HAC Dipole 2

FCC ID: HLZDME210SC

Test Report

Download: PDF
FCCID_1409407

Calibration Laboratory of
                                                                  S
                                                                   l3                                            Schweizerischer Kalibrierdienst
Schmid & Partner                                                                                                 Service suisse d‘étalonnage
  Engineering AG                                                                                                 Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                                     Swiss Calibration Service


 Accredited by the Swiss Accreditation Service (SAS)                                           Accreditation No.: SCS 108
 The Swiss Accreditation Service is one of the signatories to the EA
 Multilateral Agreementfor the recognition of calibration certificates

 Client




 Object


 Calibration procedure(s)




 Calibration date:


 Condition of the calibrated item



 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                     ID #                      Cal Date (Certificate No.)                          Scheduled Calibration
 Power meter EPM—442A                  GB37480704                08—Oct—08 (No. 217—00898)                           Oct—09
 Power sensor HP 8481A                 US37202783                08—Oct—08 (No. 217—00898)                           Oct—09
 Probe ER3DV6                          SN: 2336                  22—Dec—08 (No. ER3—2336_Dec08)                      Dec—09
 Probe H3DV6                           SN: 6065                  22—Dec—08 (No. H3—6065_—Dec08)                      Dec—09
 DAE4                                  SN 781                    20—Feb—09 (No. DAE4—781_FebO9)                      Feb—10


 Secondary Standards                   ID #                      Check Date (in house)                               Scheduled Check
 Power meter R&S NRP                   SN: 101748                23—Sep—08 (in house check Dec—08)                   In house check: Dec—10
 Power sensor R&S NRP—Z291             SN: 100711                25—Aug—08 (in house check Dec—08)                   In house check: Dec—10
 Power sensor R&S NRP—291              SN: 100712                25—Aug—08 (in house check Dec—08)                   In house check: Dec—10
 Network Analyzer HP 8753E             Us37390585                18—Oct—01 (in house check Oct—08)                   In house check: Oct—09
 RF generator E4433B                   MY 41310391               22—Nov—04 (in house check Oct—07)                   In house check: Oct—09




                                                                              Function                               Signature
 Calibrated by:                                                                  ora!                                  ;


 Approved by:




                                                                                                                    Issued: April 27, 2009
This calibration certificate shall not be reproduced exceptin full without written approval of the laboratory.

Certificate No: CD1880V3—1024_Apr09                                   Page 1 of 6


Calibration Laboratory of                                                                      Schweizerischer Kalibrierdienst
Schmid & Partner                                                                               Service suisse d‘étalonnage
      Engineering AG                                                                           Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                                                   Swiss Calibration Service


Accredited by the Swiss Accreditation Service (SAS)                                        Accreditation No.: SCS 108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreementfor the recognition of calibration certificates

References
[1]       ANSI—C63.19—2006
          American National Standard for Methods of Measurement of Compatibility between Wireless
          Communications Devices and Hearing Aids.

[2]       ANSI—C63.19—2007
          American National Standard for Methods of Measurement of Compatibility between Wireless
          Communications Devices and Hearing Aids.

Methods Applied and Interpretation of Parameters:
          Coordinate System: y—axis is in the direction of the dipole arms. z—axis is from the basis of the antenna
          (mounted on the table) towards its feed point between the two dipole arms. x—axis is normal to the other
          axes. In coincidence with the standards [1, 2], the measurement planes (probe sensor center) are
          selected to be at a distance of 10 mm above the top edge of the dipole arms.

          Measurement Conditions: Further details are available from the hardcopies at the end of the certificate.
          All figures stated in the certificate are valid at the frequency indicated. The forward powerto the dipole
          connector is set with a calibrated power meter connected and monitored with an auxiliary power meter
          connected to a directional coupler. While the dipole under test is connected, the forward power is
          adjusted to the same level.
          Antenna Positioning: The dipole is mounted on a HAC Test Arch phantom using the matching dipole
          positioner with the arms horizontal and the feeding cable coming from the floor. The measurements
          are performed in a shielded room with absorbers around the setup to reduce the reflections.
          It is verified before the mounting of the dipole under the Test Arch phantom, that its arms are perfectly in
          a line. It is installed on the HAC dipole positioner with its arms parallel below the dielectric reference
          wire and able to move elastically in vertical direction without changing its relative position to the top
          center of the Test Arch phantom. The vertical distance to the probe is adjusted after dipole mounting
          with a DASY4 Surface Check job. Before the measurement, the distance between phantom surface and
          probe tip is verified. The proper measurement distance is selected by choosing the matching section of
          the HAC Test Arch phantom with the proper device reference point (upper surface of the dipole) and the
          matching grid reference point (tip of the probe) considering the probe sensoroffset. The vertical
          distance to the probe is essential for the accuracy.
          Feed Point Impedance and Return Loss: These parameters are measured using a HP 8753E Vector
          Network Analyzer. The impedance is specified at the SMA connector of the dipole. The influence of
          reflections was eliminating by applying the averaging function while moving the dipole in the air, at least
          70cm away from any obstacles.
          E— field distribution: E field is measured in the x—y—plane with an isotropic ER3D—field probe with 100 mW
         forward power to the antenna feed point. In accordance with [1, 2], the scan area is 20mm wide, its
         length exceeds the dipole arm length (180 or 90mm). The sensor center is 10 mm (in z) above the top of
         the dipole arms. Two 3D maxima are available near the end of the dipole arms. Assuming the dipole
         arms are perfectly in one line, the average of these two maxima (in subgrid 2 and subgrid 8) is
         determined to compensate for any non—parallelity to the measurement plane as well as the sensor
         displacement. The E—field value stated as calibration value represents the maximum of the interpolated
         3D—E—field, 10mm above the dipole surface.
          H—field distribution: H—field is measured with an isotropic H—field probe with 100mW forward power to the
         antenna feed point, in the x—y—plane. The scan area and sensor distance is equivalent to the E—field
         scan. The maximum of the field is available at the center (subgrid 5) above the feed point. The H—field
         value stated as calibration value represents the maximum of the interpolated H—field, 10mm above the
         dipole surface at the feed point.



Certificate No: CD1880V3—1024_Apr09                            Page 2 of 6


1. Measurement Conditions
     DASY system configuration, as far as not given on page 1.
       DASY Version                                                    DASY4                      V4.7 B80
       DASY PP Version                                                SEMCAD                     V1.8 B186
       Phantom                                                   HAC Test Arch            SD HAC PO1 BA, #1070
       Distance Dipole Top — Probe Center                              10 mm
      Scan resolution                                            dx, dy = 5 mm               area = 20 x 90 mm
      Frequency                                               1880 MHz + 1 MHz
      Forward power at dipole connector                       20.0 dBm = 100mW
      Input power drift                                               < 0.05 dB



2. Maximum Field values

      H—field 10 mm above dipole surface                              condition           Interpolated maximum

      Maximum measured                                      100 mW forward power                 0.468 A/m
    Uncertainty for H—field measurement: 8.2% (k=2)

      E—field 10 mm above dipole surface                              condition           Interpolated maximum
      Maximum measured above high end                       100 mW forward power                 138.5 V/m
      Maximum measured above low end                        100 mW forward power                 137.8 V/m
      Averaged maximum above arm                            100 mW forward power                 138.2 V/im
    Uncertainty for E—field measurement: 12.8% (k=2)

3. Appendix

3.1 Antenna Parameters

      Frequency                               Return Loss                            Impedance
      1710   MHz                              19.0 dB                                (50.6 + j11.3 ) Ohm
      1880   MHz                              21.4 dB                                ( 53.4 + j8.2 ) Ohm
      1900   MHz                              21.9 dB                                ( 55.8 + j6.2 ) Ohm
      1950   MHz                              28.0 dB                                ( 53.3 —j2.5 ) Ohm
      2000   MHz                              19.7 dB                                ( 40.9 + j2.5 ) Ohm

3.2 Antenna Design and Handling

The calibration dipole has a symmetric geometry with a built—in two stub matching network, which leads to the
enhanced bandwidth.
The dipole is built of standard semirigid coaxial cable. The internal matching line is open ended. The antenna is
therefore open for DC signals.
Do not apply force to dipole arms, as they are liable to bend. The soldered connections near the feedpoint may
be damaged. After excessive mechanical stress or overheating, check the impedance characteristics to ensure
that the internal matching network is not affected.

After long term use with 40W radiated power, only a slight warming of the dipole near the feedpoint can be
measured.




Certificate No: CD1880V3—1024_Apr09                     Page 3 of 6


3.3 Measurement Sheets

3.3.1 Return Loss and Smith Chart


                                                            22 fApr 2009     15124120
  CHI s11       LOG       5 dB/REF —18 dB         .        21—21.350 dB     1 880.000 000 MHz




  *4                              ——|——]—]—|—l t                                     —fs        CH1 Markers

                                                                                                1:—19.006 d8
                                                                                                 1.71000 GHz
 Cor
                                                                                                3:—21.935 dB
                                                                                                 1.99000 GHz

                                                                                                4:—28.031 4B
                                                                                                 1.95000 GHz
 Avg
 16                                                                                             5:—19.679 d8
                                                                                                 2.00000 GHz




  CH2   $11     1 U FS          2: 53.412 e   $.1973 8   693.95 pH          1 880.000 000 MHz




 Del                                                                                            CH2 Markers
                                                                                                 1: 50.580 a
 Cor                                                                                                11.334 6
                                                                                                1.71000 GHz
                                                                                                 3: 55.7%2 &
                                                                                                    6.2160 6
                                                                                                1.90000 GHz
 Av                                                                                              4153.271 4
 16°                                                                                              —34229 6A
                                                                                                1.95000 GHz
                                                                                                 5t $0,904 A
                                                                                                   2318¢ 2
                                                                                                200000 GHz

              CENTER 1 880.000 DOn MHz                               SPAN 1 000.000 @aa MHz




Certificate No: CD1880V3—1024_Apr09                      Page 4 of 6


3.3.2   DASY4 H—Field Result
                                                                                Date/Time: 21.04.2009 15:03:58

Test Laboratory: SPEAG Lab 2
DUT: HAC Dipole 1880 MHz; Type: CD1880V3; Serial: 1024
Communication System: CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Medium parameters used: 0 = 0 mho/m, £, = 1; p= 1 kg/m‘
Phantom section: RF Section
Measurement Standard: DASY4 (High Precision Assessment)
DASY4 Configuration:
    *   Probe: H3DV6 — SN6065; Calibrated: 22.12.2008
    e    Sensor—Surface: (Fix Surface)
    e    Electronics: DAE4 Sn781; Calibrated: 20.02.2009
    e    Phantom: HAC Test Arch with AMCC; Type: SD HAC PO1 BA; Serial: 1070
    e    Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 186

H Scan — measurement distance from the probe sensor center to CD1880 Dipole =
10mm/Hearing Aid Compatibility Test (41x181x1):
Measurementgrid: dx=5mm, dy=5mm
Maximum value of peak Total field = 0.468 A/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, —6.30 mm
Reference Value = 0.497 A/m; Power Drift = —0.022 dB
Hearing Aid Near—Field Category: M2 (AWF 0 dB)

                                             Peak H—field in A/m

                                              Grid 1     Grid 2    Grid 3
                                              0.405 M2 0.427 M2 0.408 M2

                                              Grid 4     Grid 5    Grid 6
                                              0.441 M2 0.468 M2 0.449 M2

                                              Grid 7     Grid 8    Grid 9
                                              0.402 M2 0.431 M2 |0.412 M2


                            dB
                             0.000




                            —2.94




                            +.08




                            —8.02




                            41.8




                            q447


                                     0 dB = 0.468A/m




Certificate No: CD1880V3—1024_Apr09                        Page 5 of 6


3.3.3 DASY4 E—Field Result
                                                                                Date/Time: 22.04.2009 14:30:39

Test Laboratory: SPEAG Lab 2
DUT: HAC Dipole 1880 MHz; Type: CD1880V3; Serial: 1024
Communication System: CW; Frequency: 1880 MHz; Duty Cycle: 1:1
Medium parameters used: o = 0 mho/m, &, = 1; p = 1000 kg/m?
Phantom section: RF Section
Measurement Standard: DASY4 (High Precision Assessment)
DASY4 Configuration:
    *   Probe: ER3DV6 — SN2336; ConvF(1, 1, 1); Calibrated: 22.12.2008
    e   Sensor—Surface: (Fix Surface)
    *   Electronics: DAE4 Sn781; Calibrated: 20.02.2009
    *   Phantom: HAC Test Arch with AMCC; Type: SD HAC PO1 BA; Serial: 1070
    *   Measurement SW: DASY4, V4.7 Build 80; Postprocessing SW: SEMCAD, V1.8 Build 186

E Scan — measurement distance from the probe sensor center to CD1880 Dipole =
10mm/Hearing Aid Compatibility Test (41x181x1):
Measurement grid: dx=5mm, dy=5mm
Maximum value of peak Total field= 138.5 V/m
Probe Modulation Factor = 1.00
Device Reference Point: 0.000, 0.000, —6.30 mm
Reference Value = 155.7 V/m; Power Drift = 0.025 dB
Hearing Aid Near—Field Category: M2 (AWF 0 dB)

                                           Peak E—field in V/m

                                            Grid 1     Grid 2     Grid 3
                                            133.7 M2 138.5 M2 |134.7 M2
                                            Grid 4     Grid 5     Grid 6
                                            91.2 M3    193.6 M3   |89.4 M3

                                            Grid 7     Grid 8     Grid 9
                                            131.3 M2   137.8 M2   |133.0 M2

                           4B
                           0.000




                           —3.08




                           —163



                           6.18




                           472



                                   0 dB = 138.5V/m




Certificate No: CD1880V3—1024_Apr09                       Page 6 of 6



Document Created: 2009-10-12 10:18:34
Document Modified: 2009-10-12 10:18:34