Test Report_SAR_APPENDIX_A B

FCC ID: NDPM3A-2

RF Exposure Info

Download: PDF
FCCID_4312985

APPENDIX A. SAR System Verification Data

The plots for system verification with largest deviation for each SAR system combination are
shown as follows.


                                                               Date/Time: 23/04/2019
Test Laboratory: CERPASS TECH
Dipole Calibration for Body Tissue Pin=250mW, dist=10mm, f=2450 MHz
DUT: Dipole 2450 MHz D2450V2; Type: D2450V2; Serial: D2450V2
Communication System: CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; σ = 1.95 S/m; εr = 52.70; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)
DASY Configuration:

      Probe: EX3DV4 - SN3927; ConvF(7.61, 7.61,7.61); Calibrated: 2018/6/22;
      Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 31.0
      Electronics: DAE4 Sn1379; Calibrated: 2018/6/6
      Phantom: ELI v5.0; Type: QDOVA002AA
      DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)

Configuration/Area Scan (7x7x1): Measurement grid: dx=12mm, dy=12mm, Maximum
value of SAR (measured) = 15 W/kg
Configuration/Zoom Scan (9x9x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm,
dz=5mm, Reference Value = 105.21 V/m; Power Drift = 0.02 dB, Peak SAR (extrapolated)
= 26.4 W/kg
SAR(1 g) = 12.8 W/kg; SAR(10 g) = 5.7W/kg Maximum value of SAR (measured) = 18.7
W/kg




           0 dB = 18.7 W/kg = 12.72 dBW/kg


                                                               Date/Time: 23/04/2019
Test Laboratory: CERPASS TECH
Dipole Calibration for Body Tissue Pin=100mW, dist=10mm, f=5250 MHz
DUT: Dipole D5GHzV2; Type: D5GHzV2; Serial: D5GHzV2
Communication System: CW; Frequency: 5250 MHz
Medium parameters used: f = 5250 MHz; σ = 5.31 S/m; εr = 48.95; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration:

      Probe: EX3DV4 - SN3927; ConvF(5.44, 5.44, 5.44); Calibrated: 2018/6/22;
      Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 25.0
      Electronics: DAE4 Sn1379; Calibrated: 2018/6/6
      Phantom: ELI v5.0; Type: QDOVA002AA
      DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)

Configuration/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm, Maximum
value of SAR (measured) = 16.3 W/kg
Configuration/Zoom Scan (9x9x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm, Reference Value = 68.85 V/m; Power Drift = 0.01 dB, Peak SAR (extrapolated)
= 31.3 W/kg
SAR(1 g) = 7.67 W/kg; SAR(10 g) = 2.17 W/kg Maximum value of SAR (measured) =
17.4 W/kg




           0 dB = 17.4 W/kg = 12.41 dBW/kg


                                                               Date/Time: 24/04/2019
Test Laboratory: CERPASS TECH
Dipole Calibration for Body Tissue Pin=100mW, dist=10mm, f=5750 MHz
DUT: Dipole D5GHzV2; Type: D5GHzV2; Serial: D5GHzV2
Communication System: CW; Frequency: 5750 MHz
Medium parameters used: f = 5750 MHz; σ = 5.99 S/m; εr =48.27; ρ = 1000 kg/m3
Phantom section: Flat Section; Meas. Ambient Temp (celsius) -21℃; Input power-100mW
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007)
DASY Configuration:

      Probe: EX3DV4 - SN3927; ConvF(4.55, 4.55, 4.55); Calibrated: 2018/6/22;
      Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 25.0
      Electronics: DAE4 Sn1379; Calibrated: 2018/6/6
      Phantom: ELI v5.0; Type: QDOVA002AA
      DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)

Configuration/Area Scan (7x7x1): Measurement grid: dx=10mm, dy=10mm, Maximum
value of SAR (measured) = 18.9 W/kg
Configuration/Zoom Scan (9x9x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm, Reference Value = 71.84 V/m; Power Drift = 0.02 dB, Peak SAR (extrapolated)
= 34.5 W/kg
SAR(1 g) = 8.1 W/kg; SAR(10 g) = 2.35 W/kg Maximum value of SAR (measured) = 19.5
W/kg




          0 dB = 19.5 W/kg = 12.90 dBW/kg


APPENDIX B. SAR measurement Data

The SAR plots for highest measured SAR in each exposure configuration, wireless mode and
frequency band combination, and measured SAR > 1.5 W/kg are shown as follows.


                                                               Date/Time: 24/04/2019
Test Laboratory: CERPASS TECH
802.11b 2412MHz Edge_1
DUT: Mobile POS Terminal; Type: M3a-2
Communication System: UID 0, 2.4GHz Wi-Fi (0); Frequency: 2412MHz;
Medium parameters used: f = 2412 MHz; σ = 1.95 S/m; εr =52.73; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)
DASY5 Configuration:

      Probe: EX3DV4 - SN3927; ConvF(7.53, 7.53, 7.53); Calibrated: 2018/6/22;
      Sensor-Surface: 2.5mm (Mechanical Surface Detection), z = 1.0, 31.0
      Electronics: DAE4 Sn1379; Calibrated: 2018/6/6
      Phantom: ELI v5.0; Type: QDOVA002AA
      DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)



Configuration/Area Scan (4x9x1): Measurement grid: dx=12mm, dy=12mm
Maximum value of SAR (measured) =1.85 W/kg

Configuration/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm,
dz=5mm
Reference Value = 2.361 V/m; Power Drift = 0.65 dB
Peak SAR (extrapolated) = 3.72 W/kg
SAR(1 g) = 1.51 W/kg; SAR(10 g) = 0.546 W/kg
Maximum value of SAR (measured) =2.96 W/kg




                  0 dB = 2.96 W/kg = 4.71 dBW/kg


                                                               Date/Time: 28/04/2019
Test Laboratory: CERPASS TECH
802.11a 5240MHz Edge_1
DUT: Mobile POS Terminal; Type: M3a-2
Communication System: UID 0, 5GHz Wi-Fi (0); Frequency: 5240MHz;
Medium parameters used: f = 5240 MHz; σ = 5.26 S/m; εr = 48.60; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)
DASY5 Configuration:

      Probe: EX3DV4 - SN3927; ConvF(5.01, 5.01, 5.01); Calibrated: 2018/6/22;
      Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 25.0
      Electronics: DAE4 Sn1379; Calibrated: 2018/6/6
      Phantom: ELI v5.0; Type: QDOVA002AA
      DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)



Configuration/Area Scan (5x11x1): Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) =2.38 W/kg

Configuration/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=4mm, dy=4mm,
dz=1.4mm
Reference Value = 0 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 6.48 W/kg
SAR(1 g) = 1.42 W/kg; SAR(10 g) = 0.327 W/kg
Maximum value of SAR (measured) =3.70 W/kg




                  0 dB = 3.70 W/kg = 5.68 dBW/kg


                                                               Date/Time: 28/04/2019
Test Laboratory: CERPASS TECH
802.11a 5745MHz Edge_1
DUT: Mobile POS Terminal; Type: M3a-2
Communication System: UID 0, 5GHz Wi-Fi (0); Frequency: 5745MHz;
Medium parameters used: f = 5745 MHz; σ = 5.99 S/m; εr =47.71; ρ = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011)
DASY5 Configuration:

      Probe: EX3DV4 - SN3927; ConvF(4.46, 4.46, 4.46); Calibrated: 2018/6/22;
      Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 25.0
      Electronics: DAE4 Sn1379; Calibrated: 2018/6/6
      Phantom: ELI v5.0; Type: QDOVA002AA
      DASY52 52.8.8(1222); SEMCAD X 14.6.10(7331)



Configuration/802.11a 5240MHz Horizontal-Up(Ant 1)/Area Scan (5x11x1):
Measurement grid: dx=10mm, dy=10mm
Maximum value of SAR (measured) =2.46 W/kg

Configuration/802.11a 5240MHz Horizontal-Up(Ant 1)/Zoom Scan (9x9x7)/Cube 0:
Measurement grid: dx=4mm, dy=4mm, dz=1.4mm
Reference Value = 0 V/m; Power Drift = 0.00 dB
Peak SAR (extrapolated) = 7.65 W/kg
SAR(1 g) = 1.42 W/kg; SAR(10 g) = 0.344 W/kg
Maximum value of SAR (measured) =4.29 W/kg




                  0 dB = 4.29 W/kg =6.32 dBW/kg


APPENDIX C. Calibration Data for Probe, Dipole and DAE

Please refer to attached files.


APPENDIX D. Photographs of EUT and Setup

Please refer to attached files.



Document Created: 2019-05-23 11:52:36
Document Modified: 2019-05-23 11:52:36