Regulatory and Safety Manual
2109599-003 A




                               MAC™ 7 Resting ECG Analysis System
                               Version 1.00
                               English
                               © 2018 General Electric Company
                               All Rights Reserved.


Publication Information
This document describes version 1.00 of MAC™ 7 Resting ECG Analysis System, also referred to as the “product” or "system" or "device".
It does not apply to earlier product versions. Due to continuing product innovation, specifications in this document are subject to change
without notice.
MUSE, MAC, MARS, CASE, CardioSoft, and 12SL are trademarks owned by GE Medical Systems Information Technologies, Inc., a General
Electric Company going to market as GE Healthcare. All other trademarks contained herein are the property of their respective owners.
The MAC ™7 Resting ECG Analysis System is intended to be used, under the direct supervision of a licensed healthcare practitioner by
trained operators in a hospital or facility providing patient care.
This document provides information required for the proper use of the system. Familiarize yourself with this information and read and
understand all instructions before attempting to use this system.
NOTE:
   Illustrations in this document are provided as examples only. Depending on system configuration, screens in the document may differ
   from the screens on your system. Patient names and data are fictitious. Any similarity to actual persons is coincidental.
The document part number and revision are on each page of the document. The revision identifies the document’s update level. The
revision history of this document is summarized in the following table.


 Revision             Date                            Comment

 A                    15 August 2018                  Initial Release


To access other GE Healthcare Diagnostic Cardiology documents, go to the Common Documentation Library (CDL), located at http://
apps.gehealthcare.com/servlet/ClientServlet?REQ=Enter+Documentation+Library, and click Cardiology.
To access Original Equipment Manufacturer (OEM) documents, go to the device manufacturer's website.
Support
GE Healthcare maintains a trained staff of application and technical experts to answer questions and to respond to issues and problems
that may arise during the installation, maintenance, and use of this product.
If you require additional assistance, contact your GE Healthcare representative or GE Healthcare support at one of the following numbers:
•    North America: 1-800-558-7044
•    Europe: +49 761 45 43 -0
•    Asia: +86 21 3877 7888
Training
This document is intended as a supplement to, not a substitute for, thorough product training. If you have not received training on the use
of the product, you should request training assistance from GE Healthcare.
To see available training, go to the GE Healthcare training website (www.gehealthcare.com/training). Select Education > Product
Education-Technical > Diagnostic Cardiology. For more self-paced course offerings, tools, and reference guides you may find useful,
please visit the GE Healthcare Education Store at www.gehealthcare.com/educationstore.




2                                        MAC™ 7 Resting ECG Analysis System                                                2109599-003 A


Table of Contents


                Publication Information....................................................................... 2

1: Regulatory Information....................................................................................... 4
                      Intended Use............................................................................................................................................................ 4
                      Indications for Use................................................................................................................................................ 4
                      Contraindications................................................................................................................................................... 5
                      Prescription Device Statement........................................................................................................................ 5
                      Classification of Medical Device..................................................................................................................... 5
                      Recording ECGs during Defibrillation........................................................................................................... 5
                      Modulating Effects in Digital Systems......................................................................................................... 6
                      Biocompatibility.......................................................................................................................................................6
                      Legal Notice..............................................................................................................................................................6
                      Responsibility of the Manufacturer............................................................................................................... 7
                      Responsibility of the Purchaser/Customer................................................................................................ 7
                      Warranty Information..........................................................................................................................................7


2: Safety Information............................................................................................... 8
                      Safety Conventions............................................................................................................................................... 8
                      Safety Hazards........................................................................................................................................................ 8
                      Supplies and Accessories................................................................................................................................12


3: Product and Packaging Information...............................................................13
                      Hardware Label Locations..............................................................................................................................13


4: Equipment Identification...................................................................................17
                      Serial Number Label.......................................................................................................................................... 17
                      Device Address Label and Rating Plate................................................................................................... 17
                      Symbol Descriptions.......................................................................................................................................... 18
                      Serial Number Format...................................................................................................................................... 22
                      Unique Device Identifier...................................................................................................................................22


5: Electromagnetic Compatibility (EMC)............................................................. 23
                      Guidance and Manufacturer’s Declaration—Electromagnetic Emissions................................24
                      Guidance and Manufacturer’s Declaration—Electromagnetic Immunity.................................24
                      Recommended Separation Distances.......................................................................................................27


6: Wireless Regulations......................................................................................... 29
                      FCC Compliance...................................................................................................................................................29
                      IC Compliance.......................................................................................................................................................30
                      RED Information...................................................................................................................................................31


Regulatory Information




                                                                                       1
Regulatory Information
                Familiarize yourself with this information before attempting to use this system.
                Keep this manual with your Operator Manual and equipment at all times, and
                periodically review it.
                This section provides information about the regulatory compliance of this system.
                The system software is considered medical software. As such, it was designed and
                manufactured to the appropriate medical regulations and controls.


Intended Use
                The system is intended to acquire, analyze, display, and record electrocardiographic
                information from adult, pediatric or neonatal populations. Basic system delivers 3,6,
                or 12 lead ECG’s and interpretive analysis. Transmission and reception of ECG data
                and other clinical data to and from a central clinical information system is optional.
                The system is intended to be used under the direct supervision of a licensed
                healthcare practitioner, by trained operators in a hospital or medical professional’s
                facility.


Indications for Use
                This device is a non-invasive prescription device.
                The device is indicated for use:
                • To acquire, analyze, display and print electrocardiograms.
                • To provide interpretation of the data for consideration by a physician.
                • In a clinical setting, by a physician or by trained personnel who are acting on the
                  orders of a licensed physician. It is not intended as a sole means of diagnosis.
                  The interpretations of ECG offered by the device are only significant when used
                  in conjunction with a physician over-read as well as consideration of all other
                  relevant patient data.
                • On adult and pediatric (birth through 21 years of age) populations.



4                             MAC™ 7 Resting ECG Analysis System                            2109599-003 A


                                                                                                          Regulatory Information




Contraindications
                This system is not intended for use in the following manner:
                • During patient transport
                • With high-frequency surgical units
                • As an intra-cardiac application
                • As a sole means of diagnosis
                • As a vital signs physiological monitor


Prescription Device Statement
                CAUTION:
                   United States federal law restricts this device to sale by or on the order of a
                   physician.


Classification of Medical Device
                The device is classified as follows according to IEC 60601-1.

                Table 1: Medical Device Classifications

                Category                                          Classification

                Type of protection against electrical shock       Class I, internally powered

                Degree of protection against electrical shock     Type CF defibrillation-proof applied part

                Degree of protection against harmful ingress of   The Ingress Protection (IP) code for this device is
                solids and liquids                                IP 20.

                Degree of safety of application in the presence   Equipment is not suitable for use in the presence
                of a flammable anesthetic mixture with air or     of a flammable anesthetic mixture with air or
                with oxygen or with nitrous oxide                 with oxygen or with nitrous oxide

                Method(s) of sterilization or disinfection        Not applicable
                recommended by the manufacturer

                Mode of operation                                 Continuous operation



Recording ECGs during Defibrillation
                This equipment is protected against the effects of cardiac defibrillator discharge
                to ensure recovery, as required by test standards. The patient signal input of the
                acquisition module is defibrillation-proof. It is not necessary to remove the ECG
                electrodes prior to defibrillation.

2109599-003 A                                   MAC™ 7 Resting ECG Analysis System                                      5


Regulatory Information




                When using stainless steel or silver electrodes, a defibrillator discharge current may
                cause the electrodes to retain a residual charge causing a polarization or DC offset
                voltage. This electrode polarization blocks acquisition of the ECG signal. To avoid this
                condition, if there is a situation where a defibrillation procedure is necessary, use
                non-polarizing electrodes such as silver/silver-chloride types, which do not form a DC
                offset voltage when subjected to a DC current.
                If you use polarizing electrodes, GE Healthcare recommends disconnecting the
                leadwires from the patient before delivering the shock.
                Electrode defibrillation recovery is the ability of the electrode to allow the ECG trace
                to return after defibrillation. GE Healthcare recommends using non-polarizing
                disposable electrodes with defibrillation recovery rating as specified in AAMI
                EC12.5.2.2.4. AAMI EC12 requires that the polarization potential of an electrode pair
                does not exceed 100 mV 5 seconds after a defibrillation discharge.
                Refer to the supplies and accessories guide for this system for a list of approved
                electrodes.


Modulating Effects in Digital Systems
                This section describes the modulating effects that may occur in digital systems of the
                product.
                This device uses digital sampling techniques that may produce some variation in
                amplitudes of Q, R, and/or S waves from one heart beat to the next, which may be
                particularly noticeable in pediatric recordings. If you observe this phenomenon,
                be aware that the origin of amplitude variations is not entirely physiological. For
                measuring voltages of Q, R, and S waves, GE Healthcare advises using the QRS
                complexes with the largest deflection of the particular waves.


Biocompatibility
                The parts of the system described in this manual that come into contact with the
                patient during the intended use, including all accessories, fulfill the biocompatibility
                requirements of the applicable standards. If you have questions in this matter contact
                your GE Healthcare representative.


Legal Notice
                GE Healthcare software contains several fields that can be populated before
                performing an ECG. Some of these fields are required, others are optional and
                left to the user to assess whether they are needed to perform the exam. The field
                Race is one of these optional fields. Race has been acknowledged by the medical
                profession as useful to analyze some pathologies. You should be aware that, in some
                jurisdictions, the processing of data revealing an individual’s racial origin is subject
                to legal requirements, such as obtaining the patient’s prior consent. If you elect to


6                             MAC™ 7 Resting ECG Analysis System                             2109599-003 A


                                                                                               Regulatory Information




                collect this type of data, it is your responsibility to ensure that you comply with all
                applicable legal requirements.


Responsibility of the Manufacturer
                This section describes the responsibility of GE Healthcare as the manufacturer of your
                product.
                GE Healthcare is responsible for the safety, reliability, and performance of hardware
                supplied by GE Healthcare only if the following conditions are met:
                • Assembly operations, extensions, readjustments, modifications, or repairs are
                  performed by persons authorized by GE Healthcare.
                • The electrical installation of the room where the device is used complies with the
                  requirements of the appropriate local, state, and other government regulations.
                • The equipment is used in accordance with the instructions for use.


Responsibility of the Purchaser/Customer
                The customer is responsible for providing appropriate desks, chairs, electrical wall
                outlets, network connections, and analog phone lines, and for locating any of the
                system components described in the manuals in compliance with all local, state, and
                national codes.
                Lack of data security may compromise patient privacy. GE Healthcare recommends
                that you take appropriate steps to secure the privacy of communication on your
                network when using this product.


Warranty Information
                This device is considered GE Healthcare-supplied hardware. Only authorized GE
                Healthcare service personnel should service the device. Any unauthorized attempt to
                repair equipment under warranty voids that warranty. It is the user’s responsibility to
                report the need for service to GE Healthcare or to one of their authorized agents.




2109599-003 A                              MAC™ 7 Resting ECG Analysis System                              7


Safety Information




                                                                                                  2
Safety Information
Safety Conventions
                This section describes the safety conventions used in the documentation for the
                product.
                A Hazard is a source of potential injury to a person, property, or the system.
                The manuals for this system use the terms DANGER, WARNING, CAUTION, and NOTICE
                to point out hazards and to designate a degree or level of seriousness. Familiarize
                yourself with the following definitions and their significance.

                Table 2: Definitions of Safety Conventions

                 Safety Convention    Description

                 DANGER               Indicates an imminent hazard, which, if not avoided will result in death or
                                      serious injury.

                 WARNING              Indicates a potential hazard of unsafe practice, which, if not avoided, could
                                      result in death or serious injury.

                 CAUTION              Indicates a potential hazard or unsafe practice, which, if not avoided, could
                                      result in moderate or minor injury.

                 NOTICE               Indicates a potential hazard or unsafe practice, which, if not avoided, could
                                      result in the loss or destruction of property or data



Safety Hazards
                The following safety messages alert you to potentially hazardous conditions that
                could arise during the normal use of this product and recommend steps that can be
                taken to avoid those conditions. Safety messages pertaining to hazardous conditions
                that may arise during specific actions may also be provided during the discussion of
                those actions in this or other manuals for this product.
                WARNING:
                  EQUIPMENT MALFUNCTION - Any attempt by unauthorized personnel to service
                  the device could result in equipment malfunction and void the warranty. This
8                              MAC™ 7 Resting ECG Analysis System                                       2109599-003 A


                                                                                                 Safety Information




                   equipment contains no user-serviceable parts. Refer servicing to authorized
                   service personnel.

                CAUTION:
                  EQUIPMENT COMPATIBILITY
                   Compatibility is critical to safe and effective use of this device. Please contact
                   your local sales or service representative prior to installation to verify equipment
                   compatibility.

                WARNING:
                  PERSONAL INJURY-STUMBLING HAZARD - Patients can become entangled in the
                  cables and leadwires connected to the device, which could cause the patient to
                  stumble or trip.
                   Route cables and leadwires in a way to avoid creating a stumbling hazard: keep
                   them off the floor, and route leadwires away from the patient’s legs and the
                   healthcare provider’s work area.

                WARNING:
                  MAGNETIC AND ELECTRICAL INTERFERENCE - Magnetic and electric fields can
                  interfere with the acquisition of ECG readings.
                   Make sure that all peripheral components operated in the vicinity of the device
                   comply with the relevant EMC requirements. X-ray equipment, MRI devices,
                   radio systems (cellular phones) and so forth, are possible sources of interference
                   because they may emit higher levels of electromagnetic radiation. Verify the
                   performance of the system before use.

                WARNING:
                  EXPLOSION HAZARD - Using this device in the presence of anesthetic vapors or
                  liquids can cause explosions.
                   Do not use this device in the presence of anesthetic vapors or liquids. Only persons
                   with adequate training in the correct use of this device may use this device.

                WARNING:
                  EQUIPMENT FAILURE - Polarizing electrodes (stainless steel or silver constructed)
                  may cause the electrodes to retain a residual charge after defibrillation. A residual
                  charge blocks acquisition of the ECG signal.
                   Whenever patient defibrillation is a possibility, use non-polarizing electrodes
                   (silver-silver chloride construction) for ECG monitoring.

                CAUTION:
                  EQUIPMENT DAMAGE
                   Devices intended for emergency application must not be exposed to low
                   temperatures during storage and transport to avoid moisture condensation at the
                   application site. Wait until all moisture has vaporized before using the device.




2109599-003 A                             MAC™ 7 Resting ECG Analysis System                              9


Safety Information




                WARNING:
                  PERSONAL INJURY - Contact with patients during defibrillation can cause serious
                  injury or death.
                     Do not contact patients during defibrillation. Patient signal inputs labeled with
                     the CF symbols with paddles are protected against damage resulting from
                     defibrillation voltages. To ensure proper defibrillator protection, use only GE
                     Healthcare recommended cables and leadwires. Proper placement of defibrillator
                     paddles in relation to the electrodes is required to ensure successful defibrillation.

                CAUTION:
                  EXPLOSION HAZARD
                     Do NOT use in the presence of flammable anesthetics vapors or liquids.

                WARNING:
                  INTERPRETATION HAZARD - Results of the automated QT analysis are not
                  considered a diagnosis.
                     A qualified physician or cardiologist must review and confirm the measurements
                     and waveforms recorded by the system. It should be used only as an adjunct to
                     the clinical history, symptoms, and results of other tests.

                WARNING:
                  INTERPRETATION HAZARD - Computerized interpretation is only significant when
                  used in conjunction with clinical findings.
                     A qualified physician must overread all computer-generated tracings.

                WARNING:
                  IMPROPER USE - This is a prescriptive device.
                     This equipment is intended for use by or under the direct supervision of a licensed
                     healthcare practitioner.

                WARNING:
                  BATTERY EXPLOSION HAZARD - Batteries may explode in fires.
                     Do not dispose of the battery by fire or burning. Follow local environmental
                     guidelines concerning disposal and recycling.

                WARNING:
                  ELECTRIC SHOCK HAZARD/SYSTEM MALFUNCTION - Liquids inside a device can
                  cause electric shock or system malfunction.
                     Do not allow liquids to enter the device. If liquids enter the device, turn it off and
                     inform your service technician. Do not use the device until it is checked by a
                     service technician.

                WARNING:
                  ELECTRIC SHOCK - Improper connection of this equipment may cause electric
                  shock.
                     To avoid risk of electric shock, this equipment must only be connected to a supply
                     mains with protective earth.

10                               MAC™ 7 Resting ECG Analysis System                             2109599-003 A


                                                                                                 Safety Information




                WARNING:
                  EQUIPMENT MALFUNCTION/INTERFERENCE - Use of portable phones or other radio
                  frequency (RF) emitting equipment near the system may cause unexpected or
                  adverse operation.
                   Do not use portable phones or other electronic equipment that may emit radio
                   frequency (RF) near this system.

                WARNING:
                  EQUIPMENT MALFUNCTION/INTERFERENCE - Do not use the equipment or system
                  adjacent to, or stacked with, other equipment.
                   If adjacent or stacked use is necessary, test the equipment or system to verify
                   normal operation in the configuration in which you are using it.

                WARNING:
                  ACCESSORIES/COMPONENTS - Adding accessories or components, or modifying
                  the medical device or system, may result in increased EMISSIONS or decreased
                  IMMUNITY of the device or system.

                CAUTION:
                  ACCESSORIES/COMPONENTS - Ensure that all cables are less than 3 meters long.
                  Cable lengths greater than 3 meters may result in decreased IMMUNITY of the
                  device or system.

                CAUTION:
                  ACCESSORIES (SUPPLIES)
                   To ensure patient safety, use only parts and accessories manufactured or
                   recommended by GE Medical Systems Information Technologies. Parts and
                   accessories used must meet the requirements of the applicable IEC 60601 series
                   safety standards and essential performance standards, and/or the system
                   configuration must meet the requirements of the IEC 60601-1-1 medical electrical
                   systems standard.

                CAUTION:
                  ACCESSORIES (EQUIPMENT)
                   The use of ACCESSORY equipment not complying with the equivalent safety
                   requirements of this equipment may lead to a reduced level of safety of the
                   resulting system. Consideration relating to the choice shall include: Use of the
                   accessory in the PATIENT VICINITY; and Evidence that the safety certification of the
                   ACCESSORY has been performed in accordance to the appropriate IEC 60601-1
                   and/or IEC 60601-1-1 harmonized national standard.

                WARNING:
                  DATA LOSS - Formatting the device’s internal flash drive erases all the data in
                  memory and returns the device to its factory settings.
                   If possible, back up or record any data that you do not want to lose before
                   performing the following procedure.


2109599-003 A                            MAC™ 7 Resting ECG Analysis System                            11


Safety Information




                WARNING:
                  EQUIPMENT FAILURE AND HEALTH HAZARDS - Failure on the part of all responsible
                  individuals, hospitals or institutions, employing the use of this system, to
                  implement the recommended maintenance schedule, may cause equipment
                  failure and possible health hazards. The sole responsibility rests with the
                  individuals, hospitals, or institutions utilizing the system. The manufacturer does
                  not, in any manner, assume the responsibility for performing the recommended
                  maintenance schedule, unless an Equipment Maintenance Agreement exists.

                CAUTION:
                  DISPOSAL HAZARD
                     Disposable devices are intended for single use only. They should not be reused as
                     performance may degrade or contamination could occur.

                CAUTION:
                  DISPOSAL HAZARD - At the end of its service life, the product described in this
                  manual, as well as its accessories, must be disposed of in compliance with local,
                  state, or federal guidelines regulating the disposal of such products.
                     If you have questions concerning disposal of the product, please contact GE or its
                     representatives.

                CAUTION:
                  INTERCONNECTED DEVICES - When several items of medical equipment are
                  interconnected, summation of leakage current must meet the leakage current as
                  per IEC 60601-1.
                     Connect the device only to the GE approved supplies and accessories.

                CAUTION:
                  ISOLATION FROM SUPPLY MAINS
                     Do not position the device so that it is difficult to operate the disconnection of the
                     AC power supply of the device.


Supplies and Accessories
                This section is in regard to the supplies and accessories you may purchase for your
                product.
                You should use only supplies and accessories recommended by GE Healthcare. For a
                list of recommendations, refer to the supplies and accessories guide for this system.
                Contact GE Healthcare before using anything that is not recommended for this
                system.




12                              MAC™ 7 Resting ECG Analysis System                             2109599-003 A


                                                                                 Product and Packaging Information




                                                                                       3
Product and Packaging Information
Hardware Label Locations
                The following illustrations and table describe the labels and their location on your
                device and its packaging.




2109599-003 A                             MAC™ 7 Resting ECG Analysis System                           13




                                                                                          Product and Packaging Information




                Table 3: Label Descriptions on the Device and Packaging

                Item   Label              Location        Description

                1      Device Address Back of the         Provides regulatory and cautionary information. See
                       Label and Rating device            "Device Address Label and Rating Plate" on page 17
                       Plate                              for an explanation of the label.

                2      Country-specific   Back of the     Contains registration information for China and Latin
                       Label              device          America.

                3      Serial Number      Back of the     Identifies this device. See "Serial Number Label" on
                       Label              device          page 17 for a description of the label contents.

                4      Option Code        Back of the     Use the option codes to setup the purchased options
                       Label              device          in your system.




                5      Wireless Label     Back of the     Depending on the customer's wireless option, the label
                                          device          will differ:
                                                          Default wireless option:




                                                          FIPS wireless option:




                6      MAC Address        Back of the     Indicates the MAC address of the device.
                       Label              device




2109599-003 A                                MAC™ 7 Resting ECG Analysis System                                    15


Product and Packaging Information




                 Item   Label            Location          Description

                 7      Shipping Label   On the shipping   Indicates the shipping address where the device is
                                         package           delivered.




                 8      RED Lable        On the shipping   Contains registration information of Radio Equipement
                                         package           Directive.




                 9      Environmental                      These symbols containing warning and caution
                        Symbols                            indicators are required for shipping. For a full
                                                           description of symbols, see "Symbol Descriptions" on
                                                           page 18.

                 10     Battery Shipping On the shipping   FRAGILE—Lithium Ion batteries can cause fire if
                        Label            package           damaged.




16                              MAC™ 7 Resting ECG Analysis System                                    2109599-003 A


                                                                                        Equipment Identification




                                                                                        4
Equipment Identification
Serial Number Label
                The serial number label is in the following format:




                Table 4: Serial Number Label Format

                Item   Description

                1      Product Name

                2      Date of Manufacture in YYYY-MM-DD Format

                3      Country of Origin

                4      Device Serial Number

                5      UDI Barcode

                6      Product Part Number



Device Address Label and Rating Plate
                The device address label and rating plate is in the following format:




2109599-003 A                                 MAC™ 7 Resting ECG Analysis System                    17


Equipment Identification




                 Table 5: Device Address Label and Rating Plate Format

                 Item       Description

                 1          Product name

                 2          Manufacturer name and address

                 3          Electrical rating of the device

                 4          Symbols
                            See "Symbol Descriptions" on page 18 for a description of the symbols used on this
                            label.



Symbol Descriptions
                 The following table describes symbols or icons that are on the device or its packaging.
                 Any symbol on your device or packaging with markings in color indicates there
                 may be a danger, warning, or mandatory action. Any symbol on your device or
                 packaging that is in black and white provides additional information or may indicate
                 a caution. Familiarity with these symbols assists in the proper use and disposal of the
                 equipment.
                 For equipment symbols not shown, refer to the original equipment manufacturer
                 (OEM) manuals.

                 Table 6: Symbols, Icons, and Descriptions

                 Symbol                 Description

                                        Catalog or Orderable Part Number
                                        Indicates the manufacturer’s catalog or part number.

                                        Serial Number
                                        Indicates the manufacturer's serial number.


18                              MAC™ 7 Resting ECG Analysis System                                     2109599-003 A


                                                                                           Equipment Identification




                Symbol   Description

                         Batch Code or Lot Number
                         Indicates the manufacturer's batch code or lot number.

                         Date of Manufacture (Year-Month-Date)
                         Indicates the original manufacture date for this device.


                         Manufacturer
                         Indicates the name and address for the manufacturer of this device. It may
                         also include the date it was manufactured.

                         Rx Only
                         US Federal law restricts this device to sale by or on the order of a physician.

                         12SL
                         Indicates the device uses the Marquette™ 12SL ECG Analysis Program to
                         analyze and interpret ECG readings.

                IPxx     IP Code (Ingress Protection Rating)
                         Classifies and rates the degree of protection provided against the intrusion
                         of solid objects (such as body parts like hands and fingers, dust, accidental
                         contact), and liquids.
                         The first numeral (x) represents the degree of protection against the ingress
                         of solid objects.
                         The second numeral (x) represents the degree of protection against the
                         ingress of liquids.
                         For products with an IPxx rating, see the Classification of Medical Device in
                         this chapter for a description of that rating. Not all products have an IPxx
                         rating.

                         Regulatory Compliance Mark (RCM)
                         Indicates compliance with electrical safety, EMC, EME and
                         telecommunications requirements, as applicable to the product.
                         Required for Australia and New Zealand.

                         Wireless Communication
                         Indicates that that the equipment can be connected through wireless
                         communication.

                         Waste Electrical and Electronic Equipment (WEEE)
                         Indicates this equipment contains electrical or electronic components
                         that must not be disposed of as unsorted municipal waste but collected
                         separately. Contact an authorized representative of the manufacturer for
                         information concerning the decommissioning of your equipment.




2109599-003 A                   MAC™ 7 Resting ECG Analysis System                                         19


Equipment Identification




                 Symbol          Description

                                 Follow Instructions For Use
                                 Read and understand the operator's manual before using the device or
                                 product.
                                 As a mandatory action sign, this symbol is identified by a blue background
                                 and white symbol.

                                 This Way Up
                                 Indicates the correct upright position of the package.



                                 Keep Dry
                                 Indicates that you need to keep the container away from rain and other
                                 sources of moisture.


                                 Can Be Recycled
                                 Indicates you may recycle this material or device. Recycle or dispose of in
                                 accordance with local, state, or country laws.

                                 Temperature Limits
                                 Indicates the upper and lower temperature limits for the transportation
                                 and handling of this package. They are indicated next to the upper and
                                 lower horizontal lines.

                                 Humidity Limits
                                 Indicates upper and lower humidity limits for the transportation and
                                 handling of this package. They are indicated next to the upper and lower
                                 horizontal lines.

                                 Atmospheric Limits
                                 Indicates the upper and lower barometric pressure limitations for the
                                 transportation and handling of this package. They are indicated next to the
                                 upper and lower horizontal lines.

                                 Defibrillation-proof Type CF Applied Part
                                 Identifies a defibrillation-proof type CF applied part on medical equipment
                                 that complies with IEC 60601–1.
                                 This device meets the requirements for protection against electric shock for
                                 an earth-free (floating) applied part (one intended for contact with patients)
                                 for cardiac application.

                                 No User– or Field-serviceable Parts
                                 Do not open or disassemble the device for any reason.


                                 Protective Earth (ground)
                                 Identifies the terminal of a protective earth (ground) electrode or any
                                 terminal that is intended for connection to an external conductor for
                                 protection against electric shock in case of a fault.

20                         MAC™ 7 Resting ECG Analysis System                                      2109599-003 A


                                                                                         Equipment Identification




                Symbol   Description

                         Do Not Stack
                         Indicates that you should not stack the container or place a load on the
                         container.


                         CAUTION:
                            ELECTRIC SHOCK - Indicates the presence of hazardous energy circuits
                            or electric shock hazards.
                             To reduce the risk of electric shock hazards, do not open this enclosure.
                             Refer servicing to qualified personnel.

                         Equipotentiality
                         Connect non-grounded peripheral devices to ensure equipotential.




                         Environmental Friendly Use Period (EFUP)
                         Per Chinese standard SJ/T11364–2014, indicates the number of years from
                         the date of manufacture during which you can use the product before any
                         restricted substances are likely to leak, causing a possible environmental or
                         health hazard.
                         NOTE:
                             •     If the device contains less than the maximum concentration of
                                   restricted substances, the symbol contains a lowercase e
                             •     This is also referred to as China RoHS.

                         Fragile
                         Indicates the contents are fragile. Handle with care.


                         CAUTION:
                            SAFETY GROUND PRECAUTION - Pulling on the cable can cause the
                            cord to deteriorate resulting in electrical problems.
                             Remove the power cord from the mains source by grasping the plug.
                             DO NOT pull on the cable.

                         Contains <heavy metal chemical symbol>
                         Indicates this equipment contains heavy metal and must not be disposed
                         of as unsorted municipal waste but collected separately. The example
                         shows Lithium Ion.


                         Pushing Prohibited




2109599-003 A                 MAC™ 7 Resting ECG Analysis System                                         21


Equipment Identification




Serial Number Format
                 Each device has a serial number that uniquely identifies the device and provides
                 important information about the device. The serial number format is shown in the
                 following illustration:




                 Table 7: Serial Number Format

                 Item      Name                 Description

                 1         Product Code         A three-character code that uniquely identifies the product line.

                 2         Year Manufactured    A two-digit code identifying the year the device was manufactured.
                                                Values range from 00 to 99. For example: 00 = 2000, 04 = 2004, 17 =
                                                2017 (and so on).

                 3         Fiscal Week          A two-digit code identifying the week the device was manufactured.
                           Manufactured         Values range from 01 to 52. GE Healthcare's fiscal weeks
                                                correspond to the calendar week. For example, 01 = the first week in
                                                January.

                 4         Product Sequence     A four-digit number identifying the order in which this device was
                                                manufactured.
                                                Values range from 0001 to 9999.

                 5         Manufacturing Site   A one-letter code identifying the site where the device was
                                                manufactured.
                                                For example, F = Milwaukee, N = Freiburg, P = Bangalore, W = Wuxi,
                                                H = Helsinki, S = Mexico

                 6         Miscellaneous        A one-letter code identifying manufacturing status.
                           Characteristic       For example, P = the device is a prototype, R = the device was
                                                refurbished, U = the device was upgraded to meet the specifications
                                                of another product code, A = device is in production.



Unique Device Identifier
                 Medical devices require a unique marking for identification—the Unique Device
                 Identifier (UDI). In the event that you need the UDI for this product, check the product
                 label on the back of the device.




22                                 MAC™ 7 Resting ECG Analysis System                                     2109599-003 A


                                                                                Electromagnetic Compatibility (EMC)




                                                                                       5
Electromagnetic Compatibility (EMC)
                Before installing or using the device or system, be aware of the proximity of known
                radio frequency (RF) sources, such as the following:
                • Radio and TV stations
                • Portable and mobile RF communication devices (cell phones, two-way radios)
                • X-ray, CT, or MRI devices
                   These devices are also possible sources of interference as they may emit higher
                   levels of electromagnetic radiation.
                WARNING:
                  EQUIPMENT MALFUNCTION
                   Use of this equipment adjacent to or stacked with other equipment should be
                   avoided because it could result in improper operation. If such use is necessary, this
                   equipment and the other equipment should be observed to verify that they are
                   operating normally.Portable RF communications equipment (including peripherals
                   such as antenna cables and external antennas) should be used no closer than 30
                   cm (12 inches) to any part of the [ME EQUIPMENT or ME SYSTEM], including cables
                   specified by the manufacturer. Otherwise, degradation of the performance of this
                   equipment could result.

                WARNING:
                  EQUIPMENT MALFUNCTION OR INTERFERENCE
                   Use of this equipment adjacent to or stacked with other equipment should be
                   avoided because it could result in improper operation. If such use is necessary, this
                   equipment and the other equipment should be observed to verify that they are
                   operating normally.

                WARNING:
                  PATIENT SAFETY/EQUIPMENT FAILURE
                   Use of accessories, transducers and cables other than those specified or provided
                   by the manufacturer of this equipment could result in increased electromagnetic
                   emissions or decreased electromagnetic immunity of this equipment and result in
                   improper operation.



2109599-003 A                             MAC™ 7 Resting ECG Analysis System                           23


Electromagnetic Compatibility (EMC)




Guidance and Manufacturer’s Declaration—
Electromagnetic Emissions
                The system described in this document is intended for use in the following specified
                electromagnetic environment. It is the responsibility of the customer or user to ensure
                that this system is used in such an environment.

                Table 8: EMC Emissions Test

                 Emissions Test                  Compliance              Electromagnetic Environment

                 RF emissions (Radiated)         Group 1                 This device uses RF energy
                                                 Class B                 only for its internal function.
                 EN 55011
                                                                         Therefore, its RF emissions
                 RF emissions (Conducted)        Group 1                 are very low and are not likely
                                                 Class B                 to cause any interference in
                 •   EN 55011                                            nearby electronic equipment.

                                                                         Class B equipment is
                                                                         suitable for use in all
                                                                         establishments, including
                                                                         domestic establishments and
                                                                         those directly connected to
                                                                         the public low-voltage power
                                                                         supply network that supplies
                                                                         buildings used for domestic
                                                                         purposes.

                 Harmonic emissions              Class A

                 IEC 61000-3-2

                 Voltage fluctuations/Flicker    Complies
                 emissions

                 IEC 61000-3-3



Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity
                The system described in this document is intended for use in the following specified
                electromagnetic environment. It is the responsibility of the customer or user to ensure
                that this system is used in such an environment.




24                                MAC™ 7 Resting ECG Analysis System                          2109599-003 A


                                                                                                  Electromagnetic Compatibility (EMC)




                Table 9: EMC Immunity Test

                Immunity Test             EN60601 Test Level          Compliance Level            Electromagnetic
                                                                                                  Environment–
                                                                                                  Guidance

                Electrostatic discharge   ± 8 kV contact              ± 8 kV contact              Floors should be wood,
                (ESD)                     ± 15 kV air                 ± 15 kV air                 concrete, or ceramic
                                                                                                  tile. If the floors are
                IEC 61000-4-2
                                                                                                  covered with synthetic
                                                                                                  material, the relative
                                                                                                  humidity should be at
                                                                                                  least 30%.

                Electrical Fast           ± 2 kV for power supply     ± 2 kV for power supply     Mains power should
                Transient/burst (EFT)     lines                       lines                       be that of a typical
                                          ± 1 kV for input/output     ± 1 kV for input/output     commercial or hospital
                IEC 61000-4-4
                                          lines                       lines                       environment.

                Surge                     ± 1 kV line(s) to line(s)   ± 1 kV line(s) to line(s)   Mains power should
                IEC 61000-4-5             ± 2 kV line(s) to earth     ± 2 kV line(s) to earth     be that of a typical
                                                                                                  commercial or hospital
                                                                                                  environment.

                Voltage dips, short       < 5% Ut (> 95% dip in       < 5% Ut (> 95% dip in       Mains power should
                interruptions, and        Ut) for 0.5 cycles          Ut) for 0.5 cycles          be that of a typical
                voltage variations on                                                             commercial or hospital
                                          < 40% Ut (> 60% dip in      < 40% Ut (> 60% dip in
                power supply input                                                                environment.
                                          Ut) for 5 cycles            Ut) for 5 cycles
                lines
                                                                                                  Ut is the AC mains
                                          < 70% Ut (> 30% dip in      < 70% Ut (> 30% dip in
                IEC 61000-4-11                                                                    voltage prior to
                                          Ut) for 25 cycles           Ut) for 25 cycles
                                                                                                  application of the test
                                          < 5% Ut (> 95% dip in       < 5% Ut (> 95% dip in       level.
                                          Ut) for 5 s                 Ut) for 5 s
                                                                                                  If the user requires
                                                                                                  continued system
                                                                                                  operation during
                                                                                                  power mains
                                                                                                  interruptions, it is
                                                                                                  recommended that
                                                                                                  the system is powered
                                                                                                  from an applicably
                                                                                                  rated uninterruptible
                                                                                                  power supply or a
                                                                                                  battery.

                Power frequency           30 A/m                      30 A/m                      Power frequency
                (50/60 Hz) magnetic                                                               magnetic fields
                field                                                                             should be at levels
                                                                                                  characteristic of a
                IEC 61000-4-8
                                                                                                  typical location in a
                                                                                                  typical commercial or
                                                                                                  hospital environment.



2109599-003 A                                  MAC™ 7 Resting ECG Analysis System                                           25


Electromagnetic Compatibility (EMC)




                 Immunity Test          EN60601 Test Level     Compliance Level       Electromagnetic
                                                                                      Environment–
                                                                                      Guidance

                 Conducted RF           3 Vrms                 3 Vrms                 Recommended
                                        0.15 MHz to 80 MHz     0.15 MHz to 80 MHz     separation distance:
                 IEC 61000-4-6
                                        6 Vrms in ISM bands    6 Vrms in ISM bands    d = 12√P
                                        between 0.15 MHz and   between 0.15 MHz and
                                        80 MHz                 80 MHz

                 Radiated RF            3 V/m at 80 to 2700    3 V/m at 80 to 2700    d = 1.2 √P 80 MHz to
                                        MHz, AM Modulation     MHz, AM Modulation     800 MHz
                 IEC 61000-4-3
                                        9 to 28 V/m at 385     9 to 28 V/m at 385     d = 2.3 √P 800 MHz to
                                        to 6000 MHz, FM or     to 6000 MHz, FM or     2.5 GHz
                                        Digital Modulation     Digital Modulation
                                                                                      At 80 MHz and 800
                                                                                      MHz, the higher
                                                                                      frequency range
                                                                                      applies.

                                                                                      Recommended
                                                                                      separation distance:

                                                                                      where P is the
                                                                                      maximum output
                                                                                      power rating of
                                                                                      the transmitter in
                                                                                      watts (W) according
                                                                                      to the transmitter
                                                                                      manufacturer, and d
                                                                                      is the recommended
                                                                                      separation distance in
                                                                                      meters (m).

                                                                                      Field strengths from
                                                                                      fixed RF transmitters,
                                                                                      as determined by an
                                                                                      electromagnetic site
                                                                                      survey, should be less
                                                                                      than the compliance
                                                                                      level in each frequency
                                                                                      range.

                                                                                      Interference may
                                                                                      occur in the vicinity
                                                                                      of equipment marked
                                                                                      with the following
                                                                                      symbol:




26                               MAC™ 7 Resting ECG Analysis System                               2109599-003 A


                                                                                                 Electromagnetic Compatibility (EMC)




                Immunity Test                 EN60601 Test Level       Compliance Level           Electromagnetic
                                                                                                  Environment–
                                                                                                  Guidance

                NOTE:

                    •       Do not use portable or mobile RF communications equipment closer to any part of the
                            system, including the cables, than the recommended separation distance calculated for the
                            equation applicable to the frequency of the transmitter.

                    •       Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
                            telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot,
                            theoretically, be predicted with accuracy. To assess the electromagnetic environment
                            due to fixed RF transmitters, consider conducting an electromagnetic site survey. If the
                            measured field strength in the location the system is used exceeds the applicable RF
                            compliance level listed in this table, observe the system to verify normal operation. If
                            abnormal performance is observed, additional measures may be necessary, such as re-
                            orienting or relocating the system.

                    •       At 80 MHz and 800 MHz, the higher frequency range applies.

                    •       These guidelines may not apply in all situations. Electromagnetic propagation is affected by
                            the reflection from structures, objects, and people.




Recommended Separation Distances
                The following table provides the recommended separation distances (in meters)
                between portable and mobile RF communication equipment and the system
                described in this manual, for equipment and systems that are not life-supporting.
                The system is intended for use in the electromagnetic environment in which radiated
                RF disturbances are controlled. The customer or the user of the system can help
                prevent electromagnetic interference by maintaining the following recommended
                minimum distance between portable and mobile RF communications equipment
                (transmitters) and the system, according to the maximum output power of the
                communications equipment.

                Table 10: Recommended Separation Distances

                                                                                                                  1
                Rated                    Separation Distance (meters) According to Frequency of Transmitter
                Maximum
                                                                                         2
                Output                150 kHz to 80 MHz            80 MHz to 800 MHz              800 MHz to 2.5 GHz
                Power of           d (meters) = 3.5/V1 x √ P                                        d = 7/E1 x √ P
                                                                      d = 3.5/E1 x √ P
                Transmitter       d (meters) for V1 = 3 Vrms                                       d for E1 = 3 V/m
                in Watts (W)                                          d for E1 = 3 V/m

                                     Meters          Feet          Meters          Feet          Meters          Feet

                    0.01              0.117          0.383          0.117          0.383         0.233          0.766

                    0.1               0.369          1.210          0.369          1.210         0.738          2.421

                        1             1.167          3.828          1.167          3.828         2.333          7.655


2109599-003 A                                     MAC™ 7 Resting ECG Analysis System                                       27


Electromagnetic Compatibility (EMC)




                                                                                                             1
                 Rated                  Separation Distance (meters) According to Frequency of Transmitter
                 Maximum
                                                                                      2
                 Output              150 kHz to 80 MHz           80 MHz to 800 MHz           800 MHz to 2.5 GHz
                 Power of         d (meters) = 3.5/V1 x √ P                                    d = 7/E1 x √ P
                                                                   d = 3.5/E1 x √ P
                 Transmitter     d (meters) for V1 = 3 Vrms                                    d for E1 = 3 V/m
                 in Watts (W)                                      d for E1 = 3 V/m

                                    Meters         Feet         Meters          Feet         Meters         Feet

                      10            3.689         12.104         3.689         12.104        7.379         24.208

                     100            11.667        38.276        11.667         38.276        23.333        76.552

                 NOTE:
                    1
                      These guidelines may not apply in all situations. Electromagnetic propagation is affected by
                    absorption and reflection from structures, objects, and people.
                     2
                         At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
                 For transmitters rated at a maximum output power not listed above, estimate the recommended
                 separation distance d in meters (m) using the equitation applicable to the frequency of the
                 transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
                 to the transmitter manufacturer.




28                                 MAC™ 7 Resting ECG Analysis System                                    2109599-003 A


                                                                                               Wireless Regulations




                                                                                        6
Wireless Regulations
                The wireless and wired LAN functionality of the MAC 7 is used to retrieve ECG orders
                and send ECG reports to an ECG Management System. In addition, the wireless
                and wired LAN functionality can be used to interface to other hospital information
                systems to provide additional data to the care giver operating the electrocardiograph.
                These tasks are an adjunct to the device’s intended use of acquiring, analyzing,
                displaying and printing an electrocardiogram. Because the wireless and wired
                LAN functionality is not required for the device to fulfill its intended use, network
                performance is not critical to the performance of the device. Furthermore, the MAC 7
                does not transmit any real-time data or alarm information over the network. Network
                Quality of Service (QoS) parameters such as reliability of data transmission, latency,
                transfer rate, error rate, and priority levels are not critical to the MAC 7 functionality
                and are not specified.


FCC Compliance
                This device complies with part 15 of the FCC Rules. Operation is subject to the
                following two conditions: (1) This device may not cause harmful interference, and (2)
                this device must accept any interference received, including interference that may
                cause undesired operation.
                CAUTION:
                  Changes or modifications to this unit not expressly approved by the party
                  responsible for compliance could void the user's authority to operate the
                  equipment.

                NOTE:
                  This equipment has been tested and found to comply with the limits for a Class
                  B digital device, pursuant to part 15 of the FCC Rules. These limits are designed
                  to provide reasonable protection against harmful interference in a residential
                  installation. This equipment generates, uses and can radiate radio frequency
                  energy and, if not installed and used in accordance with the instructions, may
                  cause harmful interference to radio communications. However, there is no
                  guarantee that interference will not occur in a particular installation. If this
                  equipment does cause harmful interference to radio or television reception, which


2109599-003 A                             MAC™ 7 Resting ECG Analysis System                            29


Wireless Regulations




                       can be determined by turning the equipment off and on, the user is encouraged
                       to try to correct the interference by one or more of the following measures:
                       • Reorient or relocate the receiving antenna.
                       • Increase the separation between the equipment and receiver.
                       • Connect the equipment into an outlet on a circuit different from that to which
                         the receiver is connected.
                       • Consult the dealer or an experienced radio/TV technician for help.
                Limited by local law regulations, version for North America does not have region
                selection option.
                To satisfy FCC RF exposure requirements, a separation distance of 20 cm or more
                should be maintained between the antenna of this device and persons during device
                operation.
                To ensure compliance, operations at closer than this distance is not recommended.


IC Compliance
                This device contains licence-exempt transmitter(s)/receiver(s) that comply with
                Innovation, Science and Economic Development Canada’s licence-exempt RSS(s).
                Operation is subject to the following two conditions:
                1.      This device may not cause interference.
                2.      This device must accept any interference, including interference that may cause
                        undesired operation of the device.
                L’émetteur/récepteur exempt de licence contenu dans le présent appareil est
                conforme aux CNR d’Innovation, Sciences et Développement économique Canada
                applicables aux appareils radio exempts de licence. L’exploitation est autorisée aux
                deux conditions suivantes:
                1.      L’appareil ne doit pas produire de brouillage;
                2.      L’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage
                        est susceptible d’en compromettre le fonctionnement.
                This equipment complies with radio frequency exposure limits set forth by the
                Innovation, Science and Economic Development Canada for an uncontrolled
                environment.
                This equipment should be installed and operated with a minimum distance of 20 cm
                between the device and the user or bystanders.
                This device must not be co-located or operating in conjunction with any other
                antenna or transmitter.
                Cet équipement est conforme aux limites d'exposition aux radiofréquences définies par
                la Innovation, Sciences et Développement économique Canada pour un environnement
                non contrôlé.


30                                MAC™ 7 Resting ECG Analysis System                            2109599-003 A


                                                                                               Wireless Regulations




                Cet équipement doit être installé et utilisé avec un minimum de 20 cm de distance
                entre le dispositif et l'utilisateur ou des tiers.
                Ce dispositif ne doit pas être utilisé à proximité d’une autre antenne ou d’un autre
                émetteur.
                The device for operation in the band 5150-5250 MHz is only for indoor use to reduce
                the potential for harmful interference to co-channel mobile satellite systems;
                les dispositifs fonctionnant dans la bande de 5 150 à 5 250 MHz sont réservés
                uniquement pour une utilisation à l'intérieur afin de réduire les risques de brouillage
                préjudiciable aux systèmes de satellites mobiles utilisant les mêmes canaux;


RED Information
                The MAC 7 embedded wireless module complies with CE RED 2014/53/EU.




                This product is restricted to indoor use.

                Frequency Range         2.4 GHz frequency bands: 2.4-2.483 GHz
                                        5 GHz frequency bands: 5.15-5.35 GHz, 5.47-5.725 GHz

                Maximum RF output       20 dBm
                power




2109599-003 A                             MAC™ 7 Resting ECG Analysis System                              31


GE Medical Systems
Information Technologies, Inc.
8200 West Tower Avenue
Milwaukee, WI 53223 USA
Tel: +1 414 355 5000
Tel: +1 800 558 7044 (US Only)




GE Medical Systems Information Technologies, Inc., a General Electric Company, going to market as GE Healthcare.

www.gehealthcare.com