04T2458 SAR Attachment 5

FCC ID: NM8VIVIDA

RF Exposure Info

Download: PDF
FCCID_463591

Calibration Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland




Client




 Object(s)



 Calibration procedure(s)




 Calibration date:


 Condition of the calibrated item



 This calibration statement documents traceability of M&TE used in the calibration procedures and conformity of the procedures with the ISO/IEC
 17025 international standard.


 All calibrations have been conducted in the closed laboratory facility: environment temperature 22 +/— 2 degrees Celsius and humidity < 75%.


 Calibration Equipment used (M&TE critical for calibration)


[Model Type          _                  iD# ________                          _CalDate____                        _ Scheduled Calibration ___________
 RF generator R&S SML—O3                100698                                 27—Mar—2002                            In house check: Mar—05
 Power sensor HP 8481A                  MY41092317                             18—0ct—02                              Oct—04
 Power sensor HP 8481A                  US37202783                             30—0ct—02                              Oct—03
 Power meter EPM E442                   GB37480704                             30—0ct—02                              Oct—03
 Network Analyzer HP 8753E              US38432426                             3—May—00                               In house check: May 03




 Calibrated by:



 Approved by:

                                                                                                                                    7

                                                                                                                       Date issued: April 11, 2003



 This calibration certificate is issued as an intermediate solution until the accreditation process (based on ISO/AEC 17025 International Standard) for
 Calibration Laboratory of Schmmid & Partner Engineering AG is completed.




880—KPO301061—A                                                                                                                                Page 1 (1)


Schmid & Partner Engineering AG                        s    p    e   a   g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 245 9779
info@speag.com, http://www.speag.com




                                               DASY
                      Dipole Validation Kit

                           Type: D900V2

                         Serial:                108

                      Manufactured:             March 9, 2001
                            Calibrated:         April 10, 2003


1.     Measurement Conditions


The measurements were performed in the flat section of the SAM twin phantom filled with head
simulating solution of the following electrical parameters at 900 MHz:

       Relative Dielectricity                 42.1          +5%
       Conductivity                           0.95 mho/m    + 5%

The DASY4 System with a dosimetric E—field probe ET3DV6 (SN:1507, Conversion factor 6.6 at 900
MHz) was used for the measurements.

The dipole was mounted on the small tripod so that the dipole feedpoint was positioned below the
center marking of the flat phantom section and the dipole was oriented parallel to the body axis (the
long side of the phantom). The standard measuring distance was 15mm from dipole center to the
solution surface. The included distance holder was used during measurements for accurate distance
positioning.

The coarse grid with a grid spacing of 15mm was aligned with the dipole. The 7x7x7 fine cube was
chosen for cube integration.
The dipole input power (forward power) was 250mW + 3 %. The results are normalized to IW input
power.



2.     SAR Mcasurement with DASY4 System

Standard SAR—measurements were performed according to the measurement conditions described in
section 1. The results (see figure supplied) have been normalized to a dipole input power of 1 W
(forward power). The resulting averaged SAR—values measured with the dosimetric probe ET3DV6
SN:1507 and applying the advanced extrapolation are:


       averaged over 1 ocm‘ (1 g) oftissue:          10.5 mW/g + 16.8 % (k=2)‘

       averaged over 10 em" (10 g) of tissue:        6.76 mWi/ig + 16.2 % (k=2)‘


3.     Dipole Impedance and Return Loss

The impedance was measured at the SMA—connector with a network analyzer and numerically
transformed to the dipole feedpoint. The transformation parameters from the SMA—connector to the
dipole feedpoint are:

     Electrical delay:          1.396 ns    (one direction)
     Transmission factor:       0.991       (voltage transmission, one direction)


The dipole was positioned at the flat phantom sections according to section 1 and the distance holder
was in place during impedance measurements.


       Feedpoint impedance at 900 MHz:              Re{Z} = 51.9 0

                                                     Im {Z} = —3.1 0

       Return Loss at 900 MHz                        —28.9 dB



4.     Handling

Do not apply excessive force to the dipole arms, because they might bend. Bending of the dipole arms
stresses the soldered connections near the feedpoint leading to a damage of the dipole.


5.     Design

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is
directly connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—
signals.


6.     Power Test

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint
can be measured.


3.     Dipole Impedance and Return Loss

The impedance was measured at the SMA—connector with a network analyzer and numerically
transformed to the dipole feedpoint. The transformation parameters from the SMA—connector to the
dipole feedpoint are:

     Electrical delay:          1.396 ns    (one direction)
     Transmission factor:       0.991       (voltage transmission, one direction)


The dipole was positioned at the flat phantom sections according to section 1 and the distance holder
was in place during impedance measurements.


       Feedpoint impedance at 900 MHz:              Re{Z} = 51.9 0

                                                     Im {Z} = —3.1 0

       Return Loss at 900 MHz                        —28.9 dB



4.     Handling

Do not apply excessive force to the dipole arms, because they might bend. Bending of the dipole arms
stresses the soldered connections near the feedpoint leading to a damage of the dipole.


5.     Design

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is
directly connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—
signals.


6.     Power Test

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint
can be measured.





Document Created: 2019-10-10 07:09:56
Document Modified: 2019-10-10 07:09:56

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC