User manual

FCC ID: OU9TMB1014BS

Users Manual

Download: PDF
FCCID_4068604

                                                                                                                      Version:1.0




                                                                          User Manual
                                                             Wrist Blood Pressure Monitor TMB-1014-BS
                                                                                                         Wrist Type




                                                           Thank you very much for selecting TRANSTEK Wrist Blood Pressure Monitor
Guangdong Transtek Medical Electronics Co., Ltd.
Zone B, No.105 ,Dongli Road, Torch Development District,   TMB-1014-BS.
Zhongshan,528437,Guangdong,China                           To use the monitor correctly and safely, please read the manual thoroughly.
                                                           Please keep this manual well in order to reference in future.


                                                                                                           CATALOGUE

INTRODUCTION            ............................................................................................. 2
    General Description
    Indications for Use
    Contraindications
    Measurement Principle
    Safety Information
    LCD Display Symboll
    Monitor Components
    List
BEFOREYOU START                 ...................................................................................... 13
   Installing and Replacing the Batteries
   Setting Date, Time and Measurement Unit
   Select the User ID
   Pair-up the Blood Pressure Monitor with Your Device
MEASUREMENT             .............................................................................................. 21
     Tie the Cuff
     Start the Measurement
DATA MANAGEMENT                  ......................................................................................    25
     Recalling the Records
     Deleting the Records
INFORMATION FOR USER                   ................................................................................    27
    Tips for Measurement
    Maintenance
                                         ...............................................................................   29
ABOUT BLOOD PRESSURE
   What are systolic pressure and diastolic pressure?
   What is the standard blood pressure classification?
   Irregular heartbeat detector
   Why does my blood pressure fluctuate throughout the day?
   Why do I get a different blood pressure at home compared to the hospital?
   Is the result the same if measuring on the right wrist?
TROUBLESHOOTING                 .....................................................................................      32
SPECIFICATIONS            ...........................................................................................      34
CONTACT INFORMATION                  .................................................................................     36
 FCC STATEMENT..................................................................................................           37
COMPLIED STANDARDS LIST                     ..........................................................................     38   1
EMC GUIDANCE            .............................................................................................      40


INTRODUCTION                                                                                                                                                        INTRODUCTION

     General Description                                                              Contraindications
     Thank you for selecting TRANSTEK Blood Pressure Monitor (TMB-1014-BS).           1.The device should not be used by any person who may be suspected of,or is pregnant .
     The monitor features blood pressure measurement, pulse rate measurement          2.The device is not suitable for use on patients with implanted,electrical devices,
     and the result storage. The design provides you with two years of reliable       such as cardiac pacemakers, defibrillators.
     service.
     Readings taken by the TMB-1014-BS are equivalent to those obtained by a
     trained observer using the cuff and stethoscope auscultation method.
                                                                                      Measurement Principle
     This manual contains important safety and care information, and provides         This product uses the Oscillometric Measuring Method to detect blood pressure. Before
                                                                                      every measurement, the unit establishes a “zero point” equivalent to the atmospheric
     step by step instructions for using the product.
                                                                                      pressure. Then it starts inflating the cuff. Meanwhile, the unit detects pressure oscillation
     Read the manual thoroughly before using the product.                             generated by beat-to-beat pulsatile, which is used to determine the systolic pressure and
     Features:                                                                        diastolic pressure as well as pulse rate.
      Systolic blood pressure
      Diastolic blood pressure
      Pulse rate
      60 records for one user

     Indications for Use
     The Transtek Blood Pressure Monitor is a digital monitor intended for use in
     measuring blood pressure and heartbeat rate with a wrist circumference ranging
     from 13.5cm to 21.5 cm ( about 5⅓˝-8½˝ ).
     It is intended for adult indoor use only.




 2                                                                                                                                                                                    3


INTRODUCTION                                                                                                                                                                                  INTRODUCTION

     Safety Information                                                                                           CAUTION
     The signs below might be in the user manual, labeling or other component.
     They are the requirement of standard and using.                                                   * This device is intended for adult use in homes only.
                                                                                                       * The device is not suitable for use on neonatal patients, pregnant women,patients with implanted,
              Symbol for “THE OPERATION                  Symbol for “TYPE BF APPLIED                   electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation,
                                                                                                       peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or
              GUIDE MUST BE READ”                        PARTS”
                                                                                                       people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from
                                                         Symbol for “ENVIRONMENT                       illnesses.
               Symbol for “MANUFACTURE
                                                                                                       * The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it
               DATE”                                     PROTECTION - Electrical waste
                                                                                                       on older children.
                                                         products should not be disposed of with       * The device is not intended for patient transport outside a healthcare facility.
                                                         household waste. Please recycle where         * The device is not intended for public use.
              Symbol for “MANUFACTURER”                  facilities exist. Check with your local       * This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not
                                                         authority or retailer for recycling advice”   intended for use on extremities other than the wrist or for functions other than obtaining a blood pressure
                                                                                                       measurement.
      SN      Symbol for “SERIAL NUMBER”                 Symbol for “DIRECT CURRENT”                   * Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood
                                                                                                       pressure.Do not begin or end medical treatment without asking a physician for treatment advice.
                                                         Caution: These notes must be                  * If you are taking medication,consult your physician to determine the most appropriate time to measure
               Symbol for “RECYCLE”                      observed to prevent any damage                your blood pressure. Never change a prescribed medication without consulting your physician.
                                                         to the device.                                * Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a
                                                                                                       medicine prescribed by a doctor. Consult your doctor if you have any question about your blood
               The Green Dot is the license symbol of
                                                                                                       pressure.
               a European network of industry-funded
                                                                                                       * When the device was used to measure patients who have common arrhythmias such as atrial or
               systems for recycling the packaging
                                                                                                       ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult
               materials of consumer goods.
                                                                                                       your physician about the result.
                                                                                                       * Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.
 4                                                                                                     *Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around
                                                                                                       simultaneously, because this could cause temporary loss of function of those simultaneously-used
                                                                                                       monitoring ME equipment.
                                                                                                                                                                                                                      5


INTRODUCTION                                                                                                                                                                                              INTRODUCTION

                 CAUTION                                                                                                     CAUTION
                                                                                                                   * The patient can measure data and change batteries under normal circumstances and maintain the device
    *On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff   and its accessories according to the user manual.
    immediately. Prolonged high pressure (cuff pressure > 300mmHg or constant pressure > 15mmHg for                * To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated
    more than 3 minutes) applied to the wrist may lead to an ecchymosis.                                           interference signal or electrical fast transient/burst signal.
    *Please check that operation of the device does not result in prolonged impairment of patient blood            * The blood pressure monitor and the cuff are suitable for use within the patient environment. If you are
    circulation.                                                                                                   allergic to polyester, nylon or plastic, please don't use this device.
    * When measurement, please avoid compression or restriction of the connection tubing.                          * During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and
    * The device cannot be used with HF surgical equipment at the same time.                                       found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any
    * The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically                            potential sensization or irritation reaction.
    investigated according to the requirements of ISO 81060-2:2013.                                                * If you experience discomfort during a measurement, such as pain in the wrist or other complaints, press
    * To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.                the START/STOP button to release the air immediately from the cuff. Loosen the cuff and remove it from
    * This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing     your wrist.
    inaccurate readings, the effects of this device on the fetus are unknown.                                      * If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not
    * Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.        deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the wrist and press the
    * This unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise,      START/STOP button to stop inflation.
    the patient’s wrist and fingers will become anaesthetic, swollen and even purple due to a lack of blood.       * Before use, make sure the device functions safely and is in proper working condition. Check the device,
    * When not in use, store the device in a dry room and protect it against extreme moisture, heat, lint, dust    do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause injury,
    and direct sunlight. Never place any heavy objects on the storage case.                                        improper results, or serious danger.
    * This device may be used only for the purpose described in this booklet. The manufacturer cannot be held      * Do not wash the cuff in a washing machine or dishwasher!
    liable for damage caused by incorrect application.                                                             * The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The
    *This device comprises sensitive components and must be treated with caution. Observe the storage and          typical service life is 10000 times.
    operating conditions described in this booklet.                                                                * It is recommended that the performance should be checked every 2 years and after maintenance and
    * The maximum temperature that the applied part can be achieved is 42.5℃ while the environmental               repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air
    temperature is 40℃.                                                                                            leakage (testing at least at 50mmHg and 200mmHg).
    * The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable                * Manufacturer will make available on request circuit diagrams, component part lists, descriptions,
    anesthetic mixture with air of with oxygen or nitrous oxide.                                                   calibration instructions,etc., to assist to service personnel in parts repair.
    * Warning: No servicing/maintenance while the ME equipment is in use.                                           * The operator shall not touch output of batteries and the patient simultaneously.
    * The patient is an intended operator.
6                                                                                                                                                                                                                                  7


INTRODUCTION                                                                                                                                                                                      INTRODUCTION
                                                                                                                   LCD Display Symbol
              CAUTION
    * Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to clean the
    whole unit before and after use. Don’t use any abrasive or volatile cleaners.
    * The device doesn’t need to be calibrated within two years of reliable service.
    * If you have any problems with this device, such as setting up, maintaining or using, please contact the
    SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of
    malfunctions. The device must only be serviced, repaired and opened by individuals at authorized
    sales/service centers.
    * Please report to Transtek if any unexpected operation or events occur.
    * Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small
    parts. It is dangerous or even fatal.
    * Be careful to strangulation due to cables and hoses, particularly due to excessive length.
    * At least 30 min required for ME equipment to warm from the minimum storage temperature between
    uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the
    maximum storage temperature between uses until it is ready for intended use.
    * This equipment needs to be installed and put into service in accordance with the information provided in
    the ACCOMPANYING DOCUMENTS;                                                                                   SYMBOL   DESCRIPTION                EXPLANATION
    * Wireless communications equipment such as wireless home network devices, mobile phones, cordless                     Systolic Blood Pressure    High blood pressure
    telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a
                                                                                                                           Diastolic Blood Pressure   Low blood pressure
    distance d away from the equipment. The distance d is caculated by the MANUFACTURER from the
    80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.                                  Pulse display              Pulse in beats per minute
    * Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE.
    Otherwise, it may cause damage to the unit or danger to the user/patients.                                             Low Battery                Low battery and please replace the batteries.
    * There is no luer lock connectors are used in the construction of tubing, there is a possibility that they
    might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood                Unit                       Measurement unit of blood pressure(1mmHg=0.133kPa)
    vessel.
    * Please use the device under the environment which was provided in the user manual. Otherwise, the
    performance and lifetime of the device will be impacted and reduced.
8


 INTRODUCTION                                                                                                                   INTRODUCTION


     SYMBOL   DESCRIPTION            EXPLANATION                                                  Monitor Components
              Time                   Hour:Minute (Month/Day/Year)
                                     Blood pressure monitor is detecting an
              IHB Detector           irregular heartbeat during measurement.
              Bluet ooth             Successful Bluetooth Connection
              Error                  Data communication has failed                                                     MEM/UP
                                                                                                       SYSTOLIC
                                                                                                                       BUTTON
              User ID                The selected User ID

              Memory                 Indicate it is in the memory mode                                DIASTOLIC

              Heartbeat              Blood pressure monitor is detecting a heartbeat during                            START/STOP
                                                                                                     PULSE RATE        BUTTON
                                     measurement.
              Average value          The average value of the latest three groups bood pressure     LCD DISPLAY        TIME
                                     value                                                                             SET/DOWN
                                                                                                                       BUTTON
              Blood pressure level   Indicate the blood pressure level
              indicator




10                                                                                                                                             11


 INTRODUCTION                                                                                                                                   BEFORE YOU START


             Component List
                                                                            Installing and Replacing the Batteries
             pressure measuring system:                                     • Open the battery door.
                                                   CUFF
             1. PCBA;                              (Type BF applied part)   • Insert the batteries according to the polarity indications.
             2. Air Pipe;                                                     (Always select the authorized / specified battery:
             3. Pump;                                                          Two AAA-size batteries).
             4. Valve;                                                      • Replacing the battery door.
             5. Cuff.




                                                   BATTERY
                                                   COMPARTMENT




      List
     1) Wrist Blood Pressure Monitor TMB-1014-BS
     2) 2×AAA Batteries
     3) User manual
                                                                              The typical service life of the new and unused batteries is 138
                                                                              measurements for the operation time is 60s.

12                                                                                                                                                             13


 BEFORE YOU START                                                                                                                                              BEFORE YOU START


                                                                                           Setting Date and Time
      Replace the batteries whenever the below happen
                                                                                           Please proceed to time setting before your initial use so as to
       The                   shows                                                         ensure each piece of record are labled with a time stamp.
       The display is dim.                                                                 (Year Range: 2018-2058; Time Format: 12 Hours)
        The display does not light up                                                      1.When the monitor is off, hold pressing “SET” button for about 3
                                                                                           seconds to enter into setting mode. The blinking numeral                     MEM

                                                                                           represents [HOUR].
        CAUTION
     Do not use new and used batteries together.
     Do not use different types of batteries together.
     Do not dispose the batteries in fire. Batteries may explode or leak.
                                                                                                                                                                        SET
     Remove batteries if the device is not likely to be used for some time.
     Worn batteries are harmful to the environment. Do not dispose with daily garbage.
     Remove the old batteries from the device following your local recycling guidelines.


                                                                                            2. Press “MEM” button to change the [HOUR] . Each press will                MEM


                                                                                            increase the numeral by one in a cycling manner.




                                                                                                                                                                        SET




14                                                                                                                                                                            15


BEFORE YOU START                                                                                                                                                       BEFORE YOU START


     3.When you get the right hour,Press “SET” button again to confirm [HOUR]. Then it will   5.Repeat step 2 and 3 to confirm [MONTH], [DAY] and [YEAR].
     turn to the next step.
                                                                                                                    MEM                           MEM                           MEM
                                                                                    MEM




                                                                                                                    SET                           SET                            SET
                                                                                    SET




     4.Repeat step 2 and 3 to confirm [MINUTE].
                                                                                              6.After confirming [YEAR], the LCD will display “donE” then the monitor will turn off .

                                                                                    MEM




                                                                                    SET




16                                                                                                                                                                                      17


 BEFORE YOU START                                                                                                                                                BEFORE YOU START

     Select the User ID                                                                    Pair-up the Blood Pressure Monitor
     Before you start the measurement, please select the suitable user ID first.           with Your Device
     1.When the blood pressure monitor is off, press the                                   1.Turn on Bluetooth and the app. Make sure both are ON when pair-up is proceeding.
     MEM button for 3 seconds, then the user ID will blink.                          MEM

                                                                                           2.When the monitor is OFF, press and hold the START/STOP button to start pair-up. The
                                                                                           symbol “    + ”and user ID will be shown on the LCD alternatively, indicating pair-up is
                                                                                           proceeding.


                                                                                     SET



     2. Press the MEM button shortly to change the user
     ID between User 1 and User 2 .                                                  MEM




                                                                                           If SUCCEED, the symbol “        +    ”and user ID will be shown   on the LCD.
                                                                                     SET




     3. After selecting the suitable user ID, press SET button to confirm then the
     LCD will turn off.



18                                                                                                                                                                                    19


 BEFORE YOU START                                                                                                                                                 MEASUREMENT
                                                                      Tie the Cuff
                                                                      1. Remove all accessories (watch, bracelet,etc) from your wrist. If your physician has diagnosed you with
     If FAIL, the symbol “    +         ” will be shown on the LCD.      poor circulation in your wrist, use the other one.
                                                                      2. Roll or push up your sleeve to expose the skin.
                                                                      3. Apply the cuff to your wrist with your palm facing up.
                                                                      4. Position the edge of the cuff about 1cm~1.5cm from wrist joints.
                                                                      5. Fasten the wrist cuff around your wrist, leaving no extra room between the cuff and your skin. If the cuff is
                                                                         too loose, the measurement will not be accurate.
                                                                      6. Sit comfortably with your tested wrist resting on a flat surface. Place your elbow on a table so that the cuff is
                                                                         at the same level as your heart. Turn your palm upwards. Sit upright in a chair, and take 5-6 deep breaths.
                                                                      7. Patients with Hypertension:
                                                                         The middle of the cuff should be at the level of the right atrium of the heart; Before starting measurement,
                                                                         please sit comfortably with legs uncrossed, feet flat on the floor, back and wrist supported.
                                                                         Rest for 5 minutes before measuring.
     3.The monitor will turn off after Pair-up process is complete.      Wait at least 3 minutes between measurements. This allows your blood
                                                                         circulation to recover.
                                                                         Take the measurement in a silent room.
        Bluetooth Module No.: LS51802
                                                                         The patient must relax as much as possible and do not move and talk
        RF Frequency Range: 2402 MHz to 2480 MHz                         during the measurement procedure.
        Output Power Range: ≤4dBm                                        The cuff should maintain at the same level as the right atrium of the
                                                                         heart.
        Supply Voltage: 2 V to 3.6V
                                                                         Do not cross your legs and keep your feet on the ground.
        Transmitting Distance: 10 meters                                 Keep your back against the backrest of the chair.
                                                                         For a meaningful comparison, try to measure under similar conditions.
                                                                         For example, take daily measurements at approximately the same time,
                                                                         on the same wrist, or as directed by a physician.



20                                                                                                                                                                                       21


 MEASUREMENT                                                                                                                                              MEASUREMENT

      Start the Measurement
     1.When the monitor is off, press the START/STOP button                     2.This device will proceed to data transmission after measurement. The Bluetooth symbol
                                                              START             will blink.
     to turn on the monitor, and it will finish the whole
     measurement.(Take user 1 for example.)
                                                              STOP


     Adjust to zero.                                 Inflating and measuring.




                                                                                3.If the data transmission succeeds, the Bluetooth symbol   will light on , the LCD will display
                                                                                as pictured to the right,then the device will turn off.



     Display and save the measuring result.




22                                                                                                                                                                             23


 MEASUREMENT                                                                                                                                                                        DATA MANAGEMENT


     If the data transmission fails,the LCD will display “        ”                                          Recalling the Records                                                           MEM

     as pictured to the right, then the device will turn off.
                                                                                                              1. When the monitor is off, please press the “MEM” to show
                                                                                                              the average value of the latest three records.If the records
                                                                                                              are less than three groups,it will display the latest record first.

                                                                                                                                                                                             SET




               CAUTION
                                                                                                                                                                                             MEM
          Interference may occur in the vicinity of equipment marked with the following symbol     . And
          TMB-1014-BS may interfering vicinity electrical equipment.                                          2. Press “MEM” button or “SET” button to rotate the history
                                                                                                                                                                                             UP
          Sensitive people, including pregnant women pre-eclamptic and those who implanted medical            records.“MEM” to go forward;“SET” to go backward.
          electronic instruments, should avoid using the unit whenever possible.
          Keep the monitor at least 20 centimeters away from the human body (especially the head) when the
          data transmission is proceeding after measurement.
          To enable the data transmission function, this product should be paired to Bluetooth end at 2.4                                                                                    SET
          GHz.
          How to mitigate possible interference?                                                                                                                                            DOWN
       1. The range between the device and BT end should be reasonably close, from 1 meter to 10 meters.           CAUTION
          Please ensure no obstacles between the device and BT end so as to obtain quality connection and
          to lower the RF output range.                                                                         The most recent record (1) is shown first. Each new measurement is
       2. To avoid interference, other electronic devices (particularly those with wireless transmission /   assigned to the first (1) record. All other records are pushed back one
          Transmitter) should be kept at least 1 meter away from the monitor.
                                                                                                             digit (e.g., 2 becomes 3, and so on), and the last record (60) is dropped
                                                                                                             from the list.
24                                                                                                                                                                                                 25


 DATA MANAGEMENT                                                                                                                         INFORMATION FOR USER


     Deleting the Records                                                     Tips for Measurement
                 ABOUT
     If you did not     BLOOD
                    get the    PRESSURE
                            correct measurement, you can delete all results   Measurements may be inaccurate if taken in the following circumstances.
     by following steps below.
      1.Hold pressing “MEM” for 3 seconds when the monitor is
      in the memory recall mode ,the “dEL ALL” will blink on the
      display.


     2.Press “SET” to confirm deleting and the monitor will display                  Within 1 hour          Immediate measurement
                                                                                                                                           Within 20 minutes
                                                                                after dinner or drinking    after tea, coffee, smoking
     “dEL donE” then turn off.                                                                                                             after taking a bath

                                                                         MM
     3.If you don’t want to delete the records,
     press START/STOP” to escape.                 START
                                                  STOP


     4. If there is no record, the right display will be showed.        SET



                                                                                 When talking or
                                                                                                            In a very cold environment      When you want to
                                                                                moving your fingers
                                                                                                                                             discharge urine



26                                                                                                                                                               27


 INFORMATION FOR USER                                                                                                                                                                                 ABOUT BLOOD PRESSURE


       Maintenance                                                                                          What are systolic pressure and diastolic pressure?
                                                                                                            When ventricles contract and pump blood out of the heart, blood pressure reaches its maximum
          To obtain the best performance, please follow the instructions below.
                                                                                                            value, the highest pressure in the cycle is known as systolic pressure. When the heart relaxes
                                                                                                            between heartbeats, the lowest blood pressure is diastolic pressure.

                                                                                                            What is the standard blood pressure classification?
                                                                                                            The chart on the right is the        This chart reflects blood pressure categories defined by American Heart Association.
                                                                                                            standard blood pressure
                                                                                                                                                                                             Systolic                        Diastolic
                                                                                                            classification published by         Blood Pressure Category
                                                                                                                                                                                             mmHg (upper#)                   mmHg (lower#)
                                                                                                            American Heart Association                                              No
                                                                                                                                                                                  symbol
           Put in a dry place and       Avoid immersing it in the water.     Avoid shaking and collision.   (AHA).                                         Normal                displayed    less than 120         and      less than 80
            avoid the sunshine          Clean it with a dry cloth in case.                                                                               Elevated                             120-129              and       less than 80
                       TROUBLESHOOTING
                                                                                                            AHA Home Guideline for                 High Blood Pressure                        130-139               or          80-89
                                                                                                            Upper Limit of Normal BP              (Hypertension) Stage 1

                                                                                                              SYS      135 mm Hg                   High Blood Pressure
                                                                                                                                                  (Hypertension) Stage 2
                                                                                                                                                                                              140 or higher         or       90 or higher

                                                                                                              DIA       85 mm Hg                     Hypertensive Crisis                     Higher than 180      and/or
                                                                                                                                                  (Emergency care needed)                                                    Higher than 120



       Avoid dusty environment and        Use the slightly damp cloth          Avoid washing the cuff
                                                                                                                    CAUTION
     unstable temperature surrounding         to remove the dirt.                                             Only a physician can tell your normal BP range. Please contact a physician if your
                                                                                                              measuring result falls out of the range. Please note that only a physician can tell
                                                                                                              whether your blood pressure value has reached a dangerous point.


