T56XE1 Test Report HAC App. A

FCC ID: IHDT56XE4

Test Report

Download: PDF
FCCID_3786726

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/2/22


HAC_E_Dipole_835_180222
DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 123.4 V/m; Power Drift = -0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 109.6 V/m
Average value of Total=(109.6+107.6) / 2 = 108.6 V/m


                                         PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          106.0 V/m      109.6 V/m   108.3 V/m
                                          Grid 4 M4      Grid 5 M4   Grid 6 M4
                                          62.68 V/m      63.73 V/m   63.05 V/m
                                          Grid 7 M4      Grid 8 M4   Grid 9 M4
                                          105.9 V/m      107.6 V/m   105.9 V/m
                                                         Cursor:
                                                    Total = 109.6 V/m
                                                     E Category: M4
                                                Location: -1, -75.5, 9.7 mm




                     0 dB = 109.6 V/m = 40.80 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/2/22


HAC_E_Dipole_1880_180222
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 144.9 V/m; Power Drift = 0.10 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 93.35 V/m
Average value of Total=(91.47+93.35) / 2 = 92.41 V/m


                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          90.21 V/m      91.47 V/m   90.01 V/m
                                          Grid 4 M3      Grid 5 M3   Grid 6 M3
                                          70.88 V/m      72.25 V/m   71.49 V/m
                                          Grid 7 M3      Grid 8 M3   Grid 9 M3
                                          91.58 V/m      93.35 V/m   91.40 V/m
                                                         Cursor:
                                                    Total = 93.35 V/m
                                                     E Category: M3
                                                  Location: 0, 31, 9.7 mm




                     0 dB = 93.35 V/m = 39.40 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/3/5


HAC_E_Dipole_2450_180305
DUT: HAC Dipole 2450 MHz

Communication System: CW; Frequency: 2450 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.3 ℃

DASY5 Configuration
- Probe: EF3DV3 - SN4047; ConvF(1, 1, 1); Calibrated: 2018/1/8;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn853; Calibrated: 2017/7/19
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD2450 = 10mm & 15mm
2/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid: dx=0.5000
mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 79.25 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.62 V/m
Average value of Total=(90.72+92.62) / 2 = 91.67 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          89.74 V/m 90.72 V/m 87.84 V/m
                                          Grid 4 M3 Grid 5 M3 Grid 6 M3
                                          79.76 V/m 80.14 V/m 78.25 V/m
                                          Grid 7 M3 Grid 8 M3 Grid 9 M3
                                          91.25 V/m 92.62 V/m 89.70 V/m
                                                         Cursor:
                                                    Total = 92.62 V/m
                                                     E Category: M3
                                                Location: 0.5, 24.5, 8.7 mm




                     0 dB = 92.62 V/m = 39.33 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/2/22


HAC_E_Dipole_2600_180222
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.5 ℃

DASY5 Configuration
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn854; Calibrated: 2017/5/2
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 72.61 V/m; Power Drift = 0.05 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.36 V/m
Average value of Total=(84.42+92.36) / 2 = 88.39 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          83.91 V/m 84.42 V/m 82.63 V/m
                                          Grid 4 M3 Grid 5 M3 Grid 6 M3
                                          87.54 V/m 89.24 V/m 88.20 V/m
                                          Grid 7 M3 Grid 8 M3 Grid 9 M3
                                          90.76 V/m 92.36 V/m 90.94 V/m
                                                         Cursor:
                                                    Total = 92.36 V/m
                                                     E Category: M3
                                                  Location: 0, 20, 9.7 mm




                     0 dB = 92.36 V/m = 39.31 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/2/28


HAC_E_Dipole_2600_180228
DUT: HAC Dipole 2600 MHz

Communication System: CW ; Frequency: 2600 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
Ambient Temperature:23.3 ℃

DASY5 Configuration
- Probe: ER3DV6 - SN2358; ConvF(1, 1, 1); Calibrated: 2018/1/19;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn853; Calibrated: 2017/7/19
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD2600 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 72.50 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.18 V/m
Average value of Total=(84.26+92.18) / 2 = 88.22 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          83.74 V/m 84.26 V/m 82.47 V/m
                                          Grid 4 M3 Grid 5 M3 Grid 6 M3
                                          87.38 V/m 89.07 V/m 88.03 V/m
                                          Grid 7 M3 Grid 8 M3 Grid 9 M3
                                          90.58 V/m 92.18 V/m 90.76 V/m
                                                         Cursor:
                                                    Total = 92.18 V/m
                                                     E Category: M3
                                                  Location: 0, 20, 9.7 mm




                     0 dB = 92.18 V/m = 39.29 dBV/m



Document Created: 2018-03-06 00:03:53