                     NCL CALIBRATION LABORATORIES


                                  Calibration File No.: CP—700

                                          Client.: Wistron



CERTIFICATE OF CALIBRATIO N

              It is certified that the equipment identified below has been calibrated in the
         NCL CALIBRATION LABORATORIES by qualified personnel following recognized
                   procedures and using transfer standards traceable to NRC/NIST.



                          Equipment: Miniature Isotropic RF Probe 5200 MHz


                                    Manufacturer: APREL Laboratories
                                           Model No.: E—020
                                             Serial No.: 266

                        Calibration Procedure: SSVDRB—TP—D01—032—E020—V2
                                  Project No: QTKB—ALS—E020—CAL—5238



                                       Calibrated: 20" June 2006
                                      Released on: 22"" June 2006

                      APREL Laboratories Certified Under Laboratory 48 of SCC

This Calibration Certificate is                    Acgompanied with the Calibration Results Summary




         Released By:




                           51 SPECTRUM WAY                 Division of APREL Lab.
                           NEPEAN, ONTARIO                   TEL: (613) 820—4088
                           CANADA K2R 126                    FAX: (619) 820—4161


NCL Calibration Laboratories
Division of APREL Laboratories.



Introduction

This Calibration Report reproduces the results of the calibration performed in line
with the SSIU/DRB—TP—D01—032—E020—V2 E—Field Probe Calibration Procedure. The
results contained within this report are for APREL E—Field Probe E—020 266.

References

SSVYDRB—TP—D01—032—E020—V2 E—Field Probe Calibration Procedure
SSI—TP—011 Tissue Calibration Procedure
IEC 62209 "Human exposure to radio frequency fields from hand—held and body—
mounted wireless communication devices — Human models, instrumentation, and
procedures —Part 1 & 2: Procedure to determine the Specific Absorption Rate (SAR)
for hand—held devices used in close proximity of the ear (frequency range of 300
MHz to 3 GHz)"
1EEE 1309 Draft Standard for Calibration of Electromagnetic Field
Sensors and Probes, Excluding Antennas, from 9kHz to 40GHz


Conditions

Probe 266 was a re—calibration.

Ambient Temperature of the Laboratory:                       22 °C +/— 0.5°C
Temperature of the Tissue:                                   21 °C +/— 0.5°C

We the undersigned attest that to the best of our knowledge the calibration of
this probe  has been a ii rately conducted and that all information contained
within thisg   report has             reyiewed for accuracy.




Stuart Nicol   $ "T                    fl



      JamePors.
J#e Hones




Page 2 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Calibration Results Summary


Probe Type:                                                 E—Field Probe E—020

Serial Number:                                              266

Frequency:                                                  5200 MHz

Sensor Offset:                                              1.56 mm

Sensor Length:                                              2.5 mm

Tip Enclosure:                                              Ertalyte*

Tip Diameter:                                               <5 mm

Tip Length:                                                 60 mm

Total Length:                                               290 mm

*Resistive to recommended tissue recipes per IEEE—1528

Sensitivity in Air

Channel X:                                                  1.2 uV/(V/im)®
Channel Y:                                                  1.2 uV/(V/im)®
Channel Z:                                                  1.2 uV/i(V/im)y
Diode Compression Point:                                    95 mV




Page 3 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Sensitivity in Head Tissue

Frequency:                                                 5200 MHz

Epsilon:            36.0 (+/—5%)                           Sigma:              4.65 S/m (+/—10%)


ConvF

Channel X:          3.82

Channel Y:          3.82

Channel Z:          3.82

Tissue sensitivity values were calculated using the load impedance of the APREL
Laboratories Daq—Paq.


Boundary Effect:

Uncertainty resulting from the boundary effect is less than 2% for the distance
between the tip of the probe and the tissue boundary, when less than 2.44mm.


Spatial Resolution:

The measured probe tip diameter is 5 mm (+/— 0.01 mm) and therefore meets the
requirements of SSI/DRB—TP—D01—032 for spatial resolution.




Page 4 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Libortoris


Receiving Pattern 5200 MHz (Air)


                                  Recebing Patiern Probe sn 200 szooute




                                 ~—4—on ~#—ore _ Grd —+—Tot




Fage s of 10
"This page has been reviewed for content and atesid o on Page 2 ofthis document


NCL Calibration Laboratories
Divisio of APREL Labortoris


Isotropy Error 5200 MHz (Air)

                                             isotropy Eor
                                         Probe in ase saooutte
   Eivo t
              Buessoroos




                                   Incropy o Prote sn 2 sa00uts
 votige (it




Isotropicity Tissue:                                      0.10 48

Fage oof 10
"This page has been reviewed for contentand atestd o on Page 2 ofthisdocument


NCL Calibration Laboratories
Divisin of APREL Laboratoris


Dynamic Range

                                      Prove 2 Oynanic Range




                                                                            ib




        s        sn       we       uma     men       mm immn mmms
                                   brosrasemen




Fage? of 10
"This paghas beenreviewed forcontent and atested to on Page 2 ohidocument


NCL Calibration Laboratories
Division of APREL Laboratories.




Video Bandwidth


                                      Probe Frequency Characteristics


   S                                                                          |
   E;                             |                             |                 |                        .
   @    1                                              1                          1




   s3 1                          ||                             Nlll
                                                                1
    $
   &2                                                              \
       _3                                                                                                 HH




       —4                                                                                 \


       _5                                                                                                  L




       —6                        +                         +                          f
            1                   10                     100                        1000                    10000
                                                                                              Frequency (Hz)




Video Bandwidth at 500 Hz                            1 dB
Video Bandwidth at 1.02 KHz:                         3 dB




Page 8 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Conversion Factor Uncertainty Assessment


Sensitivity in Head Tissue

Frequency:                                                 5200 MHz

Epsilon:           36.0 (+/—5%)                            Sigma:              4.65 S/m (+/—10%)

ConvF

Channel X:         3.82                                    7%(K=2)
Channel Y:         3.82                                    7%(K=2)

Channel Z:         3.82                                    7%(K=2)

To minimize the uncertainty calculation all tissue sensitivity values were calculated
using a load impedance of 5 MC.

Boundary Effect:

For a distance of 2.5mm the evaluated uncertainty (increase in the probe sensitivity)
is less than 2%.




Page 9 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Test Equipment
The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:ANCL\Calibration Equipment\Instrument List May 2006.




Page 10 of 10