SAR Test Report Annex C_2

FCC ID: HD5-CN85L1N

RF Exposure Info

Download: PDF
FCCID_4050455

Measurement Conditions
   DASY system configuration, as far as not iven on page 1.
     DASY Version                                              DASYS                              V52.10.1
     Extrapolation                                   Advanced Extrapolation
     Phantom                                         Modular Flat Phantom
     Distance Dipole Center — TSL                              10 mm                             with Spacer

     Zoom Scan Resolution                              dx, dy, dz =5 mm
     Frequency                                         1750 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity         Conductivity
     Nominal Head TSL parameters                                  22.0 °C           40.1               1.37 mho/m
     Measured Head TSL parameters                              (22.0 + 0.2) °C   39.0 + 6 %         1.36 mho/m + 6 %

     Head TSL temperature change during test                      <0.5 °C

SAR result with Head TSL

     SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
     SAR measured                                        250 mW input power                     9.10 W/kg
     SAR for nominal Head TSL parameters                     normalized to 1W        36.3 W/kg + 17.0 % (k=2)

     SAR averaged over 10 cm* (10 g) of Head TSL                 condition
     SAR measured                                        250 mW input power                  4.82 Wikg
     SAR for nominal Head TSL parameters                     normalized to 1W         19.3 Wikg + 16.5 %(k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                               Temperature       Permittivity          Conductivity
     Nominal Body TSL parameters                                  22.0 °C           53.4               1.49 mho/m

     Measured Body TSL parameters                              (22.0 + 0.2) °C   53.6 +6 %          1.47 mho/m + 6 %
     Body TSL temperature change during test                      <0.5 °C            ————                    «++


SAR result with Body TSL

     SAR averaged over 1 cm‘ (1 g) of Body TSL                  Condition
     SAR measured                                        250 mW input power                     9.12 W/kg
     SAR for nominal Body TSL parameters                     normalized to 1W         36.8 W/kg + 17.0 %(k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                condition

      SAR measured                                       250 mW input power                  4.90 Wikg
      SAR for nominal Body TSL parameters                    normalized to 1W         19.7 Wikg + 16.5 % (k=2)



Certificate No: D1750V2—1023_Jun18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)
Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           51.0 Q — 0.5 jQ

      Return Loss                                                                       — 39.1 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           46.0 Q + 0.3 jQ

      Return Loss                                                                       —27.5 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.217 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                August 20, 2009




Certificate No: D1750V2—1023_Jun18                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                    Date: 11.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN:1023

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: £ = 1750 MHz; 0 = 1.36 S/m; &, = 39; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.5, 8.5, 8.5) @ 1750 MHz; Calibrated: 30.12.2017

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 Sn601; Calibrated: 26.10.2017

       Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

       DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 106.5 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 16.5 W/kg
SAR(I g) = 9.1 W/kg; SAR(10 g) = 4.82 W/kg
Maximum value of SAR (measured) = 14.0 W/kg




       —3.00


       —6.00


       —9.00


       —12.00


        —15.00
                  0 dB = 14.0 W/kg = 11.46 dBW/kg



Certificate No: D1750V2—1023_Jun18            Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                      11 Jun 2018    @8:41:52
         (CHI) sii     i U Fs          3t 5S1.016 a   —457.03 me 198.99 pF          1 750.000 000 MHz
                                                      ——1—_



         Del


         Ca




         Aw
         169


         H1d


          CH2




          Ca




          Ay
          16 *


          H1d

                     START 1 550.000 000 MHz                                 STOP 1 950.000 000 MHz




Certificate No: D1750V2—1023_Jun18                      Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 11.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1750 MHz; Type: D1750V2; Serial: D1750V2 — SN:1023

Communication System: UID 0 — CW; Frequency: 1750 MHz
Medium parameters used: f = 1750 MHz; 0 = 1.47 S/m; &, = 53.6; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(8.35, 8.35, 8.35) @ 1750 MHz; Calibrated: 30.12.2017

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 Sn601; Calibrated: 26.10.2017

       Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

       DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 102.3 V/m; Power Drift =—0.05 dB
Peak SAR (extrapolated) = 15.8 W/kg
SAR(I g) = 9.12 W/kg; SAR(10 g) = 4.9 W/kg
Maximum value of SAR (measured) = 13.5 W/kg


       dB
       0


       —3.20


       —6.40


       —9.60


       —12.80


        —16.00
                   0 dB = 13.5 W/kg = 11.30 dBW/kg




Certificate No: D1750V2—1023_Jun18             Page 7 of 8


Impedance Measurement Plot for Body TSL




                                                                    11 Jun 2018     @8:40:54
         CHI) sii     k s              3: 45.982 0   ©.3281 0   25.041 pH         1 750.000 000 MHz
                                                     _t*
                                               <        \           >
         %                                             [Kei~‘%
                                                                    7
         Del




          Av
          163


          H1d


          CH2




          Ca




          Av
          163



          H1d

                    START 1 $559.000 000 MHz                                STOP 1 950.000 000 MHz




Certificate No: D1750V2—1023_Jun18                    Page 8 of 8


Calibration
      .     Laboratory of                                        R\ L\ $/m
                                                                       /7                                  g       Schweizerischer Kalibrierdienst
                                                                 &         /,
Schmid & Partner                                                 m                                         c       Service suisse d‘étalonnage
  Engineering AG                                                 3 /—-/__—_\\ y                                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                      &7/fi\‘>3                                 S       swiss Calibration Service
                                                                     tA as®



Accredited by the Swiss Accreditation Service (SAS)                                                        Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      B.V. ADT (Auden)                                                                   Certificate No: D1900V2—50036_Jan18

[CALIBRATION CERTIFICATE                                                                                                                                   l
 Object                               D1900V2 — SN:5d036


 Calibration procedure(s)             QA CAL—O5.v9
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    January 18, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (Sl).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN:    104778             04—Apr—17 (No. 217—02521/02522)                       Apr—18
 Power sensor NRP—Z91                   SN:    103244             04—Apr—17 (No. 217—02521)                             Apr—18
 Power sensor NRP—Z91                   SN:    103245             04—Apr—17 (No. 217—02522)                             Apr—18
 Reference 20 dB Attenuator             SN:    5058 (20k)         07—Apr—17 (No. 217—02528)                             Apr—18
 Type—N mismatch combination            SN:    5047.2 / 06327     07—Apr—17 (No. 217—02529)                             Apr—18
 Reference Probe EX3DV4                 SN:    7349               30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN:    601                26—Oct—17 (No. DAE4—601_Oct17)                        Oct—18

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN:    GB37480704         07—Oct—15 (in   house check Oct—16)                   In       house   check:   Oct—18
 Power sensor HP 8481A                  SN:    US37292783         07—Oct—15 (in   house check Oct—16)                   In       house   check:   Oct—18
 Power sensor HP 8481A                  SN:    MY41092317         07—Oct—15 (in   house check Oct—16)                   In       house   check:   Oct—18
 RF generator R&S SMT—O6                SN:    100972             15—Jun—15 (in   house check Oct—16)                   In       house   check:   Oct—18
 Network Analyzer HP 8753E              SN:    US37390585         18—Oct—01 (in   house check Oct—17)                   In       house   check:   Oct—18


                                        Name                                   Function                                  Signature
 Calibrated by:                        Jeton Kastrati                          Laboratory Technician           :!            J



 Approved by:                           Katja Pokovic                          Technical Manager                   /C/?g/


                                                                                                                        Issued: January 18, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.


Certificate No: D1900V2—50036_Jan18                                   Page 1 of 8


Calibration Laboratory of                                     ulJn/,H
                                                             SS                 g   Schweizerischer Kalibrierdienst
         x                                                  ind       2
Schmid & Partner                                            m                   c   Service suisse d‘étalonnage
  Engineering AG                                            o mrmmomal 3            Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 {'4,/ ;/"\ \\\\3   S   Swiss Calibration Service
                                                                 amms



Accredited by the Swiss Accreditation Service (SAS)                             Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
Convr                        sensitivity in TSL / NORM x.y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
   a) IEEE Std 1528—2013, "EEE Recommended Practice for Determining the Peak Spatial—
      Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
             Communications Devices: Measurement Techniques", June 2013
    b)       EC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
             (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
             300 MHz to 6 GHz)", July 2016
    c)       EC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
             communication devices used in close proximity to the human body (frequency range of 30
             MHz to 6 GHz)", March 2010
    d)       KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e)       DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
    &        Measurement Conditions: Further details are available from the Validation Report at the end
             of the certificate. All figures stated in the certificate are valid at the frequency indicated.
             Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
             point exactly below the center marking of the flat phantom section, with the arms oriented
             paralle! to the body axis.
             Feed Point Impedance and Return Loss: These parameters are measured with the dipole
             positioned under the liquid filled phantom. The impedance stated is transformed from the
             measurement at the SMA connector to the feed point. The Return Loss ensures low
             reflected power. No uncertainty required.
             Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
             No uncertainty required.
             SAR measured: SAR measured at the stated antenna input power.
             SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
             connector.
             SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
             nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D1900V2—50036_Jan18                              Page 2 of 8


Measurement Conditions
    DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                              V52.10.0
      Extrapolation                                   Advanced Extrapolation
      Phantom                                         Modular Flat Phantom
      Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                         1900 MHz + 1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                  22.0 °C           40.0                1.40 mho/m
      Measured Head TSL parameters                              (22.0 + 0.2) °C   40.4 £ 6 %          1.39 mho/m + 6 %
      Head TSL temperature change during test                      <0.5 °C

SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition

      SAR measured                                        250 mW input power                     10.1 Wi/kg
      SAR for nominal Head TSL parameters                    normalized to 1W          40.7 Wikg x 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Head TSL                 condition

      SAR measured                                        250 mW input power                     5.28 Wi/kg
      SAR for nominal Head TSL parameters                    normalized to 1W          21.2 Wikg + 16.5 % (k=2)


Body TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Body TSL parameters                                  22.0 °C            53.3               1.52 mho/m
      Measured Body TSL parameters                              (22.0 + 0.2) °C   54.8 +6 %           1.46 mho/im +6 %
      Body TSL temperature change during test                      <0.5 °C

SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                   Condition
      SAR measured                                        250 mW input power                     9.76 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W          40.2 Wi/kg + 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition

      SAR measured                                        250 mW input power                     5.21 W/kg
      SAR for nominal Body TSL parameters                    normalized to 1W          21.2 Wikg * 16.5 % (k=2)



Certificate No: D1900V2—50036_Jan18                    Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           51.3 Q + 5.2 jQ

      Return Loss                                                                       — 25.6 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           47.7 © 4+ 5.9 jQ
      Return Loss                                                                      — 23.8 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                  1.195 ns

After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.                                                                      *



Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                 May 08, 2003




Certificate No: D1900V2—5d036_Jan18                       Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                  Date: 18.01.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:50036

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: f = 1900 MHz; 0 = 1.39 S/m; &, = 40.4; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.18, 8.18, 8.18); Calibrated: 30.12.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   e   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 109.4 V/m; Power Drift = —0.07 dB
Peak SAR (extrapolated) = 18.4 W/kg
SAR(1 g) = 10.1 W/kg; SAR(10 g) = 5.28 W/kg
Maximum value of SAR (measured) = 15.2 W/kg




       —3.00


       —6.00


        —9.00


        —12.00


        —15.00
                   0 dB = 15.2 W/kg = 11.82 dBW/kg




Certificate No: D1900V2—5d036_Jan18            Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                       18 Jan 2018    10:22:152
          CHI S11i     1 U FS          31 51.295 a    51582 &     432.08 pH          1 900.000 000 MHz
                                                 .-"-—-‘-'—T_—-'_--'




          De l

          C&




          Ay
          169


          H1d

          CH2




          CA




          Ay
          16 9



          H1d

                     START 1 700.000 000   MHz                                STOP 2 100.000 000 MHz




Certificate No: D1900V2—5d036_Jan18                      Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                  Date: 17.01.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT : Dipole 1900 MHz; Type: D1900V2; Serial: D1900V2 — SN:504036

Communication System: UID 0 — CW; Frequency: 1900 MHz
Medium parameters used: £f = 1900 MHz; 0 = 1.46 S/m; &s, = 54.8; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.15, 8.15, 8.15); Calibrated: 30.12.2017;

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAFE4 Sn601; Calibrated: 26.10.2017

   e   Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

   e   DASY52 52.10.0(1446); SEMCAD X 14.6.10(7417)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 103.6 V/m; Power Drift = —0.05 dB
Peak SAR (extrapolated) = 17.4 W/kg
SAR(1 g) = 9.76 W/kg; SAR(10 g) = 5.21 W/kg
Maximum value of SAR (measured) = 14.3 W/kg




       —3.00


       —6.00


       —9.00


       —12.00


       —15.00
                  0 dB = 14.3 W/kg = 11.55 dBW/kg




Certificate No: D1900V2—5d036_Jan18             Page 7 of 8


Impedance Measurement Plot for Body TSL—




                                                                      17 Jan 2018     16:26:31
         (CHZ S11     1 U FS          4:47.742 2        5.9297 a   496.70 pH         1 900.000 9090 MHz
                                                   a~   —~~—IT——

         Del

         CA




         Av
         16 C


         H1d

          CH2




         Ca




         Ay
         16°

         H1d

                    START 1 700.000 000 MHz                                    STOP 2 100.000 oop MHz




Certificate No: D1900V2—5d036_Jan18                       Page 8 of 8


   .    s                                                             9§409n
Calibration
      .     Laboratory of                                         §“\“\Q\__/;/§x                           S Schweizerischer Kalibrierdienst
Schmid & Partner                                                 SE%E//N&                                  c       Service suisse d‘étalonnage
  Engineering AG                                                 ;///—\\\3                                         Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                       */1//\\\\\\«                            s       Swiss Calibration Service
                                                                     "hilal?


Accredited by the Swiss Accreditation Service (SAS)                                                         Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Client      Auden                                                                               Certificate No: D2450V2—869_Jun18

|CALIBRATION CERTIFICATE
 Object                               D2450V2 — SN:869


 Calibration procedure(s)             QA CAL—O05.v10
                                      Calibration procedure for dipole validation kits above 700 MHz



 Calibration date:                    June 19, 2018


 This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (S1).
 The measurements and the uncertainties with confidence probability are given on the following pages and are part of the certificate.


 All calibrations have been conducted in the closed laboratory facility: environment temperature (22 + 3)°C and humidity < 70%.


 Calibration Equipment used (M&TE critical for calibration)


 Primary Standards                      ID #                      Cal Date (Certificate No.)                            Scheduled Calibration
 Power meter NRP                        SN: 104778                04—Apr—18 (No. 217—02672/02673)                       Apr—19
 Power sensor NRP—Z91                   SN: 108244                04—Apr—18 (No. 217—02672)                             Apr—19
 Power sensor NRP—Z91                   SN: 103245                04—Apr—18 (No. 217—02673)                             Apr—19
 Reference 20 dB Attenuator             SN:    5058 (20k)         04—Apr—18 (No. 217—02682)                             Apr—19
 Type—N mismatch combination            SN:    5047.2 / 06327     04—Apr—18 (No. 217—02683)                             Apr—19
 Reference Probe EX3DV4                 SN:    7349               30—Dec—17 (No. EX3—7349_Dec17)                        Dec—18
 DAE4                                   SN:    601                26—0ct—17 (No. DAE4—601_Oct17)                        Oct—18

 Secondary Standards                    ID #                      Check Date (in house)                                 Scheduled Check
 Power meter EPM—442A                   SN: GB37480704            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: US37202783            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 Power sensor HP 8481A                  SN: MY41092317            07—Oct—15 (in house check Oct—16)                     In house check: Oct—18
 RF generator R&S SMT—06                SN: 100972                15—Jun—15 (in house check Oct—16)                     In house check: Oct—18
 Network Analyzer HP 8753E              SN: US37390585            18—Oct—01 (in house check Oct—17)                     In house check: Oct—18


                                        Name                                   Function                                  Signature
 Calibrated by:                         Manu Seitz                             Laboratory Technician



 Approved by:                           Katja Pokovic                          Technical Manager                   ///%F



                                                                                                                        Issued: June 19, 2018
 This calibration certificate shall not be reproduced except in full without written approval of the laboratory.



Certificate No: D2450V2—869_Jun18                                     Page 1 of 8


Calibration Laboratory of                                     Cn
                                                             ¢\§.///             S   Schweizerischer Kalibrierdienst
          &                                                  &            2
Schmid & Partner                                            ila\ez_/m            c   Service suisse d‘étalonnage
  Engineering AG                                            es st                    Servizio svizzero di taratura
Zeughausstrasse 43, 8004 Zurich, Switzerland                 {«/,,/7\\\\\\\3     S   swiss Calibration Service
                                                                 Aitabs

Accredited by the Swiss Accreditation Service (SAS)                               Accreditation No.: SCS 0108
The Swiss Accreditation Service is one of the signatories to the EA
Multilateral Agreement for the recognition of calibration certificates

Glossary:
TSL                          tissue simulating liquid
ConvF                        sensitivity in TSL / NORM x,y,z
N/A                          not applicable or not measured

Calibration is Performed According to the Following Standards:
    a) IEEE Std 1528—2013, "IEEE Recommended Practice for Determining the Peak Spatial—
       Averaged Specific Absorption Rate (SAR) in the Human Head from Wireless
       Communications Devices: Measurement Techniques", June 2013
    b) IEC 62209—1, "Measurement procedure for the assessment of Specific Absorption Rate
              (SAR) from hand—held and body—mounted devices used next to the ear (frequency range of
       300 MHz to 6 GHz)", July 2016
    c) IEC 62209—2, "Procedure to determine the Specific Absorption Rate (SAR) for wireless
       communication devices used in close proximity to the human body (frequency range of 30
       MHz to 6 GHz)", March 2010
    d) KDB 865664, "SAR Measurement Requirements for 100 MHz to 6 GHz"

Additional Documentation:
    e) DASY4/5 System Handbook

Methods Applied and Interpretation of Parameters:
      e       Measurement Conditions: Further details are available from the Validation Report at the end
              of the certificate. All figures stated in the certificate are valid at the frequency indicated.
      e       Antenna Parameters with TSL: The dipole is mounted with the spacer to position its feed
              point exactly below the center marking of the flat phantom section, with the arms oriented
              parallel to the body axis.
      e       Feed Point Impedance and Return Loss: These parameters are measured with the dipole
              positioned under the liquid filled phantom. The impedance stated is transformed from the
              measurement at the SMA connector to the feed point. The Return Loss ensures low
              reflected power. No uncertainty required.
      e       Electrical Delay: One—way delay between the SMA connector and the antenna feed point.
              No uncertainty required.
      e       SAR measured: SAR measured at the stated antenna input power.
      e       SAR normalized: SAR as measured, normalized to an input power of 1 W at the antenna
              connector.
      e       SAR for nominal TSL parameters: The measured TSL parameters are used to calculate the
              nominal SAR result.
  The reported uncertainty of measurement is stated as the standard uncertainty of measurement
  multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage
  probability of approximately 95%.




Certificate No: D2450V2—869_Jun18                                Page 2 of 8


Measurement Conditions
   DASY system configuration, as far as not given on page 1.
      DASY Version                                              DASY5                              V52.10.1
      Extrapolation                                   Advanced Extrapolation
      Phantom                                          Modular Flat Phantom
      Distance Dipole Center — TSL                              10 mm                             with Spacer
     Zoom Scan Resolution                               dx, dy, dz =5 mm
      Frequency                                         2450 MHz +1 MHz



Head TSL parameters
   The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Head TSL parameters                                  22.0 °C           39.2                1.80 mho/m

      Measured Head TSL parameters                              (22.0 + 0.2) °C   38.0 + 6 %          1.87 mho/m + 6 %
      Head TSL temperature change during test                      <0.5 °C            ~——>                    ~~——

SAR result with Head TSL

      SAR averaged over 1 cm* (1 g) of Head TSL                  Condition
      SAR measured                                        250 mW input power                     13.4 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W        52.3 Wikg + 17.0 % (k=2)

      SAR averaged over 10 cm‘ (10 g) of Head TSL                 condition
      SAR measured                                        250 mW input power                     6.28 Wikg
      SAR for nominal Head TSL parameters                     normalized to 1W        24.8 Wikg # 16.5 % (k=2)


Body TSL parameters
    The following parameters and calculations were applied.
                                                                Temperature       Permittivity          Conductivity
      Nominal Body TSL parameters                                  22.0 °C           52.7                1.95 mho/m
      Measured Body TSL parameters                              (22.0 + 0.2) °C   52.3 + 6 %          2.03 mho/m + 6 %
      Body TSL temperature change during test                      «0.5 °C            ————                    ————


SAR result with Body TSL

      SAR averaged over 1 cm* (1 g) of Body TSL                  Condition
      SAR measured                                        250 mW input power                     12.9 Wi/kg
      SAR for nominal Body TSL parameters                     normalized to 1W         50.5 Wikg * 17.0 % (k=2)

      SAR averaged over 10 cm* (10 g) of Body TSL                 condition
      SAR measured                                        250 mW input power                 6.07 Wikg
      SAR for nominal Body TSL parameters                     normalized to 1W        24.0 Wikg + 16.5 %(k=2)



Certificate No: D2450V2—869_Jun18                      Page 3 of 8


Appendix (Additional assessments outside the scope of SCS 0108)

Antenna Parameters with Head TSL

      Impedance, transformed to feed point                                           52.4 Q + 7.4 jQ

      Return Loss                                                                       —22.4 dB



Antenna Parameters with Body TSL

      Impedance, transformed to feed point                                           49.0 Q + 7.7 jQ

      Return Loss                                                                       —22.1 dB




General Antenna Parameters and Design

      Electrical Delay (one direction)                                                   1.159 ns


After long term use with 100W radiated power, only a slight warming of the dipole near the feedpoint can be measured.

The dipole is made of standard semirigid coaxial cable. The center conductor of the feeding line is directly connected to the
second arm of the dipole. The antenna is therefore short—circuited for DC—signals. On some of the dipoles, small end caps
are added to the dipole arms in order to improve matching when loaded according to the position as explained in the
"Measurement Conditions" paragraph. The SAR data are not affected by this change. The overall dipole length is still
according to the Standard.
No excessive force must be applied to the dipole arms, because they might bend or the soldered connections near the
feedpoint may be damaged.


Additional EUT Data

      Manufactured by                                                                    SPEAG
      Manufactured on                                                                August 18, 2010




Certificate No: D2450V2—869_Jun18                        Page 4 of 8


DASY5 Validation Report for Head TSL

                                                                                   Date: 13.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:869

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: f = 2450 MHz; 0 = 1.87 S/m; &, = 38; p = 1000 kg/m‘
Phantom section: Flat Section
Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19—2011)

DASY52 Configuration:

   *   Probe: EX3DV4 — SN7349; ConvF(7.88, 7.88, 7.88) @ 2450 MHz; Calibrated: 30.12.2017

   e   Sensor—Surface: 1.4mm (Mechanical Surface Detection)

   e   Electronics: DAE4 Sn601; Calibrated: 26.10.2017

   &   Phantom: Flat Phantom 5.0 (front); Type: QD 000 P50 AA; Serial: 1001

   &   DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Head Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 115.1 V/m; Power Drift =—0.03 dB
Peak SAR (extrapolated) = 26.3 W/kg
SAR(I g) = 13.4 W/kg; SAR(10 g) = 6.28 W/kg
Maximum value of SAR (measured) = 22.0 W/kg




                   0 dB = 22.0 W/kg = 13.42 dBW/kg




Certificate No: D2450V2—869_Jun18              Page 5 of 8


Impedance Measurement Plot for Head TSL




                                                                13 Jun 2018  12:07:39
          (CHI sii     1 U FS          1: 52441 a   7.3555 0 477.92 pH      2450.000 000 MHz
                                                    ~——




          Del

          Ca




          Av
          159


          H1d


          CH2




          Ca




          Av
          16°

          H1d

                     START 2 250.000 000 MHz                         STOP 2 650.000 000 MHz




Certificate No: D2450V2—869_Jun18                   Page 6 of 8


DASY5 Validation Report for Body TSL

                                                                                   Date: 19.06.2018

Test Laboratory: SPEAG, Zurich, Switzerland

DUT: Dipole 2450 MHz; Type: D2450V2; Serial: D2450V2 — SN:869

Communication System: UID 0 — CW; Frequency: 2450 MHz
Medium parameters used: £ = 2450 MHz; 0 = 2.03 S/m; & = 52.3; p = 1000 kg/m3
Phantom section: Flat Section
Measurement Standard: DASYS5 (IEEE/IEC/ANSI €63.19—2011)

DASY52 Configuration:

   e   Probe: EX3DV4 — SN7349; ConvF(8.01, 8.01, 8.01) @ 2450 MHz; Calibrated: 30.12.2017

       Sensor—Surface: 1.4mm (Mechanical Surface Detection)

       Electronics: DAE4 S$Sn601; Calibrated: 26.10.2017

       Phantom: Flat Phantom 5.0 (back); Type: QD 000 P50 AA; Serial: 1002

       DASY52 52.10.1(1476); SEMCAD X 14.6.11(7439)


Dipole Calibration for Body Tissue/Pin=250 mW, d=10mm/Zoom Scan (7x7x7)/Cube 0:
Measurement grid: dx=5mm, dy=5mm, dz=5mm
Reference Value = 107.4 V/m; Power Drift = —0.09 dB
Peak SAR (extrapolated) = 25.3 W/kg
SAR(I g) = 12.9 W/kg; SAR(10 g) = 6.07 W/kg
Maximum value of SAR (measured) = 20.8 W/kg


       dB
        0


        —5.00


        —10.00


        —15.00


        —20.00


        —25.00
                   0 dB = 20.8 W/kg = 13.18 dBW/kg




Certificate No: D2450V2—869_Jun18               Page 7 of 8



Document Created: 2018-10-23 14:43:53
Document Modified: 2018-10-23 14:43:53

© 2024 FCC.report
This site is not affiliated with or endorsed by the FCC