DCD Test Report

FCC ID: LF537602

Test Report

Download: PDF
FCCID_3864553

ee




RF Test
Reports

 Report No.: FCC_RF_SL17121901-MED-023 (37602)
 Supersede Report No.: N/A

 Applicant            Medtronic, Inc.
 Product Name         Itrel 4
 Model No.            37602
 Test Standard        47 CFR 15.209
 Test Method          ANSI C63.10: 2013
 Date of test         02/27/2018
 Issue Date           02/28/2018
 Test Result          ☒ Pass              ☐ Fail
 Equipment complied with the test standard equipment               [x]
 Equipment did not comply with the test standard equipment         [ ]

 This Test Report is Issued Under the Authority of:




                             Cipher                                                        Chen Ge
                          Test Engineer                                               Engineer Reviewer
                                       This test report may be reproduced in full only
                       Test result presented in this test report is applicable to the tested sample only




                                                    Issued By:
                                               SIEMIC Laboratories
                                    775 Montague Expressway, Milpitas, CA 95035




         775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                               Visit us at: www.siemic.com; Follow us at:


                                                                       Test report No.     FCC_RF_SL17121901-MED-023 (37602)
                                                                       Page                2 of 17




                                         Laboratory Introduction
SIEMIC, headquartered in the heart of Silicon Valley, with superior facilities in US and Asia, is one of the leading independent testing
and certification facilities providing customers with one-stop shop services for Compliance Testing and Global Certifications.




In addition to testing and certification, SIEMIC provides initial design reviews and compliance management throughout a project. Our
extensive experience with China, Asia Pacific, North America, European, and International compliance requirements, assures the
fastest, most cost effective way to attain regulatory compliance for the global markets.

                                   Accreditations for Conformity Assessment
           Country/Region                  Accreditation Body                                   Scope
                 USA                            FCC, A2LA                             EMC, RF/Wireless, Telecom
              Canada                         IC, A2LA, NIST                           EMC, RF/Wireless, Telecom
               Taiwan                       BSMI, NCC, NIST                            EMC, RF, Telecom, Safety
             Hong Kong                         OFTA, NIST                                RF/Wireless, Telecom
              Australia                        NATA, NIST                              EMC, RF, Telecom, Safety
               Korea                         KCC/RRA, NIST                           EMI, EMS, RF, Telecom, Safety
               Japan                     VCCI, JATE, TELEC, RFT                        EMI, RF/Wireless, Telecom
               Mexico                    NOM, COFETEL, Caniety                      EMC, RF/Wireless, Telecom, Safety
              Europe                           A2LA, NIST                              EMC, RF, Telecom, Safety
                Israel                          MOC, NIST                              EMC, RF, Telecom, Safety


                                    Accreditations for Product Certifications
               Country                      Accreditation Body                                   Scope
                 USA                          FCC TCB, NIST                                 EMC, RF, Telecom
               Canada                          IC FCB, NIST                                 EMC, RF, Telecom
              Singapore                          iDA, NIST                                  EMC, RF, Telecom
                 EU                                 NB                                    EMC & R&TTE Directive
                Japan                         MIC (RCB 208)                                   RF, Telecom
              Hong Kong                       OFTA (US002)                                    RF, Telecom




          775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                Visit us at: www.siemic.com; Follow us at:


                                                                                                Test report No.             FCC_RF_SL17121901-MED-023 (37602)
                                                                                                Page                        3 of 17



CONTENTS
1     REPORT REVISION HISTORY ....................................................................................................................................... 4
2     EXECUTIVE SUMMARY................................................................................................................................................. 5
3     CUSTOMER INFORMATION .......................................................................................................................................... 5
4     TEST SITE INFORMATION ............................................................................................................................................ 5
5     EQUIPMENT UNDER TEST (EUT) INFORMATION .......................................................................................................... 5
    5.1    EUT Description ........................................................................................................................................................ 5
    5.2    Radio Description ...................................................................................................................................................... 6
6     SUPPORTING EQUIPMENT/SOFTWARE AND CABLING DESCRIPTION......................................................................... 7
    6.1    Supporting Equipment ............................................................................................................................................... 7
    6.2    I/O Ports ................................................................................................................................................................... 7
    6.3    Test Software Description .......................................................................................................................................... 7
7     TEST SUMMARY ........................................................................................................................................................... 8
8     MEASUREMENT UNCERTAINTY ................................................................................................................................... 9
    8.1    Radiated Emissions (9kHz to 30MHz).......................................................................................................................... 9
    8.2    Radiated Emissions (30MHz to 1GHz)......................................................................................................................... 9
    8.3    Radiated Emissions (1GHz to 40GHz) ....................................................................................................................... 10
    8.4    RF conducted measurement .................................................................................................................................... 10
9     MEASUREMENTS, EXAMINATION AND DERIVED RESULTS ....................................................................................... 11
    9.1    Antenna Requirement .............................................................................................................................................. 11
    9.2    Radiated Spurious Emissions below 30MHz .............................................................................................................. 12
ANNEX A. TEST INSTRUMENT & METHOD ........................................................................................................................ 15
ANNEX B. SIEMIC ACCREDITATION ................................................................................................................................... 16




            775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                          Visit us at: www.siemic.com; Follow us at:


                                                               Test report No.   FCC_RF_SL17121901-MED-023 (37602)
                                                               Page              4 of 17




1   Report Revision History

                Report No.                             Report Version              Description              Issue Date
     FCC_RF_SL17121901-MED-023 (37602)                     None                   Original Report           02/28/2018




      775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                          Visit us at: www.siemic.com; Follow us at:


                                                                    Test report No.   FCC_RF_SL17121901-MED-023 (37602)
                                                                    Page              5 of 17




2      Executive Summary

The purpose of this test program was to demonstrate compliance of following product
 Company:          Medtronic, Inc.
 Product:          Itrel 4
 Model:            37602

against the current Stipulated Standards. The specified model product stated above has demonstrated compliance with
the Stipulated Standard listed on 1st page.

3      Customer information

    Applicant Name                          Medtronic, Inc.
    Applicant Address                       710 Medtronic Parkway N.E., Minneapolis, MN 55432
    Manufacturer Name                       Medtronic, Inc.
    Manufacturer Address                    710 Medtronic Parkway N.E., Minneapolis, MN 55432


4      Test site information

    Lab performing tests                    SIEMIC Laboratories
    Lab Address                             775 Montague Expressway, Milpitas, CA 95035
    FCC Test Site No.                       881796
    IC Test Site No.                        4842D-2
    VCCI Test Site No.                      A0133




5      Equipment Under Test (EUT) Information

5.1      EUT Description

 Product Name                              Itrel 4
 Model No.                                 37602
 Trade Name                                Medtronic
 Serial No.                                NMX700557H
 Input Power                               Battery Power 3.0V
 Date of EUT received                      12/21/2017
 Equipment Class/ Category                 ULP-AMI-P, Class 3




           775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                               Visit us at: www.siemic.com; Follow us at:


                                                                Test report No.   FCC_RF_SL17121901-MED-023 (37602)
                                                                Page              6 of 17



5.2   Radio Description

Radio Parameters
        Operating Frequency             175 KHz
            Modulation                  OOK burst
          Channel Spacing               Single Channel Operation
           Antenna Type                 Integral loop antenna/External antenna
           Antenna Gain                 None


Channel List
          Type                Channel No.            Frequency (KHz)
        ULP-AMI                   1                        175




       775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                           Visit us at: www.siemic.com; Follow us at:


                                                                   Test report No.     FCC_RF_SL17121901-MED-023 (37602)
                                                                   Page                7 of 17




6       Supporting Equipment/Software and cabling Description

6.1      Supporting Equipment

             Supporting Equipment                                  Serial
 Item                                           Model                                Manufacturer                 Note
                 Description                                      Number
    1            InterStim ICon                    3037             N/A               Medtronic                   N/A



6.2      I/O Ports

               Connection Start                      Connection Stop                     Length / shielding Info
Item                                                                                                                      Note
               From           I/O Port                To            I/O Port            Length (m)           Shielding
N/A             N/A               N/A                N/A               N/A                 N/A                 N/A          N/A


6.3      Test Software Description

Test Item                               Software                                                Description
                                                                               Provided by manufacturer to set EUT in continuous
RF Testing                                N/A
                                                                                                    mode




          775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                              Visit us at: www.siemic.com; Follow us at:


                                                                              Test report No.        FCC_RF_SL17121901-MED-023 (37602)
                                                                              Page                   8 of 17




7      Test Summary

              Test Item                              Test standard                              Test Method/Procedure                    Pass / Fail
                                                                                                                                           ☒ Pass
        Antenna Requirement             FCC               47 CFR 15.203                FCC             ANSI C63.10 – 2013
                                                                                                                                           ☐ N/A
                                                                                                                                           ☒ Pass
     Radiated Spurious Emissions        FCC               47 CFR 15.209                FCC              ANSI C63.10: 2013
                                                                                                                                           ☐ N/A
                                                                                                                                           ☐ Pass
      AC Conducted Emissions            FCC               47 CFR 15.207                FCC              ANSI C63.10: 2013
                                                                                                                                           ☒ N/A
                     1.   All measurement uncertainties are not taken into consideration for all presented test result.
    Remark
                     2.   The applicant shall ensure frequency stability by showing that an emission is maintained within the band of operation under
                          all normal operating conditions as specified in the user’s manual.
                     3.   EUT is a DC powered device.




             775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                   Visit us at: www.siemic.com; Follow us at:


                                                                      Test report No.   FCC_RF_SL17121901-MED-023 (37602)
                                                                      Page              9 of 17




8      Measurement Uncertainty
 8.1     Radiated Emissions (9kHz to 30MHz)

         The test is to measure the radiated emissions of the EUT.

         Some error sources that can contribute to the total uncertainty:
         -   Uncertainty of the receiver
         -   Uncertainty of the antenna
         -   Uncertainty of cables
         -   Uncertainty due to the mismatches

                                              Value          Probability                Sensitivity   Expanded
                Source of Uncertainty                                        Division
                                               (dB)         Distribution                Coefficient   Uncertainty
                   Receiver Reading           0.10          Rectangular        1.732        1          0.069284
                  Cable Insertion Loss        0.21            Normal             2          1            0.105
                    Antenna Factor            0.65            Normal             2          1            0.325
                 Receiver CW accuracy         0.45          Rectangular        1.732        1         0.2598152
                       Mismatch               0.25            U-Shape          1.414        1         0.1768033
              Combined Standard Uncertainty                                                              0.935
              Expanded Uncertainty (K=2)                                                                  1.87
The total derived measurement uncertainty is +/- 1.87 dB.

 8.2     Radiated Emissions (30MHz to 1GHz)

         The test is to measure the radiated emissions of the EUT.

         Some error sources that can contribute to the total uncertainty:
         -   Uncertainty of the receiver
         -   Uncertainty of the antenna
         -   Uncertainty of cables
         -   Uncertainty due to the mismatches
         -   NSA Calibration
         -   Etc., details see the below table

                                              Value          Probability                Sensitivity    Expanded
                Source of Uncertainty                                        Division
                                               (dB)         Distribution                Coefficient   Uncertainty
                   Receiver Reading           0.12          Rectangular        1.732        1           0.069284
                  Cable Insertion Loss        0.21            Normal             2          1             0.105
                  Filter Insertion Loss       0.25            Normal             2          1             0.125
                    Antenna Factor            0.65            Normal             2          1             0.325
                 Receiver CW accuracy           0.5         Rectangular        1.732        1          0.2886836
               Pulse Amplitude Response         1.5         Rectangular        1.732        1         0.86605081
                     PRF Response               1.5         Rectangular        1.732        1         0.86605081
               Mismatch Filter - Receiver     0.25            U-Shape          1.414        1          0.1768033
                    NSA Calibration             4.0           U-Shape          1.414        1          2.8288543
              Combined Standard Uncertainty                                                            3.0059131
              Expanded Uncertainty (K=2)                                                               6.0118262
The total derived measurement uncertainty is +/- 6.00 dB.




         775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                               Visit us at: www.siemic.com; Follow us at:


                                                                           Test report No.      FCC_RF_SL17121901-MED-023 (37602)
                                                                           Page                 10 of 17



 8.3     Radiated Emissions (1GHz to 40GHz)
         The test is to measure the radiated emissions of the EUT.

         Some error sources that can contribute to the total uncertainty:
         -   Uncertainty of the receiver
         -   Uncertainty of the antenna
         -   Uncertainty of cables
         -   Uncertainty due to the mismatches
         -   VSWR Calibration
         -   Etc., details see the below table

                                              Value             Probability                      Sensitivity    Expanded
                Source of Uncertainty                                                Division
                                               (dB)            Distribution                      Coefficient    Uncertainty
                Receiver Reading              0.12             Rectangular            1.732          1          0.0692840
              Cable Insertion Loss            0.21               Normal                 2            1          0.1050000
               Filter Insertion Loss          0.25               Normal                 2            1          0.1250000
                  Antenna Factor              0.65               Normal                 2            1          0.3250000
             Receiver CW accuracy               0.5            Rectangular            1.732          1          0.2886836
           Pulse Amplitude Response             1.5            Rectangular            1.732          1          0.8660508
                  PRF Response                  1.5            Rectangular            1.732          1          0.8660508
            Mismatch Filter - Receiver        0.25               U-Shape              1.414          1          0.1768033
                VSWR Calibration                2.0              U-Shape              1.414          1          1.4144272
           Combined Standard Uncertainty                                                                          4.2363
           Expanded Uncertainty (K=2)                                                                             8.4726
         The total derived measurement uncertainty is +/- 8.47 dB.

 8.4     RF conducted measurement
         The test is to measure the RF output power from the EUT.

         Some error sources that can contribute to the total uncertainty:

         -       Uncertainty of the Reference Level Uncertainty
         -       Uncertainty of variable attenuators
         -       Uncertainty of cables
         -       Uncertainty due to the mismatches

                                         Value              Probability            Division      Sensitivity   Expanded
             Source of Uncertainty        (dB)              Distribution                         Coefficient   Uncertainty
            Reference Level            0.12              Rectangular                1.732            1          0.069284
          Cable Insertion Loss         0.21                Normal                     2              1            0.105
              Attenuator               0.25                Normal                     2              1            0.125
               Mismatch                0.25               U-Shape                   1.414            1         0.1768033
         Combined Standard Uncertainty                                                                          0.476087
         Expanded Uncertainty (K=2)                                                                             0.952174

The total derived measurement uncertainty is +/- 0.95 dB.




         775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                Visit us at: www.siemic.com; Follow us at:


                                                                     Test report No.     FCC_RF_SL17121901-MED-023 (37602)
                                                                     Page                11 of 17




9     Measurements, Examination and Derived Results

9.1     Antenna Requirement

 Spec             Requirement                                                                                        Applicable
                  An intentional radiator shall be designed to ensure that no antenna other than that furnished
                  by the responsible party shall be used with the device.

                  Antenna requirement must meet at least one of the following:
      §15.203                                                                                                             ☒
                  a) Antenna must be permanently attached to the device.
                  b) The antenna must use a unique type of connector to attach to the device.
                  c) Device must be professionally installed. The installer shall be responsible for ensuring that
                  the correct antenna is employed by the device.

      Remark      The EUT has an integrated antenna which meets the requirement.

      Result      ☒ PASS             ☐ FAIL




         775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                             Visit us at: www.siemic.com; Follow us at:


                                                                         Test report No.      FCC_RF_SL17121901-MED-023 (37602)
                                                                         Page                 12 of 17


9.2      Radiated Spurious Emissions below 30MHz
Requirement(s):
                  Item       Requirement                                                                                       Applicable
                             Except higher limit as specified elsewhere in other section, the emissions from the low-
                             power radio-frequency devices shall not exceed the field strength levels specified in the
                             following table and the level of any unwanted emissions shall not exceed the level of the
                             fundamental emission. The tighter limit applies at the band edges
 FCC 15.209          1                                                                                                                ☒
                                                                                                    Measurement
                                    Frequency range (MHz)           Field Strength (uV/m)
                                                                                                  distance (meters)
                                          0.009-0.490                       2400/F(kHz)                   300
                                          0.490-1.705                      24000/F(kHz)                    30
                                           1.705-30.0                           30                         30
                  1.       The EUT was switched on and allowed to warm up to its normal operating condition.
                  2.       The test was carried out at the selected frequency points obtained from the EUT characterisation.
                           Maximization of the emissions, was carried out by rotating the EUT, changing the antenna polarization, and
                           adjusting the antenna height in the following manner:
                           a.         Vertical or horizontal polarisation (whichever gave the higher emission level over a full rotation of
 Procedure                            the EUT) was chosen.
                           b.         The EUT was then rotated to the direction that gave the maximum emission.
                           c.         Finally, the antenna height was adjusted to the height that gave the maximum emission.
                  3.       A Quasi-peak measurement was then made for that frequency point.
                  4.       Steps 2 and 3 were repeated for the next frequency point, until all selected frequency points were measured.
                                                                                                Temperature                     21°C
 Test Date        02/27/2018                               Environmental condition              Relative Humidity               42%
                                                                                                Atmospheric Pressure            1015mbar

  Remark          N/A


      Result      ☒ Pass             ☐ Fail



Test Data      ☒ Yes                      ☐N/A
Test Plot      ☒ Yes (See below)          ☐ N/A
Test was done by Gary Chou at 10m chamber.




            775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                 Visit us at: www.siemic.com; Follow us at:


                                                                      Test report No.    FCC_RF_SL17121901-MED-023 (37602)
                                                                      Page               13 of 17


Test specification:                   Radiated Spurious Emissions
Mains Power:                          3VDC
Tested by:                            Shuo Zhang                                                                  ☒ Pass
                                                                                                 Result:
                                                                                                                  ☐ Fail
Test Date:                            02/27/2018
Remarks:                              ƒ= 100kHz – 30MHz plot, and loop antenna at 0 degree




                                                      Quasi Max Measurement


Frequency       Raw     Cable       AF        Level     Measurement      Pol       Hgt     Azt        Limit     Margin       Pass
   MHz          dBuV    Loss        dB       dBuV/m        Type         (0/90)     cm      Deg       dBuV/m      dB          /Fail

   0.55        10.28     0.39      17.65     28.32      Quasi Max         0        100      61        72.85     -44.53       Pass
   1.82        17.36      0.5      8.19      26.04      Quasi Max         0        100     125        69.54     -43.5        Pass
   0.18        28.21     0.28      27.17     55.66      Quasi Max         0        100     276        99.06     -43.39       Pass
  10.43        21.96     0.66      1.57      24.19      Quasi Max         0        100     101        69.54     -45.35       Pass
   2.12         9.72     0.51      7.08      17.31      Quasi Max         0        100     203        69.54     -52.23       Pass
  10.78        15.42     0.66      1.57      17.65      Quasi Max         0        100     102        69.54     -51.89       Pass




           775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                Visit us at: www.siemic.com; Follow us at:


                                                                      Test report No.    FCC_RF_SL17121901-MED-023 (37602)
                                                                      Page               14 of 17




Test specification:                   Radiated Spurious Emissions
Mains Power:                          3VDC
Tested by:                            Shuo Zhang                                                                  ☒ Pass
                                                                                                 Result:
                                                                                                                  ☐ Fail
Test Date:                            02/27/2018
Remarks:                              ƒ= 100kHz – 30MHz plot, and loop antenna at 90 degree




                                                      Quasi Max Measurement


Frequency       Raw     Cable       AF        Level     Measurement      Pol       Hgt     Azt        Limit     Margin       Pass
   MHz          dBuV    Loss        dB       dBuV/m        Type         (0/90)     cm      Deg       dBuV/m      dB          /Fail

   0.82          8.5     0.42      14.33     23.25      Quasi Max         90       100     203        69.29     -46.04       Pass
  10.13        14.94     0.65      1.56      17.16      Quasi Max         90       100     283        69.54     -52.38       Pass
   2.20         7.24     0.51      6.83      14.58      Quasi Max         90       100     166        69.54     -54.96       Pass
  10.83         8.34     0.66      1.56      10.56      Quasi Max         90       100     105        69.54     -58.98       Pass
  16.08         0.17     0.74      1.51       2.41      Quasi Max         90       100      40        69.54     -67.13       Pass
   0.31         6.52     0.33      22.47     29.33      Quasi Max         90       100      50        90.58     -61.26       Pass




           775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                                Visit us at: www.siemic.com; Follow us at:


                                                                Test report No.   FCC_RF_SL17121901-MED-023 (37602)
                                                                Page              15 of 17




Annex A. TEST INSTRUMENT & METHOD

           Instrument                        Model                 Serial #       Cal Cycle      Cal Due       In use

SPURIOUS EMISSIONS
     Keysight Signal Analyzer               N9030B              MY57140374          1 Year      05/24/2018
          Loop Antenna                        6512                  49120           1 Year      08/20/2018
          Pre-Amplifier                      8449B              3008A00715          1 Year      02/12/2019




      775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                           Visit us at: www.siemic.com; Follow us at:


                                                                   Test report No.       FCC_RF_SL17121901-MED-023 (37602)
                                                                   Page                  16 of 17




Annex B. SIEMIC Accreditation

        Accreditations         Document                                        Scope / Remark

ISO 17025 (A2LA)                           Please see the documents for the detailed scope

ISO Guide 65 (A2LA)                        Please see the documents for the detailed scope

TCB Designation                            A1, A2, A3, A4, B1, B2, B3, B4, C

FCC DoC Accreditation                      FCC Declaration of Conformity Accreditation

FCC Site Registration                      3 meter site

FCC Site Registration                      3 meter site

IC Site Registration                       3 meter site

IC Site Registration                       3 meter site

                                           Radio & Telecommunications Terminal Equipment:
                                           EN45001 – EN ISO/IEC 17025
EU NB
                                           Electromagnetic Compatibility:
                                           EN45001 – EN ISO/IEC 17025

Singapore iDA
                                           Phase I, Phase II
CB(Certification Body)

Vietnam MIC
                                           Please see the document for the detailed scope
CAB Accreditation

                                           (Phase II) OFCA Foreign Certification Body for Radio and Telecom
Hong Kong OFCA
                                           (Phase I) Conformity Assessment Body for Radio and Telecom

                                           Radio: Scope A – All Radio Standard Specification in Category I
Industry Canada CAB
                                           Telecom: CS-03 Part I, II, V, VI, VII, VIII




         775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                             Visit us at: www.siemic.com; Follow us at:


                                                                  Test report No.    FCC_RF_SL17121901-MED-023 (37602)
                                                                  Page               17 of 17




                                           Radio: A1. Terminal equipment for purpose of calling
Japan Recognized Certification
                                           Telecom: B1. Specified radio equipment specified in Article 38-2, Paragraph 1, Item
Body Designation
                                           1 of the Radio Law

                                           EMI: KCC Notice 2008-39, RRL Notice 2008-3: CA Procedures for EMI
                                           KN22: Test Method for EMI
                                           EMS: KCC Notice 2008-38, RRL Notice 2008-4: CA Procedures for EMS
                                           KN24, KN61000-4-2, -4-3, -4-4, -4-5, -4-6, -4-8, -4-11: Test Method for EMS

Korea CAB Accreditation                    Radio: RRL Notice 2008-26, RRL Notice 2008-2, RRL Notice 2008-10,
                                           RRL Notice 2007-49, RRL Notice 2007-20, RRL Notice 2007-21, RRL Notice 2007-
                                           80, RRL Notice 2004-68

                                           Telecom: President Notice 20664, RRL Notice 2007-30, RRL Notice 2008-7 with
                                           attachments 1, 3, 5, 6; President Notice 20664, RRL Notice 2008-7 with attachment 4

                                           LP0002, PSTN01, ADSL01, ID0002, IS6100, CNS14336, PLMN07, PLMN01,
Taiwan NCC CAB Recognition
                                           PLMN08

Taiwan BSMI CAB Recognition                CNS 13438

                                           R-3083: Radiation 3 meter site
Japan VCCI                                 C-3421: Main Ports Conducted Interference Measurement
                                           T-1597: Telecommunication Ports Conducted Interference Measurement

                                           EMC: AS/NZS CISPR 11, AS/NZS CISPR 14.1, AS/NZS CISPR22, AS/NZS
                                           61000.6.3, AS/NZS 61000.6.4

                                           Radio communications: AS/NZS 4281, AS/NZS 4268, AS/NZS 4280.1,
                                           AS/NZS 4280.2, AS/NZS 4295, AS/NZS 4582, AS/NZS 4583, AS/NZS 4769.1,
Australia CAB Recognition
                                           AS/NZS 4769.2, AS/NZS 4770, AS/NZS 4771

                                           Telecommunications: AS/ACIF S002:05, AS/ACIF S003:06, AS/ACIF S004:06
                                           AS/ACIF S006:01, AS/ACIF S016:01, AS/ACIF S031:01, AS/ACIF S038:01,
                                           AS/ACIF S040:01, AS/ACIF S041:05, AS/ACIF S043.2:06, AS/ACIF S60950.1

                                           AS/ACIF S002, AS/ACIF S003, AS/ACIF S004, AS/ACIF S006, AS/ACIF
Australia NATA Recognition                 S016,AS/ACIF S031, AS/ACIF S038, AS/ACIF S040, AS/ACIF S041, AS/ACIF
                                           S043.2




        775 Montague Expressway, Milpitas, CA 95035, USA • Phone: (+1) 408 526 1188 • Facsimile (+1) 408 526 1088

                             Visit us at: www.siemic.com; Follow us at:



Document Created: 2018-05-25 10:52:48
Document Modified: 2018-05-25 10:52:48

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC