SAR Report Test Plots

FCC ID: JOYJA53

RF Exposure Info

Download: PDF
FCCID_4183021

Report No.: DRRFCC1902-0011          FCC ID: JOYJA53             Date of issue: Feb. 7, 2019



SAR Plots

    Verification Plots


    SAR Test Plots




                              Copyright Ⓒ 2019, DT&C Co., Ltd.


                                    DT&C Co., Ltd.
           DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 - SN:1049

          Communication System: UID 0, CW (0); Frequency: 750 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 750 MHz; σ = 0.916 S/m; εr = 42.482; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.61, 6.61, 6.61); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-10; Ambient Temp: 20.3; Tissue Temp: 20.7

                       750 MHz System Head Verification (250mW)

               Area Scan (6x15x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                    Power Drift = -0.01 dB
                             Peak SAR (extrapolated) = 2.99 W/kg
                         SAR(1 g) = 2.03 W/kg; SAR(10 g) = 1.35 W/kg




              0 dB = 2.57 W/kg


                                    DT&C Co., Ltd.
           DUT: Dipole 750 MHz; Type: D750V3; Serial: D750V3 - SN:1049

          Communication System: UID 0, CW (0); Frequency: 750 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 750 MHz; σ = 0.974 S/m; εr = 57.391; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.29, 6.29, 6.29); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-10; Ambient Temp: 20.3; Tissue Temp: 20.9

                       750 MHz System Body Verification (250mW)

               Area Scan (6x15x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                    Power Drift = -0.03 dB
                             Peak SAR (extrapolated) = 3.45 W/kg
                         SAR(1 g) = 2.27 W/kg; SAR(10 g) = 1.43 W/kg




              0 dB = 2.89 W/kg


                                    DT&C Co., Ltd.
          DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 - SN:4d159

          Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 835 MHz; σ = 0.881 S/m; εr = 41.806; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.35, 6.35, 6.35); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-07; Ambient Temp: 20.5; Tissue Temp: 20.8

                       835 MHz System Head Verification (250mW)

               Area Scan (6x11x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.01 dB
                             Peak SAR (extrapolated) = 3.41 W/kg
                         SAR(1 g) = 2.23 W/kg; SAR(10 g) = 1.41 W/kg




              0 dB = 2.87 W/kg


                                    DT&C Co., Ltd.
          DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 - SN:4d159

          Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 835 MHz; σ = 1.009 S/m; εr = 54.397; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.23, 6.23, 6.23); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-07; Ambient Temp: 20.5; Tissue Temp: 20.7

                       835 MHz System Body Verification (250mW)

               Area Scan (6x11x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.01 dB
                             Peak SAR (extrapolated) = 3.56 W/kg
                         SAR(1 g) = 2.32 W/kg; SAR(10 g) = 1.45 W/kg




              0 dB = 2.96 W/kg


                                    DT&C Co., Ltd.
          DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 - SN:4d159

          Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 835 MHz; σ = 0.881 S/m; εr = 42.16; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.35, 6.35, 6.35); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-09; Ambient Temp: 20.3; Tissue Temp: 20.6

                       835 MHz System Head Verification (250mW)

               Area Scan (6x11x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.00 dB
                             Peak SAR (extrapolated) = 3.49 W/kg
                         SAR(1 g) = 2.26 W/kg; SAR(10 g) = 1.45 W/kg




              0 dB = 2.95 W/kg


                                    DT&C Co., Ltd.
          DUT: Dipole 835 MHz; Type: D835V2; Serial: D835V2 - SN:4d159

          Communication System: UID 0, CW (0); Frequency: 835 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 835 MHz; σ = 1.006 S/m; εr = 54.367; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.23, 6.23, 6.23); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-09; Ambient Temp: 20.3; Tissue Temp: 20.5

                       835 MHz System Body Verification (250mW)

               Area Scan (6x11x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.00 dB
                             Peak SAR (extrapolated) = 3.59 W/kg
                         SAR(1 g) = 2.35 W/kg; SAR(10 g) = 1.56 W/kg




              0 dB = 3.16 W/kg


                                    DT&C Co., Ltd.
        DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 - SN:5d176

           Communication System: UID 0, CW; Frequency: 1900 MHz;Duty Cycle: 1:1
        Medium parameters used: f = 1900 MHz; σ = 1.434 S/m; εr = 38.681; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3866; ConvF(8.03, 8.03, 8.03); Calibrated: 2018-05-31; Electronics: DAE4 Sn1391
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-08; Ambient Temp: 20.6; Tissue Temp: 20.9

                       1900 MHz System Head Verification (100mW)

                Area Scan (5x7x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                    Power Drift = -0.05 dB
                             Peak SAR (extrapolated) = 8.65 W/kg
                         SAR(1 g) = 4.03 W/kg; SAR(10 g) = 2.12 W/kg




              0 dB = 7.86 W/kg


                                    DT&C Co., Ltd.
        DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 - SN:5d176

          Communication System: UID 0, CW (0); Frequency: 1900 MHz;Duty Cycle: 1:1
         Medium parameters used: f = 1900 MHz; σ = 1.558 S/m; εr = 52.07; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3866; ConvF(7.68, 7.68, 7.68); Calibrated: 2018-05-31; Electronics: DAE4 Sn1391
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-08; Ambient Temp: 20.6; Tissue Temp: 21.0

                       1900 MHz System Body Verification (100mW)

                Area Scan (5x7x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.03 dB
                             Peak SAR (extrapolated) = 9.58 W/kg
                         SAR(1 g) = 4.13 W/kg; SAR(10 g) = 2.12 W/kg




              0 dB = 8.38 W/kg


                                   DT&C Co., Ltd.
         DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 - SN:920
         Communication System: UID 0, CW (0); Frequency: 2450 MHz; Duty Cycle: 1:1
        Medium parameters used: f = 2450 MHz; σ = 1.842 S/m; εr = 37.756; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3916; ConvF(7.72, 7.72, 7.72); Calibrated: 4/25/2018; Electronics: DAE4 Sn1394
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
           Phantom: SAM-twin middle_2013_09_24; Type: QD000P40CD; Serial: 1782
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-15; Ambient Temp: 20.9; Tissue Temp: 21.3

                     2450 MHz System Head Verification (100 mW)

              Area Scan (7x10x1): Measurement grid: dx=10mm, dy=10mm
         Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                    Power Drift = -0.04 dB
                             Peak SAR (extrapolated) = 11.6 W/kg
                         SAR(1 g) = 5.11 W/kg; SAR(10 g) = 2.38 W/kg




             0 dB = 12.1 W/kg


                                   DT&C Co., Ltd.
         DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 - SN:726
         Communication System: UID 0, CW (0); Frequency: 2450 MHz; Duty Cycle: 1:1
        Medium parameters used: f = 2450 MHz; σ = 1.914 S/m; εr = 51.635; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3916; ConvF(7.69, 7.69, 7.69); Calibrated: 4/25/2018; Electronics: DAE4 Sn1394
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
           Phantom: SAM-twin middle_2013_09_24; Type: QD000P40CD; Serial: 1782
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-15; Ambient Temp: 20.9; Tissue Temp: 21.0

                     2450 MHz System Body Verification (100 mW)

              Area Scan (7x10x1): Measurement grid: dx=10mm, dy=10mm
         Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = -0.07 dB
                             Peak SAR (extrapolated) = 12.1 W/kg
                         SAR(1 g) = 5.25 W/kg; SAR(10 g) = 2.48 W/kg




             0 dB = 12.8 W/kg


                                   DT&C Co., Ltd.
                                 DUT: JA53; Type: Folder
      Communication System: UID 0, GSM 850 (0); Frequency: 836.6 MHz;Duty Cycle: 1:8.3
       Medium parameters used: f = 836.6 MHz; σ = 0.882 S/m; εr = 41.785; ρ = 1000 kg/m3
                                 Phantom section: Left Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.35, 6.35, 6.35); Calibrated: 3/21/2018; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                 Test Date: 2019-01-07; Ambient Temp: 20.5; Tissue Temp: 20.8

             Left Touch, GSM850 Ch. 190, Ant Internal, Standard Battery

              Area Scan (8x17x1): Measurement grid: dx=15mm, dy=15mm
         Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                    Power Drift = 0.11 dB
                           Peak SAR (extrapolated) = 0.889 W/kg
                       SAR(1 g) = 0.630 W/kg; SAR(10 g) = 0.427 W/kg




             0 dB = 0.723 W/kg
                                                                         A1


                                   DT&C Co., Ltd.
                                 DUT: JA53; Type: Folder
   Communication System: UID 0, GSM 850_12 (0); Frequency: 836.6 MHz;Duty Cycle: 1:2.075
      Medium parameters used: f = 836.6 MHz; σ = 0.882 S/m; εr = 41.785; ρ = 1000 kg/m3
                                Phantom section: Left Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.35, 6.35, 6.35); Calibrated: 3/21/2018; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                 Test Date: 2019-01-07; Ambient Temp: 20.5; Tissue Temp: 20.8

      Left Touch, GSM850 GPRS 4 Tx Ch. 190, Ant Internal, Standard Battery

              Area Scan (8x17x1): Measurement grid: dx=15mm, dy=15mm
         Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                    Power Drift = 0.19 dB
                           Peak SAR (extrapolated) = 0.855 W/kg
                       SAR(1 g) = 0.639 W/kg; SAR(10 g) = 0.447 W/kg




             0 dB = 0.717 W/kg
                                                                        A2


                                     DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

        Communication System: UID 0, PCS 1900 (0); Frequency: 1880 MHz;Duty Cycle: 1:8.3
         Medium parameters used: f = 1880 MHz; σ = 1.417 S/m; εr = 38.761; ρ = 1000 kg/m3
                                   Phantom section: Left Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3866; ConvF(8.03, 8.03, 8.03); Calibrated: 2018-05-31; Electronics: DAE4 Sn1391
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
              Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-08; Ambient Temp: 20.6; Tissue Temp: 20.9

              Left Touch, PCS1900, Ch. 661, Ant Internal, Standard Battery

               Area Scan (8x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = 0.12 dB
                              Peak SAR (extrapolated) = 1.09 W/kg
                         SAR(1 g) = 0.657 W/kg; SAR(10 g) = 0.373 W/kg




              0 dB = 0.868 W/kg
                                                                          A3


                                     DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

    Communication System: UID 0, PCS1900_4Tx (0); Frequency: 1880 MHz;Duty Cycle: 1:2.075
       Medium parameters used: f = 1880 MHz; σ = 1.417 S/m; εr = 38.761; ρ = 1000 kg/m3
                                   Phantom section: Left Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3866; ConvF(8.03, 8.03, 8.03); Calibrated: 2018-05-31; Electronics: DAE4 Sn1391
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
              Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-08; Ambient Temp: 20.6; Tissue Temp: 20.9

       Left Touch, PCS1900 GPRS 4 Tx, Ch. 661, Ant Internal, Standard Battery

               Area Scan (8x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = 0.15 dB
                              Peak SAR (extrapolated) = 1.11 W/kg
                         SAR(1 g) = 0.683 W/kg; SAR(10 g) = 0.405 W/kg




              0 dB = 0.898 W/kg
                                                                          A4


                                     DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

      Communication System: UID 0, WCDMA 850 (0); Frequency: 836.6 MHz;Duty Cycle: 1:1
        Medium parameters used: f = 836.6 MHz; σ = 0.883 S/m; εr = 42.138; ρ = 1000 kg/m3
                                   Phantom section: Left Section

                                     DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.35, 6.35, 6.35); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                       Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-09; Ambient Temp: 20.3; Tissue Temp: 20.6

           Left Touch, WCDMA850 Ch. 4183, Ant Internal, Standard Battery

               Area Scan (8x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = 0.11 dB
                             Peak SAR (extrapolated) = 0.676 W/kg
                         SAR(1 g) = 0.481 W/kg; SAR(10 g) = 0.335 W/kg




              0 dB = 0.541 W/kg
                                                                          A5


                                     DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

    Communication System: UID 0, LTE Band 17(FCC) (0); Frequency: 710 MHz;Duty Cycle: 1:1
       Medium parameters used: f = 710 MHz; σ = 0.875 S/m; εr = 42.941; ρ = 1000 kg/m3
                                   Phantom section: Left Section

                                     DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.61, 6.61, 6.61); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                       Sensor-Surface: 3mm (Mechanical Surface Detection)
              Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-10; Ambient Temp: 20.3; Tissue Temp: 20.7

          Left Touch, LTE Band 17 Ch. 23790, Ant Internal, Standard Battery

                      Mode : BandWidth 10 MHz, QPSK, RB Size: 1

               Area Scan (8x17x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = 0.12 dB
                             Peak SAR (extrapolated) = 0.704 W/kg
                         SAR(1 g) = 0.504 W/kg; SAR(10 g) = 0.358 W/kg




              0 dB = 0.565 W/kg
                                                                         A6


                                    DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder
  Communication System: UID 0, 00_2.4 GHz W-LAN (0); Frequency: 2437 MHz; Duty Cycle: 1:1
      Medium parameters used: f = 2437 MHz; σ = 1.829 S/m; εr = 37.791; ρ = 1000 kg/m3
                                Phantom section: Right Section

                                   DASY5 Configuration:
Probe: EX3DV4 - SN3916; ConvF(7.72, 7.72, 7.72); Calibrated: 4/25/2018; Electronics: DAE4 Sn1394
                     Sensor-Surface: 2mm (Mechanical Surface Detection)
           Phantom: SAM-twin middle_2013_09_24; Type: QD000P40CD; Serial: 1782
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-15; Ambient Temp: 20.9; Tissue Temp: 21.3

     Right Touch, W-LAN(802.11b - 2.4G) Ch. 6, Ant Internal, Standard Battery

              Area Scan (10x21x1): Measurement grid: dx=12mm, dy=12mm
         Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = 0.00 dB
                            Peak SAR (extrapolated) = 0.0700 W/kg
                        SAR(1 g) = 0.037 W/kg; SAR(10 g) = 0.020 W/kg




             0 dB = 0.0541 W/kg
                                                                       A7


                                   DT&C Co., Ltd.
                                 DUT: JA53; Type: Folder
      Communication System: UID 0, GSM 850 (0); Frequency: 848.8 MHz;Duty Cycle: 1:8.3
       Medium parameters used: f = 848.8 MHz; σ = 1.022 S/m; εr = 54.278; ρ = 1000 kg/m3
                                 Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.23, 6.23, 6.23); Calibrated: 3/21/2018; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                 Test Date: 2019-01-07; Ambient Temp: 20.5; Tissue Temp: 20.7

              1cm space from Body, Rear, GSM850 Ch. 251, Ant Internal
              Area Scan (8x13x1): Measurement grid: dx=15mm, dy=15mm
         Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                    Power Drift = 0.02 dB
                            Peak SAR (extrapolated) = 1.34 W/kg
                       SAR(1 g) = 0.962 W/kg; SAR(10 g) = 0.659 W/kg




             0 dB = 1.10 W/kg
                                                                      A8


                                   DT&C Co., Ltd.
                                 DUT: JA53; Type: Folder
   Communication System: UID 0, GSM 850_12 (0); Frequency: 848.8 MHz;Duty Cycle: 1:2.075
      Medium parameters used: f = 848.8 MHz; σ = 1.022 S/m; εr = 54.278; ρ = 1000 kg/m3
                                Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.23, 6.23, 6.23); Calibrated: 3/21/2018; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                 Test Date: 2019-01-07; Ambient Temp: 20.5; Tissue Temp: 20.7

        1cm space from Body, Rear, GSM850 GPRS 4Tx Ch. 251, Ant Internal
              Area Scan (8x13x1): Measurement grid: dx=15mm, dy=15mm
         Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                    Power Drift = -0.04 dB
                            Peak SAR (extrapolated) = 1.36 W/kg
                        SAR(1 g) = 1.04 W/kg; SAR(10 g) = 0.750 W/kg




             0 dB = 1.16 W/kg
                                                                         A9


                                     DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

        Communication System: UID 0, PCS 1900 (0); Frequency: 1880 MHz;Duty Cycle: 1:8.3
         Medium parameters used: f = 1880 MHz; σ = 1.541 S/m; εr = 52.133; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3866; ConvF(7.68, 7.68, 7.68); Calibrated: 2018-05-31; Electronics: DAE4 Sn1391
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
              Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-08; Ambient Temp: 20.6; Tissue Temp: 21.0

               1 cm space from Body, Rear, PCS1900, Ch. 661, Ant Internal

               Area Scan (8x13x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = -0.04 dB
                             Peak SAR (extrapolated) = 0.923 W/kg
                         SAR(1 g) = 0.515 W/kg; SAR(10 g) = 0.284 W/kg




              0 dB = 0.721 W/kg
                                                                         A10


                                     DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

    Communication System: UID 0, PCS1900_4Tx (0); Frequency: 1880 MHz;Duty Cycle: 1:2.075
       Medium parameters used: f = 1880 MHz; σ = 1.541 S/m; εr = 52.133; ρ = 1000 kg/m3
                                   Phantom section: Flat Section

                                     DASY5 Configuration:
Probe: EX3DV4 - SN3866; ConvF(7.68, 7.68, 7.68); Calibrated: 2018-05-31; Electronics: DAE4 Sn1391
                       Sensor-Surface: 2mm (Mechanical Surface Detection)
              Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
          Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-08; Ambient Temp: 20.6; Tissue Temp: 21.0

        1 cm space from Body, Rear, PCS1900 GPRS 4 Tx, Ch. 661, Ant Internal

               Area Scan (8x13x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                      Power Drift = -0.07 dB
                              Peak SAR (extrapolated) = 1.03 W/kg
                         SAR(1 g) = 0.587 W/kg; SAR(10 g) = 0.327 W/kg




              0 dB = 0.814 W/kg
                                                                         A11


                                    DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

      Communication System: UID 0, WCDMA 850 (0); Frequency: 836.6 MHz;Duty Cycle: 1:1
        Medium parameters used: f = 836.6 MHz; σ = 1.007 S/m; εr = 54.358; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.23, 6.23, 6.23); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
            Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-09; Ambient Temp: 20.3; Tissue Temp: 20.5

           1 cm space from Body, Rear, WCDMA850 Ch. 4183, Ant Internal

               Area Scan (8x13x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = 0.00 dB
                            Peak SAR (extrapolated) = 0.972 W/kg
                        SAR(1 g) = 0.686 W/kg; SAR(10 g) = 0.470 W/kg




              0 dB = 0.785 W/kg
                                                                       A12


                                    DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder

    Communication System: UID 0, LTE Band 17(FCC) (0); Frequency: 710 MHz;Duty Cycle: 1:1
       Medium parameters used: f = 710 MHz; σ = 0.929 S/m; εr = 57.701; ρ = 1000 kg/m3
                                  Phantom section: Flat Section

                                   DASY5 Configuration:
Probe: ES3DV3 - SN3328; ConvF(6.29, 6.29, 6.29); Calibrated: 2018-03-21; Electronics: DAE4 Sn1335
                     Sensor-Surface: 3mm (Mechanical Surface Detection)
             Phantom: SAM with CRP v5.0(Right); Type: QD000P40CD; Serial: 1220
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                   Test Date: 2019-01-10 Ambient Temp: 20.3; Tissue Temp: 20.9

           1 cm space from Body, Rear, LTE Band 17 Ch. 23790, Ant Internal

                      Mode : BandWidth 10 MHz, QPSK, RB Size: 1

               Area Scan (8x13x1): Measurement grid: dx=15mm, dy=15mm
          Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
                                     Power Drift = -0.10 dB
                            Peak SAR (extrapolated) = 0.582 W/kg
                        SAR(1 g) = 0.423 W/kg; SAR(10 g) = 0.296 W/kg




              0 dB = 0.480 W/kg
                                                                        A13


                                    DT&C Co., Ltd.
                                  DUT: JA53; Type: Folder
  Communication System: UID 0, 00_2.4 GHz W-LAN (0); Frequency: 2437 MHz; Duty Cycle: 1:1
      Medium parameters used: f = 2437 MHz; σ = 1.898 S/m; εr = 51.672; ρ = 1000 kg/m3
                                Phantom section: Flat Section

                                    DASY5 Configuration:
Probe: EX3DV4 - SN3916; ConvF(7.69, 7.69, 7.69); Calibrated: 4/25/2018; Electronics: DAE4 Sn1394
                      Sensor-Surface: 2mm (Mechanical Surface Detection)
           Phantom: SAM-twin middle_2013_09_24; Type: QD000P40CD; Serial: 1782
         Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)

                  Test Date: 2019-01-15; Ambient Temp: 20.9; Tissue Temp: 21.0

       1 cm space from Body, Rear, W-LAN(802.11b - 2.4G) Ch. 6, Ant Internal
              Area Scan (10x16x1): Measurement grid: dx=12mm, dy=12mm
         Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm
                                     Power Drift = -0.08 dB
                            Peak SAR (extrapolated) = 0.117 W/kg
                        SAR(1 g) = 0.066 W/kg; SAR(10 g) = 0.037 W/kg




             0 dB = 0.0907 W/kg
                                                                       A14