User Manual

FCC ID: HFS-CI2

Users Manual

Download: PDF
FCCID_2348568

____________________________________________________________________________



        QUANTA COMPUTER INC. HUAYA PLANT
_________________________________________________________________




      Wearable ECG Monitoring
               Device
           User Manual

                                Ver. 1.0




_________________________________________________________________
2014 Quanta Computer Inc. Huaya Plant                  All rights reserved.


CI2418-01_2A


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

                                                                         INDEX
WARNINGS AND REMINDERS ............................................................................. 2
      WARNINGS: ........................................................................................................................................................ 2
      REMINDERS:....................................................................................................................................................... 3
YEAR OF MANUFACTURE ..................................................................................... 5
SUPPLIER'S DECLARATION ................................................................................. 7
REGULATORY MARKS ......................................................................................... 10
INTRODUCTION ....................................................................................................... 12
EQUIPMENT ............................................................................................................... 13
      A.NB / PC ....................................................................................................................................................... 13
EQUIPMENT OPERATING INSTRUCTIONS................................................ 15
      A. PERSONAL COMPUTER (NB/ PC) ....................................................................................................... 15
      B. QUANTA ECG MONITOR APPLICATION, NYA- DATA- CMD [FOR USE ON PERSONAL
         COMPUTERS (NB / PC)] ......................................................................................................................... 16
      C. NYA BODY MONITOR ........................................................................................................................... 21
      D. NYA BODY MONITOR ACCESSORY .................................................................................................. 25
      E. ECG ELECTRODES .................................................................................................................................. 26
      F. NYA CHARGER ........................................................................................................................................ 27
CLEANING OF THE WIRELESS HOME HEALTH MANAGEMENT
SYSTEM ........................................................................................................................ 29
PRODUCT SPECIFICATIONS ............................................................................. 29
PRODUCT LIST ......................................................................................................... 30




___________________________________________________________________________
User Manual V1.0                                                   Page: 1


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Warnings and Reminders

  Warnings:

       The NYA Quanta Wireless Home Health Management System is to be used for
        clinical assessment and reference only. The NYA Quanta Wireless Home Health
        Management System is not intended for diagnostic use.
       The NYA Quanta Wireless Home Health Management System consists entirely of
        sophisticated medical electrical devices so maintenance can only be carried out by
        professional technicians. Unauthorized disassembly of the device by the user is not
        allowed.
       The NYA Quanta Wireless Home Health Management System must be used with
        the specified "accessories" and "electrodes". The use of accessories and electrodes
        from other brands may lead to faulty readings.
       Do not allow the metal connectors to come into contact with any kind of power
        source during use.
       Before using the NYA Quanta Wireless Home Health Management System. Please
        make sure that all of the machine's indicator lights are clearly readable in the
        current usage environment.
       Damaged or faulty accessories and electrodes should not be used.
       When the NYA Quanta Wireless Home Health Management System experiences an
        error, it will automatically stop taking measurements and the corresponding
        indicator lights will blink.
       When the NYA Quanta Wireless Home Health Management System is low on
        power, it will automatically stop taking measurements and the corresponding
        indicator lights will be blink.
       When the NYA Quanta Wireless Home Health Management System is taking
        measurements, please try and avoid using any instruments or devices that may
        affect the accuracy of the readings (e.g. the use of the blood pressure meter will
        affect the measurement of the pulse).
       When the NYA Quanta Wireless Home Health Management System is taking
        measurements, please try and avoid using other electronic devices nearby. If it must
        be used with other electronic devices simultaneously, please check to make sure
        that the NYA Quanta Wireless Home Health Management System is continuing to
        take measurements normally.
       NYA-specific accessories are defined in this user guide. The use of accessories
        from other brands may damage the device.
       Do not use the NYA Quanta Wireless Home Health Management System in a
        flammable environment.




___________________________________________________________________________
User Manual V1.0                                                   Page: 2


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
  Reminders:

       Please read through this user guide carefully before using the NYA Quanta
        Wireless Home Health Management System.
       If the NYA Quanta Wireless Home Health Management System must be used to
        take measurements over an extended period of time, please inspect the contact point
        of the electrodes at least once every 8 hours to make sure that the electrodes are in
        the right position and there is no allergic skin reaction from the user.
       Do not use medical electrical equipment such as "defibrillators" when using the
        NYA Quanta Wireless Home Health Management System.
       When the NYA Quanta Wireless Home Health Management System is taking
        measurements, if a stable reading cannot be taken then stop taking the measurement
        right away (not everyone is suitable for monitoring with the NYA Quanta Wireless
        Home Health Management System).
       The NYA Quanta Wireless Home Health Management System can only take
        measurements when the subject is stationary (e.g. while sitting or lying down) or
        engaging in ordinary exercise. Any activities not permitted by the attending
        physician may affect the accuracy of the measurements.
       Do not use the NYA Quanta Wireless Home Health Management System in an
        excessively humid environment and never submerge it directly under water.
       Do not clean or wipe the NYA Quanta Wireless Home Health Management System
        with corrosive and abrasive cleaning agents.
       The NYA Quanta Wireless Home Health Management System consists entirely of
        sophisticated medical electrical devices so maintenance can only be carried out by
        professional technicians. Unauthorized disassembly of the device by the user is not
        allowed.
       When the NYA Quanta Wireless Home Health Management System is low on
        power, please insert it in the charger for re-charging as soon as possible.
       The disposal of the NYA Quanta Wireless Home Health Management System and
        its accessories should comply with the relevant local regulations.
       The NYA Quanta Wireless Home Health Management System has been tested and
        certified to international electro-magnetic compatibility (EMC) standards for
        medical equipment (EN 60601-1 and EN 60601-1-2) for the user's peace of mind.
       The NYA Quanta Wireless Home Health Management System and its accessories
        should be disposed of properly. If you are uncertain about the proper disposal
        method, please contact customer service at Quanta Computer Inc. Huaya Plant for
        assistance.
       The accuracy of the medical electrical equipment may be affected if used near
        high-frequency communications equipment.
       Battery Caution
        Risk of explosion if battery is replaced by an incorrect type.
        Dispose of used batteries according to the instructions

___________________________________________________________________________
User Manual V1.0                                                   Page: 3


      QUANTA COMPUTER INC.                   Wireless Home Health Management System
____________________________________________________________________________
     Federal Communications Commission (FCC) Statement
        15.21
        You are cautioned that changes or modifications not expressly approved by the part
        responsible for compliance could void the user’s authority to operate the
        equipment.

         This device complies with part 15 of the FCC Rules. Operation is subject to the
         following two conditions:
         1) this device may not cause interference and
         2) this device must accept any interference, including interference that may cause
         undesired operation of the device.

         15.105(b)
         This equipment has been tested and found to comply with the limits for a Class B
         digital device, pursuant to part 15 of the FCC rules. These limits are designed to
         provide reasonable protection against harmful interference in a residential
         installation.
         This equipment generates uses and can radiate radio frequency energy and, if not
         installed and used in accordance with the instructions, may cause harmful
         interference to radio communications. However, there is no guarantee that
         interference will not occur in a particular installation. If this equipment does cause
         harmful interference to radio or television reception, which can be determined by
         turning the equipment off and on, the user is encouraged to try to correct the
         interference by one or more of the following measures:

         -Reorient or relocate the receiving antenna.
         -Increase the separation between the equipment and receiver.
         -Connect the equipment into an outlet on a circuit different from that to which the
         receiver is connected.
         -Consult the dealer or an experienced radio/TV technician for help.
         .




___________________________________________________________________________
User Manual V1.0                                                   Page: 4


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Year of Manufacture

   The year of manufacture label and format of the wireless home health management
   system's body monitor, charger and ECG electrodes are shown below:

    Sensor




                                                            In the serial number on the
                                                            product label, the 7th
                                                            character starting from the
                                                            left (“14” in the shown
                                                            example) represents the
    Charger                                                sensor and charger's year of
                                                            manufacture.




       Example:
            If the serial number is “QTCI2C425001”
            The 7th characters is mean “14”
            Then the year of manufacture is 2014 AD.




___________________________________________________________________________
User Manual V1.0                                                   Page: 5


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
    ECG Electrodes
       The effective date of the ECG electrodes is labeled along the lower edge of the
       packaging in impact type (not printed) as shown below:




       Example: 220801X 2015-07
       Explanation:   220801X  Manufacturer lot number
                      2015-07  Valid before July 2015




___________________________________________________________________________
User Manual V1.0                                                   Page: 6


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Supplier's Declaration

The NYA Quanta Home Health Management System conforms to the international EN
60601-1 and EN 60601-1-2 standards for electromagnetic compatibility with medical
electrical devices and systems.

                         Compliant
  Emission Testing                        Electromagnetic Environment Guidelines
                          Standard
  The recommended electromagnetic environment for this instrument is shown in the
  table below. The customer or user should ensure that this instrument is used in the
  described environment.
  RF Emissions      Group 1            Radio-frequency is only used by the internal
  CISPR 11                             functions of this instrument. The emission
                                       energy is also very low and does not interfere
                                       with nearby electronic instruments.
  RF Emissions      Class B            This instrument is suitable for home use
  CISPR 11                             (including residential or other mixed
  Harmonic          IEC 61000-3-2      residential/public low-voltage power grids)
  Emissions
  Voltage           IEC 61000-3-3
  Fluctuations/
  Flicker
  Emissions


                                                                 Electromagnetic
                       IEC 60601-1-2          Compliant
       Testing                                                    Environment
                          testing             Standard
                                                                   Guidelines
  The recommended electromagnetic environment for this instrument is shown in the
  table below. The customer or user should ensure that it is used in the described
  environment.
  Electrostatic    ±6 kV contact        ±6 kV contact       The floor should be
  Discharge        ±8 kV air            ±8 kV air           timber, concrete or
  (ESD)                                                     ceramic tile. If the
  IEC 61000-4-2                                             floor is covered in a
                                                            synthetic material then
                                                            it should have a
                                                            relative humidity of at
                                                            least 30%.
  Power Frequency 3 A/m                 3 A/m               Power frequency and
  (50/60Hz)                                                 magnetic field in
  Magnetic Field                                            representative
  IEC 61000-4-8                                             commercial or hospital
                                                            environments.



___________________________________________________________________________
User Manual V1.0                                                   Page: 7


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

             Supplier's declaration and guidelines—Electromagnetic Immunity
                        IEC 60601-1-2        Compliant     Electromagnetic Environment
      Testing
                            testing           Standard                Guidelines
  The recommended electromagnetic environment for this instrument is shown in the table
  below. The customer or user should ensure that it is used in the described environment.
  No part of the NYA Quanta Wireless Home health management System should be placed
  near portable and mobile radio-frequency communications equipment. The recommended
  safety distance can be calculated using the radio frequency and the following formula.
  Radio-frequency 3 V/m                     [3] V        Recommended safety distance
  IEC 61000-4-6   150 kHz to 80 MHz                      d=1.17 P

  Radio-frequency 3 V/m                     [3] V/m      d=1.17 P 80 MHz to 800MHz
  IEC 61000-4-3   80 MHz to 2.5 GHz
                                                         d=2.33 P 800MHz to 2.5 GHz

                                                         P is the maximum output of the
                                                         transmitter (Unit: Watt)
                                                         d is the recommended safety
                                                         distance (Unit: Meter)

                                                         The field intensity of the fixed RF
                                                         transmitter (after an EM site
                                                         survey) should be lower than the
                                                         recommended values for the above
                                                         frequency ranges.

                                                         Interference may be produced near
                                                         instruments that have the
                                                         following symbol:




The field intensity of fixed RF transmitters (e.g. radio, mobile phone, ham radio, FM and AM
radio broadcasts and TV stations) cannot be predicted from the frequency alone. To evaluate
the EM field intensity of the environment, an EM site survey may be necessary. If the
measured field intensity for a region exceeds the above recommended range, the device
should be observed or checked to make sure it operates normally. If abnormal operation is
observed, further action such as adjusting the direction or re-positioning of the equipment
should be taken.

Attention: For frequencies between 80 MHz and 800 MHz, use the higher frequency range.

Attention: These guidelines may not be applicable to all situations. Electromagnetic waves
       can be absorbed or reflected by buildings, objects and people.


___________________________________________________________________________
User Manual V1.0                                                   Page: 8


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

                                 Recommended safety distance
The following table lists the recommended safety distance between the device and mobile RF
communications equipment.

The recommended EM environment for the NYA Quanta Wireless Home Health Management
System is where RF interference is controlled. The user should maintain the recommended
safety distance with the devices and portable/mobile RF communications equipment listed in
the following table (corresponding to the maximum output of the RF transmitter).
                        Recommended Safety Distance (corresponding to RF transmitter
                        output)
 Maximum output of 150 kHz to 80 MHz 80 MHz to 800 MHz               800 MHz to 2.5 GHz
RF transmitter (Watt)       d=1.17 P               d=1.17 P               d=2.33 P
         0.01                   0.12                  0.12                    0.23
         0.1                    0.37                  0.37                    0.74
           1                     1.2                  1.2                      2.3
          10                     3.7                  3.7                      7.4
         100                     12                    12                      23

If the maximum output of the transmitter is not shown in the above table, use the formula to
derive the recommended safety distance d (Meter) with P (Watt) being the maximum output
of the transmitter.


Attention: For frequencies between 80 MHz and 800 MHz, use the higher frequency range.


Attention: These guidelines may not be applicable to all situations. Electromagnetic waves
           can be absorbed or reflected by buildings, objects and people.




___________________________________________________________________________
User Manual V1.0                                                   Page: 9


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Regulatory Marks


The Wireless Home Health Management System conforms to the following regulatory
requirements.
Administrative Regulations on Low Power Radio Waves Radiated Devices (930322)
  Article 12
     Without permission granted by the NCC, any company, enterprise, or user is not
     allowed to change frequency, enhance transmitting power or alter original characteristic
     as well as performance to an approved low power radio-frequency devices.
  Article 14
     The low power radio-frequency devices shall not influence aircraft security and
     interfere with legal communications. If found, the user shall cease operation
     immediately until no interference is achieved.
     The said legal communications means radio communications is operated in compliance
     with the Telecommunications Act. The low power radio-frequency devices must be
     susceptible with the interference from legal communications or ISM radio wave
     radiated devices.


                    CE Mark: Indicates that the body monitor has been certified and
                             conforms to EC Directive 93/42/EEC on medical devices.


                    Type CF applied part



                    Indicates that the body monitor is classified as electrical or electronic
                    equipment requiring proper disposal (WEEE Directive)


                    Indicates the manufacturer's catalogue number
                    Attention: Catalogue number may also be referred to as the
                                reference number or reorder number.

                    Indicates the manufacture's serial number.


                    Indicates the manufacturer's name and address


                    To indicate on the rating plate that the equipment is suitable for
                    alternating current only; to identify relevant terminals.


___________________________________________________________________________
User Manual V1.0                                                  Page: 10


      QUANTA COMPUTER INC.                  Wireless Home Health Management System
____________________________________________________________________________
                 Indicates the need for the user to consult the instructions for use for
                 important cautionary information such as warnings and precautions that
                 cannot, for a variety of reasons, be presented on the medical device
                 itself

                   Indicates the need for the user to consult the instructions for use.




                   Indicates the identification of CLASS II EQUIPMENT only




___________________________________________________________________________
User Manual V1.0                                                  Page: 11


OE QUANTA COMPUTER INC.                    Wireless Home Health Management System

Introduction


     ©@ "Electrocardiogram" (ECG) Waveform


                      Required equipment                          Remarks

            Smart phone (E)+ NYA body monitor (G)

                                                        Choose one of the two
                                                        measurement options on the
          ECG electrode (I) +         Body Monitoring   left based on the usage
            Accessory (H)                 Shirt(J)      requirements, nursing
                                                        personnel‘s requirements or
                                                        user‘s personal preference.




User Manual V1.0                                                            Page: 12


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Equipment
  A.NB / PC
     Functions
     The device requires a commercially
     available personal computer (NB/PC) with
     Bluetooth capabilities. In addition to the
     basic functions of a computer, the
     installation provides an exclusive
     application (UI, as below) capable of
     performing electrocardiogram (ECG)
     measurements. Doctors and medical staff
     can effectively monitor health information
     by uploading the measurements to the main
     server via Bluetooth.


     Quanta ECG Monitor Application UI (for use on NB / PC)
     Functions
     The application (UI) can
     be installed on any
     personal computer (NB
     /PC) to provide ECG
     data measurement
     functions. With the
     Quanta ECG Monitor
     device (B), body
     monitoring accessories
     (D), ECG electrode
     patches (C) , measured
     data will be uploaded to
     the server via Bluetooth
     so doctors and medical
     staff can conduct carry
     out remote monitoring.


  B. NYA Body Monitor
     Function
     The ECG signal measured through this device is wirelessly transmitted using low-power
     Bluetooth technology to the smart phone for processing. The data is then uploaded via 3G
     wireless network to the cloud server.
     This device requires the ECG
     Electrodes (C) in order to
     take measurements.



___________________________________________________________________________
User Manual V1.0                                                  Page: 13


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

  C. NYA Body Monitor Accessory
     Function
     When the electrodes are used, this accessory is used to attach the NYA Body Monitor (B) and
     transmit the ECG signal between electrodes.




                                             Back                                    Front



  D. ECG Electrodes
     Function
     Attach to the designated location on the chest in
     order to capture physiological ECG signal for
     the NYA Body Monitor (B) for processing and
     upload.




  E. NYA Charger
     Function
     Used for charging the NYA Body
     Monitor (B). The charger has two
     slots that allow it to charge two
     Body Monitor (B) units at the
     same time.




___________________________________________________________________________
User Manual V1.0                                                  Page: 14


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Equipment Operating Instructions


  A. Personal Computer (NB/ PC)

     According to the specifications of individual personal computers (NB / PC).

     Operating Instructions:
        Start up the computer
         After starting up your personal computer (NB / PC), users can access the available
         functions.
        Shut down the computer
         To shut down your personal computer (NB / PC), click Shut Down.
        Bluetooth Functions
           Install a Bluetooth dongle on your personal computer (NB / PC).
           Plug in your Bluetooth dongle to the USB port to turn the Bluetooth
             functions on or off.
           Upon selecting and launching Bluetooth, the Bluetooth interface will be
             displayed. Click to search for the device.
           Your device name will be displayed in the Bluetooth selection field. Click on
             your device name to add it to the system.
           The configuration and connection status will be shown on top of the device
             name in your list of Bluetooth devices.




___________________________________________________________________________
User Manual V1.0                                                  Page: 15


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
  B. Quanta ECG Monitor application, NYA- Data- CMD [for use on
     personal computers (NB / PC)]
     Operating Instructions:


        Ensure that you have completed the installation of the Quanta ECG Monitor.
         Please refer to the next chapter, “G. Quanta ECG Monitor Body Monitoring
         Device" for the instructions and diagrams.
        On your computer's desktop, click the NYA-Data-CMD UI
         icon, as shown, to launch the application.
        The following window will be displayed upon launching the
         application.




            A      B




                 Button                   Function

                   A           Communication Port Settings

                   B           Create New Case

                   C           Subject & Case Management




___________________________________________________________________________
User Manual V1.0                                                  Page: 16


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
      1. Communication Port Settings
        Set up the com port to receive Vital Sign Sensor data.
        Click "System"  "Com Port Setting" to open the following window.




              Port
                The system will automatically detect and list the current Com Ports on the
                computer. Select the Port to connect to.
              Baud Rate
                Select 115200.
              Parity
                Select None.
              Date Bits
                Select 8.
              Stop Bits
                Select 1.
                After making the changes, click "Save" to complete. The changes will
                 not take effect if you cancel and exit the screen.




___________________________________________________________________________
User Manual V1.0                                                  Page: 17


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
      2. Create New Case
         Clicking "New Case" will display the following window:




         Create a New Case as follows:
           a. Click "New Case" to display the following window.




              Confirm the subject in the Subject field. Enter the Resting HR of the Subject
              and click "OK".
           b. Ensure Bluetooth has been properly connected to the Serial Port of the
              hardware.
           c. Click "Open" to open the Com Port as shown in the diagram below.


              If correctly launched, the Status will change to Connected as shown in the
              diagram below. To close the Com Port, click "Close". If during the
              measurement process, the Com Port connection disconnects due to instability,
              the system will close the Com Port automatically.


           d. Click "Start" on the left of the screen to connect to the hardware, send out the
              initializing data and prepare to receive the test data.
         e. To end the test, click "Stop" on the right of the screen. Upon clicking "Stop",
            the data automatically save to the database.
___________________________________________________________________________
User Manual V1.0                                                                  Page: 18


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
         f. Screen description:




           A. Displays the basic information of the case.
           B. Displays the time elapsed for the test.
           C. Displays the battery life of the hardware.
           D. Displays whether a Fall Event occurs. Click "Cancel" to execute Cancel
              Fall Event command.
           E. Auto ECG command. By clicking "Auto ECG", the ECG waveform will
              update every 3 seconds; HR, AAC, PAC, PAI, PAEE and battery will
              update every 15 seconds; and Steps, Stride, Speed, Calories, and
              Distance will update every 60 seconds.
                 ECG waveform display
                  Enlarge will enlarge the ECG waveform measured.




                 Change Y-axis range of ECG display
                  On the Y-axis range, right-click to
                  display a selection list. Select Edit as
                  shown in the diagram right.




___________________________________________________________________________
User Manual V1.0                                                  Page: 19


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

               Correct the two values, Min and Span, on the General Tab. Min is the
               starting value of Y-axis; Span is the length of Y-axis.




              Battery life display of monitoring device
               Under normal connection and usage, the battery life symbol will display as
               follows. Different battery bars represent different battery life.
               High battery life,   Medium battery     Low battery life,   Critical battery
               estimated usage      life, estimated    estimated usage     life, with less
               time of 12 to 16     usage time of 7    time of 2 to 6      than 1 hour of
               hours.               to 11 hours.       hours.              battery life.
                                                                           Please charge
                                                                           immediately.
                   4 bars (full)         3 bars              2 bars               1 bar
              Note: The above battery life display and usage time is for reference only, as
                     the data is obtained by testing on a new product. The actual usage time
                     will defer with different conditions such as charging conditions,
                     number of times used, environmental temperature, distance from base
                     station, connectivity, etc.

      Condition              Possible Reason                  Recommended Actions
    Data           The subject turns over or sleeps    We recommend putting on the
    disconnected   on one side causing the Quanta      physiological monitoring clothing
    or omitted     ECG Monitor to detach from the      while sleeping.
    while          accessories or causing poor
    sleeping.      connection due to pressure from
                   the body.
                   The Quanta ECG Monitor              While sleeping, put the handphone
                   monitoring device is a safe and     on your headboard or within 1 to 2
                   low power Bluetooth device. The     meters of the bedside. We
                   electric wave emitted is weak. If   recommended placing the
                   the subject sleeps on one side or   handphone at a position higher than
                   wraps one's arms around the body    the bed. Do not put the handphone
                   while sleeping, the device may      on the floor. Refrain from placing
                   easily become disconnected or       the handphone at a position lower
                   obstructed by the body.             than or at the end of the bed.


___________________________________________________________________________
User Manual V1.0                                                  Page: 20


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________


  C     Subject & Case Management
        After executing "Subject & Case", the following screen will be displayed:
         A. Click "New Subject" button.
         B. Enter Name, Sex, Birthday, Height and Weight.
         C. Click "Save All" to save into the database. If the Name entered already exists,
              the system will display the following alert message as shown in the diagram.
         D. Select the Subject Name in the "Subject Name" field. The subject's test cases
              will display in the case list below.




  C. NYA Body Monitor

       External guide to body monitor:
                                                          Male-end of magnetic clip
                          LED status light


                                                              Charger detection port


                                                              Bluetooth pairing button

                                                                       Waterproof vent




                Front of Body Monitor                       Back of Body Monitor




___________________________________________________________________________
User Manual V1.0                                                  Page: 21


      QUANTA COMPUTER INC.                    Wireless Home Health Management System
____________________________________________________________________________
       Bluetooth matching (These settings will be set by the service representative)
          (1) When the “Smart Phone” or “NYA Monitor” is used for the first time or
              updated they must be paired.
          (2) Open the smart phone and make sure that it has Bluetooth enabled (See E.
              Smart Phone Instructions).
          (3) Use the pin to press the Bluetooth pairing button on the “NYA Monitor” to
              start pairing.
          (4) During the pairing process, the LED status light will alternatively blink
              green/orange for 1 minute.
          (5) Once pairing is successful, the green LED will stay lit for 5 seconds and the
              mobile phone will show that it is connected.
          (6) If pairing fails, the LED status light on the monitor will alternatively blink
              green/orange for 1 minute then stop pairing and searching. All lights will then
              go out.
       Instructions for the Body Monitor
          (1) Clip the ECG electrode's male connector to the female connector on the
              accessory as shown below:



                                                        Wire-side
                          Device-side




          (2) Remove the paper strip from the ECG electrode. For step 1, attach the
              device-side to the lower edge of the left chest. For step 2, attach the wire-side
              to the lower edge of the right collarbone as shown below:




___________________________________________________________________________
User Manual V1.0                                                  Page: 22


      QUANTA COMPUTER INC.                    Wireless Home Health Management System
____________________________________________________________________________
         Attention: The ECG electrodes and accessory must not be worn in the following
                manner for an extended period of time. This prevents the dangling end
                from short-circuiting with any power supply.




                 Wire-side hanging loose                                Device-side hanging
                                                                        loose

          (3) Clip the male magnetic male connector on the monitor to the female connector
               on the accessory as shown below. Note: The circular red or blue mark on
               the monitor must point upwards and must not be upside-down. Please see
               below.



                     Circular mark pointing upwards




          (4) Launch the NYA-Data-CMD Application” on the PC/NB (Please see the
               previous section B. ) shown below:

                    NYA-Data-CMD
                    Application icon


               Note: Before launching the application, only one body monitor can be
               used/activated at any time. Using two monitors at the same time will
               cause the App system to fail.
          (5) Once the app is running, please see the instructions in section B.



___________________________________________________________________________
User Manual V1.0                                                  Page: 23


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

      Other body monitor information:
          (a) If the monitor has a low battery, the orange LED light will blink.
          (b) When the monitor is placed on the accessory (or body monitoring shirt), the
              green LED light on the monitor will blink for 5 seconds before going out.
          (c) During standby and measurement, the monitor's status light will not be shown
              except when power is blow.
          (d) When the monitor is being charged in the charger, the status lights will
              remain off.
          (e) As long as the monitor is not on the accessory, tapping the monitor twice will
              cause the green LED to come on. The monitor will then start transmitting data
              (the smart phone app must be running and connected). Once the transmission
              is complete, the green LED light goes out. If the smart phone does not have
              the app running, the green LED will stay on for two minutes then go out.
          (f) When the ECG electrode falls from the user or the monitor becomes detached
              from the accessory, the green LED on the monitor blinks once every 3
              seconds.




___________________________________________________________________________
User Manual V1.0                                                  Page: 24


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
  D. NYA Body Monitor Accessory

     Used for connecting the body monitor and ECG electrodes so the best ECG signal can
     be acquired. Please see the previous section "“G. NYA Body Monitor" for further
     information and graphics.

            Magnetic Clip Female Connector            Pressure Clip Female Connector
                                  ECG wiring




                     Front of accessory                   Reverse of accessory

        Attention:
            If any wear or damage to the wire
            cladding is found during use, replace
            immediately.




___________________________________________________________________________
User Manual V1.0                                                  Page: 25


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
  E. ECG Electrodes
     The ECG Electrodes are conductors placed on the skin of the chest for transmitting the
     signal generated by the beating of the heart. The signal is transmitted to the body
     monitor for amplification and processing into the ECG signal.

                  Pressure Clip Male
                          Connector




                                                   Back of electrodes
                    Front of electrodes

     Attention:
          Before attaching the ECG electrodes, make sure that the skin surface is clean.
           It may necessary to wipe the skin clean with a wet towel or alcohol swabs then
           wait for it to dry before attachment.
          Electrodes are not re-usable.
          If the skin at the attachment point shows signs of a rash/blisters/reddening or
           other discomfort, please contact the medial personnel or physician.




___________________________________________________________________________
User Manual V1.0                                                  Page: 26


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
  F. NYA Charger
     Charger function:
                                                                          Magnetic clip
                                                                       female connector
                                           Charger detection
                                           probe




             LED status light

       (1) The magnetic clip male-connector on the monitor and the magnetic clip female
           connector of the charger clips with each other. Pay attention to the direction
           when placing the monitor on the charger. The charger detection port on the
           monitor must be aligned with the charger detection probe on the charger as
           shown below.

                                                  Charger detection probe




                                                        Charger detection port
       (2) When the charger is closed and plugged into power mains, the orange LED will
           light up when it is charging normally as shown below.




                                                 During charging the orange LED stays on.

___________________________________________________________________________
User Manual V1.0                                                  Page: 27


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________

       (3) Once the monitor is fully charged, the orange LED on the charger goes out and
           the green LED comes on. Remove the charger from the socket, open the cover of
           the charger and insert finger into the ventral groove to pick up the monitor for
           use as shown below:




     The green LED lights up after charging    Insert finger into ventral groove and pick up
                                                                  monitor.

     Other charger functions
   (1) When the charger is plugged into the socket by itself (with no monitors), the orange
       LED light lights up for two seconds.
   (2) Place the monitor on the charger and connect the transformer to the power socket.
       The orange LED light on the charger will now light up.
   (3) Once the monitor has been fully charged, the orange LED is extinguished and the
       green LED slights up.




___________________________________________________________________________
User Manual V1.0                                                  Page: 28


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Cleaning of the Wireless Home Health Management System

            Item      Model Name                        Cleaning Method

    Body Monitor     NYA Target        Wipe with dry cloth

    Charger          NYA Charger       Wipe with dry cloth
    Accessory (S)    NYA Patch S
    Accessory (M)    NYA Patch M       Wipe with damp cloth
    Accessory (L)    NYA Patch L


Product Specifications

    HR Monitor          Measuring Lead : Lead I
                        HR Range : 30~240
                        Frequency response : 0.1~40Hz
                        Heart rate accuracy : ±3 or ±3%
                        Sampling rate : 256Hz
    Activity          G-sensor (3 axis) : ±8g
                      G-sensor sampling rate : 52Hz

    Radio             Bluetooth 3.0+EDR
                      Tx: 0 dbm
                      Transmission range : 10 meters

    Battery           Lithium polymer battery 3.7V/140 mAh
                      Battery life : about 15 hours under measuring mode (Based on
                       link with iPhone and App on)
    Operating         5°C ~ 40°C
    Temperature




___________________________________________________________________________
User Manual V1.0                                                  Page: 29


      QUANTA COMPUTER INC.             Wireless Home Health Management System
____________________________________________________________________________
Product List

  Part Name/Model      Part Description




___________________________________________________________________________
User Manual V1.0                                                  Page: 30



Document Created: 2014-07-18 18:03:50
Document Modified: 2014-07-18 18:03:50

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC