Test Report

FCC ID: OU9LS417-F01

Test Report

Download: PDF
FCCID_3373510

Report Number: 68.930.17.019.01




                                                               FCC - TEST REPORT



Report Number                                    :     68.930.17.019.01                                 Date of Issue:                        April 14, 2017


 Model                                           :     LS417-F, LS418-F

 Product Type                                    :     Lifesense Band 2, Lifesense Band 2S

 Applicant                                       :     Guangdong Transtek Medical Electronics Co., Ltd

 Address                                         :     Zone A, No. 105, Dongli Road, Torch Development District

                                                       528437 Zhongshan, Guangdong, PEOPLE'S REPUBLIC OF CHINA

 Production Facility                             :     Guangdong Transtek Medical Electronics Co., Ltd

 Address                                         :     Zone B, No. 105, Dongli Road, Torch Development District

                                                       528437 Zhongshan, Guangdong, PEOPLE'S REPUBLIC OF CHINA


Test Result                                      :      Positive                       Negative



Total pages including
Appendices                                       :     28




TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch is a subcontractor to TÜV SÜD Product Service GmbH according to the principles outlined in
ISO 17025.

TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch reports apply only to the specific samples tested under stated test conditions. Construction of
the actual test samples has been documented. It is the manufacturer's responsibility to assure that additional production units of this model are manufactured with
identical electrical and mechanical components. The manufacturer/importer is responsible to the Competent Authorities in Europe for any modifications made to the
production units which result in non-compliance to the relevant regulations. TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch shall have no
liability for any deductions, inferences or generalizations drawn by the client or others from TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
issued reports.

This report is the confidential property of the client. As a mutual protection to our clients, the public and ourselves, extracts from the test report shall not be
reproduced except in full without our written approval.




EMC_SZ_FR_21.00 FCC                                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                                Page 1 of 28
                                           Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



1 Table of Contents
1     Table of Contents ................................................................................................................................................................ 2
2     Details about the Test Laboratory..................................................................................................................................... 3
3     Description of the Equipment under Test ........................................................................................................................ 4
4     Summary of Test Standards .............................................................................................................................................. 5
5     Summary of Test Results ................................................................................................................................................... 6
6     General Remarks ................................................................................................................................................................ 7
7     Test Setups .......................................................................................................................................................................... 8
8     Systems test configuration ................................................................................................................................................. 9
9     Technical Requirement ........................................................................................................................................................ 10
    9.1       Conducted peak output power................................................................................................................................ 10
    9.2       6dB bandwidth .......................................................................................................................................................... 13
    9.3       Power spectral density............................................................................................................................................. 15
    9.4       Spurious RF conducted emissions ........................................................................................................................ 17
    9.5       Band edge ................................................................................................................................................................. 21
    9.6       Spurious radiated emissions for transmitter ......................................................................................................... 23
10        Test Equipment List ...................................................................................................................................................... 27
11        System Measurement Uncertainty ............................................................................................................................. 28




EMC_SZ_FR_21.00 FCC                                       TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                                           Page 2 of 28
                                             Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                                  Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



2 Details about the Test Laboratory
Details about the Test Laboratory
Test Site 1

Company name:         TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch
                      Building 12&13, Zhiheng Wisdomland Business Park,
                      Nantou Checkpoint Road 2, Nanshan District,
                      Shenzhen City, 518052,
                      P. R. China

FCC Registration      502708
Number:

IC Registration       10320A-1
Number:

Telephone:            86 755 8828 6998
Fax:                  86 755 828 5299




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 3 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



3 Description of the Equipment under Test
Description of the Equipment Under Test

Product:                          Lifesense Band 2S

Model no.:                        LS418-F

FCC ID:                           OU9LS417-F01

Options and accessories:          NIL

Rating:                           5VDC, 0.5A (Charging by USB port)
                                  3.8VDC, 75mAh (Supplied by the internal Li-ion rechargeable
                                  battery)

RF Transmission                   2402-2480MHz
Frequency:
No. of Operated Channel:          40

Modulation:                       GFSK

Antenna Type:                     Integrated Antenna

Antenna Gain:                     -2.0dBi

Description of the EUT:           The Equipment Under Test (EUT) is a Lifesense Band 2S with
                                  Bluetooth function operating at 2.4GHz




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 4 of 28
                          Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20               Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



4 Summary of Test Standards

                                         Test Standards
FCC Part 15 Subpart C       PART 15 - RADIO FREQUENCY DEVICES
10-1-2016 Edition           Subpart C - Intentional Radiators

All the test methods were according to KDB558074 DTS Measurement Guidance and ANSI
C63.10 (2013).




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 5 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



5 Summary of Test Results

                                            Technical Requirements
FCC Part 15 Subpart C
Test Condition                                                                             Pages           Test Result         Test
                                                                                                                               Site
§15.207              Conducted emission AC power port                                          --                N/A            --

§15.247(b)(1)        Conducted peak output power                                              10                Pass          Site 1
§15.247(e)           Power spectral density*                                                  15                Pass          Site 1

§15.247(a)(2)        6dB bandwidth                                                            13                Pass          Site 1

§15.247(a)(1)        20dB bandwidth and 99% Occupied                                           --                N/A            --
                     Bandwidth
§15.247(a)(1)        Carrier frequency separation                                              --                N/A            --

§15.247(a)(1)(iii)   Number of hopping frequencies                                             --                N/A            --

§15.247(a)(1)(iii)   Dwell Time                                                                --                N/A            --

§15.247(d)           Spurious RF conducted emissions                                          17                Pass          Site 1

§15.247(d)           Band edge                                                                21                Pass          Site 1

§15.247(d) &         Spurious radiated emissions for                                          23                Pass          Site 1
§15.209 &            transmitter and receiver
§15.203              Antenna requirement                                                    See                 Pass            --
                                                                                           note 1

Remark 1: N/A – Not Applicable.
Note 1: The EUT uses an Integrated Antenna, which gain is -2.0dBi. According to §15.203, it is
considered sufficiently to comply with the provisions of this section.




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 6 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



6 General Remarks
Remarks

This submittal(s) (test report) is intended for FCC ID: OU9LS417-F01 complies with Section
15.209, 15.247 of the FCC Part 15, Subpart C.

This report is for the BT 4.0 part.

SUMMARY:

All tests according to the regulations cited on page 5 were

 - Performed

 - Not Performed

The Equipment under Test

 - Fulfills the general approval requirements.

 - Does not fulfill the general approval requirements.


Sample Received Date:                       April 10, 2017

Testing Start Date:                         April 10, 2017

Testing End Date:                           April 14, 2017


- TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch -

Reviewed by:                             Prepared by:                                            Tested by:




    _______________________                      _______________________                          _______________________
         Phoebe Hu                                    Aaron Lai                                      Leon Zhang
     EMC Project Manager                          EMC Project Engineer                           EMC Test Engineer




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 7 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



7 Test Setups
Below 1GHz




Above 1GHz




7.2 Conducted RF test setups


  Measuring                   EUT
  Receiver




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 8 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



8 Systems test configuration
Auxiliary Equipment Used during Test:

            DESCRIPTION            MANUFACTURER                    MODEL NO.(SHIELD)                     S/N(LENGTH)
                  ---                         ---                                ---                             ---

Test software: nRF go studio test tool, which used to control the EUT in continues transmitting
mode.

The system was configured to channel 0, 19, and 39 for the test.




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 9 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



9 Technical Requirement
9.1
9.1 Conducted peak output power

Test Method
   1. Use the following spectrum analyzer settings:
      RBW > the 6 dB bandwidth of the emission being measured, VBW≥3RBW, Span≥3RBW
      Sweep = auto, Detector function = peak, Trace = max hold.
   2. Add a correction factor to the display.
   3. Allow the trace to stabilize. Use the marker-to-peak function to set the marker to the
      peak of the emission. The indicated level is the peak output power.

Limits
According to §15.247 (b) (1), conducted peak output power limit as below:

                      Frequency Range                        Limit                          Limit
                            MHz                               W                             dBm
                         2400-2483.5                          ≤1                             ≤30

Test result as below table

                                                      Conducted Peak
                      Frequency                        Output Power                               Result
                          MHz                                   dBm
                Bottom channel 2402MHz                          -3.15                               Pass
                Middle channel 2440MHz                          -3.27                               Pass
                 Top channel 2480MHz                            -3.53                               Pass




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 10 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



                                                        2402MHz




                                                        2440MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 11 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



                                                        2480MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 12 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




9.2 6dB bandwidth
Test Method
   1. Use the following spectrum analyzer settings:
      RBW=100K, VBW≥3RBW, Sweep = auto, Detector function = peak, Trace = max hold
   2. Use the automatic bandwidth measurement capability of an instrument, may be
      employed using the X dB bandwidth mode with X set to 6 dB, care shall be taken so that
      the bandwidth measurement is not influenced by any intermediate power nulls in the
      fundamental emission that might be ≥ 6 dB.
   3. Allow the trace to stabilize, record the X dB Bandwidth value.
Limit

                                                        Limit [kHz]

                                                            ≥500

Test result
           Frequency                   6dB bandwidth                      99 bandwidth
                                                                                                                   Result
              MHz                           kHz                               kHz
     Bottom channel 2402MHz                    699.00                          1072.40                              Pass
     Middle channel 2440MHz                    694.60                          1072.40                              Pass
      Top channel 2480MHz                      694.60                          1072.40                              Pass

                                                        2402MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 13 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



6 dB Bandwidth

                                                        2440MHz




                                                        2480MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 14 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




9.3 Power spectral density
Test Method
This procedure shall be used if maximum peak conducted output power was used to
demonstrate compliance:
   1. Set analyzer center frequency to DTS channel center frequency. RBW=3kHz,
       VBW≥3RBW, Span=1.5 times DTS bandwidth, Detector=Peak, Sweep=auto,
       Trace= max hold.
   2. Allow trace to fully stabilize, use the peak marker function to determine the maximum
       amplitude level within the RBW .
   3. Repeat above procedures until other frequencies measured were completed.
Limit

                                                       Limit [dBm]

                                                              ≤8

Test result
                                                        Power spectral
                      Frequency                            density                                Result
                          MHz                                   dBm
                Bottom channel 2402MHz                         -21.67                               Pass
                Middle channel 2440MHz                         -21.76                               Pass
                 Top channel 2480MHz                           -22.01                               Pass

                                                        2402MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 15 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




                                                        2440MHz




                                                        2480MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 16 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




9.4 Spurious RF conducted emissions
Test Method
   1. Establish a reference level by using the following procedure:
      a. Set RBW=100 kHz. VBW≥3RBW. Detector =peak, Sweep time = auto couple, Trace
         mode = max hold.
      b. Allow trace to fully stabilize, use the peak marker function to determine the maximum
         PSD level.
   2. Use the maximum PSD level to establish the reference level.
      a. Set the center frequency and span to encompass frequency range to be measured.
      b. Use the peak marker function to determine the maximum amplitude level. Ensure that
         the amplitude of all unwanted emissions outside of the authorized frequency band
         (excluding restricted frequency bands) are attenuated by at least the minimum
         requirements, report the three highest emissions relative to the limit.
   3. Repeat above procedures until other frequencies measured were completed.

Limit

                                    Frequency Range                          Limit (dBc)
                                          MHz
                                        30-25000                                   -20




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 17 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




Spurious RF conducted emissions

                                                        2402MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 18 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




Spurious RF conducted emissions

                                                        2440MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 19 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




Spurious RF conducted emissions

                                                        2480MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 20 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




9.5 Band edge
Test Method
   1 Use the following spectrum analyzer settings:
      Span = wide enough to capture the peak level of the in-band emission and all spurious
      RBW = 100 kHz, VBW≥RBW, Sweep = auto, Detector function = peak, Trace = max
      hold.
   2 Allow the trace to stabilize, use the peak and delta measurement to record the result.
   3 The level displayed must comply with the limit specified in this Section.

Limit
                                    Frequency Range                          Limit (dBc)
                                          MHz
                                        30-25000                                   -20

Test result

                                                           2402MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 21 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




Band edge

                                                           2480MHz




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 22 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




9.6 Spurious radiated emissions for transmitter

Test Method
   1: The EUT was place on a turn table which is 1.5m above ground plane for above 1GHz
   and 0.8m above ground for below 1GHz at 3-meter chamber room for test. The table was
   rotated 360 degrees to determine the position of the highest radiation.
   2: The EUT was set 3 meters away from the interference – receiving antenna, which was
   mounted on the top of a variable – height antenna tower.
   3: The height of antenna is varied from one meter to four meters above the ground to
   determine the maximum value of the field strength. Both horizontal and vertical polarizations
   of the antenna are set to make the measurement.
   4: For each suspected emission, the EUT was arranged to its worst case and then the
   antenna was tuned to heights from 1 meter to 4 meters and the rotatable table was turned
   from 0 degrees to 360 degrees to find the maximum reading.
   5: Use the following spectrum analyzer settings According to C63.10:
   For Above 1GHz
       Span = wide enough to capture the peak level of the in-band emission and all spurious
       RBW = 1MHz, VBW≥RBW for peak measurement and VBW = 10Hz for average
       measurement, Sweep = auto, Detector function = peak, Trace = max hold.
   For Below 1GHz
   Use the following spectrum analyzer settings:
       Span = wide enough to capture the peak level of the in-band emission and all spurious
       RBW = 100 KHz, VBW≥RBW for peak measurement, Sweep = auto, Detector function =
       peak, Trace = max hold.
Note:
       1: The resolution bandwidth and video bandwidth of test receiver/spectrum analyzer is
       120 KHz for Quasi-peak detection (QP) at frequency below 1GHz.
       2: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 3MHz for peak detection (PK) at frequency above 1GHz.
       3: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 3MHz for RMS Average ((duty cycle ˂ 98%) for Average detection (AV) at
       frequency above 1GHz, then the measurement results was added to a correction factor
       (20log(1/duty cycle)).
       4: The resolution bandwidth of test receiver/spectrum analyzer is 1MHz and the video
       bandwidth is 10Hz (duty cycle ˃ 98%) for Average detection (AV) at frequency above
       1GHz.




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 23 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




Limit

The radio emission outside the operating frequency band shall be at least 20 dB below that in
the 100 kHz bandwidth within the band that contains the highest level of the desired power.
Radiated emissions which fall in the restricted bands, as defined in section15.205, must comply
with the radiated emission limits specified in section 15.209.

              Frequency                 Field Strength                 Field Strength                       Detector
                  MHz                        uV/m                         dBμV/m
                  30-88                       100                            40                                 QP
                 88-216                       150                           43.5                                QP
                216-960                       200                            46                                 QP
               960-1000                       500                            54                                 QP
              Above 1000                      500                            54                                 AV
              Above 1000                     5000                            74                                 PK




EMC_SZ_FR_21.00 FCC                     TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 24 of 28
                           Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



Spurious radiated emissions for transmitter

According to C63.10, if the peak (or quasi-peak) measured value complies with the average limit,
it is unnecessary to perform an average measurement, so AV emission value did not show in
below table if the peak value complies with average limit.

Transmitting spurious emission test result as below:

2402MHz (30MHz – 1GHz)
                                    Emission
                   Frequency                           Polarization            Limit         Detector       Result
                                      Level
                      MHz            dBuV/m                                  dBμV/m
                     938.76           33.56              Horizontal           46.00              QP           Pass
                     954.34           35.85               Vertical            46.00              QP           Pass

2402MHz (Above 1GHz)
                                    Emission
                   Frequency                           Polarization            Limit         Detector       Result
                                      Level
                      MHz            dBuV/m                                  dBμV/m
                     7206.56          47.32              Horizontal           74.00              PK           Pass
                     7205.15          48.55               Vertical            74.00              PK           Pass

2440MHz (30MHz – 1GHz)
                                    Emission
                  Frequency                              Polarization             Limit        Detector        Result
                                      Level
                     MHz             dBuV/m                                     dBμV/m
                      --                --                Horizontal              --               QP           Pass
                      --                --                 Vertical               --               QP           Pass

2440MHz (Above 1GHz)
                                    Emission
                   Frequency                           Polarization            Limit         Detector       Result
                                      Level
                      MHz            dBuV/m                                  dBμV/m
                     7319.06          56.25              Horizontal           74.00              PK           Pass
                     7319.06          41.18              Horizontal           54.00              AV           Pass
                     7319.06          48.03               Vertical            74.00              PK           Pass

Remark:
      (1) Data of measurement within this frequency range shown “-” in the table above means the reading of
          emissions are attenuated more than 20db below the permissible limits or the field strength is too small
          to be measured.
      (2) “*” means the emission(s) appear within the restrict bands shall follow the requirement of section
          15.205.




EMC_SZ_FR_21.00 FCC                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 25 of 28
                            Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                 Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01




2480MHz (30MHz – 1GHz)
                                    Emission
                  Frequency                              Polarization             Limit        Detector        Result
                                      Level
                     MHz             dBuV/m                                     dBμV/m
                      --                --                Horizontal              --               QP           Pass
                      --                --                 Vertical               --               QP           Pass

2480MHz (Above 1GHz)
                                    Emission
                   Frequency                           Polarization            Limit         Detector       Result
                                      Level
                      MHz            dBuV/m                                  dBμV/m
                     7439.06          52.67              Horizontal           74.00              PK           Pass
                     7440.00          51.18               Vertical            74.00              PK           Pass

Remark:
      (1) Data of measurement within this frequency range shown “-” in the table above means the reading of
          emissions are attenuated more than 20db below the permissible limits or the field strength is too small
          to be measured.
      (2) “*” means the emission(s) appear within the restrict bands shall follow the requirement of section
          15.205.




EMC_SZ_FR_21.00 FCC                      TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 26 of 28
                            Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20                 Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



10 Test Equipment List
                                            List of Test Instruments

      DESCRIPTION               MANUFACTURER                     MODEL NO.               SERIAL NO.              CAL. DUE DATE
     EMI Test Receiver          Rohde & Schwarz                     ESR 3                  101782                  2017-7-15
           LISN                 Rohde & Schwarz                    ENV4200                 100249                  2017-7-15
           LISN                 Rohde & Schwarz                    ENV216                  100326                  2017-7-15
           ISN                  Rohde & Schwarz                     ENY81                  100177                  2017-7-15
           ISN                  Rohde & Schwarz                   ENY81-CA6                101664                  2017-7-15
                                                                 TK9420(VT94
     High Voltage Probe          Rohde & Schwarz                                            9420-58                   2017-7-15
                                                                     20)
      RF Current Probe           Rohde & Schwarz                    EZ-17                  100816                     2017-7-15
          Attenuator            Shanghai Huaxiang                  TS2-26-3              080928189                    2017-7-17
        Test software            Rohde & Schwarz                    EMC32               Version9.15.00                   N/A
       Signal Analyzer           Rohde & Schwarz                    FSV40                  101030                     2017-7-15

C - Conducted RF tests
   Conducted peak output power
   6dB bandwidth
   Power spectral density*
   Spurious RF conducted emissions
   Band edge




EMC_SZ_FR_21.00 FCC                    TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 27 of 28
                          Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20               Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299


Report Number: 68.930.17.019.01



11 System Measurement Uncertainty
For a 95% confidence level, the measurement expanded uncertainties for defined systems, in
accordance with the recommendations of ISO 17025 were:

                                  System Measurement Uncertainty
                             Test Items                                                        Extended Uncertainty
  Uncertainty for Radiated Spurious Emission 25MHz-3000MHz                              Horizontal: 4.98dB;
                                                                                        Vertical: 5.06dB;
  Uncertainty for Radiated Spurious Emission 3000MHz-                                   Horizontal: 4.95dB;
  18000MHz                                                                              Vertical: 4.94dB;
  Uncertainty for Radiated Spurious Emission 18000MHz-                                  Horizontal: 5.14dB;
  40000MHz                                                                              Vertical: 5.12dB;
  Uncertainty for Conducted RF test with TS 8997                                        Power level test involved: 2.06dB
                                                                                        Frequency test involved:
                                                                                        1.16×10-7




EMC_SZ_FR_21.00 FCC                   TÜV SÜD Certification and Testing (China) Co., Ltd. Shenzhen Branch                Page 28 of 28
                         Building 12&13, Zhiheng Wisdomland Business Park, Nantou Checkpoint Road 2, Nanshan District,
Release 2014-03-20              Shenzhen City, 518052, P. R. China Tel. +86 755 8828 6998, Fax: +86 755 8828 5299



Document Created: 2017-04-21 10:15:11
Document Modified: 2017-04-21 10:15:11