SAR Test report Appendix A

FCC ID: N7NEM7455

RF Exposure Info

Download: PDF
FCCID_3899266

Test Laboratory: BTL Inc.                 Date: 2018/5/30


System Check_B750_0530


DUT: Dipole 750 MHz D750V3;SN:1095;


Communication System: UID 0, CW (0); Frequency: 750 MHz; Duty Cycle: 1:1
                                                                               3
Medium parameters used: f = 750 MHz; σ = 0.957 S/m; εr = 55.552; ρ = 1000 kg/m
Ambient Temperature:23.2 ℃;    Liquid Temperature:22.5    ℃


DASY Configuration:


       Probe: EX3DV4 - SN7396; ConvF(10.3, 10.3, 10.3); Calibrated: 2018/5/29;
       Sensor-Surface: 4mm (Mechanical Surface Detection), z = 1.0, 31.0
       Electronics: DAE4 Sn1390; Calibrated: 2018/5/11
       Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1222
       DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)



Area Scan (8x14x1): Interpolated grid: dx=15 mm, dy=15 mm
Maximum value of SAR (interpolated) = 2.25 W/kg


Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 53.38 V/m; Power Drift = -0.02 dB
Peak SAR (extrapolated) = 3.04 W/kg
SAR(1 g) = 2.09 W/kg; SAR(10 g) = 1.39 W/kg
Maximum value of SAR (measured) = 2.25 W/kg


Test Laboratory: BTL Inc.                 Date: 2018/5/31


System Check_B835_0531


DUT: Dipole 835 MHz D835V2;SN:4d160;


Communication System: UID 0, CW (0); Frequency: 835 MHz; Duty Cycle: 1:1
                                                                               3
Medium parameters used: f = 835 MHz; σ = 0.977 S/m; εr = 55.621; ρ = 1000 kg/m
Ambient Temperature:23.4 ℃;    Liquid Temperature:22.4    ℃


DASY Configuration:


       Probe: EX3DV4 - SN7396; ConvF(9.89, 9.89, 9.89); Calibrated: 2018/5/29;
       Sensor-Surface: 4mm (Mechanical Surface Detection), z = 1.0, 31.0
       Electronics: DAE4 Sn1390; Calibrated: 2018/5/11
       Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1222
       DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)



Area Scan (7x13x1): Interpolated grid: dx=15 mm, dy=15 mm
Maximum value of SAR (interpolated) = 2.57 W/kg


Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 56.26 V/m; Power Drift = -0.01 dB
Peak SAR (extrapolated) = 3.40 W/kg
SAR(1 g) = 2.34 W/kg; SAR(10 g) = 1.53 W/kg
Maximum value of SAR (measured) = 2.54 W/kg


Test Laboratory: BTL Inc.                 Date: 2018/6/1


System Check_B1750_0601


DUT: Dipole 1750 MHz D1750V2;SN:1101;


Communication System: UID 0, CW (0); Frequency: 1750 MHz; Duty Cycle: 1:1
                                                                                3
Medium parameters used: f = 1750 MHz; σ = 1.499 S/m; εr = 52.347; ρ = 1000 kg/m
Ambient Temperature:23.5 ℃;    Liquid Temperature:22.3     ℃


DASY Configuration:


       Probe: EX3DV4 - SN7396; ConvF(8.2, 8.2, 8.2); Calibrated: 2018/5/29;
       Sensor-Surface: 4mm (Mechanical Surface Detection), z = 1.0, 31.0
       Electronics: DAE4 Sn1390; Calibrated: 2018/5/11
       Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1222
       DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)



Area Scan (5x7x1): Interpolated grid: dx=15 mm, dy=15 mm
Maximum value of SAR (interpolated) = 10.6 W/kg


Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 92.72 V/m; Power Drift = -0.03 dB
Peak SAR (extrapolated) = 15.0 W/kg
SAR(1 g) = 9.11 W/kg; SAR(10 g) = 5.07 W/kg
Maximum value of SAR (measured) = 10.2 W/kg


Test Laboratory: BTL Inc.                 Date: 2018/6/2


System Check_B1900_0602


DUT: Dipole 1900 MHz D1900V2;SN:5d179;


Communication System: UID 0, CW (0); Frequency: 1900 MHz; Duty Cycle: 1:1
                                                                                3
Medium parameters used: f = 1900 MHz; σ = 1.545 S/m; εr = 52.512; ρ = 1000 kg/m
Ambient Temperature:23.4 ℃;    Liquid Temperature:22.5     ℃


DASY Configuration:


       Probe: EX3DV4 - SN7396; ConvF(7.91, 7.91, 7.91); Calibrated: 2018/5/29;
       Sensor-Surface: 4mm (Mechanical Surface Detection), z = 1.0, 31.0
       Electronics: DAE4 Sn1390; Calibrated: 2018/5/11
       Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1222
       DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)



Area Scan (5x6x1): Interpolated grid: dx=15 mm, dy=15 mm
Maximum value of SAR (interpolated) = 11.8 W/kg


Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 96.76 V/m; Power Drift = -0.10 dB
Peak SAR (extrapolated) = 17.2 W/kg
SAR(1 g) = 9.78 W/kg; SAR(10 g) = 5.18 W/kg
Maximum value of SAR (measured) = 11.0 W/kg


Test Laboratory: BTL Inc.                 Date: 2018/6/3


System Check_B2600_0603


DUT: Dipole 2600 MHz D2600V2;SN:1067;


Communication System: UID 0, CW (0); Frequency: 2600 MHz; Duty Cycle: 1:1
                                                                                3
Medium parameters used: f = 2600 MHz; σ = 2.234 S/m; εr = 52.085; ρ = 1000 kg/m
Ambient Temperature:23.2 ℃;    Liquid Temperature:22.2     ℃


DASY Configuration:


       Probe: EX3DV4 - SN7396; ConvF(7.3, 7.3, 7.3); Calibrated: 2018/5/29;
       Sensor-Surface: 4mm (Mechanical Surface Detection), z = 1.0, 31.0
       Electronics: DAE4 Sn1390; Calibrated: 2018/5/11
       Phantom: ELI v5.0; Type: QDOVA002AA; Serial: TP:1222
       DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)



Area Scan (6x6x1): Interpolated grid: dx=12 mm, dy=12 mm
Maximum value of SAR (interpolated) = 15.7 W/kg


Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 94.65 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) = 30.0 W/kg
SAR(1 g) = 13.7 W/kg; SAR(10 g) = 6.03 W/kg
Maximum value of SAR (measured) = 15.6 W/kg