Test Report HAC RF Appendix A

FCC ID: MSQZ01QD

Test Report

Download: PDF
FCCID_3958368

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/7/9


HAC_E_Dipole_835
DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.3 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1399; Calibrated: 2017/11/16
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 130.1 V/m; Power Drift = 0.00 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 112.5 V/m
Average value of Total=(109.8+112.5) / 2 = 111.15 V/m

                                         PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          108.0 V/m      109.8 V/m   107.9 V/m
                                          Grid 4 M4      Grid 5 M4   Grid 6 M4
                                          64.50 V/m      65.95 V/m   65.11 V/m
                                          Grid 7 M4      Grid 8 M4   Grid 9 M4
                                          110.5 V/m      112.5 V/m   111.0 V/m
                                                       Cursor:
                                                  Total = 112.5 V/m
                                                    E Category: M4
                                              Location: -0.5, 70.5, 9.7 mm




                    0 dB = 112.5 V/m = 41.02 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/7/9


HAC_E_Dipole_1880
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.3 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1399; Calibrated: 2017/11/16
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 147.4 V/m; Power Drift = 0.03 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 94.67 V/m
Average value of Total=(88.89+94.67) / 2 = 91.78 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          87.69 V/m      88.89 V/m   87.30 V/m
                                          Grid 4 M3      Grid 5 M3   Grid 6 M3
                                          72.47 V/m      73.67 V/m   72.74 V/m
                                          Grid 7 M3      Grid 8 M3   Grid 9 M3
                                          93.31 V/m      94.67 V/m   92.48 V/m
                                                        Cursor:
                                                   Total = 94.67 V/m
                                                    E Category: M3
                                                Location: 0.5, 31, 9.7 mm




                    0 dB = 94.67 V/m = 39.52 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2018/7/9


HAC_E_Dipole_2600
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.3 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1399; Calibrated: 2017/11/16
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (0);SEMCAD X Version 14.6.10 (7417)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 76.31 V/m; Power Drift = 0.03 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 96.49 V/m
Average value of Total=(87.99+96.49) / 2 = 92.24 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          86.80 V/m      87.99 V/m   86.90 V/m
                                          Grid 4 M3      Grid 5 M3   Grid 6 M3
                                          90.42 V/m      92.64 V/m   91.43 V/m
                                          Grid 7 M3      Grid 8 M3   Grid 9 M3
                                          94.61 V/m      96.49 V/m   94.70 V/m
                                                        Cursor:
                                                   Total = 96.49 V/m
                                                    E Category: M3
                                                Location: 0, 20.5, 9.7 mm




                    0 dB = 96.49 V/m = 39.69 dBV/m



Document Created: 2018-07-27 17:51:18
Document Modified: 2018-07-27 17:51:18