SAR Test Report 2

FCC ID: NOI-W410

RF Exposure Info

Download: PDF
FCCID_644614

  A
   #

ADT CORM


APPENDIX C: PHOTOGRAPHS OF SYSTEM VALIDATION




                            C-1


APPENDIX D: SYSTEM CERTIFICATE & CALIBRATION

D1: SAM PHANTOM


.    Schmid & Partner
    '-Englneermg AG 8 0                                                     _
       Zoughnuutrnu 43, 8004 Zurlch. 5wltz¢rland, Phone +41 1 245 97 00. Fax +41 9 245 97 79 '
                                                                                                          Can)




       Certificate of conformity _/Flrst'A_rtlclé Inspection _                                                                              2k
        Item                ‘                .     SAM Twin Phantom V40
        Type No _:                                 QD 000 P40 CA .
        Series No                                  TP—1150 and higher _:                        2
        Manufacturer / Origin                    «] Untersee Composites
             Ds                          +         Hauptstr. 69         .
                                             CH—8559 Fruthwilen
                                       ~__.| Switzerland .___
       Tests

       The series production processused allows the hmntaflon to test of first articles.
      . Complele tests were made on the pre—series Type No. QD 000 P40 AA, Serial No. TP—1001 and on the .
        series first article Type No. QD 000 P40 BA, Serial No. TP-1 006. Certain parameters have beeri retested
       using further series umts (called samp(es)                                   .                                                               ©

        Test                     —|   Requirement                 202                   Details           —                    Units tested
        Shape                         Compliance with the geometry              >       ITIS CAD File (*)                   ~.| First article, —
                                      according to the CAD model.                                     .                        Samples
        Material thickness            Compliant with the requirements                   2mm +/— 0.2mm in                       First article,
                            _         according to the standards                        specific areas                          Samples .
        Material                . . | Dielectricparameters for required                 200 MHz — 3 GHz                       : Material
        parameters                    frequencnes                       .               Relative permittivity <5                sample® _
                                                          _                             Loss tangent—< 0.05.                   TP 104—5
        Material resistivity          The material has been tested to be                 Liquid type HSL 1800 .                Pre—series,
                       «              compatible with theliquids definedin               fand others according to .             First article
                                      the standards                                     the standard.                                           .

       Standards

       [1] «CENELEC EN 50361
       [2] IEEE—P1528—200x% draft 6.5
       [3]     IEC PT 62209 draft 0.9                                                                     '                    ‘
     >(       Efl]\e ITIS CAD file is derived from [2]
                                                    andis also wnthln the tolerance requirements of the shapes of .0
               1} and [3]

       Conformity                                        §        (

       Based on the sample tests above, we certify thatthis item is in compliance with the uncertainty
       reqmrements of SAR measurements specified in standard [1] and draft standards [2] and [3].



       Date        ‘                                 28.02.2002   >

                                                                                    Schmid & Partner
     . Signature / Stamp                                                    Engineering AG
                                                                                12_:':;?'hnuutrnu 43, CH—800
                                                                                                             4 Zurich
                                                                                          1 245 97 00, Ffax 4411 245
                                                                                                                      97 79 °

                   FW

       Doc No 881 —00.000 P40 CA —B                                   uC                    —                         <o0          Page      C1 (1)_


  /A
   »
apr corm


Schmid & Partner Engineering AG                             s                e        a

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 246 9779
info@speag.com, http://www.speag.com




                                           IMPORTANT NOTICE
USAGE OF PROBES IN ORGANIC SOLVENTS
Diethylene Gycol Monobuthy Ether(the basis for liquids above 1 GHz), as many other organic
solvents, is a very effective softener for synthetic materials. These solvents can cause irreparable
damage to certain SPEAG products, except those which are explicitly declared as compliant with
organic solvents.

Compatible Probes:
— ET3DV6
— ET3DV6GR
— ES3DVx
— ER3DV6
— H3DV6




                                                     to solventslongerthan necessary for the
                                                     ifteruse with warm water andstoreddry.



$    p_     o_    a     _0
Schmid & Partner Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 $700, Fax+41 1 245 9779
info@apeag.com, http/www.speag.com




Schmid & Partner Enginecring AG




Technical Note 01.06.15—1                                                              June 2002


Schmid & Partner Engineering AG                                s            e       a

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 245 9779
info@speag.com, http://www.speag.com




INTRODUCTION OF NEW PROBE CONFIGURATION FILES (*.cf5)

Dear Customer,

Please be informed that as of November 2004, probe configuration files have been change to the new
version, i.e., *.cf5 configuration files. Together with the DASY4 v4.4 Build 3 software version, the
frequency compensation is applied to measured data providing extension of frequency validity of the
probe ConvF from + 50 MHz to £ 100 MHz with the same accuracy.


