Ex 2 Part 95 Rpt

FCC ID: OU5340MT

Test Report

Download: PDF
FCCID_155207

                                                        Intertek Testing Services
                                                        ETL SEMKO       &



                                                         FCC Part 95 (H) Test Report
                                                                     for
                                                            GE Medical Systems
                                                                        on the
                                                              WMTS Transmitter
                                                            Model: 342MCCN—XXX
                                                             FCC ID: OUS340MT


                                                                Job No: J20046260
                                                         Date of Test: May 1 to 10, 2001


                                                             Report No.: 20462601
                                                          Date of Report: May 25, 2001



                                                            nviad
im                     /!’,———'--.\\                    NVLAP Laboratory Code: 200201—0
     .| Tested by:     !                     x          Suresh Kondapalli                          Review Date:         .
                                                                                                                                 F
       Reyiewer:                       |                David Chernomordik, Ph.D.                  Review Date:
e       c:(lfi,l//:fl/' %“OHAMZ EMC Site Manager                                                                         5—/?//0/'




       Accreditation logos in this area if any apply to testing reported. Note that limitations on scope of accreditation must be stated. For
       example, NIST Handbook 150, section 285.8, requires: "Use of the NVLAP logo...should be accompanied by the following:
       "Accredited by the National Voluntary Laboratory Accreditation Program for the specific scope of accreditation under Lab Code
       200201—0."



       All services undertaken are subject to the following general policy: Reports are submitted for exclusive use of the client to whom
       they are addressed. Their significance is subject to the adequacy and representative character of the samples and to the
       comprehensiveness of the tests, examinations or surveys made. This report shall not be reproduced except in full, without written
       consent of Intertek Testing Services, NA Inc. This report must not be used to claim product endorsement by NVLAP, NIST nor
       any other agency of the U.S. Government.




                    € b                                   Intertek Testing Services NA, Inc.
                                                            1365 Adams Court, Mento Park, CA 94025
                                           Telephone 650—463—2900 Fax 650—463—2910     Home Page www.etisemko.com
                     Liste?


                                                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                                                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT



                                                  TABLE OF CONTENTS


1.0     Job Description....................................................................................................................... 2
        1.1    Client Information ....................................................................................................... 2
        1.2    Equipment under test (EUT) ........................................................................................ 2
        1.3    Test plan reference....................................................................................................... 3
        1.4    Description of the EUT................................................................................................ 3
        1.5    System test configuration............................................................................................. 4
               1.5.1 System block diagram & Support equipment ................................................... 4
        1.6    Justification ................................................................................................................. 5
        1.7    Mode(s) of operation ................................................................................................... 5
        1.8    Modifications required for compliance......................................................................... 5

2.0     Test Summary ....................................................................................................................... 6

3.0     RF Power Output ................................................................................................................... 7
        3.1   Test Description........................................................................................................... 7
        3.2   Test Procedure............................................................................................................. 7
        3.3   Test Results ................................................................................................................. 7

4.0     Occupied Bandwidth, Necessary Bandwidth......................................................................... 9
        4.1    Test Description........................................................................................................... 9
        4.2    Test Procedure............................................................................................................. 9
        4.3    Test Results ................................................................................................................. 9

5.0     Spurious emissions at antenna terminal .............................................................................. 12
        5.1    Test description ......................................................................................................... 12
        5.2    Test Procedure........................................................................................................... 12
        5.3    Test Result................................................................................................................. 12

6.0     Field Strength of Radiated Emissions.................................................................................. 16
        6.1     Test description ......................................................................................................... 16
        6.2     Test Procedure........................................................................................................... 16
        6.3     Test Results ............................................................................................................... 16

7.0     Frequency Stability .............................................................................................................. 19
        7.1   Test description ......................................................................................................... 19
        7.2   Test Procedure........................................................................................................... 19
              7.2.1 Frequency Stability vs. Temperature............................................................... 19
              7.2.2 Frequency Stability vs. Voltage ..................................................................... 19
        7.3   Test Results ............................................................................................................... 20

8.0     List of test equipment........................................................................................................... 21



File: 20462601                                                 Version 1.0                                                       Page 1 of 21
FCC Part 95 (H)


                                                             1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                            Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

1.0     Job Description

1.1     Client Information

The EUT has been tested at the request of:

Company:                         GE Medical Systems
                                 61 Barnes park North,
                                 Wallingford, CT 06492 USA

Name of contact:                 Lenny Bienasz
Telephone:                       (203) 949-8225
Fax:                             (203) 949-2536


1.2     Equipment under test (EUT)

Equipment type:                  WMTS Transmitter

Model number(s):                 342MCCN-XXX


Part or serial number:           Not Labeled

Manufacturer:                    SAME as above.

Use of Product :                 Medical Telemetry

Production is planned:           [X] Yes, [ ] No

Technical Specifications:

                          Type of Emission                         F3D
                          Max. modulation frequency (M)            0.5 kHz
                          Max. Allowed deviation (D)               5 kHz, typically 3 kHz
                          RF Output Power                          4 mW
                          Means for variation of operating power   Fixed
                          Frequency Range                          608 to 614MHz
                          Max. number of Channels                  238
                          Channel separation                       25 kHz
                          Antenna                                  Detachable rod antenna
                          External input                           Ultrasound, ECG and UA




File: 20462601                                 Version 1.0                               Page 2 of 21
FCC Part 95 (H)


                                                              1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                             Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

EUT receive date:                April 30, 2001

EUT received condition:          Good condition prototype

Test start date:                 May 1, 2001

Test end date:                   May 4, 2001


1.3     Test plan reference

FCC Part 2.1093, FCC Part 95 (H)

1.4     Description of the EUT

The 340M OB Telemetry System is comprised of a transmitter and receiver to provide an RF link
between patient worn transducers and fetal monitor, allowing patient to freely move about while staff
monitor fetal heart rate and uterine contractions remotely.




File: 20462601                                 Version 1.0                                   Page 3 of 21
FCC Part 95 (H)


                                                                   1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                                  Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

1.5       System test configuration

1.5.1     System block diagram & Support equipment

The diagram shown below details the placement of the equipment under test during the testing

                                                U, 2 m
          Helical
          Antenna                                                           Headset
                                                                              1



                                                            Ultrasound
                                                U, 1 m
                                                               3

                                                                                            Fetal
                  EUT                      U, 1 m              ECG                        Simulator


                                                                                              2
                                           U, 1 m           UA




                                                      U, 1 m                                  Remote
                                                                                               Push
                                                                                              Botton

                                                                                                  4

 S:       Shielded                    U:    Unshield                         F:      With Ferrite Core


 Item #               Description               Manufacturer             Model No.           Serial No.
    1        Headset                             Radio Shack              33.1160            Not Labeled
    2        Maternal Fetal Simulator                GE                     325              Not Labeled
    3        5700 Transducer ( Two               Corometrics                                 Not Labeled
             numbers)
      4      Remote push button                     Switch Craft            3919             Not Labeled




File: 20462601                                  Version 1.0                                       Page 4 of 21
FCC Part 95 (H)


                                                              1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                             Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

1.6     Justification

The system was configured for testing in a typical manner in accordance with ANSI C63.4 standard.
During radiated emission testing, the peripheral locations were varied with respect to the EUT to
maximize the emissions.


1.7     Mode(s) of operation

Transmitting data from patient worn transducer to fetal monitor.


1.8     Modifications required for compliance

No modifications were implemented by Intertek Testing Services.




File: 20462601                                  Version 1.0                               Page 5 of 21
FCC Part 95 (H)


                                                              1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                             Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

2.0      Test Summary

 FCC RULE          DESCRIPTION OF TEST                        RESULT                  PAGE
      2.1046      RF Power Output                              5 mW                      7

      2.1047      Modulation Characteristics              Not Applicable                 -

      2.1049      Occupied Bandwidth                           Passed                    9

      2.1053      Spurious emissions at antenna                Passed                   12
                  terminal
  95.1115(a)      Field Strength of fundamental                Passed                   16

  95.1115(b)      Field Strength of undesired                  Passed                   16
                  emissions
      15.107      AC Line Conducted                      Not Applicable.                 -
                  Emissions                         The EUT is battery operated
      2.1055      Frequency Stability                                                   20
                  Vs. Temperature                             0.4 ppm
                  Vs. Voltage                                 0.12 ppm
      2.1093      RF Exposure                                  Passed              Separate SAR
                                                                                      report




File: 20462601                                  Version 1.0                                  Page 6 of 21
FCC Part 95 (H)


                                                            1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                           Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

3.0     RF Power Output

3.1     Test Description

 Requirement:                       FCC § 2.1046
 RF Output Power



3.2     Test Procedure

The antenna was disconnected from the transmitter and a spectrum analyzer was connected to the antenna
terminal through a short cable. The output power was read out and the cable loss was setup as OFFSET.

In addition, ERP was measured by the substitution method using half-wave dipole (Robert’s antenna).


3.3     Test Results

See plot 3.1on the next page. The measured power output is 6.8 dBm (4.8 mW).
The maximum ERP measured equals 7.2 dBm or 5.2 mW.




File: 20462601                               Version 1.0                                   Page 7 of 21
FCC Part 95 (H)


                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                       Plot 3.1




File: 20462601                        Version 1.0                               Page 8 of 21
FCC Part 95 (H)


                                                             1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                            Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT


4.0     Occupied Bandwidth, Necessary Bandwidth

4.1     Test Description

 Requirement:                        FCC §2.1049
 Emission Bandwidth:                 <11 kHz



4.2     Test Procedure

The antenna was disconnected from the transmitter and the short cable was connected to the transmitter
RF output. The RF output was connected to the input of the spectrum analyzer.

The resolution bandwidth of the spectrum analyzer was setup at least 10 times higher then the necessary
bandwidth of the transmitter. The spectrum analyzer reading obtained is used as a reference for occupied
bandwidth measurements.

The resolution bandwidth of the spectrum analyzer was setup to 100 Hz and the spectrum of the
transmitting signal was recorded.

4.3     Test Results

See on the following pages the occupied bandwidth plots:

         Plot Number                    Description
                  4.1             Reference level
                  4.2             20-dB bandwidth

The Necessary Bandwidth was calculated using the formula:

               B = 2 × (M +D),
Where M is the highest modulation frequency, D is the deviation.

As per 342MCCN-XXX specification, M=0.5 kHz, D=5 kHz. Therefore B=11 kHz.




File: 20462601                                 Version 1.0                                  Page 9 of 21
FCC Part 95 (H)


                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                       Plot 4.1




File: 20462601                        Version 1.0                              Page 10 of 21
FCC Part 95 (H)


                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                       Plot 4.2




File: 20462601                        Version 1.0                              Page 11 of 21
FCC Part 95 (H)


                                                                1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                               Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

5.0     Spurious emissions at antenna terminal

5.1     Test description

 Parameter:                           FCC §2.1051
 Requirement:



5.2     Test Procedure

The test was performed according the procedure outlined in ANSI C63.4 Standard.

The antenna was disconnected from the transmitter and a spectrum analyzer was connected to the antenna
terminal through a short cable. The cable loss was setup as OFFSET.

The frequency range up to tenth harmonic of the fundamental frequency was investigated.

5.3     Test Result

See plots 5.1 – 5.3 for spurious emissions

      Data Sheet Number                      Description
                  5.1              Conducted Spurious Emissions, 1.0 MHz to 1 GHz
                  5.2              Conducted Spurious Emissions,1 GHz to 2.4 GHz
                  5.3              Conducted Spurious Emissions, 2.4 GHz to 10 GHz




File: 20462601                                    Version 1.0                              Page 12 of 21
FCC Part 95 (H)


                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                       Plot 5.1




File: 20462601                        Version 1.0                              Page 13 of 21
FCC Part 95 (H)


                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                       Plot 5.2




File: 20462601                        Version 1.0                              Page 14 of 21
FCC Part 95 (H)


                                                    1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                   Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                       Plot 5.3




File: 20462601                        Version 1.0                              Page 15 of 21
FCC Part 95 (H)


                                                               1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                              Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT


6.0     Field Strength of Radiated Emissions

6.1     Test description

 Parameter:                            FCC § 95.1115
 Requirement:                          < 200 mV/m at fundamental frequency,
                                       < 500 µV/m undesired radiation


6.2     Test Procedure

The test was performed according the procedure outlined in ANSI C63.4 Standard.

The transmitter was placed on a wooden turntable. The measurement antenna was placed at a distance of
3 meters from the EUT. During the tests, the antenna height and polarization as well as EUT azimuth
were varied in order to identify the maximum level of emissions from the EUT.

The frequency range up to tenth harmonic of the fundamental frequency was investigated.

6.3     Test Results

Please see the following pages for the test result.


      Data Sheet Number                    Description
                  6.1               Radiated Emissions at Low Channel 608 MHz
                  6.2               Radiated Emissions at High Channel 614 MHz




File: 20462601                                   Version 1.0                              Page 16 of 21
FCC Part 95 (H)


                                                                              1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                                         Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                                      Data Sheet 6.1

  Radiated Emissions
       Test Data

Company:     GE Medical Inc                             Model #:     340T                    Standard           FCC § 95.1115
EUT:          WMTS Transmitter                          FCC ID #     OU340MT                 Limit
Project #:   J20042625                                  Test Date:   May 04,20 01            Test Distance          3       meters
Test Mode:   Transmitter@608.02MHz                      Engineer:    Suresh                  Duty                   0       dB
                                                                                             Relaxation

                  Antenna Used                  Pre-Amp Used                        Cable Used                   Transducer Used
Number:            8        7      2            2         8               0          21       0         0           0
Model:        EMCO      EM LPA-     EMCO     CDI P950 CDI_P1000         None       Grn_M+L   None     None        None
               3115       25         3143

 Frequency Reading Detector Ant Amp. Ant. Pol.             Ant.        Pre-Amp     Insert.   D. C.     Net        Limit     Margin
                             .                            Factor                    Loss      F.                  @3m
    MHz       dB(µV)      P/A/Q    #    #       H/V       dB(1/m)        dB          dB       dB     dB(µV/m)    dB(µV/m)     dB
  608.02       81.9      Peak     7     0       V          19.1          0.0         1.7      0.0    102.7       106.0       -3.3
 1216.04       23.4      Ave.     8     0       V          25.0          0.0         2.5      0.0     50.9        54.0       -3.1
 1824.06       51.5      Ave.     8     8       V          26.7         29.3         3.1      0.0     52.0        54.0       -2.0
 2432.08       49.3      Ave.     8     8       V          29.1         28.5         2.3      0.0     52.2        54.0       -1.8
 3041.00       44.1      Ave.     8     8       V          31.3         28.0         2.5      0.0     49.9        54.0       -4.1
 3648.12       40.2      Ave.     8     8       V          33.0         27.8         2.7      0.0     48.1        54.0       -5.9
 4256.14       36.5      Ave.     8     8       V          34.5         27.9         2.9      0.0     46.0        54.0       -8.0
 4864.16       28.4      Ave.     8     8       V          34.0         28.1         3.2      0.0     37.5        54.0      -16.5
 5472.18       28.1      Ave.     8     8       V          35.4         28.3         3.5      0.0     38.7        54.0      -15.3
 6080.20       29.0      Ave.     8     8       V          37.1         28.3         3.9      0.0     41.7        54.0      -12.3

Notes:       a) D.C.F.:Distance Correction Factor
             b) Insert. Loss (dB) = Cable A + Cable B + Cable C .
             c) Net (dB) = Reading + Antenna Factor - Pre-amp + Insert. Loss. - Transducer Loss - Duty Relaxation (transmitter
             only).
             d) Negative signs (-) in Margin column signify levels below the limits.
             e) All other emissions not reported are below the equipment noise floor which is at least 10 dB below the limits.




File: 20462601                                    Version 1.0                                         Page 17 of 21
FCC Part 95 (H)


                                                                              1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                                         Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

                                                      Data Sheet 6.2


  Radiated Emissions
       Test Data

Company:     GE Medical Inc                             Model #:     340T                    Standard       FCC 95.1115
EUT:          WMTS Transmitter                          S/N #:       OU340MT                 Limits
Project #:   J20042625                                  Test Date:   May 04,20 01            Test Distance_     3     meters
Test Mode:   Transmitter@613.97MHz                      Engineer:    Suresh K                Duty               0     dB
                                                                                             Relaxation

                  Antenna Used                  Pre-Amp Used                        Cable Used                  Transducer Used
Number:            8        7     11            2         8               0          21       0         0          0
Model:        EMCO      EM LPA-   LPB-2520A CDI P950     CDI P1000      None       Grn_M+L   None     None       None
               3115       25

 Frequency Reading Detector Ant Amp. Ant. Pol.             Ant.        Pre-Amp     Insert.   D. C.     Net       Limit     Margin
                                                          Factor                    Loss      F.                 @3m
    MHz       dB(µV)      P/A/Q    #    #       H/V       dB(1/m)        dB          dB       dB     dB(µV/m)   dB(µV/m)     dB
  613.97   84.5      Peak 11 0               V             18.9          0.0         1.7     0.0     105.1       106        -0.9
 1227.95   23.3       Ave. 8         0       V             25.0          0.0         2.5     0.0      50.8       54.0       -3.2
 1841.92   46.5       Ave. 8         8       V             26.7         29.3         3.1     0.0      47.0       54.0       -7.0
  2455.9   41.8       Ave. 8         8       V             29.1         28.5         2.3     0.0      44.7       54.0       -9.3
  3069.9   40.2       Ave. 8         8       V             31.3         28.0         2.5     0.0      46.0       54.0       -8.0
  3683.8   38.8       Ave. 8         8       V             33.0         27.8         2.7     0.0      46.7       54.0       -7.3
  4297.8   36.6       Ave. 8         8       V             34.5         27.9         2.9     0.0      46.1       54.0       -7.9
  4911.8   28.8       Ave. 8         8       V             34.0         28.1         3.2     0.0      37.9       54.0      -16.1
  5525.7   33.8       Ave. 8         8       V             36.6         28.3         3.7     0.0      45.8       54.0       -8.2
  6139.7   30.7       Ave. 8         8       V             37.1         28.3         3.9     0.0      43.4       54.0      -10.6
Notes:   a) D.C.F.:Distance Correction Factor
             b) Insert. Loss (dB) = Cable A + Cable B + Cable C .
             c) Net (dB) = Reading + Antenna Factor - Pre-amp + Insert. Loss. - Transducer Loss - Duty Relaxation (transmitter
             only).
             d) Negative signs (-) in Margin column signify levels below the limits.
             e) All other emissions not reported are below the equipment noise floor which is at least 10 dB below the limits.




File: 20462601                                    Version 1.0                                         Page 18 of 21
FCC Part 95 (H)


                                                                1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                               Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

7.0     Frequency Stability

7.1     Test description

 Requirement:                        FCC § 2.1055
 Frequency Tolerance:



7.2     Test Procedure

The ppm frequency error of the transmitter was calculated by:

             MCF    
ppm error =      − 1 ⋅ 10 6
             ACF    
Where MCF is the Measured Carrier Frequency in MHz
      ACF is the Assigned Carrier Frequency in MHz


7.2.1   Frequency Stability vs. Temperature

The equipment under test was connected to an external DC power supply and the RF output was
connected to a frequency counter via feedthrough attenuator. The EUT was placed inside the temperature
chamber.

After the temperature stabilized for approximately 20 minutes, the frequency of the output signal was
recorded from the counter.


7.2.2   Frequency Stability vs. Voltage

At room temperature (25 ±5 oC), an external variable DC power supply was connected to the EUT. The
frequency of the transmitter was measured for 115%, 100% and 85% of the nominal operating input
voltage.




File: 20462601                                 Version 1.0                                  Page 19 of 21
FCC Part 95 (H)


                                                            1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                           Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

7.3       Test Results



                             Frequency Stability vs Temperature
                     ACF (MHz): 610.999790                                     Limit: N/A
          Temperature, C                  MCF (MHz)                            PPM Error
                50                        610.999600                              0.19
                40                        610.999740                              0.05
                30                        610.999710                              0.04
                20                        611.000015                              .022
                10                        611.000190                              0.40
                0                         611.000505                              0.40
               -10              Unit turned off by thermal sensor                   -
               -20                              -                                   -


                            AFC (MHz) 611.11                                      Limit: 2 ppm
             %                  Voltage                    MCF (MHz)               PPM Error
            115                   6.9                      610.999823                0.003
            100                  6.0V                      610.999913                 0.12
             85                   5.1                      610.999798                0.001
      Battery Endpoint            4.0                      610.999798                0.001

Maximum frequency deviation equals 0.4 ppm




File: 20462601                               Version 1.0                               Page 20 of 21
FCC Part 95 (H)


                                                             1365 Adams Court, Menlo Park, CA 94025
GE Medical Systems., Model No: 342MCCN-XXX                            Date of Test: May 1 to 10, 2001
FCC ID: OU5340MT

8.0     List of test equipment


        EQUIPMENT            MANUFACTURER              MODEL           SERIAL        CAL.         CAL.
                                                      NUMBER          NUMBER      INTERVAL         DUE
Spectrum Analyzer w/85650    Hewlett Packard       8566B             2416A00317       12          4/6/02
QP Adapter                                                           2043A00251
Spectrum Analyzer w/8650     Hewlett Packard       8568B             1912A0053        12         2/23/02
QP Adaptor                                                           2521A01021
Spectrum Analyzer            Tektronix             2784              B3020108         12         8/4/01
Bi-Log Antenna               EMCO                  3143              9509-1160        12         6/8/01
Log Periodic Antenna         EMCO                  LPA-25            1079             12         7/05/02
Double-ridged Horn Antenna   EMCO                  3115              9107-3712        12         3/17/02
Pre-Amplifier                CDI                   P950              ITS009           12         10/6/01
Pre-Amplifier                CDI                   P1000             N/A              12         10/6/01
Power Meter                  Hewlett Packard       8900D             3607U00673       12         7/31/01




File: 20462601                                 Version 1.0                                 Page 21 of 21
FCC Part 95 (H)