RF Exposure Part 3

FCC ID: IHDT56XD7

RF Exposure Info

Download: PDF
FCCID_3817434

           m \n, Collsboration with
          a_‘/"/‘/, s p e a q
           eaaagee>"                CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2218       Fax: +86—10—62304633—2209
        E—mail: cttl@chinattl.com       Hittp:/www.chinattl.cn

Glossary:
DAE                            data acquisition electronics
Connector angle                 information used in DASY system to align probe sensor X
                               to the robot coordinate system.


Methods Applied and Interpretation of Parameters:
e DC Voltage Measurement. Calibration Factor assessed for use in DASY
   system by comparison with a calibrated instrument traceable to national
   standards. The figure given corresponds to the full scale range of the
   voltmeter in the respective range.

e    Connector angle: The angle of the connector is assessed measuring the
     angle mechanically by a tool inserted. Uncertainty is not required.

e    The report provide only calibration results for DAE, it does not contain other
     performance test results.




Certificate No: Z17—97153                               Page 2 of 3


           mm\©                     In Collaboration with

          ‘7/"7‘J, s p e a g
           aaaayey>""               CALIBRATION LABORATORY
        Add: No.51 Xueyuan Road, Haidian District, Beijing, 100191, China
        Tel: +86—10—62304633—2218        Fax: +86—10—62304633—2209
        E—mail: cttl@chinattl.com        Http:/www.chinattl.cn



DC Voltage Measurement
    A/D — Converter Resolution nominal
        High Range:       1LSB =       6.A1uV , full range =   —100...+300 mV
        Low Range:        1LSB =        61nV ,  full range =   ~Avsllll. +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors                    X                              Y                    Z.
      High Range                    404.898 + 0.15% (k=2)        405.241 + 0.15% (k=2)   404.618 + 0.15% (k=2)
      Low Range                     3.98840 + 0.7% (k=2)         3.99650 + 0.7% (k=2)    3.99854 + 0.7% (k=2)


Connector Angle


     Connector Angle to be used in DASY system                                                41.5° +1 °




Certificate No: Z17—97153                               Page 3 of 3


Phons sa1 1248 9700,Fan «d
inotspaay com, itspeon com



                                  IMPORTANT NOTICE


USAGE OF THE DAE4
The DAE unt is a delcate, high procision insrument and requires carefi treatment by the user. There are no
serviceable part inside the DAE. Specil attention shallbe given to he folowng ponts
Battery Exchange: The batley cover af the DAEA uniis lased using a serew, over lightening the serow may
cause the threads insde the DAE to wear out
Shipping of the DAE: Refore shipping the DAE to SPEAG for calbration, remove the batteris and pack the
DAE in an anitatie bag, This anistale bag shall en be packed nto larger box orcontainer which protects the
DAE from impacts during transporation. The package shallbe marked to indeate that a fragie instrument is
inside
EStop Fallures: Touch detecion may be malfinclioning due to broken magnels in e Estop. Rough handing
of the Etop may lead to damage of these magnels. Touch and calsionerars are often caused by dust and dit
accumulated in the Estop, To prevent Estop falure, the customer shall alvays maunt the probe to the DAE
earefily and keepthe DAE unt n a nor—dusty environimenti not used for measurements
Repalr: Minorrepats are performad at no extra cost duting the annval calbration, However, SPEAG reserves
the right to chargefor any repair especialy i rough unprofessional handing caused the defect
DASY Confiquration Files: Since the exact values of the DAE input resitances, as measured during the
eallration procedure ofa DAE un‘t, are not used by the DASY sofware, a norinal value o200 MOhm is ghen
in the corresponding conturaton fle:
important Note:
Warranty and calibration is void if the DAE unitis disassembled partly or fully by the
(Customer.

Important No
Never attompt to grease or ail the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by cortified SPEAG personnel only and is part of the annual
calibration procedure.


Important Not
[To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
\while turning the locking nut of the connector. The same care shall be used when
(disconnecting the probe from the DAE.


Scimid & Parner Engincemng
TNBRor0315AD DAES doe                                                                         11.12.2009


      r
Calibration Laboratory of                         «>                              SciweisrischrKaltrelonnt
Schmid & Partner                                  %                             z Servicesuisse tétstonnage
  Engineering AG                                  1oets                             Servzto avizero ttrntun
‘ghausstrase 4, 0004 zurich, Suttsatand           *fams                         . swis Caltration Serice

Accuiedby o Svis AccuctaionSevics (SAG)                                  reeredtston: SCS 0108
‘Tre Suiss AceredttionSrvieis oneofthesignstois ts the EA
HtatiterstAgreomentor threcogniton ofcalbratonconifetes
ce Sporton (Auden)                                                       Cetticas io: DAE4—1279_Jan18
CALIBRATION CERTIFICATE
 Oe                           DAE4 — SD 000 Do# BM — SN: 1279

 Cattaion pocedin(s           OA CAL—06.v20
                              Calibration procedure for the data acquisiion electronics (DAE)


 Catbaton ts                 January 03, 2018



 Thcalbaton corfeat decnats h ricnaviy o natena stardards,utich roatc h hyicaunte o messirenens S
 ‘The mossirements anh ureaininis wih conidanc obubityangiven on h otowingpages and e patofthe cotfeate
 Alcalbatons have boon condeto n thecosed aboratoylacieniorment temportiee(2 + 2and ramity «701
 CaltraionEnpmentused (UATE cntea for catbadon)
 Piman Stnduts                 o«                 calDate(contcwNo)                    scteded Catbraton
 Kathny MutinaarTypu 2001      it oiere           Sr¥eit Noz                           Aupis
 Sncondar Stamdarts            +                 tect Dare intouse)                    Scteded Creck
 Aco DNE Caitaten im           se is ons MA 1001.Osn17 in house crec                   in rousechesk sinto
 cittrBorvet                   se uis 0s M¥ 1002 Osvint7 n hoosectuct)                 Inhouse cteck ints




                               Name                         Functon                     Sqoauee
 Cattredby                     Atan Gaing                   LateratoyTechicin          A 77

 Appromby                      Snkim                        Dovuty mavage
                                                                                        i ,l/.(lgj UMMV'
                                                                                       issuet Jardarya, 2018
Ti cattatoncotfeatshal nebe mprdized ces n ul wthot wetenaeprovl t e latortary


Corticato No: DAE#—1270_Jant                        Page 1ot5


Calibration Laboratory of                                                 Schwoizerischor Kliriertanst
Schmid & Partner                                                        g Servien uisneialonnage
  Engineering AG                                                          Sorvito svizoro ditarurn
Hghousstraise i, 004 Zurich, Switzetand                                 S.—   swiss Calbration Sonvice

Accrudto by h SnisAccediaton Serice (GAS)                                Aestedtaton No: SCS 0108
The Suiss Aceredtation Service is oneofthesignntris to tEA
MulatoralAgroomentforthe recognitonofcalration corifentos

Glossary
Dae                     data acquisition electronics
Connector angle         information used in DASY system to align probe sensor X to the robot
                        coordinate system.
Methods Applied and Interpretation of Parameters
  * DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
        comparison with a calibrated instrument traceable to national standards. The figure given
        corresponds to the full scale range of the voltmeter in the respective range.
   * Gonnector angle: The angle of the connector is assessed measuring the angle
     mechanically by a tool inserted. Uncertainty is not required.
   *   The following parameters as documented in the Appendix contain technical information as a
       resultfrom the performance test and require no uncertainty.
       *    DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
            the nominal callbration voltage. Influence of offset voltageis included in this
            measurement.
       *    Common mode sensitivity: Influence of a positive or negative common modevoltage on
            the differential measurement.
       *    Channel separation:Influence of a voltage on the neighbor channels not subject to an
            input voltage.
        * AD Converter Values with inputs shorted: Values on the internal AD converter
          corresponding to zero input voltage
        * Input Offset Measurement: Output veltage and statistical results over a large number of
            zero voltage measurements.
        *   Input Offset Current: Typical value for information; Maximum channel input offset
            current, not considering the input resistance.
        *Input resistance: Typical value for information: DAE input resistance at the connector,
         during internal auto—zeroing and during measurement,
       * Low Battery Alarm Voltage: Typical valuefor information. Below this voltage, a battery
         alarm signal is generated.
       + Power consumption: Typical value for information. Supply currents in various operating
            modes.




Coriicate No: DAEA—1270_Janta                        Page 2015


DC Voltage Measurement
   A/D — Converter Resolution nominal
       High Range:          wse»       ety ,         fullrange = ~100.,4300 mV
       Low Range:           ise        stov .        fullrange 1......c3mV
   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring tme: 3 seo
     Callbration Factors            x                       v                     z
     High Range            403.996 + 0.02% (k—2) 4nson+ 0.00% (k=2) 404.618 2 0.02% (ke2)
     Low Range             a.o4027 + 1.50% (ke2) a.000t0 + 1.500%(k=2)   a.ceoon + 1.50% (ce2)

Connector Angle
     Gonnector Angle to be used in DASY system                               ssroee1®




Cerifeate No: DAE4—1270.Jant                     Page3or$


Appendix (Additional assessments outside the scope of SCS0108)
1. DC Voltage Linearity
     High Rango                               Reading (iV)       |_Ditterence (1¥)        Eiror (%)
     Ghannel x      + Input                     ressor64                10e                aco
     Channel X    + input                       2000480                280                 oo1
     Ghannel X____— input                       —rs09870               207                Eo
     Channel ¥    + input                       reosor3                 17e                oco
     Channel ¥     _ _+ input                   2000336                 rar               oo
     Ghannal Y     _ _— input                  —eoooe.7e               ce                  oo1
     Channel 2    + input                      19o007.co                186                oco
     Channel 2     _ _+ input                   20001.30               osr                —o00
     Channel 2     _ _— input                  2000042                 235                 oo
     Low Range                                Reaing (2V)        __|_Difference (iY)      Eror ()
     Ghannel X___+ input                        2ocnas                  ase                oos
     Ghannel X    + input                        z00e                    ts                asa
     Channel X____— input                       —r9082                   rar              se
     Channel Y    + input                       2oo1.79                 ogo                oce
     Channel ¥    + input                        zo18e                  oce                oo«
     Channel Y     _ — input                    19807                   ao1               00
     Ghannel 2    + input                       200213                  ase               oo3
     Channel 2    + input                        20117                  ws1
     Channel 2    > input                       1008                    omn               o«

2. Common mode sensitivity
   DASY measurement parameters: Auto Zero Time: 3 see; Measuring time: 3 seo
                    Common mode                     High Range                   Low Range
                    Input Vottage (m¥)        Average Reading (1¥)           Average Reading (1¥)
    Ghannel X                    200                   as51                       848
                                —200                   zes                         zore
    Channel ¥                    2o                     sao                         sar
                                —200                    se                         a15
    Channel2                     200                    osr                        c
                                200                     z90                        zse

3. Channel separation
   DASY measurement parameters: Auto Zere Time: 3 see; Messiring tme: see
                      Input Voltage (m¥)__|      Channe!X (@¥)     ChannelY (i¥)       ChannelZ (@¥)
    Channel xX                200                      —               se                  385
    Channel ¥                 200                    asr                                   4so
    Channelz                  200                    103                  sso                —

Cerifcate No: DAE#—1270_Jante                           Pagedot5


4. AD—ConverterValues with inputs shorted
   DASY mensurement parameters: Auto Zero Time: 3 see; Measuring tme: 3 sec
                                             High Range (L98)             Low Range (L98)
    Channel X                                       1so00                         18110
    Channel ¥                                       1sose                         17200
    Channel z.                                      1sros                         isriz

5. Input Offset Measurement
   DASY measurement parameters: Asto Zoro Time: 3 sec; Measuring time: 3 seo
   inour ton
                         Average (i¥)      min. Offset (i¥) max. offset yy)       * Z"“;;'“""
    Channol X                105                oz               2s                 oae
    Channal ¥                028                use              oss                o«2
    Channel 2               s08                 —1ss             205                asr

6. Input Offset Current
    Nominalinputcroutry ffset current on alchannels: «251A

7. Input Resistance (Tpicalvalues fo infomation)
                                            Zeroing (kOhm)                Measuring (Whm)
    Channel xX                                    2o                             2o
    Channol‘Y                                     200                            2o
    Channel2                                      200                            200

8. Low Battery Alarm Voltage(pica! values for information)
    Typical values                           Alarm Lovel (VDG)
    Supply (+ Vee)                                                   «9
    Supply (—Vee)                                                    2s

9. Power Consumption (Typicalvalues for nformaton)
    Typical values                      Switehed off (ma) Stand by (m)         |_Transmiting (ma)
    Supply (+ Veo)                           w00               «s                      w
    Supply (—Vee)                            —oo1                s                     s




Gorlicate No: DAEH—1270_Jantg                       PageSols


                                 ;a
                                  uctusiom T& mm t
                    TH causranion Lasoratory                                     7 /\            & CNAs a’lwflon
            Add: No 51 Xueyuan Road. Haidian Distrct Beiing, 100191, China        % '»@\“e                        CNAS L570
            Tel: +86—10—62304633—2218          Facc +86—10—62304633—2200
            E—mail: ct@chinatl.com             chinatl.en
      Client             Auden                                              Certificate No: Z17—97052



Object                                    EX3DV4 — SN:3763

Calibration Procedure(s)                  EF—Z11—004—01

                                          Calbration Procedures for Dosimetric E—field Probes

Calibration date:                         May 05, 2017

This calibration Certificate documents the traceabiity to national standards, which realize the physical units of
measurements(S)). The measurements and the uncertainties with confidence probability are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facility: environment temperature(@22:3)C and
humidity<70%.


Calibration Equipment used (M&TE critical for calibration)
Primary Standards                       ID #          Cal Date(Calibrated by, Certificate No.)      Scheduled Calibration
 Power Meter        NRP2         101919                 27—Jun—16 (CTTL, No.J16X04777)               Jun—17
 Power sensor NRP—Z291           101547      27—Jun—16 (CTTL, No.J16X04777)                          Jun—17
 Power sensor NRP—Z291           101548      27—Jun—16 (CTTL, No.J16X04777)                          Jun—17
 Reference10dBAttenuator         18N5OW—1048 13—Mar—16(CTTL.No.J16X01547)                            Mar—18
 Reference20dBAttenuator         18N5OW—200B 13—Mar—16(CTTL, No.J16X01548)                           Mar—18
 Reference Probe EX3DV4          SN 7433     26—Sep—16(SPEAGNo.EX3—7433_Sep16)                       Sep—17
 DAE4                            SN 549      13—Dec—16(SPEAG, No.DAE4—549_Dec16)                      Dec—17

 Secondary Standards     iD #                           Cal Date(Calibrated by, Certificate No.)    Scheduled Calibration
 SignalGeneratorMG3700A 6201052605                      27—Jun—16 (CTTL, No.J16X04776)               Jun—17
 Network Analyzer ES071G MY46110673                     13—Jan—17 (CTTL, No.J17X00285)
                               Name                           Function
Calibrated by:                   Yu Zongying                 SAR Test Engineer

Reviewed ty                      Lin Hao                     SAR Test Engineer
Approved by:                     Qi Dianyuan                 SAR Project Leader

                                                                                      Issued: May 06, 2017
This callbration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: Z17—97052                               Page 1 of 11


                                   causranon Lasoratory
         Add: No.51 Xueyuan Road. Haidian District, Beiing, 100191, China
         Tel: +86—10—62304633—2218      Fax: +86—10—62304033—2200
         E—mail:ct@chinat.com            Htphoowchinatilen

Glossary:
TSL             tissue simulating liquid
NORMxy,z        sensitivity in free space
ConvF           sensitivity in TSL / NORMxy.z
DCP             diede compression point
CF              crest factor (1/duty_cycle) of the RF signal
AB.C.D          modulation dependent linearization parameters
Polarization ®  ©® rotation around probe axis
Polarization 0  8 rotation around an axis that is in the plane normal to probe axis (at measurement center), i
                8=0 is normal to probe axis
Connector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatia—Averaged
  Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
  Measurement Techniques", June 2013
b) IEC 62208—1, Procedure to measure the Specific Absorption Rate (SAR) for hand—held devices used
     in close proximity to the ear(frequency range of 300MHz to 3GHz), February 2005
c) IEC 62208—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency rangeof 30 MHz to 6 GHz), March
   2010
d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
* NORMxy,z: Assessed for E—field polarization 0=0 (fs900MHz in TEM—cell; {> 1800MHz: waveguide}.
     NORMxy,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
      E" feld uncertainty inside TSL (see below ConvF).
*    NORM(xy2 = NORMx.y,2" frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
      frequency response is included in the stated uncertainty of ConvF.
*     DCPx,y,2: DCP are numerical linearization parameters assessed based on the data of power sweep
      (no uncertainty required). DCP does not depend on frequency nor media.
«_    PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
      characteristics.
*     Axyz; Buy.z: Coyz;VRxy,z:A,B,C are numericallinearization parameters assessed based on the
      data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
       media, VR is the maximum calibration range expressed in RMS voltage across the diode.
*      ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
       Transfer Standard for fs800MHz) and inside waveguide using analytical field distributions based on
       power measurements for f >B0OMHz. The same setups are used for assessment of the parameters
       applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
       These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
       The sensitivity in TSL corresponds to NORMx,y,z* Conywhereby the uncertainty corresponds to
       that given for ConvF. A frequency dependent ConvF is used in DASY version 4.4 and higher which
       allows extending the validity from«50MHz to*100MHz.
*      Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a fiat
       phantom exposed by a patch antenna.
*      Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the
       probe tip (on probe axis). No tolerance required.
*      Connector Angle: The angle is assessed using the information gained by determining the NORMx
       (no uncertainty required).



Certificate No: Z17—97052                                Page 2 of 11


                                causranon LasorAtory
       Add: No.51 Xueyun Road, Haidian Distict, Beiing. 100191, China
       Tel: +86—10—62304633—2218     Fac: +86—10—62304633—2209
       E—mail: ctl@chinatl.com        ftpohoww chinat.en




                     Probe EX3DV4


                                           SN: 3753


                                       Calibrated: May 05, 2017

                              Calibrated for DASY/EASY Systems
                                     (Note: non—compatible with DASY2 system!)




Certificate No: Z17—97052                            Page 3 of 11


        Add: No.51 Xueyuan Road. Haidian District, Beiing, 100|9| China
        Tel: +86—10—62304633.2218     Faxc +86—
        E—mail: cl@chinat.com          Htpshwsewchinatton

     DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3753

Basic Calibration Parameters
                                   Sensor X               Sensor Y              Sensor Z        Unc (k22)
 Norm(gVi(Vim))*                   0.47                   0.30                  0.46            £10.0%
 DGP(mV)®                          1014                   107.2                 104.5

Modulation Calibration Parameters
 uin         Communication                       A           C            c            p      VR         Unc®
             System Name                         dB          dB—pV                     dB     mV         (k22)
 0           Cw                         x        0.0         0.0          1.0          0.00   185.8__|   22.0%
                                        Y        0.0         0.0          1.0                 1414
                                        2        0.0         0.0          1.0                 182.6


     The reported uncertainty of measurement is stated as the standard uncertainty of
     Measurement multiplied by the coverage factor k=2, which for a normal distribution
     Corresponds to a coverage probability of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E*—field uncertainty inside TSL (see Page 5 and Page 6).
*2 Numericallinearization parameter: uncertainty not required.
© Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: Z17—97052                              Pages of 11


            i1 in Colliboretion vith
            uxT"TI a
            Clz CAUIERATION LABORATORY
          Add: No.51 Xueyuan Road. Haidian Distrct Beijing. 100191, China
          Te+86—10—62304633—2218        Fax: +86—10—62301633—2209
          E—mal: et@chinatt.com          Hiphwwwchinatilen

      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3753

Calibration Parameter Determined in Head Tissue Simulating Media
                                                                                                         6
 t ukz) Pe:';:::;y — c°"‘::7:‘)’f' ConvF X ConvF Y ConvF 2Z Alpha® D(::’:; :i':;‘
    Tso     419          use         oz      o2      o4z      oso_| 080 £121%
    a35     a15          o.80        ais     ais     ais      of2 148| 121%
    2oo     a15          o7          a20     a20     920      015| 142 £121%
   1750     401           137        si6     816     _|_a16_| o2     112 £121%
   1900     0.0           140        77e     7e      17e      om     115 £121%
   2000           40.0                 140             7.80             7.80    7.80    0.19      1.18       £12.1%
   2450           302                 180              Tzs              12      128     os1_|     on         £121%
   2600           300                 196              Tzo              120     120     061       068        £121%
   3500           arg                 291              Tor              1or     702|    oS6       085        £133%
   5250           350                 am               s25              525     525     045       120        £133%
   5600           355                 507              a7s              475     475     045       130        £133%
   5750           En                  5.22             are              a76_|   ar6_|   045       130        £133%

© Frequency validity above300 MHz of +100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for Conv assessments at 30, 64, 128,
150 and 220 MHiz respectively. Above 5 GHz frequency validity can be extended to + 110 MHz.
"Atfrequency below 3 GHz, the validity of tissue parameters (e and 0) can be relaxed to +10% if iquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (c and 0) is
restricted to £5%. The uncertainty is the RSS of the Conv uncertainty for indicated target tissue parameters
GAlphalDepzh are determined during callbration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than 2 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z17—97052                             Page 5 of 11


                                    In Colleboration with
            %“TTL a
          Add: No.51 Xueyuan Road. Haidian Distric, Beijing. 100191 China
          Tel: +86—10—62304633—2218     Fax:+86—10—62301633—2209
          E—mailetl@chinatt.com           itp:fwwnchinattien

      DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3753

Calibration Parameter Determined in Body Tissue Simulating Media
                                                                                                        x
 t mb P"::’;‘::;y, c°"‘::,°:‘)’fy ConvF X ConvF Y ConvF Z Alpha® D(:f:; :L':;;
    Tso           555                 096              a4s                 ads   o43     oso       080 £121%
    ses           552                 oar              ar                  am    am      oir       130 £121%
    200           s50                 105              ars                 ars   am      orm       116 |£121%
   1750           534                 149              Ter                 1er   187     020       118 £121%
   1900           s33                 152              Tss                 156   758     017       132 £121%
   2000           sa3                 152              Teo                 Tso   760     o7        141 £121%
   2450           527                 195              Tar                 12    122     02e       146 £121%
   2600           s25                 216              T14                 714   714     031|      126 £121%
   3500           513                 331              64s                 648   646     oss       oss £133%
   5250           489                 5.36             487                 487   487     055       120 £133%
   5600           455                 577              a2r                 42    422     oss       130| £133%
   5750           483                 5.94             as2                 as2   452     055       1.60 £133%

© Frequency validty above 300 MHz of £100MHz only applies for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency valiity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for Conv assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency validity can be extended to £ 110 MHz
"At frequency below 3 GHz, the validity of tssue parameters (¢ and 0) can be relaxed to +10% if iquid compensation
formula is applied to measured SAR values. At frequencies above 3 GHz, the valdity of tissue parameters (c and 0is
restricted to +5%. The uncertainty is the RSS of the ConvF uncertainty for indicated target tissue parameters.
©Alpha/Depth are determined during callbration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GH at any distance larger than half the probe tip diameter from the boundary.




   Certificate No: Z17—97052                                Page 6 of 11


                       SS in Coliibcretion vith
                       \igmzzzme~~                          CAUIBRATION LABORATORY
      Add: No.31 Xuyuan Road. Haidian District, Beiing. 100191, China
      Tel: +86—10—62304633—2218                                   Fac +86—10—62304033—2200
      E—mailett@chinatt.com                                       Hitpilhiwschinattlen


                                                      Frequency Response of E—Field
                                                      (TEM—Cell: ifi110 EXX, Waveguide: R22)
                                        s
                                        a
                                        2
      Frequency response (normalized)
                                        s
                                        6
                                        ob
                                        n
                                        s
                                        2
                                        s
                                        o
                                        s
                                        b
                                        s
                                        e
                                        s
                                        \
                                        e
                                            0




                                                                             T               T       y          T
                                        o
                                        in




                                                0           500            1000            1500    2000       2500   3000
                                                                  E9                                      _
                                                                  Tem                     Limra)          S
                                                    Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certiicate No: Z17—97052                                                          Page 7 of 11


       Add: No31 Xueyuan Road, Haidian District, Beiing, 100191, China
       ‘Tel: +86—10—62304633—2218         Fac: +86—10—62304033—2200
       E—mail:cil@chinatl.com              Itphnwwchinattlen


                              Receiving Pattern (®), 0=0°

             f=600 MHz, TEM                                                  f=1800 MHz, R22


                                                                                       s Q_.\’. a
                                                                                           /X




                     +               +                +                  m———i————i———
                   150              100              EJ               0           so        100      150

                             I~«—[1O0MHz           _—+—60OMHz             ——— 1800MHz   ———2500MHz
                            Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




Certificate No: Z17—97052                                 Page 8 of 11


         SW1 io Colsboration vith

        i/                          causranion Lasorstory
       Add: No.51 Xueyuan Road. Haidian Distrct, Beijing, 100191, China
       "Tel: +86—10—62304633—2218          Fax +86—10—62301633—2200
       E—mail: ctl@chinat.com               Atpihowwchinatlcn


                               Dynamic Range f(SARneaq)
                                                 (TEM cell, f = 900 MHz)



                10‘ 4



           E 10°4
           m
           C
           R
           C
           ¥%
           a
           &

                10°



                10°                          T                   T                  T                  T           1
                      10°                   10°                 10°                10‘             10°         10°
                                                                 SAR[mW/cm‘]
                                                   EW—not compensated     —@— compensated




               24                      T                    T             r    T                   T       :   1
                10                    10                   10°                410                 10           10
                                                                SAR[mWicm ]
                             [___E®—\not compensated                                    —s— compensated _|
                                Uncertainty of Linearity Assessment: £0.9% (k=2)
Certificate No: Z17—97052                                Page 9 of 11


        r\’ in Collboration with
        e=‘7"7‘J, a
       Add: No.51 Xueywan Road. Haidian Distrc, Beiing, 100191, China
       Tel: +86—10—62304633—2218      Fax: +86—10—62304633—2209
       E—mail: cti@chinatl.com          lip:/wwvchinati.en

                        Conversion Factor Assessment
     f=900 MHz, WGLS R9(H_convF)                                  f=1750 MHz, WGLS R22(H_convF)
             «s                                                               z0              y

             # k——————————                                                    s                                    _
                                            <                                           |                    T
             30             ———}—
                                                                              mm        :                      _
           5*                      t        1154                        |%i             |                      L
                                   I

           F
           $i»                              t      —!—                  § 100      |—¥:                —/—
           §.! AX__| _‘                                                 § ta00                    |                —
                                                                                                                       |
             as                        imcicct                                *TR T 1
             sn
                  o       a        ©        w       m        im
                            Htral                                                                      aton)
                            _                                                               [—<t—measires
                                                                                             _—anricd



                      Deviation from Isotropy in Liquid




                              sz
                      zis




                              04
                              «o8
                              s
                               ,1550




                                       no   oe0 om o4 am          0           o2   o4   o%   om   10
                              Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z17—97052                            Page 10 of 11


        <TTL LnLn_
       Add: No.51 Xueywan Road. Haidian Distict, Beling, 100191, China
       Tel: +86—10—62304633—2218      Fax: +86—10—62304633—2209
       E—mai:etd@chinat.com           Hitpihewwchinat.cn

  DASY/EASY — Parameters of Probe: EX3DV4 — SN: 3753


    Other Probe Parameters
   Sensor Arrangement                                                    Triangular

   Connector Angle (°)                                                         36.9

   Mechanical Surface Detection Mode                                       enabled

   Optical Surface Detection Mode                                           disable

    Probe Overall Length                                                    337mm

    Probe Body Diameter                                                      10mm

   Tip Length                                                                 9mm

    Tip Diameter                                                            2.5mm

    Probe Tip to Sensor X Calibration Point                                   1mm

    Probe Tip to Sensor Y Calibration Point                                   1mm

    Probe Tip to Sensor Z Calibration Point                                   1mm

    Recommended Measurement Distance from Surface                            14mm




Certificate No: Z17—97052                             Page 11 of 11


                                   5in              hon wih                        «J3             m xn
                ETTL v‘a mem@GNASE®"
                                         CAHBRATION tABoRiONN                     KZ7zy                           cmmnon
             Add: No.51 Xueyuan Road, Haid          ict.Belling, 100191, China     »/“;"\                         CNAS LO570
             Tel: «6—10—62304633—2218        Fax:   +86—10—62301633—2200
             E—mail:cl@chinattl.com              tipohinkd      L
      Client            Sporton InternationalINC                            Certificate No:    217—97152

CALIBRATION CERTIFICATE

Object                                    ES3DV3 — SN:3203

Calibration Procedure(s)                  EF—211—004—01

                                          Callbration Procedures for Dosimetric E—field Probes

Calbration date:                          September 25, 2017

This calibration Certificate documents the traceabiity to national standards, which realize the physical units of
measurements(S)). The measurements and the uncertainties with confidence probabilty are given on the following
pages and are part of the certificate.

All calibrations have been conducted in the closed laboratory facilty: environment temperature(@22s3)© and
humidity<70%.

Calibration Equipment used (M&TE critical for callbration)
Primary Standards                    D#             Cal Date(Calibrated by, Certificate No.)        Scheduled Calibration
 Power Meter      NRP2            101919              27—Jun—17 (CTTL, No.J17X05857)                Jun—18
 Power sensor      NRP—291        101547              27—Jun—17 (CTTL, No.J17X05857)                Jun—18
 Power sensor      NRP—291        101548              27—Jun—17 (CTTL, No.J17X05857)                Jun—18
 Reference10dBAttenuator          18N50W—10d8         13—Mar—16(CTTLNo.J16X01547)                   Mar—18
 Reference20dBAttenuator          18N50W—200B         13—Mar—16(CTTL, No.J16X01548)                 Mar—18
 Reference Probe EX3DV4           SN 7433             26—Sep—16(SPEAG,No.EX3—7433_Sep16)            Sep—17
 DAE4                             SN 549              13—Dec—16(SPEAG, No.DAE4—549_Dec16)           Dec—17


 Secondary Standards              ID #                Cal Date(Calibrated by, Certiicate No.)      Scheduled Calibration
 SignalGeneratorMG3700A        6201052605             27—Jun—17 (CTTL, No.J17X05858)                Jun—18
 Network Analyzer ES071C       MY46110673             13—Jan—17 (CTTL, No.J17X00285)                Jan—18
                              Name                         Function                                  Signature
Calbrated by:                    Yu Zongying                SAR Test Engineer

Reviewed by:                     Zhao Jing                 SAR Test Engineer                           gz/

Approved by:                     Oii Dian
                                      Dianyuan              SAR Project
                                                                  roject LLeader                %

                                                                                     Issued: September 27, 2017
This callbration certificate shall not be reproduced except in full without written approval of the laboratory.



     Certificate No: Z17—97152                             Page 1 of 11


                              "— in Collaborationwith

                                 causranon Lagorarory
        Add: No.S1 Xueywan Road. Hnidian District, BeiJing, 100191, China
            +o10—ea30d633—2218        Fav +86—10—02304603—2200
        E—mall citl@chinatl.com       Hips/hnwnchinatl.on
Glossary:
TSL                   tissue simulating liquid
NORMxy.z              sensitivity in free space
ConvF                 sensitivity in TSL / NORMxy,z
DCP                   diede compression point
CF                    crest factor (1/duty_cycle) of the RF signal
AB.C,D                modulation dependent linearization parameters
Polarization ®   ©® rotation around probe axis
Polarization 0   8 rotation around an axis that is in the plane normal to probe axis (at measurement center)
                 8=0 is normal to probe axis
Con nector Angle information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—Averaged
   Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices:
  Measurement Techniques®, June 2013
b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate (SAR) from
   hand—held and body—mounted devices used next to the ear (frequency range of 300 MHz to 6 GHz)°,
   July 2016
c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication
   devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)*, March
   2010
d) K DB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"
Methods Applied and Interpretation of Parameters:
     NORMs,y,z: Assessed for E—field polarization 6=0 (fs900MHz in TEM—cell; {> 1800MHz: waveguide),
     NORMx.y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not effect the
      E" field uncertainty inside TSL (see below ConvF)
     NORM(Dx,y.z        IORMx.y,2" frequency_response (see Frequency Response Chart). This
     linearization is implemented in DASY4 software versions later than 4.2. The uncertainty of the
     frequency response is included in the stated uncertainty of ConvF.
     DCPx.y,z: DCP are numerical linearization parameters assessed based on the data of power sweep
     (no uncertainty required). DCP does not depend on frequency nor media.
     PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on thesignal
     characteristics.
     Axy.z; Bx.y.z; Cxy,2z; VRXy.2:A,B,C are numerical linearization parameters assessed based on the
     data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
     media. VR is the maximum calibration range expressed in RMS voltage across the diode.
     ConvF and Boundary Effect Parameters: Assessed in flat phantom using E—field (or Temperature
     Transfer Standard for f5800MHz) and inside waveguide using analytical field distributions based on
     power measurements for f »800MHz. The same setups are used for assessment of the parameters
     applied for boundary compensation (alpha, depth) of which typical uncertainty valued are given.
     These parameters are used in DASY4 software to improve probe accuracy close to the boundary.
     The sensitivity in TSL corresponds to NORMx,y,z* Conv whereby the uncertainty corresponds to
     that given for ConvF. A frequency dependent Convr is used in DASY version 4.4 and higher which
     allows extending thevalidity from::50MHz tos100MHz.
     Spherical isotropy (3D deviation from isotropy); in a field of low gradients realized using a flat
     phantom exposed by a patch antenna.
     Sensor Offset: The sensor offset corresponds to theoffset of virtual measurement center from the
     probe tip (on probe axis). No tolerance required.
     Connector Angle: The angle is assessed using the information gained by determining the NORMx
     (no uncertainty required)


Certi ficate No: Z17—97152                             Page 2 of11


        Tel: «86—10—            P +86—10—62301633—2200
        E—mail: ul              Mitpihwwschinathen




                       Probe ES3DV3


                                    SN: 3293


                             Calibrated: September 25, 2017

                            Calibrated for DASY/EASY Systems
                               (Note: non—compatible with DASY2 system!)




Certificate No: 217—97152                  Page 3 of11


                                °in Collboration with

                                      causranon Lasoratory
          Add: No.51 Xueyuan Road. Haidian District, Beijing. 100191, China
          Tel: +86—10—62304633—2218       Fax: +86—10—62304633—2200
          E—mail: ctl@chinatl.com         Hipcwschinatlen

      DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3293

Basic Calibration Parameters
                                      Sensor X               SensorY                Sensor Z        Un (k=2)
 Norm(pVi(Vim)®)                       112                   0.90                   0.76            £10.0%
 DCP(mV)®                              1055                   108.7                 109.8

Modulation Calibration Parameters
  uin          Communication                        A            B            c            p      VR         Une®
               System Name                          dB           dByuV                     dB     mV         (k=2)
  0            ow                          x        |0.0         0.0          1.0          0.00   |286.1     12.8%
                                           vy       _|0.0        0.0          1.0                 254.6
                                           2        0.0          0.0          1.0                 232.5


      The reported uncertainty of measurement is stated as the standard uncertainty of
      Measurement muftiplied by the coverage factor k=2, which for a normal distribution
      Corresponds to a coverage probability of approximately 95%.


* The uncertainties of Norm X, Y, Z do not affect the E*—field uncertainty inside TSL (see Page 5 and Page 6).
* Numerical linearization parameter: uncertainty not required.
© Uncertainly is determined using the max. deviation from linear response applying rectangular distribution
and is expressed for the square of the field value.




Certificate No: 21797152                                 Page 4 of 11


                                 " in Collaborationwith
                   TTL a
           Add: No.51 Xueyuan Rood, Haidian Distrct Beijing, 100191, China
           ‘els rb6—10—62304633—2218     Paxc +86—10—02304633—2200
           t—mail: l @rchinatl.com        Hipi/nwchinatilen

       DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3293

Calibration Parameter Determined in Head Tissue Simulating Media
                                                                                                           &
 t mkz© Pe:';:::;y g c""‘::f:‘)’i" ConvF X ConvF Y GonvF 2 Alpha® r‘::""':',                                   :‘k:;:
    T50            a10                 se               a4ar           647   647|       060        115|        +121%
    835            415                 0.90             630            630   630|       041|       147_|       +121%
    900            a15                 nor              6x             es    es         041|       149         |+121%
   1450            405                 120              545            545   545        032        187         121%
   1750            401                 137              a32            532   532        06         125         +121%
   1900            a0.0                140              s23            523   523        o6§        123         +121%
   2000            20.0                140              ase            408   498        043        160         £121%
   2300            305                 167              aso            490   490|       090        110|        £121%
   2450            302                 180              ars            47    a73        o%8        114|        121%
   2600            300                 196              aso            450   450|       080        110|        +121%

© Frequency validity above 300 MHz of £100MHz only apples for DASY v4.4 and higher(Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of Conve uncertainty at calibration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for ConvF assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency valdity can be extended to + 110 MHz
"At frequency below 3 GHz, the validty of tissue parameters (c and 0) can be relaxed to 210% if liquid compensation
formula is applied to measured SAR values. At frequencies above3 GHz, the validity of tissue parameters (e and 0) is
restricted to +5%. The uncertainty is the RSS of the Conv uncertainty for indicated target tissue parameters.
° Alpha/Depth are determined during callbration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is alwaysless than + 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHzat any distance farger than half the probe tip diameter from the boundary.




  Certificate No: Z17—97152                             Page S of 11


           Add: No.S1 Xueyuan Road. Haidian District, Beiling, 100191, China
           Tels +86—10—62304633—2218      Fac +86—10—6230403—2200
           E—mail: etl@@chinattl.com       Hitpi/wnchinatiien

       DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3293


Calibration Parameter Determinedin Body Tissue Simulating Media
                                                                                                         s
 f uHz® Fe::"::::;y ,              c°"°:;f:‘;','y ConvF X ConvF Y ConvF 2 Alpha® D:“":)                      :i::')
    Tso            55.5                 0.96             643             643   643     o0         125        +121%
    35             55.2                 o7               e1g             a19   a19     040        1160       £121%
   1750            534                  149              s0s             505   505     o6é        127        £121%
   1900            533                  152              ase             48e   ase     om         117        +121%
   2450            527                  195              aso             ase   ase     ore        131        +121%
   2600            525                  216              a16             a16   416     080        115|       £121%

© Frequency validity above 300 MHz of +100MHz only apples for DASY v4.4 and higher (Page 2), else it is restricted to
#50MHz. The uncertainty is the RSS of ConvF uncertainty at callbration frequency and the uncertainty for the indicated
frequency band. Frequency validity below 300 MHz is + 10, 25, 40, 50 and 70 MHz for Convr assessments at 30, 64, 128,
150 and 220 MHz respectively. Above 5 GHz frequency valldity can be extended to # 110 MHz.
"At frequency below 3 GHz, the validity of tssue parameters (¢ and 0) can be relaxed to £10% if liquid compensation
formula is applied to measured SAR values. Atfrequencies above 3 GHz,the validity of tissue parameters ( and 0) is
restricted to £5%. The uncertainty is the RSS of the Convr uncertainty for indicated target tissue parameters
©Aipha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary
effect after compensation is always less than : 1% for frequencies below 3 GHz and below + 2% for the frequencies
between 3—6 GHz at any distance larger than half the probe tip diameter from the boundary.




  Certificate No: Z17—97152                               Page 6 of 11


                                                        °in Collborationwith
                                  u/"7°"J, a
                                                           causRAnon LaBORArORY
                 Add: No.51 Xueyuan Road, Haidian District, Bejing. 100191, China
                 Tel: +86—10—62304633—2218     Fax +86—10.62304633—2200
                   E—mail:tl@@chinatl.com                       Bipi/twwchinatt.en

                                                    Frequency Response of E—Field
                                                     (TEM—Cell: ifi110 EXX, Waveguide: R22)
       Frequency response (normalized)
                                         s
                                         oi
                                         s
                                         n
                                         —
                                         s
                                         s
                                         o
                                         o
                                         e
                                         e
                                         e
                                         s
                                         ~
                                         e
                                         a
                                         8
                                         in




                                              0           500          1000          1500           2000         2500   3000
                                                                E23                  f [(MHz               oo
                                                                TEM                   I        ul          Raz

                                                  Uncertainty of Frequency Response of E—field: £7.4% (k=2)




Certificate No: 217—97152                                                  Page 7 of 11


                     rot3—2218       Fix: +86—10—62304033—2200
        E—mail:   l@chinatt.com      MipeAwwncchinatilen
                              Receiving Pattern (®), 0=0°

              f=600 MHz, TEM                                           f=1800 MHz, R22




                                                t                               t     t
                     «1so         «100         «o                  o     so    100   150


                            Uncertainty of Axial Isotropy Assessment: £1.2% (k=2)




Certificate No: Z17—97152                           Page 8 of 11


          r\' in Collaboration with
          ‘TTL! p_ e a g
        Add: No.51 Xueyuan Road. Haidian District, Beljng, 100191, China
        Tel: +86—10—62304638—2218      Fan+86—10—62304633—2200
        E—mail: e@chinat.com            Hipuhiosscchinat
                                Dynamic Range f(SARneaq)




                                                                                                            i
                     10°                      10°                           10°                 10         10°
                                                              SAR[mW/cm‘]
                                                FW—}not compensated               —@~ compensated




                                                                     1:)'         5
                                                           SAR[mWiem‘
                            [__T—=—Inot compensated                                   —e— compensated _]
                                Uncertainty of Linearity Assessment: £0.9% (k=2)

Certificate No: 217—97152                             Page 9 of 11


        Add: No.31 Xueyuan Road. Haidian Distrit, Beijing. 100191, China
        Tel: +86—10—62304633—2218      Fax        n—6e2301633—2200
        E—mal: ctl@chinat.com             pilwwchinailco

                          Conversion Factor Assessment

       f=835 MHz, WGLS R9(H_convF)                                        f=1750 MHz, WGLS R22(H_convF)
                    aso                                                     se
                    ase

                    aso
          sanpgyy




                    200

                    1so
                    100

                    ase

                    aco
                                       zoo
                            [~<e—messired——wrabed




                            10
                             o8
                             cs
                             o«
                             02
                             oo
                             02
                             o4
                             «s
                             «8
                              14




                                     40 am com omm ze                 o    o7 o« om om   io
                            Uncertainty of Spherical Isotropy Assessment: £3.2% (K=2)



Certificate No: Z17—97152                             Page 10 of 11


        Add: No.51 Xueyuan Road. HaidianDisrict, Beijing, 100191, China
        Tels +86—10—62304033—2218      Fax: +86—10—62301603—2200
        —mail: ctl@chinatl.com         Httpshwchinatt.en

    DASY/EASY — Parameters of Probe: ES3DV3 — SN: 3293


     Other Probe Parameters
    Sensor Arrangement                                                    Triangular

    Connector Angle (°)                                                          7.2

    Mechanical Surface Detection Mode                                       enabled

    Optical Surface Detection Mode                                          disable

    Probe Overall Length                                                    337mm

    Probe Body Diameter                                                      10mm

    Tip Length                                                               10mm

    Tip Diameter                                                               4mm

    Probe Tip to Sensor X Calibration Point                                    2mm

    Probe Tip to Sensor Y Calibration Point                                    2mm

    Probe Tip to Sensor Z Calibration Point                                    2mm

    Recommended Measurement Distance from Surface                             3mm




Certificate No: 217—97152                            Page 11 of 11


         FCC SAR Test Report                                                Report No. : FA7D2507-01


Appendix F.            Reference Report

Please refer to Sporton report number FA7D2507 and FA7D2507-03 which is issued separately.




Sporton International (Kunshan) Inc.
TEL : +86-512-57900158 / FAX : +86-512-57900958                              Issued Date : Apr. 16, 2018
FCC ID : IHDT56XD7                                Page F1 of F1             Form version. : 170509



Document Created: 2018-04-16 10:50:33
Document Modified: 2018-04-16 10:50:33

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC