DTS (BLE) Report

FCC ID: F6OINBODYBAND2

Test Report

Download: PDF
FCCID_3234526

                                                                Page 1 of 31                             Report No. : W16DR-D001




        ELECTROMAGNETIC EMISSION
           COMPLIANCE REPORT
 FOR LOW-POWER, NON-LICENSED TRANSMITTER

      Test Report No.                 : W16DR-D001

      AGR No.                         : A16NA-275

      Applicant                       : InBody Co., Ltd.

      Address                         : InBody Bldg., 54, Nonhyeon-ro 2-gil, Gangnam-gu, Seoul, 135-960, Korea

      Manufacturer                    : InBody Co., Ltd.

      Address                         : 15, Heugam-gil, Ipjang-myeon, Seobuk-gu, Cheonan-si, Chungcheongnam-do, 31025,
                                       KOREA

      Type of Equipment               : Body Composition Analyzer

      FCC ID.                         : F6OINBODYBAND2

      Model Name                      : InBodyBAND2

      Serial number                   : N/A

      Total page of Report            : 31 pages (including this page)

      Date of Incoming                : November 24, 2016

      Date of issue                   : December 02, 2016



     SUMMARY
      The equipment complies with the regulation; FCC PART 15 SUBPART C Section 15.247
      This test report only contains the result of a single test of the sample supplied for the examination.
      It is not a generally valid assessment of the features of the respective products of the mass-production.




      Reviewed by:                                                            Approved by:
                          Ki-Hong, Nam / Asst, Chief Engineer                                  Keun-Young, Choi / Vice President
                          ONETECH Corp.                                                        ONETECH Corp.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


oneETECH
         Ns                                                                 Page 2 of 31                                             Report No.: W16DR—D001

                                                                             CONTENTS

                                                                                                                                                                         PAGE

1. VERIFICATION OF COMPLIANCE                                                                                                                                                    5

2. TEST SUMMARY                                                                                                                                                                  6

  2.1 TEST ITEMS AND RESULTS o. 22.00.00.00. 00000000000 000en0 en en ennnnennennnnennennennennennenennnnennnnnennennennnnennnnnnennnnennnennnnnnennnnenennennea 6

  2.2 ADDITIONS, DEVIATIONS, EXCLUSIONS FROM STANDARDS:.........22.00.00.00.0000000000000en0n0ennennennnneenennnare ennrennennennennenennennen 6

  2.3 RELATED SUBMITTAL(S) / GRANT(S) .. 22.00.0020.000.0000000000000000n0en0ennnenne en nnenennennenne en narennenne en nne en en en en en narennennennennenennennea 6

  2.4 PURPOSE OF THE TEST ..
  2.5 TEST METHODOLOGY...

  PA c ol oce18in e 6

3. GENERAL INFORMATION                                                                                                                                                           7

   3.1 PRODUCT DESCRIPTION....                                                                                                                                               2227

   3.2 ALTERNATIVE TYPE(S)/MODEL(S); ALSO COVERED BY THIS TEST REPORT...                                                                                                        .7

4. EUT MODIFICATIONS                                                                                                                                                             7

5. SYSTEM TEST CONFIGURATION                                                                                                                                                     8

   5.1 JJUSTIFICATION .. 22. 00.0000 0000000000 n0 en en enr en n en en en en en en en en nnenne en en en nnenennennennennnnennennnnennnnennennenne en snn nnenneenenennennenne 8

   5.2 PERIPHERAL EQUIPMENT a.2.22.22.00.0.00.00.000000000000en0en0en0 en en nnenenennennennennenne ennenneannarennennenneaneaneaneaneannnnnenneaneaneanenennennenne 8

   5.3 MODE OF OPERATION DURING THE TEST ..........00.00.00.0000000000e00 en en nennennennnenneenenenenne en en en en nne en en nnrennennennenneenenennennenne 8

   5.4 CONFIGURATION OF TEST SYSTEM ... 22.00.00.000.00000000000n0en0ennennnnernnr en en en en nnnr en en nnnenne en en en en en en en en nnsnnennenneenenennennenne 8

6. PRELIMINARY TEST                                                                                                                                                              9

   6.1 AC POWER LINE CONDUCTED EMISSIONS TESTS.........02.00.00.00000000000e0en0enennenennareenrenrenrenrenrenrenrenrenennrennenneeneenenennennenne 9

   6.2 GENERAL RADIATED EMISSIONS TESTS ........02.00.00.00.00000000000eneeenenenr en nenrenrenennerevrenvenranreneenreneenennrenneeneeneeneenenennennenne 9

7. MINIMUM 6 DB BANDWIDTH                                                                                                                                                      10

  7.1 OPERATING ENVIRONMENT ... 22.00.0. 00000000000 00000en0 en en ennnne en enennenne en en en anenennenneannnnnennennennennnnnnennnnnnnnennennennennenennena 10
  7.2 TEST SETAUP a.2, .000 n0 en ennenne en en nnen en en en nnnnne en en en en en nne en nnennnennennnnennenennennennennennnennennennennennnnennennennennnennennennennenennena 10

  7.3 TEST EQUIPMENT USED ..2.22.00.00.00.000 000000000000 en 00e en nn n nnnarennennennennennenennennennenarennennenneaneanannneanennnnennennennennennenennena 10

  7.A TEST DATAG... .000 n nenenerenneenrenrenranenenr en en narenne en en en en en en en en en nne en nnnnenennennennennennnennennennennnnnnennennennennnennennennennenennena 11

8. MAXIMUM PEAK OUTPUT POWER                                                                                                                                                   13

   8.1 OPERATING ENVIRONMENT ... 22.00.0.00.00.00000000000en0en0ennenene en en narenne en en en anenennennennenarennenneane en nnnnennnnennennrennennnnnnenennenne 13

   8.2 TEST SETAUP 22220000000 n0 en ene en en nnr en en nare en en en en en en en en nnnnenne en en nne en nne en nn en nnennennnnnnennnnennnnennennennennenennenne 13

   8.3 TEST EQUIPMENT USED ... 22.00.00.00.00.00 00000000000 0000000 en0 en n nnennennennenne n e ennennennennrennenneanennennennennennenarennennennennennenennennea 13

   8.4 TEST DATA....

9. 100 KHZ BANDWIDTH OUTSIDE THE FREQUENCY BAND                                                                                                                                16

   9.1 OPERATING ENVIRONMENT .......2.. 0000000000000 en en en eneenvenvenvenmeeneeneennennennanennennennenneennennnnennennennennennennenneennennennennencnnennen 16

Tt should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                                                     EMC—003 (Rev.2)
ONETECH Corp.: 43—14, Jinsaegol—gil, Chowol—eup, Gwangiu—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799.9599)


oneETECH
         Ns                                                                     Page 3 of 31                                              Report No.: W16DR—D001

   9.2 TEST SET—UP FOR CONDUCTED MEASUREMENT ... 2..00.00.00.00.00000000000000en0en0eneneeneenenarenne en en en en en en en en narennennennennenennenne 16
   9.3 TEST SET—UP FOR RADIATED MEASUREMENT....0..00.00.00.00.00000000000000en0en0ennenenneneenenarenne en en en en en en en narennennennnnennenennenne 16

   9.4 TEST EQUIPMENT USED .                                                                                                                                                         .16

   9.5 TEST DATA FOR CONDUCTED EMISSION .........00.00.00.0000000000n0en0 en ennenenne en en en nnennennnennenne en en en en en en en revvnnennennenennenne 17

   9.6 TEST DATA FOR RADIATED EMISSION ... .2.02.000.00.00.0000000000000en0ennenarenneene en en nnne en en enennnenne en en enenne en nnennennennnennennennenennenne 22
      9.6.1 Radiated Emission which fall in the ReSLriCL@d BOMA.......................clccccivcerreerrenrensensensensenrenneennarenrenrenrnennennen 22

      9.6.2 SPUTiOUS & HQPMORIG RQditted EMAISSIOM .................cscouncsensensensensenenrrnnenrennrrnsennennennennennennennennenerennennennennenennvnnena 23

10. PEAK POWER SPECTRAL DENSITY                                                                                                                                                       24

   10.1 OPERATING ENVIRONMENT ...2..22.0..00.00000000000en en en enmenmenmenmenmenmeeneennennennennnnnnennennennnennennennennennennennennennennnennennennennancnncnne 24
   10.2 TEST SET—DP a.22.00.00 000000 n0 en en nnennenennr en en enenenenne en en en en en en en narenne en en en nnnnenne en en nnrennnnennennennenn en nnennenarenne en en nnennnenne 24

   10.3 TEST EQUIPMENT USED n.2.22.00.00.00 0000000000000 000000 en ennennnnnnnnnnennenne en en en en en en nerennenne en en en en en en nerenne en en en anenennenne 24
   10.4 TEST DATA :0000 0e n en ennnneenrenennnr en en eneneneene en en en en en en en nnnenn en en en nnnnenne en en nnrennnne en en nnenn en nnennennrenne en en nnennnenne 25

11. RADIATED EMISSION TEST                                                                                                                                                            27

   11L.1 OPERATING ENVIRONMENT 12. 22. 00.00.00.00.000000000000000en0 en ennnneeneenenenenne en en en nnenennenneenennnennane en en en en en nnanennnennennennennennnenne 27
   1L.2 TEST SETADP 220. 0000000 e cce ennnnennennenennr en en enenenenne en en en en en en en narenne en en en nnnnenne en en nnrennnnennennennenn en nnennennrennennennennennnenne 27

   11.3 TEST EQUIPMENT USED n.2.22.00.00.0000.000000000000 0000000 n0 en en ennnn en ennennnnennnnenn en nnennennrennennenneanenne en anennenerennenneaneanennenennenne 27

   11.4 TEST DATA FOR TRANSMITTING MODE ..2.222.00.00.00.0000000000000000eneoeeneeneenrenenrenrenennrennrenrenreneenrenreneeneenennnennenneeneenennenennenne 28

      11.4.1 Test datafor 30 MHz ~ 1 GHz ....                                                                                                                                      ...28
      11.4.2 TeSt dGL@fOP BOIOW 30 MHZ ............c.ccsucsensenvensensensensrnrenerrnrrnnrenrnnenennennrnnrnvenerennennennennennennvnnennennenerennennennenennvnneia 31

      11.4.3 TeSt d@L@fOP @BOVG 1 GHZ ............c.coulccnsenvensensensensrnnensrrnsennrnnrnnennenennennennennennrennennennennennennvnnennennenerennennennenennvnneia 31




Tt should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                                                           EMC—003 (Rev.2)
ONETECH Corp.: 43—14, Jinsaegol—gil, Chowol—eup, Gwangiu—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799.9599)


                                                                 Page 4 of 31                            Report No. : W16DR-D001




                                                             Revision History

Issued Report No.      Issued Date               Revisions                                                               Effect Section

  W16DR-D001            December 02, 2016        Initial Issue                                                           All




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 5 of 31                             Report No. : W16DR-D001


1. VERIFICATION OF COMPLIANCE
Applicant           : InBody Co., Ltd.
Address             : InBody Bldg., 54, Nonhyeon-ro 2-gil, Gangnam-gu, Seoul, 135-960, Korea
Contact Person : Kyung Keun, Kim / Manager
Telephone No. : +82-2-2182-8968
FCC ID              : F6OINBODYBAND2
Model Name          : InBodyBAND2


Brand Name          :
Serial Number : N/A
Date                : December 02, 2016

 EQUIPMENT CLASS                                            DTS – DIGITAL TRNSMISSION SYSTEM
 E.U.T. DESCRIPTION                                        Body Composition Analyzer
 THIS REPORT CONCERNS                                       Original Grant
 MEASUREMENT PROCEDURES                                     ANSI C63.10: 2013
 TYPE OF EQUIPMENT TESTED                                   Pre-Production
 KIND OF EQUIPMENT
                                                            Certification
 AUTHORIZATION REQUESTED
 EQUIPMENT WILL BE OPERATED
                                                            FCC PART 15 SUBPART C Section 15.247
 UNDER FCC RULES PART(S)
 Modifications on the Equipment to            Achieve
                                                            None
 Compliance
 Final Test was Conducted On                                3 m, Semi Anechoic Chamber

-. The above equipment was tested by ONETECH Corp. for compliance with the requirement set forth in the FCC Rules
   and Regulations. This said equipment in the configuration described in this report, shows the maximum emission levels
   emanating from equipment are within the compliance requirements.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 6 of 31                             Report No. : W16DR-D001


2. TEST SUMMARY

2.1 Test items and results
      SECTION                                                 TEST ITEMS                                               RESULTS

  15.247 (a) (2)           Minimum 6 dB Bandwidth                                                                 Met the Limit / PASS
  15.247 (b) (3)           Maximum Peak Conducted Output Power                                                    Met the Limit / PASS
  15.247 (d)               100 kHz Bandwidth Outside the Frequency Band                                           Met the Limit / PASS
  15.247 (d)               Radiated Emission which fall in the Restricted Band                                    Met the Limit / PASS
  15.247 (e)               Peak Power Spectral Density                                                            Met the Limit / PASS
  15.209                   Radiated Emission Limits                                                               Met the Limit / PASS
  15.207                   Conducted Limits                                                                           N/A (See Note)
  15.203                   Antenna Requirement                                                                   Met requirement / PASS
Note: This test is not performed because the EUT does not transmit and receive during charging..

2.2 Additions, deviations, exclusions from standards
No additions, deviations or exclusions have been made from standard.

2.3 Related Submittal(s) / Grant(s)
Original submittal only

2.4 Purpose of the test
To determine whether the equipment under test fulfills the requirements of the regulation stated in FCC PART 15
SUBPART C Section 15.247.

2.5 Test Methodology
Both conducted and radiated testing was performed according to the procedures in ANSI C63.10: 2013. Radiated testing
was performed at a distance of 3 m from EUT to the antenna.

2.6 Test Facility
The Onetech Corp. has been designated to perform equipment testing in compliance with ISO/IEC 17025.
The Electromagnetic compatibility measurement facilities are located at 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si,
Gyeonggi-do, 12735, Korea
-. Site Filing:
VCCI (Voluntary Control Council for Interference) – Registration No. R-4112/ C-4617/ G-10666 / T-1842
IC (Industry Canada) – Registration No. Site# 3736A-3
-. Site Accreditation:
KOLAS (Korea Laboratory Accreditation Scheme) - Accreditation NO. KT085
FCC (Federal Communications Commission) - Accreditation No. KR0013
RRA (Radio Research Agency) – Designation No. KR0013


It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 7 of 31                             Report No. : W16DR-D001


3. GENERAL INFORMATION

3.1 Product Description
The InBody Co., Ltd., Model InBodyBAND2 (referred to as the EUT in this report) is a Body Composition Analyzer. The
product specification described herein was obtained from product data sheet or user’s manual.

  Device Type                             Body Composition Analyzer
  Temperature Range                       -15 °C ~ 60 °C
  Operating Frequency                     2 402 MHz ~ 2 480 MHz
  RF Output Power                         1.42 dBm
  Number of Channel                       40 Channel
  Modulation Type                         GFSK
  Antenna Type                            Chip Antenna
                                          Marker: NORDIC SEMICONDUCTOR
  USED RF CHIP
                                          Model Name: nRF51822
  Antenna Gain                            1.99 dBi
  List of each Osc. or crystal
                                          32.768 MHz, 16 MHz
  Freq.(Freq. >= 1 MHz)

3.2 Alternative type(s)/model(s); also covered by this test report.
-. None


4. EUT MODIFICATIONS
-. None




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 8 of 31                             Report No. : W16DR-D001


5. SYSTEM TEST CONFIGURATION

5.1 Justification
This device was configured for testing in a typical way as a normal customer is supposed to be used. During the test, the
following components were installed inside of the EUT.
      DEVICE TYPE                         MANUFACTURER                             MODEL/PART NUMBER                         FCC ID

         Main Board                          InBody Co., Ltd.                                N/A                                N/A
            OLED                                    N/A                                      N/A                                N/A
            Battery                                 N/A                                      N/A                                N/A

5.2 Peripheral equipment
Defined as equipment needed for correct operation of the EUT, but not considered as tested: None

5.3 Mode of operation during the test
For the testing, software used to control the EUT for staying in continuous transmitting is programmed.
For final testing, the EUT was set at 2 402 MHz, 2 440 MHz, and 2 480 MHz to get a maximum emission levels from the
EUT. The EUT was moved throughout the XY, XZ, and YZ planes and the worst case is “XZ” axis, but the worst data was
recorded in this report.

5.4 Configuration of Test System
Line Conducted Test:                     It is not need to test this requirement, because the EUT shall be operated by DC battery.


Radiated Emission Test:                  Preliminary radiated emissions test were conducted using the procedure in ANSI C63.10:
                                         2013 to determine the worse operating conditions. Final radiated emission tests were
                                         conducted at 3 meter Semi Anechoic Chamber.
                                         The turntable was rotated through 360 degrees and the EUT was tested by positioned
                                         three orthogonal planes to obtain the highest reading on the field strength meter. Once
                                         maximum reading was determined, the search antenna was raised and lowered in both
                                         vertical and horizontal polarization.