28                                                                                                                                                                                                                                             29


 ABOUT BLOOD PRESSURE                                                                                                                                                           ABOUT BLOOD PRESSURE

       Irregular Heartbeat Detector                                                                          Why does my blood pressure fluctuate throughout the day?
     An irregular heartbeat is detected when a heartbeat rhythm varies while the device is measuring         1. Individual blood pressure varies multiple times everyday. It is also
     systolic pressure and diastolic pressure. During each measurement, blood pressure monitor will          affected by the way you tie your cuff and your measurement position, so
     keep a record of all the pulse intervals and calculate the average value of them. If there are two or   please take the measurement under the same conditions.
     more pulse intervals , the difference between each interval and the average is more than the            2.If the person takes medicine, the pressure will vary more.
     average value of ±25% , or there are four or more pulse intervals ,the difference between each
     interval and the average is more than the average value of ±15%, then the irregular heartbeat           3.Wait at least 3 minutes for another measurement.
     symbol will appear on the display with the measurement result.
                                                                                                             Why do I get a different blood
                  CAUTION                                                                                    pressure at home compared
         The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular
         heartbeat was detected during measurement. Usually this is NOT a cause for concern.                 to the hospital?
         However, if the symbol appears often, we recommend you seek medical advice. Please                                                                                 What you need to pay attention
                                                                                                             The blood pressure is different even throughout the day        to when you measure your blood
         note that the device does not replace a cardiac examination, but serves to detect pulse
                                                                                                             due to weather, emotion, exercise etc, Also, there is the      pressure at home:
         irregularities at an early stage.
                                                                                                             “white coat” effect, which means blood pressure usually
                                                                                                             increases in clinical settings.                                If the cuff is tied properly.
                                                                                                                                                                            If the cuff is too tight or too loose.
                                                                                                             Is the result the same if                                      If the cuff is tied on the wrist.
                                                                                                             measuring on the right                                         If you feel anxious.
                                                                                                                                                                            Taking 2-3 deep breaths before
                                                                                                             wrist?                                                         beginning will be better for
                                                                                                             It is ok for both wrists, but there will be                    measuring.
                                                                                                             some different results for different                           Advice: Relax yourself for 4-5
                                                                                                             people. We suggest you measure the                             minutes until you calm down.
                                                                                                             same wrist every time.
30                                                                                                                                                                                                               31


 TROUBLESHOOTING                                                                                                                                                                  TROUBLESHOOTING

   This section includes a list of error messages and frequently
 asked questions for problems you may encounter with your blood                                                PROBLEM SYMPTOM             CHECK THIS           REMEDY
 pressure monitor. If the products not operating as you think it                                                                                                Make sure that phone’s Bluetooth is on or
                                                                                                                          Error +         Data communication
 should, check here before arranging for servicing.                                                             Error                     has failed            within the distance range
                                                                                                                          shows
                                                                                                                message
                                                                                                                          EExx,shows on   A calibration error   Retake the measurement.If the problem
     PROBLEM SYMPTOM                  CHECK THIS                  REMEDY                                                  the display.    occurred.             persists,contact the retailer or our customer
                                     Batteries are exhausted.     Replace with new batteries                                                                    service department for further assistance.
                  Display is dim or                                                                                                                             Refer to the warranty for contact information
      No power                       Batteries are inserted       Insert the batteries correctly                                                                and return instructions.
                  will not light up.
                                     incorrectly.
                                                                                                                                                                Relax for a moment.
      Low               Show on                                                                                Warning                    Out of measurement    Refasten the cuff and then
                                    Batteries are low.            Replace with new batteries                              “out ” shows
      batteries         the display                                                                            message                    range                 measure again. If the
                                                                                                                                                                problem persists, contact
                  E 01 shows                                      Refasten the cuff and then measure again.                                                     your physician.
                                     The cuff is not secure.
                                                                  Refasten the cuff and then measure again.
                  E 02 shows         The cuff is very tight

                                     The pressure of the
                  E 03 shows                                      Relax for a moment and then measure again.
                                     cuff is excess.
                                    The monitor detected motion, movement can affect the measurement.
                  E 10 or           talking,or the pulse is too poor Relax for a moment and then measure
                  E 11 shows        while measuring.                 again.

                                     The measurement process      Loosen the clothing on the arm and then
      Error       E 20 shows         does not detect the pulse    measure again.
                                     signal.
      message
                  E 21 shows         Measure incorrectly.         Relax for a moment and then measure again.

32                                                                                                                                                                                                              33


             EMC GUIDANCE
 SPECIFICATIONS                                                                                                                                                               SPECIFICATIONS


          Power supply          Battery powered mode: 2*AAA batteries
                                                                                               Mode of operation      Continuous operation
          Display mode          Digital LCD V.A.36mmx41mm
       Measurement mode         Oscillographic testing mode                                   Degree of protection    Type BF applied part

                                Rated cuff pressure:                                         Device Classification    Internally Powered ME Equipment
                                0mmHg~299mmHg(0kPa ~ 39.9kPa)
       Measurement range        Measurement pressure:                                                                 IP22: The first number 2: Protected against solid foreign
                                SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)                                                  objects of 12,5mm Ф and greater. The second number:
                                DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)                                                  Protected against vertically falling water drops when enclosure
                                Pulse value: (40-199)beat/minute                                IP Classification     titled up to 15º. Vertically falling drops shall have no harmful
                                                                                                                      effects when the enclosure is titled at any angle up to 15º on
            Accuracy            Pressure:5°C-40°C within±3mmHg(0.4kPa)
                                                                                                                      either side of the vertical.
                                Pulse value:±5%
                                A temperature range of :+5°C to +40°C                          Software Version       A01
                                A relative humidity range of 15% to 90%, non-condensing,
        Working condition       but not requiring a water vapour partial pressure greater
                                than 50 hPa                                                 WARNING: No modification of this equipment is allowed.
                                An atmospheric pressure range of : 700 hPa to 1060 hPa
                                Temperature:-20°C to +60°C
     Storage & transportation   A relative humidity range of ≤ 93%, non-condensing,
            condition           at a water vapour pressure up to 50hPa
     Measurement perimeter
                                About 13.5cm-21.5cm
        of the upper arm
             Weight             Approx.120g(Excluding the batteries)
       External dimensions      Approx.80mm×65mm×22mm(Excluding the cuff)
           Attachment           2*AAA batteries,user manual

34                                                                                                                                                                                        35


 CONTACT INFORMATION                                                                                                                                                              FCC Statement


     Contact Information                                                                              FCC Statement
                                                                                                   FCC ID:OU9TMB1014BS
     For more information about our products, please visit www.transtek.cn.you can get customer
     service, usual problems and customer download, transtek will serve you anytime.              This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
                                                                                                  conditions: (1) this device may not cause harmful interference, and (2) this device must accept any
                                                                                                  interference received, including interference that may cause undesired operation. Caution: The user is
       Manufactured by: Guangdong Transtek Medical Electronics Co., Ltd.                          cautioned that changes or modifications not expressly approved by the party responsible for
       Company: Guangdong Transtek Medical Electronics Co., Ltd.                                  compliance could void the user's authority to operate the equipment.
       Address: Zone B, No.105 ,Dongli Road, Torch Development District,                          NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device,
                Zhongshan,528437,Guangdong,China                                                  pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
       Authorized European Representative:                                                        against harmful interference in a residential installation. This equipment generates, uses and can
       Company: MDSS - Medical Device Safety Service GmbH                                         radiate radio frequency energy and, if not installed and used in accordance with the instructions, may
       Address: Schiffgraben 41, 30175 Hannover, Germany                                          cause harmful interference to radio communications. However, there is no guarantee that interference
                                                                                                  will not occur in a particular installation. If this equipment does cause harmful interference to radio or
                                                                                                  television reception, which can be determined by turning the equipment off and on, the user is
                                                                                                  encouraged to try to correct the interference by one or more of the following measures:
                                                                                                  -- Reorient or relocate the receiving antenna.
                                                                                                  -- Increase the separation between the equipment and receiver.
                                                                                                  -- Connect the equipment into an outlet on a circuit different from that to
                                                                                                  which the receiver is connected.
                                                                                                  -- Consult the dealer or an experienced radio/TV technician for help.
                                                                                                  FCC Radiation Exposure Statement:
                                                                                                  This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.
                                                                                                  This transmitter must not be co-located or operating in conjunction with any other antenna or
                                                                                                  transmitter.

36                                                                                                                                                                                                        37


COMPLIED STANDARDS LIST                                                                                                                                                COMPLIED STANDARDS LIST

        Complied Standards List
                                                                                                                                EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers -
                                                                                                                              Part 3: Supplementary requirements for electro-mechanical blood
                               EN ISO 14971:2012 / ISO 14971:2007 Medical devices -
       Risk management        Application of risk management to medical devices
                                                                                                                              pressure measuring systems
                                                                                                                                IEC 80601-2-30:2009+A1:2013 Medical electrical equipment-
                                                                                                                              Part 2-30: Particular requirements for the basic safety and
                                EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices.
                                                                                                                              essential performance of automated non-invasive
                              Symbols to be used with medical device labels, labelling and
       Labeling               information to be supplied. Part 1 : General requirements
                                                                                                                              sphygmomanometers

                                                                                                                                EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test
                               EN 1041:2008 Information supplied by the manufacturer of                                       procedures to determine the overall system accuracy of automated
       User manual            medical devices                                                        Clinical investigation   non-invasive sphygmomanometers
                                                                                                                                ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2:
                                EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical
                                                                                                                              Clinical validation of automated measurement type
                              electrical equipment - Part 1: General requirements for basic safety
                              and essential performance                                                                         EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013
       General Requirements     EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical                                    Medical electrical equipment - Part 1-6: General requirements for
       for Safety             equipment - Part 1-11: General requirements for basic safety and       Usability                basic safety and essential performance - Collateral standard:
                              essential performance - Collateral standard: Requirements for                                   Usability
                              medical electrical equipment and medical electrical systems used in                               IEC 62366-1:2015 Medical devices - Part 1: Application of
                                                                                                                              usability engineering to medical devices
                              the home healthcare environment

                                EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical
                                                                                                     Software life-cycle        EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015 Medical
                                                                                                                              device software - Software life-cycle processes
       Electromagnetic        equipment - Part 1-2: General requirements for basic safety and        processes
       compatibility          essential performance - Collateral standard: Electromagnetic                                     ISO 10993-1:2009 Biological evaluation of medical devices-
                              disturbances - Requirements and tests                                                           Part 1: Evaluation and testing within a risk management process
                                                                                                                               ISO 10993-5:2009 Biological evaluation of medical devices -
       Performance              EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part                                     Part 5: Tests for in vitro cytotoxicity
                              1: Requirements and test methods for non-automated measurement         Bio-compatibility         ISO 10993-10:2010 Biological evaluation of medical devices -
       requirements           type                                                                                            Part 10: Tests for irritation and skin sensitization
  38                                                                                                                                                                                              39


EMC GUIDANCE                                                                                                                                                           EMC GUIDANCE

                                                                                 Table 1
     EMC Guidance                                                                          Guidance and manufacturer’s declaration – electromagnetic emissions

 1)This product needs special precautions regarding EMC and needs to be
 installed and put into service according to the EMC information provided,        The device is intended for use in the electromagnetic environment specified below. The customer or the
 and this unit can be affected by portable and mobile RF communications           user of the device should assure that it is used in such an environment.

 equipment.                                                                       Emissions test                Compliance            Electromagnetic environment - guidance
 2)* Do not use a mobile phone or other devices that emit electromagnetic
 fields, near the unit. This may result in incorrect operation of the unit.
                                                                                  RF emissions                                       The device uses RF energy only for its internal
 3)Caution: This unit has been thoroughly tested and inspected to assure                                        Group 1              function. Therefore, its RF emissions are very low
                                                                                  CISPR 11
 proper performance and operation!                                                                                                   and are not likely to cause any interference in nearby
 4)* Caution: This machine should not be used adjacent to or stacked with                                                            electronic equipment.

 other equipment and that if adjacent or stacked use is necessary, this          RF emissions
 machine should be observed to verify normal operation in the configuration in   CISPR 11
                                                                                                                Class B              The device is suitable for use in all establishments,
 which it will be used.                                                                                                              other than domestic and those directly connected to
                                                                                                                                     the public low-voltage power supply network that
                                                                                  Harmonic emissions                                 supplies buildings used for domestic purposes.
                                                                                                                Class A
                                                                                  IEC 61000-3-2

                                                                                 Voltage fluctuations/
                                                                                 flicker emissions              Complies
                                                                                 IEC 61000-3-3

40                                                                                                                                                                                            41


EMC GUIDANCE                                                                                                                                                                                                       EMC GUIDANCE

 Table 2
                       Guidance and manufacturer’s declaration – electromagnetic immunity                              Voltage dips,           0%UT; 0.5 cycle             0% UT ; 0.5 cycle         Mains power quality should be
                                                                                                                       short interruptions     At 0°, 45°, 90°, 135°,      At 0°, 45°, 90°, 135°,    that of a typical commercial or
     The device is intended for use in the electromagnetic environment specified below. The customer or the            and voltage             180°,225°,270° and 315°     180°,225°,270° and        hospital environment.
     user of the device should assure that it is used in such an environment.                                          variations on           0%UT ; 1 cycle              315°
                                                                                                                       power supply            and                         0% UT ; 1 cycle
                                                                                                                       input lines             70%UT ; 25/30 cycles        and
                                                                                  Electromagnetic                                              Single phase: at 0°         70% UT ; 25/30 cycles
     Immunity test            IEC 60601 test level        Compliance level                                             IEC 61000-4-11
                                                                                  environment - guidance                                       0% UT ; 300 cycle           Single phase: at 0°
                                                                                                                                                                           0% UT ;300 cycle
                                                                                  Floors should be wood, concrete
     Electrostatic                                                                or ceramic tile. If floors are       Power frequency                                                               Power frequency magnetic fields
                              ±8 kV contact                                                                                                                                                          should be at levels characteristic
     discharge (ESD)                                      ±8 kV contact           covered with synthetic material,     (50Hz/60Hz)
                              ±15 kV air                                          the relative humidity should be at   magnetic field                  30 A/m                 30 A/m                 of a typical location in a typical
     IEC 61000-4-2                                        ±15 kV air                                                                                                                                 commercial or hospital
                                                                                  least 30%.                                                           50Hz/60Hz              50Hz/60Hz
                                                                                                                       IEC 61000-4-8                                                                 environment.
     Electrical fast         power supply lines:         power supply lines:
                                                                                  Mains power quality should be        NOTE       UT is the a.c. mains voltage prior to application of the test level.
     transient/burst         ±2 kV                       ±2 kV                    that of a typical commercial or
     IEC 61000-4-4           input/output lines:                                  hospital environment.
                             ±1 kV

     Surge                   line(s) to line(s): ±1 kV   line(s) to line(s): ±1 kV Mains power quality should be
                             line(s) to earth: ±2 kV
     IEC61000-4-5                                                                  that of a typical commercial or
                             100 kHz repetition          100 kHz repetition        hospital environment.
                             frequency                   frequency



42                                                                                                                                                                                                                                        43


 EMC
 TableGUIDANCE
      5                                                                                                                                                                                                    EMC GUIDANCE

     Table 3

       Guidance and manufacturer’s declaration – electromagnetic immunity                                                                                                                           determined by an
                                                                                                                                                                                                    electromagnetic site
     The device is intended for use in the electromagnetic environment specified below. The customer or the                                                                                         survey, a should be less
     user of the device should assure that it is used in such an environment.                                                                                                                       than the compliance levelb
                                                                                                                                                                                                    in each frequency range.
     Immunity test      IEC 60601           Compliance       Electromagnetic environment - guidance                                                                                                 Interference may occur in
                        Test level          level                                                                                                                                                   the vicinity of equipment
                                                                                                                                                                                                    marked with the following
      Conducted RF       150 kHz to         150 kHz to       Portable and mobile RF communications                                                                                                  symbol:
      IEC 61000-4-6      80 MHz:            80 MHz:          equipment should be used no closer to any part
                         3 Vrms             3 Vrms           of the device, including cables, than the
                         6Vrms (in ISM      6Vrms (in ISM    recommended separation distance calculated
                         and amateur        and amateur      from the equation appropriate for the frequency
                         radio bands)       radio bands)     of the transmitter.                                  NOTE 1      At 80 MHz and 800 MHz, the higher frequency range applies.
                         80% Am at 1kHz     80% Am at 1kHz   Recommended separation distances:
                                                                                                                  NOTE 2      These guidelines may not apply in all situations. Electromagnetic propagation is affected by
                                                             d=0.35 P ;
                                                                                                                  absorption and reflection from structures, objects and people.
                                                             d=1.2 P
                                                                                                                   a Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
     Radiated RF        10V/m, 80% Am      10V/m, 80% Am     80 MHz to 800 MHz: where, P is the maximum
     IEC 61000-4-3      at 1kHz            at 1kHz           d=1.2 P                 output power rating of the      telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
                                                             800 MHz to 2.7 GHz: transmitter in watts (W)            cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
                                                             d=2.3 P                 according to the                fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
                                                                                     transmitter manufacturer,       strength in the location in which the device is used exceeds the applicable RF compliance level
                                                                                     d is the recommended            above, the device should be observed to verify normal operation. If abnormal performance is
                                                                                     separation distance in          observed, additional measures may be necessary, such as re-orienting or relocating the device.
                                                                                     meters (m). Field             b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
                                                                                     strengths from fixed RF
                                                                                     transmitters, as
                                                                                                                                                                                                                                 45
44


 Table 5
 EMC GUIDANCE                                                                                                                                                                                            EMC GUIDANCE

     Table 4

      Recommended separation distances between portable and mobile RF communications                                            Guidance and manufacturer’s declaration - electromagnetic immunity
      equipment and the device.

      The device is intended for use in an electromagnetic environment in which radiated RF disturbances are      The device is intended for use in the electromagnetic environment specified below. The customer or the
      controlled. The customer or the user of the device can help prevent electromagnetic interference by         user of the device, should assure that it is used in such an environment.
      maintaining a minimum distance between portable and mobile RF communications equipment
      (transmittters) and the device as recommended below, according to the maximum output power of the          Radiated RF Test        Band a)      Service a)      Modulation b) Modulation b)    Distance (m) IMMUNITY
                                                                                                                 IEC61000-4-3 Frequency (MHz)                                       (W)                             TEST
      communications equipment.                                                                                                                                                                                     LEVEL
                                                                                                                 (Test          (MHz)
                                                                                                                                                                                                                    (V/m)
       Rated maximum output         Separation distance according to frequency of transmitter (m)                specifications
        power of transmitter                                                                                     for              385   380-390       TETRA           Pulse
               (W)                 150 kHz to 80 MHz        80 MHz to 800 MHz         800 MHz to 2.7 GHz         ENCLOSURE                            400             modulation b)          1.8         0.3           27
                                        d = 3.5 P                d = 1.2 P                  d = 2.3 P            PORT                                                 18Hz
                                                                                                                 IMMUNITY to
                                                                                                                                                      GMRS 460,       FM c) ± 5kHz
                0.01                   0.12                         0.12                        0.23             RF wireless                                                                 2           0.3           28
                                                                                                                                  450   380-390       FRS 460         deviation 1kHz
                                                                                                                 communicatio
                 0.1                   0.37                         0.38                        0.73                                                                  sine
                                                                                                                 ns equipment)
                  1                    1.2                          1.2                         2.3                               710   704-787       LTE Band        Pulse
                                                                                                                                  745                 13,             modulation b)      0.2             0.3           9
                 10                    3.8                          3.8                         7.3                                                   17              217Hz
                                                                                                                                  780
                 100                   12                           12                          23
                                                                                                                                                      GSM             Pulse
                                                                                                                                  810       800-960
      For transmitters rated at a maximum output power not listed above, the recommended separation                                                   800/900,        modulation b)      2               0.3           28
                                                                                                                                                      TETRA 800,      18Hz
      distance d in metres (m) can be estimated using the equation applicable to the frequency of the                             870
                                                                                                                                                      iDEN 820,
      transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the                                        CDMA 850,
      transmitter manufacturer.                                                                                                   930                 LTE Band 5
      NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
      NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
46    absorption and reflection from structures, objects and people.                                                                                                                                                        47


 Table 5
 EMC GUIDANCE


                      1720     1700-      GSM 1800;        Pulse
                                          CDMA 1900;                        2               0.3           28
                               1990                        modulation b)
                                          GSM 1900;        217Hz
                      1845
                                          DECT;
                                          LTE Band 1, 3,
                      1970                4,25; UMTS

                      2450     2400-      Bluetooth,       Pulse             2               0.3           28
                               2570       WLAN,            modulation b)
                                          802.11           217 Hz
                                          b/g/n, RFID
                                          2450, LTE
                                          Band 7
                      5240     5100-      WLAN             Pulse             0.2             0.3           9
                               5800       802.11           modulation b)
                      5240                a/n              217 Hz
                      5785

     NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and
     the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC
     61000-4-3.
     a) For some services, only the uplink frequencies are included.
     b) The carrier shall be modulated using a 50% duty cycle square wave signal.
     c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
     represent actual modulation, it would be worst case.

     The MANUFACTURER should consider reducing the minimum separation distance, based on RISK
     MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum
     separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated
     using the following equation: E=

     Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY
48   TEST LEVEL in V/m.                                                                                           47



Document Created: 2018-10-30 16:31:35
Document Modified: 2018-10-30 16:31:35

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC