LCS190923018AEB_SAR_REPORT_2

FCC ID: O55503719

RF Exposure Info

Download: PDF
FCCID_4507272

  COMOSAR E—Field Probe Calibration Report



                                              Ref : ACR.281.2.18.SATU.A




 SHENZHEN LCS COMPLIANCE TESTING
        LA BORATORY LTD.
 1F., XINGYUAN INDUSTRIAL PARK, TONGDA ROAD,
                  BAO‘AN BLVD
BAO‘AN DISTRICT, SHENZHEN, GUANGDONG, CHINA
MVG COMOSAR DOSIMETRIC E—FIELD PROBE
        SERIAL NO.: SN 31/17 EPGO324
                            Calibrated at MVG US
                    2105 Barrett Park Dr. — Kennesaw, GA 30144




                          $ ng
                           14@\\?            ACCREDITED
                              Toh uks         romenmerengrerznss
                         Calibration Date: 10/08/2018

  ie
  ‘ ;SIUHIHHI}'.‘


  | This document presents the method and results from an accredited COMOSAR Dosimetric E—Field
    Probe calibration performed in MVG USA using the CALISAR / CALIBAIR test bench, for use
    with a COMOSAR system only.         All calibration results are traceable to national metrology
  ‘ institutions


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                  FCC ID: 055503719                               Report No.: LCS190923018AEB




           mvgq                     COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR.281.2.18.SATU.A




                                              Name                              Function                          Date                  Signature
           Prepared by :                 Jérome LUC                       Product Manager                    10/8/2018                      fi

           Checked by :                  Jérome LUC                        Product Manager                   10/8/2018                      fi/

           Approved by :           Kim RUTKOWSKI                           Quality Manager                   10/8/2018                fum   PutthowiPd




                                                                                   Customer Name
                                                                                  Shenzhen LCS
                                                  Distribution :                Compliance Testing
                                                                                   Laboratory Ltd.




               Issue                   Date                                                        Modifications
                 A                  10/8/2018                 Initial release




                                                                         Page: 2/10

                            This document shall not be reproduced, except infull orin part, without the written approval ofMVG.
                            The information contained herein is to be used onlyfor the puposefor whichit is submitted and is not to
                            be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 74 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                     FCC ID: 055503719                               Report No.: LCS190923018AEB




           nc                          COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR.281.2.18.SATU.A




                                                               TABLE OF CONTENTS




           1      DeVi@@ UNG@P T@St ... 0000000 meemeeee n en en en neneeee en en en nee en en ennerne en ererne en 4
           2      PFOGUCt D@SCMIDMOM ..... ... ... .ccscc meenenemenmnenenemeemn en en en en erne en eren en eneenee nc es 4

                2.1      General Information                                                                                                           4
           3      MEASU@ME@Nt MtROG ............ .l iesecee mm mmmmmeemememmeerenmemeemenenneenmnerneenneene en en eneeneeees 4

                3.1      Linearity                                                                                                                     4
                3.2      Sensitivity                                                                                                                   5
                3.3      Lower Detection Limit                                                                                                         5

                3.4      Isotropy                                                                                                                      5
                3.3      Boundary Effect                                                                                                               3

           t      MCASUM@MANUUINCENAIALY :——mmmmmmmmmmmmrnemcrmecommenmmescmonmmamenmemmnim 3
           $      Calibration Messurement ReSUIS : ——mmmummmmemmemmmmmmmmmnnemenmmoenmeanuns 6
                5.1      Sensitivity in air                                                                                                            6
                5.2      Linearity                                                                                                                     7
                5.3      Sensitivity in liquid                                                                                                         7
                5.4      Isotropy                                                                                                                      8

           6.     Tistrof EQUIPMENE ——mmmmummmenmmenpemonnanmmnmmmenmonmememmemenemenren 10




                                                                            Page: 3/10

                               This document shall not be reproduced, except infull orin part, without the written approval ofMVG.
                               The information contained herein is to be used onlyfor the puposefor whichit is submitted and is not to
                               be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 75 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                   FCC ID: 055503719                               Report No.: LCS190923018AEB




                                     COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                   Ref: ACR.281.2.18.SATU.A




           1              DEVICE UNDER TEST

                                                                 Device Under Test
                 Device Type                                                    COMOSAR DOSIMETRIC E FIELD PROBE
                 Manufacturer                                                   MVG
                 Model                                                          SSE2
                 Serial Number                                                  SN 31/17 EPGO324
                 Product Condition (new / used)                                 New
                 Frequency Range of Probe                                       0.15 GHz—6GHz
                 Resistance of Three Dipoles at Connector                       Dipole 1: R1=0.189 MQ
                                                                                Dipole 2: R2=0.203 MQ
                                                                                Dipole 3: R3=0.218 MQ

           A yearly calibration interval is recommended.

           2              PRODUCT DESCRIPTION

           2.1          GENERAL INFORMATION

           MVG‘s COMOSAR E field Probes are built in accordance to the IEEE 1528, OET 65 Bulletin C and
           CEIVIEC 62209 standards.




                                        Figure 1 —MVG COMOSAR Dosimetric Efield Dipole

                 Probe Length                                                   330 mm
                 Length of Individual Dipoles                                   2 mm
                 Maximum extemal diameter                                       8 mm
                 Probe Tip Extemal Diameter                                     2.5 mm
                 Distance between dipoles / probe extremity                     1 mm


                    3     MEASUREMENT METHOD
           The IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEIIEC 62209 standards provide
           recommended practices for the probe calibrations, including the performance characteristics of
           interest and methods by which to assess their affect. All calibrations / measurements performed meet
           the fore mentioned standards.

                    3.1   LINEARITY
           The evaluation of the linearity was done in free space using the waveguide, performing a power
           sweep to cover the SAR range 0.01 W/kg to 100W/kg.


                                                                          Page: 4/10

                            This documentshall not be reproduced, except in full or in part, without the written approval ofMVG.
                            The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                            be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenzhen LCS Compliance Testing Laboratory Ltd.
                                                                Page 76 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                     FCC ID: 055503719                                 Report No.: LCS190923018AEB




                                       COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                   Ref: ACR.281.2.18.SATU.A




                   3.2        SENSITIVITY
           The sensitivity factors of the three dipoles were determined using a two step calibration method (air
           and tissue simulating liquid) using waveguides as outlined in the standards.

                   3.3    LOWER DETECTION LIMIT
           The lower detection limit was assessed using the same measurement set up as used for the linearity
           measurement. The required lower detection limit is 10 mW/kg.

                   3.4        ISOTROPY
           The axial isotropy was evaluated by exposing the probe to a reference wave from a standard dipole
           with the dipole mounted under the flat phantom in the test configuration suggested for system
           validations and checks. The probe was rotated along its main axis from 0 — 360 degrees in 15 degree
           steps. The hemispherical isotropy is determined by inserting the probe in a thin plastic box filled
           with tissue—equivalent liquid, with the plastic box illuminated with the fields from a half wave dipole.
           The dipole is rotated about its axis (0°—180°) in 15° increments. At each step the probe is rotated
           about its axis (0%—360°).

                   3.5        BOUNDARY EFFECT
           The boundary effect is defined as the deviation between the SAR measured data and the expected
           exponential decay in the liquid when the probe is oriented normal to the interface. To evaluate this
           effect, the liquid filled flat phantom is exposed to fields from either a reference dipole or waveguide.
           With the probe normal to the phantom surface, the peak spatial average SAR is measured and
           compared to the analytical value at the surface.

           4             MEASUREMENT UNCERTAINTY

           The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEMIEC
           62209 standards were followed to generate the measurement uncertainty associated with an E—field
           probe calibration using the waveguide technique. All uncertainties listed below represent an
           expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor
           of k=2, traceable to the Internationally Accepted Guides to Measurement Uncertainty.
            Uncertainty analysis of the probe calibration in waveguide

                                                            Uncertainty            Probability               hor                  4          Standard
                    ERREOR SOURCES                           value (%)             Distribution             sisor                C        Uncertainty (%)

            Incident or forward power                           3.00%               Rectangular                 \/?:    |         1            1.732%

            Reflected power                                     3.00%               Rectangular                 \/g     |         3            1.732%

            Liquid conductivity                                 5.00%               Rectangular                 \/?:    |         1            2.887%

            Liquid permittivity                                 4.00%               Rectangular                 \/g     |         l            2.309%

            Field homogeneity                                   3.00%               Rectangular                 \/5     ‘         3            1.732%

            Field probe positioning                             5.00%               Rectangular                 \6      |         1            2.887%



                                                                            Page: 5/10

                              This documentshall not be reproduced, except in full or in part, without the written approval ofMVG.
                              The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                              be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenzhen LCS Compliance Testing Laboratory Ltd.
                                                                Page 77 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                    FCC ID: 055503719                               Report No.: LCS190923018AEB




           mvgq                      COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR.281.2.18.SATU.A



            Field probe linearity                             3.00%                Rectangular               \6       |        1            1.732%

              Combined standard uncertainty                                                                                                 5.831%

                   Expanded uncertainty                                                                                                      12.0%
                 95 %confidence level k= 2                                                                                                    C *



                   5     CALIBRATION MEASUREMENT RESULTS


                                                              Calibration Parameters
               Liquid Temperature                                                 21°C
               Lab Temperature                                                    21°C
               Lab Humidity                                                       45 %


                   5.1     SENSITIVITY IN ARR



                                               Normx dipole            Normy dipole            Normz dipole
                                                1 (uV/(V/m)) 2 (uV/(V/m)) 3 (uV/(V/m))
                                                       0.80                   0.83                   0.68


                                                DCP dipole 1           DCP dipole 2            DCP dipole 3
                                                       (mV)                  (mV)                    (mV)
                                                        95                    90                      93


           Calibration curves ei=f(V) (i=1.2,3) allow to obtain H—field value using the formula:
                                                                 E=fE/ +E," +E,"


                                                       Calibration curves
                                    63—                                                            ces

                                    so0—                                               z_
                                € 400
                                §                                //                                                 Dipole 1
                                72                                                                                  Dipole 2
                                © ong                &                                                              Dipcle 3

                                    100     ,/
                                      .
                                      ooo        oor       omm       o0           on     010
                                                                                                         ;
                                                                                                       013
                                                                    Voltage (V)




                                                                          Page: 6/10

                             This document shall not be reproduced, except infull orin part, without the written approval ofMVG.
                             The information contained herein is to be used onlyfor the puposefor whichit is submitted and is not to
                             be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 78 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                                          FCC ID: 055503719                               Report No.: LCS190923018AEB




           mvG                                              COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR.281.2.18.SATU.A




                             5.2                   LINEARITY

                                                                        Linearity
                                          1.00—
                                          0.76
                                          0.50
                   Linearity Enior (dB)




                                          0.25
                                          0.00—
                                          A.25—
                                          .50
                                          0.76
                                          4.007,
                                               0                 00    qs0      ato       250   ado       sto
                                                                         E—Field (V/m)

                  LinearityI+/1 13% (++0.050B


                             5.3                   SENSITIVITY IN LQUID

               Liquid                                 Frequency              Permittivity        Epsilon (S/m)             ConvF
                                                        MHz +/—
                                                       100MHz)
               HL450                                     450                     42.17                0.86                   1.56
               BL450                                       450                   57.65                0.95                   1.60
               HL750                                       750                   40.03                0.93                   1.45
              BL750                                       750                    56.83                1.00                   1.50
              HL850                                       835                    42.19                0.90                   1.55
              BL850                                       835                    54.67                1.01                   1.59
              HL900                                       900                    42.08                1.01                   1.54
              BL900                                       900                    55.25                1.08                   1.60
              HL1800                                      1800                   41.68                1.46                   1.65
              BL1800                                      1800                   53.86                1.46                   1.68
              HL1900                                      1900                   38.45                1.45                   1.86
              BL1900                                      1900                   §1322                1.56                   1.93
              HL2000                                      2000                   38.26                 1.38                  1.83
              BL2000                                      2000                   52.170                1.51                  1.89
              HL2300                                      2300                   39.44                 1.62                 1.95
              BL2300                                      2300                   54.52                 117                  2.01
              HL2450                                      2450                   37.50                 1.80                 1.91
              BL2450                                      2450                   5322                 1.89                   1.95
              HL2600                                      2600                   39.80                1.99                   1.89
              BL2600                                      2600                   5252                 2.23                   1.94
              HL5200                                      5200                   35.64                4.67                   1.50
              BL5200                                      5200                   48.64                5.51                   1.56
              HL5400                                      5400                   36.44                4.87                   1.44
              BL5400                                      5400                   46.52                5.71                   1.47
              HL 5600                                     5600                   36.66                5.17                   1.48
              BL5600                                      5600                   46.179               5.77                   1.53
              HL 5800                                     5800                   35.31                5.31                   1.50
              BL5800                                      5800                   47.04                6.10                   1.55

           LOWER DETECTION LIMIT: ImW/kg
                                                                                                  Page: 7/10

                                                    This document shall not be reproduced, except infull orin part, without the written approval ofMVG.
                                                    The information contained herein is to be used onlyfor the puposefor whichit is submitted and is not to
                                                    be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 79 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                      FCC ID: 055503719                           Report No.: LCS190923018AEB




           mvG                      COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                  Ref: ACR.281.2.18.SATU.A




                   5.4      ISOTROPY
           HL900 MHz
           — Axial isotropy:                                  0.05 dB
           — Hemispherical isotropy:                          0.07 dB


                                                              isotropy curves



                                                06                                                 t

                                                o«                                                «=
                                                02

                                               0

                                                02

                                               04

                                                06

                                                08

                                               104
                                                     fo a8 as o+! g2 co o2 or os 08 10i



           HL1800 MHz
           — Axial isotropy:                                  0.06 dB
           — Hemispherical isotropy:                          0.07 dB


                                                              isotropy curves

                                                                                             Dirole a0
                                                                                             Dicole 30
                                                                                             Deole at




                                                10
                                                     to as os o« o2 oo oz o+ os os to




                                                                                Page: 8/10

                            This document shall not be reproduced, except infull orin part, without the written approval ofMVG.
                            The information contained herein is to be used onlyfor the puposefor whichit is submitted and is not to
                            be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 80 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                    FCC ID: 055503719                            Report No.: LCS190923018AEB




           I nVvG                   COMOSAR E—FIELD PROBE CALIBRATION REPORT                                                 Ref: ACR.281.2.18.SATU.A




           HLS600 MHz
           — Axial isotropy:                               0.06 dB
           — Hemispherical isotropy:                       0.10 dB


                                                            Isotropy curves




                                               404                              V
                                                 fo as os o1 92 c0 o2 or os os 10




                                                                              Page: 9/10

                            This document shall not be reproduced, except infull orin part, without the written approval ofMVG.
                            The information contained hereinis to be used onlyfor the puposefor whichit is submitted and is not to
                            be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 81 of 144


SHENZHEN LCS COMPLIANCE TESTING LABORATORY LTD.                                    FCC ID: 055503719                                Report No.: LCS190923018AEB




                                                                                                                                 Ref: ACR.281.2.18.SATU.A
           LDAC                        COMOSAR E—FIELD PROBE CALIBRATION REPORT




                   6    LIST OF EQUIPMENT



                                                            Equipment Summary Sheet

               Equipment                 Manufacturer /              Identification No.                   Current                  Next Calibration
               Description                   Model                                              ‘|    Calibration Date                   Date

               Flat Phantom                       mve                  SW306g.gamy, NValldaled
                                                                                    required.
                                                                                               Noewal                            mlldalec
                                                                                                                                 required.
                                                                                                                                                 No: cal
                                                   .                                                 Validated. No cal           Validated.      No    cal
          COMOSAR Test Bench                   Version 3                         NA                  required.                   required.

             Network Analyzer            Rhodeg‘vsAChwa'z                   SN100132                        o22o16                       022019
              Reference Probe                     MVG                   EP 94 SN 37/08                      10/2017                      10/2019

                 Multimeter                 Keithley 2000                    1188656                        01/2017                      01/2020

             Signal Generator              Agilent E4438C                 MY49070581                        01/2017                      01/2020

                       ie                                                                            Characterized prior to \Characterized prior to
                  Ampilfier                  Aethersomins                     $NO4s                  test. No cal required. |test. No cal required.

                Power Meter                   HP E4418A                   US38261498                        01/2017                      01/2020

               Power Sensor                 HP ECP—E26A                   US37181460                        01/2017                      01/2020

            Directional Coupler            Narda 4216—20                       o1386                 Characterized pripr to |Characterized priprto
                                                                                                     test. No cal required. |test. No cal required.
                Waveguide                  Mega Industries            Osgy7—158.13.742 required.
                                                                                       Validated. No cal                         Validated.
                                                                                                                                 required.
                                                                                                                                            No cal
           Waveguide Transition            Mega Industries            Osgy7—158.13.701 required.
                                                                                       Validated. No cal                         Validated.
                                                                                                                                 required.
                                                                                                                                            No cal
          Waveguide Termination|            Mega Industries           Osgy7—158—13.701 Validated.
                                                                                       required.
                                                                                                  No cal                         Validated.
                                                                                                                                 required.
                                                                                                                                            No cal
          Temperature / HUMIIIRY|— control Company
                  ensor
                                                                           150798832                        112017                       1112020




                                                                           Page: 10/10

                              This document shall not be reproduced, except infull or in part, without the written approval ofMVG.
                              The information contained herein is to be used onlyfor the puposefor which it is submitted and is not to
                              be released in whole orpart without written approval ofMVG.




   This report shall not be reproduced except in full, without the written approval of Shenthen LCS Compliance Testing Laboratory Ltd.
                                                                Page 82 of 144


                                         (/(.\
                                  SATIMO
                                  The microwave vision company_A




     SAR Reference Dipole Calibration Report



                                        Ref : ACR.287.3.14.SATU.A




 SHENZHEN LCS COMPLIANCE TESTING
        LABORATORY LTD.
1F., XINGYUAN INDUSTRIAL PARK, TONGDA ROAD,
                 BAO‘AN BLVD
BAO‘AN DISTRICT, SHENZHEN, GUANGDONG, CHINA
  SATIMO COMOSAR REFERENCE DIPOLE
                     FREQUENCY: 750 MHZ
                     SERIAL NO.: SN 07/14 DIP 0G750—302


                          Calibrated at SATIMO US
                 2105 Barrett Park Dr. — Kennesaw, GA 30144
                            w¥ 0pe



                          e CR
                           S\ /z


                          Tz
                           ZrAanS                 (accrebitep
                             CUAaj Wwluk®®          cafttrationcent szz0e02

                                       10/01/2018

   Summary:

   This document presents the method and results from an accredited SAR reference dipole calibration
   performed in SATIMO USA using the COMOSAR test bench. All calibration results are traceable
   to national metrologyinstitutions.


                                                                                                                   Ref: ACR.287.3.14.SATU.A
sArImo                   SAR REFERENCE DIPOLE CALIBRATION REPORT




                                  Name                              Function                          Date                  Signature
Prepared by:                 Jérome LUC                       Product Manager                    10/14/2018                   2/5’

Checked by :                 Jérome LUC                        Product Manager                   10/14/2018                   //%/

Approved by :          Kim RUTKOWSKI                           Quality Manager                   10/14/2018                Jum Pautthowh?




                                                                       Customer Name
                                                                      Shenzhen LCS
                                      Distribution :                Compliance Testing
                                                                      Laboratory Ltd.




   Issue                   Date                                                         Modifications
     A                 10/14/2018                 Initial release




                                                             Page: 2/11

                This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                The information contained herein is to be used onlyforthe purposefor which it is submitted and is not to
                be released in whole orpart without written approval ofSATIMO.




                                                           = mes csw


     (( [      \               SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   Ref: ACR287 3.14.SATU.A
sArImo

                                                       TABLE OF CONTENTS




 )      IMfOducHO®:—mmmeormmmnmenememrmmmmnareroemenrarerermmarnmenmmemymmonmmernmnm 4

B       DCVIC@ URCCF TESE ... ... s cmemimecmmmmmmmnnmncrmensecirmmmecmnnemmntrenmrermmrocremnmmmsnteernnccnczcne 4
3       PFOUUGCE DeSCFIPLION ——.. vemmememmmmommenmmmenmmnmiereemmenenenmmmmeminmmmnrnememenisiimmmercer 4
      3.1       General Information                                                                                                           4

#:          MeASUROMeNENCHAIOU —:——+—msmemscommmrenecncerormemiermsnctisesmemseretemecmmetimmmemisrime $
      4.1       Return Loss Requirements                                                                                                      5

      4.2       Mechanical Requirements                                                                                                       5
5       MEASUT@ME@Nt UNCETEANEY ... ... en eeemmeemeememeeeeeee n n en en en ernernenneernernerne en en enene. 5

      5.1       Return Loss                                                                                                                   $

      5.2       Dimension Measurement                                                                                                         §
      5.3       Validation Measurement                                                                                                        5

6       CalibrAtiON MEASUT@MENE RESUIS .............2.nene e en en reeneeneemneeneen e 6
      6.1       Return Loss and Impedance                                                                                                     6

      6.2       Mechanical Dimensions                                                                                                         6
7       Validation measurement ...

      7.1       Head Liquid Measurement                                                                                                       7

      7.2       SAR Measurement Result With Head Liquid                                                                                       7

      7.3       Body Liquid Measurement                                                                                                       9
      7.4       SAR Measurement Result With Body Liquid                                                                                       9
        LAE OT EqUHDMIGME : 000000 rrmomevormmsecrroreeoeroremeerememrmeeaecararanmmmyrerares §1
50




                                                                   Page: 3/11

                      This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                      The information contained herein is to be used onlyforthe purposefor which it is submitted and is not to
                      be released in whole orpart without written approval ofSATIMO.




                                                                 t ces csw —


    (((   \               SAR REFERENCE DIPOLE CALIBRATION REPORT                                                   KiMexesioAugatodt



1             INTRODUCTION


This document contains a summary of the requirements set forth by the IEEE 1528, OET 65 Bulletin

C and CEVIEC 62209 standards for reference dipoles used for SAR measurement system validations
and the measurements that were performed to verify that the product complies with the fore
mentioned standards.




2             DEVICE UNDER TEST




                                                     Device Under Test
          Device Type                                       COMOSAR 750 MHz REFERENCE DIPOLE
          Manufacturer                                      Satimo
          Model                                             $ID750
          Serial Number                                     SN 07/14 DIP 0G750—302
          Product Condition (new / used)                    New

A yearly calibration interval is recommended.

3             PRODUCT DESCRIPTION

3.1       GENERAL INFORMATION

Satimo‘s COMOSAR Validation Dipoles are built in accordance to the IEEE 1528, OET 65 Bulletin
C and CEIIEC 62209 standards. The product is designed for use with the COMOSAR test bench
only.




                                Figure 1 — Satimo COMOSAR Validation Dipole




                                                              Page: 4/11

                This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO.
                The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                be released in whole orpart without written approval ofSATIMO.


 ((( . \                  SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR.287.3.14.SATU.A
SATIMO

       4     MEASUREMENT METHOD
The IEEE 1528, OET 65 Bulletin C and CELIIEC 62209 standards provide requirements for
reference dipoles used for system validation measurements. The following measurements were
performed to verify that the product complies with the fore mentioned standards.

       4.1   RETURN LOSS REQUIREMENTS
The dipole used for SAR system validation measurements and checks must have a return loss of —20
dB or better. The return loss measurement shall be performed against a liquid filled flat phantom,
with the phantom constucted as outlined in the fore mentioned standards.

       4.2   MECHANICAL REQUIREMENTS
The IEEE Std. 1528 and CEIM/IEC 62209 standards specify the mechanical components and
dimensions of the validation dipoles, with the dimensions frequency and phantom shell thickness
dependent. The COMOSAR test bench employs a 2 mm phantom shell thickness therefore the
dipoles sold for use with the COMOSAR test bench comply with the requirements set forth for a 2
mm phantom shell thickness.


5            MEASUREMENT UNCERTAINTY

All uncertainties listed below represent an expanded uncertainty expressed at approximately the 95%
confidence level using a coverage factor of k=2, traceable to the Internationally Accepted Guides to
Measurement Uncertainty.

       5.1   RETURNLOSS
The following uncertainties apply to the return loss measurement:
                      Frequency band                                    Expanded Uncertainty on Return Loss
                        400—6000MHz                                                               0.1 dB

       5.2   DIMENSION MEASUREMENT
The following uncertainties apply to the dimension measurements:
                         Length (mm)                                        Expanded Uncertainty on Length

                              3 — 300                                                            0.05 mm



       5.3   VALIDATION MEASUREMENT
The guidelines outlined in the IEEE 1528, OET 65 Bulletin C, CENELEC EN50361 and CEIIEC
62209 standards were followed to generate the measurement uncertainty for validation
measurements.
                         Scan Volume                                       Expanded Uncertainty
                                  1 g                                                   20.30%
                                 10 g                                                   20.1 %

                                                              Page: 5/11
                This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO.
                The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                be released in whole orpart without written approval ofSATIMO.


 (((                         SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR287 3.14.SATU.A
sArImo
     /


         6     CALIBRATION MEASUREMENT RESULTS

         6.1     RETURN LOSS AND IMPEDANCE
                                                                  Frequency, MHz
                              650660       680      700     720     740      760        780    800      820   840850
                             0—


                             ,5_




                        C                                     1 __ __
                        o
                        —> 20—
                        &




                            40—

  Frequency (MHz)                         Return Loss (dB)                      Requirement (dB)                          Impedance
        750                                    —34.80                                 —20                               50.7 Q + 1.6 jG2

         6.2       MECHANICAL DIMENSIONS

 Frequency MHz                            L mm                                          h mm                                   d mm

                             required             measured                 required            measured            required           measured

        300                 420.0 £1 %.                                   250.0 £1 %.                             6.35 +1 %.

        450                 290.0 £1 %.                                   166.7 £1 %.                             6.35 £1 %.
        750                 176.0 +1 %.              PASS                 100.0 £1 %.                PASS         6.35 £1 %.             PASS
        835                 161.0 +1 %.                                   89.8 +1 %.                               3.6 +1 %.
        900                 149.0 £1 %.                                   83.3 +1%.                                3.6 +1 %.

        1450                89.1 +1 %.                                    51.7 £1 %.                               3.6 +1 %.
        1500                80.5 £1 %.                                    50.0 £1 %.                               3.6 +1 %.

        1640                79.0 1 %.                                     45.7 £1%.                                3.6 £1 %.
        1750                75.2 £1%.                                     42.9 £1%.                                3.6 +1 %.

        1800                72.0 £1 %.                                    41.7 £1%.                                3.6 £1 %.

        1900                68.0 +1 %.                                    39.5 £1%.                                3.6 £1 %.
        1950                66.3 +1 %.                                    38.5 £1%.                                3.6 £1 %.

        2000                64.5 41 %.                                    37.5 £1%.                                3.6 £1 %.
        2100                61.0 +1 %.                                    35.7 £1%.                                3.6 +1 %.

        2300                55.5 41 %.                                    32.6 +1 %.                               3.6 £1 %.

        2450                51.5 #1%.                                     30.4 £1%.                                3.6 +1 %.
        2600                48.5 +1 %.                                    28.8 £1%.                                3.6 +1 %.

        3000                41.5 £1%.                                     25.0 £1 %.                               3.6 £1 %.

        3500                 37.0%1%.                                     26.4 £1%.                                3.6 41 %.

        3700                 34.7%1%.                                     26.4 £1%.                                3.6 +1 %.


                                                                  Page: 6/11

                   This document shall not be reproduced, except in full orin part, without the written approval ofSATIMO.
                   The information contained hereinis to be used onlyforthepurposefor whichit is submitted and is not to
                   be released in whole orpart without writtenapproval ofSATIMO.




  omm     e es hopeecennnn en nemvapc en geeesg en nerrernen nne m e nen ues mvir en en nee uy eermernarners ce ie eoemapeemnrnen n n emereg eee c reee e y se


 ((( . \                    SAR REFERENCE DIPOLE CALIBRATION REPORT                                                  Ref: ACR.287.3.14.SATU.A
SATIMO

           7     VALIDATION MEASUREMENT
The IEEE Std. 1528, OET 65 Bulletin C and CEIIEC 62209 standards state that the system
validation measurements must be performed using a reference dipole meeting the fore mentioned
return loss and mechanical dimension requirements. The validation measurement must be performed
against a liquid filled flat phantom, with the phantom constructed as outlined in the fore mentioned
standards. Per the standards, the dipole shall be positioned below the bottom of the phantom, with
the dipole length centered and parallel to the longest dimension of the flat phantom, with the top
surface of the dipole at the described distance from the bottom surface of the phantom.

           7.1   HEAD LIQUID MEASUREMENT

                           Fre’?/lu:zncy          Relative permittivity (€,‘)             Conductivity {0) S/m

                                                  required          measured            required          measured

                               300               45.3 15 %                             0.87 £5 %
                               450               43.5 25 %                              0.87 £5 %

                               750               41.9 25 %             PASS             0.89 £5 %            PASS
                               835               41.5 25 %                              0.90 £5 %
                               900               41.5 25 %                              0.97 £5 %
                              1450               40.5 45 %                              1.20 25 %

                              1500               40.4 15 %                              123125 %
                              1640               40.2 25 %                              1.31 15 %
                              1750               40.1 25 %                              1.37 25 %
                              1800               40.0 25 %                              1.40 25 %
                              1900               40.0 25 %                              1.40 25 %
                              1950               40.0 25 %                              140 £5 %

                              2000               40.0 25 %                              140 25 %
                              2100               39.8 45 %                              149 £5 %
                              2300               39.5 25 %                              1.67 25 %
                              2450               39.2 25 %                              1.80 25 %
                              2600               39.0 25 %                              1.96 25 %

                              3000               38.5 15 %                              2.40 £5 %
                              3500               37.9 25 %                              2.91 +5 %


           7.2    SAR MEASUREMENT RESULT WITH HEAD LIQUID
The IEEE Std. 1528 and CEVIEC 62209 standards state that the system validation measurements
should produce the SAR values shown below (for phantom thickness of 2 mm), within the
uncertainty for the system validation. All SAR values are normalized to 1 W forward power. In
bracket, the measured SAR is given with the used input power.

Software                                                                 OPENSAR V4
Phantom                                                                  SN 20/09 SAM71
Probe                                                                    SN 18/11 BPG122
Liquid                                                                   Head Liquid Values: eps‘ : 42.1 sigma : 0.89
Distance between dipole center and liquid                                15.0 mm
Area scan resolution                                                     dx=8mm/dy=8mm

                                                                Page: 7/11

                  This document shall not be reproduced, except infull or in part, without the written approval ofSATIMO.
                  The information contained herein is to be used onlyfor the purposefor which it is submitted and is not to
                  be released in whole orpart without written approval ofSATIMO.




                                                             t nan c     y —


 ((/.\                            SAR REFERENCE DIPOLE CALIBRATION REPORT                                                        LRenestoAtgArodt
SATIMO

Zoon Scan Resolution                                                      dx=8mm/dy=8m/dz=5mm
Frequency                                                                 750 MHz
Input power                                                               20 dBm
Liquid Temperature                                                        71 °C
Lab Temperature                                                           21 °C
Lab Humidity                                                              45 %




                                  F
                                      'e&":z"“’         1 g SAR (W/ke/w)                    10 g SAR (W/kg/w)
                                                   required           measured            required                 measured

                                          300         2.85                                  1.94

                                          450         4.58                                  3.06

                                          750         8.49            8.38 (0.84)           5.55                   5.53 (0.55)
                                          835         9.56                                  6.22

                                          900         10.9                                  6.99
                                          1450         29                                       16

                                          1500        30.5                                  16.8

                                          1640        34.2                                  18.4

                                          1750        36.4                                  19.3

                                          1800        38.4                                  20.1

                                          1900        $9.7                                  20.5
                                          1950        40.5                                  20.9

                                       2000           41.1                                  21.1

                                       2100           43.6                                  21.9

                                       2300           48.7                                  23.3
                                       2450           52.4                                   24

                                          2600        55.3                                  24.6

                                          3000        63.8                                  25.7

                                          3500        67.1                                   25


                       oreeetaomn                                                                    7
                    Sutsen nesice meros           _termos_|                                     ’;



                                                                                                                           JR
                                                                                                1 l'\

                                                                                                10
     7
     Zunters                                                                                a.l§::           |NA
   rar         o»                                                                           & os                     [A
                ]                                                                           &                             in
                                                                                                as
   we}   gol)                                                                                   a2—,                                   W                  t
                     [ ximet B Yieet                                                                 8   2    4    608    10 12 WZIL"‘::I 18 20 22 M m 2 30




                                                                 Page: 8/11

                    This document shall not be reproduced, except in full or in part, without the written approval ofSATIMO.
                    The information contained herein is to be used onlyforthe purposefor which it is submitted and is not to
                    be released in whole orpart without written approval ofSATIMO.




                                                               t cesc c —



Document Created: 2019-11-04 18:19:11
Document Modified: 2019-11-04 18:19:11

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC