FCC ID: H8NWLF010

RF Exposure Info

Download: PDF
FCCID_470919

D3: DAE


Calibration Laboratory of
Schmid & Partner
  Engineering AG
Zeughausstrasse 43, 8004 Zurich, Switzerland




Client


 CALIBRATION CERTIFICATE                                                   mmmmmmmmemmmem


 Object(s)                           DAE3 — SN:579

 Calibration procedure(s)            QA CAL—O6.v3
                                     Calibration procedure for the data acquisition unit (DAE)



 Calibration date:                   August 15, 2003

 Condition of the calibrated item    In Tolerance (according to the specific calibration document)


 This calibration statement documents traceability of M&TE used in the calibration procedures and conformity of the procedures with the ISO/IEC
 17025 international standard.


 All calibrations have been conducted in the closed laboratory facility: environmenttemperature 22 +/— 2 degrees Celsius and humidity <75%.


 Calibration Equipment used (M&TE critical for calibration)


 Model Type                           ID #                                  Cal Date                              Scheduled Calibration
 Fluke Process Calibrator Type 702    SN: 6295803                           3—Sep—01                              Sep—03




                                      Name                                  Function                              Signature

 Calibrated by:                       Philipp Storchenegger                 Technician                   7:}w/{L ,L/‘

 Approved by:                         Fin Bomholt                           R&D Director             ”7? /{W
                                                                      he                             A    C   l        P
                                                                                                                  Date issued: August 15, 2003




 This calibration certificate is issued as an intermediate solution until the accreditation process (based on ISO/IEC 17025 International Standard)
 for Calibration Laboratory of Schmid & Partner Engineering AG is completed.


DAE 3 SN: 579                                                                 DATE: 15.08.2003

1. Cal Lab. Incoming Inspection & Pre Test

    Modification Status                Note Status here > > > >                    BC
    Visual Inspection                  Note anomalies......................       None

     Pre Test                          Indication                               Yes/No
     Probe Touch                       Function                                   Yes
     Probe Collision                   Function                                   Yes
     Probe Touch&Collision             Function                                   Yes


2. DC Voltage Measurement

   A/D — Converter Resolution nominal
      High Range:      1LSB =    6. 1uV ,              full range =      400 mV
      Low Range:          1LSB=         61inV ,        full range =        4 mV

   DASY measurement parameters: Auto Zero Time: 3 sec; Measuring time: 3 sec

     Calibration Factors                       X                  Y                 Z
     High Range rounded to 7 digits    |_404.5370401__|_404.5593911__|        404.3923437
     Low Range rounded to 6 digits         3.9686              3.9584             3.95
     Connector Angle to be used       in DASY System             311°


    High Range                             Input         Reading in V          % Error
    Channel X + Input                     200mV            199999.6             0.00
                                           20mV             19998.2             —0.01
    Channel X       — Input                 20mV             —19995.3           —0.02
    Channel Y       + Input               200mV             199999.8            0.00
                                           20mV              19998.3            —0.01
    Channel Y       — Input                20mV             —19993.6            —0.03
    Channel Z       + Input               200mV             200000.6              0.00
                                            20mV             19997.8            —0.01
    Channel Z       — Input                 20mV             —19994.3           —0.03

    Low Range                              Input         Reading in uV         % Error
    Channel X      + Input                  2mV              1999.99              0.00
                                          0.2mV               199.66            —0.17
    Channel X      — Input                0.2mV              —200.21              0.11
    Channel Y      + Input                  2mV              1999.89            —0.01
                                          0.2mV               199.20            —0.40
    Channel Y      — Input                0.2mV              —201.14              0.57
    Channel Z      + Input                 2mV               1999.99            0.00
                                          0.2mV               199.18            —0.41
    Channel Z      — Input                0.2mV              —202.26              113
                                                                                     Page 2 of 4



Document Created: 2004-09-13 19:16:20
Document Modified: 2004-09-13 19:16:20