SAR Report Appendix E Probe 5800 Body

FCC ID: MAU023

RF Exposure Info

Download: PDF
FCCID_785892

                     NCL CALIBRATION LABORATORIES


                                 Calibration File No.: CP—703

                                          Client.: Wistron



CERTIFICATE OF CALIBRATIO N

              it is certified that the equipment identified below has been calibrated in the
         NCL CALIBRATION LABORATORIES by qualified personnel following recognized
                   procedures and using transfer standards traceable to NRC/NIST.



                          Equipment: Miniature Isotropic RF Probe 5800 MHz
                                           BODY Calibration
                                  Manufacturer: APREL Laboratories
                                           Model No.: E—020
                                              Serial No.: 266


                        Calibration Procedure: SSYDRB—TP—D01—032—E020—V2
                                Project No: QTKB—ALS—E020—CAL—5238



                                      Calibrated: 20" June 2006
                                     Released on: 22"" June 2006
                      APREL Laboratories Certifed Under Laboratory 48 of SCC

This Calibration Certificate i                       cbou|panied with the Calibration Results Summary




         Released By:



                         NQL CALIBRATION LABORATORIES
                            51 SPECTRUM WAY                Division of APREL Lab.
                            NEPEAN, ONTARIO                 TEL: (613) 820—4088
                            CANADA K2R 126                  FAX: (613) 820—4161


NCL Calibration Laboratories
Division of APREL Laboratories.



Introduction

This Calibration Report reproduces the results of the calibration performed in line
with the SSIU/DRB—TP—D01—032—E020—V2 E—Field Probe Calibration Procedure. The
results contained within this report are for APREL E—Field Probe E—020 266.

References

SSVDRB—TP—D01—032—E020—V2 E—Field Probe Calibration Procedure
IEEE 1528 "Recommended Practice for Determining the Peak Spatial—Average
Specific Absorption Rate (SAR) in the Human Body Due to Wireless
Communications Devices: Experimental Techniques"
SSI—TP—011 Tissue Calibration Procedure
IEC 62209 "Human exposure to radio frequency fields from hand—held and body—
mounted wireless communication devices — Human models, instrumentation, and
procedures —Part 1 & 2: Procedure to determine the Specific Absorption Rate (SAR)
for hand—held devices used in close proximity of the ear (frequency range of 300
MHz to 3 GHz)"
IEEE 1309 Draft Standard for Calibration of Electromagnetic Field
Sensors and Probes, Excluding Antennas, from 9kHz to 40GHz


Conditions

Probe 266 was a re—calibration.

Ambient Temperature of the Laboratory:                       22 °C +/— 0.5°C
Temperature of the Tissue:                                   21 °C +/— 0.5°C

We the      undersigned Attest that to the best of our knowledge the calibration of
                                  qurately conducted and that all information contained
                                   n reviewed for accuracy.




Stuart Nicol
              V
                   N                 V

       faznPan.
Jes{e Hones




Page 2 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Calibration Results Summary


Probe Type:                                                 E—Field Probe E—020

Serial Number:                                              266

Frequency:                                                  5800 MHz

Sensor Offset:                                              1.56 mm

Sensor Length:                                              2.5 mm

Tip Enclosure:                                              Ertalyte*

Tip Diameter:                                               <5 mm

Tip Length:                                                 60 mm

Total Length:                                               290 mm

*Resistive to recommended tissue recipes per IEEE—1528

Sensitivity in Air

Channel X:                                                  1.2 uV/(V/im)
Channel Y:                                                  1.2 uV/(V/im)
Channel Z:                                                  1.2 uV/(V/im)
Diode Compression Point:                                    95 mV




Page 3 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Sensitivity in Body Tissue

Frequency:                                                  5800 MHz

Epsilon:            48.2 (+/—10%)                           Sigma:             6.0 S/im (+/—10%)


ConvF

Channel X:          4.2

Channel Y:          4.2

Channel Z:          4.2

Tissue sensitivity values were calculated using the load impedance of the APREL
Laboratories Dag—Paq.


Boundary Effect:

Uncertainty resulting from the boundary effect is less than 2% for the distance
between the tip of the probe and the tissue boundary, when less than 2.44mm.


Spatial Resolution:

The measured probe tip diameter is 5 mm (+/— 0.01 mm) and therefore meets the
requirements of SSI/DRB—TP—D01—032 for spatial resolution.




Page 4 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laborions


Receiving Pattern 5800 MHz (Air)


                                 RecebingPattorn Probe sn 20e stoouits




                                —4—om —m—Ge +Cs —+—Tot




Pare sot 10
"This page has been reviewed for contentand atested o on Page 2 ofhisdocument


NCL Calibration Laboratories
Division of APREL Labortonis


Isotropy Error 5800 MHz (Air)

                                              isctrapy tner
                                          Probe in zse seooutne

          ce
          ti                                                          se
                                Z                                 2
           {
    &$Gi
       3$
    IE a


        3i                                                            se
         q;                   e                                       ssm
                                                                       e
          j
             s               m                    zo                  e


|
                                    TeotropyPotProve snsesecoure

    g.X                       [                                       >
    }_/                                 A
    $*               4                                                L
      $                           A             A                                X
                                                 Degrees i0)




Isotropicity in Tissue:                                            0.10 48

Fage Gof 10
"This page has beenreviewed for contentand atestd o on Page 2 of this document


NCL Calibration Laboratories
Division of APREL Laboraones


Dynamic Range

                                        Probe 288 Dynanic Range
     wae




      on
        in        en        wa       uns      ome       mm wmun emem
                                     trvermsoimen




Pare Tof 10
"This page has been reviewed for contentand atestd o on Page 2 ofthis docoment


NCL Calibration Laboratories
Division of APREL Laboratories.




Video Bandwidth


                                          Probe Frequency Characteristics
  Probe output level (dB)




                            —6
                                 1   10                  100                  1000               10000

                                                                                     Frequency (Hz)




Video Bandwidth at 500 Hz                              1 dB
Video Bandwidth at 1.02 KHz:                           3 dB




Page 8 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Conversion Factor Uncertainty Assessment


Frequency:                                                  5800MHz

Epsilon:            48.2 (+/—10%)                           Sigma:              6.0 S/m (+/—10%)

ConvF

Channel X:          4.2                                     7%(K=2)

Channel Y:          4.2                                     T%(K=2)

Channel Z:          4.2                                     T%(K=2)

To minimize the uncertainty calculation all tissue sensitivity values were calculated
using a load impedance of 5 MQ.

Boundary Effect:

For a distance of 2.4mm the evaluated uncertainty (increase in the probe sensitivity)
is less than 2%.




Page 9 of 10
This page has been reviewed for content and aftested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Test Equipment
The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:ANCL\Calibration Equipment\Instrument List May 2004.




Page 10 of 10
This page has been reviewed for content and attested to on Page 2 of this document.