RF Exposure Appendix A

FCC ID: IHDT56XR3

RF Exposure Info

Download: PDF
FCCID_4071408

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.2

System Check_Head_835MHz

DUT: D835V2 — SN:4d091

Communication System: UID 0, CW (0), Frequency: 835 MHz;Duty Cycle: 1: 1
Medium: HSL_850 Medium parameters used: £ = 835 MHz; 6 = 0.925 S/m; s, =42.477; p = 1000
kg/m>
Ambient Temperature : 23.5 °C; Liquid Temperature : 22.8 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3935; ConvF(10.36, 10.36, 10.36); Calibrated: 2017.12.14;
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
— Phantom: SAM1; Type: SAM; Serial: TP—1839
— Measurement SW: DASY52, Version 52.8 (8), SEMCAD X Version 14.6.10 (7372)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) =2.51 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 48.21 V/m; Power Drift =—0.00 dB
Peak SAR (extrapolated) =3.15 W/kg
SAR(I g) = 2.17 W/kg; SAR(10 g) = 1.44 W/kg
Maximum value of SAR (measured) =2.52 W/kg




    —3.61



    —1.22



    —10.82



    —14.43



    —18.04
               0 dB =2.51 W/kg =4.00 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.7

System Check_Head_1750MHz

DUT: D1750V2 — SN:1069

Communication System: UID 0, CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: HSL_1750 Medium parameters used: £=1750 MHz; 0 = 1.345 S/m; s = 41.143; p= 1000
kg/m3
Ambient Temperature : 23.2 °C; Liquid Temperature : 22.6 °C

DASY 5 Configuration:
— Probe: EX3DV4 — SN3935; ConvF(8.85, 8.85, 8.85); Calibrated: 2017.12.14;
— Sensor—Surface: 1.Amm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
— Phantom: SAM1; Type: SAM; Serial: TP—1839
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Sean (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 12.9 Wkg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Mcasurementgrid: dx=8mm, dy=8mm, dz=5mm
Reference Value=84.44 V/m; Power Drift = 0.08 dB
Peak SAR (extrapolated) = 16.4 W/kg
SAR(1 g) = 8.9 W/kg; SAR(1O g) = 4.66 W/kg
Maximum value of SAR (measured) = 12.8 W/kg

   dB
   0


    —3.50


    —6.99


    —10.49


    13.98


    —17.48
               0 dB = 12.8 W/kg =11.07 dBWi/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.8

System Check_Head_1900MHz

DUT: D1900V2 — SN: 50118

Communication System: UID 0, CW; Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: HSL_1900 Medium parameters used: £= 1900 MHz; 0 = 1.428 S/m; s = 41.108; p= 1000
kg/m3
Ambient Temperature : 23.2 °C; Liquid Temperature : 22.7 °C

DASY 5 Configuration:
— Probe: EX3DV4 — SN3935; ConvF(8.41, 8.41, 8.41); Calibrated: 2017.12.14;
— Sensor—Surface: 1.Amm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
— Phantom: SAM1; Type: SAM; Serial: TP—1839
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Sean (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 14.2 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Mcasurementgrid: dx=8mm, dy=8mm, dz=5mm
Reference Value=87.91 V/m; Power Drift=0.01 dB
Peak SAR (extrapolated) = 18.0 Wkg
SAR(1 g) = 9.76 Wi/kg; SAR(1O g) = 5.05 W/kg
Maximum value of SAR (measured) =14.1 W/kg

   dB
   0


    —3.60


    —7.20


    —10.80


    —14.40


    —18.00
               0 dB =14.1 W/kg=11.49 dBWi/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.22

System Check_Head_2450MHz

DUT: D2450V2 — SN:840

Communication System: UID 0, CW (0); Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: HSL_2450 Medium parameters used: £=2450 MHz; 0 = 1.848 S$/m; £,= 39.012; p = 1000
kg/m3
Ambient Temperature : 23.5 °C; Liquid Temperature : 22.7 °C

DASY S Configuration:
—   Probe: EX3DV4 — SN3935; ConvF(7.87, 7.87,7.87); Calibrated: 2017.12.14;
—   Sensor—Surface: 1.Amm (Mechanical Surface Detection)
—   Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
—   Phantom: SAM2; Type: SAM; Serial: TP—1842
—   Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 20.8 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, daz=Smm
Reference Value=88.69 V/m; Power Drift =0.03 dB
Peak SAR (extrapolated) = 28.0 Wkg
SAR(1 g) = 13.39 W/kg; SAR(10 g) = 5.88 W/kg
Maximum value of SAR (measured) = 19.96 W/kg

     dB
     0


      ~4.83


      —9.66


      —14.49


      "19.32


      —24.15
               0 dB =9.96 Wi/kg =9.98 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.24

System Check_Head_2600MHz

DUT: D2600V2 — SN:1061

Communication System: UID 0, CW; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: HSL_2600 Medium parameters used: £=2600 MHz; 0 =2.021 S/m; s = 38.375; p= 1000
kg/m3
Ambient Temperature : 23.4 °C; Liquid Temperature : 226 °C

DASY 5 Configuration:
— Probe: EX3DV4 — SN3935; ConvF(7.67, 7.67, 7.67); Calibrated: 2017.12.14;
— Sensor—Surface: 1.Amm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
— Phantom: SAM2; Type: SAM; Serial: TP—1842
— Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) =21.6 Wkg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Mcasurementgrid: dx=5mm, dy=5mm, dz=5mm
Reference Value=83.89 V/m; Power Drift = 0.03 dB
Peak SAR (extrapolated) =31.8 W/kg
SAR(1 g) = 13.6 W/kg; SAR(10 g) = 6.89 W/kg
Maximum value of SAR (measured) =21.9 W/kg

   dB
   0


    ~4.90


    —9.79


    —14.69


    —19.58


    —24.48
               0 dB = 22.9 Wikg =13.60 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.4

System Check_Body_835MHz

DUT: D835V2 — SN:4d091

Communication System: UID 0, CW; Frequency: 835 MHz;Duty Cyole: 1: 1
Medium: MSL_850 Medium parameters used: £ = 835 MHz; 5 = 0.983 S/m; s, = 54.264; p = 1000
kg/m>
Ambient Temperature : 23.5 °C; Liquid Temperature : 22.8 °C

DASY5 Configuration:
— Probe: EX3DV4 — SN3935; ConvF(10.33, 10.33, 10.33); Calibrated: 2017.12.14;
— Sensor—Surface: 1.4mm (Mechanical Surface Detection)
— Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
— Phantom: SAM2; Type: SAM; Serial: TP—1842
— Measurement SW: DASY52, Version 52.8 (8), SEMCAD X Version 14.6.10 (7372)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) =2.86 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm
Reference Value = 49.12 V/m; Power Drift = 0.04 dB
Peak SAR (extrapolated) = 3.34 W/kg
SAR(I g) = 2.29 W/kg; SAR(10 g) = 1.52 W/kg
Maximum value of SAR (measured) = 2.88 W/kg




    —3.74



    —1 48



    —11.21



    —14.95



    —18.69
               0 dB = 2.88 W/kg = 4.59 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.20

System Check_Body_1750MHz

DUT: D1750V2 — SN:1069

Communication System: UID 0, CW; Frequency: 1750 MHz;Duty Cycle: 1:1
Medium: MSL_1750 Medium parameters used: £= 1750 MHz; a= 1.491 S/m; e = 54.729; p= 1000
kg/m3
Ambient Temperature : 23.3 °C; Liquid Temperature : 22.5 °C

DASY S Configuration:
—   Probe: EX3DV4 — SN3935; ConvF(8.71, 8.71, 8.71); Calibrated: 2017.12.14;
—   Sensor—Surface: 1.Amm (Mechanical Surface Detection)
—   Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
—   Phantom: SAM2; Type: SAM; Serial: TP—1842
—   Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.49 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, daz=Smm
Reference Value=84.72 V/m; Power Drift =0.13 dB
Peak SAR (extrapolated) = 16.3 W/kg
SAR(1 g) = 9.35 W/kg; SAR(1O g) = 5.08 W/kg
Maximum value of SAR (measured) = 12.90 W/kg

     dB
     0


      —3.29


      —6.59


      —9.88


      13118


      "16.47
               0 dB =9.90 Wikg =9.96 dBW/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.21

System Check_Body_1900MHz

DUT: D1900V2 — SN: 50118

Communication System: UID 0, CW (0); Frequency: 1900 MHz;Duty Cycle: 1:1
Medium: MSL_1900 Medium parameters used: £= 1900 MHz; a= 1.537 S/m; e = 53.469; p = 1000
kg/m3
Ambient Temperature : 23.2 °C; Liquid Temperature : 22.6 °C

DASY S Configuration:
—   Probe: EX3DV4 — SN3935; ConvF(8.3, 8.3, 8.3); Calibrated: 2017.12.14;
—   Sensor—Surface: 1.Amm (Mechanical Surface Detection)
—   Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
—   Phantom: SAM1; Type: SAM; Serial: TP—1839
—   Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (61x61x1): Interpolated grid: dx=1.500 mm, dy=1.500 mm
Maximum value of SAR (interpolated) = 13.9 W/kg

Pin=250mW/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, daz=Smm
Reference Value=83.85 V/m; Power Drift =0.10 dB
Peak SAR (extrapolated) = 17.5 Wkg
SAR(1 g) = 9.67 Wi/kg; SAR(1O g) = 5.05 W/kg
Maximum value of SAR (measured) = 13.7 W/kg

     dB
     0


      —3.62


      —7.24


      —10.86


      "14.48


      1810
               0 dB = 10.7 Wikg =10.29 dBWi/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.23

System Check_Body_2450MHz

DUT: D2450V2 — SN:840

Communication System: UID 0, CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: MSL_2450 Medium parameters used: £= 2450 MHz; a=2.003 S/m; e = 53.104; p = 1000
kg/m3
Ambient Temperature : 23.4 °C; Liquid Temperature : 22.7 °C

DASY S Configuration:
—   Probe: EX3DV4 — SN3935; ConvF(7.99, 7.99, 7.99); Calibrated: 2017.12.14;
—   Sensor—Surface: 1.Amm (Mechanical Surface Detection)
—   Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
—   Phantom: SAM2; Type: SAM; Serial: TP—1842
—   Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) = 18.3 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, daz=Smm
Reference Value=81.82 V/m; Power Drift =—0.09 dB
Peak SAR (extrapolated) =23.1 W/kg
SAR(1 g) = 12.1 W/kg; SAR(1O g) = 5.87 W/kg
Maximum value of SAR (measured) =17.9 W/kg

     dB
     0


      ~4.41


      —8.83


      ~13.24


      —17.66


      —22.07
               0 dB =17.9 W/kg =12.53 dBWi/kg


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab   Date: 2018.10.22

System Check_Body_2600MHz

DUT: D2600V2 — SN:1061

Communication System: UID 0, CW (0); Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: MSL_2600 Medium parameters used: £= 2600 MHz; a=2.231 S/m; e = 52.422; p = 1000
kg/m3
Ambient Temperature : 23.4 °C; Liquid Temperature : 22.8 °C

DASY S Configuration:
—   Probe: EX3DV4 — SN3935; ConvF(7.71, 7.71, 7.71); Calibrated: 2017.12.14;
—   Sensor—Surface: 1.Amm (Mechanical Surface Detection)
—   Electronics: DAE4 Sn1338; Calibrated: 2017.12.4
—   Phantom: SAM2; Type: SAM; Serial: TP—1842
—   Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331)


Pin=250mW/Area Scan (71x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm
Maximum value of SAR (interpolated) =21.7 W/kg

Pin=250mW/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=Smm, daz=Smm
Reference Value=83.27 V/m; Power Drift =0.03 dB
Peak SAR (extrapolated) =28.1 Wkg
SAR(1 g) = 14.2 W/kg; SAR(1O g) = 6.55 W/kg
Maximum value of SAR (measured) =21.5 W/kg

     dB
     0


      ~4.66


      —9.33


      13.99


      —18.66


      —23.32
               0 dB =21.5 Wikg =13.32 dBWi/kg