Appendix E Probe 1900 head

FCC ID: MAU027

RF Exposure Info

Download: PDF
FCCID_946550

            NCL CALIBRATION LABORATORIES


                        Calibration File No.: CP-761

                                  Client.: Wistron



C E R T I F I C A T E OF C A L I B R A T I O N

      It is certified that the equipment identified below has been calibrated in the
   NCL CALIBRATION LABORATORIES by qualified personnel following recognized
          procedures and using transfer standards traceable to NRC/NIST.



                 Equipment: Miniature Isotropic RF Probe 1900 MHz


                          Manufacturer: APREL Laboratories
                                   Model No.: E-020
                                     Serial No.: 266




               Calibration Procedure: SSI/DRB-TP-D01-032-E020-V2
                   Project No: QTKB-WISB-ALS-E020-16CAL-5290


                                             th
                               Calibrated: 9 July 2007
                                             th
                              Released on: 13 July 2007




                NCL CALIBRATION LABORATORIES
                   51 SPECTRUM WAY                 Division of APREL Lab.
                   NEPEAN, ONTARIO                   TEL: (613) 820-4988
                   CANADA K2R 1E6                   FAX: (613) 820-4161


NCL Calibration Laboratories
Division of APREL Laboratories.




Introduction
This Calibration Report reproduces the results of the calibration performed in line
with the SSI/DRB-TP-D01-032-E020-V2 E-Field Probe Calibration Procedure. The
results contained within this report are for APREL E-Field Probe E-020 266.

References
SSI/DRB-TP-D01-032-E020-V2 E-Field Probe Calibration Procedure
IEEE 1528 “Recommended Practice for Determining the Peak Spatial-Average
Specific Absorption Rate (SAR) in the Human Body Due to Wireless
Communications Devices: Experimental Techniques”
SSI-TP-011 Tissue Calibration Procedure

Conditions
Probe 266 was a new probe taken from stock prior to calibration.

Ambient Temperature of the Laboratory:                        22 °C +/- 0.5°C
Temperature of the Tissue:                                    21 °C +/- 0.5°C




Page 2 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Calibration Results Summary

Probe Type:                                                E-Field Probe E-020

Serial Number:                                              266

Frequency:                                                  1900 MHz

Sensor Offset:                                              1.56 mm

Sensor Length:                                              2.5 mm

Tip Enclosure:                                              Ertalyte*

Tip Diameter:                                               <5 mm

Tip Length:                                                 60 mm

Total Length:                                               290 mm

*Resistive to recommended tissue recipes per IEEE-1528

Sensitivity in Air
Channel X:                                                  1.2 µV/(V/m)2
Channel Y:                                                  1.2 µV/(V/m)2
Channel Z:                                                  1.2 µV/(V/m)2

Diode Compression Point:                                    95 mV




Page 3 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Sensitivity in Head Tissue

Frequency:                                                  1900 MHz

Epsilon:            40.0 (+/-5%)                            Sigma:              1.40 S/m (+/-5%)


ConvF

Channel X:          5.5

Channel Y:          5.5

Channel Z:          5.5

Tissue sensitivity values were calculated using the load impedance of the APREL
Laboratories Daq-Paq.


Boundary Effect:
Uncertainty resulting from the boundary effect is less than 2% for the distance
between the tip of the probe and the tissue boundary, when less than 2.44mm.


Spatial Resolution:
The measured probe tip diameter is 5 mm (+/- 0.01 mm) and therefore meets the
requirements of SSI/DRB-TP-D01-032 for spatial resolution.




Page 4 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Receiving Pattern 1900 MHz (Air)



                                      Receiving Pattern Probe s/n 266 1900MHz




                                                              0
                                                350 1.2           10
                                            340                         20
                                      330                                     30
                                                         1
                                320                                                 40
                            310                         0.8                              50
                          300                           0.6                                   60

                      290                               0.4                                    70

                     280                                0.2                                        80

                    270                                  0                                          90

                     260                                                                           100

                      250                                                                      110

                          240                                                                 120
                            230                                                          130
                                220                                                 140
                                      210                                     150
                                            200                         160
                                                  190             170
                                                          180


                                            Ch1         Ch2       Ch3          Tot




Page 5 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Isotropy Error 1900 MHz (Air)

                                                          Isotropy Error
                                                      Probe s/n 266 1900MHz


                         10
                          9
                          8
                          7
                          6
                          5
                          4
                          3
           Error [dB]




                          2
                          1
                          0
                         -1
                         -2
                         -3
                         -4
                         -5
                         -6
                         -7
                         -8
                         -9
                        -10
                                0   100                        200                       300




                                                Isotropy Plot Probe s/n 266 1900MHz

                   0.8


                   0.7


                   0.6
   Voltage (mV)




                   0.5


                   0.4


                   0.3


                   0.2


                   0.1


                        0
                            0             100                              200                   300


                                                             Degrees (x10)

                                                             Ch1     Ch2         Ch3




Isotropicity in Tissue:                                                                0.10 dB


Page 6 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Dynamic Range

                                                      Probe 266 Dynamic Range


                      10000




                       1000




                        100
   Voltage [microV]




                                                                                                                         Uncompensated
                         10
                                                                                                                         Compensated



                          1




                        0.1




                       0.01
                          1.00   10.00   100.00   1,000.00       10,000.00   100,000.00   1,000,000.00   10,000,000.00
                                                  Electric Field [(V/m)^2]




Page 7 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Video Bandwidth


                                     Probe Frequency Characteristics

      1


      0


      -1


      -2


      -3


      -4


      -5


      -6
           1                    10                     100                     1000               10000

                                                                                      Frequency (Hz)




Video Bandwidth at 500 Hz                            1 dB
Video Bandwidth at 1.02 KHz:                         3 dB




Page 8 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.


Conversion Factor Uncertainty Assessment

Frequency:                                                  1900MHz

Epsilon:            40.0 (+/-5%)                            Sigma:              1.40 S/m (+/-5%)

ConvF

Channel X:          5.5                                     7%(K=2)

Channel Y:          5.5                                     7%(K=2)

Channel Z:          5.5                                     7%(K=2)

To minimize the uncertainty calculation all tissue sensitivity values were calculated
using a load impedance of 5 MΩ.

Boundary Effect:

For a distance of 2.4mm the evaluated uncertainty (increase in the probe sensitivity)
is less than 2%.




Page 9 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.


Test Equipment
The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:\NCL\Calibration Equipment\Instrument List May 2004.




Page 10 of 10
This page has been reviewed for content and attested to on Page 2 of this document.



Document Created: 2007-07-11 15:39:16
Document Modified: 2007-07-11 15:39:16

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC