 13 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




                                                      Low channel 2402MHz




                  Date:14.APR.2019 16:40:15




                                                    Middle channel 2440MHz




                  Date:14.APR.2019 16:42:05




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 14 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




                                                     High channel 2480MHz




                  Date:14.APR.2019 16:43:59




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 15 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




9.3 Power spectral density
Test Method
This procedure shall be used if maximum peak conducted output power was used to
demonstrate compliance:
   1. Set analyzer center frequency to DTS channel center frequency. RBW=10kHz,
       VBW≥3RBW, Span=1.5 times DTS bandwidth, Detector=Peak, Sweep=auto,
       Trace= max hold.
   2. Allow trace to fully stabilize, use the peak marker function to determine the maximum
       amplitude level within the RBW .
   3. Repeat above procedures until other frequencies measured were completed.
Limit

                                                    Limit [dBm/3kHz]

                                                              ≤8

Test result
                                                        Power spectral
                      Frequency                            density                                 Result
                          MHz                               dBm/10kHz
                 Top channel 2402MHz                          -8.01                                  Pass
                Middle channel 2440MHz                        -8.51                                  Pass
                Bottom channel 2480MHz                        -8.92                                  Pass




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 16 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01



                                                      Low channel 2402MHz




                  Date:14.APR.2019 16:40:21




                                                    Middle channel 2440MHz




                  Date:14.APR.2019 16:42:11




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 17 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




                                                     High channel 2480MHz




                  Date:14.APR.2019 16:44:05




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 18 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




9.4 6 dB Bandwidth and 99% Occupied Bandwidth
Test Method
      1. Use the following spectrum analyzer settings:
      RBW=100K, VBW≥3RBW, Sweep = auto, Detector function = peak, Trace = max hold
      2. Use the automatic bandwidth measurement capability of an instrument, may be
      employed using the X dB bandwidth mode with X set to 6 dB, care shall be taken so that
      the bandwidth measurement is not influenced by any intermediate power nulls in the
      fundamental emission that might be ≥ 6 dB.
      3. Allow the trace to stabilize, record the X dB Bandwidth value.

Limit

                                                        Limit [kHz]

                                                             ≥500


Test result
               Frequency               6dB bandwidth                    99 bandwidth
                                                                                                             Result
                  MHz                       kHz                             kHz
         Bottom channel 2402MHz                 1340                           2398                            Pass
         Middle channel 2440MHz                 1520                           2422                            Pass
          Top channel 2480MHz                   1416                           2338                            Pass




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 19 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




6 dB Bandwidth

                                                      Low channel 2402MHz




                  Date:14.APR.2019 16:40:08




                  Date:14.APR.2019 16:39:57




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 20 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




                                                    Middle channel 2440MHz




                  Date:14.APR.2019 16:41:58




                  Date:14.APR.2019 16:41:47




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 21 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01



                                                     High channel 2480MHz




                  Date:14.APR.2019 16:43:52




                  Date:14.APR.2019 16:43:40




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 22 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




9.5 Spurious RF conducted emissions

Test Method
   1. Establish a reference level by using the following procedure:
      a. Set RBW=100 kHz. VBW≥3RBW. Detector =peak, Sweep time = auto couple, Trace
         mode = max hold.
      b. Allow trace to fully stabilize, use the peak marker function to determine the maximum
         PSD level.
   2. Use the maximum PSD level to establish the reference level.
      a. Set the center frequency and span to encompass frequency range to be measured.
      b. Use the peak marker function to determine the maximum amplitude level. Ensure that
         the amplitude of all unwanted emissions outside of the authorized frequency band
         (excluding restricted frequency bands) are attenuated by at least the minimum
         requirements, report the three highest emissions relative to the limit.
   3. Repeat above procedures until other frequencies measured were completed.

Limit

                                    Frequency Range                           Limit (dBc)
                                          MHz
                                        30-25000                                    -20




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 23 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




Spurious RF conducted emissions


                                                             2402MHz




                  Date:14.APR.2019 16:40:46




                  Date:14.APR.2019 16:40:58




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 24 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01



                                                             2440MHz




                  Date:14.APR.2019 16:42:26




                  Date:14.APR.2019 16:42:38




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 25 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




                                                             2480MHz




                  Date:14.APR.2019 16:44:30




                  Date:14.APR.2019 16:44:41




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 26 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




9.6 Band edge
Test Method
   1 Use the following spectrum analyzer settings:
      Span = wide enough to capture the peak level of the in-band emission and all spurious
      RBW = 100 kHz, VBW≥RBW, Sweep = auto, Detector function = peak, Trace = max
      hold.
   2 Allow the trace to stabilize, use the peak and delta measurement to record the result.
   3 The level displayed must comply with the limit specified in this Section.

Limit
                                    Frequency Range                           Limit (dBc)
                                          MHz
                                        30-25000                                    -20




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 27 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




Band edge testing

                                                                2402MHz




                  Date:14.APR.2019 16:40:31




                                                                2480MHz




                  Date:14.APR.2019 16:44:15




EMC_SZ_FR_21.00 FCC                         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 28 of 34
                              Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                   Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




9.7 Spurious radiated emissions for transmitter

Test Method
   1: The EUT was place on a turn table which is 1.5m above ground plane for above 1GHz
   and 0.8m above ground for below 1GHz at 3 meter chamber room for test. The table was
   rotated 360 degrees to determine the position of the highest radiation.
   2: The EUT was set 3 meters away from the interference – receiving antenna, which was
   mounted on the top of a variable – height antenna tower.
   3: The height of antenna is varied from one meter to four meters above the ground to
   determine the maximum value of the field strength. Both horizontal and vertical polarizations
   of the antenna are set to make the measurement.
   4: For each suspected emission, the EUT was arranged to its worst case and then the
   antenna was tuned to heights from 1 meter to 4 meters and the rotatable table was turned
   from 0 degrees to 360 degrees to find the maximum reading.
   5: Use the following spectrum analyzer settings According to C63.10:
   For Above 1GHz
       Span = wide enough to capture the peak level of the in-band emission and all spurious
       RBW = 1MHz, VBW≥RBW for peak measurement and VBW = 10Hz for average
       measurement, Sweep = auto, Detector function = peak, Trace = max hold.
   For Below 1GHz
   Use the following spectrum analyzer settings:
       Span = wide enough to capture the peak level of the in-band emission and all spurious
       RBW = 100 KHz, VBW≥RBW for peak measurement, Sweep = auto, Detector function =
       peak, Trace = max hold.
Note:
       1: The resolution bandwidth and video bandwidth of test receiver/spectrum analyzer is
       120 KHz for Quasi-peak detection (QP) at frequency below 1GHz.
       2: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 3MHz for peak detection (PK) at frequency above 1GHz.
       3: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 3MHz for RMS Average ((duty cycle ˂ 98%) for Average detection (AV) at
       frequency above 1GHz, then the measurement results was added to a correction factor
       (20log(1/duty cycle).
       4: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 10Hz (duty cycle ˃ 98%) for Average detection (AV) at requencyabove1GHz




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 29 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




Limit

The radio emission outside the operating frequency band shall be at least 20 dB below that in
the 100 kHz bandwidth within the band that contains the highest level of the desired power.
Radiated emissions which fall in the restricted bands, as defined in section15.205, must comply
with the radiated emission limits specified in section 15.209.

              Frequency                 Field Strength                  Field Strength                       Detector
                 MHz                         uV/m                          dBμV/m
                 30-88                        100                             40                                 QP
                88-216                        150                            43.5                                QP
                216-960                       200                             46                                 QP
               960-1000                       500                             54                                 QP
              Above 1000                      500                             54                                 AV
              Above 1000                     5000                             74                                 PK




EMC_SZ_FR_21.00 FCC                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 30 of 34
                           Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




Spurious radiated emissions for transmitter

Transmitting spurious emission test result as below:

Low channel 2402MHz Test Result

                               Emission                                                                            Correct
   Frequency    Frequency                        Polarization          Limit        Detector         Margin                   Result
                                 Level                                                                              factor
     Band
                  MHz           dBuV/m                               dBμV/m                         dBuV/m         (dB/m)
       30-       886.13*         27.04                  H              46               QP           18.96           -15.8     Pass
    1000MHz      877.19*         31.07                  V              46               QP           14.93           -15.9     Pass
                  4804*          43.87                  H              74               PK           30.13            2.7      Pass
     1000-          --             --                   H              54               AV             --              --      Pass
   25000MHz       4804*          41.80                  V              74               PK           32.20            2.7      Pass
                    --             --                   V              54               AV             --              --      Pass


Middle channel 2440MHz Test Result

                               Emission                                                                            Correct
   Frequency    Frequency                        Polarization          Limit        Detector         Margin                   Result
                                 Level                                                                              factor
     Band
                  MHz           dBuV/m                               dBμV/m                         dBuV/m         (dB/m)
       30-          --             --                   H             43.5              QP             --              --      Pass
    1000MHz         --             --                   H              46               QP             --              --      Pass
                  4880*          47.90                  H              74               PK           26.10            5.2      Pass
     1000-          --             --                   H              54               AV             --              --      Pass
   25000MHz       4880*          48.25                  V              74               PK           25.75            5.2      Pass
                    --             --                   V              54               AV             --              --      Pass




EMC_SZ_FR_21.00 FCC                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 31 of 34
                           Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




High channel 2480MHz Test Result

                                Emission                                                                            Correct
   Frequency     Frequency                        Polarization          Limit        Detector         Margin                   Result
                                  Level                                                                              factor
     Band
                    MHz          dBuV/m                               dBμV/m                         dBuV/m         (dB/m)
       30-            --            --                   H             43.5              QP             --              --      Pass
    1000MHz           --            --                   H              46               QP             --              --      Pass
                    4960*         44.02                  H              74               PK           29.98            3.3      Pass
      1000-           --            --                   H              54               AV             --              --      Pass
    25000MHz        4960*         43.32                  V              74               PK           30.68            3.3      Pass
                      --            --                   V              54               AV             --              --      Pass


Remark:
      (1) “*” means the emission(s) appear within the restrict bands shall follow the requirement of section
          15.205.
      (2) Data of measurement within this frequency range shown “--” in the table above means the reading of
          emissions are the noise floor or attenuated more than 10dB below the permissible limits or the field
          strength is too small to be measured.
      (3) Level=Reading Level + Correction Factor
          Above 1GHz: Corrector factor = Antenna Factor + Cable Loss- Amplifier Gain
          Below 1GHz: Corrector factor = Antenna Factor + Cable Loss
          (The Reading Level is recorded by software which is not shown in the sheet)




EMC_SZ_FR_21.00 FCC                       TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 32 of 34
                            Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                 Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




10 Test Equipment List
                                           List of Test Instruments
Radiated Spurious Emission Test
         Description            Manufacturer                     Model no.                    Serial no.             Cal. due date
       Signal Analyzer       Rohde & Schwarz                      FSV40                        101031                 2019-7-6
  Trilog Super Broadband
                                Schwarzbeck                     VULB 9163                         708                     2019-7-13
        Test Antenna
        Horn Antenna         Rohde & Schwarz                   HF907                           102295                     2019-7-13
       Wideband Horn                                       QWH-SL-18-40-K-
                                  Q-PAR                                                         12827                     2019-7-12
          Antenna                                               SG
        Pre-amplifier        Rohde & Schwarz                  SCU 18                           102230                     2019-7-6
        Pre-amplifier        Rohde & Schwarz                 SCU 40A                           100432                     2019-7-6
       Fully Anechoic
                                   TDK                             8X4X4                           --                     2020-7-7
          Chamber
        Test software        Rohde & Schwarz                       EMC32                 Version 9.15.00                    N/A

RF Conducted Test
        Description            Manufacturer                      Model no.                    Serial no.             cal. due date
      Signal Analyzer        Rohde & Schwarz                      FSV40                        101030                  2019-7-6




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                  Page 33 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.19.0006.01




11 System Measurement Uncertainty

For a 95% confidence level, the measurement expanded uncertainties for defined systems, in
accordance with the recommendations of ISO 17025 were:

                               System Measurement Uncertainty
                          Test Items                           Extended Uncertainty
  Uncertainty for Radiated Spurious Emission 25MHz-             Horizontal: 4.80dB;
  3000MHz                                                        Vertical: 4.87dB;
  Uncertainty for Radiated Spurious Emission 3000MHz-           Horizontal: 4.59dB;
  18000MHz                                                       Vertical: 4.58dB;
  Uncertainty for Radiated Spurious Emission 18000MHz-          Horizontal: 5.05dB;
  40000MHz                                                       Vertical: 5.04dB;
  Uncertainty for Conducted RF test with TS 8997           RF Power Conducted: 1.16dB
                                                              Frequency test involved:
                                                                  0.6×10-7 or 1%




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 34 of 34
                         Building 12&13, Zhiheng W isdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299



Document Created: 2019-07-10 04:22:39
Document Modified: 2019-07-10 04:22:39