RF Exposure Appendix D 3

FCC ID: IHDT56XP1

RF Exposure Info

Download: PDF
FCCID_4060473

Exapve— snsos                                                                                                      January 31, 2018



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Calibration Parameter Determined in Head Tissue Simulating Media
                  Relative   Conductivity                                                                   Depth®           Une
    {(Mits)©_|_Pormittivity®    (Sim)"    ConvEX ConvEY |_ConvEZ Apha®                                       |__(mm)        (ke2)
        750             410               o0              1072         i072          ioza         oze         i1         +120%
        sas             415               0.90            1020         1020          1020         oas         o%o        +120%
       1750             40.               137             865          865           805          oze         oss        £120%
       1900             40.0              140             a41          sa1           sa1          oz1         ose        £120%
       2000             40.0              140             833          833           a33          oso_|       om_|       £120%
       2300             so5               167             789          789           789          ost         os         £120%
       2450             so2               180             740          749           7ao          oss         om         £120%
       2600             30.0              196             731          731           ran          om          108        +120%
       s2so             350               an              520          5.20          520          oso         180        a131%
       se00             355               sor             459          459           4so          o40         180        +131%
       s7so             354                522            474           474          474          o40         180        £181%

© rrequency valtyabove 300 Nz o 2 100 Mz only apls for DASY v6.t and Nigher(see Page 2), ols t s resticted to + 50 MIHz. The
uncertaiy s the RSS of the Coovuncertaint at calbrtion requency and the unceainy orthe incleated requency band. Frequency valdty
below 300 itz is + 10,25,40, 50 and 70 MHe forConyassessmenta at 2064, 128,180 and 220Mz respectvaly. Above 5 GHz fequency
yataty can be extended t 2 110 itz
* Atfrequencies blow 3 Git, the valdty of ssue parameters (and o) can be relaved o : 10% i lquid compensation formula s appled to
measured SAR values. A requencies above 3 GHi, the valdty of ssue parameters(sand o)is resticted o + 8%. The uncerialnty is the RSS of
the Conv: uncertaint for ndlealed targettasue paramoters
* Apha(Depth are daternined during calbraton. SPEAG warrants tat h remiaining deviaton due t t boundary efect after compensation is
always less than # 1% fo frequencies below 3 GHtz and blow 2 2% forfrequencies betwoen 3.6 GHz at any distancelarge than halthe probe to
ciameterrom the boundry.




Cortficate No: EX3—3054_Jan18                                Page S of 11


exsove— sNgoss                                                                                                   January 31, 2018



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Calibration Parameter Determined in Body Tissue Simulating Media
               Relative  Conductivity                                                                     Depth®          Une
     1(mits)©      Pormittivity"        (Sim"          ConvFX       ConvEY       ConvEZ       Aipha®       |__(mm)       (e2)
       750             555                0.96          10.21         10.21        10.21        041        0.84        £12.0 %

       835             552                0.97          10.02         10.02        10.02        0.40       0.85        £12.0 %

       1750            sea                149            a31          a31          a31          oss        oss         +120%
       1900            533                1.52           803          8.03         8.03         041        0.85        £12.0 %

       2300            52.9               1.81           774          7.74         1.74         0.46       0.80        £12.0 %

       2450            527                195            7.53         753          783         0.34        o.8         £12.0 %

       2600            525                2.16           6.92         6.92         6.92         027        1.05        2120 %

       5250            489                5.36           462          462          462         035         1.90        £19.1 %

       5600            485                5.77           4.05         4.05         405         0.40        1.90        £19.1 %

       5750            48.3               5.94           4.18         4.18         4.18        0.40        1.90        £13.1%



© Frequency valdty above 300 itz f : 100 NHz anly appliesfor DASY vi.4 and Nighor (soe Page 2), lse t is restreted to : 50 Mtz.The
uncertaint s the RSS ofthe Con uncertanty at calbraton frequency and the uncerainy forthe inicated reauency band. Frequency valdty
below 300 Mte is 2 10,2540. 50 and 70 MHz for Comv assessmentsat 20 64, 128, 150 and 220 Ntz respectivaly. Above 5 GHz freauency
yaldity can bo antendad t 2 110 itz
* Atrequencies blow 3 GHz,the valdty oftssue pararmeters ( and ) can be eloxed o : 10% i lquid componsaton formula s appled to
measured SAR values. Atrequencies above 3 GH, thevalty f issc parameters(sand a)is resttod o 2 5%. The uncetainy i e RSS of
the Gon uncertaint ornclcate trgettssue paramoters
* AphaDepth are deterninad during caltraton. SPEAG warrans that t remiining deviaton ds to the boundary afect ltcompensation is
always less than:: 19 fo frequencias below 3 GHtzand below : 2% fofrequencies betean 36 GHz at any istance large than hall he probe t
iameterom the boundary




Certficate No: EX3—3054_Jan18                               Page 6 of 11


Exsove— sngost                                                                                         January 31, 2018

                                            Frequency Response of E—Field
         Frequency response (normalized)         (TEM—Cell:if110 EXX, Waveguide: R22)




                                           Uncertainty of Frequency Response of E—field: # 6.3%(k=2)




Cortficate No: EX3—3054_Jan18                                     Page 7 of 11


Exsove— ssas4                                                                                        January 31, 2018



