            NCL CALIBRATION LABORATORIES


                       Calibration File No: DC-1401
                     Project Number: ISL-D5200-5640


C E R T I F I C A T E OF C A L I B R A T I O N

      It is certified that the equipment identified below has been calibrated in the
   NCL CALIBRATION LABORATORIES by qualified personnel following recognized
          procedures and using transfer standards traceable to NRC/NIST.



                              ISL Body Validation Dipole


                          Manufacturer: APREL Laboratories
                             Part number: ALS-D-5200-S-2
                                 Frequency: 5200 MHz
                              Serial No: 5200-230-00802


                                     Customer: ISL


                                           th
                             Calibrated: 25 January 2012
                                             th
                            Released on: 25 January 2012


NCL Calibration Laboratories
Division of APREL Laboratories.




Conditions
Dipole 5200-230-00802 was a re-calibration.

Ambient Temperature of the Laboratory:                  22 °C +/- 0.5°C
Temperature of the Tissue:                              21 °C +/- 0.5°C




                                                                                             2
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.


Calibration Results Summary
The following results relate the Calibrated Dipole and should be used as a quick
reference for the user.

Mechanical Dimensions

Length:                 23.6 mm
Height:                 14.0 mm

Electrical Specification

     S11 R/L                  -23.67 dB
      SWR                       1.14 U
    Impedance                  57.17 Ω

System Validation Results

 Frequency 1 Gram             10 Gram        Peak
 5200 MHz   67.35              22.23        199.16




                                                                                             3
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.


Introduction
This Calibration Report has been produced in line with the SSI Dipole Calibration
Procedure SSI-TP-018-ALSAS. The results contained within this report are for
Validation Dipole 235-00801. The calibration routine consisted of a three-step
process. Step 1 was a mechanical verification of the dipole to ensure that it meets
the mechanical specifications. Step 2 was an Electrical Calibration for the Validation
Dipole, where the SWR, Impedance, and the Return loss were assessed. Step 3
involved a System Validation using the ALSAS-10U, along with APREL E-030 130
MHz to 26 GHz E-Field Probe Serial Number 215.

References
SSI-TP-018-ALSAS Dipole Calibration Procedure
SSI-TP-016 Tissue Calibration Procedure
IEEE 1528 “Recommended Practice for Determining the Peak Spatial-Average
Specific Absorption Rate (SAR) in the Human Body Due to Wireless
Communications Devices: Experimental Techniques”
IEC-62209 “Human exposure to radio frequency fields from hand-held and body-
mounted wireless communication devices – Human models, instrumentation, and
procedures”
Part 1: “Procedure to determine the Specific Absorption Rate (SAR) for hand-held
devices used in close proximity of the ear (frequency range of 300 MHz to 3 GHz)”
IEC-62209 “Human exposure to radio frequency fields from hand-held and body-
mounted wireless communication devices – Human models, instrumentation, and
procedures”
Part 2 Draft: “Procedure to determine the Specific Absorption Rate (SAR) for hand-
held devices used in close proximity of the ear (frequency range of 30 MHz to 6
GHz)”

Conditions
Ambient Temperature of the Laboratory:                  22 °C +/- 0.5°C
Temperature of the Tissue:                              20 °C +/- 0.5°C

This was a recalibration.
Dipole Calibration uncertainty
The calibration uncertainty for the dipole is made up of various parameters presented below.

Mechanical                        1%
Positioning Error                 1.22%
Electrical                        1.7%
Tissue                             2.2%
Dipole Validation                 2.2%
TOTAL                              8.32% (16.64% K=2)

                                                                                             4
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Dipole Calibration Results
Mechanical Verification

   APREL               APREL             Measured            Measured
    Length              Height            Length              Height
   23.6 mm             14.0 mm           23.4 mm             15.4 mm


Tissue Validation


 Body Tissue 5200 MHz             Measured

 Dielectric constant, εr            47.16
 Conductivity, σ [S/m]               5.14




                                                                                             5
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Electrical Calibration

       Test                     Result
     S11 R/L                  -23.67 dB
      SWR                       1.14 U
    Impedance                  57.17 Ω

The Following Graphs are the results as displayed on the Vector Network Analyzer.

S11 Parameter Return Loss




                                                                                             6
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.



SWR




                                                                                             7
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Smith Chart Dipole Impedance




                                                                                             8
This page has been reviewed for content and attested to by signature within this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Test Equipment
The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:\NCL\Calibration Equipment\Instrument List May 2011.




                                                                                             9
This page has been reviewed for content and attested to by signature within this document.



Document Created: 2012-01-26 14:32:08
Document Modified: 2012-01-26 14:32:08

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC