Appendix E ISL-E-CAL-266-HB-1298

FCC ID: MAU047

RF Exposure Info

Download: PDF
FCCID_1776976

            NCL CALIBRATION LABORATORIES

                     Calibration File No.: 1289-1298

                                       Client.: ISL

C E R T I F I C A T E OF C A L I B R A T I O N

      It is certified that the equipment identified below has been calibrated in the
   NCL CALIBRATION LABORATORIES by qualified personnel following recognized
          procedures and using transfer standards traceable to NRC/NIST.



                       Equipment: Miniature Isotropic RF Probe
                                   Record of Calibration
                                       Head and Body
                          Manufacturer: APREL Laboratories
                                     Model No.: E-020
                                       Serial No.: 266


    Calibration Procedure: D01-032-E020-V2, D22-012-Tissue, D28-002-Dipole
                           Project No: ISL-E020-5612

                              Calibrated: 8th August 2011
                             Released on: 15th August 2011

                              Approved By:         Stuart Nicol




                NCL CALIBRATION LABORATORIES
                  303 Terry Fox Drive, Suite 102          Division of APREL
                                Kanata, Ontario           TEL: (613) 435-8300
                             CANADA K2K 3J1               FAX: (613) 435-8306


NCL Calibration Laboratories
Division of APREL Inc.


Introduction

This Calibration Report reproduces the results of the calibration performed in line with the references
listed below. Calibration is performed using accepted methodologies as per the references listed
below. Probes are calibrated for air, and tissue and the values reported are the results from the
physical quantification of the probe through meteorgical practices.

Calibration Method

Probes are calibrated using the following methods.

<1000MHz
TEM Cell for sensitivity in air
Standard phantom using temperature transfer method for sensitivity in tissue

>1000MHz
Waveguide* method to determine sensitivity in air and tissue
*Waveguide is numerically (simulation) assessed to determine the field distribution and power

The boundary effect for the probe is assessed using a standard flat phantom where the probe output
is compared against a numerically simulated series of data points

References

    o   IEEE Standard 1528 (2003) including Amendment 1
        IEEE Recommended Practice for Determining the Peak Spatial-Average Specific Absorption
        Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
        Techniques
    o   EN 62209-1 (2006)
        Human Exposure to RF Fields from hand-held and body-mounted wireless communication
        devices - Human models. instrumentation, and procedures-Part 1: Procedure to measure
        the Specific Absorption Rate (SAR) for hand-held mobile wireless devices
    o   IEC 62209-2 Ed. 1.0 (2010-03)
        Human exposure to RF fields from hand-held and body-mounted wireless devices - Human
        models, instrumentation, and procedures - Part 2: specific absorption rate (SAR) for
        wireless communication devices (30 MHz - 6 GHz)
    o   TP-D01-032-E020-V2 E-Field probe calibration procedure
    o   D22-012-Tissue dielectric tissue calibration procedure
    o   D28-002-Dipole procedure for validation of SAR system using a dipole
    o   IEEE 1309 Draft Standard for Calibration of Electromagnetic Field Sensors and Probes,
        Excluding Antennas, from 9kHz to 40GHz




Page 2 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.


Conditions

Probe 266 was a recalibration.
**The probe was received in good working order, although at 1900MHz the uncertainty was higher
than our standard (see note**)

Ambient Temperature of the Laboratory:               22 °C +/- 1.5°C
Temperature of the Tissue:                           21 °C +/- 1.5°C
Relative Humidity:                                   < 60%

Primary Measurement Standards

Instrument                                  Serial Number                      Cal due date
Power meter Anritsu MA2408A                 90025437                           Nov.4, 2011
Power Sensor Anritsu MA2481D                103555                             Nov 4, 2011
Attenuator HP 8495A (70dB)                  1944A10711                         Sept. 14, 2011
Network Analyzer Anritsu MT8801C            MB11855                            Feb. 8, 2012

Secondary Measurement Standards

Signal Generator Agilent E4438C -506 MY55182336                                June 7, 2012


Attestation

The below named signatories have conducted the calibration and review of the data which is
presented in this calibration report.




Page 3 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.



Probe Summary


Probe Type:                                     E-Field Probe E020

Serial Number:                                  266

Frequency:                                      As presented on page 5

Sensor Offset:                                  1.56

Sensor Length:                                  2.5

Tip Enclosure:                                  Composite*

Tip Diameter:                                   < 2.9 mm

Tip Length:                                     55 mm

Total Length:                                   289 mm

*Resistive to recommended tissue recipes per IEEE-1528

Sensitivity in Air

Channel X:                                                  1.2 µV/(V/m)2
Channel Y:                                                  1.2 µV/(V/m)2
Channel Z:                                                  1.2 µV/(V/m)2

Diode Compression Point:                                    95 mV




Page 4 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.


Calibration for Tissue (Head H, Body B)

                                                                                 Tolerance
                  Tissue        Measured        Measured       Calibration                    Conversion
Frequency                                                                       Uncertainty
                   Type          Epsilon         Sigma         Uncertainty                      Factor
                                                                                  for 5%*
  450 H            Head              X               X               X               X            X
   450 B           Body              X               X               X               X            X
  750 H            Head              X               X               X               X            X
   750 B           Body              X               X               X               X            X
  835 H            Head            42.35           0.94             3.5             3.4          6.6
  835 B            Body            56.65           1.018            3.5             3.4          6.8
  900 H            Head            41.35           0.98             3.5             3.4          6.4
  900 B            Body            56.08           1.05             3.5             3.4          6.7
  1450 H           Head              X               X               X               X            X
  1450 B           Body              X               X               X               X            X
  1500 H           Head              X               X               X               X            X
  1500 B           Body              X               X               X               X            X
  1640 H           Head              X               X               X               X            X
  1640 B           Body              X               X               X               X            X
  1750 H           Head              X               X               X               X            X
  1750 B           Body              X               X               X               X            X
  1800 H           Head            40.56           1.37             3.5             2.7           5.4
  1800 B           Body            52.16           1.56             3.5             2.7           5.5
  1900 H           Head            38.19           1.43             3.5             2.7           5.5
  1900 B           Body            51.87           1.59             3.5             2.7           5.3
  2000 H           Head              X              X                X               X             X
  2000 B           Body              X              X                X               X             X
  2100 H           Head              X              X                X               X             X
  2100 B           Body              X              X                X               X             X
  2300 H           Head              X              X                X               X             X
  2300 B           Body              X              X                X               X             X
  2450 H           Head            38.06           1.87             3.5             3.5            5
  2450B            Body            50.52           2.04             3.5             3.5          4.55
  2600 H           Head              X               X               X               X            X
  2600 B           Body              X               X               X               X            X
  3000 H           Head              X               X               X               X            X
  3000 B           Body              X               X               X               X            X
  3600 H           Head              X               X               X               X            X
  3600 B           Body              X               X               X               X            X
  5200 H           Head              X               X               X               X            X
  5200 B           Body              X               X               X               X            X
  5600 H           Head              X               X               X               X            X
  5600 B           Body              X               X               X               X            X
  5800 H           Head              X               X               X               X            X
  5800 B           Body              X               X               X               X            X



Page 5 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.



Boundary Effect:

Uncertainty resulting from the boundary effect is less than 2.1% for the distance between the tip of
the probe and the tissue boundary, when less than 0.58mm.


Spatial Resolution:

The spatial resolution uncertainty is less than 1.5% for 4.9mm diameter probe.
The spatial resolution uncertainty is less than 1.0% for 2.5mm diameter probe.

DAQ-PAQ Contribution

To minimize the uncertainty calculation all tissue sensitivity values were calculated using a load
impedance of 5 MΩ.

Boundary Effect:

For a distance of 0.58mm the worst case evaluated uncertainty (increase in the probe sensitivity) is
less than 2.1%.

NOTES:

*The maximum deviation from the centre frequency when comparing the lower to upper range is
listed.

**The 1900MHz body calibration had an assessed deviation of +8.52%. This value has been
normalized to be within 1.4% assessed deviation. As the deviation measured would have contributed
to a higher SAR value APREL can conclude that SAR measurements made with this probe will have
been within typical uncertainty of 10% and that this would not contribute to SAR value which if
corrected would yield a higher than reported value.




Page 6 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.




Receiving Pattern Air


                                  Receiving Pattern Probe E020 S/N 266




                                                        0
                                              3501.2          10
                                        340                         20
                                  330                                     30
                                                    1
                               320                                              40
                             310                 0.8                                 50
                         300                     0.6                                      60
                       290                       0.4                                       70
                      280                        0.2                                           80

                     270                            0                                          90

                      260                                                                      100
                       250                                                                 110
                         240                                                              120
                             230                                               130
                               220                                           140
                                  210                                     150
                                        200                         160
                                              190             170
                                                        180


                                        Ch1         Ch2        Ch3             Tot




Page 7 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.




Isotropy Error Air

                                                 Isotropy Error
                                                 Probe S/N 266


            10
             9
             8
             7
             6
             5
             4
             3
         ] 2
         B
         d
         [ 1
         r 0
         o
         rr -1
          E -2
            -3
            -4
            -5
            -6
            -7
            -8
            -9
           -10

                   0           100                        200                          300




                                           Isotropy Plot Probe S/N 266


      0. 8


      0. 7


      0. 6
    )
    V
        0. 5
    (m
     e
     g
     a
     tl 0. 4
      o
      V
      0. 3


      0. 2


      0. 1


         0

               0                     100                                   200               300
                                                         D egrees (x10)


                                                            Ch1                  Ch2
                                                            Ch3




Isotropicity Tissue:                                                      0.10 dB




Page 8 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.




Dynamic Range

                                          Dynamic Range Probe S/N 266


          10000




          1000



    ]
    V       100
    o
    r
    ic
     m
     [
     e                                                                                                                 Uncompensated
     g      10
     a
     lt                                                                                                                Compensated
      o
      V

              1




            0.1




           0.01

                  1.00   10.00   100.00      1,000.00          10,000.00   100,000.00   1,000,000.00   10,000,000.00

                                          Electric F ield [ (V/m )^2]




Page 9 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Inc.




Video Bandwidth

                                  Probe Frequency Characteristics
    )
    B
    d
    ( 1
    l
    e
    v
    e
    l 0
    t
    u
    p
    t
    u -1
    o
    e
    b
    o
    r -2
    P

     -3

     -4

     -5

     -6
           1                 10                   10 0                 1000                 100 00
                                                                                Frequency (Hz)




Video Bandwidth at 500 Hz                                1 dB
Video Bandwidth at 1.02 KHz:                             3 dB




Test Equipment

The test equipment used during Probe Calibration, manufacturer, model number and, current
calibration status are listed and located on the main APREL server R:\NCL\Calibration
Equipment\Instrument List May 2011.




Page 10 of 10
This page has been reviewed for content and attested to on Page 2 of this document.