5.5 Antenna Requirement
For intentional device, according to section 15.203, an intentional radiator shall be designed to ensure that no antenna other
than that furnished by the responsible party shall be used with the device.
Antenna Construction:
The antenna of the EUT is a Chip Antenna on the main board in the EUT, so no consideration of replacement by the user.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 9 of 31                               Report No. : W16DR-D001


6. PRELIMINARY TEST

6.1 AC Power line Conducted Emissions Tests
During Preliminary Tests, the following operating mode was investigated

                        Operation Mode                                    The Worse operating condition (Please check one only)

                    It is not need to test this requirement, because the power of the EUT is supplied by battery.

6.2 General Radiated Emissions Tests
During Preliminary Tests, the following operating modes were investigated

                        Operation Mode                                    The Worse operating condition (Please check one only)

                       Transmitting Mode                                                               X




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 10 of 31                               Report No. : W16DR-D001


7. MINIMUM 6 dB BANDWIDTH

7.1 Operating environment
Temperature                          :       22.0 °C
Relative humidity                    :       43.2 % R.H.

7.2 Test set-up
The antenna output of the EUT was connected to the spectrum analyzer. The resolution bandwidth is set to 100 kHz, and
peak detection was used. The 6 dB bandwidth is defined as the total spectrum over which the power is higher than the peak
power minus 6 dB.




                             EUT                                                            Spectrum analyzer



7.3 Test equipment used
       Model Number               Manufacturer                 Description                        Serial Number      Last Cal.
-     FSV40                      Rohde & Schwarz              Signal Analyzer                    101009             May 31, 2016 (1Y)
All test equipment used is calibrated on a regular basis.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 11 of 31                              Report No. : W16DR-D001


7.4 Test data
-. Test Date                    : November 28, 2016 ~ November 29, 2016
-. Test Result                  : Pass

 CHANNEL            FREQUENCY(MHz)                  MEASURED VALUE (kHz)                        LIMIT (kHz)           MARGIN (kHz)
      Low                   2 402.00                             695.10                              500                    195.10
    Middle                  2 440.00                             694.30                              500                    194.30
     High                   2 480.00                             694.30                              500                    194.30
Remark. Margin = Measured Value - Limit




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 12 of 31                            Report No. : W16DR-D001




                                                                 Middle Channel




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


oneETECH
        Ns                                                      Page 13 of 31                             Report No.: W16DR—D001



8. MAXIMUM PEAK QOUTPUT POWER

8.1 Operating environment
Temperature                          :      22.0 °C

Relative humidity                    :      43.2 % RH.

8.2 Test set—up
The antenna output of the EUT was connected to the spectrum analyzer.
The resolution bandwidth is set to Z2 DTS Bandwidth, the video bandwidth is set to 3 times the resolution bandwidth.




                               EUT                                                         Spectrum Analyzer




8.3 Test equipment used
       Model Number               Manufacturer                 Description                      Serial Number        Last Cal.

M —__FSV40                        Rohde & Schwarz              Signal Analyzer                  101009               May 31, 2016 (1Y)
All test equipment used is calibrated on a regular basis.




Tt should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                    EMC—003 (Rev.2)
ONETECH Corp.: 43—14, Jinsaegol—gil, Chowol—eup, Gwangiu—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799.9599)


                                                                Page 14 of 31                              Report No. : W16DR-D001


8.4 Test data
-. Test Date                    : November 28, 2016 ~ November 29, 2016
-. Test Result                  : Pass

                          FREQUENCY                   MEASURED VALUE                            LIMIT                   MARGIN
  CHANNEL
                               (MHz)                           (dBm)                            (dBm)                      (dB)
     LOW                      2 402.00                           1.42                            30.00                     28.58
   MIDDLE                     2 440.00                           1.05                            30.00                     28.95
     HIGH                     2 480.00                           0.40                            30.00                     29.60
Remark. Margin = Limit – Measured Value (=Receiver Reading + Cable Loss)




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 15 of 31                            Report No. : W16DR-D001




                                                                Middle Channel




                                                                  High Channel



It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 16 of 31                               Report No. : W16DR-D001


9. 100 kHz BANDWIDTH OUTSIDE THE FREQUENCY BAND

9.1 Operating environment
Temperature                          :       22.0 °C
Relative humidity                    :       43.2 % R.H.


9.2 Test set-up for conducted measurement
The antenna output of the EUT was connected to the spectrum analyzer. The resolution and video bandwidth is set to 100
kHz, and peak detection was used.



                             EUT                                                            Spectrum analyzer



9.3 Test set-up for radiated measurement
The radiated emissions measurements were performed on the 3 m semi anechoic chamber. The EUT was placed on
turntable approximately 1.5 m above the ground plane.

The frequency spectrum from 30 MHz to 26.5 GHz was scanned and maximum emission levels at each frequency recorded.
The system was rotated 360°, and the antenna was varied in the height between 1.0 m and 4.0 m in order to determine the
maximum emission levels. This procedure was performed for horizontal and vertical polarization of the receiving antenna.


9.4 Test equipment used
       Model Number              Manufacturer               Description                            Serial Number     Last Cal.
-     FSV40                     Rohde & Schwarz            Signal Analyzer                        101009            May 31, 2016 (1Y)
-     ESU                       Rohde & Schwarz            EMI Test Receiver                      100261            Apr. 06, 2016 (1Y)
-     310N                      Sonoma Instrument          Pre-Amplifier                          312544            Apr. 05, 2016 (1Y)
-     SCU-18                    Rohde & Schwarz            Pre-Amplifier                          102209            May 31, 2016 (1Y)
-     DT3000-3t                 Innco System               Turn Table                             DT3000/093        N/A
-     MA-4000XPET               Innco System               Antenna Master                         MA4000/509        N/A
-     VULB9163                  Schwarzbeck                TRILOG Broadband Antenna               9163-421          Apr. 15, 2016 (1Y)
-     BBHA9120D                 Schwarzbeck                Horn Antenna                           BBHA9120D295 Aug. 31, 2015 (2Y)
-     BBHA9170                  Schwarzbeck                Horn Antenna                           BBHA9170178       Aug. 31, 2015 (2Y)
All test equipment used is calibrated on a regular basis.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 17 of 31                            Report No. : W16DR-D001


9.5 Test data for conducted emission




                                                                  Low Channel




                                                                Middle Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 18 of 31                            Report No. : W16DR-D001




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 19 of 31                            Report No. : W16DR-D001




                                                                  Low Channel




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 20 of 31                            Report No. : W16DR-D001




                                                                Middle Channel




                                                                Middle Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 21 of 31                            Report No. : W16DR-D001




                                                                  High Channel




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 22 of 31                              Report No. : W16DR-D001


9.6 Test data for radiated emission

9.6.1 Radiated Emission which fall in the Restricted Band
-. Test Date                    : November 28, 2016 ~ November 29, 2016
-. Resolution bandwidth         : 1 MHz for Peak and Average Mode
-. Video bandwidth              : 1 MHz for Peak Mode, 10 Hz for Average Mode
-. Measurement distance         :3m
-. Duty Cycle                   : 100 %
-. Result                       : PASSED

Frequency        Reading         Detector        Ant. Pol.         Ant.         Cable          Amp       Total       Limits         Margin
   (MHz)         (dBμV)           Mode             (H/V)          Factor         Loss          Gain    (dBμV/m)    (dBμV/m)          (dB)

                                                          Test Data for Low Channel

                   54.95           Peak               H                                                  53.62        74.00         20.38
                   45.66         Average              H                                                  44.33        54.00          9.67
  2 381.64                                                         27.47        11.36          40.16
                   48.22           Peak               V                                                  46.89        74.00         27.11
                   39.15         Average              V                                                  37.82        54.00         16.18

                                                          Test Data for Low Channel

                   56.01           Peak               H                                                  54.68        74.00         19.32
                   46.84         Average              H                                                  45.51        54.00          8.49
  2 400.00                                                         27.47        11.36          40.16
                   48.31           Peak               V                                                  46.98        74.00         27.02
                   39.54         Average              V                                                  38.21        54.00         15.79

                                                          Test Data for High Channel

                   49.57           Peak               H                                                  48.24        74.00         25.76
                   40.66         Average              H                                                  39.33        54.00         14.67
  2 483.50                                                         27.47        11.36          40.16
                   47.61           Peak               V                                                  46.28        74.00         27.72
                   38.56         Average              V                                                  37.23        54.00         16.77

                                                   Tabulated test data for Restricted Band
Remark: “H”: Horizontal, “V”: Vertical
             Margin (dB) = Limits (dBμV/m) - Total Level (dBμV/m)
             Total Level = Reading + Antenna Factor + Cable Loss – Pre-Amplifier Gain




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer


It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 23 of 31                               Report No. : W16DR-D001


9.6.2 Spurious & Harmonic Radiated Emission
-. Test Date                    : November 28, 2016 ~ November 29, 2016
-. Resolution bandwidth         : 1 MHz for Peak and Average Mode for the emissions fall in restricted band,
                                 100 kHz for Peak Mode for the emissions outside restricted band
-. Video bandwidth              : 1 MHz for Peak Mode, 10 Hz for Average Mode
-. Frequency range              : 1 GHz ~ 26.5 GHz
-. Measurement distance         :3m
-. Duty Cycle                   : 100 %
-. Result                       : PASSED

 Frequency         Reading        Detector Ant. Pol.            Ant.       Cable       Amp          Total        Limits             Margin
    (MHz)           (dBμV)         Mode          (H/V)        Factor        Loss       Gain       (dBμV/m)      (dBμV/m)             (dB)
                                                         Test Data for Low Channel

                     47.51          Peak            H                                               53.71         74.00             20.29
                     38.66        Average           H                                               44.86         54.00              9.14
  4 804.00                                                     30.70       16.10       40.60
                     47.29          Peak            V                                               53.49         74.00             20.51
                     38.61        Average           V                                               44.81         54.00              9.19
                                                        Test Data for Middle Channel

                     46.99          Peak            H                                               53.59         74.00             20.41
                     38.26        Average           H                                               44.86         54.00              9.14
  4 880.00                                                     30.90       16.30       40.60
                     47.21          Peak            V                                               53.81         74.00             20.19
                     38.51        Average           V                                               45.11         54.00              8.89
                                                         Test Data for High Channel
                     47.26          Peak            H                                               54.16         74.00             19.84
                     38.71        Average           H                                               45.61         54.00              8.39
  4 960.00                                                    31.00       16.50       40.60
                     47.60          Peak            V                                               54.50         74.00             19.50
                     38.26        Average           V                                               45.16         54.00              8.84
                                                   Tabulated test data for Restricted Band
Remark: “H”: Horizontal, “V”: Vertical
            Margin (dB) = Limits (dBμV/m) - Total Level (dBμV/m)
            Total Level = Reading + Antenna Factor + Cable Loss – Pre-Amplifier Gain




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer



It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


oneETECH
        Ns                                                      Page 24 of 31                              Report No.: W16DR—D001


10. PEAK POWER SPECTRAL DENSITY

10.1 Operating environment
Temperature                          :      22.0 °C

Relative humidity                    :      43.2 % RH.

10.2 Test set—up
The antenna output of the EUT was connected to the spectrum analyzer.
The resolution bandwidth is set to 3 kHz = RBW ©100 kHz , the video bandwidth is set to 3 times the resolution

bandwidth.




                             EUT                                                           Spectrum analyzer



10.3 Test equipment used
       Model Number               Manufacturer                 Description                       Serial Number        Last Cal.

M —__FSV40                        Rohde & Schwarz              Signal Analyzer                   101009               May 31, 2016 (1Y)

All test equipment used is calibrated on a regular basis.




Tt should not be reproduced except in full, withoutthe written approval of ONETECH Corp.                                    EMC—003 (Rev.2)
ONETECH Corp.: 43—14, Jinsaegol—gil, Chowol—eup, Gwangiu—si, Gyeonggi—do, 12735, Korea (TEL: 82—31—799—9500, FAX: 82—31—799.9599)


                                                                Page 25 of 31                              Report No. : W16DR-D001


10.4 Test data
-. Test Date                    : November 28, 2016 ~ November 29, 2016
-. Test Result                  : Pass
-. Operating Condition          : Continuous transmitting mode
  CHANNEL             FREQUENCY(MHz)                   MEASURED VALUE (dBm)                          LIMIT (dBm)          MARGIN (dB)
      Low                     2 402.00                               -11.66                              8.00                  19.66
    Middle                    2 440.00                               -13.57                              8.00                  21.57
      High                    2 480.00                               -13.80                              8.00                  21.80
Remark. Margin = Limit – Measured value




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer




                                                                  Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 26 of 31                            Report No. : W16DR-D001




                                                                Middle Channel




                                                                  High Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 27 of 31                            Report No. : W16DR-D001


11. RADIATED EMISSION TEST

11.1 Operating environment
Temperature                          :       22.0 °C
Relative humidity                    :       43.2 % R.H.

11.2 Test set-up
The radiated emissions measurements were on the 3 m semi anechoic chamber. The EUT and other support equipment were
placed on a non-conductive turntable above the ground plane. The interconnecting cables from outside test site were
inserted into ferrite clamps at the point where the cables reach the turntable.
The frequency spectrum from 30 MHz to 26.5 GHz was scanned and emission levels maximized at each frequency recorded.
The system was rotated 360°, and the antenna was varied in height between 1.0 m and 4.0 m in order to determine the
maximum emission levels. This procedure was performed for both horizontal and vertical polarization of the receiving
antenna.


11.3 Test equipment used
       Model Number              Manufacturer               Description                        Serial Number         Last Cal.
-     FSV40                     Rohde & Schwarz            Signal Analyzer                    101009                May 31, 2016 (1Y)
-     ESU                       Rohde & Schwarz            EMI Test Receiver                  100261                Apr. 06, 2016 (1Y)
-     310N                      Sonoma Instrument          Pre-Amplifier                      312544                Apr. 05, 2016 (1Y)
-     SCU-18                    Rohde & Schwarz            Pre-Amplifier                      102209                May 31, 2016 (1Y)
-     DT3000-3t                 Innco System               Turn Table                         DT3000/093            N/A
-     MA-4000XPET               Innco System               Antenna Master                     MA4000/509            N/A
-     VULB9163                  Schwarzbeck                TRILOG Broadband Antenna           9163-421              Apr. 15, 2016 (1Y)
-     BBHA9120D                 Schwarzbeck                Horn Antenna                       BBHA9120D295 Aug. 31, 2015 (2Y)
-     BBHA9170                  Schwarzbeck                Horn Antenna                       BBHA9170178           Aug. 31, 2015 (2Y)
All test equipment used is calibrated on a regular basis.




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 28 of 31                            Report No. : W16DR-D001


11.4 Test data for Transmitting Mode

11.4.1 Test data for 30 MHz ~ 1 GHz
Humidity Level                    : 43.2 % R.H.                                                                      Temperature: 22.0 °C
Limits apply to                   : FCC CFR 47, PART 15, SUBPART C, SECTION 15.247
Result                            : PASSED

EUT                               : Body Composition Analyzer                                                   Date: November 28, 2016
Detector                          : CISPR Quasi-Peak (6 dB Bandwidth: 120 kHz)
Operating condition               : Low Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 29 of 31                            Report No. : W16DR-D001



Operating condition               : Middle Channel




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 30 of 31                              Report No. : W16DR-D001



Operating condition               : High Channel




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)


                                                                Page 31 of 31                              Report No. : W16DR-D001


11.4.2 Test data for Below 30 MHz
-. Test Date                    : November 28, 2016
-. Resolution bandwidth         : 200 Hz (from 9 kHz to 0.15 MHz), 9 kHz (from 0.15 MHz to 30 MHz)
-. Frequency range              : 9 kHz ~ 30 MHz
-. Measurement distance         :3m
-.Operating mode                : Transmitting mode

Frequency Reading Ant. Pol. Ant.                            Angle     Ant. Factor       Cable         Emission     Limits           Margin
  (MHz)   (dBμV)   (H/V) Height (m)                          (°)       (dB/m)           Loss       Level(dBμV/m) (dBμV/m)            (dB)



                                            It was not observed any emissions from the EUT.




11.4.3 Test data for above 1 GHz
-. Test Date                    : November 28, 2016
-. Resolution bandwidth         : 1 MHz for Peak and Average Mode
-. Video bandwidth              : 1 MHz for Peak Mode, 10 Hz for Average Mode
-. Frequency range              : 1 GHz ~ 26.5 GHz
-. Measurement distance         :3m
-.Operating mode                : Transmitting mode

Frequency Reading Ant. Pol. Ant.                            Angle     Ant. Factor       Cable         Emission     Limits           Margin
  (MHz)   (dBμV)   (H/V) Height (m)                          (°)       (dB/m)           Loss       Level(dBμV/m) (dBμV/m)            (dB)



                                            It was not observed any emissions from the EUT.




                                                                                            Tested by: Tae-Ho, Kim / Senior Engineer




It should not be reproduced except in full, without the written approval of ONETECH Corp.                                   EMC-003 (Rev.2)
ONETECH Corp.: 43-14, Jinsaegol-gil, Chowol-eup, Gwangju-si, Gyeonggi-do, 12735, Korea (TEL: 82-31-799-9500, FAX: 82-31-799-9599)



Document Created: 2016-12-16 19:06:52
Document Modified: 2016-12-16 19:06:52

© 2025 FCC.report
This site is not affiliated with or endorsed by the FCC