SAR Report Appendix 1

FCC ID: LDK-HS561

RF Exposure Info

Download: PDF
FCCID_4019881

PROBE CALIBRATION CERTIFICATES


           esw© .( Collsboration vith
           @=‘7"7‘J,   a
           Wiimmees>"" CALIBRATION LABORATORY
        Add: No.51 XueyuanRoad, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2218     Fax: +86—10—62304633—2209
        E—mail: ettl@chinatl.com      Hitp//wwis.chinattl.en

Glossary:
TSL                   tissue simulating liquid
NORMx,y,z             sensitivity in free space
ConvF                 sensitivity in TSL / NORMx,y,z
DCP                   diode compression point
CF                    crest factor (1/duty_cycle) of the RF signal
A,B,.C.D              modulation dependent linearization parameters
Polarization ®        ® rotation around probe axis
Polarization 0        8 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                   8=0 is normal to probe axis
Connector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
   Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
   Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
   July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
    devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
e   NORMx,y,z: Assessed for E—field polarization 6=0 (f900MHz in TEM—cell; {>1800MHz: waveguide).
     NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
     E" —field uncertainty inside TSL (see below ConvF).
e    NORM(Px,y.2 = NORMx,y,z* frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
     frequency response is included in the stated uncertainty of ConvE.
e    DCPx,y,2: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
*    PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
     characteristics.
e    Axy2; Bxyz: Cxy,2;VRx,y,z:A,B,C are numerical linearization parameters assessed based on the
     data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum calibration range expressed in RMS voltage across the diode.
*    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
     Transfer Standard for f800MHz) and inside waveguide using analytical field distributions based on
     power measurements for f >800MHz. The same setups are used for assessmentof the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
     that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
     allows extending the validity from50MHz toz100MHz.
e    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
     phantom exposed by a patch antenna.
e    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
     probe tip (on probe axis). No tolerance required.
e    Connector Angle: The angle is assessed using the information gained by determining the NORMx
     (no uncertainty required).

Certificate No: Z17—97167                             Page 2 of11


           se© n Collsboretion with
          /T        a
          Wiaaee»~" CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Ha         District, Beijing, 100191, China
        Tel:       0—62304633—2218      ax:    +86—10—62304633—2209
        E—mail: cttl @chinattl.com     Hitp:/www.chinattl.en




                      Probe EX3DV4


                                            SN: 7431


                                 Calibrated: September 30, 2017

                             Calibrated for DASY/EASY Systems
                                      (Note: non—compatible with DASY2 system!)




Certificate No: 217—97167                              Page 3 of 11


                                      In Collaboration with
                            T, s             _p     e         a       g

        Tel: +86—1      04     2218        Fax: +86—10—   01633—2209
        E—mail: cttl@chinattl.com            ps/www.chinattl.en



     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7431

Basic Calibration Parameters
                                       Sensor X                Sensor Y           Sensor Z     Unc (k=2)
 Norm{uV/(Vim)®) *                     0.39                    0.35               0.39        £10.0%
 DCP(mV)®                              103.6                   106.0              103.3


Modulation Calibration Parameters
 uiD         Communication                           A            B        ¢          D      VR        Une®
             System Name                             dB           dB{uV               dB     mV        (k=2)
 0           cw                            x         0.0          0.0      1.0        0.00   156.2     42.9%
                                           v         0.0          0.0      1.0               146.1
                                           2         0.0          0.0     | 1.0              155.7



     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.



* The uncertainties of Norm X, Y, Z do not affect the E*—field uncertainty inside TSL (see Page 5 and Page 6).
® Numerical linearization parameter: uncertainty not required.
€ Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z17—97167                                  Page 4 of11


               e® .( Collsboretion with
              x7J a
              Wez CALIBRATION LABORATORY
             Add: No.51 Xueyuan Road, Haidian District, Beijing. 100191, China
             Tel: +86—10—62304633—2218.     Fax: +86—10—62304633—2209
             E ails ettl @chinattl.com      Hitp://wwws.chinattl.en



       DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7431

Calibration Parameter Determined in Head Tissue Simulating Media
                           T                      wz                                                         G
  f [MHz]®      Pe’::'nli:il\‘::y £   Com:;lcr:\)npty    ConvF X       ConvF Y   ConvF Z   Alpha®   D(:’:;       (L:::;t)

    750               41.9                 0.89            10.04         10.04    10.04     0.35     0.80        £12.1%
    900               41.5                 0.97            9.72           9.72    9.72      0.16     1.33        |£121%
   1750               40.1                 1.37            8.36           8.36    8.36      0.22     1.09        |£121%
   1900               40.0                 1.40            8.24           8.24    8.24      0.25     1.02        £12.1%
   2450               39.2                 1.80            7.86           7.86    7.86      0.59     0.70        £12.1%
   2600               39.0                 1.96            7.38           7.38    7.38      0.67     0.67        £12.1%
   5200               36.0                4.66             5.92           5.92    5.92      0.40     1.35        +13.3%
   5300               35.9                4.76             5.60           5.60    5.60      0.40     1.35        +13.3%
   5600               35.5                5.07             4.99           4.99    4.99      0.40     1.35        £13.3%
   5800               35.3                5.27             5.05           5.05    5.05      0.40     140         |£13.3%



& Frequency validity above 300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
*50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
"At frequency below 3 GHz, thevalidity of tissue parameters (e and 0) can be relaxed to +10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and 0) is
restricted to +5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
GA\pha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensationis always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




  Certificate No: Z17—97167                                Page 5 of11


             3e® in Colleboration with
             ex"]"[°J,   a
             Wimzzeee>"* CALIBRATION LABORATORY
           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel:  +86    2304633—2218      Fax: +86—10—62304633—2209
           E—mail: ettl@chinattl.com      Hitp:/rww.chinattl.en



      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7431

Calibration Parameter Determined in Body Tissue Simulating Media
                        :                     fow                                                        6

 f [MHz]® P;:':;':;y — c°"1;7{:‘)":y ConvF X ConvF Y ConvF Z Alpha® D(:’r:; (L:(':;t)
    750       555          0.95       011     1011    1011    040     0.85 £121%
    900       55.0          1.05       9.85    9.85    986    o27     1.08 £121%
   1750       534           149        8.23    8.23    823    o2      101 £121%
   1900       53.3          1.52       8.00    8.00    800    014     1.89 £121%
   2450       52.7          1.95       7.62    7.62    T62    038     1.05 £121%
   2600       525          216         7.31    731     731    os7     075 £+121%]
   5200       49.0         5.30        5.22    5.22    522    050     1.60 +133%
   5300       48.9         5.42        4.93    4.93    493    050     1.65 £133%
   5600       48.5         5.77        4.40    4.40    440    050     1.70 £133%
   5800       48.2         6.00        4.43    4.43    443    082     1.80 £133%

© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
" At frequency below 3 GHz, the validity of tissue parameters (£ and 0) can be relaxed to +10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
©Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probetip diameter from the boundary.




   Certificate No: Z17—97167                              Page 6 of11


                                          A              In Collaboration with

                              =7"7J,s    a
                               Wimzeee>" CALIBRATION LABEORATORY
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2218      Fax: +86—10—62304633—2209
             E—mail: citl@chinattl.com                        Hiutp:/wwa.chinatl.en


                                                   Frequency Response of E—Field
                                                   (TEM—Cell: ifi110 EXX, Waveguide: R22)
        Frequency response (normalized)




                                          0.5       +                    T                 T     T            T      T
                                                                      1000            1500      2000         2500   3000

                                                             Ca—                      f [MHz]          <%~
                                                             TEM                                       R22



                                                Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z17—97167                                                    Page 7 of11


                  w®                      in collsboration with

                 ®=‘7‘7°J,  a
                 Wimmmes>"* CALIBRATION LABORATORY
               Add: No.51 Xueyuan Road, Haidia     istrict. Beijing, 100191, China
               Te        0—62304633—2218       Fax: +86—10—623040633—2209
               E—mail: cttl@chinattl.com       Uitp:/wwis.chinattl.en



                                       Receiving Pattern (®), 0=0°

                     f=600 MHz, TEM                                                  f=1800 MHz, R22
   ErrorfdB]




                            u               r                                                           u
                          —150            ~100             «50                 0       5o     IlI)O    150
                                                         Rollf®]
                                    [E—<100MHz __—+—600MHz _ ———1800MHz__———2500MHz]
                                    Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




Certificate No: Z17—97167                                        Page 8 of11


                 é                                ~"      in Collsboration with

              ma7"7J, a
                         CWimzese>"*                      CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian Di                                           0191, China
        1                            6—10    04633—2218         f                     3—22
        E—mail: ettl@chinatt.com                                                  .n



                                                       Dynamic Range f(SARpneaq)
                                                                    (TEM cell, f= 900 MHz)
                  Input Signal[pV]




                                            10°                10°                    10°                 10‘         10°         10
                                                                                       SAR[mW/cm*]
                                                                       —i—Inot compensated          —@— compensated
            Error[dB]




                                                                                  T                   T
                                                                       10°         , 10            10                            10°
                                                                          SAR[mW/cm ]
                                                  [___—=—Inot compensated               —e— compensated                     _|
                                                     Uncertainty of Linearity Assessment: £0.9% (k=2)
Certificate No: Z17—97167                                                    Page 9 of11


           B© in Collsbboration with
                 TTL a
          Timmzeee>"*                     CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel:     10—62304633—2218                Fax: +86—10—62304633—2209
        E—mail: cttl@chinattl.com                Hitp:www.chinattl.en



                             Conversion Factor Assessment
      f=900 MHz, WGLS R9(H_convF)                                                         f=1750 MHz, WGLS R22(H_convF)
              400                                                                              3000


              aso                                                                                               |
                                                                                               2500             j
              aco                                                                                          \
                                                                                               2000
            gz 50                                                                          s
            ©                                                                              s
            §20                                                                            § 1500                   |
            &0 150                                                                         &
                                                                                           6
                                                                                             1000                       a—                                 —
              100                                                                                               |
                                                                                               50o
              aso                                                                                                            \
              100                                                                              200
                     o            20        «o           so         so          100                   o0       dn            m    s          4e   sn   i       To
                                                 2{mm}                                                                                2mn}
                                       —+—measured       —lnllib‘ll                                                      ~+—measuted     —II\I&EGI‘




                          Deviation from Isotropy in Liquid
                         7 Axis




                                                    al",
                                                               00                                                            i
                                                     Aus"           250                                                 y
                                                                          300
                                                                                  350

                                              —10    —o80     —oo .040 .020           0    020        o40      o%0       080     10
                                   Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z17—97167                                           Page 10 of11


          se© in Collsboration vilth
          ©7"7‘]/, s p e a q
           ns                       CALIBRATION LABORATORY
        Add: No.51 Xucy          ad. Haidian District, Beijing, 100191, China
        Tel: +86—10—062304633—2218         Fax: +86—10—62304633—2209
        E—mail: cttl@chinattl.com          Hup:/wwas.chinattl.en



    DASY/EASY — Parameters of Probe: EX3DV4 — SN: 7431


     Other Probe Parameters
    Sensor Arrangement                                                          Triangular

    Connector Angle (°)                                                               87.2

    Mechanical Surface Detection Mode                                             enabled

    Optical Surface Detection Mode                                                disable

    Probe Overall Length                                                          337mm

    Probe Body Diameter                                                             10mm

    Tip Length                                                                       9mm

    Tip Diameter                                                                   2.5mm

    Probe Tip to Sensor X Calibration Point                                          1mm

    Probe Tip to Sensor Y Calibration Point                                          1mm

    Probe Tip to Sensor Z Calibration Point                                          1mm

    Recommended Measurement Distance from Surface                                  14mm




Certificate No: Z17—97167                                Page 11 of 11


DIPOLE CALIBRATION CERTIFICATES

















Document Created: 2018-09-26 16:22:14
Document Modified: 2018-09-26 16:22:14

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC