14.8. D835V2 - SN 4d142

FCC ID: N7NMC7355

RF Exposure Info

Download: PDF
FCCID_2058258

Calibration Laboratory of                                                                                  S       Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Muitilateral Agreement for the recognition of calibration certificates

Client




 Object



 Calibration procedure(s)




 Calibration date:




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (St1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 * 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter EPM—442A                   GB37480704                05—Oct—11 (No. 217—01451)                             Oct—12
 Power sensor HP 8481A                  US37292783                05—Oct—11 (No. 217—01451)                             Oct—12
 Reference 20 dB Attenuator             SN: 5058 (20k)            27—Mar—12 (No. 217—01530)                             Apr—13
 Type—N mismatch combination            SN: 5047.2 / 06327        27—Mar—12 (No. 217—01533)                             Apr—13
 Reference Probe ES3DOV3                SN: 3205                  30—Dec—11 (No. ES3—3205_Dect1)                        Dec—12
 DAE4                                   SN: 601                   27—Jun—12 (No. DAEA4—601_Jun12)                       Jun—13


 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power sensor HP 8481A                  MY41092317                18—Oct—02 (in house check Oct—11)                     In house check: Oct—13
 RF generator R&S SMT—O6                100005                    04—Aug—99 (in house check Oct—11)                     in house check: Oct—13
 Network Analyzer HP 8753E              US37390585 $4206           18—Oct—01 (in house check Oct—11)                    In house check: Oct—12




                                        Name                                                                             Signat
 Calibrated by:



 Approved by:



                                                                                                                        issued: October 5, 2012
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D835V2—4d142_Oct12                                     Page 1 of 8


Calibration Laboratory of                                                         Schweizerischer Kalibrierdienst
Schmid & Partner                                                                  Service suisse d‘étalonnage
  Engineering AG                                                                  Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                      Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                            Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                         tissue simulating liquid
ConvrF                       sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2003, "IEEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
      Communications Devices: Measurement Techniques", December 2003
   b) IEC 62209—1, "Procedure to measure the Specific Absorption Rate (SAR) for hand—held
      devices used in close proximity to the ear (frequency range of 300 MHz to 3 GHz)",
       February 2005
   c) Federal Communications Commission Office of Engineering & Technology (FCC OET),
      "Evaluating Compliance with FCC Guidelines for Human Exposure to Radiofrequency
       Electromagnetic Fields; Additional Information for Evaluating Compliance of Mobile and
       Portable Devices with FCC Limits for Human Exposure to Radiofrequency Emissions",
      Supplement C (Edition 01—01) to Bulletin 65

Additional Documentation:
    d) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    e    Measurement Conditions: Further details are available from the Validation Report at the end
         of the certificate. All figures stated in the certificate are valid at the frequency indicated.
    e    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
         point exactly below the center marking of the flat phantom section, with the arms oriented
         parallel to the body axis.
    e    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
         positioned under the liquid filled phantom. The impedance stated is transformed from the
         measurement at the SMA connector to the feed point. The Return Loss ensures low
         reflected power. No uncertainty required.
    e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
         No uncertainty required.
         SAR measured: SAR measured at the stated antenna input power.
         SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
         connector.
    e    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the —
         nominal SAR result.

 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probability of approximately 95%.



Certificate No: D835V2—4d142_Oct12                               Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                               V52.8.2
      Extrapolation                                   Advanced Extrapolation

      Phantom                                          Modular Flat Phantom
      Distance Dipole Center — TSL                              15 mm                             with Spacer

     Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                          835 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                  22.0 °C           41.5                0.90 mho/m
      Measured Head TSL parameters                              (22.0 + 0.2) °C   41.0 #£6 %          0.90 mho/m + 6 %
      Head TSL temperature change during test                      <0.5 °C            ——~~                     =———


SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                   2.37 mW / g
      SAR for nominal Head TSL parameters                    normalized to 1W         9.45 mW /g + 17.0 % (k=2)


      SAR averaged over 10 cm‘* (10 g) of Head TSL                condition
      SAR measured                                        250 mW input power                     1.56 mW / g
      SAR for nominal Head TSL parameters                    normalized to 1W         6.23 mW /g £ 16.5 % (k=2)



Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Body TSL parameters                                  22.0 °C            55.2               0.97 mho/m

      Measured Body TSL parameters                              (22.0 + 0.2) °C    53.3 +6 %          1.00 mho/im + 6 %
      Body TSL temperature change during test                      <0.5 °C                                      _=——


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
      SAR measured                                        250 mW input power                     245 mW / g
      SAR for nominal Body TSL parameters                    normalized to 1W         9.50 mW / g + 17.0 % (k=2)


      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                        250 mW input power                     1.61 mW / g
      SAR for nominal Body TSL parameters                    normalized to 1W         6.29 mW / g + 16.5 % (k=2)




Certificate No: DB835V2—4d142_Oct12               v    Page 3 of 8


Appendix

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           53.2 O — 2.0 jQ
      Return Loss                                                                       — 28.7 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           48.6 Q — 4.7 jQ
      Return Loss                                                                       —26.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.425 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                March 27, 2012




Certificate No: D835V2—4d142_Oct12                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                  Date: 03.10.2012

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 40142

Communication System: CW; Frequency: 835 MHz
Medium parameters used: £= 835 MHz; 0 = 0.9 mho/m; s = 41; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

       Probe: ES3DV3 — SN3205; ConvF(6.07, 6.07, 6.07); Calibrated: 30.12.2011;

        Sensor—Surface: 3mm (Mechanical Surface Detection)

       Electronics: DAE4 S$n601; Calibrated: 27.06.2012

       Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

       DASY52 52.8.2(969); SEMCAD X 14.6.6(6824)


Dipole Calibration for Head Tissue/Pin=250 mW, d=15mm 2/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=>5mm, dz=5mm
Reference Value = 57.215 V/m; Power Drift = —0.03 dB
Peak SAR (extrapolated) = 3.504 mW/g
SAR(1 g) = 2.37 mW/g; SAR(10 g) = 1.56 mW/g
Maximum value of SAR (measured) = 2.76 W/kg




        3.00



        6.00


       |—9.00



        —12.00


        —15.00
                   0 dB = 2.76 W/kg = 8.82 dB W/kg




Certificate No: D835V2—4d142_Oct12             Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                           3 Octf 2012    11:07:36
                ~   S$41.
                     .      1 UOFS          1159.223
                                                   * 4   —1.9590   4
                                                         florriodsdbid 97.2398 pF        $35.000 900
                                                                                                  .  MHz




          De 1


          Cog.




          Pre
          159


          Hid.

          CHZ



          Bel


          Cor




          Prz
          16

          H1d
                                635.000   d@@ MHz                                TOP 1 190.000 009 MHz




Certificate No: D835V2—4d142_Oct12                          Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 04.10.2012

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 — SN: 4d142

Communication System: CW; Frequency: 835 MHz
Medium parameters used: f = 835 MHz; 0 = 1 mho/m; s, = 53.3; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2007)

DASY52 Configuration:

   &   Probe: ES3DV3 — SN3205; ConvF(6.02, 6.02, 6.02); Calibrated: 30.12.2011;

   e    Sensor—Surface: 3mm (Mechanical Surface Detection)

   e _ Electronics: DAF4 S$n601; Calibrated: 27.06.2012

   e   Phantom: Flat Phantom 4.9L; Type: QDOOOP49AA; Serial: 1001

   e   DASY52 52.8.2(969); SEMCAD X 14.6.6(6824)


Dipole Calibration for Body Tissue/Pin=250 mW, d=15mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 55.416 V/m; Power Drift = 0.01 dB
Peak SAR (extrapolated) = 3.566 mW/g
SAR(1 g) = 2.45 mW/g; SAR(10 g) = 1.61 mW/g
Maximum value of SAR (measured) = 2.85 W/kg




                   0 dB = 2.85 W/kg =9.10 dB W/kg




Certificate No: D835V2—4d142_Oct12             Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                           4 Oct 2012       @8:55:18
          CH si1         i4 ouors            11 48.563 &   —4.673§% n   48.781 pF          835.000000 MHz

                                                                                                              CH1 Markers
                                                                                                               3©60.397 a
                                                                                                                  32623 ¢



                                                    Tom,
                                                                                                              1989.000 MHz


                                        LoioA@IRDGSw
          Cor:




                                             s
          Ax                                54
          153

          H1q

          CH2Z



                                                                                                              CH2Z Markers

          Cor
                                                                                                              2:—10,.528 dB
                                                                                                               9909.000 MHz




          Ax
          16"

          H1d

                   ‘ST        E35.000        FH=z                                   ‘STOP 4 109.000 0BB MHz




Certificate No: D835V2—4d142_Oct12                           Page 8 of 8



Document Created: 2012-10-24 16:01:04
Document Modified: 2012-10-24 16:01:04