Test Report HAC Appendix A

FCC ID: IHDT56XC4

Test Report

Download: PDF
FCCID_3776471

Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/1/13


HAC_E_Dipole_835_180113
DUT: HAC-Dipole 835 MHz

Communication System: CW; Frequency: 835 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
A ie Te per re:23.4 ℃

DASY5 Configuration
- Probe: ER3DV6 - SN2480; ConvF(1, 1, 1); Calibrated: 2017/12/15;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn778; Calibrated: 2017/5/22
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD835 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x361x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 129.1 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 113.4 V/m
Average value of Total=(113.4+112.4) / 2 = 112.9 V/m


                                         PMF scaled E-field
                                          Grid 1 M4 Grid 2 M4 Grid 3 M4
                                          110.4 V/m      113.4 V/m   112.3 V/m
                                          Grid 4 M4      Grid 5 M4   Grid 6 M4
                                          64.29 V/m      65.23 V/m   64.06 V/m
                                          Grid 7 M4      Grid 8 M4   Grid 9 M4
                                          111.6 V/m      112.4 V/m   109.7 V/m
                                                         Cursor:
                                                    Total = 113.4 V/m
                                                     E Category: M4
                                                Location: -0.5, -78, 9.7 mm




                     0 dB = 113.4 V/m = 41.09 dBV/m


Test Laboratory: Sporton International Inc. SAR/HAC Testing Lab        Date: 2018/1/13


HAC_E_Dipole_1880_180113
DUT: HAC Dipole 1880 MHz

Communication System: CW; Frequency: 1880 MHz;Duty Cycle: 1:1
Medium: Air Medium parameters used: σ = 0 S/m, εr = 1; ρ = 0 kg/m3
A ie Te per re:23.4 ℃

DASY5 Configuration
- Probe: ER3DV6 - SN2480; ConvF(1, 1, 1); Calibrated: 2017/12/15;
- Sensor-Surface: (Fix Surface)
- Electronics: DAE4 Sn778; Calibrated: 2017/5/22
- Phantom: HAC Test Arch with AMCC; Type: SD HAC P01 BA;
- Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7373)


E Scan - measurement distance from the probe sensor center to CD1880 = 10mm &
15mm/Hearing Aid Compatibility Test at 15mm distance (41x181x1): Interpolated grid:
dx=0.5000 mm, dy=0.5000 mm
Device Reference Point: 0, 0, -6.3 mm
Reference Value = 145.6 V/m; Power Drift = 0.01 dB
PMR not calibrated. PMF = 1.000 is applied.
E-field emissions = 92.45 V/m
Average value of Total=(91.53+92.45) / 2 = 91.99 V/m

                                         PMF scaled E-field
                                          Grid 1 M3 Grid 2 M3 Grid 3 M3
                                          91.14 V/m 91.53 V/m 88.46 V/m
                                          Grid 4 M3 Grid 5 M3 Grid 6 M3
                                          71.38 V/m 72.54 V/m 71.72 V/m
                                          Grid 7 M3 Grid 8 M3 Grid 9 M3
                                          90.71 V/m 92.45 V/m 90.72 V/m
                                                         Cursor:
                                                    Total = 92.45 V/m
                                                     E Category: M3
                                                 Location: 0, 30.5, 9.7 mm




                     0 dB = 92.45 V/m = 39.32 dBV/m



Document Created: 2018-04-07 21:24:08
Document Modified: 2018-04-07 21:24:08