RF Exposure Info-TMB-1598-BS

FCC ID: OU9TMB1598-B

RF Exposure Info

Download: PDF
FCCID_3771675

                        RF exposure Estimation for TMB—995—BT

 Introduction
 TMB—1598—BS, TMB—1598—BA are Blood Pressure Monitor, which contain Bluetooth function
 inside.
 FCC ID: OU9TMB1598—B
Limit and Guidelines on Exposure to Electromagnetic Fields
According to §15.247(e)(i) and §1.1307(b)(1), systems operating under the provisions of this
section shall be operated in a manner that ensure that the public is not exposed to radio
frequency energy level in excess of the Commission‘s guideline.

According to KDB 447498 DO1 Mobile Portable RF Exposure v05r02, no SAR required if power
is lower than the flowing threshold:
The 1—g and 10—g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation
distances < 50 mm are determined by:
 [(max. power of channel, including tune—up tolerance, mW)/(min. test separation distance, mm)]
 [Vf(GHz)] < 3.0 for 1—g SAR and < 7.5 for 10—g extremity SAR, where
     * f(GHz) is the RF channel transmit frequency in GHz
     * Power and distance are rounded to the nearest mW and mm before calculation25
     * The result is rounded to one decimal place for comparison
     + 3.0 and 7.5 are referred to as the numeric thresholds in the step 2 below

The test exclusions are applicable only when the minimum test separation distance is < 50 mm
and for transmission frequencies between 100 MHz and 6 GHz. When the minimum test
separation distance is < 5 mm, a distance of 5 mm according to 5) in section 4.1 is applied to
determine SAR test exclusion.

‘Calculation method

[({max. power of channel, including tune—up tolerance, mW)/(min. test separation distance,
mm)] +[Vf(GHz)] < 3.0

Conducted Power + tune up tolerance = 1.465mW
Distance = 5 mm
f = 2.440 GHz

[1.465/5] * SQRT(2.440) =0.458
0.458 < 3.0
Therefore, excluded from SAR testing.


TUV SUD China, Guangzhou Branch

Reviewed by:                                                                           Prepared By:

 ’___-1——'——7



Tony Liu / Project Reviewer                                                       Ke‘yin Ouyang / Project               Handler
Date: 2018—03—06                                                                              Date: 2018—03—06




                         TUV SUD Certification and Testing (China) Co., Ltd. Guangzhou Branch
5F, Communication Building, 163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478



Document Created: 2018-03-06 16:44:31
Document Modified: 2018-03-06 16:44:31