Calibration Laboratory of                                                                                          Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                   Service suisse d‘étalonnage
  Engineering AG                                                                                                   Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                       Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                              Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multitateral Agreementfor the recognition of calibration certificates

Client       CCS—TW (Auden)                                                                      Certificate No: H3—6163_Apr10

CALIBRATION CERTIFICATE
 Object                                H3DV6 — SN:6163


 Calibration procedure(s)              QA CAL—03.v5 and QA CAL—25.v2
                                       Calibration procedure for H—field probes optimized for close near field
                                       evaluations in air


 Calibration date:                     April 26, 2010




 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 & 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                       Cal Date (Certificate No.)                           Scheduled Calibration
 Power meter E4419B                     GB41293874                 1—Apr—10 (No. 217—01136)                            Apr—11
 Power sensor E4412A                    MY41495277                 1—Apr—10 (No. 217—01136)                             Apr—11
 Power sensor E4412A                    MY41498087                 1—Apr—10 (No. 217—01136)                             Apr—11
 Reference 3 dB Aftenuator              SN: 55054 (3c)             30—Mar—10 (No. 217—01159)                            Mar—11
 Reference 20 dB Attenuator             SN: 55086 (20b)            30—Mar—10 (No. 217—01161)                            Mar—11
 Reference 30 dB Attenuator             SN: $5129 (30b)            30—Mar—10 (No. 217—01160)                            Mar—11
 Reference Probe H3DV6                  SN: 6182                   3—0ct—09 (No. H3—6182_0ct09)                         Oct—10
 DAE4                                   SN: 789                    23—Dec—09 (No. DAE4—789_Dec09)                       Dec—10


 Secondary Standards                     ID #                      Check Date (in house)                                Scheduled Check
 RF generator HP 8648C                   US3642U01700              4—Aug—99 (in house check Oct—09)                     In house check: Oct—11
 Network Analyzer HP 8753E              US37390585                 18—0ct—01 (in house cht_ack Oct—09)                  In house check: Oct10


                                        Name                                    Function                                Signature
 Calibrated by:                         Jeton Kastrati                          Laboratory Technician



 Approved by:                           Katia Pokovie                           Technical Manager             /@%_\



                                                                                                                        Issued: April 27, 2010
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: H3—6163_Apr10                                         Page 1 of 10


Calibration Laboratory of                                     «il77                          Schweizerischer Kalibrierdienst
Schmid & Partner                                            ilfié                             Service suisse d‘étalonnage
   Engineering AG                                           Ts                               Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                  ns                             Swiss Calibration Service



Accredited by the Swiss Accreditation Service (SAS)                                       Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates



Glossary:
NORMx,y,z                    sensitivity in free space
DCP                          diode compression point
CF                           crest factor (1/duty_cycle) of the RF signal
A, B,C                       modulation dependentlinearization parameters
Polarization q               «p rotation around probe axis
Polarization 8               8 rotation around an axis thatis in the plane normal to probe axis (at measurement center),
                             .e., 8 = 0 is normal to probe axis
Connector Angle              information used in DASY system to align probe sensor X to the robot coordinate system

Calibration is Performed According to the Following Standards:
    a)   IEEE Std 1309—2005, " IEEE Standard for calibration of electromagnetic field sensors and probes, excluding
         antennas, from 9 kHz to 40 GHz", December 2005.

Methods Applied and Interpretation of Parameters:
         NORMx,y,z: Assessed for E—field polarization 8 = 0 for XY sensors and 8 = 90 for Z sensor(f < 900 MHz in
         TEM—cell; f> 1800 MHz: R22 waveguide).
         X,Y,2(0)_adata2= X,Y,Z2_a0ata2* frequency_response (see Frequency Response Chart).
         DCPx,y,z; DCP are numerical linearization parameters assessed based on the data of power sweep with CW
         signal (no uncertainty required). DCP does not depend on frequency nor media.
         Ax,y,2; Bx,y,z; Cx,y,z, VRx,y,z: A, B, C are numerical linearization parameters assessed based on the data of
         power sweep for specific modulation signal. The parameters do not depend on frequency nor media. VR is the
         maximum calibration range expressed in RMS voltage across the diode.
         Spherical isotropy (3D deviation from isotropy}: in a locally homogeneous field realized using an open
         waveguide setup.
         Sensor Offset: The sensor offset corresponds to the offset of virtual measurement center from the probe tip
         (on probe axis). No tolerance required.
         Connector Angle: The angle is assessed using the information gained by determining the X_aOa1a2 (no
         uncertainty required).




Certificate No: H3—6163_Apr10                                   Page 2 of 10


H3DV6 SN:6163                                                                      April 26, 2010




                                  Probe H3DVC6

                                             SN:6163
                             Manufactured:                       January 1, 2005
                             Last calibrated:                    April 27, 2009
                             Recalibrated:                       April 26, 2010

                                      Calibrated for DASY Systems
                                       (Note: non—compatible with DASY2 system!)




  Certificate No: H3—6163_Apr10                  Page 3 of 10


 H3DV6 SN:6163                                                                                                                                 April 26, 2010


 DASY — Parameters of Probe: H3DVG6 SN:6163


 Basic Calibration Parameters
                                                                                               Sensor X]      Sensor Y        Sensor 2 |Unc (k=2)

Norm (A/m / \/(pV))                                                                      a0| 2.65E—3|         2.60E—3         3.14E—3         £5.1%
Norm (A/m / \/(pV))                                                                      all   1.04E—5|       3.78E—5         —1.32E4|]       £5.1%
Norm (A/m / \/(pV))                                                                      a2| —4.99E—5|        —6.56E—5        1.50E—5         £5.1%
DCP (my)*                                                                                         82.3           95.5           83.3

Modulation Calibration Parameters
UID             Communication System Name                               PAR                        A               B              C            VR           Unc®
                                                                                                   dB           dBuV                           mV           (k=2)
10000           CW                                                       0.00        x                 0.00            0.00           1.00|    300        £15%
                                                                                     Y                 0.00            0.00           1.00     300
                                                                                     Z                 0.00            0.00           1.00     300




 The reported uncertainty of measurement is stated as the standard uncertainty of measurement multiplied
 by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of
 approximately 95%.




* numerical linearization parameter: uncertainty not required

& Uncertainty is determined using the maximum deviation from linear response applying recatangular distribution and is expressed for the square of the field value.




      Certificate No: H3—6163_Apr10                                      Page 4 of 10


H3DV6 SN:6163                                                                                                                      April 26, 2010


                                                                     Frequency Response of H—Field
                                                                          (TEM—Cell:ifi110 EXX, Waveguide R22)


                                                          14


                                                          1.3


                                                          1.2
                        Frequency response (normalized)




                                                          14


                                                          10                                       ::I:Ifl:_:§|
                                                          0.9


                                                          0.8


                                                          0.7


                                                          0.6
                                                                0         500         1000       1500   2000      2500      3000
                                                                                              £ [MHz]
                                                                                xy:          —@—TEM     —B—R22
                                                                                z:           —6—TEM     —R—R22



                                                                    Uncertainty of Frequency Response of H—field: * 6.3% (k=2)




  Certificate No: H3—6163_Apr10                                                         Page 5 of 10


H3DV6 SN:6163                                                                                   April 26, 2010



                                     Receiving Pattern (¢), 9 = 90°

                        f = 600 MHz, TEM ifi110EXX                  f= 2500 MHz, WG R22




                     —#—X     —#—Y    —@—Z   —O—iot              ~@—X   ~@—¥   ~@—20   ~O—tot




                                     Receiving Pattern (¢), 9 = 0°
                      f= 600 MHz, TEM ifi110EXX                    f= 2500 MHz, WG R22




                                             —O—tot                                    ~O—tot




  Certificate No: H3—6163_Apr10                   Page 6 of 10


H3DV6 SN:6163                                                                                                        April 26, 2010



                                             Receiving Pattern (¢), 9 = 90°

                  1.0
                  0.8
                  g'j                                                                                —6—300 MHz
            mon                                                                                      —BI—600 MHz
            io 0.2                                                                                   —&—750 MHz
            2 oo
             8 o»                                                                                    —BI— 1800 MHz
             4 o4                                                                                    —i—2500 MHz

                —0.6
                —0.8
                —1.0
                        1          7              13          19            25      31
                                                            (AW!




                                            Uncertainty of Axial Isotropy Assessment: £ 0.5%(k=2)



                                                   ivi                 —
                                              Receiving Pattern (¢), 8 = 0°

                    1.0
                    0.8
                    0.6
                    0.4                                                                                —0—300 MHz
              ®     o2                                                                                 —BI—600 MHz
              2 o0                                                                                     —4—750 MHz
              2     .02                                                                                —BI— 1800 MHz
              4 .04                                                                                    —&— 2500 MHz
                    —0.6
                    —0.8
                    —1.0
                            0          60           120       180           240     300

                                                            $11




                                            Uncertainty of Axial Isotropy Assessment: £ 0.5% (k=2)




   Certificate No: H3—6163_Apr10                             Page 7 of 10


H3DV6 SN:6163                                             April 26, 2010




Other Probe Parameters

Sensor Arrangement                                           Rectangular
Connector Angle (°)                                               —255.2
Mechanical Surface Detection Mode                               enabled
Optical Surface Detection Mode                                  disabled
Probe Overall Length                                            337 mm
Probe Body Diameter                                              10 mm
Tip Length                                                       20 mm
Tip Diameter                                                     6.0 mm
Probe Tip to Sensor X Calibration Point                           3 mm
Probe Tip to Sensor Y Calibration Point                           3 mm
Probe Tip to Sensor Z Calibration Point                           3 mm




    Certificate No: H3—6163_Apr10         Page 10 of 10