Calibration Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland




Client




 Object(s)


 Calibration procedure(s)




 Calibration date:


 Condition of the calibrated item



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probabilty are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facilty: environ ment temperature 22 +/ 2 degrees Celsius and humidity < 75%.


 Calibration Equipment used (M&TE critical for calibration)


 Model Type                              ID #                      Cal Date (Calibrated by, Certificate No.)            Scheduled Calibration
 Power meter EPM E44198                  GB41293874                5—May—04 (METAS, No 251—00388)                        May—05
 Power sensor E4412A                     MY41495277                5—May—04 (METAS, No 251—00388)                        May—05
 Reference 20 dB Attenuator              SN: 5086 (20b)            3—May—04 (METAS, No 251—00389)                        May—05
 Fluke Process Calibrator Type 702       SN: 6295803               8—Sep—03 (Sintrel SCS No. E—030020)                  Sep—04
 Power sensor HP 8481A                   MY41092180                18—Sep—02 (SPEAG, in house check Oct—03)             In house check: Oct 05
 RF generator HP 8684C                   US3642U01700              4—Aug—99 (SPEAG, in house check Aug—02)              In house check: Aug—05
 Network Analyzer HP 8753E               US37390585                18—Oct—01 (SPEAG, in house check Oct—03)             In house check: Oct 05




 Calibrated by:




                                                                                                                         Date issued: May 24, 2004



 This calibration certificate is issued as an intermediate solution until the accreditation process (based on ISO/IEC 17025 InternationalStandard) for
 Calibration Laboratory of Schmid & Partner Engineering AG is completed.




880—KPO301061—A                                                        Page 1 of 8


Probe ET3DVC6

           SN:1590
 Manufactured:                   March 19, 2001
 Last calibrated:                May 15, 2003
 Recalibrated:                   May 24, 2004

     Calibrated for DASY Systems
      (Note: non—compatible with DASY2 system!)




                 Page 2 of 8


ET3DV6 SN: 1590                                                                                    May 24, 2004


DASY — Parameters of Probe: ET3DV6 SN:1590


Sensitivity in Free Space                                                        Diode Compression"®

            NormX                      1.85 uV/(V/im)                           DCPX        91     mV
            NormY                      2.01 uV/(V/im)                           DCPY        91     mV
            NormZ                      1.73 uV/i(V/im)                          DCPZ        91     mV

Sensitivity in Tissue Simulating Liquid (Conversion Factors)
Plese see Page 7.



Boundary Effect

Head                        900 MHz           Typical SAR gradient: 5 % per mm


           Sensor Centerto Phantom Surface Distance                              3.7 mm 4.7 mm
           SARpe [%6]             Without Correction Algorithm                     8.0     4.4

           SARpe [%]              With Correction Algorithm                        0.1     0.2


Head                      1800 MHz            Typical SAR gradient: 10 % per mm


           Sensor Center to Phantom Surface Distance                             3.7 mm   4.7 mm
           SARpe [%]              Without Correction Algorithm                    12.2     8.5
           SARp, [%]              With Correction Algorithm                        0.2     0.1



Sensor Offset

            Probe Tip to Sensor Center                                    2.7    mm
           Optical Surface Detection                                      in tolerance




 The reported uncertainty of measurement is stated as the standard uncertainty of
 measurement multiplied by the coverage factor k=2, which for a normal distribution
 corresponds to a coverage probability of approximately 95%.

* numerical linearization parameter: uncertainty not required




                                                            Page 3 of 8


ET3DV6 SN:1590                                                                      May 24, 2004



                                  Frequency Response of E—Field
                                         ( TEM—Cell:ifi110, Waveguide R22)


                           1.50


                           1.40


                           1.30


                           1.20
      frequency response

                           e o




                           0.70


                           0.60


                           0.50
                                   500      1000       1500        2000   2500   3000
                                                    f [MHz]
                                         —@—TEM                  —O—R22




                                                   Page 4 of 8


ET3DV6 SN: 1590                                                                                 May 24, 2004




                                   Receiving Pattern (¢ ), 0 = 0°

                  f= 900 MHz, TEM cell ifi110                               f= 1800 MHz, WG R22




            290                    ,               70



              1          >         %fi.
      227;
            " [ Y&’}'{i‘fi‘&%?fl
                 t
                     ';‘;:/ Pn 3 \5:.8\‘.: 19:0

                       y tA
                  SSA             4‘\}” 199
                         o[|Le         60




                                                                        —@—X   —@—¥   —#—2     —O—Tot




              1.00
              0.80
                                                                                               —0— 30 MHz
              0.60
              0.40                                                                             —®— 100 MHz
                                                                                               —&— 300 MHz
error{dB]




              0.20
              0.00                                                                             —>—900 MHz
             —0.20                                                                             —®— 1800 MHz
             —0.40                                                                             —&—2500 MHz
             —0.60
             —0.80
             ~1.00
                     0       60             120         180           240       300      360




                                              Axial Isotropy Error < £ 0.2 dB




                                                        Page 5 of 8


ET3DV6 SN: 1590                                                                    May 24, 2004



                           Dynamic Range f(SARneaa)
                                     (Waveguide R22 )



       1.E+7



       1.E+6



       1.E+5



       1.E+4



       1.E+3



       1.E+2



       1.E+1



       1.E+0
          0.0001   0.001      0.01         0.1            3.   10.        100.
                                        mW/cm*

                   —®—not compensated                —®—compensated




                      0.01           0.1              1              10          100
                                           mW/cm*



                             Probe Linearity Error < * 0.2 dB




                                       Page 6 of 8


ET3DV6 SN: 1590                                                                                                               May 24, 2004

                                        Conversion Factor Assessment

                           f= 900 MHz, WGLS R9 (head)                                           f= 1750 MHz, WGLS R22 (head)
                                                                                             25.0




                                                                           SAR[mW/cm*] / W
   SAR[mW/cm*] / W
                     ho
                     o
                     aa
                     in




                     1.0 4
                                                                                              5.0 —
                     0.5 7

                     0.0 +


                                        z[mm]                                                                         z[mm]

                       —@— Analytical   —O—Measuremets                                         —@— Analytical         —O—Measuremets




f [MHz] Validity [MHz]®                  Tissue Permittivity    Conductivity                        __Alpha   Depth     __ConvF Uncertainty
835                       750—950        Head    41.5+5%        0.90+5%                               0.68     1.64       6.71 £11.9% (k=2)

1750                      1700—1800      Head    40.0%£5%       140 +5%                               0.43     2.67       5.28 +9.7% (k=2)

1900                      1850—1950      Head    40.0%£5%       140 %5%                               0.46     2.81       5.03 £+9.7% (k=2)

2450                      2400—2500      Head    39.2+5%        1.80+5%                               0.81     1.95       4.44 +9.7% (k=2)




835                       750—950         Body   55.2%5%         0.97%5%                              0.49     1.99       6.54 £11.9% (k=2)

1750                      1700—1800       Body §3.3%5%          1.52%5%                               0.50     2.87       4.68 £9.7% (k=2)
1900                      1850—1950       Body   53.3%5%        1.52%5%                               0.52     2.93       4.58 +9.7% (k=2)

2450                      2400—2500       Body   52.7%5%          1.95%5%                             0.91     1.78       4.22 £9.7% (k=2)




® The total standard uncertainty is calculated as root—sum—square of standard uncertainty of the Conversion Factor at calibration
frequency and the standard uncertainty for the indicated frequency band.




                                                            Page 7 of 8


ET3DV6 SN: 1590                                                                              May 24, 2004



                   Deviation from Isotropy in HSL
                                    Error (0, $), f = 900 MHz




                                                                                Error [dB]




              ©—1.00—0.80 ©—0.80—0.60 M—0.60—0.40 M—0.40—0.20 ©—0.20—0.00
              ©0.00—0.20   ©0.20—0.40   ©M0.40—0.60   M0.60—0.80   ©0.80—1.00




                            Spherical Isotropy Error < * 0.4 dB




                                          Page 8 of 8


Schmid & Partner Engineering AG                                 s      p       e   a     g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 245 9779
info@speag.com, http://www.speag.com




                         Additional Conversion Factors
                                               for Dosimetric E—Field Probe



                  Type:                                               ET3DV6


                  Serial Number:                                        1590



                  Place of Assessment:                                Zurich


                  Date of Assessment:                               May 25, 2004



                  Probe Calibration Date:                         May 24, 2004



       Schmid & Partner Engineering AG hereby certifies that conversion factor(s) ofthis probe
       have been evaluated on the date indicated above. The assessment was performed using
       the FDTD numerical code SEMCAD of Schmid & Partner Engineering AG. Since the
       evaluation is coupled with measured conversion factors, it has to be recalculated yearly,
       i.e., following the re—calibration schedule of the probe. The uncertainty of the numerical
       assessment is based on the extrapolation from measured value at 900 MHz or at 1800
       MHz.




                  Assessed by:




ET3DV6—SN:1590                                   Page 1 of 2                           May 25, 2004


Schmid & Partner Engineering AG                              s       E       e         a     g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 1 245 9700, Fax +41 1 245 9779
info@speag.com, http://www.speag.com




Dosimetric E—Field Probe ET3DV6 SN:1590
Conversion factor (+ standard deviation)


150 MHz                   ConvF            9.1 + 8%                    s = 52.3 + 5%
                                                                       0 =0.76 + 5% mho/m
                                                             (head tissue)

300 MHz                    ConvF           7.9 +8%                     &= 45.3 £ 5%
                                                                       0=0.87 + 5% mho/m
                                                             (head tissue)

450 MHz                    ConvF           7.5#8%                      €;= 43.5 £ 5%
                                                                       0=0.87 + 5% mho/m
                                                             (head tissue)

150 MHz                    ConvF           8.8 £8%                     &,= 61.9 + 5%
                                                                       0 = 0.80 + 5% mho/m
                                                             (body tissue)

450 MHz                    ConvF           7.1 *8%                     & = 56.7 + 5%
                                                                       0=0.94 + 5% mho/m
                                                             (body tissue)




Important Note:

For numerically assessed probe conversion factors, parameters Alpha and Delta in the
DASY software must have the following entries: Alpha = 0 and Delta= 1.

Please see also Section 4.7 of the DASY4 Manual.




ET3DV6—SN:1590                                 Page 2 of 2                                 May 25, 2004