SAR Report Appendix E Probe 5200 Body

FCC ID: MAU023

RF Exposure Info

Download: PDF
FCCID_785890

                     NCL CALIBRATION LABORATORIES


                                  Calibration File No.: CP—701

                                          Client.: Wistron




CERTIFICATE OF CALIBRATION

              It is certified that the equipmentidentified below has been calibrated in the
         NCL CALIBRATION LABORATORIES by qualified personnel following recognized
                   procedures and using transfer standards traceable to NRC/NIST.



                          Equipment: Miniature Isotropic RF Probe 5200 MHz
                                             BODY Calibration
                                   Manufacturer: APREL Laboratories
                                           Model No.: E—020
                                              Serial No.: 266


                        Calibration Procedure: SSUDRB—TP—D01—032—E020—V2
                                Project No: QTKB—ALS—E020—CAL—5238



                                      Calibrated: 20" June 2006
                                     Released on: 22"" June 2006
                      APREL Laboratories Certified Under Laboratory 48 of SCC

This Calibration Certificate is                    Accompanied with the Calibration Results Summary




         Released By:



                                          1        TT
                           51 SPECTRUM WaAY                Division of APREL Lab.
                            NEPEAN, ONTARIO                  TEL: (613) 820—4088
                            CANADA K2R 126                      FAX:(613) 820—4161


NCL Calibration Laboratories
Division of APREL Laboratories.



Introduction

This Calibration Report reproduces the results of the calibration performed in line
with the SSIU/DRB—TP—D01—032—E020—V2 E—Field Probe Calibration Procedure. The
results contained within this report are for APREL E—Field Probe E—020 266.

References

SSUVDRB—TP—D01—032—E020—V2 E—Field Probe Calibration Procedure
IEEE 1528 "Recommended Practice for Determining the Peak Spatial—Average
Specific Absorption Rate (SAR) in the Human Body Due to Wireless
Communications Devices: Experimental Techniques"
SSI—TP—011 Tissue Calibration Procedure
IEC 62209 "Human exposure to radio frequency fields from hand—held and body—
mounted wireless communication devices — Human models, instrumentation, and
procedures —Part 1 & 2: Procedure to determine the Specific Absorption Rate (SAR)
for hand—held devices used in close proximity of the ear (frequency range of 300
MHz to 3 GHz)"
IEEE 1309 Draft Standard for Calibration of Electromagnetic Field
Sensors and Probes, Excluding Antennas, from 9kHz to 40GHz


Conditions

Probe 266 was a re—calibration.

Ambient Temperature of the Laboratory:                        22 °C +/— 0.5°C
Temperature of the Tissue:                                    21 °C +/— 0.5°C

We the undersigned attest that to the best of our knowledge the calibration of
this probe has beemaccurately conducted and that all information contained
within this|report has \been reviewed for accuracy.




Page 2 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Calibration Results Summary


Probe Type:                                                E—Field Probe E—020

Serial Number:                                             266

Frequency:                                                 5200 MHz

Sensor Offset:                                             1.56 mm

Sensor Length:                                             2.5 mm

Tip Enclosure:                                             Ertalyte*

Tip Diameter:                                              <5 mm

Tip Length:                                                60 mm

Total Length:                                              290 mm

*Resistive to recommended tissue recipes per IEEE—1528

Sensitivity in Air

Channel X:                                                 1.2 uV/(V/im)®
Channel Y:                                                 1.2 uV/(V/im)
Channel Z:                                                 1.2 uV/(V/im)®

Diode Compression Point:                                   95 mV




Page 3 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.




Sensitivity in Body Tissue

Frequency:                                                 5200 MHz

Epsilon:           43.0 (+/—10%)                           Sigma:              5.75 S/m (+/—10%)


ConvF

Channel X:         4.4

Channel Y:          4.4

Channel Z:          4.4

Tissue sensitivity values were calculated using the load impedance of the APREL
Laboratories Daq—Paq.


Boundary Effect:

Uncertainty resulting from the boundary effect is less than 2% for the distance
between the tip of the probe and the tissue boundary, when less than 2.44mm.


Spatial Resolution:

The measured probe tip diameter is 5 mm (+/— 0.01 mm) and therefore meets the
requirements of SSI/DRB—TP—D01—032 for spatial resolution.




Page 4 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Divisio of APREL Laborionis


Receiving Pattern 5200 MHz (Air)



                                 Recebing Patorn Probe sn 20e saooutte




                                —+—om ~#—ore _> ons —+—Tat




Fage Sof 10
"This page has been reviewed fr cntent and atested o on Page2 ofthis document


NCL Calibration Laboratories
Division of APREL Laborstones


Isotropy Error 5200 MHz (Air)

                                             novony Enor
                                         Probe in see seocute


         f
    a iij        s2
         | rrr
    i4i             :       >
                            >
      $3       z         5>
       i4
        ii
     4
           o 100      xo      xo




                                   Incropy is Prote sn se sacoumc




                                                Dogrees (td)


Isotropicity in Tissue:                                            0.10 48

Fage cof 10
"This page has beenreviewed for content and atested o on Page 2 ofhis document


NCL Calibration Laboratories
Division of APREL Libortors


Dynal             Range

                                        Probe 2Dynanic Range
             um
Youre pwon




                                                                                 —e—crevoms




                                     tevnesnien




PareTot 10
"This page has been reviewed for content and atested o on Page 2 ofhisdocument


NCL Calibration Laboratories
Division of APREL Laboratories.




Video Bandwidth


                                      Probe Frequency Characteristics
   Probe output level (dB)




                             1   10                     100                    1000               10000

                                                                                      Frequency (Hz)




Video Bandwidth at 500 Hz                            1 dB
Video Bandwidth at 1.02 KHz:                         3 dB




Page 8 of 10
This page has been reviewed for content and attested to on Page 2 of this document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Conversion Factor Uncertainty Assessment

Frequency:                                                 5200MHz

Epsilon:            43.0 (+/—10%)                          Sigma:              5.75 S/m (+/10%)
ConvF

Channel X:          4.4                                    7%(K=2)

Channel Y:          4.4                                    7%(K=2)
Channel Z:          4.4                                    T%(K=2)

To minimize the uncertainty calculation all tissue sensitivity values were calculated
using a load impedance of 5 MGQ.

Boundary Effect:

For a distance of 2.4mm the evaluated uncertainty (increase in the probe sensitivity)
is less than 2%.




Page 9 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.


NCL Calibration Laboratories
Division of APREL Laboratories.



Test Equipment
The test equipment used during Probe Calibration, manufacturer, model number
and, current calibration status are listed and located on the main APREL server
R:ANCL\Calibration Equipment\Instrument List May 2004.




Page 10 of 10
This page has been reviewed for content and attested to on Page 2 ofthis document.