SOFTWARE INSTALLATION
The new software is available for download from the DASY4 homepage
http://w w w.dasy4.com/updates/updates.html
(via: DASY4 / Support & Downloads / Downloads / Software Updates)

The two new executable files:
DASY4.4 B3 EXE (item 3 within the DASY 4.4 Upgrade section)
SEMCAD V1.8 B130 EXE (item 4 within the DASY 4.4 Upgrade section)

To access and download the files, please use:
Username: dasy4
Password: woweool


Pleasedownloadthe wo executable
                     c          : files andreplace theexistingDASY4/SEMCAD
                                                                       CAD software
                                                                           8        _
           :                       DASY4directoryon your PCwith thesenewreleases at your earliest — >




Best regards
Your SPEAG DASY4 support team


 $    p        s   a     g
 Schmid & Partner Enginsering AG
 Zeughausstrasse 43, 8004 Zurich, Switzerland
 Phone +41 1 245 9700, Fax +41 1 245 9770
 info@speag.com, http:iwww.speag.com

Schmid & Partner Engineering AG




Technical Note 12.11.04—1                                                              November 2004


Calibration Laboratory of                                                                                       Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                Service suisse d‘étalonnage
   Engineering AG                                                                                               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                    Swiss Calibration Service


Accredited by the Swiss Federal Office of Metrology and Accreditation                         Accreditation No.: SCS 108
‘The Swiss Accreditation Service is one ofthe signatories to the EA
Multilateral Agreement for the recognition of calibration certificates




 Calipration procedure(s)




 Calibration date:


 Condition of the calibrated item



 This callbration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All callbrations have been conducted in the closed laboratory facility: environment temperature (22 # 3)°C and humidity < 70%.

 Calibration Equipment used (M&TE critical for calibration)

 Primary Standards                    ID #                      Cal Date (Calibrated by, Certificate No.)           Scheduled Calibration
 Power meter E44198                   GB41203874                5—May—04 (METAS, No. 251—00388)                     May—05
 Power sensor E4412A                  MY41495277                5—May—04 (METAS, No. 251—00388)                     May—05
 Reference 3 dB Aftenuator            SN: 85054 (3c)            10—Aug—04 (METAS, No. 251—00403)                    Aug—05                            ,
 Reference 20 dB Attenuator           SN: $5086 (206)           3—May—04 (METAS, No. 251—00389)                     May—05                        .
 Reference 30 dB Attenuator           SN: S5129 (30b)           10—Aug—04 (METAS, No. 251—00404)                    Aug—05                        ‘
 Reference Probe ESSDV2               SN: 3013                  8—Jan—04 (SPEAG, No. ES3—3013_Jan04)                Jan—05
 DAE4                                 SN: 617                   26—May—04 (SPEAG, No. DAE4—617_MayQ4)               May—05

 Secondary Standards                  ID #                      Check Date (in house)                               Scheduled Check
 Power sensor HP 8481A                MY41092180                18—Sep—02 (SPEAG, in house check Oct—03)            In house check: Oct 05
 RF generator HP 8648C                US3642U01700              4—Aug—99 (SPEAG, in house check Dec—03)             In house check: Dec—05
 Network Analyzer HP 8753E            US37390585                18—Oct—01 (SPEAG, in house check Nov—04)            In house check: Nov 05

                                      Name                                   Function                               Signature
 Calibrated by:



Approved by:



                                                                                                                    Issued: December 29, 2004
This calibration certificate shall not be reproduced except in full without written approvalof thelaboratory.


Certificate No: ET3—1790_Dec04                                      Page 1 of 9


Calibration
      .     Laboratory              of                                           S   Schweizerischer Kalibrierdienst
Schmid & Partner                                                                 C   Service suisse d‘étalonnage
  Engineering AG                                                                     Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                     S   Swiss Calibration Service



Accredited by the Swiss Federal Office of Metrology and Accreditation            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Glossary:
TSL                          tissue simulating liquid
NORMx,.y.z                   sensitivity in free space
ConF                         sensitivity in TSL / NORMx,y,z
DCP                          diode compression point
Polarization q               @ rotation around probe axis
Polarization 8               8 rotation around an axis that is in the plane normal to probe axis (at
                             measurement center), i.e., 8 = 0 is normal to probe axis

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", December 2003
   b) CENELEC EN 50361, "Basic standard for the measurement of Specific Absorption Rate
      related to human exposure to electromagnetic fields from mobile phones (300 MHz — 3
      GHz), July 2001

Methods Applied and Interpretation of Parameters:
   * NORMx,y,z: Assessed for E—field polarization 8 = 0 (f < 900 MHz in TEM—cell; f > 1800 MHz:
     R22 waveguide). NORMx,y,z are only intermediate values, i.e., the uncertainties of
     NORMx,y,z does not effect the E*—field uncertainty inside TSL (see below ConvF).
    *    NORM(Px,y,z = NORMx,y,z * frequency_response (see Frequency Response Chart). This
         linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of
         the frequency response is included in the stated uncertainty of ConvF.
    *    DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of                            a
         power sweep (no uncertainty required). DCP does not depend on frequency nor media.
    *    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or
         Temperature Transfer Standard for f < 800 MHz) and inside waveguide using analytical field
         distributions based on power measurements for f > 800 MHz. The same setups are used for
         assessment of the parameters applied for boundary compensation (alpha, depth) of which
         typical uncertainty values are given. These parameters are used in DASY4 software to
         improve probe accuracy close to the boundary. The sensitivity in TSL corresponds to
         NORMx,y,z * ConvF whereby the uncertainty corresponds to that given for ConvF. A
         frequency dependent ConvF is used in DASY 4.3 B17 and higher which allows extending
         the validity from £ 50 MHz to £ 100 MHz.
    *    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a
         flat phantom exposed by a patch antenna.
    *    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center
         from the probe tip (on probe axis). No tolerance required.

Certificate No: ET3—1790_DecQ4                                   Page 2 of 9


    ET3DV6 SN:1790                                                                December 20, 2004




                        Probe ET3DVGO

                                            SN:1790
                                 Manufactured:                   May 28, 2003
                                 Last calibrated:                August 29, 2003
                                 Repaired:                       October 5, 2004
                                 Recalibrated:                   December 20, 2004

                                     Calibrated for DASY Systems
                                      (Note: non—compatible with DASY2 system!)




Certificate No: ET3—1790_DecQ4                   Page 3 of 9


    ET3DV6 SN:1790                                                                                      December 20, 2004


    DASY — Parameters of Probe: ET3DV6 SN:1790

    Sensitivity in Free Space*                                                                 Diode CompressionB

               NormX                     2.07 + 9.9%                uV/(V/im)                 DCP x            94 mV
               NormY                     2.04 £9.9%                 uV/(V/im)®                DCP Y            94 mV
               NormZ                     1.178 £9.9%                uV/(V/imy                 DCP Z            94 mV

    Sensitivity in Tissue Simulating Liquid (Conversion Factors)
    Please see Page 8.



    Boundary Effect
    TSL                       900 MHz           Typical SAR gradient: 5 % per mm


              Sensor Center to Phantom Surface Distance                                       3.7 mm    4.7 mm
              SARpe [%]             Without Correction Algorithm                                  9.2    4.9
              SARpe [%]             With Correction Algorithm                                     0.1    0.2


    TSL                      1750 MHz           Typical SAR gradient: 10 % per mm


              Sensor Center to Phantom Surface Distance                                       3.7 mm 4.7 mm
              SARpe [%]             Without Correction Algorithm                                 12.6    8.4
              SARpe [%]             With Correction Algorithm                                     0.5    0.0


    Sensor Offset
              Probe Tip to Sensor Center                                                 2.7 mm




    The reported uncertainty of measurement is stated as the standard uncertainty of
    measurement multiplied by the coverage factor k=2, which for a normal distribution
    corresponds to a coverage probability of approximately 95%.


   * The uncertainties of NormX,Y,Z do not affect the E*—field uncertainty inside TSL (see Page 8).
   ® Numerical linearization parameter: uncertainty not required.




Certificate No: ET3—1790_Dec04                                Page 4 of 9


     ET3DV6 SN:1790                                                                                           December 20, 2004



                                                         Frequency Response of E—Field
                                                               (TEM—Cell:ifi110 EXX, Waveguide: R22)
               Frequency response (normalized)




                                                 0       500        1000        1500        2000       2500    3000
                                                                            f [MHz]
                                                     [         —e—TEM                    —o—R22    |


                                                     Uncertainty of Frequency Response of E—field: £ 6.3% (k=2)




Certificate No: ET3—1790_Dec04                                             Page 5 of 9


    ET3DV6 SN:1790                                                                 December 20, 2004



                                   Receiving Pattern (¢), 8 = 0°

                  f= 600 MHz, TEM ifif10EXX                           f= 1800 MHz, WG R22




                                                                                           —0—30 MHz
                                                                                           —BI— 100 MHz
     Error [dB]




                                                                                           —~>—600 MHz
                                                                                           —BI— 1800 MHz
                                                                                           —i&— 2500 MHz




                  0        60          120        180           240      300         360
                                                  $1


                                 Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




Cerfificate No: ET3—1790_Dec04                    Page 6 of 9


    ET3DV6 SN:1790                                                                                 December 20, 2004


                                                  Dynamic Range f(SARpeaq)
                                                        (Waveguide R22, f = 1800 MHz)



                                  1.2+7


                                  1.E+6



                                  1.E+5
              Input Signal [mV]




                                  1.E+4



                                  1.E+3


                                  1.E+2


                                  1.E+1



                                  1.E+0
                                     0.0001     0.001     0.01         0.1           1        10     100
                                                                 SAR [mWicm‘]
                                              —®— not compensated               —®~— compensated
               Error [dB]




                                   0.001         0.01            0.1             1            10           100
                                                                 SAR [mW/icm‘]


                                                 Uncertainty of Linearity Assessment: £ 0.6% (k=2)




Certificate No: ET3—1790_Dec04                                    Page 7 of 9


    ET3DV6 SN:1790                                                                                           December 20, 2004



                                       Conversion Factor Assessment

                     f= 900 MHz, WGLS R9 (head)                                    1= 1750 MHz, WGLS R22 (head)
               3.5                                                                30.0

               3.0       —                                                        25.0

               2.5 7
           3                                                                € 200 —
          "€
          5
             20              ?k
                                                                           C
                                                                           8 150
           €                                                                S
           £15
           C
                                                                            £&
                                                                            < 10.0
          8 10                                                              n                   ‘!ga%
               0.5                      %e                                         5.0                   &

               0.0                                            1                    0.0                        Cfeeraiee;
                                                                                                                   T     toront»
                     0            20           40             60                         0           20            40            60
                                       z[mm]                                                              z[mm]

               —®— Analytical          —O—Measurements                              —®— Analytical        —O—Measurements




    f [MHz] Validity [mHz]®              TSL   Permittivity        Conductivity      Alpha      Depth        ConvF Uncertainty

    835         £ 50 / £ 100            Head   41.5%5%             0.90£5%               0.72     1.69         6.94 £11.0% (k=2)

    900         £ 50 / £ 100            Head   41.5%5%             0.97%5%               0.63    1.84          6.68 £11.0% (k=2)

    1750        £50/£100                Head   40.1%25%            1.37+5%               0.55    240           5.51 £11.0% (k=2)

    1900        #50/#100                Head   40.0%25%            1.40£5%               0.53    2.51          5.26 £11.0% (k=2)

    2450        £50/£100                Head   39.2%5%             1.80£5%               0.63    2.30          4.74 #11.8% (k=2)




    835         £ 50/ £ 100             Body   55.215%             0.97%5%               0.50    2.07          6.65 2 11.0% (k=2)

    900         £ 50 / £ 100            Body   55.0%5%             1.05#5%               0.50   2.13           6.35 £11.0% (k=2)

    1750        £50/£100                Body   53.4%5%              149%5%               0.52    2.81         4.83 £11.0% (k=2)
    1900        £50/ 100                Body   53.3+5%              1.52%5%              0.52   2.96          4.71 #11.0% (k=2)

    2450        £50/£100                Body   §2.7%5%             1.95%5%               0.59   2.33          4.35 £11.8% (k=2)




   © The validity of £ 100 MHz only applies for DASY v4.4 and higher (see Page 2). The uncertainty is the RSS
   of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band.




Cerfificate No: ET3—1790_Dec04                            Page 8 of 9


    ET3DV6 SN:1790                                                                                            December 20, 2004



                                 Deviation from Isotropy in HSL
                                                 Error (6, 8), f = 900 MHz




                                                                                                 Error [dB]




                           MI—1.00——0.80 ©1—0.80——0.60 MI—0.60~0.40 Hi—0.40——0.20 ©A—0.20—0.00
                             0.00—0.20   ©10.20—0.40   ©0.40—0.60    ©0.60—0.80   MI0.80—1.00




                             Uncertainty of Spherical Isotropy Assessment: 2.6% (k=2)




Certificate No: ET3—1790_Dec04                         Page 9 of 9


D3: DAE


                                                                                                                                                    /A—
                                                                                                                                                »
                                                                                                                                             aor corn




  Calibration Laboratory of                                                                                       Schwelzerischer Kallbrierdionst
  Schmid & Partner                                                                                                Service sulsse d‘étalonnage
     Engineering AG                                                                                               Servizio svizzero di taratura
  Zoughausstrasse 43, 8004 Zurich, Switzeriand                                                                    Swiss    Callbration Service

  Aceredited by the Swiss Federal Offce of Metrology and Accreditaion                                                     SCs 108
  The Swiss Accreditation Service is one of the signatories to the EA
  Multilateral Agreement for the recognition of callbration certificates
  Client       ADT (Auden)                                                                     Certificate No: DAE3—510_AugO5

  [CALIBRATION CERTIFICATE                                                                                                                                |
   Object                               DAE3 — SD 000 D03 AA — SN: 510

   Callbration procedure(s)             QA CAL—O6.v12
                                        Calibration procedure for the data acquisition unit (DAE)



   Calibration date:                    August 17, 2005

   Conditon of the calibrated item      In Tolerance

   This calbration certifate documents the traceabiltyto national standards, which realize the physical units of measurements (1).
   The measurements and the uncertainties with confidence probabilty are gven on the following pages and are partof the certficate.

   All ealbrations have been conducted in the closed laboratory facilty: environment temperature (22 # 3)°C and humidity < 70%.

   Callbration Equipment used (M&TE criical for callbration)

   Primary Standards                 1#                            Gal Date (Callbrated byCertficate No.)                 Scheduled Calibration
   Fluke Process Caltorator Type 702 SN: 6205803                   7—Sep—04 (Sintrel, No.€—040073)                        Sep—05


   Secondary Standards                   1#                 Check Date (in houso)                                         Scheduled Check
   Gallbrator Box V1.1                   SE UMS 008 AB 1002 29—/un—05 (SPEAG, in house check)                             in house check Jun—06




                                         Name                                  Function                                   Signature
                                                                               Technician                                  <5         eliscs        oo
                                                                                                              o           ee
   Caltbrated by:                        Eic Halnfold




   Approved by:                          Fin Bomhott                            R&D Director


                                                                                                                          Issued: August 17, 2005
   This callbration certiicate shall not be reproduced excapt in full wihout writen approval of the laboratory.

‘ Certificate No: DAE3—510_Augo5                                      Page1 of 5


                                                                                                                           P
                                                                                                                        aor corn




  Calibration Laboratory of                                                       [/alse        &   Schweizerischer Kalibriordionst
  Schmid & Partner                                                               (f                 Service sulsse d‘étalonnage




                                                                                      €

                                                                                                o
                                                                                                    Sorvizio avizzoro di taratura




                                                                                 w


                                                                                            N
    Engineering AG
  Zoughausstrasse 43, 8004 Zurich, Switzeriand                                        ory       S   swiss Calibration Service

  Aceredited by the Swiss Federal Office of Metrology and Accreditation                         Accreditation No.: SCS 108
  The Swiss Accroditation Service is one of the signatories to the EA
  Multilatoral Agroomont for the recognition of callbration certificates



  Glossary
  DAE                          digital acquisition electronics
  Connector angle              information used in DASY system to align probe sensor X to the robot
                               coordinate system.

  Methods Applied and Interpretation of Parameters
      *     DC Voitage Measurement: Calibration Factor assessed for use in DASY system by
           comparison with a calibrated instrumenttraceable to national standards. The figure given
           corresponds to the full scale range of the voltmeterin the respective range.
      *    Connector angle: The angle of the connector is assessed measuring the angle
           mechanically by a tool inserted. Uncertainty is not required.
      *    The following parameters contain technical information as a result from the performance
           test and require no uncertainty.
      *    DC Voltage MeasurementLinearity: Verification of the Linearity at +10% and —10% of the
           nominal calibration voltage. Influence of offset voltageis included in this measurement.
      «_   Common mode sensitivity: Influence of a positive or negative common mode voltage on the
           differential measurement.
      *    Channel separation: Influence of a voltage on the neighbor channels not subject to an input
           voltage.
      «_   AD Converter Values with inputs shorted: Values on the internal AD converter
           corresponding to zero input voltage
      *    Input Offset Measurement: Output voltage and statisticalresults over a large number of
           zero voltage measurements.
      *    Input Offset Current: Typical value for information; Maximum channel input offset current,
           not considering the input resistance.
      *    Input resistance: DAE input resistance at the connector, during internal auto—zeroing and
           during measurement.
      *    Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
           alarm signal is generated.
      *    Power consumption: Typical value for information. Supply currents in various operating
           modes.




‘ Certificate No: DAE3—510_Aug05                                   Page 2 of 5


                                                                                                           /a—
                                                                                                         apr corn




   DC Voltage Measurement
      ATD — Converter Resolution nominal
          High Range:         1LsB =           6.uV ,      full range    100...+300 mV
         Low Range:        1LSB =         6inV,         full range           .+3mV
      DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

        Calibration Factors                x                         Y                     2Z
        High Range                403.415 + 0.1% (k=2) 403.485 + 0.1% (k=2)      |_403.851 £0.1% (k=2)
        Low Range                 3.95575 + 0.7% (k=2) 3.93400 + 0.7% (k=2)      3.95875 + 0.7% (k=2)



  Connector Angle

        Connector Angle to be used in DASY system                                        44°%1°




‘ Certiicate No: DAE3—510_Aug05                         Page3 of 5


                                                                                                                /A
                                                                                                              aor corn




  Appendix
  1. DC Voltage Linearity
        High Range                             Input (uV)            Reading (V)               Error (%)
        Channel X       + Input                 200000                 200000.5                 0.00
        Channel X       + Input                  20000                 20004.73                 0.02
        Ghannel X       — Input                  20000                 —19990.82               —0.05
        Channel Y       + Input                 200000                 200000.4                 0.00
        Channel Y       + Input                  20000                 20002.02                 0.01
        Channel Y       — Input                  20000                 —19995.68               —0.02
        Channel Z       + Input                 200000                 200000.2                 0.00
        Channel Z       + Input                  20000                 20006.25                 0.03
        Channel Z       — Input                  20000                 —19996.09               —0.02


        Low Range                              Input (uV)            Reading (V)               Error (%)
        Channel X       + Input                    2000                 2000.1                  0.00
        Channel X       + Input                     200                 200.36                  0.18
        Channel X       — Input                     200                 —200.01                 0.00
        Channel Y       + Input                    2000                 1999.9                  0.00
        Channel Y       + Input                     200                 199.23                 —0.38
        Channel Y       — Input                     200                 —200.56                 0.28
        Channel Z       + Input                    2000                  2000                   0.00
        Channel Z       + Input                     200                 199.08                 —0.46
        Channel Z       — Input                     200                 —200.99                 0.50


  2. Common mode sensitivity
      DASY measurement parameters: Auto Zero Time: 3 sec: Measuring time: 3 see
                          Common mode                       High Range                     Low Range
                          Input Voltage (mV)         Average Reading (uV)              Average Reading (1V)
       Channel X                      200                      17.56                         16.61
                                     —200                     —15.84                        —17.03
       Channel Y                      200                      14.35                         14.26
                                     —200                     —15.35                        —15.86
       Channel Z                      200                      —8.76                         —8.81
                                     —200                      741                           746


  3. Channel separation
      DASY measurement parameters: Auto Zero Time: 3 sec; Measuringtime: 3 sec
                         Input Voltage (mV)     ChannelX(uV)           ChannelY (uw)       Channel Z (V)
       Channel X                 200                  —                    317                  o64
       Channel Y                 200                o23                      .                  a.72
       Channel Z                 200                .73                    —0.05                  —

‘ Certificate No: DAE3—510_Augo5                          Page4 of 5


                                                                                                                 /A—
                                                                                                                  »
                                                                                                               apr corn




   4. AD—Converter Values with inputs shorted
       DASY measurementparameters: Auto Zero Time: 3 sec; Measuring time: 3 see
                                                High Range (LSB)              Low Range (LSB)
       Channel X                                             15976                            16467
       Channel Y                                             16203                            16387
       Channel 2                                             16171                            16258

  5. Input Offset Measurement
      DASY measurementparameters: Auto Zero Time: 3 sec; Measuring time: 3 see
      input t0Mc
                         Average (uV)      min. Offset (V) max. Offset (uv)                    5* '(’:“;;
       Channel X                   o.21                 —0.66                  0.72                 o2r
       Channel Y                   117                  —3.01                  0.16                 0.33
       Channel 2                   142                  —2.98                 —0.81                 0.29

  6. Input Offset Current
       Nominal Input circuitry offset current on all channels: <25fA

  7. Input Resistance
                                                    Zeroing (MOhm)                      Measuring (MOhm)
       Channel X                                         0.2001                               201.1
       Channet Y                                         0.2001                               199.6
       Channel Z                                         0.2000                               200.5

  8. Low Battery Alarm Voltage (verified during pre test)
       Typical values                               Alarm Level (VDC)
       Supply (+ Vee)                                                            +7.9
       Supply (—Vee)                                                             .6

  9. Power Consumption (veriied during pre test)
       Typical values                         Switched off (mA)        Stand by (mA)       Transmitting (mA)
       Supply (+ Vee)                                +0.0                   +6                    +14
       Supply (— Vee)                               —0.01                   —8                     —a




‘ Certifcate No: DAE3—510_AugO$                             Page 5 of 5


  /A
   »
apr corm


                                                                                                                                            /a—
                                                                                                                                      apr corn




Calibration Laboratory of                                                                                 s    Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                         _ Service suisse d‘étalonnage
  Engineering AG                                                                        $ @6               Servizio svizzero ditaratura
Zoughausstrasse 43, 8004 Zurich, Switzerland                                            C                s Swiss Calibration Service

Aceredited by the Swiss Federal Offce of Metrology and Accreditation                         Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agroement for the recognition of callbration cortificatos
Client        ADT (Auden)                                                                    Cortificate No: D2450V2—716_AugOs

CALIBRATION CERTIFICATE
 Object                               D2450V2 — SN: 716

 Calibration procedure(s)             QA CAL—05.v6
                                      Calibration procedure for dipole validation kits



 Calibration date:                   August 23, 2005

 Gondition of the calibrated item     In Tolerance


 This calbration certficate documents the traceabiity to national standards, which realize the physical units of measurements (S).
 The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the certficate.
 All callations have been conducted in the closed laboratory facilty: environment temperature (22 + 3)°C and humidity < 70%.

 Calbration Equipment used (M&TE crtial for callbration)

 Primary Standards                     io #                     Gal Date (Calibrated by Crtificate No.)            Scheduled Calteration
 Power meter EPM E442                  GBsra80704               12:0ct04 (METAS, No. 251—00412)                    Oct0s
 Power sensor HP 8481A                 ussrz02703               12:0ct—04 (METAS, No. 251—00412)                   Oct.0s
 Reference 20 dB Attenuator            SN: 086 (209)            11—Aug—05 (METAS, No 251—00498)                    Aug—06
 Reference 10 dB Attenuator            SN: 50472 (10)           11—Aug—05 (METAS, No 251—00498)                    Aug.—06
 Reference Probe ESSDV2                sn 3025                  20—Cct.04 (GPEAG, No. ES3—3025_Octo4)              Oct0s
 Dars                                  Sn 601                   O7—Jan—05 (SPEAG, No. DAE4—601_Jan05)              Jan—06

 Secondary Standards                   D#                       Check Date (n house)                               Scheduled Check
 Power sensor HP 8481A                 Mys1002317               18—0ct—02 (GPEAG, in house chack Oct—03)           In house check: Oct.05
 RF generator R&S SML—03               100898                   27—ar—02 (GPEAG, in house check Dec—03)            in house check: Dec—05
 Network Analyzer HP 8763E             ussr3s0ses 54206         18—0c—01 (GPEAG, in house check Nov—04)            in house check: Nov—05

                                      Name                                   Function                              Signature
 Callbrated
              M   by:                 Mike Moil                              Labo
                                                                                  S3    Technician
                                                                                                                   tirokt
Approved by:                          Katja Pokovie                         Technical Manager
                                                                                                                 Ze [%
                                                                                                                  Issued: August 25, 2005
 This calibration certficate shall not be reproduced except in full wihout writen approval o the laboratory.

Certificate No: D2450V2—716_AugO5                                   Page 1 of 9


                                                                                                            /A—
                                                                                                          »
                                                                                                       apr corn




Calibration Laboratory of                                                     & Schweizerischer Kalibrierdionst
Schmid & Partner                                                              g   Sorvice suisso d‘étalonnage
   Engineering AG                                                                 Servizio svizzero di taratura
Zoughausstrasse 43, 8004 Zurich, Switzerland                                  S   swiss Caltbration Service

Aceredited by the Swiss Federal Offce of Metrology and Accreditation          Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of callbration cortificatos

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM xy,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques®, December 2003
   b) CENELEC EN 50361, "Basic standard for the measurement of Specific Absorption Rate
      related to human exposure to electromagnetic fields from mobile phones (300 MHz — 3
         GHz), July 2001
    0) Federal Communications Commission Office of Engineering & Technology (FCC OET),
         "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
         Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
         Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
         Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
   d) DASY4 System Handbook
Methods Applied and Interpretation of Parameters:
     Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
        Antenna Parameters with TSL: The dipole is mounted with the spacer to positionits feed
        point exactly below the center marking of the flat phantom section, with the arms oriented
        paralle! to the body axis.
        Feed Point Impedance and Return Loss: These parameters are measured with the dipole
        positioned under the liquid filled phantom. The impedance stated is transformed from the
        measurement at the SMA connectorto the feed point. The Return Loss ensures low
        reflected power. No uncertainty required.
        Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
        No uncertainty required.
        SAR measured: SAR measured at the stated antenna input power
        SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
        connector.
        SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
        nominal SAR result.


Certiicate No: D2450V2—716_Aug0$                                Page 2 of 9


                                                                                                                     /S—
                                                                                                                      #
                                                                                                                   aor corn




Measurement Conditions
    DASY system configuration, as far as not given on page 1.
       DASY Version                                             DASY4                                   v4.6
      Extrapolation                                    Advanced Extrapolation
      Phantom                                         Modular Flat Phantom V$.0
      Distance Dipole Center— TSL                               10 mm                                with Spacer
      Area Scan resolution                                 dx, dy = 15 mm
      Zoom Scan Resolution                                dx, dy, dz = 5 mm
      Frequency                                          2450 MHz & 1 MHz



Head TSL parameters
    The following parameters and calculations were appled.
                                                                Temperature           Permittivity        Conductivity
      Nominal Head TSL parameters                                     22.0 °C            30.2              1.80 mhofm
      Measured Head TSL parameters                              (22.0 £0.2) °C        38.5 16 %         1.73 mhoim + 6 %
      Head TSL temperature during test                          (21.0202) °C          38.4 £6 %         1.77 mhoim +6 %


SAR result with Head TSL

      SAR averaged over 1 cm° (1 g) of Head TSL                  condition
      SAR measured                                         250 mW input power                     13.5 mw /g
      SAR normalized                                         normalized to 1W                     54.0 mW /g
      SAR for nominal Head TSL parameters ‘                  normalized to 1W            54.4 mW / g + 17.0 % (k=2)


      SAR averaged over 10 cm (10 g) of Head TSL                 condition
      SAR measured                                         250 mW input power                     6.23 mW /g
      SAR normalized                                         normalized to 1W                     24.9 mW /g
      SAR for nominal Head TSL parameters ‘                  normalized to 1W            25.1 mW / g + 16.5 % (k=2)




" Correction to nominal TSL parameters according to d), chapter "SAR Sensitivities®

Certificate No: D2450V2—716_AugO5                       Page 3 of 9


                                                                                                                           /S—
                                                                                                                        #
                                                                                                                     aor corn




    Body TSL parameters
        The following parameters and calculations were applied.
                                                                     Temperature           Pormittivity        Conductivity
           Nominal Body TSL parameters                                    22.0 °C             52.7              1.95 mho/m
           Measured Body TSL parameters                             (22.0 £0.2) °C         52.5 £ 6 %        2.03 mhoim + 6 %
           Body TSL temperature during test                         (21.5 £0.2) °C         52.5 £ 6 %        2.06 mhoim + 6 %



    SAR result with Body TSL

          SAR averaged over 1 cm° (1 g) of Body TSL                   condition
          SAR measured                                          250 mW input power                      13.6 mW /g
          SAR normalized                                          normalized to 1W                      54.4 mW /g
          SAR for nominal Body TSL parameters °                   normalized to 1W            53.1 mW / g + 17.0 % (k=2)

          SAR averaged over 10 m (10 g) of Body TSL                   condition
          SAR measured                                          250 mW input power                      6.28 mW /g
          SAR normalized                                          normalized to 1W                      25.1 mW /g
          SAR for nominal Body TSL parameters *                   normalized to 1W            24.5 mW / g 4 16.5 % (k=2)




|
    * Correction to nominal TSL parameters according to d), chapter "SAR Sensitivities:"

    Certificate No: D2450V2—716_Augo$                        Page4 of 9


                                                                                                                      /A—
                                                                                                                     #
                                                                                                                  aor corn




Appendix
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           s3.7 n + 2.5 in
      Retun Loss                                                                         —27.3 dB

Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           50.5 n + 4.0 j
      Return Loss                                                                       —27.9 dB



General Antenna Parameters and Design

     [ Electrical Detay (one direction)                      |                           1.147 ns

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                             September 10 , 2002




Certificate No: D24G0V2—716_Aug05                        Page 5 of 9


                                                                                                        /A—
                                                                                                       D
                                                                                                    apr corn




DASY4 Validation Report for Head TSL

                                                                                 Date/Time: 23.08.2005 17:40:59

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN716
Communication System: CW—2450; Frequency: 2450 MHz; Duty Cyecle: 1:1
Medium: HSL U1O0 BB;
Medium parameters used: £= 2450 MHz; 0 = 1.77 mho/m; s = 38.4; p = 1000 kg/m?
Phantom section: Flat Section
Measurement Standard: DASY4 (High Precision Assessment)
DASY4 Configuration:
    *   Probe: ES3DV2 — SN3025; ConvF(4.4, 4.4, 4.4); Calibrated: 29.10.2004
    *   Sensor—Surfice: 4mm (Mechanical Surface Detection)
    *   Electronics: DAE4 Sn601; Calibrated: 22.07.2004
    *   Phantom: Flat Phantom 5.0 (front); Type: QDO0OPSOAA
    *   Measurement SWDASY4, V4.6 Build 9; Postprocessing SW: SEMCAD, V1L8 Build 151

Pin = 250 mW; d = 10 mm/Area Scan (41x61x1):
Measurementgrid: dx=15mm, dy—=1 5mm
Maximum value of SAR (interpolated) = 17.5 mW/g

Pin = 250 mW; d = 10 mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 87.4 V/m; Power Drift = 0.097 dB
Peak SAR (extrapolated) = 28.0 W/kg
SAR(I g) = 13.5 mW/g; SAR(1O g) = 6.23 mW/g
Maximum value of SAR (measured)= 15.3 mW/g




        —16.0


        —24.0


        ~32.0


        ~40.0

                   0 4B = 153mwWig

Certificate No: D2450V2—716_AugO$                   Page 6 of 9


                                                                                                 /A—
                                                                                                  »
                                                                                               aor corn




    Impedance Measurement Plot for Head TSL


                                                          22 mug 2005 serasiis
               san      a ure           asaruse 5L2
                                                assore 16494 pH       2 450.000 c00WHz




         ny
         1e°



         cn2




         ty
         1t°


                     Start 2 250.000 co0 miz                          sToe 2 650.000 a00 riz




    Certificate No: D2450V2—716_Aug05                   Page 7 of 9

|


                                                                                                              /A—
                                                                                                             D
                                                                                                          apr corn




     DASY4 Validation Report for Body TSL
                                                                                       Date/Time: 23.08.2005 16:16:42

    Test Laboratory: SPEAG, Zurich, Switzerland

    DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN716
    Communication System: CW—2450; Frequency: 2450 MHz; Duty Cycle: 1:1
    Medium: MSL 2450;
    Medium parameters used: £= 2450 MHz; 0 = 2.05 mho/m; s = 52.5; p = 1000 kg/ml
    Phantom section: Flat Section
    Measurement Standard: DASY4 (High Precision Assessment)
    DASY4 Configuration:
        *   Probe: ES3DV2 — SN3025; ConvEF(4.13, 4.13, 4.13); Calibrated: 29.10.2004
        *   Sensor—Surface: 4mm (Mechanical Surface Detection)
        *   Electronics: DAE4 Sn601; Calibrated: 22.07.2004
        *   Phantom: Flat Phantom 5.0 (back); Type: QDO00PSOAA
        *   Measurement SW; DASY4, V4.6 Build 9; Postprocessing SW: SEMCAD, V1L8 Build 151

    Pin = 250 mW; d = 10 mm/Area Scan (81x81x1):
    Measurementgrid: dx=15mm, dy=1 5mm
    Maximum value of SAR (interpolated) = 15.4 mW/g

    Pin = 250 mW; d = 10 mm/Zoom Scan (7x7x7)/Cube 0:
    Measurement grid: dx=3mm, dy=5mm, dz=Smm
    Reference Value = 89.2 V/m; Power Drift=—0.055 dB
    Peak SAR (extrapolated) = 28.3 W/kg
    SAR(I g) = 13.6 mW/g; SAR(10 g) = 6.28 mWi/g
    Maximum value of SAR (measured) = 15.6 mW/g

            dB
            0.00


            ~8.00



            ~16.0


            —24.0


            —32.0



            —40.0

                       0 4B = 15.6mWig



|
    Certificate No: D2450V2—716_Aug0$                    Page 8 of 9


                                                                                               aor corn




    Impedance Measurement Plot for Body TSL


                                                               22 mug 2e05 d4rssrar
          cen
          CH)   s     i u rs            a se.451 a   see52026210
                                                     «as52e z6213 pH         2 450.000 o0 4z




          fve
          16



          cna



          ton




                                                                              &

                    Sthet 2 250.000 no0 niz                            Stor 2 ese.000 soe iz




|
    Certificate No: D2450V2—716_Augo5                    Page9 of 9



Document Created: 2006-04-06 08:41:44
Document Modified: 2006-04-06 08:41:44

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC