SAR Evaluation FCC Annex A

FCC ID: N5A6847

RF Exposure Info

Download: PDF
FCCID_4085394

                Test report no.: 18-1-0044501T06a-C3
                Issue Date:8/11/2018




                             Annex A
                     System performance check

1. System Performance Check for Head and Body Tissue simulating liquid

•   System Performance Check 2450 MHz Head
•   System Performance Check 2450 MHz Body
•   System Performance Check 2450 MHz Body (2018-10-24)
•   System Performance Check 2450 MHz Head (2018-11-08)


                                                              Date: 29.06.2018

Test Laboratory: Cetecom Essen

System Performance Check 2450 MHz Head250mW

DUT: D2450V2; Type: D2450V2; Serial: 993

Communication System: UID 0, CW (0); Communication System Band: D2450
(2450 MHz); Frequency: 2450 MHz;
Medium parameters used: f = 2450 MHz; σ = 1.831 S/m; εr = 40.165; ρ = 1000
kg/m3
Phantom section: Flat Section

DASY Configuration:

   •   Probe: EX3DV4 - SN3739; ConvF(7.34, 7.34, 7.34); Calibrated: 15.06.2018;
   •   Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 31.0
   •   Electronics: DAE4 Sn1233; Calibrated: 16.02.2017
   •   Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD;
       Serial: 1639
   •   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)


Configuration/Body/Area Scan (7x7x1): Measurement grid: dx=15mm,
dy=15mm

Maximum value of SAR (measured) = 18.8 W/kg

Configuration/Body/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 75.58 V/m; Power Drift = -0.17 dB
Maximum value of SAR (measured) = 17.3 W/kg




       0 dB = 18.8 W/kg = 12.74 dBW/kg


                                                               Date: 29.06.2018

Test Laboratory: Cetecom Essen

System Performance Check 2450 MHz Body 250mW

DUT: D2450V2; Type: D2450V2; Serial: 993

Communication System: UID 0, CW (0); Communication System Band: D2450
(2450 MHz); Frequency: 2450 MHz;
Medium parameters used: f = 2450 MHz; σ = 2.02 S/m; εr = 52.291; ρ = 1000
kg/m3
Phantom section: Flat Section

DASY Configuration:

   •   Probe: EX3DV4 - SN3739; ConvF(4.9, 4.9, 4.9); Calibrated: 15.06.2018;
   •   Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 23.0
   •   Electronics: DAE4 Sn1233; Calibrated: 16.02.2017
   •   Phantom: Twin-SAM right V5.0 (30deg); Type: QD 000 P40 CD; Serial:
       1640
   •   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)


Configuration/Body/Area Scan (7x7x1): Measurement grid:
dx=15mm, dy=15mm

Maximum value of SAR (measured) = 13.2 W/kg

Configuration/Body/Zoom Scan (7x7x7)/Cube 0: Measurement grid:
dx=5mm, dy=5mm, dz=5mm
Reference Value = 105.3 V/m; Power Drift = -0.13 dB
Maximum value of SAR (measured) = 14.8 W/kg




              0 dB = 14.8 W/kg = 11.70 dBW/kg


                                                               Date: 24.10.2018

Test Laboratory: Cetecom Essen

System Performance Check 2450 MHz Body 250mW

DUT: D2450V2; Type: D2450V2; Serial: 993

Communication System: UID 0, CW (0); Communication System Band: D2450
(2450 MHz); Frequency: 2450 MHz;
Medium parameters used: f = 2450 MHz; σ = 1.999 S/m; εr = 53.181; ρ = 1000
kg/m3
Phantom section: Flat Section

DASY Configuration:

   •   Probe: EX3DV4 - SN3739; ConvF(4.9, 4.9, 4.9); Calibrated: 15.06.2018;
   •   Sensor-Surface: 1.4mm (Mechanical Surface Detection), z = 1.0, 23.0
   •   Electronics: DAE4 Sn1233; Calibrated: 09.08.2018
   •   Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD;
       Serial: 1639
   •   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)

Configuration/Body/Area Scan (7x7x1): Measurement grid: dx=15mm,
dy=15mm

Maximum value of SAR (measured) = 15.4 W/kg

Configuration/Body/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 75.58 V/m; Power Drift = -0.13 dB
Peak SAR (extrapolated) = 26.4 W/kg
SAR(1 g) = 13.5 W/kg; SAR(10 g) = 6.41 W/kg




             0 dB = 15.4 W/kg = 11.64 dBW/kg


                                                              Date: 08.11.2018

Test Laboratory: Cetecom Essen

System Performance Check 2450 MHz Head 2018-11-08

DUT: D2450V2; Type: D2450V2; Serial: 993

Communication System: UID 0, CW (0); Communication System Band: D2450
(2450 MHz); Frequency: 2450 MHz;
Medium parameters used: f = 2450 MHz; σ = 1.857 S/m; εr = 40.054; ρ = 1000
kg/m3
Phantom section: Flat Section

DASY Configuration:

   •   Probe: EX3DV4 - SN3739; ConvF(7.24, 7.24, 7.24); Calibrated: 15.06.2018;
   •   Sensor-Surface: 3mm (Mechanical Surface Detection), z = 1.0, 31.0
   •   Electronics: DAE4 Sn1233; Calibrated: 09.08.2018
   •   Phantom: Twin-SAM V5.0 (30deg probe tilt); Type: QD 000 P40 CD;
       Serial: xxxx
   •   DASY52 52.8.8(1258); SEMCAD X 14.6.10(7373)


Configuration/Head/Area Scan (5x7x1): Measurement grid: dx=15mm,
dy=15mm
Maximum value of SAR (measured) = 68.2 W/kg

Configuration/Head/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm,
dy=5mm, dz=5mm
Reference Value = 111.6 V/m; Power Drift = -0.13 dB
Peak SAR (extrapolated) = 109 W/kg
SAR(1 g) = 52.5 W/kg; SAR(10 g) = 24.5 W/kg




             0 dB = 68.2 W/kg = 18.34 dBW/kg



Document Created: 2018-11-08 11:10:12
Document Modified: 2018-11-08 11:10:12

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC