Test Report RF Exp Part 4

FCC ID: IHDT56WH2

RF Exposure Info

Download: PDF
FCCID_3418723

Schmid & Partner Engineering AG                               s       p        e       a        g

Zeughausstrasse 43, 8004 Zurich, Switzerland
Phone +41 44 245 9700, Fax +41 44 245 9779
info@speag.com, http://www.speag.com




                                        IMPORTANT NOTICE


USAGE OF THE DAE 4

The DAE unit is a delicate, high precision instrument and requires careful treatment by the user. There are no
serviceable parts inside the DAE. Special attention shall be given to the following points:

Battery Exchange: The battery cover of the DAE4 unit is closed using a screw, over tightening the screw may
cause the threads inside the DAE to wear out.

Shipping of the DAE: Before shipping the DAE to SPEAG for calibration, remove the batteries and pack the
DAE in an antistatic bag. This antistatic bag shall then be packed into a larger box or container which protects the
DAE from impacts during transportation. The package shall be marked to indicate that a fragile instrument is
inside.

E—Stop Failures: Touch detection may be malfunctioning due to broken magnets in the E—stop. Rough handling
of the E—stop may lead to damage of these magnets. Touch and collision errors are often caused by dust and dirt
accumulated in the Estop. To prevent E—stop failure, the customer shall always mount the probe to the DAE
carefully and keep the DAE unit in a non—dusty environment if not used for measurements.

Repair: Minor repairs are performed at no extra cost during the annual calibration. However, SPEAG reserves
the right to charge for any repair especially if rough unprofessional handling caused the defect.

DASY Configuration Files: Since the exact values of the DAE input resistances, as measured during the
calibration procedure of a DAE unit, are not used by the DASY software, a nominal value of 200 MOhm is given
in the corresponding configuration file.

Important Note:
Warranty and calibration is void if the DAE unit is disassembled partly or fully by the
Customer.


Important Note:
Never attempt to grease or oil the E—stop assembly. Cleaning and readjusting of the E—
stop assembly is allowed by certified SPEAG personnel only and is part of the annual
calibration procedure.


Important Note:
To prevent damage of the DAE probe connector pins, use great care when installing the
probe to the DAE. Carefully connect the probe with the connector notch oriented in the
mating position. Avoid any rotational movement of the probe body versus the DAE
while turning the locking nut of the connector. The same care shall be used when
disconnecting the probe from the DAE.



Schmid & Partner Engineering


TN_BRO40315AD DAE4.doc                                                                                 11.12.2009


Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                              Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client       Sony Mobile CN (Vitec)                                                              Certificate No: DAE4—1437_Jul16

CALIBRATION CERTIFICATE
 Object                                DAEA4 — SD 000 DO4 BM — SN: 1437


 Calibration procedure(s)              QA CAL—O6.v29
                                       Calibration procedure for the data acquisition electronics (DAE)



 Calibration date:                     July 12, 2016




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                           Scheduled Calibration
 Keithley Multimeter Type 2001          SN: 0810278               09—Sep—15 (No:17153)                                 Sep—16


 Secondary Standards                    ID #                      Check Date (in house)                                Scheduled Check
 Auto DAE Calibration Unit              SE UWS 053 AA 1001        O5—Jan—16 (in house check)                           In house check: Jan—17
 Calibrator Box V2.1                    SE UMS 006 AA 1002        O5—Jan—16 (in house check)                           In house check: Jan—17




                                        Name                                   Function                                 Signature
 Calibrated by:                         Dominique Steffen                      Technician                             %



 Approved by:                           Fin Bomholt                             Deputy Technical Manager               A   ‘/                      .
                                                                                                                        tov   \                    A


                                                                                                                       Issued: July 12, 2016
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: DAE4—1437_Jul16                                       Page 1 of 5


     «       w                                                 \\‘\Hl”l’/,
Callbratlon Laboratory of                                    S\\\\QJ/////’;         Schweizerischer Kalibrierdienst
Schmid & Partner                                            iE\E—/m                 Service suisse d‘étalonnage
   Englnee ring AG                                          T oys                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 é/,fi@                  Swiss Calibration Service
                                                                Alululs®



Accredited by the Swiss Accreditation Service (SAS)                              Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates


Glossary
DAE                          data acquisition electronics
Connector angle              information used in DASY system to align probe sensor X to the robot
                             coordinate system.

Methods Applied and Interpretation of Parameters
    e     DC Voltage Measurement: Calibration Factor assessed for use in DASY system by
         comparison with a calibrated instrument traceable to national standards. The figure given
         corresponds to the full scale range of the voltmeter in the respective range.
    e    Connector angle: The angle of the connector is assessed measuring the angle
         mechanically by a tool inserted. Uncertainty is not required.
    e    The following parameters as documented in the Appendix contain technical information as a
         result from the performance test and require no uncertainty.
         e       DC Voltage Measurement Linearity: Verification of the Linearity at +10% and —10% of
                 the nominal calibration voltage. Influence of offset voltage is included in this
                 measurement.
         e       Common mode sensitivity: Influence of a positive or negative common mode voltage on
                 the differential measurement.
         e       Channel separation: Influence of a voltage on the neighbor channels not subject to an
                 input voltage.
         e       AD Converter Values with inputs shorted: Values on the internal AD converter
                 corresponding to zero input voltage
         e       Input Offset Measurement. Output voltage and statistical results over a large number of
                 zero voltage measurements.
         e       Input Offset Current: Typical value for information; Maximum channel input offset
                 current, not considering the input resistance.
         e       Input resistance: Typical value for information: DAE input resistance at the connector,
                 during internal auto—zeroing and during measurement.
         e       Low Battery Alarm Voltage: Typical value for information. Below this voltage, a battery
                 alarm signal is generated.
         e       Power consumption: Typical value for information. Supply currents in various operating
                 modes.




Certificate No: DAE4—1437_Jul16                                 Page 2 of 5


DC Voltage Measurement
    A/D — Converter Resolution nominal
        High Range:          1LSB =     6.1uV ,       full range =  —100...+300 mV
        Low Range:           1LSB =      61nV ,       full range =   ~1....... +3mV
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec


      Calibration Factors              X                           Y                      Z

      High Range             403.845 + 0.02% (k=2)   403.377 + 0.02% (k=2)   403.789 + 0.02% (k=2)
      Low Range              3.95013 + 1.50% (k=2)   3.93811 + 1.50% (k=2)   3.90441 + 1.50% (k=2)



Connector Angle

      Connector Angle to be used in DASY system                                       64.0 ° +1 °




Certificate No: DAE4—1437_Jul16                      Page 3 of 5


Appendix (Additional assessments outside the scope of SCS0108)
1. DC Voltage Linearity
      High Range                                Reading (uV)          Difference (uV)               Error (%)
      Channel X        + Input                   199994.09                   —2.50                  —0.00
      Channel X        + Input                    20005.91                    4.62                   0.02
      Channel X       — Input                    —20001.55                   —0.46                   0.00
      Channel Y       + Input                    199996.67                    0.10                   0.00
      Channel Y       + Input                    20004.93                     3.69                   0.02
      Channel Y       — Input                    —20002.77                   —1.71                  0.01
      Channel Z       + Input                    199996.72                   —0.44                  —0.00
      Channel Z       + Input                    20003.16                     1.91                  0.01
      Channel Z       — Input                    —20004.52                   —3.47                  0.02


      Low Range                                 Reading (uV)         Difference (uV)                Error (%)
      Channel X       + Input                     2001.13                    0.13                   0.01
      Channel X       + Input                      201.55                    0.06                   0.03
      Channel X       — Input                     —198.43                    0.12                   —0.06
     Channel Y        + Input                     2001.36                    0.44                   0.02
     Channel Y        + Input                     201.04                     —0.44                  —0.22
     Channel Y        — Input                     —199.19                    —0.72                  0.36
     Channel Z        + Input                     2001.25                    0.37                   0.02
     Channel Z        + Input                      199.92                    —1.40                  —0.69
     Channel Z        — Input                     —199.48                    —0.93                  0.47



2. Common mode sensitivity
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                        Common mode                        High Range                            Low Range
                        Input Voltage (mV)             Average Reading (uV)                 Average Reading (uV)

     Channel X                           200                      —3.71                           —5.33
                                        — 200                      6.91                           4.56
     Channel Y                           200                      21.85                          21.92
                                        — 200                     —25.58                         —24.66
     Channel Z                           200                      —10.13                          —9.77
                                        — 200                      7.10                           7.65



3. Channel separation
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
                        Input Voltage (mV)         Channel X (uV)          Channel Y (uV)       Channel Z (uV)

     Channel X                    200                        —                  —0.23                —1.73
     Channel Y                    200                   4.79                         —               1.47
     Channel Z                    200                   6.01                    2.75                      —


Certificate No: DAE4—1437_Jul16                              Page 4 of 5


4. AD—Converter Values with inputs shorted
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

                                                    High Range (LSB)                        Low Range (LSB)

     Channel X                                              16227                                16144

     Channel Y                                              16369                                16590

     Channel Z                                              16351                                16972



5. Input Offset Measurement
    DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec
    Input 10MQ

                            Average (uV)        min. Offset (uV)         max. Offset (uv)        °** E(’:\‘,’;a“m
     Channel X                    —0.45               —~1.47                      1.29                  0.40
     Channel Y                    0.47                —0.65                       1.66                  0.40
     Channel Z                    —2.20               —3.64                   —1.19                     0.41



6. Input Offset Current
     Nominal Input circuitry offset current on all channels: <25fA



7. Input Resistance (Typical values for information)
                                                  Zeroing (kOhm)                           Measuring (MOhm)

     Channel X                                             200                                   200
     Channel Y                                             200                                   200
     Channel Z                                             200                                   200



8. Low Battery Alarm Voltage (Typical values for information)
     Typical values                                Alarm Level (VDC)
     Supply (+ Vec)                                                                +7.9

     Supply (— Vec)                                                                 —7.6



9. Power Consumption (Typical values for information)
     Typical values                         Switched off (mA¥)       Stand by (mA¥)           Transmitting (mA)

     Supply (+ Vec)                                +0.01                     +6                        +14
     Supply (— Vec)                                —0.01                     —8                        —9




Certificate No: DAE4—1437_Jul16                            Page 5 of 5


Calibration Laboratory of                                                               Scinizeiicher Kattvarstont
Schmid & Partner




                                                                                 wo o
                                                                                        Service uie duttomnage
 Engineering AG                                                                         Sorvin ies on
Znvghivasvanse 430004 zarich, Sutriand                                                  uies Cattvaton Servce
Accndtn by t Swas ecrudten Savcn(5A3)                                              Accredtaton : SCS 0108
Te Swias Accrudtaton Servic i on o th igntares ts he EA
Mutistert Agreemontfortherecogniton of calbraton conitcates
ciew      Sporto—KS(Audon)                                               CaniicasNo: EX3—3054_Nov16

|CALIBRATION CERTIFICATE
oves                           Exsove — sNi3o54

Cattraton poceduts)            OACAL—O10, OA CAL:tave,                              »
                               ‘Callbration procedure fordosimatric E—feld probes


Caitraton dn                   November 28, 2016

is caltrate crtcate dcsments the racenbay o natonat tardaisuc eatc t pysca unts t memurements(3)
Th namrenent and e srcnianin w contderce robsbiny e gven on t ftowngpage an e pr o h crtcan
Adcatiaters hav on condetn i th ow on lt eniorment anperaire 2 227C an hanity«70%
CalratonEqeintvan TE erteltocaraton)

 Priman Standarts            o                         GitDate (Gontrai o)                Scomaded Catteaton
 rov nawe n                 Entun                      018 t 2w0z               weer
 Pove sese ts#251           sn wn                      oeterts m 2irz00         heerr
 Fome seme 1.251            sns                        0ct 2rrazam              tecr
 HRetwence 20 0b Aloeatc    sn Sser‘ oo                oit 217ccom              hectr
 Retwerce Pobe Essove       sn aots                    Eremmremconemben
 mee                        si ow                      Zpre1s in Dibito Dects) Deeis
 SmcondaySardugs            u:                         Chexbas nrong            soetiet en
 Pomenow tattt0             sn couman                  ho2 n rowmechook istD Intomecieck anB
 Foverseme Eiti2h           sic mvaremonr              oApcil n howeched Jn 10— inromectect nis
 Poverseme Eotizn           sn oomsen                  otAnc18 inhome ces hn19— ntomeciec: hn
 it qonpaowes               s ussotzuorni0             240intowe dad in—| Intomecte ani8
 netvors Araie ve esE       sn ussrasous               180001 in owctex 0s15) Intowse ceac Oc17
                               Rure                        Fade                   Span
Caitaie by                     LotiOpnec                      LatentoyTachicon           ‘ '%,


Aosomty                        wilrolove                      Tibiampee                             fé&

                                                                                           hnNoventer 22016
Th cattraten cutteae stt se h merroducnd merst in t uthut w aporn o t iatortoy


Corttcate No: Ex3—305¢ Novio                         Page t ot 11


      4
Callbration Laboratory of                                                        .: Seeicnschor Katierdinst
Schmid & Partner                                                                 . Servie sisse titsonnage
  Engineering AG                                                                 & Seride ies at amre
irvghausstnase 43. 1004 2ic, Sntentand                                               SuiseCatiraton Serice

Acendta by t un AcrudatenServen(A8)                                               Acedtatontis: SCS 0108
The Swiss Accradtaton Sarice i on oft nigntorestohe A
MitimertAgreament fore recognton of cabratonconicates
Glossary:
Ts                     tssue simulatng lould
NoRMxya                sensitviy in ree space
Comt                   sensibvty in TSL / NORMyz
ocp                    iode compression point
or                     erest factor(tiduty_cle)ofthe RF signal
A8.c.0                 moxiiation dependantncarzation parameters
Polarzation o          « otation around probe s
Polariaton 0           8 rotation around an anis that is n the plane normal t probe ax‘s(at measuremont conter,
                       t, 0 =0 is normalto probe axis
Comnector Angle        information used in DASY systom to algn probe sensor X to the robot coortinate system
Calibration is Performed According to the Following Standards:
   a).IEEE Sid 15282013, 1EEE Recommended Practicefor Deternining the Poak Spatla—Averaged Specite
       Absorption Rate (GARin the Human Head from Wireless Communicatons Devices: Measuremont
       Technques®,Jine 2013
   6). 120 62200—1, *Procedure to measure the Speciic Absorption Rate (SAR) forhand—held devices used n cose
       prowimitt the ea(Requency range o 300 Mz to 3 GHzF, February 2005
   e). 120 62200—2, "Procedureto determine the Specifc Absorplon Rate (GAR) or wreless communication devices
       used in close proximit to the human body (requency range o 30 MHe to 6 Giic), March 2010
   «) KoB 656. SAR Measuremont Requrementsfor 100 Mz t 6 GHtz"
Mofllodl Applied and Interpretation of Parameters:
        NORMrqca: Assessed forE—fld polaizaion & =0 t 900 MEtz in TEM—cel;1> 1900 Mite:R22 wavequide}.
        NORM3.z are only intermediate valus, . the uncertainies of NORK»y.z daes not affec the E—fld
        unceraint nside TS (see below Con?)
   *    NORMfMry.z = NORkixy:z *requency.response (see Frequency Response Chart) This ncarization is
        implamented in DASY4 sotware versionslatethan 4.2.The uncertaity f the requency response is included
        in the stated uncertainty of GomvF:
   +    0GPxy.z: DCP are numerical insarization parameters assessed based on the data of power seep with CW
        signal(no uncertaintyrequired). DCP does notdepend on frequency nor media.
   +    PAFR: PAR is the Reak to Average Ratlo thais not calbrated but determined based on the signal
        chracterstics
   +.   AujieBycz Ciez: Diy.z: Viiey.z:A, . G, D are numerical inearization parameters assossed based on
        the data of power swaes for speciic modulrton signal, The parameters do not depend on frequency nor
        media. VR s the maximum calbration range expressed in RMS vatage across the dode
   *    Gorveand Boundary Elfect Paramaters: Assessed in flat phantom using E—eld (or Temporature Transter
        Standard fof 800 MiHz) and inside waveguide using analylalfld distibutions based on power
        measuremantsfo 1 > 800 MHz. The same setups are used for assessment of the parameters appled for
        boundary compensation (alpha, depth)of which typical uncertaintyvalues are given, These parameters aro
        used in DASY4 sottware toimprove probe accuracy close to the boundary. The senstily in TSL coresponds
        to NORxy.z * Can: whereby the uncertainty corresponds t tat aven for Cony. A freauency denendont
        Conis used in DASY version 4.4 and higher which alows xtending th valdty om 2 50 MHe to 2 100
        lb
   *    Sphriea!isotray (3D deviaton rom isotrapy) in fld of ow graclents realized usinga t phantom
        expased by a patch antenna.
   *    Sensor Offet:The sensor oifset coresponds tothe ffsatof virval measurement entr m the probe to
        (on probe axis)No tolerance required
   *    Comnetor Angl: The angle is assessed using the information gained by deternining the NORA (ro
        uncertainty requred)

Gortieat No: Ex01084 Nort6                        Page 20¢ 11


wxaove— snsost                                                             Novembor 2n, 2018




                   Probe EX3DV4

                                      SN:3954

                              Manufactured:           August 6, 2013
                              Repaired:               November 21, 2016
                              Calibrated:             November 28, 2016



                             Calibrated for DASY/EASY Systems
                                (Note: non—compattle with DASYz system!)




Certfcats No: ©XG—3084 Noi                    Page 0t t1


waove~ swsoss                                                                      Novembor 282010


DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954
Basic Calibration Parameters
                                 SensorX            Sensory              Sensorz         Wnepe®
 No @Vivim?)"                      ost                  oas               ose            si0r%
 ©GF (my)                          w5                   mork              sro
Modulation Calibration Parameters
 uo         Communicaion Systom Name           x            a     c       a        w       we‘
                                               «o         enviv           as       mv      (en
6           ow                             x|_oo         |_oo     1o_|   600       mao    »5%
                                           v|_os           oo     10               nos
                                           2oo             oo     i0               mas

 The reported uncertainty of measurement is stated as the standard uncertainty of measurement
 multipled by the coverage factor k=2, which for a normal distribution corresponds to a coverage
 probabilty of approximately 95%.

" The urosabtos o Norm XY 2o e atet h E‘dad unc eide T9L o Pagen ad 6.
©Phrmaeat iesansaton paanater incetanty et muend
* Unceanty adterinad usng t mae. devaton um Wrc esnseaspng en duitutonant n mssn o i scare o ha
tuitvaie




Cortfcats No: Ex—084 Novtd                  Page 4 11


exsove— smause                                                                      Novermbor20, 2016



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Callbration Parameter Determined in Head Tissue Simulating Media
  uy           P-mut:'hy’    cw(“nmw      ComEX     ComeY     |_ComtZ   Aiha     D(.:r@"     (‘u’-“:;
      150         a10           ose        1oss      tose      tose      osr      os2      +120%
      sss         415           aso        1ose      ross      ioss      ome      142      a120%
      sco         «15           oar        1035      toss      toss      oss      oss      +120%
     17so         «01           137        ase       sse       ass       om       ost      s120%
     1900         «00           140        ss2       ase       ase       oar      i2t      s120%
     2000         «0            140        sas       ses       ez        om       im       et20%
     2300         sos           107        res       zes       res       or       115      «120%
     2450         so2           180        zus       ras       ras       oso      om       £120%
     200          o             198        mm        rar       ror       |oar     is       aro%
     asoo         sro           201        m0        zio       zio       om       120      en
     soso         sso           an         sos       _|_sos    sos       oss      iso      arn1% |
     seoo         sss           sor        «70       azo       aro       oso      180      eisi%
     srso         asa           s2e        aso       aso       iss       os       im       ani%

©Irequeney vakayahor 0 c t 100 Mb on apute o DA vet and grer sePage 2)ols s etici t# 0t The
vecetant n h S t heCm unceiany t calealonfecuecyand ho uncntant t in nicate haqurcy bant reavercyty
befow 300 ht n a 10254090 and 70e o Cordansesnant a 3. M,120, 10and 220 it respeciniy ow 9 lt ropuercy
td can b edandeits s 110t
‘Athocuencs tlow3Gith valdy f ie ooo co) cab alne o 10% 1 ui conponatan m s ovote o
rnoannd SA vtam.At uces abov G.th vniy o ue paranetars(cand on ented t# 9. Thuncetany te RSS of
theGori urcetant to intcin getoowpwnates
*Aorwoedth mdtemiead ving caaton 5P€AG waran iat e remaringdevaton unto hboondry lc ate compersalon is
aayton
d          an 11 o Heauerciotw 3 C ad blow 21 orbaguercs uoo 20 GHeatany marc age hn hie pate t




Centicai No: Ex—s084Norte                     Page S1


exsove~ snaose                                                                         Novembor 20, 2016



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Calibration Parameter Determined in Body Tissue Simulating Media

    «ue        p."..:'.m,r    cw’           ComX      ComEY      |_Gomvtz   Alpha®   m.         (!-";J
      150          sss           ass         1ose      tose    tose          o1       om      120%
      ss           see           osr         1032      tos2    tos           om       im      r120%
     17so          sea           149         ase       ass     es            om       om      «120%
     1000          sas           is2         ao1       aon     sor           oso      o       »120%
     2300          seo           n1          reo       _zso    rso           oas      om      «120%
     2450          ser           195         mss       tss     rss           oar      om      «120%
     2000          sas           216         z0s        _|_zos  zos          o«       ost     r120%
     ssoo          sis           as          ers       srs     ers           om       iz      ai21%
     soso          «so           s3e         aso       aso |aso              oas      180     ers1%
     seoo          «s            sir         ase       ase     ase           oso      180     a121%
     srso          «s3           soe         aos       sos     aos           oss      is      et21%

5 equency uky aheve 200 t 100 t on aptes o DASY t and hghr se Page 21 i t i reivctoto + 0M The
«ncetant n o ASG o the Con? ncerany acaliatonbecuncy andtheunconnty t t indatat hocurcyband Frpueroy altty
ow 200 ht a 16254050 and 7 te o C annesumenis t 20. 6,128, 10 ant 20 Ne eseaivty Abowe 8 it rapurcy
ywdycan t denan i 110ht
* t hecuenc blaw3 6t ne vahdyof ie arne ( o ) con t ow t 109 i compernaton tm in apste t
«remrnSA vaum. 2t becurcim acov C in ind o taou arel i n o) enc t 1. Th ncwiany t SS of
toeGooi ncntanty forindcin ge t ovanuts
* AprDoothar detenined suing caitaten SPEAG vrartsat h renonngdevaton ie t h bounda ols ater cepruato is
atraye leaitea19 t Reauerce i 3 Gie and biow a 2s o hnguencen etron 20 GHeat onitarcn y on hatito mobe to
      mite tom  bocnday




Certfcate No EXG—054Norte                       Page Satt1


exaove~suaost                                                                                     Novemtor28, 2016

                                          Frequency Response of E—Field
        Frequency response (normalzed)        (TEM—Collif110 EXX, Waveguide: R22)




                                         Uncertainty of Frequency Response of Eld: 2 6.3% (es2)




Cerifcate No: ©G—3054 Novte                                   Page 7 of 11


exaove~ snaose                                                                          November 2n, 2018



                               Receiving Pattern (¢), 3 = 0°

                   1°600 MHz,TEM                                      (=1800 MHz,R22




                                                                          wel /




                                                                                           <ia
                                               ®




                                                                                  <4
                                                           !
    Enc io




                 alo            Eo
                                                        noiri
                   w                       «ht                   wifihe                wl
                             Uncertainty of Axial Isotropy Ausossmont: + 0.3% (ke2)




Certteate No: Exs—3954ovt6                         Page ot 11


Cl                                                                               Novnbor28. 208



                               Dynamic Range f(SARncaq)
                                       (TEM coll , fo.r= 1900 MHz)




                              Uncertainty of Linearity Assossmont: £0.6% (ke2)



Corticate No: Ex32054.Novt6                     Page tot 11


exciove— suaose                                                                    Noverbor20, 2016

                          Conversion Factor Assessment
               1 = aas hMzWGLS R (.comt)                     1= 1900 MitcWGLS Rz2 _cont)




                              uP




                                                                        i
                         Deviation from Isotropy in Liquid
                                      Error (6, 8), £=900 MHz




                      «0 sos os o« mz oo o2 o4 os o8 10
                       Uncertainty of Spherica Isotropy Assossmonts # 2.6% (ke2)


Corticate No: Ex3.205¢.ovt6                  Page 10 of 11


wxaove— saose                                               Novermtor 20,2016



DASY/EASY — Parameters of Probe: EX3DV4 — SN:3954

Other Probe Parameters
 Sensor Arangement
     nector Angje ()
Mechanica Surface Detecton Mode
 OplcalSurlace Detection Mode
 Probe OveraLengh
Probe Body Diamoter
 To Lngh
 To Damefer
 Frobe Tisto Sensor X Galbraton Poin
 Probe Tisto Sensor V Gallbration Point
 Probe Tipto Sensor Z Calbraon Point
 Recommended Measurement Distance from Surface                      14 mm




Certtcato No: 084Novld                      Page 11 of 11


Calibration Laboratory of
Schmid & Partner                                                                                            S SChWeizenscher Kaiib」 erdienst
                                                                                                             C SerViCe suisse d'6tabnnage
                                                                                                            S ttξ :』 橘11:職 m
 Engineering AG
Zeughausstrasse 43, 8004 Zurich, Swikerland


Accrediied by the Swiss Accreditation Service (SAS)                                                         AccrednaJon No.:SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

CIient                                                                                          cerrificate No:    EX3-3857_May1 6




Thts calibration certificate documents the traceability to national standards, which realize the physical units of measurements (Sl).
The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


All calibrations have been conducted in the closed Iaboratory facility: envrronment temperature (22      !3)'C    and humidity < 70%.


Calibration Equipment used (M&TE critical for calibration)



 Primary Standards                   ID                                 Cal Date (Certificate No.)                    Scheduled Ca ibration
 Power meter NRP                     SN:104778                          06‑Ap「 16(No 217‑02288/02289)                 Ap「 17
 Power sensor NRP-291                SN:103244                          06‑Ap● 16(No 217‑02288)                       Apr-17
 Power sensor NRP-291                SN:103245                          06‑Ap「 16(No 217‑02289)                       Ao卜 17
 Reference 20 dB Attenuator          SNI S5277(20x)                     05‑Ap「 16(No 217‑02293)                       Ap卜 17
 Reference Probe ES3D∨ 2             SN:3013                            31-Dec-15 (No. ES3-3013 Dec15)                Dec‑16
 DAE4                                SN:660                             23-Dec-15 (No. DAE4-660 Dec15)                Dec‑16


 Secondary Standards                 ID                                 Check Date (in house)                         Scheduled Check
 Power meter 844198                  SN:GB41293874                      06‑Apr‑16(No 217‑02285/02284)                 ln house check: Jun-16
 Power sensor E44124                 SN:MY41498087                      06 Ap卜 16(No 217‑02285)                        n house check: Jun-16
 Power sensor E4412A                 SN:000110210                       06 Ap「 16(No 217‑02284)                        n house cheDk: Jun-16
 RF generator HP 8648C               SN:US3642U01700                    04‑Au9‑99(in houSe check Apr‑13)               n house check: Jun-16
 Network Analvzer HP 8753E           SN:US37390585                      18‑Oct‑01(ln house check Oct‑15)               n house check: Oct-16

                                          Name                                 Function
Calibrated by:                         Jeton Kastrati                          ta,9,tory
                                                           餡
Approved by:                              Katja Pokovic                        Technical Manager



                                                                                                                        lssued: May 31,2016
This calibration certificate shall not be reproduced except in full without written approval of the laboratory.




Certificate No: EX3-3857-Mayl 6                                       Page 1 of   11


Calibration Laboratory of




                                                                                           S C S
                                                                                                   Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                   Service suisse d'6talonnage
 Engineering AG                                                                                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland
                                                                                                   Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                        Accreditation     tto.: SGS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
NORMx,y,z                    sensitivity in free space
ConvF                        sensitivity in TSL / NORMx,y,z
DCP                          diode compression point
CF                           crest factor (1/duty_cycle) of the RF signal
A,B,C,D                      modulation dependent linearization parameters
Polarization rp              rp rotation around probe axis
Polarization $               $ rotation around an axis that is in the plane normal to probe axis (at measurement center),
                             i.e., $ = 0 is normal to probe axis
Connector Angle              information used in DASY system to align probe sensor X to the robot coordinate system
Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528-2013,       "1EEE Recommended Practice for Determining the Peak Spatial-Averaged Specific
         Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement
         Techniques", June 2013
    b)   IEC 62209-1, "Procedure to measure the Specific Absorption Rate (SAR) for hand-held devices used in close
         proximity to the ear (frequency range of 300 MHz to 3 GHz)", February 2005
    c)   IEC 62209-2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless communication devices
         used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)", March 2010
    d)   KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Methods Applied and lnterpretation of Parameters:
    e    NORMx,y,z; Assessed for E-field polarization $ = 0 (f < 900 MHz in TEM-cell; f > 1800 MHz R22 waveguide).
         NORMx,y,z are only intermediate values, i.e., the uncertainties of NORMx,y,z does not affect the E2-field
         uncertainty inside TSL (see below ConvF).
    .     NORM(|X,y,Z = NORMx,y,z * frequency-response (see Frequency Response Chart). This linearization is
         implemented in DASY4 software versions later than 4.2. The uncertainty of the frequency response is included
         in the stated uncertainty of ConvF.
    .    DCPx,y,z: DCP are numerical linearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
    t     PAR: PAR is the Peak to Average Ratio that is not calibrated but determined based on the signal
         characteristics
    .    Ax,y,z; Bx,y,z; Cx,y,z; Dx,y,z; VRx,y,z: A, B, C, D are numerical linearization parameters assessed based on
         the data of power sweep for specific modulation signal. The parameters do not depend on frequency nor
         media. VR is the maximum calibration range expressed in RMS voltage across the diode.
    .    ConvF and Boundary Effect ParameteSs:Assessed in flat phantom using E-field (or Temperature Transfer
         Standard for f < 800 MHz) and inside waveguide using analytical field distributions based on power
         measurements for f > 800 MHz. The same setups are used for assessment of the parameters applied for
         boundary compensation (alpha, depth) of which typical uncertainty values are given. These parameters are
         used in DASY4 software to improve probe accuracy close to the boundary. The sensitivity in TSL corresponds
                         *
         to. NORMx,y,z ConvF whereby the uncertainty corresponds to that given for ConvF. A frequency dependent
         ConvF is used in DASY verslon 4.4 and higher which allows extending the validity from t 50 MHz to 100          t
         MHz.
    o    Spherical isotropy (3D deviation from isotropy): in a field of low gradients realized using a flat phantom
         exposed by a patch antenna.
    .    Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.
    .    Connector Angle: The angle is assessed using the information gained by determining the NORMx (no
         uncertainty required).


Certificate No: EX3-3857-May'l 6                                Page 2 of   11


EX3D∨ 4‑SN:3857
                                                                                 May 25,2016




                       Probe EX3DV4
                                           SN:3857

                                   Manufactured: January 23, 2012
                                   Calibrated:              May 25,2016



                                  Calibrated for DASY/EASY Systems
                                     (Note: non-compatible with DASY2 system!)




Certificate No: EX3-3857_Mayl 6                    Page3of11


EX3D∨ 4‑SN:3857                                                                                                                   May 25,2016



DASYノ EASY‐ Parameters of Probe:EX3DV4‐ SN:3857
Basic Calibration Parameters
                                                Sensor X                         Sensor Y                    Sensor Z              Unc (k=2)
    Norm (uVi(Vim)2)A                             0.18                            0.44                         0.47                ±10.1%
    DCP (mV)B                                     97.6                            97.5                          983

Modulation Calibration Parameters




                                                                                                                            V m
                                                                                                                            R V
    U:D




                                                                                                                 D
                Communication System Name                                   A            B           C                                 Unc'




                                                                                                                d
                                                                                                                B
                                                                           dB        dBr/uV                                            (k=2)
    0          CW                                                 X        00            00                    000        1356        ±30%
                                                                  Y        00            00         10                    1287
                                                                  Z        0.0           00         10                    1286


    The.reported uncertainty of_measurement is stated as the standard uncertainty of measurement
    multip.lie..d by the coverage. tqtql k=2, which for a normal distribution correspohds to a coverage
    probability of approximately 95%.

A
  The uncertainties of Norm x,Y,Z do not affect the E2-field uncertainty inside TSL (see Pages 5 and 6).
B
   Numerical linearization parameter: uncertainty not required.
' Uncertainty is determined using the max. deviatioh from linear response applying rectangular distribution and is expressed for the square of the
field value.




Certificate No: EX3-3857_Mayl 6                                    Page4of11


EX3DV4‑SN:3857                                                                                                                 ヽ4ay 25,2016




DASYノ EASY‐ Parameters of Probel EX3DV4‐ SN:3857

Galibration Parameter Determined in Head Tissue Simulating Media
                                         ∞ γ




                                                                                                                                              n刊
                                                                                                                                              c
                                          器 町




                                                                                                                                        U k
                          Relative                                                                                  Depthl
                c       permittivitv F
      f (MHz)                                               ConvF X       ConvF Y        ConvF Z       Aloha    G
                                                                                                                     (mm)
        750                 41.9              089             968            968             9.68        037         080          ± 120%

        835                 415               090             932            932             932         0.44        082          ± 120%

        900                 41.5              097             9.08           9.08            908         043         080          ± 120%

        1750                40.1              137             800            800             8.00        033         0.80         ± 120%

        1900                400               1.40            7.85           7.85            785         032         080          ± 120%

       2000                 400               140             783            783             783         0.27        080          ± 12.0%

       2300                 39.5              1.67            744            744             7.44       0.19         1.03         ± 12.0%

       2450                 39.2              180             7.19           719             719        036          080          ± 120%

       2600                 390               196             7.08           7.08            708        030          1.20        ± 12.0%

       3500                 379              291              6.87           687             687        030          1.30        ± 13.1%

       3700                 37.7             312              642            642             6.42       025          1.30        ± 131%

       5250                 35,9             471              515            515             5.15       030          180         ± 131%

                                                              470                            470        040


                                                                                                                                  +一
       5600                 35.5             5.07                           4.70                                     180               131 %
       5750                 354              5.22             500            5.00            500        040          180         ± 131%


t Frequency validity above 300 MHz of t 100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to + 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the.uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10,25,40,50 and 70 MHzfor ConvF assessments at 30, 64,128,150 and 220MHz respectively. Above 5 GHzfrequency
validity can be extended to t 1 10 MHz.
F
                                                                                         !
  At frequencies below 3 GHz, the validity of tissue parameters (a and o) can be relaxed lo 10% if liquid compensation formula is applied to
measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and a) is restricted to t 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
'Alpha/Depth are determined during calibration. SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
                    t
always less than 1% for frequencies below 3 GHz and below + 27o for frequencies between 3-6 GHz at any distance largeithan half the probe tip
diameter from the boundary.




Certificate No: EX3-3857_Mayl 6                                  Page 5 ofll


EX3D∨ 4‑SN:3857                                                                                                                May 25,2016




DASYノ EASY‐ Parameters of Probe:EX3DV4‐ SN:3857

Calibration Parameter Determined in Body Tissue simulating Media
                                          ∞ γ
                                           器 町
                          Relative                                                                                  Depthl             Unc
                c
      f (MHz)           Pernlittivitv F                     ConvF X       ConvF Y        ConvF Z       Aioha G       (mmヽ              (k=2)

        750                  555              096             945            945           945           050         080          ± 120%

        835                  552              097             9.25           925           925           0.34        099          ± 12.0%

        1750                 534              149             781            781                         036         080




                                                                                                                                  +一
                                                                                           781                                         12.0%
        1900                53.3              152             755            755           7.55          035         080          ± 120%

       2300                 52.9              1.81            7.31           7.31          7.31          038         0.87         ± 12.0%

       2450                 527               195             7.23           723           723          0.33         0.80         ± 12.0%

       2600                 525               216             7.13           7.13          7.13         0.28         080          ± 12.0%

       3500                 51.3              331             641            641           641          028          140         ±13.1%

       3700                 510               3.55            6.26           626           626          030          1.40        ±13.1%

       5250                 489              536              427           4.27           4.27         050          190         ± 131%

       5600                 48.5             5.77             376            3.76          3.76         055          1.90        ±13.1%

       5750                 483              594              397            397           397          060          190         ± 131%



" Frequency validity above 300 MHz of t   100 MHz only applies for DASY v4.4 and higher (see Page 2), else it is restricted to t 50 MHz. The
uncertainty is the RSS of the ConvF uncertainty at calibration frequency and the uncertainty for the indicated frequency band. Frequency validity
below 300 MHz is + 10,25,40,50 and 70 MHzfor ConvF assessments at 30, 64,128,150 and 220MHz respectively. Above 5 GHzfrequency
validity can be extended to t 110 MHz.
  At frequencies below 3 GHz, the validity of tissue parameters (s and o) can be relaxed to t 10% if liquid compensation formula is applied to
F

measured SAR values. At frequencies above 3 GHz, the validity of tissue parameters (e and a) is restricted to t 5%. The uncertainty is the RSS of
the ConvF uncertainty for indicated target tissue parameters.
" Alpha/Depth are determined during calibration" SPEAG warrants that the remaining deviation due to the boundary effect after compensation is
                    t
always less than 17o for frequencies below 3 GHz and below t 2% for frequencies between 3-6 GHz at any distance larger than half the probe tip
diameter from the boundary.




Certificate No: EX3-3857_Mayl 6                                   Page6of11


EX3D∨ 4‑SN:3857
                                                                                                                              May 25,2016

                                                       Frequency RespOnse of E‐ Field
                                                              (TEM¨ Ce:::ifil1 0 EXX,Waveguide:R22)




                                             15

                                             14
              ︵で①N識鮨属




                                             1




                                             12
            O C︶ 00 颯O Q00 ﹂ 一O C O コ σO﹂﹂




                                             11

                                             10

                                             09

                                             08

                                             07

                                             0

                                             0.51         │                        │          !
                                                  0
                                                                         I                                     │




                                                        500                      1500         2000
                                                                                                                         1




                                                                       1000                                  2500      3000
                                                                                                           回R

                                                                                  fIMHZl
                                                                                                            2
                                                                                                            2


                                                                」
                                                                TttM


                                                      Uncertainty of Frequency Response of E-field:   t   6.3% (k=2)




Certificate No: EX3-3857_May1 6                                               PageT of   11


EX3D∨ 4‑SN:3857
                                                                                                                              May 25,2016




                                     Receiving Pattern(φ ),s=0°

                       卜600 MHz,丁 匡M                                                         ← 1800      MHz,R22




                                                                                       111                              11,




                                                                                                                                一T
                                                                                                              一
                            ・




                                                                                                              0一
                            一X




                                         ■                燃                                          ■




                                                                                                                Y
                                         Y                Z                     TOt              x




                 05
        ご己 ﹄
           2﹂




                                         `       ´
                                             ■
             Ш




                        ―                                                                    ー
                 ‑05
                            一
                            劃咄




                                                                   Ro‖   [・ ]


                                                      回
                                                     600 MHz
                                                                                        」
                                                                                      1800 MHz                 2・J:ざ   鷺
                                                                                                                       JHz


                                  uncertainty of Axialisotropy Assessment:±                      0.5% (k=2)




Certificate No: EX3-3857_Mayl 6                                Page8of11


EX3D∨ 4‑SN:3857                                                                                      May 25,2016




                                        Dynamic Range f(SARhead)
                                                      (TEM G€l! , fsy2l= 1900 MHz)

                                   0
                    一
                    >コ﹈焉 cpの 一
                             っQ三

                                   0




                                           10-r              10-1
                                                                    SAR [mWcm3]
                                                      __J
                                           not compensated
                                                                                         E
                                                                                    compensated




                                            lo‐   2         1o‐ 1       100
                                                                    SAR [mWlcm3]

                                           not
                                                      iat
                                                 compensated
                                                                                        E
                                                                                   compensated

                                       Uncertainty of Linearity Assessment:         t   0.6% (k=2)




Certificate No: EX3-3857_Mayl 6                                     Page9of11


EX3DV4‑SN:3857                                                                                                                                            Nlay 25,2016


                                   Conversion Factor Assessment




                                                                                                     √キ■ 一
                 f=835 MHz,VVOLS R9(H conVF)                                                                  f=1900 MHz,V√ GLS R22(H̲conVF)




                                                                                                                一
                \




                                                                                                        ¨
                                                                                           ● ´一● ざ
       3251




                                                                                                         ︲
       ≧2ol




                                                                                      ︐︐



                                                                                                         ・
                                                                                                         o
                                                                                                         ︲       ︑
                                                                                           一γ




                                                                                                         ト
       誘 :




                                                                                           ′
         ,5'


          10,
                                                                                                                                \    `
                                                      :)│..、 ヽ̲̲1                                                                    \ 、
          05■                                                                                                                           、
                                                                                                                                         、      ̲̲.̲̲̲1
          001ニ ユ 上二̲̲二
                                                                                │



                                                L=二   L.二              二
                                                              ̲̀饉 T上
                                                         30   "             =                                                               為     30謳          40
                                       ZImml

                         a轟 載                                                                                            コ                        1コ
                                                     爛                                                               神       1) al              :"aswed


                                Deviation from lsotropy in Liquid
                                                         Error(φ ,3),f〓                              900 MHz

                             10
                             08
                             06
                           c 04
                           0
                           罵 02
                           3
                           0‐
                             00
                                02
                                04
                                …
                                ‑06
                                08
                                ‐
                                ‑10
                                   0




                            ‑1.0       ‑0.8    ‑06       ‑0.4      ‑02              0.0                  02   0.4        0.6         03   1.0
                                uncertainty of Sphericalisotropy Assessment:± 2.6% (k=2)




Certificate No:EX3‐ 3857̲Nlay1 6                                           Page 10 of                1   1


EX3DV4‑SN:3857                                            ヽ4ay 25,2016




DASYノ EASY‐ Parameters of Probe:EX3DV4‐ SN:3857

0ther Probe Parameters
 Sensor Arrangement                                       Triangular
 Connector Angle (')                                             140
 Mechanical Surface Detection Mode                          enabled
 Optical Surface Detection Mode                             disabled
 Probe Overall Length                                       337 mm
 Probe Body Diameter                                          10 mm
 Tip Length                                                   9mm
 Tip Diameter                                                25mm
 Probe Tip to Sensor X Calibration Point                       l mm
 Probe Tip to Sensor Y Calibration Point                       lmm
 Probe Tip to Sensor Z Calibration Point                       lmm
 Recommended Measurement Distance from Surface               1.4mm




Certincate No:EX3‐   3857̲May16              Page11of11



Document Created: 2017-06-01 16:15:15
Document Modified: 2017-06-01 16:15:15

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC