HAC RF Appendix A Test Report

FCC ID: IHDT56XS2

Test Report

Download: PDF
FCCID_4190735

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2019/1/11

HAC_E_Dipole_835

DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
                    23.5

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1) @ 835 MHz; Calibrated: 2018/1/19
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 138.0 V/m; Power Drift = -0.05 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 119.6 V/m
Average value of Total=(109.4+119.6) / 2 = 114.5 V/m

                                          PMF scaled E-field
                                           Grid 1 M4 Grid 2 M4 Grid 3 M4
                                           105.7 V/m 109.4 V/m 108.8 V/m
                                           Grid 4 M4 Grid 5 M4 Grid 6 M4
                                           66.55 V/m 67.49 V/m 66.44 V/m
                                           Grid 7 M4 Grid 8 M4 Grid 9 M4
                                           119.0 V/m 119.6 V/m 116.1 V/m
                                                         Cursor:
                                                    Total = 119.6 V/m
                                                     E Category: M4
                                                 Location: 1.5, 72, 9.7 mm




                      0 dB = 119.6 V/m = 41.55 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2019/1/11

HAC_E_Dipole_1880

DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
                    23.5

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1) @ 1880 MHz; Calibrated: 2018/1/19
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 132.5 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 85.52 V/m
Average value of Total=(83.55+85.52) / 2 = 84.535 V/m

                                          PMF scaled E-field
                                           Grid 1 M3 Grid 2 M3 Grid 3 M3
                                           82.47 V/m 83.55 V/m 81.85 V/m
                                           Grid 4 M3 Grid 5 M3 Grid 6 M3
                                           65.17 V/m 66.52 V/m 65.79 V/m
                                           Grid 7 M3 Grid 8 M3 Grid 9 M3
                                           83.73 V/m 85.52 V/m 84.01 V/m
                                                         Cursor:
                                                    Total = 85.52 V/m
                                                     E Category: M3
                                                  Location: 0, 31, 9.7 mm




                      0 dB = 85.52 V/m = 38.64 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2019/1/16


HAC_E_Dipole_2450
DUT: HAC Dipole 2450 MHz

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2480; ConvF(1, 1, 1) @ 2450 MHz; Calibrated: 2018/12/10
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD2450 = 10mm & 15mm
2/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000 mm,
dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 82.48 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.85 V/m
Average value of Total=(91.78+92.85) / 2 = 92.315 V/m

                                          PMF scaled E-field
                                           Grid 1 M3 Grid 2 M3 Grid 3 M3
                                           90.81 V/m      91.78 V/m   88.33 V/m
                                           Grid 4 M3      Grid 5 M3   Grid 6 M3
                                           80.32 V/m      80.98 V/m   78.48 V/m
                                           Grid 7 M3      Grid 8 M3   Grid 9 M3
                                           91.73 V/m      92.85 V/m   89.91 V/m
                                                         Cursor:
                                                    Total = 92.85 V/m
                                                     E Category: M3
                                                Location: 0.5, 24.5, 8.7 mm




                     0 dB = 92.85 V/m = 39.36 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab      Date: 2019/1/11

HAC_E_Dipole_2600

DUT: HAC Dipole 2600 MHz

Communication System: CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
                    23.5

DASY5 Configuration:
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2018/1/19
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn1424; Calibrated: 2018/1/18
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 72.22 V/m; Power Drift = 0.02 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 89.15 V/m
Average value of Total=(84.26+89.15) / 2 = 86.705 V/m

                                          PMF scaled E-field
                                           Grid 1 M3 Grid 2 M3 Grid 3 M3
                                           83.80 V/m 84.26 V/m 82.21 V/m
                                           Grid 4 M3 Grid 5 M3 Grid 6 M3
                                           83.82 V/m 85.76 V/m 85.05 V/m
                                           Grid 7 M3 Grid 8 M3 Grid 9 M3
                                           87.38 V/m 89.15 V/m 87.95 V/m
                                                         Cursor:
                                                   Total = 89.15 V/m
                                                     E Category: M3
                                               Location: -0.5, 20.5, 9.7 mm




                      0 dB = 89.15 V/m = 39.00 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab       Date: 2019/1/16


HAC_E_Dipole_2600
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration:
- Probe: ER3DV6 - SN2480; ConvF(1, 1, 1) @ 2600 MHz; Calibrated: 2018/12/10
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2018/6/14
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.10 (1);SEMCAD X Version 14.6.11 (7439)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 73.74 V/m; Power Drift = 0.05 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 94.68 V/m
Average value of Total=(87.53+94.68) / 2 = 91.105 V/m


                                          PMF scaled E-field
                                           Grid 1 M3 Grid 2 M3 Grid 3 M3
                                           87.12 V/m      87.53 V/m   85.39 V/m
                                           Grid 4 M3      Grid 5 M3   Grid 6 M3
                                           89.35 V/m      91.36 V/m   90.32 V/m
                                           Grid 7 M3      Grid 8 M3   Grid 9 M3
                                           93.15 V/m      94.68 V/m   93.22 V/m
                                                         Cursor:
                                                    Total = 94.68 V/m
                                                     E Category: M3
                                                  Location: 0, 20, 9.7 mm




                     0 dB = 94.68 V/m = 39.53 dBV/m



Document Created: 2019-03-01 11:03:44