Users Manual

FCC ID: HFSMH3

Users Manual

Download: PDF
FCCID_3605223

Quanta Pulse Oximeter


            Model: Q-spo-wu1




MH3418-01                      Version: 3A


Contents
Safety Instructions.................................................................................... 3
Introduction ............................................................................................... 5
  Indication for Use ............................................................................................... 5
  Package Contents .............................................................................................. 5
  Product Overview ............................................................................................... 6
  Product Requirements ........................................................................................ 6

Getting Started .......................................................................................... 7
  Fitting the Finger Sleeve .................................................................................... 7
  Charging the Device ........................................................................................... 8
  Installing the APP ............................................................................................... 9
  Creating User Profile .......................................................................................... 9

Using the Device ..................................................................................... 11
  Pairing the Device with Your Smartphone ........................................................ 11
  Performing Measurement ................................................................................. 12
  Viewing Real-time Readings ............................................................................ 13
  Switching Measurement Mode ......................................................................... 14
  Viewing History ................................................................................................. 15
  Editing User Profile ........................................................................................... 16
  Clear Data ........................................................................................................ 17
  Reset the Device .............................................................................................. 17
  Exit APP ........................................................................................................... 17

Specifications ......................................................................................... 18
  Pulse Oximeter Specifications .......................................................................... 18
  Charger Specifications ..................................................................................... 19

Additional Information ............................................................................ 20
  Troubleshooting ................................................................................................ 20
  Cleaning ........................................................................................................... 20
  Customer Support ............................................................................................ 21
  EU Representative ........................................................................................... 21
  Manufacture Date ............................................................................................. 22
  Supplier’s Declaration ...................................................................................... 23
  Federal Communications Commission (FCC) Statement . 錯誤! 尚未定義書籤。
  Rugulatory Marks ............................................................................................. 29


MH3418-01                                                                                                       EN-2


Safety Instructions
Read the Safety Instructions thoroughly before using this device. Failure to follow
these instructions can cause measuring abnormality or device malfunction.
 The Quanta Pulse Oximeter is to be used for sleep assessment and personal
   reference only. The Quanta Pulse Oximeter is not intended for diagnostic use.
 The Quanta Pulse Oximeter consists entirely of sophisticated medical electrical
   parts so maintenance can only be carried out by professional technicians.
   Unauthorized disassembly of the device by the user is not allowed.
 Do not allow the metal connectors to come into contact with any kind of power
   source during use.
 When the Quanta Pulse Oximeter is low on power, it will automatically stop
   taking measurements and the corresponding indicator lights will blink. Please
   charge the sensor as soon as it shows a low battery state.
 When the Quanta Pulse Oximeter is taking measurements, please try and avoid
   using any instruments or devices that may affect the accuracy of the readings
   (e.g. the use of a blood pressure meter will affect the measurement of the
   pulse).
 The light (the infrared light is invisible) emitted from the device is harmful to the
   eyes, so service technician or user should not stare at the light.
 Do not clip Quanta Pulse Oximeter on edema or tender tissue.
 Do not use the Quanta Pulse Oximeter in a flammable environment.
 Quanta Pulse Oximeter is not a treatment device.
 Pay attention to ensure that the Quanta Pulse Oximeter is not swallowed by
   pets or children.
 Please read through this user manual carefully before using the Quanta Pulse
   Oximeter.
 If the Quanta Pulse Oximeter must be used to take measurements over an
   extended period of time, please inspect the contact point of the emitter and
   receiver that they are in the right position.
 When the Quanta Pulse Oximeter is taking measurements and a stable reading
   cannot be taken then stop taking measurements right away as not everyone is
   suitable to sensor using the Quanta Pulse Oximeter.
 The Quanta Pulse Oximeter can only take measurements when the subject is
   stationary (e.g. while sitting or lying down). Any activities not permitted by the
   attending physician may affect the accuracy of the measurements.
 Do not use the Quanta Pulse Oximeter in an excessively humid environment
   and never submerge it directly under water.
 Do not clean or wipe the Quanta Pulse Oximeter with corrosive and abrasive
   cleaning agents.
 The Quanta Pulse Oximeter has been tested and certified to international
   electro-magnetic compatibility (EMC) standards for medical equipment (EN
   60601-1 and EN 60601-1-2).
MH3418-01                                                                      EN-3


 The Quanta Pulse Oximeter and its accessories should be disposed of properly.
 Battery Caution: There is a risk of explosion if the battery for the Quanta Pulse
  Oximeter is replaced by an incorrect type.
 The Quanta Pulse Oximeter is intended for adult use.
 Quanta Pulse Oximeter does not provide with a low SpO2 alarm condition or
  pulse rate physiological alarm condition.
 Quanta Pulse Oximeter does not support low perfusion detection.
 Clean the skin of Quanta Pulse Oximeter sensor and charger with 75% alcohol;
  and clean the finger sleeve in wash machine.
 If you wear Quanta Pulse Oximeter too long, it may cause uncomfortable by
  long time pressure. Please check if you have any on the finger every eight
  hours.
 Quanta Pulse Oximeter Sensor is a compact all-in-one design with Emitter /
  Receiver embedded inside. So that it doesn’t have to verify the compliance of
  monitor, probe extenders and cable before use.
 Quanta Pulse Oximeter has been validated and tested for compliance with
  international standard ISO 80601-2-61.
 The self life of Quanta Pulse Oximeter is 2 years.
 Quanta Pulse Oximeter is calibrated to display functional oxygen saturation.
 The pulse rate accuracy is verified by the root-mean-square(rms) method which
  is the difference between pulse rate data recorded by Quanta Pulse Oximeter
  and with pulse rate data set by the SpO2 simulator.
 Do not use any unauthorized oximeter simulator to assess the accuracy of
  Quanta Pulse Oximeter. And if there has any question please contact Quanta
  Customer Support.
 The SpO2 accuracy validation, Quanta Pulse oximeter clinical trial was
  performed in accordance with ISO 80601-2-61 and FDA guidance. The study
  population is healthy volunteers who consent to induced hypoxia and arterial
  blood sampling more than 200 data points. Participated subjects is including
  Asian, Caucasian, African, Male, and Female. Their ages are from 21 to 50
  years old. The statistically study result indicated that the SpO2 accuracy (±Arms)
  of Quanta Pulse Oximeter was less than ±3 % in the SpO2 range 70% ~100%
  complied with the accuracy requirement.
 When the finger sleeve is not viscous, or it is recommended to use finger sleeve
  for a month if it’s used every day.




MH3418-01                                                                 EN-4


Introduction
Indication for Use
The Quanta Pulse oximeter is intended for use in the non-invasive measurement of
patient blood oxygen (SpO2) and pulse rate in healthcare institute, and home
environments. The Quanta Pulse oximeter will process SpO2 signals and calculate
pulse rate. SpO2 value and pulse rate data will be transferred to a display device
via Bluetooth, and further displayed on the display device screen with a dedicated
App called Q-SPO2 APP. The device should be worn on the index finger, middle
finger and ring finger along with a finger sleeve and it’s for adult use only. The
Quanta Pulse oximeter is sold to professional organizations such as home care
givers, clinics, or trained patient rather than to the general public. Unauthorized
installation by users is not allowed.


Package Contents
The following items come with your package. If any of the items is missing, please
contact your dealer.




     Pulse Oximeter                  Charger                  Finger Sleeve




        Adapter                   User Manual

Adaptor: Manufacture by ASIAN POWER DEVICES INC.
         Model No: WB-10G05FG
         Input: 90-264V,47~63Hz
         Output: 5V/2A



MH3418-01                                                                  EN-5


Product Overview
Pulse Oximeter

    Power button
       with LED                                                              Pogo charging
                                                                             port




          Probe
        (sensor)




Charger




                                               Pogo pin




Product Requirements
In order to properly use the Quanta Pulse Oximeter the following items are
required:
   The Qunata Pulse Oximeter Sensor
   Finger Sleeve
   A Bluetooth-enabled Android Smartphone* (with Android version 5.0 or above and a
    display resolution of 1920x1080 or 2560x1440)
   The Quanta Pulse Oximeter App Q-SPO2*.

* Items not included in the product package.




MH3418-01                                                                     EN-6


Getting Started
Fitting the Finger Sleeve
Before inserting the device into the finger sleeve, observe the openings on the
finger sleeve.
To fit the finger sleeve, do the following:
    1. With the charging port is facing outwards, insert the bottom side of the
        device halfway into the bottom opening of the finger sleeve.




   2. Insert the upper side of the device into the upper opening of the finger
      sleeve. Then push the device firmly into the finger sleeve until it fully seated.




   3. Use the Velcro strip to fit the device snugly around the finger sleeve.




MH3418-01                                                                       EN-7


Charging the Device
It is recommended that you charge your device prior to first use. It takes
approximately 1.5 hours to fully charge.
To charge the battery, do the following:
   1. Connect the other end of the charger to the USB port of the adapter.
   2. Detach the Velcro strip and flip it over to another side of the finger sleeve so
      that you can access the pogo charging port.
   3. Align and attach the pogo pin of the charger with the charging port of the
      device.




   4. Plug the adapter to a wall outlet.
        The Power LED lights orange while the battery is charging.
        Once the Power LED lights green, this indicates the battery is fully
         charged.


Note: You can also directly charge the device before fitting it into the finger sleeve.




MH3418-01                                                                       EN-8


Installing the APP
To obtain the readings from the device, you have to install Q-SPO2 APP first on
your smartphone.
    1. Make sure your smartphone is connected to the Internet.
    2. Launch Google Play Store application and search for Q-SPO2 APP.
    3. Follow the on-screen instructions to download and install Q-SPO2 APP on
       your smartphone.

Note: In order to install the Q-SPO2 app your smartphone will need at least 8MB of
storage capacity available.
.


Creating User Profile
The first time you launch Q-SPO2 APP, you will be prompted to create a user
profile.
    1. Launch Q-SPO2 APP (         ). The welcome screen appears.




    2. Enter your personal data into the required entry fields.




        IMPORTANT: It is important to enter the correct resting heart rate to
                   ensure the reading accuracy.

MH3418-01                                                                        EN-9


        To determine your resting heart rate, do one of the following:
         Measure your heart rate after at least 5 minutes of resting.
         Obtain a more accurate heart rate measurement performed by a trained
           professional.
   3. Tap Done to save the data. The main screen appears.




Note: You can also modify your user profile later via Profile setting. Refer to page
       16.




MH3418-01                                                                    EN-10


Using the Device
Pairing the Device with Your Smartphone
The device connects and transfers data wirelessly to your smartphone via
Bluetooth.




IMPORTANT: Before pairing, make sure both devices are within Bluetooth range
           (10m).

To pair the device with your smartphone, do the following:
   1. Press the Power button for one second (flash green LED) to turn the device
       on.




   2. On your smartphone, launch Q-SPO2 APP (           ).
      Note: By default, the Bluetooth function will automatically turn on. Otherwise,
            go to Settings > Bluetooth and set the setting to ON to turn on the
            Bluetooth function.

   3. Tap     > Pairing.




MH3418-01                                                                    EN-11


   4. Tap the device to pair.




Note: Not all smartphones can pair with the device directly through Q-SPO2 app.
In this case, you will need to perform the pairing process via Settings app (go to
Settings > Bluetooth, and then select the device to pair).


Performing Measurement
Before performing the measurement, make sure you have successfully paired your
smartphone with the device.
   1. Insert your index finger fully into the finger sleeve. Make sure your finger is
      accurately placed on the probe of the device.




       Note: Keep your finger and hand still during measurement.
   2. On your smartphone, launch Q-SPO2 APP and the real-time readings will
      appear on the screen.
   3. Press the Power button for one second (flash orange LED) to turn the
      device off.
      Note: when you start to measure SpO2 value, make sure to connect
      Q-SPO2 APP every time.




MH3418-01                                                                    EN-12


Viewing Real-time Readings
The main screen displays a real-time PPG waveform, SpO2 value, and pulse rate
which you can present to a medical professional for evaluation.




Note:
 If the SpO2 level falls below 90%, this indicates a need for supplemental oxygen.
  If this persists, please consult your doctor for proper evaluation.
 The SpO2 level and pulse rate would update every three seconds. And it has no
  alarm functions.
 The device will trigger smartphone to vibrate when the battery level is low ( ).
 If the current time displays “     “ and these two icons ( / ) appear on the
  screen, this indicates one of the following:
   The device is turned off.
   The distance between the device and your smartphone exceeds the
       Bluetooth transmission range.
 When Quanta Pulse Oximeter is connected to smart phone. And the sensor
  detects unstable signal or data updating over 30 seconds, it would show below
  message.




MH3418-01                                                                  EN-13


Switching Measurement Mode
This device supports both spot-check and continuous monitoring of SpO2 and
pulse rate. By default, the measurement mode is set to Continuous monitoring.
To switch to spot-check measurement, do the following:
   1. Tap   .
   2. Tap Spot Check.




      In spot-check mode, the sensor performs measurement every 12~18
      seconds and then turns off emitter for a while (12~51 seconds) for power
      saving.




MH3418-01                                                                 EN-14


Viewing History
Tap     > History to view the recorded data that includes a scrollable and scalable
PPG waveform.
The information appears in History main screen as following:




Quick Search
Slide the timeline scroll bar to quickly select the desired time. The numbers on the
scroll bar represent the hour of the day shown, and each point on the scroll bar
represents 10 minutes interval.




MH3418-01                                                                    EN-15


Switching Date
Tap the date to switch to the Calendar view screen.
In Calendar view, the date with the recorded data is marked by a dot(). Tap < or >
to go to other months.
Select the desired date and tap Done to view the recorded data on that day.




 Selected                                                                    Today’s
     date                                                                    date




                                     Recorded data


Note: Q-SPO2 app will store the recorded data up to 7 days.


Editing User Profile
Tap     > Profile to edit the profile settings.




Make the necessary changes and the system will automatically update the user
profile.




MH3418-01                                                                  EN-16


Clear Data
Tap     > Clear Data to delete the user profile and all recorded data.




A confirmation message appears on the screen. Tap Yes to clear all data.


Reset the Device
Tap     > Reset Device to restore the factory default settings.




A confirmation message appears on the screen. Tap OK to reset the device.


Exit APP
Tap     > Quit to stop Q-SPO2 app from running in the background.




A confirmation message appears on the screen. Tap OK to exit Q-SPO2 app.
Note: Once Q-SPO2 app is launched, it keeps running in the background to ensure
the data synchronization will continue when you use the device.



MH3418-01                                                                  EN-17


Specifications
Pulse Oximeter Specifications
Item                       Specifications
                           Continuous SPO2 data acquisition and calculation
                           Pulse Rate:
                            <100, ±3    1/min
Sensor
                            >=100, ±3%      1/min
                           SpO2 Value:
                            Declared Accuracy Range, 70%~100% ±3digits
                           Red 660 nanometer @ 1.91 mWwmaximum average
Emitter wave length
                           IR 905 nanometer @ 0.92 mWwmaximum average
                           Activity Status detection by G-sensor
                            G-sensor (3 axis): ±8g
G-sensor
                            G-sensor sampling rate: 52Hz
                            G-sensor accuracy: ±0.0156g
                           Bluetooth 4.0 BLE
Connectivity
                           Transmit distance: 10 meters (open space)
USB                        USB2.0
Battery                    3.7V/80mAh
                            Continuous measuring: > 5 hours
Battery Life                Spot checking: > 7 hours
                            Charging time: 1.5 hours
Operating Temperature /     Operating temperature: 5⁰C ~ 40⁰C
Humidity                    Operating humidity: 10% ~ 95% (non-condensing)
Storage & transportation    Temperature: -20⁰C ~ 60⁰C
Temperature / Humidity      Humidity: 10% ~ 95% (non-condensing)
Atmospheric Pressure
                           800 hPa ~ 1013 hPa
Range
                           IP22 (Vertically dripping water shall have no harmful
IP classification
                           effect)
Altitude                   2000m
Weight                     11g
Dimensions                 14.6 x 22.5 x 4.1 mm



MH3418-01                                                                  EN-18


Charger Specifications
Item                    Specifications
Input                   5V/0.5A
Operating Temperature   5°C ~ 40°C
Storage Temperature     -20°C ~ 60°C
Weight                  18g
Dimensions              30.7 x 17 x 20.64 mm




MH3418-01                                      EN-19


Additional Information
Troubleshooting
If you encounter any problems when using the device, try the following solutions.
Problem                Possible Cause          Solution
                       The battery is low.     Charge the battery and try again.
The device cannot
be turned on.                                  Please contact Quanta Customer
                       Device malfunction.
                                               Support.
                       The device is turned    Press the Power button to turn the
                       off.                    device on.
                       The Bluetooth           Go to Settings > Bluetooth and
Q-SPO2 APP             function on your        set the setting to ON to turn on the
cannot detect the      smartphone is           Bluetooth function.
device.                disabled.
                       Both devices are out    Keep both devices within 10
                       of the Bluetooth        meters.
                       transmission range.
                       Finger might not be     Retry by inserting the finger into
                       placed accurately on    the finger sleeve until it touches the
                       the probe.              probe (sensor) area.
SpO2 or pulse rate
                                             Remove the device from the finger
does not display on    The probe (sensor) is
                                             sleeve. Then wipe the probe area
the main screen.       dirty.
                                             using a soft dry cloth.
                       Excessive movement Keep your finger, hand, and body
                       during measurement. still during measurement.



Cleaning
It is important to perform cleaning once devices are grease and dirt.
Item                       Cleaning Method
Pulse Oximeter Sensor      Wipe the external surface with a soft dry cloth.
Charger                    Wipe the external surface with a soft dry cloth.
                            Hand wash. If you are using a washing machine,
                             place the finger sleeve inside a laundry bag.
Finger Sleeve               Do not use bleach.
                            Line drying. Do not dry the finger sleeve in the dryer.



MH3418-01                                                                     EN-20


Customer Support
For additional technical information, contact Quanta Customer Support
Department.
               Quanta Computer Inc.(QCI)
               Address:
                  No. 188, Wenhua 2nd Rd., Guishan Dist.,
                  Taoyuan City 333, Taiwan
               TEL: +886-3-327-2345
               FAX: +886-3-318-4207
               Email: MedicalSales@quantatw.com


EU Representative
               EU Representative: MedNet GmbH
               Address: Borkstrasse 10, 48163 Münster, Germany




MH3418-01                                                               EN-21


Manufacture Date
The year of manufacture label and format of the wireless home health
management system's sensor, charger is shown below:




In the serial number on the product label, the 6th and 7th character starting from the left represents
the year of manufacture, while the 8th and 9th character represents the week of manufacture. In
the example shown the “17” represents a manufacture year of 2017 and the “10” represents the
10th week of 2017.




MH3418-01                                                                                    EN-22


Supplier’s Declaration
The Quanta Pulse Oximeter conforms to the international EN 60601-1 and
EN 60601-1-2 standards for electromagnetic compatibility with medical electrical
devices and systems.


   Guidance and manufacturer’s declaration-electromagnetic emissions
The Q-spo-wu1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Q-spo-wu1 should assure that it is used in such an
environment.
  Emission test      Compliance           Electromagnetic environment-guidance
RF emissions         Group 1          The Q-spo-wu1 uses RF energy only for its internal
CISPR 11                              function. Therefore, its RF emissions are very low
                                      and are not likely to cause any interference in
                                      nearby electronic equipment.
RF emissions         Class B          The Q-spo-wu1 is suitable for use in all
CISPR 11                              establishments, including domestic establishments
Harmonic             Class A          and those directly connected to the public
emissions                             low-voltage power supply network that supplies
IEC 61000-3-2                         buildings used for domestic purposes.
Voltage              Compliance
fluctuations
/flicker emissions
IEC 61000-3-3




    Guidance and manufacturer’s declaration-electromagnetic immunity
The Q-spo-wu1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the Q-spo-wu1 should assure that it is used in such an
environment.
 Immunity test          IEC 60601          Compliance           Electromagnetic
                         test level          level           environment-guidance
Electrostatic        + 6 kV contact       + 6 kV contact   Floors should be wood,
discharge(ESD)       + 8 kV air           + 8 kV air       concrete or ceramic tile. If
IEC 61000-4-2                                              floors are covered with
                                                           synthetic material, the relative
                                                           humidity should be at least
                                                           30%
Electrical fast      + 2kV for power      + 2kV for        Mains power quality should
transient/burst      supply lines         power supply     be that of a typical
IEC 61000-4-4        + 1kV for            lines            commercial or hospital
                     input/output lines   Not applicable   environment.
Surge IEC            + 1kV line(s) to     + 1kV            Mains power quality should
61000-4-5            line(s)              differential     be that of a typical
                     + 2kV line(s) to     mode             commercial or hospital
                     earth                Not applicable   environment.
MH3418-01                                                                           EN-23


Voltage Dips,         <5% UT(>95%           <5% UT(>95%    Mains power quality should
short interruptions   dip in UT) for 0,5    dip in UT) for be that of a typical
and voltage           cycle                 0,5 cycle      commercial or hospital
variations on         40% UT(60% dip        40% UT(60%     environment. If the user of the
power supply          in UT) for 5          dip in UT) for 5
                                                           Q-spo-wu1
input lines IEC       cycles                cycles         requires continued operation
61000-4-11            70% UT(30% dip        70% UT(30%     during power mains
                      in UT) for 25         dip in UT) for interruptions, it is
                      cycles                25 cycles      recommended that the
                      <5% UT(>95%           <5% UT(>95%    Q-spo-wu1 be powered from
                      dip in UT) for 5 s    dip in UT) for 5
                                                           an uninterruptible power
                                            s              supply or a battery.
Power               3 A/m                   3 A/m          The Q-spo-wu1 power
frequency(50, 60                                           frequency magnetic fields
Hz) magnetic field                                         should be at levels
IEC 61000-4-8                                              characteristic of a typical
                                                           location in a typical
                                                           commercial or hospital
                                                           environment.
NOTE       UT is the a.c. mains voltage prior to application of the test level.




       Guidance and manufacturer’s declaration-electromagnetic immunity
 The Q-spo-wu1 is intended for use in the electromagnetic environment specified
 below.
 The customer or the user of the Q-spo-wu1 should assure that is used in such
 and environment.
  Immunity test         IEC 60601          Compliance             Electromagnetic
                         test level          level             environment-guidance
                                                           Portable and mobile RF
                                                           communications equipment
                                                           should be used no closer to
                                                           any part of the Q-spo-wu1
                                                           including cables, than the
                                                           recommended separation
                                                           distance calculated from the
                                                           equation applicable to the
                                                           frequency of the transmitter.
 Conducted RF         3 Vrms           3 Vrms
 IEC 61000-4-6        150 KHz to                           Recommended separation
                      80 MHz                               distance:
                                                           d = 1,2 √P
 Radiated RF                           3 V/m
                      3 V/m
MH3418-01                                                                         EN-24


 IEC 61000-4-3     80MHz to 2,5                    d = 1,2 √P 80MHz to 800
                   GHz                             MHz
                                                   d = 2,3 √P 800MHz to 2,5
                                                   GHz
                                                   Where P is the maximum
                                                   output power rating of the
                                                   transmitter in watts (W)
                                                   according to the transmitter
                                                   manufacturer and d is the
                                                   recommended separation
                                                   distance in metres (m).

                                                   Field strengths from fixed RF
                                                   transmitters, as determined by
                                                   an electromagnetic site
                                                   survey, a should be less than
                                                   the compliance level in each
                                                   frequency range.b

                                                   Interference may occur in the
                                                   vicinity of equipment marked
                                                   with the following symbol:



 NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.
 NOTE 2. These guidelines may not apply in all situations. Electromagnetic
 propagation is affected by absorption and reflection from structures, objects and
 people.
 a. Field strengths from fixed transmitters, such as base stations for radio
 (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
 radio broadcast and TV broadcast cannot be predicted theoretically with
 accuracy. To assess the electromagnetic environment due to fixed RF
 transmitters, an electromagnetic site survey should be considered. If the
 measured field strength in the location in which the Q-spo-wu1 is used exceeds
 the applicable RF compliance level above, the Q-spo-wu1 should be observed to
 verify normal operation. If abnormal performance is observed, additional
 measures may be necessary, such as re-orienting or relocating the Q-spo-wu1.
 b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
 than 3 V/m.




MH3418-01                                                                 EN-25


Recommended safety distance
The following table lists the recommended safety distance between the device and
mobile RF communications equipment.
                  Recommended separation distance between
     portable and mobile RF communications equipment and the Q-spo-wu1
The Q-spo-wu1 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
Q-spo-wu1 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the Q-spo-wu1 as recommended below, according to the
maximum output power of the communications equipment.
      Rated         Separation distance according to frequency of transmitter
   maximum                                     m
 output power     150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
 of transmitter        d =1,2√P            d =1,2√P             d =2,3√P
        W
       0,01              0,12                 0,12                  0,23
        0,1              0,38                 0,38                  0,73
         1               1,2                   1,2                   2,3
        10               3,8                   3,8                   7,3
       100                12                   12                    23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.




MH3418-01                                                                  EN-26


            CE Mark: Indicates that the body sensor has been certified
            and conforms to EC Directive 93/42/EEC on medical
            devices.

            Type BF applied part

            Indicates that the body sensor is classified as electrical or
            electronic equipment requiring proper disposal (WEEE
            Directive).
            Indicates the manufacturer's catalogue number.
            Attention: Catalogue number may also be referred to as
            the reference number or reorder number.

            Indicates the manufacture's serial number.


            Indicates the manufacturer's name and address.


            To indicate on the rating plate that the equipment is suitable
            for alternating current only; to identify relevant terminals.

            Indicates the need for the user to consult the instructions for
            use for important cautionary information such as warnings
            and precautions that cannot, for a variety of reasons, be
            presented on the medical device itself.

            Indicates that it is mandatory to refer to the instruction
            manual before using this device.

            Indicates no alarm function.


            Indicates the international protection marking.




MH3418-01                                                                   EN-27


Federal Communications Commission (FCC) Statement
The FCC ID is HFSMH3.

15.21
You are cautioned that changes or modifications not expressly approved by the
part responsible for compliance could void the user’s authority to operate the
equipment.

15.105(b)
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC rules. These limits are designed to
provide reasonable protection against harmful interference in a residential
installation.
This equipment generates uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:

-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help.

This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
1) this device may not cause interference and
2) this device must accept any interference, including interference that may cause
undesired operation of the device.


RF Radiation Exposure Statement:
To comply with the FCC RF exposure compliance requirements, this device and its
antenna must not be co-located or operating in conjunction with any other antenna
or transmitter.
For body worn operation, this device has been tested and meets FCC RF exposure
guidelines when used with an accessory that contains no metal and that positions
the device a minimum of 5 mm from the body. Use of other accessories may not
ensure compliance with FCC RF exposure guidelines

MH3418-01                                                                      EN-28


Regulatory Marks
The Quanta Pulse Oximeter conforms to the following regulatory requirements.
Administrative Regulations on Low Power Radio Waves Radiated Devices
(930322)


Article 12
Without permission granted by the NCC, any company, enterprise, or user is not
allowed to change frequency, enhance transmitting power or alter original
characteristic as well as performance to an approved low power radio-frequency
devices.
第十二條 經型式認證合格之低功率射頻電機,非經許可,公司、商號或使用者均不
得擅自變更頻率、加大功率或變更原設計之特性及功能。


Article 14
The low power radio-frequency devices shall not influence aircraft security and
interfere with legal communications. If found, the user shall cease operation
immediately until no interference is achieved.
The said legal communications means radio communications is operated in
compliance with the Telecommunications Act. The low power radio-frequency
devices must be susceptible with the interference from legal communications or
ISM radio wave radiated devices.
第十四條 低功率射頻電機之使用不得影響飛航安全及干擾合法通信;經發現有干擾現象
時,應立即停用,並改善至無干擾時方得繼續使用。前項合法通信,指依電信法規定作業之
無線電通信。低功率射頻電機須忍受合法通信或工業、科學及醫療用電波輻射性電機設備之
干擾。