                                   Receiving Pattern (¢), 9 = 0°

                         £2600 MHz,TEM                                       11800 MHz,R22
                          x«



                                      <«
           w




                                                 ®




                                                                      ®



                                                                                  ®



                                                                                            <«



         as}—
                puacd        p ud auea ue d uc ues
                        ho         —ho          4o             d           4          1do             tho
                                                             roir
                         \hile                 shtte                      wtthe                  wWitke
                                 Uncertainty of Axial Isotropy Assessment: £0.5% (k=2)




Certficate No: EX3—3054_Jan18                          Page 8 of 11


Exsove— h3054                                                                                      January 31, 2018




                                           Dynamic Range f(SARneaq)
                                                  (TEM cell , fo,s= 1900 MHz)


                                     3
                 Input Signal (uV)




                                                          i                     bbdidifenstcd 4j
                                                        .:'(;V           10         10
                                                                 SAR [mWlems]




                                                    +
                                             not compansated                       compensates
                                         Uncertainty of Linearity Assessment: # 0.6% (=2)




Cortficate No: EX3—3054_Jan8                                     Page 9 of 11


Exgove— sNaos4                                                                                        January 31, 2018

                                Conversion Factor Assessment
                = 835 MHz.WGLS RO (H_conve)                                1‘= 1900 MHz.WGLS R22 (H_conve)




           &                     a.   &A       &  m         +          s    4       ow «en
                                                                                        0.          »   s      .
                                «pe
                       n
                    wiliew                    a.
                                            witte                                  8
                                                                                 wolde              »
                                                                                                  wotte

                             Deviation from Isotropy in Liquid
                                              Error (6, 8), f= 900 MHz




                         «10    —08   .08    o4       —o2       oo   o2     o4    o8   08    10
                             Uncertainty of Spherical Isotropy Assessment:£2.6% (k=2)


Cortficate No: EX3—3054_Jan18                          Page 10 of 11


exsove— sNgose                                               January 31, 2018


DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Other Probe Parameters
 Sensor Arrangement                                              "Trianguiar
 Connector Angle (*)                                                   Toa
 Mechanical Surface Detection Mode                                 enabled
 Optical Surface Detection Mode                                    disabled
 Probe Overall Length                                              337 mm
 Probe Body Diameter                                                10 mm
 Tip Leng                                                            9 mm
 Tip Diameter                                                       25 mm
 Probe Tip to Sensor X Calibration Point                              1 mm
 Probe Tip to Sensor Y Callbration Point                              1 mm
 Probe Tip to Sensor 2 Calibration Point                              1 mm
 Recommended Measurement Distance from Surtace                      14 mm




Certficate No: EX3—3954_Jan18                Page 11 of 11


                                                         .
                                          in Collaboration with
                                                                                       $
                                                                                       4P ,'l
                                                                                      & \\\/
                                                                                              Vn& A 4 B{AaJ
                                                                                                    IEE“

                     TTL CALIBRATION
                         a                                                                              CNASz#
                                                             LABORATORY                4 /_\\ §          v CALIBRATION
             Add: No.51 Xueyuan Road. Haidian District, Beijing, 100191, China          "/.,,fi \a"                        CNAS LO570
             Tel: +86—10—62304633—2512           Fax: +86—10—62304633—2504
             E—mail: cttl@chinattl.com           Hitp://www.chinattl.on
      Client              HMN TESTING                                           Certificate No: 2Z18—60230
CALIBRATION CERTIFICATE

Object                                      ES3DV3 — SN:3088

Calibration Procedure(s)                    FF—211—004—01

                                            Calibration Procedures for Dosimetric E—field Probes

Calibration date:                           July 19, 2018


This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(S1). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘c and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)
Primary Standards                         ID #          Cal Date(Calibrated by, Certificate No.)           Scheduled Calibration
 Power Meter        NRP2           101919                    20—Jun—18 (CTTL, No.J18X05032)                   Jun—19
 Power sensor       NRP—Z91        101547                    20—Jun—18 (CTTL, No.J18X05032)                   Jun—19
 Power sensor       NRP—Z91        101548                    20—Jun—18 (CTTL, No.J18X05032)                   Jun—19
 Reference10dBAttenuator           18N5OW—10dB               09—Feb—18(CTTL, No.J18X01133                     Feb—20
 Reference20dBAttenuator           18N5OW—20dB               09—Feb—18(CTTL, No.J18X01132                     Feb—20
 Reference Probe EX3DV4            SN 3846                   25—Jan—18(SPEAG,N0.EX3—3846 Jan18)               Jan—19
 DAE4                              SN 777                    15—Dec—17(SPEAG, No.DAE4—77"_Dec17)              Dec—18


 Secondary Standards               ID #                      Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 SignalGeneratorMG3700A            6201052605                21—Jun—18 (CTTL, No.J18X05033)                 Jun—19
 Network Analyzer E5071C           MY46110673                14—Jan—18 (CTTL, No.J18X00561)                 Jan —19
                                Name                              Function                                   Signature
Calibrated by:                   Yu Zongying                     SAR Test Engineer                      s /ssfl/q‘fig()

Reviewed by:                     Lin Hao                         SAR Test Engineer                        Hfij{, ~

Approved by:                     Qi Dianyuan                     SAR Project Leader                     ‘a‘g/&/
                                                                                          Issued: July 20, 2018
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



    Certificate No: Z18—60230                                    Page 1 of 11


                                       in Collaboration with

                    TTL a
                                       CALIBRATION LABORATORY

           Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com        Http://www.chinattl.on

Glossary:
TSL                      tissue simulating liquid
NORMx,y,z                sensitivity in free space
ConvF                    sensitivity in TSL / NORMx,y,z
DCP                      diode compression point
CF                       crest factor (1/duty_cycle) of the RF signal
A,B,.C,.D                modulation dependent linearization parameters
Polarization ®           ® rotation around probe axis
Polarization 0           0 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                         0=0 is normal to probe axis
Connector Angle           information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
  Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
  Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
    July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
    2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GFz"
Methods Applied and Interpretation of Parameters:
e    NORMx,y,z: Assessed for E—field polarization 0=0 (fs900MHz in TEM—cell; > 1800MHz: waveguide).
     NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
      E" —field uncertainty inside TSL (see below ConvF).
e    NORM(P)x,y,z = NORMx,y.z* frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than «.2. The uncertainty of the
     frequency response is included in the stated uncertainty of ConvF.
*    DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
e    PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
     characteristics.
e    Ax,y,z; Bxyz; Cx,yz;VRx,y,z:A,B,C are numerical linearization parameters assessed based on the
     data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum calibration range expressed in RMS voltage across the diode.
e    ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
     Transfer Standard for f<800MHz) and inside waveguide using analytical field distributions based on
     power measurements for f >800MHz. The same setups are used for assessment of the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
     that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
     allows extending the validity from+50MHz to£+100MHz.
e    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
     phantom exposed by a patch antenna.
e    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
     probe tip (on probe axis). No tolerance required.
e    Connector Angle: The angle is assessed using the information gained by determining the NORMx
     (no uncertainty required).



Certificate No: Z18—60230                                 Page 2 of 11


                                    in Collaboration with

                 TTL a              CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: ettl@chinattl.com        Hitp://www.chinattl.on




                       Probe ES3DV3


                                              SN: 3088


                                          Calibrated: July 19, 2018

                                Calibrated for DASY/EASY Systems
                                        (Note: non—compatible with DASY2 system!)




Certificate No: Z18—60230                               Page 3 of11


                                     in Collaboration with

                  TTL a
                                     CALIBRATION LABORATORY

         Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com        Hitp:/www.chinattl.en

      DASY/EASY — Parameters of Probe: ES$3DV3 — SN: 3088

Basic Calibration Parameters
                                      Sensor X                Sensor Y             Sensor Z         Unc (k=2)
 Norm(uV/(V/m))A                      1.31                    1.25                 1.18             +£10.0%
 DCP(mV)®                             105.2                   104.9                106.4


Modulation Calibration Parameters
  UID         Communication                         A             B          6             )      VR          Unc®
              System Name                           dB            dB/uV                    dB     mV          (k=2)
  0           Cw                          X         0.0           0.0        1.0            .00   2175.5      £3.0%
                                          Y         0.0           0.0        1.0                  260.8
                                          2         0.0           0.0        1.0                  251.3



      The reported uncertainty of measurement is stated as the standard uncertainty of
      Measurement multiplied by the coverage factor k=2, which for a normal distribution
      Corresponds to a coverage probability of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E®—field uncertainty insice TSL (see Page 5 and Page 6).
® Numerical linearization parameter: uncertainty not required.
& Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z18—60230                                 Page 4 of 11


                                       in Collaboration with

                   TTL                 s       p__e_           a     q
                                       CALIBRATION LABORATORY
           Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
           Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
           E—mail: cttl@chinattl.com        Hitp://www.chinattl.en

       DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3088


Calibration Parameter Determined in Head Tissue Simulating Media
                        7                      ez                                                          6
 f [IMHz]® Pe::Ii:tti?lliin C°"‘:;Z:‘)’f” ConvF X ConvF Y ConvF Z Alpha® D(:’::;                               :i:czt)
    900            415                  0.97              6.13           613    613       0.43      153        £121%
   1810            40.0                 1.40              5.05           5.05   505       0.46      153        £121%
   1900            40.0                 1.40              5.07           5.07   5o7       o.68      123        £121%
   2450            39.2                 1.80              4.65           4.65   465       o83       118        112 1%
   2600            39.0                 1.96              445            445    445       097       1.08       112 1%

& Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and t gher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency ind the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz or ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended tc £ 110 MHz.
" At frequency below 3 GHz, the validity of tissue parameters (e and 0) can be rela:» ad to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the valid y of tissue parameters (e and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicatec target tissue parameters.
GAIpha/Depth are determined during calibration. SPEAG warrants that the remaini g deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz ar 1 below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the b undary.




  Certificate No: Z18—60230                               Page 5 of 11


                                        CALIBRATION LABORATORY
            Add: No.31 Xucyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
            E—mail: cttl@chinattl.com       Hitp://www.chinattl.en

       DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3088


Calibration Parameter Determined in Body Tissue Simulating Media
                         s                      hm                                                        G
 f [MHz]®      P::";:;:;),F         C°"‘:;/°r:‘)'fy     ConvF X ConvF Y ConvF Z Alpha® D(':np:;               :i':;t)
    900             55.0                 1.05             6.15           615    615      048        146       £121%
   1810             53.3                 1.52             4.82           4.82   482      0.63       130       £121%
   1900             53.3                 1.52             4.71           4.71   4a7i     064        130       £121%
   2450             52.7                 1.95             4.41           4.41   441      O0.85      120       +121%
   2600             52.5                 216              415            415    415      0.90       113       £121%

© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
* aAt frequency below 3 GHz, the validity of tissue parameters (¢ and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (¢ and 0) is
restricted to +5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
& Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




  Certificate No: Z18—60230                               Page 6 of 11


                                                                in Collaboration with

                                              TTL a
                                                                CALIBRATION LABORATORY
         Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
         Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
         E—mail: ettl@chinattl.com      Hitp://www.chinattl.on


                                                          Frequency Response of E—Field
                                                          (TEM—Cell: ifi110 EXX, Waveguide: R22)
                                          1.5 —

                                          «14. 1..
                                          183.
        Frequency response (normalized)




                                          12|
                                          111.
                                          ~
                                              o
                                                   }
                                                   i
                                          o
                                              «©
                                                   J
                                          o
                                              Co
                                                   }
                                                   _i
                                          o
                                              ~
                                          J ;      L
                                          &
                                          ®




                                          0.5               $    j       +       d                    +    ;     +     ;     t     T
                                                    0           500           1000           1500         2000        2500       3000
                                                                      E*3                   f [MHz
                                                                      Tem                    (MFiz]              2s
                                                        Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z18—60230                                                          Page 7 of 11


                                        in Collaboration with

                    TTL a               CALIBRATION LABORATORY
            Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
            Tel: +86—10—62304633—2512         Fax: +86—10—62304633—2504
            E—mail: ettl@chinattl.com         Hitp://www.chinattl.en

                                   Receiving Pattern (®), 6=0°

                  f=600 MHz, TEM                                                 f=1800 MHz, R22
ErrorfdB]




                    amy mgen Wr‘*"m ~#—»@—0—*® fogepeck '*""‘““’"t:':'?-tz:u




                      T                 T               T                  T            T         T           T
                    ~150             —100              50                  0           50        100         150

                                            uce             Ees     Rollf*]                            seq
                              [~+—100MHz               _ —_ 600MHz             _ —_ 1800MHz   —_ 2500MHz]
                                 Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




   Certificate No: Z18—60230                                Page 8 of 11


                                                         In Collaboration with

                                        TTL CALIBRATION
                                            a           LABORATORY
        Add: No.51 Xucyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512                             Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com                             Hitp://www.chinattl.en


                                                        Dynamic Range f(SARn—q4)
                                                                  (TEM cell, f                        900 MHz)
                     Input Signal[uV]




                                                    2                    1                           00
                                               10                   10                           1

                                                                                       SAR[mW/cm*]
                                                                       E not compensated                  —®— compensated
         Error(dB]




                                                                  ~@—+ U <L—C|—O—0—#+:9 s s o 0 o L000
                                               —s —s—#A~4—#—#4~#~ —m




                                                           e
                                        10 °                   10                           10                        10        10°
                                                                                  SAR[mW/cm ]
                                                          @— not compensated            * sompensated                       |
                                                                                                                            )
                                                                           ffy Assessment +% 3.9% (k=2)
                                                        Uncertainty of Li nearl

Certificate No: Z18—60230                                                    Page 9 of 11


          ETTL s\ea a                       CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com      Http://www.chinattl.en


                              Conversion Factor Assessment

       f=900 MHz, WGLS R9(H_convF)                                                     f=1810 MHz, WGLS R22(H_convF)
              4.00 ;                                                                        20.00 [

              250 }                                                                                  .
                   k                                                                        2500 [t————           (——               four   omfromemntemmcem
              £B0 J
                  [A                                                                        20 00
            a 250 |:—+—                     t                —=                         a
            a         &                                                                 7

            $.a &
            4i 1.50. i#
                                                                                        §
                                                                                        6
                                                                                            10.00



                                                                                             5.00


                                                                                            9 00
                                                                                                                 20    30           40      50     50     7O
                                                 2{mm]                                                                      zimm]
                                   __—#—measured _ —analytical _                                                —#—measured      —— analytical _




                       Deviation from Isotropy in Liquid
                                    10

                                    08

                                    06

                                    0 4

                                    0.2
                          7 Axis




                                     0.0

                                    —0 2
                                    —0 4
                                     06
                                     —0 8
                                      .18




                                                10   .080   .060   .040   .020     0    020         040   060    8o   10
                                   Uncertainty of Spherical Isotropy Assessmen:t: £3.2% (K=2)



Certificate No: Z18—60230                                          Page 10 of 11


                                    CALIBRATION LABORATORY
        Add: No.31 Xucyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2512      Fax: +86—10—62304633—2504
        E—mail: cttl@chinattl.com       Hitp://www.chinattl.en

    DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3088


     Other Probe Parameters
     Sensor Arrangement                                                     Triangular

    Connector Angle (°)                                                            4.7

     Mechanical Surface Detection Mode                                        enabled

    Optical Surface Detection Mode                                            disable

    Probe Overall Length                                                      337mm

    Probe Body Diameter                                                         10mm

    Tip Length                                                                  10mm

    Tip Diameter                                                                 4mm

    Probe Tip to Sensor X Calibration Point                                      2mm

    Probe Tip to Sensor Y Calibration Point                                      2mm

    Probe Tip to Sensor Z Calibration Point                                      2mm

    Recommended Measurement Distance from Surface                                3mm




Certificate No: Z18—60230                             Page 11 of 11


                                                                                         “\l 1 ||I t                       o


                 33‘                    In Collabaration with                     .,&‘\\//"/zz A
                                                                                  $yz                                   AFiAn
               w‘/"7‘/, s p                             e       a       q         MCNAsg                             gfifi
                                        CALIBRANRON LABORATORY                     i3                  &
            Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
                                                                                   ze\y Nee Cauisration
                                                                                        "/,,m‘\ w                       CNAS LO570
            Tel: +86—10—62304633—2218          Fax: +86—10—62304633—2209                    mm
            E—mail: cttl@chinattl.com          Hitp://www.chinattl.cn
      Client           Sporton International INC                             Certificate No: 217—97257
   AHIBRATIONCERTIFICATE
Object                                    EX3DV4 — SN:3935

Calibration Procedure(s)                  FF—211—004—01
                                          Calibration Procedures for Dosimetric E—field Probes

Calibration date:                         December 14, 2017

This calibration Certificate documents the traceability to national standards, which realize the physical units of
measurements(Sl). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.


All calibrations have been conducted in the closed laboratory facility: environment temperature(22+3)‘c and
humidity<70%.

Calibration Equipment used (M&TE critical for calibration)
Primary Standards                       ID #          Cal Date(Calibrated by, Certificate No.)             Scheduled Calibration
 Power Meter        NRP2          101919                27—Jun—17 (CTTL, No.J17X05857)                      Jun—18
 Power sensor       NRP—Z291      101547                27—Jun—17 (CTTL, No.J17X05857)                       Jun—18
 Power sensor       NRP—Z91       101548                 27—Jun—17 (CTTL, No.J17X05857)                      Jun—18
 Reference10dBAttenuator 18NS5OW—10dB                     13—Mar—16(CTTL,No.J16X01547)               Mar—18
 Reference20dBAttenuator 18N5OW—20dB                     13—Mar—16(CTTL, No.J16X01548)               Mar—18
 Reference Probe EX3DV4 SN 7464                          12—Sep—17{SPEAG,No.EX3—7464_Sep17)          Sep—18
 DAE4                    SN 549                          13—Dec—16(SPEAG, No.DAE4—549_Dec16)         Dec —17
 DAE4                    SN 1524                         13—Sep—17(SPEAG, No.DAE4—1524_Sep17)        Sep —18
 Secondary Standards     ID #                            Cal Date(Calibrated by, Certificate No.) Scheduled Calibration
 SignalGeneratorMG3700A 6201052605                       27—Jun—17 {CTTL, No.J17X05858)            Jun—18
 Network Analyzer E5071C          MY46110673             13—Jan—17 (CTTL, No.J17X00285)                    Jan —18
                                 Name                           Function

Calibrated by:                   Yu Zongying                   SAR Test Engineer

Reviewed by:                     Lin Hao                      SAR Test Engineer

Approved by:                     Qi Dianyuan                   SAR Project Leader

                             \                                                            iIssued: December 16, 2017
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: Z17—97257                                Page 1 of 11


                              6
           .?\                     in Collaboration with
           7/77 s                                 e a
                                   CALIBRANTON LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing,
                                                            100191, China
        Tel: +86—10—62304633—2218     Fax: +86—10—62304633—2209
        E—mail: cttl@chinattl.com     Http:/www.chinattl.cn


Glossary:
TSL                  tissue simulating liquid
NORMx,y,z            sensitivity in free space
ConvF                sensitivity in TSL / NORMx,y,z2
DCP                  diode compression point
CF                   crest factor (1/duty_cycle) of the RF signal
A,8B,C,D             modulation dependent linearization parameters
Polarization ®  © rotation around probe axis
Polarization 0  0 rotgtion around an axis that is in the plane normal to probe axis (at measurement
                                                                                                      center)
                0=0 is normal to probe axis
Connector Angle information used in DASY system to align probe sensor X to the
                                                                                   robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial
                                                                                               —Averaged
   Specific Absorption Rate (SAR) in the Human Head from Wireless Communications
                                                                                                Devices:
   Measurement Techniques", June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   gard£8?l6d and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)",
    uly
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
     NORMx,y,z: Assessed for E—field polarization 0=0 (fs900MHz in TEM—cell; f> 1800MHz: waveguide).
     NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
     E‘ —field uncertainty inside TSL (see below ConvF).
     NORM(Px,y,z = NORMx,y,z* frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
     frequency response is included in the stated uncertainty of ConvF.
     DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
     PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
     characteristics.
     Ax,y,z; Bx,y,z; Cx,y z; VRx,y,z:A,B,C are numerical linearization parameters assessed based on the
     data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum calibration range expressed in RMS voltage across the diode.
     ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
     Transfer Standard for fs800MH2z) and inside waveguide using analytical field distributions based on
     power measurements for f >800MHz. The same setups are used for assessment of the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitivity in TSL corresponds to NORMx,y,z* ConvF whereby the uncertainty corresponds to
     that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
     allows extending the validity from+50MHz to+100MHz.
     Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat
     phantom exposed by a patch antenna.
     Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
     probe tip (on probe axis). No tolerance required.
     Connector Angle: The angle is assessed using the information gained by determining the NORMx
     {no uncertainty required).

Certificate No: Z17—97257                             Page 2 of 11


                 a®
                                          in Collaboration with
                =7"7~ a
                                         CALIBRATON LaBOorArory
             Add: No.51 Xueyuan
                                  Road, Haidian District,
             Tel: +86—1 0—62304633—22                     Beijing, 100191, China
                                      18     Fax: +86—10
             E—mail: cttl@chinattl.c                      —62304633—2209
                                    om        Hittp://www.chinattl.cn




                            Probe EX38DV4


                                                SN: 3935


                                     Calibrated: December 1 4, 2017

                                Calibrated for DASY/EASY Syst
                                                              ems
                                         (Note: non—compatible with DASY2
                                                                          system!)




Certificate No: Z17—97257
                                                      Page 3 of 11


                                   In Collaboration with


          E‘T./]Z fmmnou?.fimzmm
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2218     Fax: +86—10—62304633—2209
        E—mail: cttl@chinattl.com      Http://www.chinattl.en


     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3935

Basic Calibration Parameters
                                    Sensor X                Sensor Y              Sensor Z         Unc (k=2)
Norm(uV/(V/m)‘)*                    0.48                    0.54                  0.49             £10.0%
DCP(my)®                            104.4                   104.3                 106.1


Modulation Calibration Parameters
 UID         Communication                        A            B            C             D      VR         unc ®
             System Name                          dB           dB\uV                      dB     myV        (k=22)
 0           Cw                          X        0.0          0.0          1.0           0.00   162.8      |+2.2%
                                         Y¥        0.0         0.0          1.0                  176.1
                                         2         0.0         0.0          1.0                  165.6


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E*—field uncertainty inside TSL (see Page 5 and Page 6).
® Numerical linearization parameter: uncertainty not required.
E Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z17—97257                                Page 4 of 11


                                ® |n, Collaboration with
            «/TL s                                  e_ a
             \Wiaagee»"               CALIBRATION LABORATORY
          Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
          Tel: +86—10—62304633—2218       Fax: +86—10—62304633—2209
          E—mail: cttl@chinattl.com       Http://www.chinattl.on


      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3935

Calibration Parameter Determined in Head Tissue Simulating Media
                       .                     sls                                                       6
 f [MHz]® Pe'::":t':i"‘,’;yF C°"‘:;Z:‘)"Fty ConvF X ConvF Y ConvF Z Alpha® D(:::; ::(:‘;t)
    750         41.9              0.89       10.6s   10.68   10.68   0.40 0.70 +121%
    835         41.5              0.90       10.36   10.36   10.36   010 167 +121%
    900         41.5              0.97       1022    1022    1022    016     129 £121%
   1750         40.1              1.37        8.85    8.85    8.85   0.22    1.08 £121%
   1900         40.0              1.40        8.41    8.41    841    0.26 0.94 +121%
   2000         40.0              140         8.41    8.41    841    0.28 0.90 £121%
   2300         39.5              1.67        8.39    8.39    8.39   0.44 0.79 +12.1%
   2450         39.2              1.80        7.87    7.987   7.87   0.53 0.73 £121%
   2600         39.0              1.96        7.67    7.67    7.67   0.60 0.68 £121%
   5200         36.0              4.66        5.91    5.91    591    0.35 150 £13.3%
   5300         35.9              4.76        5.63    5.63    5.63   0.35    150 £13.3%
   5500         35.6              4.96        5.29    5.29    5.29   0.35 1.60 £13.3%
   5600         35.5              5.07        5.08    5.08    5.08   0.35    1.60 £13.3%
   5800         35.3              5.27        515     515     515    0.40    140 £13.3%

C Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
£+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
" At frequency below 3 GHz, the validity of tissue parameters (s and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (€ and 0) is
restricted to +£5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
© Aipha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z17—97257                              Page 5 of 11


            ?@                       in Collaboration with
            7/"7L s__p                                             a
            paaar»" cnumnon?mofimm
          Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
          Tel: +86—10—62304633—2218     Fax: +86—10—62304633—2209
          E—mail: cttl@chinattl.com      Http://www.chinattL.en


      DASY/EASY — Parameters of Probe: EX3DV4 ~— SN: 3935

Calibration Parameter Determined in Body Tissue Simulating Media
                                             a                                                         G
 f [MHz]® P::i::“”;yF c°"°:;7r:‘)'ity ConvF X ConvF Y ConvF Z Alpha® D(::’;h) ::(’:;t)
    750       55.5         0.96        10.65   10.65   10.65   040 0.80 +121%
    835       55.2         0.97        1033    10.33   10.33   017     141 £12.1%
   1750       53.4         1.49         8.71    8.71    871    0.26 0.99 £121%
   1900       53.3         1.52         8.30    8.30    8.30   016 1.39 +121%
   2300       52.9         1.81         8. 10   8.10    810    0.32    116 £121%
   2450       52.7         1.95         7.99    7.99    799    0.29 125 +121%
   2600       52.5         2. 16        7.71    7.71    771    0.39 0.95 £121%
   5200       49.0         5.30         5.41    5.41    541    0.45 1.30 +13.3%
   5300       48.9         5.42         5.20    5.20    5.20   0.40    1.30 £13.3%
   5500       48.6         5.65         4.62    4.62    462    0.40    1.70 +13.3%
   5600       48.5         5.77         4.51    4.51    451    0.45    1.55 £13.3%
   5800       48.2         6.00         4.64    4.64    464    0.58 1.07 £13.3%

© Frequency validity above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
+50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is £ 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
" At frequency below 3 GHz, the validity of tissue parameters (¢ and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (£ and 0) is
restricted to £5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
© Aipha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z17—97257                              Page 6 of 11


         _;% in Collaboration with
         n                               s p e a g
              iaaagee>"                  CALIBRATION LAEORATORY
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2218                Fax: +86—10—62304633—2209
       E—mail: ctt!@chinattl.com                Http:/www.chinattl.on



                             Frequency Response of E—Field
                             (TEM—Cell: ifi110 EXX, Waveguide: R22)

              1.5 —
                    devinnencbennennnhannnnnndanencnndrnennenbrennnncdunennnnnhrnennnnnhonnnnnonhnenenendnnennnndonnnnnnde
       pgg
              14 Jrcsssfpernccoccciendiecceecfaceverfiaiceedecaveetfacasecaihaceveecavecelacceecancceck
                  esdfecccenderrrcenderersrnferenererfenencrerfecsnnnethreverribactnnenspranemondinevninferetenecocn
        @ 1.3 C cesccssc scccdocccccncdoncncnondonnnnencdnnnnnnenhnennnnnnhnnenemechen dncnedoecce en cd.s.
        'E              ........ i...--,-.-:.................. J: ........ deseae2s, i ........ kese2e.>. :L ........ E....----.i........ Jesesases 42.


        § 124. kschocfeooo o oo e
        00          {eeeaddeePporononnfioresseneferecesesfonsrensefunesnrechsrenosederencesedrrncneredocs
        § 11 4@—kgo—o oL {———————— fosskll>> fusslcll>— pe.——————een.~——esle———dfeell._} {——
        o    }........}\.----------------- puollkll> {......> {......_._. flspeelel}>>Recsen ensls e} feke l {——
        2 1.0 4.———pe >bg——4—=—— duzzzezzzizsaasas                                            —@—@¥—@i—— e==frncck
        o      .cc‘busscs se de se sene ds sl se huslllls fuulllll. puulkl—>> besllll}— besllll!>hovllll..helll.l. dhuislll}>
        & 0.9 4|esns edec se slsRe.l.~.~. ds feeel}}> fese———>> pesll}.>> fellll.}.p.ulll. .fuaill... flls
        0                   1                       i                                                  1 1 1        i         ;
        tw     ap enb n ns ncesn n ns a cessn ce dn mssn nen d en en en n f snn enb c se nn > s
        n     aoh ons on encb esn n ns thene ce ns donnnneen dosnon ons dosnenena ons se se nhe esn es
        3 °8
        &
                                 f                    f                               f                     f                     TCO
                    ks cl .. }.>>Rosasa=s ces e .>.>>feve s enc edes se es en des ns nssef en esn aafe en es e}> Recasassopesesc}>sfecsecesedes se ces 4. es
        $ 0.7 4/esncb s en > >feensseneden sns ende ns nnasp en en hen en k >+> Rbeneneneh en —< >
        o               l..l222.huls.ll..fon en en enn ee en e docnnenendensennenbessessecpree enbonenenecbesseeneceneeeeedesenec en dee.
        2
        u‘:
                                 |                    f                               bollldllll
              0.6 wap ce c > >>> §' """""""""E""""“‘“"""""""f """" :" """" Finininbnintuinint sinininiatnbnints': """""""""""
                                                                                                                                  ol
                    1   ...........................é................................... ::. -------- l--------\--------.i----------------- 44«




              0-5                I         l          ll        r          E          l          ;          I          €          '          C          l


                    0                   500                 1000                  1500                  2000                  2500                  3000
                                                E@3                              f [MHz]                            wa
                                                TEM                                                                 R22

                        Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: Z17—97257                                           Page 7 of 11


          ;fi’                      in Collaboration with
          77L s_p                                 e a
           \Wiape"                CALIERATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2218     Fax: +86—10—62304633—2209
        E—mail: ctti@chinattl.com      Http://www.chinattl.on



                               Receiving Pattern (®), 0=0°

              f=600 MHz, TEM                                                  f=1800 MHz, R22




                      —150           —100           «50              0            50          100         1560
                                                                  Rollf*]
                               [—*<~]100MHz       __—+—600MHz            __—+— 1800MHz   __—+— 2500MHz]
                             Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




Certificate No: Z17—97257                                 Page 8 of 11


                                                ®
                                                                                  In Collaborauon wi
                              TTL s
                                                                                  CALIBRATION LABORATORY
      Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, ChiIng
      Tel +86—10—62304633—2218      Fax +86—10—62304633—2209
      E—mail: cttl@chiinalttl.com   Htt /Iwww.ch inattl.en



                                                                    Dynamic Range f(SARneaq)
                                                                                                    (TEM cell, f 900 MHz)

                                                                                      carppet mcC
                                                                                               i
                                                                                      a~tntarrrrt
          [afjfreubig indyj




                                                                                                              *~rrrcersecmcansn=mer




                                                                                                                                                                                                   LELL L OC 222222 EGLR EOE O00020 OAELEREA
                                                                                                                                                                                                    Fitp 4 ob oto o un e ~ n d 4ih 6 oto ote e dn e n n d sds 4 ofe ofe n dn ~ R +
                                       f————— 48— 9b > ds 4s
                                   4




                                                                                                                                                                                                            ¥oY
                                                                                                                        apes ue c oo d nB 4 ofe sio o uge <o n 4 BE Bb 6 5o ob n oiz e cn



                                                                                                                        —=

                                                                                                                        a—
                                                                                                                         +
                                                                                                                         +

                                                                                                                         +
                                                                                                                         «
                                                                                                                         a
                                                                                                                         +
                                                                                                                         +
                                                                                                                         a

                                                                                                                         +

                                                                                                                         i
                                                                                                                           2422 2s d dE E4 ofo cho o ue e on d E6EA 4 oo ob o d e 2e e
                                                                                                                           s 4s n n 4 $ndin t obe ofs n to e in d BE e# ns ob n dn 9+ n
                                                                                                                           santncarrerrom mc usn nc erenpromomcuto—~




                                                                                                                                                                                                                     03
                                                                                                                              a




                                                                                                                                                                                                                              oOs S00
                                                                                                                                      rev e




                                                                                                                                                                                                                                                         oo icae
                                                                                                                                                                +




                                                                                                                                                                                                                                                                 +
                                                                                                                                                                          +0




                                                                                                                                  s>=~4>~—}>>                                                                                                                       42




                                                                                                                                                                r‘l'l]llll



                                                                                                                                                                                             10°
                                                                                                                        SAR[{mW/cm*
                                                                                                      —M—)not compensated                                                       —C@— compensated

                         2    =




                              2000002 Jhe—>
                                                    dpsccosems esn es 4ee ce >s




                                                                                                              10°                                                                           10‘
                                                                                                                     SAR[mW/cm*]
                                                                                   [—=—|not compensated            —e— compensated                                                                                                                                                   _|
                                                                                  Uncertainty of Linearity Assessment £0.9% (k=2)
Certificate No 717—97257                                                                                   Page 9 of 11


                                       In Collaboration with

        w77‘ s p e a g
                                       CALIBRANON LABORATORY
      Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
      Tel: +86—10—62304633—2218                     Fax: +86—10—62304633—2209
      E—mail: ctti@chinattl.com                     Http://www.chinattl.on



                        Conversion Factor Assessment
     f=850 MHz, WGLS R9(H_convrF)                                                                         f=1750 MHz, WGLS R22(H_convPr)
            3.50                                                                                                             30.00


            3.00 \                                                                                                           25.00 |
            2.50
                                                                                                                             20.00 \




                                                                                                               SAR[WKkgJ/W
          %2.00 t XA                                                                                                                \.
          S                                                                                                                  15.00   \g
          3 [|
            1.50

                                                                                                                             10.00
            1.00

            0.50                                                                                                              5.00
            0.00                                _               PebmeERSRSE        e3nn00+                                    809 LuzuLecl_ L |, rifrevptrrmetonsmennenls
                   0        20             40              so                 so               100                               0    400  200  30   400      500      s0   To
                                          z{mm]                                                                                                                   z{mm]
                                 —@—measured __——analytical




                       Deviation from Isotropyin Liquid
                                                       _________________                             :                                 ----........._____\__.._"-’
                                           -------                                                   t                                                            ;
                                 1.0

                                 0.8
                                 0.6
                                  0.4
                                  0.2
                                   0.0
                       7 Axis




                                  —0.2
                                   —0.4
                                   —0.6
                                    —0.8
                                    18
                                                     100
                                                           160
                                                     *4             200
                                                        Aris~                  2so
                                                                                             300
                                                                                                         350
                                                —1.0 —0.80 —0.60 040 —0.20                                O              020         040O.60 080               10
                                 Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z17—97257                                                      Page 10 of 11


         £®
                                  in Collaboration with


            aages" cmmnouimgrom
       Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
       Tel: +86—10—62304633—2218     Fax: +86—10—62304633—2209
       E—mail: cttt@chinattl.com      Http://www.chinattlen


  DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3935


    Other Probe Parameters
   Sensor Arrangement                                                      Triangular

   Connector Angle (°)                                                          42.2

   Mechanical Surface Detection Mode                                         enabled

   Optical Surface Detection Mode                                            disable

   Probe Overall Length                                                      337mm

   Probe Body Diameter                                                         10mm

   Tip Length                                                                   9mm

   Tip Diameter                                                               2.8mm

   Probe Tip to Sensor X Calibration Point                                      1mm

   Probe Tip to Sensor Y Calibration Point                                      1imm

    Probe Tip to Sensor Z Calibration Point                                     1imm

    Recommended Measurement Distance from Surface                             1.4mm




Certificate No: Z17—97257                              Page 11 of 11



Document Created: 2019-04-11 13:07:00
Document Modified: 2019-04-11 13:07:00

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC