RF Exposure Part 5

FCC ID: IHDT56XS2

RF Exposure Info

Download: PDF
FCCID_4287342

           r‘                       In Collaboration with

                                    CALIBRATION LABORATORY
           y
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—623
         E—mail: cttl@chinatt.com       hitp://Iwww.chinattl.en




DASYS Validation Report for Body TSL                                     Date: 03.26.2019
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN: 1090
   Communication System: UID 0, CW; Frequency: 1750 MHz; Duty Cycle: 1:1
   Medium parameters used: {= 1750 MHz; 0 = 1.449 S/m; &, = 54.97; p = 1000 kg/m3
   Phantom section: Center Section
DASYS Configuration:

             *      Probe: EX3DV4 — SN3617; ConvF(8.03, 8.03, 8.03) @ 1750 MHz; Calibrated:
                    1/31/2019
             *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
             +    Electronics: DAE4 Sn1331; Calibrated: 2/6/2019
             *«   Phantom: MFP_V5.1C ; Type: QD 000 P51CA; Serial: 1062
             *«   Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12
                  (7450)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 93.13 V/m; Power Drift = —0.04 dB
        Peak SAR (extrapolated) = 16.8 W/kg
        SAR(I g) =9.21 W/kg; SAR(10 g) = 4.89 W/kg
        Maximum value of SAR (measured) = 14.2 W/kg

           dB
           0


            —3.26


            —6.52


            —9.77


           13.03


           16.29
                       0 dB = 14.2 W/kg = 11.52 dBW/kg




Certificate No: Z19—60084                            Page 7 of8


           !\N                       In Collaboration with

           -\/
                   TTIL a
                        CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com       hitp://www.chinattl.cn


         Impedance Measurement Plot for Body TSL


      Tri si1 Log Hay 10.00ds7 ref 0. ooode (ri]
        5990 53 —1.7500000 onz =23.18¢ ds
        40. 00

        30.00
        20.00
        10. 00
        0.000)}
       —10.00                                                                                 A
       ~20.00
       —30.00
       ~40.00

       ~50.00                                                       a                    —‘
     BMR s11 smith (R+J®) scale 1.000U [EX oe1]
       >   1.7500000 ahz.   43.873 n —2.1946 o    41.4947(\




                                                    .                  e
                                                          _4
    [1 sumise oz                                          wow 100 he         Ston 11655 oc IRal




Certificate No: Z19—60084                            Page 8 of 8


                      2e 5                   in Collsboration with                  ‘\\\‘&j"]o’       _i FBWAH
                  &77J, s p e a g                                                  s              *                     AREX
                 V                       CALIBRATION LABORATORY                      Af& CNAS Botk
                                                                                              CALIBRATION
                                                                                 5
               Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China 3777
               Tel:   +86—10—62304633—2079       Fax: +86—10—62304633—2504             "m“                              CNAS LO570
               E—mail: ett@chinattl.com          http=//www.chinattl.en

         Client              Sporton                                            Certificate No:       Z19—60085

CALIBRATION CERTIFICATE

Object                                        D1900V2 — SN: 5d170

Calibration Procedure(s)                      FF—211—003—01

                                              Calibration Procedures for dipole validation kits

Calibration date:                             March 26, 2019

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)C and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)


Primary Standards                   ID #                 Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 Power Meter          NRP2          106277              20—Aug—18 (CTTL, No.J18X06862)                            Aug—19
 Power sensor          NRPBS        104291              20—Aug—18 (CTTL, No.J18X06862)                            Aug—19
 Reference Probe EX3DV4             SN 3617             31—Jan—19(SPEAG,No.EX3—3617_Jan19)                        Jan—20
 DAE4                               SN 1331             06—Feb—19(SPEAG,No.DAE4—1331_Feb19)                       Feb—20


 Secondary Standards                ID #                Cal Date(Calibrated by, Certificate No.)         Scheduled Calibration
 Signal Generator E4438C            MY49071430           23—Jan—19 (CTTL, No.J19X00336)                           Jan—20
 NetworkAnalyzer ES071C             MY46110673           24—Jan—19 (CTTL, No.J19K00547)                           Jan—20



                                    Name                             Function                                Signature
Calibrated by:                     Zhao Jing                   SAR Test Engineer                                  /{,

Reviewed by:                       Lin Hao                     SAR Test Engineer                           #{4
Approved by:                       Qi Dianyuan                 SAR Project Leader                          Bs_
                                                                                         Issued: March 29, 2019
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z19—60085                                  Page 1 of 8


            m:                      In Collaboration with

          a=‘/"/"J, a
                                    CALIBRATION LABORATORY
           ‘/
         Add: No.51 XueyuanRoad, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
         E—mail: ct@chinattl.com         hitp://www.chinattl.cn


 lossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y,z
 N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
* Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
«_    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
*     Feed Point Impedance and Return Loss: These parameters are measured with the dipole
      positioned underthe liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to the feed point. The Return Loss ensures low
      reflected power. No uncertainty required.
*     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
e     SAR measured: SAR measured at the stated antenna input power.
«_    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
&     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.




Certificate No: Z19—60085                              Page 2 of 8


           Ti ies
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
              86—10—62304633—2079        Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com       http://www.chinattl.en


 Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                         DASY52                                   52.10.2.1495

      Extrapolation                              Advanced Extrapolation

      Phantom                                     Triple Flat Phantom 5.1C

      Distance Dipole Center — TSL                         10 mm                                     with Spacer

     Zoom Scan Resolution                            dx, dy, dz = 5 mm

     Frequency                                       1900 MHz + 1 MHz


Head TSL parameters
   The following parameters and calculations were applied.
                                                           Temperature              Permittivity          Conductivity
     Nominal Head TSL parameters                              22.0 °C                   40.0              1.40 mho/m

     Measured Head TSL parameters                         (22.0 £.0.2) °C            40.5 x 6 %         1.44 mhom + 6 %

     Head TSL temperature change during test                  <1.0 °C                    sus                  se

SAR result with Head TSL
     SAR averaged over 1_C/7"_(1 g) of Head TSL                         Condition
     SAR measured                                                  250 mW input power                  9.90 Wikg

     SAR for nominal Head TSL parameters                             normalized to 1W            39.0 Wikg £ 18.8 % (k=2)

     SAR averaged over 10 Ci2" (10 g) of Head TSL                       Condition
     SAR measured                                                  250 mW input power                  5.12 Wikg

     SAR for nominal Head TSL parameters                             normalized to 1W            20.3 Wikg & 18.7 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                           Temperature              Permittivity         Conductivity

     Nominal Body TSL parameters                              22.0 °C                  53.3              1.52 mho/m

     Measured Body TSL parameters                         (22.0 £ 0.2) °C            54.5 46 %         1.56 mhom + 6 %

     Body TSL temperature change during test                  <1.0 °C                   ——                   ——
SAR result with Body TSL
     SAR averaged over 1_Ci7°_(1 g) of Body TSL                         Condition
     SAR measured                                                  250 mW input power                  10.1 Wikg

     SAR for nominal Body TSL parameters                             normalized to 1W          40.0 Wikg £ 18.8 % (k=2)

     SAR averaged over 10 Cmm (10 g) of Body TSL                        Condition
     SAR measured                                                  250 mW input power                 5.28 Wikg

     SAR for nominal Body TSL parameters                             normalized to 1W          21.0 Wikg £ 18.7 % (k=2)

Certificate No: 19—60085                               Page 3 of 8


           fi                         I;Collubomtionewith a


                                     CALIBRATION LABORATORY

         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com       hitp://www.chinattl.en


  Appendix (Additional assessments outside the scope of CNAS LO570)

  Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                   51.70+ 6.73j0

      Return Loss                                                              ~23.30B


 Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                   47.80+ 6.72)0
      Return Loss                                                              —22.80B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                          1.066 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm ofthe dipole. The antennais therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole armsin order to improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole lengthis still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend orthe soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: Z19—60085                              Page 4 of 8


           b‘                       In Collaboration with
          e=x‘/"]7"J, a
           V                        CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: ett@chinattl.com        hitp://www.chinattl.en


DASY5 Validation Report for Head TSL                                     Date: 03.26.2019
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50170
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1:1
   Medium parameters used: {= 1900 MHz; 0 = 1.441 S/m; s, = 40.48; p = 1000 kg/m3
   Phantom section: Center Section
DASYS Configuration:

            *    Probe: EX3DV4 — SN3617; ConyF(8.14, 8.14, 8.14) @ 1900 MHz; Calibrated:
                   1/31/2019
            *    Sensor—Surface; 1.4mm (Mechanical Surface Detection)
            +    Electronics: DAE4 Sn1331; Calibrated: 2/6/2019
            *«   Phantom: MFP_V5.1C ; Type: QD 000 P51CA; Serial: 1062
            *«   Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12
                 (7450)
        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid:
        dx=Smm, dy=5mm, dz=5mm
        Reference Value = 97.54 V/m; Power Drift = 0.08 dB
        Peak SAR (extrapolated) = 18.9 W/kg
        SAR(I g) = 9.9 W/kg; SAR(10 g) = 5.12 W/kg
        Maximum value of SAR (measured) = 15.6 W/kg

           dB
           0


           —3.57


           T13


           10.70


           14.26


           17.83
                      0 dB = 15.6 W/kg = 11.93 dBW/kg




Certificate No: 19—60085                              Page 5 of 8


                ziies
              Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
              Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
              E—mail: cttl@chinatl.com        hitp://www.chinattl.en




              Impedance Measurement Plot for Head TSL


      Tri siT tog mag 10. 00dey nef 0. ooods (F1]
        590 r531.3000000 onz =23.342 do
        40. oo
         30. 00

         20.00
        10.00

        o.oonJ                                                                                a
       —10. 00

       ~20.00

       ~30.00
       ~40. 00
       ~50.00                                                              x
     MB si1 smith (RtJ>) scale 1.000u [r1. pe1]
       »1..     1.9000000 aiz    S1.689 n   6.7277 n   S63.55—pi_




    T Suti76%                                                  iew 100 e          Sten 211 Ga B




Certificate No: Z19—60085                                   Page 6 of 8


           !\"                      In Collaboration with
          a_‘/"/"J, a
          \/        CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com       http=//www.chinattl.en




DASYS Validation Report for Body TSL                                  Date: 03.26.2019
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN: 50170
   Communication System: UID 0, CW; Frequency: 1900 MHz; Duty Cycle: 1: 1
    Medium parameters used: {= 1900 MHz; 0 = 1.56 $/m; &, = 54.52; p = 1000 kg/m3
  Phantom section: Right Section
DASYS Configuration:

            *    Probe: EX3DV4 — SN3617; ConvF(7.78, 7.78, 7.78) @ 1900 MHz; Calibrated:
                 1/31/2019
            *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +    Electronics: DAE4 Sn1331; Calibrated: 2/6/2019
            *«   Phantom: MFP_V5.1C ; Type: QD 000 P51CA; Serial: 1062
            +    Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12
                 (7450)

        System Performance Check/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Mcasurement grid:
        dx=5mm, dy=5mm, dz=5mm
        Reference Value = 95.48 V/m; Power Drift = 0.04 dB
        Peak SAR (extrapolated) = 18.6 W/kg
        SAR(I g) = 10.1 W/kg; SAR(10 g) = 5.28 W/kg
        Maximum value of SAR (measured) = 15.7 W/kg

           dB
           0


           —3.45


           6.90


           10.36


           13.81

                         C
           17.26                                  asimmmmnnmmiamenescarsaccucmecass
                      0 dB = 15.7 W/kg = 11.96 dBW/kg




Certificate No: Z19—60085                             Page 7 of 8


           r\*                       in Collaboration with
                         L_L_GC_54..9........
                   T‘TZ, CALIBRATION
           !\,                       LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Pax: +86—10—62304633—2504
         E—mail: cttl@chinatt.com       hitp://www.chinattl.en


         Impedance Measurement Plot for Body TSL


      Tri iT Log nay 15.00de7 ner 0. oode []
        5990 5t ~1.5000000 onz =22. 841. db
       40. 00

        30. 00

       20.00
       20. 00
       0.000))                                                                              «
      —10.00                                                                                f—
      —20.00 \\/q//
      ~30.00



                                                                       k           —
                 smith (RHJX) scale 1.000U [F2. vel]
       »2 1.9000000 anz 47. 796 n 6. 7152 n 562.50—pi




    1 surti7ot                                            wew 100 he         stop 21 one IRai




Certificate No: Z19—60085                              Page 8 of 8


                  !\n                       InColleboration with                  &\\‘L““/"m,, A * BWaT
                                                                                               CA A EinA
                                                                       lChina
                Adt: No.51 Xueyuan Rond. Haidian District, Beljimp, 100194        %/%zany
                                                                                      NN¥                      sgfimou
                Tel: +86—10—62304633—2070       Fas +#6—10—62300633.2504           "/,,m‘\\“\                         CNAS LO570
                E—mail ttl@chinattL.com         hitps/iwwn.chinatl.on
         Client           Sporton                                             Certificate No:       21860365
 s      AtIEs        ¥                                                                          —
I—                                                                            S                                   a           3

 Object                                       ©2300V2 — SN: 1058

 Calbration Procedure(s)                     EF—Z11—00301

                                             Calibration Procedures for dipole validation kits

 Calibration date:                            September 20, 2018


 This calibration Certificate documents the traceability to national standards, which realize the physical units of
 measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
 pages and are part of the certificate.


 All callbrations have been conducted in the closed laboratory facility: environment temperature@22s3)© and
 humidity<70%.


 Calbration Equipment used (M&TE critical for callbration)


 Primary Standards                  1D #                Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
     Power Meter NRVD               102083             01—Nov—17 (CTTL, No.J17X08756)                          Oct—18
     Power sensor NRV—Z5            100542             01—Nov—17 (CTTL, No.J17X08756)                          Oct18
     Reference Probe EX3DV4         SN 7514            27—Aug—18(SPEAG,No.EX3—7514_Aug18)                      Aug—19
     DAE4                           SN 1555            20—Aug—18(SPEAGNo.DAE4—1555_Aug18)                      Aug—19

     Secondary Standards            ID #               Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
     Signal Generator E4438C        MY49071430          23—Jan—18 (CTTL, No.118X00560)                         Jan—19
     NetworkAnalyzer ESO71C         MY46110873         24—Jan—18 (CTTL, No.J18X00561)                          Jan—19



                                     Name                          Function                                Signature
 Callbrated by:                     Zheo Jing                 SAR Test Engineer                            /gA

 Reviewed by:                       Lin Hao                    SAR Test Engineer                     > ‘fifj{f"'                    |
 Approved by:                       Oi Dianyan                 SAR Project Leader                     e~
                                                                                  Issued: September 23, 2018
 This callbration certificate shall not be reproduced except in full without written approval of the laboratory.



        Certificate No: Z18—60365                             Page 1 of 8


          é TL&S_
                                     Distict, Beljing. 100191, China
                       633—2070    Fac: +86—10—62304633—2504
                     inattl.com    Aipotiwwchinaten


                        tissue simulating liquid
                        sensitivity in TSL / NORMx.y,z
                       not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
     Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
     communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
     6GHz)‘, July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDBS65664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
& Measurement Conditions: Further details are available from the Validation Report at the end
   of the certificate. All figures stated in the certificate are valid at the frequency indicated.
&     Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented.
      paralle! to the body axis.
«—    Feed Point Impedance and Retum Loss: These parameters are measured with the dipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurement at the SMA connector to thefeed point. The Return Loss ensures low
      reflected power. No uncertainty required.
*     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
&     SAR measured: SAR measured at the stated antenna input power.
«—    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
&     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result,


     The reported uncertainty of measurement is stated as the standard uncenaimy_of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.




Certificate No: 718—60365                       Page 2 of 8


          r\‘                       in Colleboration with
              a
          ‘,L Callsranon uaporkrory
        Add: No.51 Xueyan Rood, HaidianDistrict,Beijing, 100191, China
        Tel: +86—10—62304033—2079     Fax: «6—10.02304603.2504
        R—mail:eti@chinant.com        hitgsfwwchinatLen

Measurement Conditions
  system confiquration. as far as not given on page 1
     DASY Version                                      DASY52                                         52.10.1.1476        7
     Extrapolation                             Advanced Extrapolation
     Phantom                                   Triple Flat Phanitom 5.1C
     Distance Dipole Center — TSL        (             10 mm                                       with Spacer            —'
     Zoom Scan Resolution                         dx, dy, dz =5 mm                 "
   [Froqulncy                                     2300 MHz + 1 Niz               I
Head TSL parameters
   The following parameters and calculations were applied
                                                        Temperature              Pormittivity              Conductivity
     Nominal Head TSL parameters                            220°C                      39.5                 1.87 mhoim
     Measured Head TSL parameters                     (@2.0202)°C                 40.3 26 %             l 1.72 mhoim 28 %
   Eiud TSL temperature change during test                  «10 °C                      ~~
SAR result with Head TSL
   rsm averaged over 1_cnt‘_(1 g) of Head TSL                        Condition                T
                                                               250 mW input power                       12.3 mW /g

     SAR for nominal Head TSL parameters                         normalized to 1W                 48.7 mW 1g2 18.8 % (k=2)

     SAR averaged over 10 cm"_(10 g) of Hoad TSL |                   Condition
     SAR measured                                              250 mW input power                       5.82 mW 1g

     SAR for nominal Head TSL parameters                         normalized to 1W                 23.2 mW 1g x 18.7 % (ke2)
Body TSL parameters
  ‘The following parameters and calculations were applied.




     Body TSL temperature change during tost
SAR result with Body TSL
   [ SAR averaged over 1_cn"_(1 g) of Body TSL                       Gondition
     SAR measured                                                250 mW input power                     11. mW i9

     SAR for nominal Body TSL parameters                          normalized to 1W                47.6 mW 1g £ 18.8 % (k=2)

     SAR averaged over 10 cnt‘_(10 g) of Body TSL                    Gondition                |                           7|
     SAR measured                                                250 mW input power           I         5.04 mW1 g
    2
     SAR for nominal Body TSL parameters                          normalized to 1W



Cortificate No: Z18—6036$                          Page 3 of 8


               TTL                 in Collaboration with


      Add: No.51 Xueyun Road, Haidian District, Beling, 100191, China
      Tel: +86—10—62304633—2070       Fax: +86—10—62304633—2504
      E—mail: ctl@zchinatt.com        hi(/wwechinattlen


Appendix (Additional assessments outside the scope of CNAS LO570)

Antenna Parameters with Head TSL

   Impedance, transformed to feed point                                 45.60—2.030

  LRell.ll’n Loss                                                         —25.90B


Antenna Parameters with Body TSL

  I Impedance, transformed to feed point                                44.00— 0.85i0
  rfielum Loss                                                             —23.80B

General Antenna Parameters and Design

   Electrical Delay (one direction)                         \             1,034 ns



After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
connected to the second arm of the dipole. The antennais therefore short—circuited for DC—signals. On some
of the dipoles, small end caps are added to the dipole arms in order to improve matching when loaded
according to theposition as explained in the "MeasurementConditions" paragraph. The SAR data are not
affected by this change. The overall dipole length is stil according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered
connections near the feedpoint may be damaged.



Additional EUT Data

   [Menufactured by                                             [             sreac




Certificate No: 218—60365                           Page 4 ofs


         fi                          \;Callfiwavonewith 3



       Add: No.51 Xueywan Road. Haidian Distct, Beijing. 100191, China
       Tels«86—10—62304633—2070       Faxc «86—10—2304623.2504
       F—mail;ctl@chinat.com          iitpWwchinaten

DASYS Validation Report for Head TSL                                    Date: 09.20.2018
Test Laboratory: CTTL, Beijing. China
DUT: Dipole 2300 MHz; Type: D2300V2; Seri        2300V2 — SN: 1055
   Communication System: UID 0, CW; Frequen        00 MHz; Duty Cycle: 1:1
   Medium parameters used: f= 2300 MHz; a = 1.718 $/m; e;= 40.3; p = 1000 kg/m3
  Phantom section: CenterSection
DASY5 Configuration:

           *    Probe: EX3DV4 — SN7514; ConvF(7.42, 7.42, 7.42) @ 2300 MHz; Calibrated:
                $/727/72018
           *    Sensor—Surface: 1.4mm(Mechanical Surface Detection)
           «+   Electronics: DA      1555; Calibrated: 8/20/2018
           *    Phantom: MFP_V5.1C ; Type: QD 000 PS1CA; Serial: 1062
           +    Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11
                (7439)

       Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/(Cube 0: Measurement grid: dx=3mm.
       dy=Smm, dz=3mm
       Reference Value = 97.59 V/m; Power Drift=—0.01 dB
       Peak SAR (extrapolated) = 25.4 W/ke
       SAR(I g) = 12.3 W/kg; SAR(IO g) = 5.82 Wikg
       Maximum value ofSAR (measured) = 20.4 W/ke

          dB
          0


           418


           —8.36


           12.53


           16.71


           —20.09
                      0 dB =20.4 W/kg = 13.10 dBW/kg




Certificate No: Z18—60365                           Page 5 of 8


       Add: No.51 Xueyuan Road, Haidian Distict, Beijing, 100191, China
       "el: +80—10—62304633—2070     Faxc +86—10—62304633—2504
       E—mail: ettl@chinatt.com        hitp?wwwchinattl.cn




       Impedance MeasurementPlot for Head TSL


     ¥er Sit toy wag 16.00807     ses.cood(FL]
      seie m s soonn ons               ses
      40.00
      30.00
      20.00
      10.00
      .000
    —20.00
    20.00
    —30.00
    »s0.00
    ~so.c0
   MBR 513. sntzh (rags) scate 2.0000 [E oel]
     sr. 223000000 ouz. 45.598 a —a.om0 n m.o5tern ~.




Certificate No: Z18—60365                            Page 6 of8


         GTi ¥%as                    CAIBRATION LaBORATORY
       Add: No.51 Xueyuan Raad, Haidian Disteict, Reljing. 100191, China
       ‘Tel +R6—10—62304633—2070      Faxc +86—10—6230463,
       E—mail ctl@chinatl.com          hupiliwwchinaitlc

DASYS Validation Report for Body TSL                                                     Date: 09.20,2018
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2300 MHz; Type: D2300V2; Serial: D2300V2— SN: 1055
  Communication System: UID 0, CW; Frequency:                     2300 MHz:; DutyCycle: 1:1
  Medium parameters used: (= 2300 MHz; a= 1.834 S/m; sr= 54.44; p = 1000 kg/m3
  Phantom section: Right Section
DASYS Configuration:

           *    Probe: EX3DV4 — SN7S14; ConvF(7.25, 7.25, 7.25) @ 2300 MHz: Calibrated:
                $272018
           +    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
           *    Electronics: DAE4 Sn1555; Calibrated: 8/20/2018
           *    Phantorn: MFP_V5.1C ; Type: QD 000 PS1CA; Serial: 1062
           +    Measurement SW: DASY52, Version 52.10 (1); SEMCAD X Version 14.6.11
                (7439)

       Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=3mm.
       dy=Smm, dz=Smm
       Reference Value = 97.37 V/m; Power Drift=—0.01 dB
       Peak SAR (extrapolated) = 23.9 Wikg
       SAR(I g) =11.9 W/kg; SAR(LO g) = 5.64 Wikg
       Maximum value of SAR (measured) = 19.4 W/ke

          da
          0

           413

           —8.26


           —12.39

           418.52


           —20.65                                  =z                          mm w es 4
                      0 dB =19.4 W/kg = 12.88 dBW/kg




Certificate No: 718—60365                            Page 7 of8


        !\                          in Collaboration with
         iTTL
            sn —e—2_4__.
       Add: No.51 Xueyuan Road, Haidian District. Bejing. 100191, China
       Tel: +86—10—62304633—2079      Faxc+86—10.62304633.2504
       E—mail: ctl@chinat.com         hiptwwwchinattlen

       Impedance Measurement Plot for Body TSL


   [Te Ei toe mas 15. 0087 ner o. sonds [ey
     S0.00
        10,00
        s0.00
        20.00
       20.00
        o.c0o
      —20.00
      —20.00
      —20.00
      —10.00
      —so.00
     1BB si2 sntth 140 Scale 2.0000 en oet]
       on 2.3000000 onz ac.oro o ~ess.e2 mo 80.8i_




Certificate No: 218—60365                           Page 8 of 8


                  !\“                       in Collaboration with                     s\\“I"u      A EYA
                    ‘/"/"‘/, CSA_HL?_LS_ \\/ CNA                                                                    E4WA




                                                                                   8 wl
               Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China     E/—\g                     sCALIBRATION




                                                                                     i
               Tel: +86—10—62304633—2079        Fax: +86—10—62304633—2504                 '@“\e    v CNAS LO570
                E—mail: cttl@chinattl.com       hitp://www.chinattl.en

       Client               Sporton                                            Certificate No:     219—60087

CALIBRATION CERTIFICATE

Object                                        D2450V2 — SN: 908

Calibration Procedure(s)                      FF—211—003—01

                                              Calibration Procedures for dipole validation kits

Calibration date:                             March 25, 2019

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22«3)C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)


Primary Standards                    ID #               Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 Power Meter         NRP2            106277            20—Aug—18 (CTTL, No.J18X06862)                          Aug—19
 Power sensor        NRPBS           104291            20—Aug—18 (CTTL, No.J18X06862)                          Aug—19
 Reference Probe EX3DV4              SN 3617           31—Jan—19(SPEAG,No.EX3—3617_Jan19)                      Jan—20
 DAE4                                SN 1331           06—Feb—19(SPEAG,No.DAE4—1331_Feb19)                     Feb—20

 Secondary Standards                 ID #              Cal Date(Calibrated by, Certificate No.)       Scheduled Calibration
 Signal Generator E4438C             MY49071430         23—Jan—19 (CTTL, No.J19X00336)                         Jan—20
 NetworkAnalyzer E5071C              MY46110673        24—Jan—19 (CTTL, No.J19X00547)                          Jan—20



                                     Name                           Function                              Signature
Calibrated by:                     Zhao Jing                  SAR Test Engineer                           VZ\ i/

Reviewed by:                        Lin Hao                   SAR Test Engineer                         Tflfi -7}{7

Approved by:                       Qi Dianyuan                SAR Project Leader                      _éé’./fi__/

                                                                                 Issued: March 28, 2019
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z19—60087                                 Page 1 of 8


        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       hitp://www.chinattl.on


Glossary:
TSL                       tissue simulating liquid
ConvF                      sensitivity in TSL / NORMx,y,z
N/A                       not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
   Communications Devices: Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for assessmentof specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
   communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
   6GHz)", July 2016
c) IEC 62209—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to 6GHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
*    Measurement Conditions: Further details are available from the Validation Report at the end
     of the certificate. All figures stated in the certificate are valid at the frequency indicated.
&    Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
     point exactly below the center marking of the flat phantom section, with the arms oriented
     parallel to the body axis.
*    Feed Point Impedance and Return Loss: These parameters are measured with the dipole
     positioned under the liquid filled phantom. The impedance stated is transformed from the
     measurement at the SMA connector to the feed point. The Return Loss ensures low
     reflected power. No uncertainty required.
e    Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
     No uncertainty required.
*    SAR measured: SAR measured at the stated antenna input power.
*    SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
     connector.
&    SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
     nominal SAR result.


    The reported uncertainty of measurement is stated as the standard uncertainty of
    Measurement multiplied by the coverage factor k=2, which for a normal distribution
    Corresponds to a coverage probability of approximately 95%.




Certificate No: Z19—60087                            Page 2 of 8


           mm:                         in Collaboration with
           _/"7"‘/                     s      p_      e        a      q
          -\/                          CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com      hitp://www.chinattl.en


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                        DASY52                                     52.10,2.1495

      Extrapolation                                Advanced Extrapolation
      Phantom                                      Triple Flat Phantom 5.1C

      Distance Dipole Center— TSL                         10 mm                                       with Spacer

     Zoom Scan Resolution                             dx, dy, dz = 5 mm

     Frequency                                        2450 MHz +1 MHz

Head TSL parameters
   The following parameters and calculations were applied.
                                                           Temperature               Permittivity          Conductivity

     Nominal Head TSL parameters                               22.0 °C                  39.2               1.80 mho/m

     Measured Head TSL parameters                         (22.0 £0.2) °C              39.6 1 6 %         1.84 mho/m £ 6 %

     Head TSL temperature change during test                   <1.0 °C                   —                     ——
SAR result with Head TSL
     SAR averaged over 1 _Cm"_(1 g) of Head TSL                          Condition
     SAR measured                                                   250 mW input power                   13.3 Wikg
     SAR for nominal Head TSL parameters                             normalized to 1W             52.8 Wikg + 18.8 % (k=2)
     SAR averaged over 10 cm‘ (10 g) of Head TSL                         Condition

     SAR measured                                                   250 mW input power                  6.07 Wikg
     SAR for nominal Head TSL parameters                             normalized to 1W             24.2 Wikg # 18.7 % (k=2)
Body TSL parameters
   The following parameters and calculations were applied.
                                                           Temperature               Permittivity         Conductivity

     Nominal Body TSL parameters                               22.0 °C                  52.7               1.95 mho/m
     Measured Body TSL parameters                         (22.0 £0.2) °C              53.8 16 %         2.00 mho/m + 6 %

     Body TSL temperature change during test                   <1.0 °C                   ——                    ———
SAR result with Body TSL
   SAR averaged over 1_cm_(1 g) of Body TSL                             Condition
   SAR measured                                                     250 mW input power                  12.8 Wikg
     SAR for nominal Body TSL parameters                             normalized to 1W           50.8 Wikg £ 18.8 % (k=2)

     SAR averaged over 10. cm" (10 g) of Body TSL                        Condition
     SAR measured                                                   250 mW input power                  5.91 Whg

     SAR for nominal Body TSL parameters                            normalized to 1W            23.6 Wikg £ 18.7 % (k=2)

Certificate No: Z19—60087                             Page 3 of 8


           o8                           in Collaboration with
          =TTL                          2
                                        CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fax: +86—10—62304633—2504
            ail: cttl@chinattl.com        http://www.chinattl.cn


  Appendix (Additional assessments outside the scope of CNAS L0570)

 Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                   57.30+ 5.18 jQ

      Return Loss                                                              —21.60B


 Antenna Parameters with Body TSL

     Impedance, transformed to feed point                                    52.60+ 5.81 jQ

     Return Loss                                                               —24.10B


 General Antenna Parameters and Design

     Electrical Delay (one direction)                                          1.020 ns




 After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can
 be measured.

 The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly
 connected to the second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some
 of the dipoles, small end caps are added to the dipole arms in orderto improve matching when loaded
 according to the position as explained in the "Measurement Conditions" paragraph. The SAR data are not
 affected by this change. The overall dipole length is still according to the Standard.
 No excessive force must be applied to the dipole arms, because they might bend or the soldered
 connections near the feedpoint may be damaged.



 Additional EUT Data

     Manufactured by                                                             SPEAG




Certificate No: 719—60087                              Page 4 of 8


           filil                       in Collaboration with

                                      a
           zen                        CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079      Fax: +86—10—62304633—2504
        E—mail: cttl@chinatt.com       hitp://www.chinattl.en

DASY5 Validation Report for Head TSL                                     Date: 03.25.2019
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 908
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1:1
   Medium parameters used: {= 2450 MHz; 0 = 1.841 S/m; &, = 39.63; p = 1000 kg/m3
  Phantom section: Right Section
DASYS Configuration:

            *    Probe: EX3DV4 — SN3617; ConvF(7.62, 7.62, 7.62) @ 2450 MHz; Calibrated:
                 1/31/2019
            *«   Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +    Electronics: DAE4 Sn1331; Calibrated: 2/6/2019
            *«   Phantom: MFP_V5.1C ; Type: QD 000 P51CA; Serial: 1062
            *«   Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12
                 (7450)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=Smm, dz=5mm
        Reference Value = 96.04 V/m; Power Drift = 0.04 dB
        Peak SAR (extrapolated) = 28.3 W/kg
        SAR(I g) = 13.3 W/kg; SAR(10 g) = 6.07 W/kg
        Maximum value of SAR (measured) =22.4 W/kg

           dB
           0


           ~4.59


           —9.18


           113.77


           18.36


           —22.95
                      0 dB =22.4 W/kg = 13.50 dBW/kg




Certificate No: Z19—60087                            Page 5 of 8


          Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
          Tel: +86—10—62304633—2079     Fax: +86—10—62304633—2504
          E—mail: cttl@chinattl.com          hitp://www.chinattl.en




          Impedance Measurement Plot for Head TSL


      Tri i1 Log Mag 10.00d8/         ref 0.000d8 [F1]
         so. oo    >1   2.4500000 anz ~21.595 do
        40. 00

         30.00
         20.00
        10.00

        0.0009                                                                          (€
       —10.00                                                                           h
       ~20.00

       ~30. 00
       ~40.00
       ~50.00                                                              A
    INIRR si1 smith (Ref>) scate 1.000u (e1. oel]
        »1.. 24500000 cnz      57,283 o.   5.1842 o   336.78—pf
                                                         .




    [1: Start 225 Gte                                        irew 100 He       fimflnseullfil




Certificate No: 19—60087                                 Page 6 of 8


                                      In Collaboration with

                 TTL CALIBRATION
                     a
            \ppzame~             LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2079       Fasc: +86—10—62304633—2504
        E—mail: ettl@chinattl.com       hitp=//www.chinattl.en


DASYS Validation Report for Body TSL                                     Date: 03.25.2019
Test Laboratory: CTTL, Beijing, China
DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN: 908
   Communication System: UID 0, CW; Frequency: 2450 MHz; Duty Cycle: 1: 1
   Medium parameters used: £= 2450 MHz; 0 = 2.003 S/m; &, = 53.78; p = 1000 kg/m3
   Phantom section: Center Section
DASYS Configuration:

            *    Probe: EX3DV4 — SN3617; ConvF(7.79, 7.79, 7.79) @ 2450 MHz; Calibrated:
                 1/31/2019
            *    Sensor—Surface: 1.4mm (Mechanical Surface Detection)
            +    Electronics: DAE4 Sn1331; Calibrated: 2/6/2019
            +    Phantom: MFP_V5.1C ; Type: QD 000 P51CA; Serial: 1062
            *    Measurement SW: DASY52, Version 52.10 (2); SEMCAD X Version 14.6.12
                 (7450)

        Dipole Calibration/Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm,
        dy=5mm, dz=5mm
        Reference Value = 95.51 V/m; Power Drift = —0.02 dB
        Peak SAR (extrapolated) =27.1 W/kg
        SAR(I g) = 12.8 W/kg; SAR(10 g) = 5.91 W/kg
        Maximum value of SAR (measured) =21.4 W/kg

           dB
           0


           445


           —8.90


           13.34


           417.79

                         L.                       T
           22.24
                      0 dB =21.4 W/kg = 13.30 dBW/kg




Certificate No: 719—60087                            Page 7 of 8


           fi\n“;                         ;Collaboration with a


         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2079         Fax: +86—10—62304633—2504
         E—mail: cttl@chinattl.com         hitp://www.chinattl.on

         Impedance Measurement Plot for Body TSL


      Ti SiT tog hay 10. ooda7 ner 5:0008 [FLY
       5909. sa—2.2500000 onz =24.108 ds
        40. 00
        30.00

        20.00

        10.00
        0.000))                                                                             C
       —10.00

       ~20.00

       ~30.00
       ~40. 00
       —50.00                        —                —                a
     MBR 511 smith (ReJ®) scale 10000 [e1 oe1]
       »a 2.4500000 onz      52.615 0.   5.8142 o 377.}04( . mss




    (1 sur 225 oc                                         IFW 100 He         Sten 2.65 Gic TR




Certificate No: Z19—60087                             Page 8 of 8


                                                                                   U hm #B%8
               2‘
            Add: No.51 Xueyuan Road, Haidian Distrct, Bejing. 100191, Chine
                                                                              EsmCNas®""
                                                                                  ‘-,,fi\;                        CALIBRATION
            Tel: +86—10—62304633—2079      Faxc «86—10—62304633—2504               Yahsteit®                     CNAS LO670
             E—mail: etl@chinat.com          hips/fwwnchinatn
      Client           Sporton                                           Certificate No:         218—60490
                                                                              i    \V‘

                                                                          esAE                         ty s hib xB
Object                                   ©2600V2 — SN: 1061

Calibration Procedure(s)                 FEZ11—003—01

                                         Calibration Procedures for dipole validation kits

Calbration date:                         December 7, 2018

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S!), The measurements and the uncertainties with confidenceprobability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(@2«3)C and
humidity<70%.

Calibration Equipment used (M&TE eritical for callbration)


Primary Standards                                   Cal Date(Calibrated by, Certificate No.)          Scheduled Calibration
 Power Meter NRVD                102196            07—Mar—18 (CTTL, No.J18X01510)                            Mar—19
 Power sensor      NRV—Z5        100596            07—Mar—18 (CTTL, No.J18X01510)                            Mar—19
 Reference Probe EX3DV4          SN 7514           27—Aug—18(SPEAGNo.2X3—7514_Aug18)                         Aug—19
 Dak4                            SN 1555           20—Aug—18(SPEAG,No.DAE4—1555_Aug18)                       Aug—18

 Secondary Standards             D#                Cal Date(Calibrated by, Certificate No.)           Scheduled Calibration
 Signal Generator E4438C         MY49071430 23—Jan—18 (CTTL, No.J18X00560)                                    Jan—19
 Network Analyzer E5071C         MY46110673.— 24—Jan—18 (CTTL, No.J18X00561)                                  Jan—19


                                  Name                        Function                                     Signature
Calibrated by:                   Zhao Jing                SAR Test Engineer                                 €1,,
Reviewed by:                     Lin Hao                  SAR Test Engineer                               ‘FFQ:?%
Approved by:                     @Sieryien                SAR Project Leader                        fi%\/
                                                                                         Issued: December 10, 2018
This callbration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: Z18—60490                            Page 1 of 8


          Tt $Ea s
        Add: No.51 Xueyuan Road, Heidian District, Beiing, 100191, China
          el: +86—10—62304633—2070    Fanc +§6—10—62304633—2504
             it ctl@chinatcom           hitps(iwrichinatten

Glossary:
TSL                       tissue simulating liquid
ConvF                     sensitivity in TSL / NORMx,y,z
N/A                      not applicable or not measured


Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak
   Spatial—Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
     Communications Devices: Measurement Techniques‘, June 2013
b) IEC 62209—1, "Measurement procedure for assessment of specific absorption rate of human
   exposure to radio frequency fields from hand—held and body—mounted wireless
     communication devices— Part 1: Device used next to the ear (Frequency range of 300MHz to
     6GHz)", July 2016
c) IEC 62208—2, "Procedure to measure the Specific Absorption Rate (SAR) For wireless
   communication devices used in close proximity to the human body (frequency range of
   30MHz to GGHz)", March 2010
d) KDB865664, SAR Measurement Requirements for 100 MHz to 6 GHz

Additional Documentation:
e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
+     Measurement Conditions: Furtherdetails are available from the Validation Report at the end
      of the certificate. All figures stated in the certificate are valid at thefrequency indicated.
«—    Antenna Parameters with TSL: The dipole is mounted with the spacerto position its feed
      point exactly below the center marking of the flat phantom section, with the arms oriented
      parallel to the body axis.
*«    Feed Point Impedance and Return Loss: These parameters are measured with thedipole
      positioned under the liquid filled phantom. The impedance stated is transformed from the
      measurementat the SMA connector to the feedpoint. The Return Loss ensures low
      reflected power. No uncertainty required.
*     Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
      No uncertainty required.
&     SAR measured: SAR measured at thestated antenna input power.
+     SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
      connector.
&     SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
      nominal SAR result.


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.




Certificate No: Z18—60490                             Page 2 of8



Document Created: 2019-05-29 19:33:08
Document Modified: 2019-05-29 19:33:08

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC