                                                               FCC - TEST REPORT



Report Number                                     :      64.790.19.01904.01                             Date of Issue:               2019-07-26


Model                                             :      SBM 69

Product Type                                      :      Blood Pressure Monitor

Applicant                                         :      Guangdong Transtek Medical Electronics Co.,Ltd

Address                                           :      Zone A, No.105, Dongli Road, Torch Development District, 528437

                                                  :      Zhongshan, Guangdong, China

Production Facility                               :      Guangdong Transtek Medical Electronics Co.,Ltd

Address                                           :      Zone B, No.105, Dongli Road, Torch Development District, 528437

                                                  :      Zhongshan, Guangdong, China




Test Result                                       :      ◼ Positive                       Negative




Total pages including
Appendices                                        :                  31

TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch is a subcontractor to TÜV SÜD Product Service GmbH according to the principles outlined
in ISO 17025.

TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch reports apply only to the specific samples tested under stated test conditions. Construction
of the actual test samples has been documented. It is the manufacturer's responsibility to assure that additional production units of this model are manufactured with
identical electrical and mechanical components. The manufacturer/importer is responsible to the Competent Authorities in Europe for any modifications made to the
production units which result in non-compliance to the relevant regulations. TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch shall have no
liability for any deductions, inferences or generalizations drawn by the client or others from TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
issued reports.

This report is the confidential property of the client. As a mutual protection to our clients, the public and ourselves, extracts from the test report shall not be
reproduced except in full without our written approval.




Report Number: 64.790.19.01904.01                                                                                                                    Page 1 of 31
                                           TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
             5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


1 Table of Contents
1     Table of Contents ................................................................................................................................ 2
2     Details about the Test Laboratory .................................................................................................... 3
3     Description of the Equipment Under Test ....................................................................................... 4
4     Summary of Test Standards .............................................................................................................. 5
5     Summary of Test Results ................................................................................................................... 6
6     General Remarks ................................................................................................................................ 7
7     Test Setups .......................................................................................................................................... 8
8     Systems test configuration ................................................................................................................. 9
9     Technical Requirement .................................................................................................................... 10
    9.1      Conducted peak output power ................................................................................................. 10
    9.2      Power spectral density .............................................................................................................. 13
    9.3      6 dB Bandwidth and 99% Occupied Bandwidth.................................................................... 16
    9.4      Spurious RF conducted emissions.......................................................................................... 20
    9.5      Band edge ................................................................................................................................... 24
    9.6      Spurious radiated emissions for transmitter .......................................................................... 26
10        Test Equipment List....................................................................................................................... 30
11        System Measurement Uncertainty .............................................................................................. 31




Report Number: 64.790.19.01904.01                                                                                                      Page 2 of 31
                                      TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
           5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


2 Details about the Test Laboratory
Details about the Test Laboratory
Test Site 1

Company name:              TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
                           Building 12 & 13, Zhiheng Wisdomland Business Park, Nantou Checkpoint
                           Road 2, Nanshan District
                           Shenzhen 518052
                           P.R. China
Telephone:                 86 755 8828 6998
Fax:                       86 755 828 5299

FCC Registration           514049
No.:




Report Number: 64.790.19.01904.01                                                                                         Page 3 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


3 Description of the Equipment Under Test

Product:                                  Blood Pressure Monitor

Model no.:                                SBM 69
FCC ID:                                   OU9SBM69-B01
Options and accessories:                  N/A

Rating:                                   6VDC (4×1.5V AAA batteries)

RF Transmission                           2402MHz-2480MHz
Frequency:
No. of Operated Channel:                  40

Modulation:                               GFSK

Antenna Type:                             Internal Antenna

Antenna Gain:                             0 dBi

Description of the EUT:                   The Equipment Under Test (EUT) SBM 69 is Blood Pressure
                                          Monitor with Bluetooth operated at 2.402-2.48GHz




Report Number: 64.790.19.01904.01                                                                                            Page 4 of 31
                                   TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
          5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


4 Summary of Test Standards

                                                Test Standards
FCC Part 15 Subpart C              PART 15 - RADIO FREQUENCY DEVICES
10-1-2017 Edition                  Subpart C - Intentional Radiators

All the test methods were according to KDB558074 v05r02 DTS Measurement Guidance and
ANSI C63.10 (2013).




Report Number: 64.790.19.01904.01                                                                                         Page 5 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


5 Summary of Test Results
                Technical Requirements
FCC Part 15 Subpart C
                                                                                                   Test             Test Result
Test Condition
                                                                                                   Site          Pass Fail N/A
§15.207                           Conducted emission AC power port                              ---

§15.247 (b) (1)                   Conducted peak output power                                   Site 1

§15.247(a)(1)                     20dB bandwidth                                                ---

§15.247(a)(1)                     Carrier frequency separation                                  ---

§15.247(a)(1)(iii)                Number of hopping frequencies                                 ---

§15.247(a)(1)(iii)                Dwell Time                                                    ---
                                  6dB bandwidth and 99% Occupied
§15.247(a)(2)                                                                                   Site 1
                                  Bandwidth
§15.247(e)                        Power spectral density                                        Site 1

§15.247(d)                        Spurious RF conducted emissions                               Site 1

§15.247(d)                        Band edge                                                     Site 1
                                  Spurious radiated emissions for
§15.247(d) & §15.209                                                                            Site 1
                                  transmitter
                                                                                                See
§15.203                           Antenna requirement
                                                                                                note 1

Note 1: N/A=Not Applicable.
Note 2: The EUT uses a Integrated antenna, which gain is 0 dBi. In accordance to §15.203, it is
considered sufficiently to comply with the provisions of this section.




Report Number: 64.790.19.01904.01                                                                                          Page 6 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


6   General Remarks

This submittal(s) (test report) is intended for FCC ID:; OU9SBM69—BO1 complies with Section
15.209, 15.247 of the FCC Part 15, Subpart C rules.
SBM 69 is Blood Pressure Monitor with Bluetooth 4.2. The TX and RX range is 2402MHz—
2480MHz for 4.2.

SUMMARY:

All tests according to the regulations cited on page 5 were

® — Performed

C — Not Performed

The Equipment under Test

& — Fulfilis the general approval requirements.

0O — Does not fulfill the general approval requirements.



Sample Received Date:                              May 20, 2019

Testing Start Date:                                May 23, 2019

Testing End Date:                                  May 25, 2019


— TUV SUD Certification and Testing (China) Co., Ltd. Guangzhou Branch —


                                                                                               Test by:           ZasWitigh




Report Number: 64.790.19.01904.01                                                                                           Page 7 of 31
                                TUV SUD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       SF, Communication Building, 163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +586 20 3832 0668 FAX: +86 20 3832 0478


7 Test Setups
7.1 AC Power Line Conducted Emission test setups

N/A

7.2 Radiated test setups

Below 1GHz




Above 1GHz




7.2 Conducted RF test setups


  Measuring                          EUT
  Receiver




Report Number: 64.790.19.01904.01                                                                                         Page 8 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


8 Systems test configuration

Auxiliary Equipment Used during Test:

             DESCRIPTION                 MANUFACTURER                   MODEL NO.(SHIELD)                   S/N(LENGTH)
                Notebook                         Lenovo                             X220                            ---
               Mobile phone                      Huawei                             P10                             ---

Test software: EMI_Test_Tool.exe, which used to control the EUT in continues transmitting
mode.

The system was configured to channel 0, 19, and 39 for the test.




Report Number: 64.790.19.01904.01                                                                                         Page 9 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


9 Technical Requirement

9.1 Conducted peak output power

Test Method
   1. Use the following spectrum analyzer settings:
       RBW > the 6 dB bandwidth of the emission being measured, VBW≥3RBW, Span≥3RBW
       Sweep = auto, Detector function = peak, Trace = max hold.
   2. Add a correction factor to the display.
   3. Allow the trace to stabilize. Use the marker-to-peak function to set the marker to the
       peak of the emission. The indicated level is the peak output power.
Limits
According to §15.247 (b) (1), conducted peak output power limit as below:

                           Frequency Range                        Limit                          Limit
                                 MHz                               W                             dBm
                              2400-2483.5                          ≤1                             ≤30

Test result as below table

                                                                    Conducted Peak
                                     Frequency                       Output Power                 Result
                                        MHz                              dBm
                              Bottom channel 2402MHz                     5.35                      Pass
                              Middle channel 2440MHz                     5.19                      Pass
                               Top channel 2480MHz                       4.83                      Pass




Report Number: 64.790.19.01904.01                                                                                         Page 10 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                      Low channel 2402MHz




                 Date:25.M AY.2019 18:13:54



                                                     Middle channel 2440MHz




                 Date:25.M AY.2019 18:16:37




Report Number: 64.790.19.01904.01                                                                                         Page 11 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                      High channel 2480MHz




                 Date:25.M AY.2019 18:18:15




Report Number: 64.790.19.01904.01                                                                                         Page 12 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


9.2 Power spectral density
Test Method
This procedure shall be used if maximum peak conducted output power was used to
demonstrate compliance:
   1. Set analyzer center frequency to DTS channel center frequency. RBW=3kHz,
       VBW≥3RBW, Span=1.5 times DTS bandwidth, Detector=Peak, Sweep=auto,
       Trace= max hold.
   2. Allow trace to fully stabilize, use the peak marker function to determine the maximum
       amplitude level within the RBW .
   3. Repeat above procedures until other frequencies measured were completed.
Limit

                                                             Limit [dBm]

                                                             ≤8dBm/3KHz

Test result
                                                              Power spectral
                           Frequency                             density                              Result
                            MHz                                       dBm
                  Bottom channel 2402MHz                             -12.26                             Pass
                  Middle channel 2440MHz                             -12.71                             Pass
                   Top channel 2480MHz                               -13.15                             Pass




Report Number: 64.790.19.01904.01                                                                                         Page 13 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                      Low channel 2402MHz




                 Date:25.M AY.2019 18:14:00



                                                     Middle channel 2440MHz




                 Date:25.M AY.2019 18:16:43




Report Number: 64.790.19.01904.01                                                                                         Page 14 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                      High channel 2480MHz




                 Date:25.M AY.2019 18:18:21




Report Number: 64.790.19.01904.01                                                                                         Page 15 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


9.3 6 dB Bandwidth and 99% Occupied Bandwidth
Test Method
      1. Use the following spectrum analyzer settings:
      RBW=100K, VBW≥3RBW, Sweep = auto, Detector function = peak, Trace = max hold
      2. Use the automatic bandwidth measurement capability of an instrument, may be
      employed using the X dB bandwidth mode with X set to 6 dB, care shall be taken so that
      the bandwidth measurement is not influenced by any intermediate power nulls in the
      fundamental emission that might be ≥ 6 dB.
      3. Allow the trace to stabilize, record the X dB Bandwidth value.

Limit

                                                               Limit [kHz]

                                                                   ≥500


Test result
           Frequency                           6dB bandwidth                  99% bandwidth
                                                                                                                       Result
              MHz                                   kHz                           kHz
     Bottom channel 2402MHz                            1264                            2250                              Pass
     Middle channel 2440MHz                            1400                            2386                              Pass
      Top channel 2480MHz                              1476                            2454                              Pass




Report Number: 64.790.19.01904.01                                                                                          Page 16 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


6 dB Bandwidth

                                                      Low channel 2402MHz




                 Date:25.M AY.2019 18:13:36




                 Date:25.M AY.2019 18:13:48




Report Number: 64.790.19.01904.01                                                                                         Page 17 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                     Middle channel 2440MHz




                 Date:25.M AY.2019 18:16:19




                 Date:25.M AY.2019 18:16:31




Report Number: 64.790.19.01904.01                                                                                         Page 18 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                      High channel 2480MHz




                 Date:25.M AY.2019 18:17:57




                 Date:25.M AY.2019 18:18:08




Report Number: 64.790.19.01904.01                                                                                         Page 19 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


9.4 Spurious RF conducted emissions

Test Method
   1. Establish a reference level by using the following procedure:
      a. Set RBW=100 kHz. VBW≥3RBW. Detector =peak, Sweep time = auto couple, Trace
         mode = max hold.
      b. Allow trace to fully stabilize, use the peak marker function to determine the maximum
         PSD level.
   2. Use the maximum PSD level to establish the reference level.
      a. Set the center frequency and span to encompass frequency range to be measured.
      b. Use the peak marker function to determine the maximum amplitude level. Ensure that
         the amplitude of all unwanted emissions outside of the authorized frequency band
         (excluding restricted frequency bands) are attenuated by at least the minimum
         requirements, report the three highest emissions relative to the limit.
   3. Repeat above procedures until other frequencies measured were completed.

Limit

                                            Frequency Range                        Limit (dBc)
                                                  MHz
                                                30-25000                                 -20




Report Number: 64.790.19.01904.01                                                                                          Page 20 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


Spurious RF conducted emissions


                                                              2402MHz




                 Date:25.M AY.2019 18:15:17




                 Date:25.M AY.2019 18:15:28




Report Number: 64.790.19.01904.01                                                                                         Page 21 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                              2440MHz




                 Date:25.M AY.2019 18:16:58




                 Date:25.M AY.2019 18:17:09




Report Number: 64.790.19.01904.01                                                                                         Page 22 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


                                                              2480MHz




                 Date:25.M AY.2019 18:19:37




                 Date:25.M AY.2019 18:19:48




Report Number: 64.790.19.01904.01                                                                                         Page 23 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


9.5 Band edge
Test Method
   1 Use the following spectrum analyzer settings:
      Span = wide enough to capture the peak level of the in-band emission and all spurious
      RBW = 100 kHz, VBW≥RBW, Sweep = auto, Detector function = peak, Trace = max
      hold.
   2 Allow the trace to stabilize, use the peak and delta measurement to record the result.
   3 The level displayed must comply with the limit specified in this Section.

Limit
                                            Frequency Range                        Limit (dBc)
                                                  MHz
                                                30-25000                                 -20




Report Number: 64.790.19.01904.01                                                                                          Page 24 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


Band edge testing

                                                                2402MHz




                 Date:25.M AY.2019 18:14:10




                                                                2480MHz




                 Date:25.M AY.2019 18:18:31




Report Number: 64.790.19.01904.01                                                                                         Page 25 of 31
                                  TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


9.6 Spurious radiated emissions for transmitter

Test Method
   1: The EUT was place on a turn table which is 1.5m above ground plane for above 1GHz
   and 0.8m above ground for below 1GHz at 3 meter chamber room for test. The table was
   rotated 360 degrees to determine the position of the highest radiation.
   2: The EUT was set 3 meters away from the interference – receiving antenna, which was
   mounted on the top of a variable – height antenna tower.
   3: The height of antenna is varied from one meter to four meters above the ground to
   determine the maximum value of the field strength. Both horizontal and vertical polarizations
   of the antenna are set to make the measurement.
   4: For each suspected emission, the EUT was arranged to its worst case and then the
   antenna was tuned to heights from 1 meter to 4 meters and the rotatable table was turned
   from 0 degrees to 360 degrees to find the maximum reading.
   5: Use the following spectrum analyzer settings According to C63.10:
   For Above 1GHz
       Span = wide enough to capture the peak level of the in-band emission and all spurious
       RBW = 1MHz, VBW≥RBW for peak measurement and VBW = 10Hz for average
       measurement, Sweep = auto, Detector function = peak, Trace = max hold.
   For Below 1GHz
   Use the following spectrum analyzer settings:
       Span = wide enough to capture the peak level of the in-band emission and all spurious
       RBW = 100 KHz, VBW≥RBW for peak measurement, Sweep = auto, Detector function =
       peak, Trace = max hold.
Note:
       1: The resolution bandwidth and video bandwidth of test receiver/spectrum analyzer is
       120 KHz for Quasi-peak detection (QP) at frequency below 1GHz.
       2: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 3MHz for peak detection (PK) at frequency above 1GHz.
       3: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 3MHz for RMS Average ((duty cycle ˂ 98%) for Average detection (AV) at
       frequency above 1GHz, then the measurement results was added to a correction factor
       (20log(1/duty cycle).
       4: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 10Hz (duty cycle ˃ 98%) for Average detection (AV) at requencyabove1GHz




Report Number: 64.790.19.01904.01                                                                                         Page 26 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


Limit

The radio emission outside the operating frequency band shall be at least 20 dB below that in
the 100 kHz bandwidth within the band that contains the highest level of the desired power.
Radiated emissions which fall in the restricted bands, as defined in section15.205, must comply
with the radiated emission limits specified in section 15.209.

                Frequency                    Field Strength                 Field Strength                     Detector
                    MHz                           uV/m                         dBμV/m
                    30-88                          100                            40                               QP
                   88-216                          150                           43.5                              QP
                  216-960                          200                            46                               QP
                 960-1000                          500                            54                               QP
                Above 1000                         500                            54                               AV
                Above 1000                        5000                            74                               PK




Report Number: 64.790.19.01904.01                                                                                          Page 27 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


Spurious radiated emissions for transmitter

According to C63.10, if the peak (or quasi-peak) measured value complies with the average limit,
it is unnecessary to perform an average measurement, so AV emission value did not show in
below table if the peak value complies with average limit.

Transmitting spurious emission test result as below:

Below 1GHz
                                      Emission                                                                      Correct
   Frequency      Frequency                            Polarization         Limit       Detector       Margin                    Result
                                       Level                                                                         factor
      Band
                     MHz              dBuV/m                              dBμV/m                          dB          (dB)
                  71.979444            24.45                  H            40.00            QP          15.55         -32.0        Pass
                  360.015556           30.50                  H            46.00            QP          15.50         -25.0        Pass
       30-        877.995556           38.08                  H            46.00            QP           7.92         -15.9        Pass
    1000MHz       71.979444            26.71                  V            40.00            QP          13.29         -32.0        Pass
                  287.373333           33.04                  V            46.00            QP          12.96         -24.9        Pass
                  884.785556           39.33                  V            46.00            QP           6.67         -15.8        Pass

Above 1GHZ
Low channel 2402MHz Test Result
                                      Emission                                                                       Correct
  Frequency        Frequency                          Polarization          Limit       Detector       Margin                     Result
                                        Level                                                                         factor
     Band
                     MHz               dBuV/m                             dBμV/m                          dB           (dB)
                 1252.062500            37.65                 H             74              PK           36.35         -12.1       Pass
                 1787.750000            31.75                 H             74              PK           42.25         -10.1       Pass
                 7206.093750            39.90                 H             74              PK           34.10           5.0       Pass
    1000-        17625.468750           50.27                 H             74              PK           23.73          21.2       Pass
  25000MHz       1232.125000            35.98                 V             74              PK           38.02         -12.0       Pass
                 1825.812500            32.98                 V             74              PK           41.02          -9.8       Pass
                 7206.093750            39.77                 V             74              PK           34.23           5.0       Pass
                 17602.031250           50.00                 V             74              PK           24.00          21.2       Pass

Middle channel 2440MHz Test Result

                                       Emission                                                                     Correct
   Frequency       Frequency                           Polarization          Limit       Detector      Margin                    Result
                                         Level                                                                       factor
      Band
                     MHz                dBuV/m                             dBμV/m                         dB          (dB)
                 1236.062500             38.54                H              74              PK          35.46        -12.0        Pass
                 1793.437500             31.24                H              74              PK          42.76        -10.1        Pass
                 7023.750000             37.99                H              74              PK          36.01          5.6        Pass
     1000-       17758.125000            50.16                H              74              PK          23.84         21.3        Pass
   25000MHz      1234.750000             37.16                V              74              PK          36.84        -12.0        Pass
                 1826.687500             32.45                V              74              PK          41.55         -9.8        Pass
                 7320.000000             37.19                V              74              PK          36.81          5.2        Pass
                 17186.718750            49.69                V              74              PK          24.31         20.4        Pass




Report Number: 64.790.19.01904.01                                                                                         Page 28 of 31
                                TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
       5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


High channel 2480MHz Test Result

                                        Emission                                                                     Correct
   Frequency        Frequency                           Polarization          Limit       Detector      Margin                    Result
                                          Level                                                                       factor
      Band
                      MHz                dBuV/m                             dBμV/m                         dB          (dB)
                  1228.500000             37.74                H              74              PK          36.26        -12.1        Pass
                  1803.562500             30.47                H              74              PK          43.53        -10.0        Pass
                  7184.062500             36.03                H              74              PK          37.97          5.1        Pass
     1000-        17868.750000            49.58                H              74              PK          24.42         21.4        Pass
   25000MHz       1243.562500             37.33                V              74              PK          36.67        -12.1        Pass
                  2157.687500             32.89                V              74              PK          41.11         -7.9        Pass
                  7010.625000             38.48                V              74              PK          35.52          5.4        Pass
                  17648.437500            49.30                V              74              PK          24.70         21.2        Pass


Remark:
      (1) Data of measurement within this frequency range shown “--” in the table above means the reading of
          emissions are the noise floor or attenuated more than 10dB below the permissible limits or the field
          strength is too small to be measured.
      (2) Above 1GHz: Corrector factor = Antenna Factor + Cable Loss- Amplifier Gain
          Below 1GHz: Corrector factor = Antenna Factor + Cable Loss
      (3) The Reading Level is recorded by software which is not shown in the sheet




Report Number: 64.790.19.01904.01                                                                                          Page 29 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


10 Test Equipment List
                                                  List of Test Instruments
Radiated Emission Test
           Description                      Manufacturer                 Model no.              Serial no.              cal. due date
        EMI Test Receiver                 Rohde & Schwarz                 ESR 26                 101269                   2019-7-6
  Trilog Super Broadband Test
                                             Schwarzbeck                VULB 9163                   707                  2019-6-28
            Antenna
          Horn Antenna                    Rohde & Schwarz                  HF907                 102294                  2019-6-28
         Pre-amplifier                    Rohde & Schwarz                 SCU 18                102230                   2019-6-28
       Signal Generator                   Rohde & Schwarz                 SMY01              839369/005                  2019-6-28
          Attenuator                          Agilent                     8491A              MY39264334                  2019-6-28
   3m Semi-anechoic chamber                    TDK                        9X6X6                   ----                    2020-7-7
         Test software                    Rohde & Schwarz                 EMC32             Version 9.15.00                 N/A

Conducted Emission Test
         Description                      Manufacturer                   Model no.            Serial no.               cal. due date
     EMI Test Receiver                  Rohde & Schwarz                   ESR 3                101782                    2019-7-6
            LISN                        Rohde & Schwarz                  ENV432                101318                    2019-7-6
         Attenuator                    Shanghai Huaxiang                 TS2-26-3            080928189                   2019-7-6
        Test software                   Rohde & Schwarz                   EMC32             Version9.15.00                  N/A

RF conducted test
         Description                   Manufacturer                 Model no.                     Serial no.             cal. due date
       Signal Analyzer               Rohde & Schwarz                 FSV40                         101030                  2019-7-6
                                                                 OSP120/OSP-SBM
      RF Switch Module               Rohde & Schwarz                                          101226/100851                2019-7-6
                                                                      6957
        Power Splitter                    Weinschel                   1580                          SC319                  2019-7-5
       10dB Attenuator                    Weinschel                  4M-10                          43152                  2019-7-6
                                                                                                   Version                   N/A
        Test software                Rohde & Schwarz                     EMC32
                                                                                                   10.38.00
        Test software                     Tonscend               System for BT/WIFI              Version 2.6                  N/A




Report Number: 64.790.19.01904.01                                                                                          Page 30 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478


11 System Measurement Uncertainty
For a 95% confidence level, the measurement expanded uncertainties for defined systems, in accordance with the
recommendations of ISO 17025 were:

                                        System Measurement Uncertainty
                                Test Items                                                             Extended Uncertainty
  Uncertainty for Conducted Emission 150kHz-30MHz (for test using                                            2.51 dB
  High Voltage Probe TK9420(VT9420))
  Uncertainty for Radiated Spurious Emission 25MHz-3000MHz                                            Horizontal: 4.80dB;
                                                                                                        Vertical: 4.87dB;
  Uncertainty for Radiated Spurious Emission 3000MHz-18000MHz                                         Horizontal: 4.59dB;
                                                                                                        Vertical: 4.58dB;
  Uncertainty for Radiated Spurious Emission 18000MHz-40000MHz                                        Horizontal: 5.05dB;
                                                                                                        Vertical: 5.04dB;
  Uncertainty for Conducted RF test with TS 8997                                                Power level test involved: 1.16dB
                                                                                                   Frequency test involved:
                                                                                                         0.6×10-7 or 1%




Report Number: 64.790.19.01904.01                                                                                          Page 31 of 31
                                 TÜV SÜD Certification and Testing (China) Co., Ltd. Guangzhou Branch
        5F, Communication Building,163 Pingyun Rd, Huangpu Ave. West Guangzhou, P.R.China TEL: +86 20 3832 0668 FAX: +86 20 3832 0478



Document Created: 2019-08-21 10:51:19
Document Modified: 2019-08-21 10:51:19

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC